Model 2000A-EU
Thermal Conductivity Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING
SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
P/N M66182
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
07/22/05
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/ ECO # 05-0131
MANAGER.
i
Teledyne Analytical Instruments
Model 2000A-EU
All Rights Reserved. No part of this manual may be reproduced, transmitted, tran-
scribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free
from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an autho-
rized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumenta-
tion related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent danger,
it has no control over process conditions, and it can be misused. In particular, any alarm or
control systems installed must be tested and understood, both as to how they operate and
as to how they can be defeated. Any safeguards required such as locks, labels, or redun-
dancy, must be provided by the user or specifically requested of Teledyne at the time the
order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufactur-
er or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Thermal Conductivity Analyzer
The instrument for which this manual was supplied may incorporate one or more
options not supplied in the standard instrument. Commonly available options are listed
below, with check boxes. Any that are incorporated in the instrument for which this
manual is supplied are indicated by a check mark in the box.
iii
Model 2000A-EU
GROUND
Protective Earth
DANGER
COMBUSTIBLE GAS USAGE WARNING
iv
Thermal Conductivity Analyzer
Table of Contents
1 Introduction
1.1 Overview ........................................................................ 1-1
1.2 Typical Applications ....................................................... 1-2
1.3 Main Features of the Analyzer ....................................... 1-2
1.4 Model Designations ....................................................... 1-3
1.5 Front Panel (Operator Interface) ..................................... 1-3
1.6 Recognizing Difference Between LCD & VFD ............... 1-5
1.7 Rear Panel (Equipment Interface) .................................. 1-5
1.8 Gas Connections ........................................................... 1-7
2 Operational Theory
2.1 Introduction .................................................................... 2-1
2.2 Sensor Theory ............................................................... 2-1
2.2.1 Sensor Configuration ............................................... 2-1
2.2.2 Calibration ............................................................... 2-2
2.2.3 Effects of Flowrate and Gas Density ........................ 2-3
2.2.4 Measurement Results .............................................. 2-3
2.3 Electronics and Signal Processing ................................ 2-3
2.4 Temperature Control ...................................................... 2-5
3 Installation
3.1 Unpacking the Analyzer ................................................. 3-1
3.2 Mounting the Analyzer ................................................... 3-1
3.3 Electrical Connections (Rear Panel) .............................. 3-3
3.3.1 Primary Input Power ............................................... 3-4
3.3.2 Primary Input Power ............................................... 3-4
3.3.3 50-Pin Equipment Interface Connector .................. 3-4
3.3.3.1 Analog Outputs .............................................. 3-5
3.3.3.2 Alarm Relays ................................................. 3-6
3.3.3.3 Digital Remote Cal Inputs .............................. 3-7
3.3.3.4 Range ID Relays ........................................... 3-9
3.3.3.5 Network I/O .................................................... 3-9
3.3.3.6 Remote Valve Connector ............................... 3-9
3.3.4 RS-232 Port ........................................................... 3-10
3.4 Gas Connections ........................................................... 3-11
3.4.1 Sample System Design ......................................... 3-13
3.4.2 Pressure and Flow Rate Regulation ...................... 3-13
3.4.3 VENT Exhaust ....................................................... 3-14
3.4.4 SAMPLE Gas......................................................... 3-14
v
Model 2000A-EU
4 Operation
4.1 Introduction .................................................................... 4-1
4.2 Using the Data Entry and Function Buttons ................... 4-1
4.3 The System Function ..................................................... 4-4
4.3.1 Setting the Display ................................................. 4-5
4.3.2 Setting up an Auto-Cal ........................................... 4-5
4.3.3 Password Protection .............................................. 4-6
4.3.3.1 Entering the Password ................................... 4-7
4.3.3.2 Installing or Changing the Password ............. 4-7
4.3.4 Logging Out ........................................................... 4-9
4.3.5 System Self-Diagnostic Test .................................. 4-9
4.3.6 The Model Screen ................................................. 4-10
4.3.7 Checking Linearity with ALGORITHM ................... 4-10
4.4 The Zero and Span Functions ....................................... 4-11
4.4.1 Zero Cal ................................................................. 4-12
4.4.1.1 Auto Mode Zeroing ........................................ 4-12
4.4.1.2 Manual Mode Zeroing .................................... 4-13
4.4.1.3 Cell Failure .................................................... 4-14
4.4.2 Span Cal ................................................................ 4-14
4.4.2.1 Auto Mode Spanning ..................................... 4-15
4.4.2.2 Manual Mode Spanning ................................. 4-15
4.5 The Alarms Function ...................................................... 4-16
4.6 The Range Function ...................................................... 4-18
4.6.1 Manual (Select/Define Range) Screen .................. 4-19
4.6.2 Auto (Single Application) Screen ........................... 4-19
4.6.3 Precautions ............................................................ 4-21
4.7 The Analyze Function .................................................... 4-22
4.8 Programming ................................................................. 4-22
4.8.1 The Set Range Screen .......................................... 4.23
4.8.2 The Curve Algorithm Screen ................................. 4-25
4.8.2.1 Checking the Linearization ............................ 4-25
4.8.2.2 Manual Mode Linearization ........................... 4-26
4.8.2.3 Auto Mode Linearization ................................ 4-27
4.9 Special Function Setup .................................................. 4-28
4.9.1 Output Signal Reversal .......................................... 4.28
vi
Thermal Conductivity Analyzer
Maintenance
5.1 Routine Maintenance ..................................................... 5-1
5.2 System Self Diagnostic Test ........................................... 5-1
5.3 VFD Display .................................................................. 5-2
5.4 Fuse Replacement......................................................... 5-2
5.5 Major Internal Components ............................................ 5-3
5.6 Cell, Heater, and/or Thermistor Replacement ................ 5-5
5.6.1 Removing the Cell Compartment........................... 5-5
5.6.2 Removing and Replacing the Cell Block ............... 5-6
5.6.3 Removing the Heater and/or Thermocouple .......... 5-7
5.7 Cleaning ........................................................................ 5-4
5.8 Phone Numbers ............................................................. 5-5
Appendix
A-1 Specifications ................................................................ A-1
A-2 Recommended 2-Year Spare Parts List ......................... A-3
A-3 Drawing List ................................................................... A-4
A-4 19-Inch Relay Rack Panel Mount ................................... A-4
A-5 Calibration Procedure for TG Application........................... A-5
vii
Model 2000A-EU
viii
Thermal Conductivity Analyzer Introduction 1
Introduction
1.1 Overview
The Analytical Instruments Model 2000 Thermal Conductivity Ana-
lyzer is a versatile microprocessor-based instrument for measuring a com-
ponent gas in a background gas, or in a specific mixture of background
gases. 2000A-EU Analyzer complies with all of the requirements of the
Comonwealth of Europe (CE) for Radio Frequency Interference and Elec-
tromagnetic Interfaces (RFI/EMI) protection. It compares the thermal
conductivity of a sample stream with that of a reference gas of known
composition. The 2000 can—
• measure the concentration of one gas in a mixture of two gases.
• measure the concentration of a gas in a specific mixture of
background gases.
• measure the purity of a sample stream containing a single
impurity or a mixture of impurities.
The standard 2000 is preprogrammed with automatic linearization
algorithms for a large number of gases and gas mixtures. The factory can
add to this data base for custom applications, or the sophisticated user can
add his own unique application.
This manual covers the Model 2000A-EU General Purpose flush-
panel and rack-mount units only. These units are for indoor use in a
nonhazardous environment.
Many of the Model 2000 features covered in this manual are optional,
selected according to the customers specific application. Therefore, the user
will find much here that does not apply to his instrument. This is unavoid-
able due to the number of possible combinations of features available. We
have endeavored to make the manual as usable and convenient as possible,
in light of this flexibility.
Operational Theory
2.1 Introduction
The analyzer is composed of two subsystems:
1. Thermal Conductivity Sensor
2. Electronic Signal Processing, Display and Control.
The sensor is a thermal conductivity comparator that continuously
compares the thermal conductivity of the sample gas with that of a refer-
ence gas having a known conductivity.
The electronic signal processing, display and control subsystem
simplifies operation of the analyzer and accurately processes the sampled
data. A microprocessor controls all signal processing, input/output, and
display functions for the analyzer.
2.2.2 Calibration
Because analysis by thermal conductivity is not an absolute measure-
ment, calibration gases of known composition are required to fix the
upper and lower parameters (“zero” and “span”) of the range, or ranges, of
analysis. These gases must be used periodically, to check the accuracy of
the analyzer.
During calibration, the bridge circuit is balanced, with zero gas
against the reference gas, at one end of the measurement range; and it is
sensitized with span gas against the reference gas at the other end of the
measurement range. The resulting electrical signals are processed by the
analyzer electronics to produce a standard 0-1V, or an isolated 4–20 mA
dc, output signal, as described in the next section.
Installation
are four mounting holes—one in each corner of the rigid frame. Figure
3-1a contains the hole pattern dimensions. See the outline drawing, at the
back of this manual for overall dimensions.
On special order, a 19" rack-mounting panel can be provided. For
rack mounting, one or two 2000A series analyzers are flush-panel mounted
on the rack panel. See Figure 3-1b for dimensions of the mounting panel.
6.7"
8.75
5.75
Examples:
The analog output signal has a voltage which depends on gas concen-
tration relative to the full scale of the range. To relate the signal output to
the actual concentration, it is necessary to know what range the instrument
is currently on, especially when the analyzer is in the autoranging mode.
The signal output for concentration is linear over the currently se-
lected analysis range. For example, if the analyzer is set on a range that
was defined as 0–10 % hydrogen, then the output would be as shown in
Table 3-2.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will close quickly.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Probe connector (paragraph 3.3.3.6) provides
signals to operate the zero and span gas valves synchro-
nously. However, if you have the –C Internal valve option,
which includes zero and span gas inputs, the 2000A automati-
cally regulates the zero, span and sample gas flow.
Operation
4.1 Introduction
Although the Model 2000 is usually programmed to your application at the
factory, it can be further configured at the operator level, or even, cautiously,
reprogrammed. Depending on the specifics of the application, this might include
all or a subset of the following procedures:
• Setting system parameters:
• Establish a security password, if desired, requiring Operator
to log in.
• Establish and start an automatic calibration cycle, if desired.
• Routine Operation:
• Calibrate the instrument.
• Choose autoranging or select a fixed range of analysis.
• Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).
• Program/Reprogram the analyzer:
• Define new applications.
• Linearize your ranges.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.
The Enter button is used to accept any new entries on the VFD screen.
The Escape button is used to abort any new entries on the VFD screen that are
not yet accepted by use of the Enter button.
Figure 4-1 shows the hierarchy of functions available to the operator via the
function buttons. The six function buttons on the analyzer are:
• Analyze. This is the normal operating mode. The analyzer
monitors the thermal conductivity of the sample, displays the
percent or parts-per-million of target gas or contamination, and
warns of any alarm conditions.
• System. The system function consists of nine subfunctions.
Four of these are for ordinary setup and operation:
• Setup an Auto-Cal
• Assign Passwords
• Log out to secure system
• Initiate a Self-Test
Three of the subfunctions do auxiliary tasks:
• Checking model and software version
• Adjust LCD screen contrast Contrast Function is DISABLED
(Refer to Section 1.6)
• Display more subfunctions
Two of these are for programming/reprogramming the analyzer:
• Define gas applications and ranges (Refer to programming
section, or contact factory.)
• Use the Curve Algorithm to linearize output. (Refer to
programming section, or contact factory.)
• Zero. Used to set up a zero calibration.
• Span. Used to set up a span calibration.
• Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching,
and/or failsafe.
• Range. Used to set up three analysis ranges that can be
switched automatically with autoranging or used as individual
fixed ranges.
Any function can be selected at any time by pressing the appropriate button
(unless password restrictions apply). The order as presented in this manual is
appropriate for an initial setup.
Each of these functions is described in greater detail in the following proce-
dures. The VFD screen text that accompanies each operation is reproduced, at
System
MORE
Show Model
MODEL and Version
Auto/Manual Zero in
Zero Zero Select Progress
Analyze
Analyze Sample
If you cannot read anything on the display after first powering up:
1. Observe LED readout.
a. If LED meter reads 8.8.8.8.8., go to step 3.
b. If LED meter displays anything else, go to step 2.
2. Press button twice to turn Analyzer OFF and ON again.
LED meter should now read 8.8.8.8.8.. Go to step 3.
ZERO in Ød Øh off
SPAN in Ød Øh off
Press < > arrows to blink SPAN (or ZERO), then press Enter again. (You
won’t be able to set OFF to ON if a zero interval is entered.) A Span
Every ... (or Zero Every ...) screen appears.
Zero schedule: OFF
Day: Ød Hour: Øh
Use ∆ ∇ arrows to set an interval value, then use < > arrows to move to the
start-time value. Use ∆ ∇ arrows to set a start-time value.
To turn ON the SPAN and/or ZERO cycles (to activate AUTOCAL):
Press System again, choose AUTOCAL, and press Enter again. When the
ZERO/SPAN values screen appears, use the < > arrows to blink the SPAN
(or ZERO) and press Enter to go to the next screen. Use < > to select OFF/
ON field. Use ∆ ∇ arrows to set the OFF/ON field to ON. You can now turn
these fields ON because there is a nonzero span interval defined.
Enter the password using the < > arrow keys to move back and forth
between the existing password letters, and the ∆ ∇ arrow keys to change the
letters to the new password. The full set of 94 characters available for password
use are shown in the table below.
Characters Available for Password Definition:
A B C D E F G H I J
K L M N O P Q R S T
U V W X Y Z [ ¥ ] ^
_ ` a b c d e f g h
i j k l m n o p q r
s t u v w x y z { |
} → ! " # $ % & ' (
) * + ' - . / 0 1 2
3 4 5 6 7 8 9 : ; <
= > ? @
When you have finished typing the new password, press Enter. A verifica-
tion screen appears. The screen will prompt you to retype your password for
verification.
Enter PWD To Verify:
AAA
Use the arrow keys to retype your password and press Enter when
finished. Your password will be stored in the microprocessor and the system will
immediately switch to the Analyze screen, and you now have access to all
instrument functions.
If all alarms are defeated, the Analyze screen appears as:
Ø.Ø % H2 in N2
RØ1: Ø 1ØØ Anlz
If an alarm is tripped, the second line will change to show which alarm it is:
Ø.Ø % H2 in N2
AL1
NOTE:If you log off the system using the LOGOUT function in the
system menu, you will now be required to re-enter the pass-
word to gain access to Alarm, and Range functions.
Power: OK Analog: OK
Cell: 2 Preamp: 3
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting for number-code information. The results
screen alternates for a time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
The OUTPUT value is the output of the linearizer. It should be the ACTU-
AL concentration of the span gas being simulated.
If the OUTPUT value shown is not correct, the linearization must be
corrected. Press ESCAPE to return to the previous screen. Select and EN-
TER SET UP to Calibration Mode screen.
Select algorithm
mode : AUTO
There are two ways to linearize: AUTO and MANUAL: The auto mode
requires as many calibration gases as there will be correction points along the
curve. The user decides on the number of points, based on the precision re-
quired.
The manual mode only requires entering the values for each correction
point into the microprocessor via the front panel buttons. Again, the number of
points required is determined by the user.
Note: Shut off the gas pressure before connecting it to the analyzer,
and be sure to limit pressure to 40 psig or less when turning it
back on.
Readjust the gas pressure into the analyzer until the flowrate through the
sensor settles between 50 to 200 cc/min (approximately 0.1 to 0.4 scfh).
Then, and whenever Slope is less than 0.01 for at least 3 min, instead of
Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are
software steps in the zeroing process that the system must complete, AFTER
settling, before it can go back to Analyze. Software zero is indicated by S
Zero in the lower right corner.
###### % O2 N2
4 Left=##### SZero
The zeroing process will automatically conclude when the output is within
the acceptable range for a good zero. Then the analyzer automatically returns to
the Analyze mode.
Generally, you have a good zero when Slope is less than 0.05 ppm/s for
about 30 seconds.
Once zero settling completes, the information is stored in the analyzer’s
memory, and the instrument automatically returns to the Analyze mode.
arrow keys to toggle between AUTO and MAN span settling. Stop when
AUTO appears, blinking, on the display.
Select span
mode: AUTO
Press Enter to move to the next screen.
Span Val: 2Ø.ØØ %
<ENT> To begin span
Use the < > arrow keys to toggle between the span concentration value
and the units field (%/ppm). Use the ∆ ∇ arrow keys change the value and/or the
units, as necessary. When you have set the concentration of the span gas you are
using, press Enter to begin the Span calibration.
#### % O2 N2
Slope=##### Span
The beginning span value is shown in the upper left corner of the display.
As the span reading settles, the screen displays and updates information on
Slope. Spanning automatically ends when the span output corresponds, within
tolerance, to the value of the span gas concentration. Then the instrument auto-
matically returns to the analyze mode.
Once the span has begun, the microprocessor samples the output at a
predetermined rate. It calculates the difference between successive samplings
and displays this difference as Slope on the screen. It takes several seconds for
the first Slope value to display. Slope indicates rate of change of the Span
reading. It is a sensitive indicator of stability.
##### % O2 Air
Slope=#### Span
When the Span value displayed on the screen is sufficiently stable, press
Enter. (Generally, when the Span reading changes by 1 % or less of the range
being calibrated for a period of ten minutes it is sufficiently stable.) Once Enter
is pressed, the Span reading changes to the correct value. The instrument then
automatically enters the Analyze function.
Note: If all three ranges are currently defined for different applica-
tion gases, then the above screen does not display (because
mode must be manual). Instead, the LCD goes directly to the
following screen.
If above screen displays, use the ∆∇ arrow keys to Select MANUAL, and
press Enter.
Select range to run
> Ø1 Ø2 Ø3 CAL<
Use the < > keys to select the range: 00, 01, 02, or 03. (04 is for future
expansion.) Then press Enter.
Gas use: O2 N2
Range: Ø 16 %
Use the < > keys to toggle between the Range: low-end field and the
Range: high-end field. Use the ∆ ∇ keys to change the values of the fields.
Press Escape to return to the previous screen to select or define another
range.
Press Enter to return the to the Analyze function.
Note: If all three ranges are currently defined for different applica-
tion gases, then the above screen does not display (because
mode must be manual). Instead, the VFD goes directly to the
following screen.
If above screen displays, use the ∆∇ arrow keys to Select AUTO, and
press Enter.
Select auto ranging
Gas use: O2 N2
Use the ∆∇ arrow keys to change the application (gas use:).
Press Escape to return to the previous screen to select or define another
range.
Press Enter to return the to the Analyze function.
4.6.3 Precautions
The Model 2000 allows a great deal of flexibility in choosing ranges for
automatic range switching. However, there are some pitfalls that are to be
avoided.
right corner, the abbreviation Anlz indicates that the analyzer is in the Analyze
mode. If there is an * before the Anlz, it indicates that the range is linearized.
19.3 % O2 Air
R: ØØ:Ø 17 *Anlz
If the concentration detected is overrange, the first line of the display blinks
continuously.
4.8 Programming
Use the < > arrow keys again to move the blinking to APPLICATION and
press Enter.
Select rng to set appl:
> Ø1 Ø2 Ø3 CAL <
Use the ∆∇ arrow keys to increment/decrement the range number to 0, 1,
2, or 3, and press Enter.
Imp: Ø2 Bck: N2
FRØ TO1ØØ %
Use the < > arrow keys to move to Imp: (impurity), Bck: (background),
FR: (from—lower end of range), TO: (to—upper end of range), and PPM or %.
Use the ∆∇ arrow keys to increment the respective parameters as desired.
Press Enter to accept the values and return to Analyze mode. (See note
below.) Repeat for each range you want to set.
Note: The ranges must be increasing from low to high, for example,
if Range 1 is set to 0–10 % and Range 2 is set to 0–100 %, then
Range 3 cannot be set to 0–50 % since that makes Range 3
lower than Range 2.
Ranges, alarms, and spans are always set in either percent or ppm units, as
selected by the operator, even though all concentration-data outputs change
from ppm to percent when the concentration is above 9999 ppm.
Note: When performing analysis on a fixed range, if the concentra-
tion rises above the upper limit as established by the operator
for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output
continue to read regardless of the analog output range.
To end the session:
If started with the RS-232, send:
st<enter>
st<enter>
to the analyzer from the computer.
If started through the front panel, turn the instrument off and back on.
Press the Enter key twice to return to the Analyze mode.
Select algorithm
mode : AUTO
There are two ways to linearize: AUTO and MANUAL: The auto mode
requires as many calibration gases as there will be correction points along the
curve. The user decides on the number of points, based on the precision re-
quired.
The manual mode only requires entering the values for each correction
point into the microprocessor via the front panel buttons. Again, the number of
points required is determined by the user.
Note: Before performing section 4.8.2 or 4.8.2.3, you must check to
ensure that your calibration gases or points are between low
end and high end of the range setup. All correction points
must be between Zero and Span concentrations. Do not enter
Zero and Span points as part of the correction.
Repeat the above procedure for each of the data points you are setting (up
to nine points: 0-8). Set the points in unit increments. Do not skip numbers. The
linearizer will automatically adjust for the number of points entered.
When you are done, Press ESCAPE. The message, Completed. Wait
for calculation, appears briefly, and then the main System screen returns.
To end the session:
If started with the RS-232, send:
st<enter>
st<enter>
to the analyzer from the computer.
If started through the front panel, turn the instrument off and back on.
Press the Enter key twice to return to the Analyze mode.
st<enter>
st<enter>
to the analyzer from the computer.
If started through the front panel, turn the instrument off and back on.
Press the Enter key twice to return to the Analyze mode.
NOTE: If the unit has a range or ranges that specified >0 unit setup
for inverting.
The steps are:
1. Press RANGE key.
2. Use LEFT/RIGHT key to move to the range that is speci-
fied as inverting output.
3. Press and hold DOWN key for approximate 5 to 7 seconds.
NOTE: This setup will be identified only when performing GAS TEST
or calculation. The formula 1 will determine the range(s) is
required polarity coding.
If VFD negative with it proper span gas or the calculation is not satisfied,
set the S1 accordingly to the table below:
Close S1-5 range 1
Close S1-6 range 2
Close S1-7 range 3
Close S1-8 cal range
Press I/O to restart the system.
NOTE: This section apply during GAS TEST routine for the unit that
has more than one range install with nonlinear output applica-
tion.
The steps are as follows:
1. Set unit range to lowest range reading.
2. Using the computer generated settings for the controller,
adjust the controller settings for the maximum span gas
output.
3. Press SPAN key. Select AUTO mode and setup the setting
to span level. Press ENTER key to span.
4. Set range switch to next range.
5. Press SPAN key.
Maintenance
(Continued)
Preamp
0 OK
1 Zero too high
2 Amplifier output doesn't match test input
3 Both Failed
Cell
0 OK
1 Failed (open filament, short to ground, no
power.)
2 Unbalance (deterioration of filaments, blocked
tube)
CAUTION: The front and rear panels and all parts of the instru-
ment case must be tightly closed for CE mark com-
pliance to be in effect.
The 2000A contains the following major components:
• Analysis Section
Cell Compartment
Cell Block
• Power Supply
• Preamp and Motherboard with Microcontroller
• Display Board and Displays
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
• Rear Panel Board.
X X X
X X
X X
X X X
Figure 5-3: Rear Panel Retaining Screws
To detach the rear panel, remove only those screws marked with an X.
c. Turn the uncovered Cell Block assembly over so that the bottom
faces you. The black rectangular block with four screws is the
Heater Block. Separate the Heater Block and Insulator from the
Cell Block by removing the four screws. Leave the Heater Block
electrical connections connected.
d. Remove the four screws from each of the black plates that hold
the Cell. The Cell is sandwiched between the plates. You should
now be able to slide the Cell free.
e. Leave the electrical connections connected at the Cell. Unlace the
cabling, and unplug the grey Cell cable at the Preamplifier PCB
connector, J3. (See Figure 5-4, and/or drawings at the rear of this
manual.) The Preamplifier PCB can be more easily accessed by
removing the analyzer's rear panel. (See Section 5.5.)
f. Replace the cell by reversing the above procedure, steps a
through e.
CAUTION: The larger duct is for the Heater element, and the
smaller duct is for the Thermocouple.
5.7 Cleaning
If instrument is unmounted at time of cleaning, disconnect the instru-
ment from the power source. Close and latch the front-panel access door.
Clean outside surfaces with a soft cloth dampened slightly with plain clean
water. Do not use any harsh solvents such as paint thinner or benzene.
For mounted instruments, DO NOT wipe the front panel while the
instrument is controlling your process. Clean the front panel as prescribed in
the above paragraph.
Appendix
A-1 Specifications
System Power
Requirements: 110 VAC, 50-60Hz
Dimensions: 7.5”H x 10.8 “w X 13.7”D
Cell Material: Nickel plated brass block with nickel alloy
filaments and stainless steel plates
O/P Interface: Full duplex RS-232, implement a subset of
Tracs Command
Mounting:
Standard: General purpose flush panel mounting
Options: General purpose rack mounted to contain
either one or two in a 19” rack mountable
plate
Relative Humidity: Up to 99%
Altitude: 1,609 m
* Other configurations, including a totally
explosion-proof, are available
Note: Orders for replacement parts should include the part number (if
available) and the model and serial number of the instrument for
which the parts are intended.
Orders should be sent to:
TELEDYNE Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91749-1580
Phone (626) 961-9221, Fax (626) 961-2538
TWX (910) 584-1887 TDYANYL COID
or your local representative.
The following instructions show how to calibrate each range in the analyzer
independently. If all ranges must be calibrated using the Cal range, go to the cal
range calibration section and following the instructions.
CALIBRATE RANGE 1
1 Select the RANGE function on the main menu and place the analyzer in Range
1. DO THIS, EVEN IF ANALYZER IS IN RANGE 1.
2 Feed zero gas to analyzer and purge for the time needed to get readings that are
leveled. Feed CO2 to the analyzer
3 Press ZERO. You have the option to select AUTO or MANual. AUTO lets the
analyzer adjust itself and may take longer while in MANual mode the operator
has to adjust the zero reading on the display manually using the Up and Down
arrows.
4 Press ZERO select manual press the arrow > hold until OK is displayed in the
upper right corner of the VFD display.
5 If you choose manual, you first will do the Coarse zero adjustment using the Up
and Down arrows. Adjust until display reads as close to zero as you can then
press Enter. Now do the Fine zero adjustment using the Up and Down arrows
until the display reads as close to zero as possible then press Enter. Now the
analyzer will do the final step which is the Software zero adjustment. No input
is required and the analyzer will return automatically to the Analyze mode when
done.
6 When the analyzer finishes the zero, feed the span gas and purge for the time
needed to get readings that are leveled. Feed Air to the analyzer
7 Press the SPAN button, press the arrow > hold it until OK appears on the upper
right corner of the VFD display.
8 You have the option to select AUTO or MANual. AUTO lets the analyzer
adjust itself and may take longer while in MANual mode the operator has only
to press Enter when he thinks the reading is stable enough. Set the Span value
to 100.00%.
9 The analyzer should end the span by itself if AUTO was selected or when you
press the Enter button if MANual was selected.
CALIBRATE RANGE 2
1 Select the RANGE function on the main menu and place the analyzer in Range
2. DO THIS, EVEN IF ANALYZER IS IN RANGE 2.
2 Feed zero gas to analyzer and purge for the time needed to get readings that are
leveled. Feed CO2 to the analyzer
3 Press ZERO. You have the option to select AUTO or MANual. AUTO lets the
analyzer adjust itself and may take longer while in MANual mode the operator
has to adjust the zero reading on the display manually using the Up and Down
button.
4 Press ZERO select manual press the arrow > hold until OK is displayed in the
upper right corner of the VFD display.
5 If you choose manual, you first will do the Coarse zero adjustment using the Up
and Down arrows. Adjust until display reads as close to zero as you can then
press Enter. Now do the Fine zero adjustment using the Up and Down arrows
until the display reads as close to zero as possible then press Enter. Now the
analyzer will do the final step which is the Software zero adjustment. No input
is required and the analyzer will return automatically to the Analyze mode when
done.
6 When the analyzer finishes the zero, feed the span gas and purge for the time
needed to get readings that are leveled. Feed H2 to the analyzer.
7 Press the SPAN button, press the arrow > hold it until OK appears on the upper
right of the VFD display.
8 You have the option to select AUTO or MANual. AUTO lets the analyzer
adjust itself and may take longer while in MANual mode the operator has only
to press Enter when he thinks the reading is stable enough. Set the Span value
to 100.00%.
9 The analyzer should end the span by itself if AUTO was selected or when you
press the Enter button if MANual was selected.
CALIBRATE RANGE 3
1 Select the RANGE function on the main menu and place the analyzer in Range
3. DO THIS, EVEN IF ANALYZER IS IN RANGE 3.
2 Feed zero gas to analyzer and purge for the time needed to get readings that are
leveled. Feed H2 to the analyzer
3 Press ZERO. You have the option to select AUTO or MANual. AUTO lets the
analyzer adjust itself and may take longer while in MANual mode the operator
has to adjust the 100% reading on the display manually using the Up and Down
button.
4 Press ZERO select MANual, press arrow > hold until OK is displayed in the
upper right corner of the VFD display.
5 you choose manual, you first will do the Coarse zero adjustment using the Up
and Down arrows. Adjust until display reads as close to 100.00% as you can
then press Enter. Now do the Fine zero adjustment using the Up and Down
button until the display reads as close to 100.00% as possible then press Enter.
Now the analyzer will do the final step which is the Software zero adjustment.
No input is required and the analyzer will return automatically to the Analyze
mode when done.
6 When the analyzer finishes zero, feed span gas and purge for the time needed
to get readings that are leveled. Feed 90% H2 in air to the analyzer.
7 Press the SPAN button, press the arrow > hold until OK appears on the upper
right corner of the VFD display.
8 You have the option to select AUTO or MANual. AUTO lets the analyzer
adjust itself and may take longer while in MANual mode the operator has only
to press Enter when he thinks the reading is stable enough. Set the Span value
to 80.00%.
9 The analyzer should end the span by itself if AUTO was selected or when you
press the Enter switch if MANual was selected.
1 Select the RANGE function on the main menu and place the analyzer in the Cal
Range.
2 Feed zero gas to analyzer and purge for the time needed to get readings that are
leveled. Feed H2 to the analyzer
3 Now select the ZERO function from the main menu and press Enter.
If you want to calibrate only the Cal Range:
select MANual.
If you want to calibrate all ranges:
select AUTO.
4 If you choose manual, you first will do the Coarse zero adjustment using the Up
and Down arrows. Adjust until display reads as close to zero as you can then
press Enter. Now do the Fine zero adjustment using the Up and Down button
until the display reads as close to zero as possible then press Enter. Now the
analyzer will do the final step which is the Software zero adjustment.
Now the analyzer will do the final step which is the Software zero adjustment.
No input is required and the analyzer will return automatically to the Analyze
mode when done.
5 When the analyzer finishes the zero, feed the span gas and purge for the time
needed to get readings that are leveled. Feed 10% Air in H2 (90% H2 in air)
to the analyzer
6 Now select the SPAN function from the main menu and press Enter.
7 The analyzer should end the span by itself if AUTO was selected or when you
press the Enter if MANual was selected.
8 This range is now calibrated, and if it was AUTO calibrated, all ranges have
been calibrated.