You are on page 1of 2

Truth & Consequences

Coating Adhesion Testing in Accordance


with ASTM D4541ÑSticky Business
By Mark Schilling, Corrosion Probe, Inc.

O
n a recent coating categorize the test area (as com-
project, adhesion pared with the job requirement
testing was to be of one test per 1,000 square feet).
done in accordance with ASTM As with SSPC-PA 2, the aim is to
D4541. The specified require- statistically categorize an area,
ments were that a minimum of not to seek out and find a low
one test be done per 1,000 value.
square feet of coating, and one ASTM D4541-02 contains a
test per 100 square feet of coat- section that addresses precision
ing repairs. The coating specifica- and has a table that shows the
tion itself required a minimum expected variance for laboratory
adhesion of 1,000 psi. However, test results (intra-laboratory—
the coating specification was tai- done in the same lab, as well as
lored to a particular 100% vol- inter-laboratory—a comparison
ume solids polyurethane for which the adhesive vs. cohesive failure should be between different labs). The variability is
product data sheet advertised an adhe- noted, as well as the relative areas, and it presented in three different forms. The
sion value of 2,000 psi. Prior to com- sometimes is, but it’s that discrete quanti- first is the coefficient of variation, which
mencing work, it was decided that the tative value which people covet. is a decimal or percentage determined by
coating needed to test out at a minimum Although the test result is a single dividing the standard deviation of a sam-
of 1,900 psi. It seemed reasonable that numerical value, any such test is only an ple by the mean (average). The second is in
the coating adhesion should be at least indication of relative performance. There terms of degrees of freedom, which
95% of the advertised value (95% of is more to coating adhesion testing than relates to the t-distribution for the sample.
2,000 psi = 1,900 psi). Many tests were to using a tool to get a numerical result that (Think in terms of the familiar bell curve.)
2,000 psi or beyond without adhesion can be used as a pass/fail criterion. The These are terms that only a statistician
failure. Unfortunately for the contractor, ASTM D4541 test method was specified, could love. Thankfully, the third approach
many tests did not meet the minimum but it provides information that was com- is more straightforward and amenable to
requirement of 1,900 psi. A few failed pletely ignored on this project. coatings professionals—a maximum
tests of very small total area meant the The test reflects both material and acceptable percentage difference.
contractor had to reblast and reapply instrumental parameters. Different tools The table shows that depending on
coating to large areas. Boy, was he stuck! and different adhesives give different val- instrument type, the acceptable percent-
It’s imperative that a coating adhere ues. Differences in temperature, humidi- age difference between two values may
well to the substrate, and adhesion test- ty, and other factors complicate matters. be between about 25% and 58%. That
ing seems like a straightforward task. But Even the rigidity of the substrate affects provides a clear indication that the test is
coating adhesion is extremely complex the test results. For example, all else quite variable, but an example helps. For
and poorly understood. Simple adhesion equal, test values are generally higher on a type IV instrument used in the same
testing such as probing and attempting to 1
⁄4-inch-thick steel compared to 1⁄8-inch- laboratory, Table 1 of ASTM D4541-02
lift coating with a putty knife or pocket thick steel. gives the coefficient of variation as 8.5%,
knife is subjective. Pull-off tensile adhe- The principal problem is that the test there are 48 degrees of freedom, and the
sion testing is conceptually simple, and it method suffers from a lack of repro- maximum acceptable difference is 29%.
provides a single discrete quantity—the ducibility. Section 6.1.1 of ASTM D4541 The difference in percent relative to two
force required to cause fracture. The says that at least three replications are test results is determined by the simple
nature of the coating fracture in terms of usually required in order to statistically equation on page 19.
18 JPCL / April 2004 / PCE www.paintsquare.com
being used seems reasonable, if one an adhesion value of 1,000 psi. In fact, a
(X1 - X2) doesn’t know anything about coatings coating with a lower adhesion value can
X 100 as materials, the test instruments, and be the better performer. PSI is a conve-
(X1 + X2) / 2 the ASTM D4541 standard test nient and familiar unit of measure.
method. Coating adhesion is not at all Coatings may be applied by the square
Now consider two test values: 2,100 like the tensile strength of steel or the foot but they do not perform or fail by
psi and 1,600 psi. Plugging these into the compressive strength of concrete. A the square anything. Don’t get stuck
equation gives a result of 27%, which is coating with an adhesion value of 2,000 chasing magic numbers.
less than the 29% maximum acceptable psi is not twice as good as a coating with
difference given in the ASTM table. The
stated precision of the test method is so
poor that neither of these two values,
2,100 psi or 1,600 psi, can be declared to
be the one that is correct. With a very
small sampling of only a few measure-
ments, these two values should rightful-
ly be regarded as pretty much the same
result. That’s not common sense, but
that’s precision for you, or lack thereof.
Accuracy and precision are some-
times thought of as being more or less
the same thing when in fact, they are
very different. Accuracy is a question
of being on target. Precision is a ques-
tion of repeatability and reproducibility.
For example, when throwing darts and
aiming for the bull’s eye, if one throws
three darts with each landing one inch
from the center of the bull’s eye but
spaced at the corners of an equilateral
triangle, one has good accuracy, but pre-
cision is poor. The darts are all close to
target, but they are not close to each
other. In fact, they are on all sides of the
target. In contrast, if all three darts land
in the double 20 slot, accuracy is poor
but precision is excellent. For good
accuracy and good precision, one needs
to put all three darts close to the bull’s
eye, and close to each other.
The various types of tensile pull-off
testers can be calibrated for good accu-
racy. The shortcoming of the test
method, however, is that the precision is
not very good. It’s not very precise in the
laboratory. Don’t expect it to be better
under field conditions.
Specifying a minimum tensile pull-off
adhesion value of 1,900 psi on this pro-
ject to get within 5% of the advertised
adhesion of 2,000 psi for the coating

www.paintsquare.com JPCL / April 2004 / PCE 19

You might also like