You are on page 1of 11

dental

materials
Dental Materials 19 (2003) 188±198
www.elsevier.com/locate/dental

In vitro corrosion behavior of bioceramic, metallic, and bioceramic±


metallic coated stainless steel dental implants
M.H. Fathi a,*, M. Salehi a, A. Saatchi a, V. Mortazavi b, S.B. Moosavi b
a
Department of Materials Engineering, Isfahan University of Technology, Isfahan 84154, Iran
b
School of Dentistry, Isfahan University of Medical Sciences, Isfahan, Iran
Received 4 July 2001; revised 26 November 2001; accepted 18 December 2001

Abstract
Objectives: The most common metals and alloys used in dentistry may be exposed to a process of corrosion in vivo that make them
cytotoxic. The biocompatibility of dental alloys is primarily related to their corrosion behavior. The aim of this work was to evaluate the
corrosion behavior and thus the biocompatibility of the uncoated and coated stainless steels and compare the effect of type of coatings on
corrosion behavior.
Methods: Three types of coatings, hydroxyapatite (HA), titanium (Ti), and a double-layer HA/Ti on AISI 316L stainless steel were made.
HA coating was produced using plasma-spraying technique and Ti coating was made using physical vapor deposition process. In order to
perform a novel double-layer composite coating, a top layer of HA was plasma-sprayed over a physical vapor deposited Ti layer on AISI
316L stainless steel. Structural characterization techniques including XRD, SEM and EDX were used to investigate the microstructure,
morphology and crystallinity of the coatings. Electrochemical potentiodynamic tests were performed in physiological solutions in order to
determine and compare the corrosion behavior of the coated and uncoated specimens as an indication of biocompatibility.
Results: Double-layer HA/Ti coating on AISI 316L SS had a positive effect on improvement of corrosion behavior. The decrease in
corrosion current densities was signi®cant for these coated specimens and was much lower than the values obtained for uncoated and single
HA coated specimens. Ti coating on AISI 316L SS also has a bene®cial effect on corrosion behavior. The results were compared with the
results of corrosion behavior of HA coated commercially pure titanium (cpTi) and uncoated cpTi.
Signi®cance: These results demonstrated that the double-layer HA/Ti coated 316L SS can be used as an endodontic implant and two goals
including improvement of corrosion resistance and bone osteointegration can be obtained simultaneously.
q 2003 Academy of Dental Materials. Published by Elsevier Science Ltd. All rights reserved.
Keywords: AISI 316L stainless steel; Corrosion; HA coating; Ti coating

1. Introduction biocompatibility of dental alloys is primarily related to


their corrosion behavior [7]. The higher the corrosion of
Biocompatibility is the ability of a material to perform an alloy, the more its elements will be released and the
with an appropriate host response in a speci®c application risk of unwanted reactions in the oral tissues may be
[1]. This means that the tissue of the patient that comes into increased. These unwanted reactions include unpleasant
contact with the materials does not suffer from any toxic, tastes, irritation, allergy or other reaction [7].
irritating, in¯ammatory, allergic, mutagenic, or carcino- Surgical implants are usually made of metallic materials,
genetic action [2±5]. Since the oral environment is particu- such as austenitic stainless steel, cobalt±chromium alloys,
larly favorable for the biodegradation of metals due to its and titanium and its alloys. Among all the metallic materi-
ionic, thermal, microbiological, and enzymatic properties, it als, the austenitic stainless steels are the most popular
can be presumed that the patient is exposed to a certain materials because of their relatively low cost, ease of fabri-
extent to the products of the corrosion process [6]. The cation and reasonable corrosion resistance [8]. However, the
austenitic stainless steels are prone to localized attack in
long-term applications due to the aggressive biological
* Corresponding author. Tel.: 198-311-8912750; fax: 198-311-
8912752.
effects. The corrosion products include iron, chromium,
E-mail addresses: fathi@cc.iut.ac.ir, fathimoh@yahoo.com nickel and molybdenum, etc. Ions can accumulate in tissues
(M.H. Fathi). surrounding the implant or be transported to distant parts of
0109-5641/03/$30.00 + 0.00 q 2003 Academy of Dental Materials. Published by Elsevier Science Ltd. All rights reserved.
PII: S01 09- 5641( 02) 0002 9-5
M.H. Fathi et al. / Dental Materials 19 (2003) 188±198 189

Fig. 1. XRD patterns of the coatings: (a) plasma-sprayed HA coating on cpTi; (b) plasma-sprayed HA coating on 316L SS; (c) double-layer HA/Ti coating on
316L SS.

the body [9]. It has been demonstrated that metallic ions these ions in vivo not only causes toxic effects in mouse
resulting from the in vitro corrosion of austenitic stainless testicular seminiferous epithelium but also alterations in the
steels cause alteration of the expression of human lympho- spleen cellular population [11].
cyte-surface antigens and inhibit the immune response as A number of investigations have demonstrated that metal
assessed by lymphocyte proliferation [10]. The presence of ions can be released from metallic materials as the result of
190 M.H. Fathi et al. / Dental Materials 19 (2003) 188±198

Fig. 1. (continued)

corrosion [2,3,8±10,12±14]. Local adverse tissue reactions logical results of hydroxyapatite coated/uncoated metallic
or elicit allergy reactions caused by metallic implants implants in animals [36]. In this previous work, endodontic
originate from the release of metal ions. This release of implants consisting of plasma-sprayed HA coated and
ions depends upon the corrosion rate of the alloy and the uncoated substrates of various materials, namely 316L
solubility of the ®rst formed corrosion products [15]. Inten- stainless steel and Co±Cr±Mo alloy, were prepared. These
sive investigations have been carried out on biomaterials in implants were subsequently implanted in the mandibular
order to study their corrosion behavior, ion release and its canine of cats. After a healing period of 4 months, osteo-
effects [14,16±25], and toxicity and immunity of alloys integration evaluation and histopathological interpretation
[6,16±18,26±29]. The studies have been more intensive were carried out by using scanning electron microscopy
for stainless steels due to their corrosion behavior [14,19± (SEM). The results showed that different substrates had
29]. It has been suggested that metal ions associated with pronounced effects on the histopathological response to
316L stainless steel are toxic to osteogenic cells, affecting different HA coated implants [36].
their proliferation and differentiation [19]. Stainless steel The aim of the present work was to evaluate the corrosion
corrosion products above certain concentrations may disturb behavior and therefore biocompatibility of the uncoated and
the normal behavior of osteoblast-like bone marrow cell coated stainless steels. In order to design and produce
cultures [20±22,25]. Advanced techniques have been used a desired coating for improvement of histopathological
for determining toxicity and tissue reaction around metal response, tissue reaction and bone osteointegration around
implants [24,27,29]. For this purpose, in vivo animal tests dental implants, the effects of three types of coatings on the
have been performed [28]. behavior of substrates were compared.
Ion release rates in vitro can be used to determine the
proportionality of release in vivo [30±32] but long-term in
vivo studies have been made and results indicate that metal 2. Materials and methods
ion release increases with the exposure time [33,34].
Release of metal ion in vivo analysis did not demonstrate 2.1. Substrates and coatings
any consistent differences in the concentrations of metallic
elements next to stressed and nonstressed miniplates and Commercially pure titanium (cpTi) and AISI 316L stain-
screws of stainless steels [35]. less steel were used as substrates. The composition (wt%) of
A previous in vivo study by the present authors has shown stainless steel was C 0.03, Si 0.80, Mn 1.2, Cr 17.55, Ni
that the type of metallic implants had signi®cant effects on 13.65, Mo 3.1, P # 0.040, S # 0.030 and Fe as the balance.
the clinical success, bone tissue response and histopatho- The composition (wt%) of cpTi was Ti . 99.5%.
M.H. Fathi et al. / Dental Materials 19 (2003) 188±198 191

Fig. 2. SEM micrograph of the coatings: (a) plasma-sprayed HA coating surface showing partially melted with some porosity and microcracks; (b) the top
layer of a double-layer HA/Ti coating shows some porosity and microcracks; (c) a physical vapor deposited Ti coating on a 316L SS.

Ti coating (3±7 mm thickness) was made using physical (Philips XL 30) were used to study the microstructure and
vapor deposition process on AISI 316L stainless steel. Crys- morphology of the coatings. EDX analysis was utilized to
talline hydroxyapatite coatings with 40±60 mm were pro- estimate the ratio of Ca/P of the HA coating and the compo-
duced using plasma-spraying technique on two types of sition of the coatings.
substrate, i.e. AISI 316L stainless steel and commercially Commercially pure titanium (cpTi) and AISI 316L stain-
pure Ti. A double-layer composite coating composed of a less steel (316L SS) were selected as the sample materials.
hydroxyapatite top layer (40±60 mm thickness) and a Ti Coupons of two types of alloys were cut and surface
under layer (3±5 mm thickness) was prepared using a preparations were carried out by wet grinding with a series
plasma-spraying and a physical vapor deposition process, of SiC papers to 4000-grit followed by thorough cleaning in
respectively, on AISI 316L stainless steel. distilled water, degreasing with acetone and air drying.
An electrochemical corrosion polarization test cell was
2.2. Experimental procedure used for in vitro potentiodynamic corrosion tests in physio-
logical solutions. Graphite was used as the counter electrode
X-ray diffraction (XRD) technique (Philips X'Pert-MPD and saturated calomel electrode (SCE) as the reference elec-
System with a Cu ka wavelength of 1.5418 A Ê ) was used to trode. The physiological solutions were the Ringer's solu-
analyze the crystalline structure and phases present in the tion and physiological normal saline (0.9 wt% NaCl) and
coatings. This technique was also used to estimate the the potentiodynamic corrosion tests were performed at
percentage of the crystallinity of the HA coating. SEM 37 ^ 1 8C. Dynamic polarization curves were recorded at a
and energy dispersive X-ray analysis (EDX) techniques potential scanning rate of 0.5 mV s 21 initiated at 2250 mV
192 M.H. Fathi et al. / Dental Materials 19 (2003) 188±198

Fig. 3. SEM EDX of the coatings: (a) HA coating, desirable Ca/P is obtained; (b) and (c) HA layer and Ti layer, respectively, of a double-layer HA/Ti coating.

below the open circuit potential and the atmosphere was ti®ed by the Tafel equation. DE ˆ b log…i=I0 †: The expanded
open to air. Five replicate tests of each group of specimens form of this equation shows linearity where b is slope and
were performed. log i0 is the Y intercept. The scans are plotted as potential, E
To compare the corrosion behavior between the uncoated (mV, SCE), vs. log current density, log i (mA cm 21), and are
and HA coated cpTi, the uncoated 316L SS, HA coated often referred to as Evans Diagrams. The reverse scan was
316L SS and double-layer HA/Ti coated 316L SS, speci- obtained by reversing the direction of the scan, thus obtain-
mens were dynamically polarized in two different types ing a reverse Evans Diagram. Curve shape, potentials at
of physiological solutions. Potentiodynamic polarization current discontinuities and absolute values are all important
curves were determined at 37 ^ 1 8C using a EG and G in characterizing corrosion behavior. The second test
model 263A potentiostat/galvanostat interfaced with a involved potentiodynamic or continuous scanning but over
computer and a recorder. The tests were started after a a much smaller potential range than anodic polarization.
steady open-circuit potential was attained (not more than Typically, a scan is started 10 mV anodic to the corrosion
^5 mV drift in 5 min). The anodic and cathodic polariza- potential and terminates 10 mV cathodic to it. The scan is
tion curves were obtained for each specimen and corrosion plotted directly as E vs. i and is referred to as a linear
potentials and corrosion current densities were determined polarization plot. The slope of the linear segment of the
by Tafel extrapolation and linear polarization methods. The DE vs. Di plot is expressed by the formula:
mean value and standard deviations of the results were
…DE=Di†DE!0 ˆ RP ˆ B=icorr
calculated.
It is noticeable that two kinds of electrochemical tests where RP is polarization resistance (kV cm 2), (DE/Di) is
were utilized in this investigation. The ®rst test was referred slope (mV cm 2/mA) and
to as a potentiodynamic anodic polarization-scan because it
B…constant† ˆ b a bc =2:3…ba 1 bc †
is dynamic or continuous. It represented a pro®le of the
behavior of the alloy when scanned in a forward direction where ba and bc are Tafel slopes for the anodic and cathodic
from anodic to cathodic potential and spans a range of 1 V pro®les, respectively. One way to ®nd the value of the equa-
or more. The straight line segments of these scans are iden- tion involves the determination of Tafel bc and ba from
M.H. Fathi et al. / Dental Materials 19 (2003) 188±198 193

Fig. 3. (continued)

potentiodynamic polarization curves and calculation of the porous surface and some microcracks. Fig. 2(c) is a SEM
resulting value of B. micrograph showing a physical vapor deposited Ti coating
on a 316L SS. The porosity of the coating is less than HA
coating but microcraks cannot be seen.
3. Results The result of EDX analysis of HA coating surface is
shown in Fig. 3(a). The Ca/P ˆ 2 is the desired ratio. Fig.
3.1. Structural characterization 3(b) and (c) show EDX analysis of HA layer and Ti layer of
double-layer HA/Ti coating.
The XRD patterns of the plasma-sprayed HA coatings on
cpTi and 316L SS are shown in Fig. 1(a) and (b), respec- 3.2. Corrosion behavior
tively. In the plot of intensity vs. 2u , there are numerous
sharp peaks and a low background indicative of highly crys- The potentiodynamic polarization curves of the uncoated,
talline HA coating. Fig. 1(c) also shows the XRD pattern of HA coated, Ti coated, double-layer HA/Ti coated 316L SS
a double-layer HA/Ti coating on 316L SS which is very and HA coated/uncoated cpTi in the normal saline solution
similar to the XRD pattern of HA coating on 316L SS. are shown in Fig. 4. The similar curves that are obtained in
The SEM micrograph showing a typical plasma-sprayed Ringer's solution are plotted in Fig. 5. These curves are
HA coating on a metallic substrate is shown in Fig. 2(a). The selected because their extracted data were the most nearest
structure appears to be partially melted and porous and to the mean values of the current densities of each group of
microcracks can be observed scattered throughout the struc- specimens.
ture, probably as a result of a very rapid cooling rate during The corrosion current densities of various specimens
the deposition process. Fig. 2(b) is a SEM micrograph show- were determined from the potentiodynamic polarization
ing a typical double-layer HA/Ti coating on 316L SS curves (Figs. 4 and 5) by Tafel extrapolation method.
substrate which also shows that the coating consists of a These results align with corrosion potentials, summarized
194 M.H. Fathi et al. / Dental Materials 19 (2003) 188±198

Fig. 3. (continued)

in Tables 1 and 2 in normal saline solution and Ringer's resistance of metallic substrate, i.e. 316L SS and cpTi and
solution, respectively. The standard deviations of the corro- decreased the corrosion current density of each type of
sion current densities are shown in parentheses in Tables 1 coated metallic substrate. The polarization curve, (b) in
and 2. The corrosion current densities as determined by Fig. 4, of the HA coated 316L SS was shifted to the left
linear polarization method are also shown in Tables 1 and 2. upper area, compared with curve (a), which is the uncoated
According to Table 1 and Fig. 4, the uncoated 316L SS 316L SS specimen. It means that the HA coated 316L SS
possesses lower corrosion resistance and thus higher corro- (Ecorr ˆ 2105 mV, icorr ˆ 68 nA) was more corrosion resis-
sion current density (icorr ˆ 265 nA cm 22) in the normal tant than the uncoated 316L SS (Ecorr ˆ 2174 mV,
saline solution. The similar trend can be observed in the icorr ˆ 265 nA). The corrosion behavior of the HA coated
Ringer's solution (Table 2). The HA coating decreased the cpTi (Ecorr ˆ 2256 mV, icorr ˆ 43 nA) was better than
corrosion current density of the HA coated 316L SS uncoated cpTi (Ecorr ˆ 2334 mV, icorr ˆ 72 nA) up to
(icorr ˆ 68 nA cm 22) and HA coated cpTi (icorr ˆ 43 200 mV, as manifested by a shift of the polarization curve
nA cm 22). The double-layer HA/Ti coating had a sig- (f) to the left upper of the uncoated cpTi curve (e). A similar
ni®cant effect on the corrosion behavior of 316L SS so trend can be observed in the Ringer's solution (Fig. 5,
that the corrosion current density of HA/Ti coated 316L curves (a), (b), (e) and (f), Table 2). Therefore, single HA
SS decreased (icorr ˆ 44 nA cm 22) and was exactly similar coating on metallic substrate caused signi®cant changes on
to single HA coated cpTi (icorr ˆ 43 nA cm 22) in normal corrosion behavior of substrate. This result was in agree-
saline solution. The Ti coating could decrease the corrosion ment with the results of the other researchers [37].
current density of the Ti coated 316L SS in both types of the Single HA coating could decrease the corrosion current
physiological solutions (Table 1 and 2). density of 316L stainless steel substrate in physiological
solutions (Tables 1 and 2 and Figs. 4 and 5) but this effect
was not adequately complete. This was due to the fact that
4. Discussion the structure and surface morphology of single HA coating
consists of a porous surface and some microcracks (Fig. 2)
The single HA coating had a positive effect on corrosion and could not act as a barrier to the release of metal ions
M.H. Fathi et al. / Dental Materials 19 (2003) 188±198 195

Fig. 5. Potentiodynamic polarization curves of: (a) the uncoated; (b) HA


Fig. 4. Potentiodynamic polarization curves of: (a) the uncoated; (b) HA coated; (c) Ti coated; (d) double-layer HA/Ti coated 316L SS; (e) the HA
coated; (c) Ti coated; (d) double-layer HA/Ti coated 316L SS; (e) the HA
coated cpTi; (f) uncoated cpTi in the Ringer's solution at 37 8C.
coated cpTi; (f) uncoated cpTi in the normal saline solution (0.9 wt% NaCl)
at 37 8C.
pathological responses to uncoated 316L SS and uncoated
from metallic implant to physiological solutions and vitallium were not acceptable [36]. These results are in
corrosive environment. The HA coating acted simply as a agreement with the studies that have been carried out on
semi-mechanical barrier. The corrosion rate in this case was endodontic implants [16,17]. Vitallium was originally
drastically reduced, although in the microcracks and pores used but its biocompatibility was questioned when it was
of the HA coating the substrate came into contact with the shown to undergo surface corrosion. This is an important
solution and some corrosion occurred. factor in the success of an endodontic implant. Stainless
Earlier work by the present authors showed that the histo- steel and vitallium have generally lost acceptance and are

Table 1 Table 2
Mean values (standard deviation) of corrosion current densities and corro- Mean values (standard deviation) of corrosion current densities and corro-
sion potentials in normal saline solution (0.9 wt% NaCl) at 37 8C sion potentials in Ringer's solutions at 37 8C

Material Ecorr (mV) Icorr (nA/cm 2) Material Ecorr (mV) Icorr (nA/cm 2)

Tafel Linear Tafel Linear

Uncoated 316L SS 2174(20) 265 (16) 194 (16) Uncoated 316L SS 2195(20) 218 (17) 181 (20)
Uncoated cpTi 2334(11) 72 (7) 63 (9) Uncoated cpTi 2303(5) 71 (6) 61 (6)
HA coated 316L SS 2105(10) 68 (2) 52 (1) HA coated 316L SS 2110(21) 57 (3) 49 (1)
HA coated cpTi 2256(10) 43 (6) 34 (7) HA coated cpTi 2215(14) 40 (4) 29 (3)
HA/Ti coated 316L SS 2153(15) 44 (5) 32 (3) HA/Ti coated 316L SS 2182(20) 43 (5) 32 (4)
Ti coated 316L SS 288(20) 27 (5) 26 (5) Ti coated 316L SS 261(15) 24 (5) 26 (2)
196 M.H. Fathi et al. / Dental Materials 19 (2003) 188±198

Many attempts have been made for improvement and


optimization of implant surface conditions by creating a
bioceramic coating on implant surface [39±43]. In this
case, double- or multi-layer coatings have been recently
considered by a number of researchers' works [37,44±46]
for improvement of biomedical applications and offering
potential clinical bene®ts in orthopedic and dental surgery.
Bioceramic coatings such as Al2O3, TiO2 and ZrO2 on 316L
SS have already been considered to increase corrosion resis-
tance [43,47±49].
The in vitro electrochemical polarization corrosion test
results indicated that novel double-layer HA/Ti coating
have a bene®cial and desired effect on corrosion behavior
of 316L SS. This can improve the corrosion resistance (Figs.
4 and 5 and Tables 1 and 2) and subsequently can decrease
the release of metallic ions and deleterious effects. There-
fore, two goals could be obtained simultaneously by the
formation of the double-layer HA/Ti coating. First, the
improvement of 316L SS corrosion resistance and decreas-
ing of ion release and consequently prevention of tissue
damages, in¯ammation and irritation and obtain desirable
histopathological response. The second improvement is in
terms of bone osteointegration and bone bonding with the
coated implant.
The Ti coating had a positive effect on corrosion resis-
tance of 316L SS. The polarization curve, (c) in Fig. 4, of
the Ti coated 316L SS was shifted to the left upper
area, compared with curve (a), which is the uncoated
316L SS specimen. The Ti coated 316L SS (Ecorr ˆ 288
mV, icorr ˆ 30 nA) was more corrosion resistant than the
Fig. 6. Periapical tissue response to HA coated 316L SS: (a) in¯ammatory
uncoated 316L SS. The type of physiological solution had
connective tissue can be observed ( £ 25); (b) the same picture ( £ 100).
no effect on this trend.
It is noticeable that the passive current density (ip) of the
no longer available for clinical use as endodontic implants Ti coated 316L SS (curve (c) in Figs. 4 and 5) was similar to
[16,17]. HA/Ti coated 316L SS (curve (d) in Figs. 4 and 5). This
On the other hand, earlier work showed that the type of means that the Ti responded primarily to environment and
metallic substrates which were HA plasma-sprayed coated corrosion reaction was essentially between Ti and solution.
and used for endodontic implants in animals had signi®cant
effects on histopathological response [36]. Statistical analyses
of the signi®cance of the differences between histopathologi-
5. Conclusion
cal response to single HA coated vitallium and single HA
coated 316L SS were carried out by Kruskal±Wallis method.
Double-layer HA/Ti coating can have a bene®cial and
More tissue damage, in¯ammation and irritation have been
desired effect on corrosion behavior of 316L SS and
observed around single HA coated 316L SS (Fig. 6) in
decrease the corrosion current density that is a distinct
comparison with single HA coated vitallium [36]. This was
advantage for prevention of ion release. Corrosion current
owing to the inability of the single HA coating to act as a
density of double-layer HA/Ti coated 316L SS was similar
barrier to the release of metal ions to animal's body on the
and equal to single HA coated cpTi and it suggests that
316L SS implants. The bone osteointegration around single
double-layer HA/Ti coated 316L SS can be used as endo-
HA coated vitallium alloy and single HA coated 316L SS
dontic implant.
were almost the same and from the point of the clinical
success and osteointegration no signi®cant difference were
observed [36]. This result is also in agreement with the
other researchers [38]. It has been observed that there are no Acknowledgements
differences between osteointegration when stainless steel is
used instead of Ti alloy as metallic substrate, con®rming The authors are grateful for the support of this research by
that bone responds primarily to coating [38]. Isfahan University of Technology.
M.H. Fathi et al. / Dental Materials 19 (2003) 188±198 197

References [24] Milosev I, Strehblow HH. The behavior of stainless steels in physio-
logical solution containing complexing agent studied by X-ray photo-
[1] Williams DF. De®nitions in biomaterials. Proceeding of a Consensus electron spectroscopy. J Biomed Mater Res 2000;52:404±12.
Conference of the European Society for Biomaterials, England. Vol. [25] Fernandes MH, Costa MA. Proliferation/differentiation of osteo-
4, New York: Elsevier; 1986. blastic human alveolar bone cell cultures in the presence of stainless
[2] Vahey JW, Simonian PT, Conrad EU. Carcinogenicity and metallic steel corrosion products. J Mater Sci: Mater Med 2000;11(3):141±53.
implants. Am J Orthod Dentofacial Orthop 1995;24:319±24. [26] Grimsdottir MR, Hensten-Pettersen A, Kulmann A. Cytotoxic effect
[3] Arvideon K, Cottler-Fox M, Friberg V. Cytotoxic effects of Co±Cr of orthodontic appliances. Eur J Orthodon 1992;14:47±53.
alloys on ®broblast derived form human gingiva. Scand J Dent Res [27] Shettlemore MG, Bundy KJ. Toxicity measurement of orthopedic
1986;95:356±63. implant alloy degradation products using a bioluminescent bacterial
[4] Jacobsen N, Hensten-Pettersen A. Occupational heath problems and assay. J Biomed Mater Res 1999;45(4):395±403.
adverse patient reactions in orthodontics. Eur J Orthodon 1989;11: [28] Pereira MC, Pereira ML, Sousa JP. Individual study of chromium in
254±64. the stainless steel implants degradation: an experimental study in
[5] Lin Sun Z, Wataha JC, Hanks CT. Effects of metal ions on osteoblast- mice. BioMetals 1999;12:275±80.
like cell metabolism and differentiation. J Biomed Mater Res 1997; [29] Uo M, Watari F, Yokoyama A, Matsuno H, Kawasaki T. Tissue
34:29±37. reaction around metal implants observed by X-ray scanning analytical
[6] Locci P, Lilli C, Marinucci L, Calvitti M, Belcastro S, Bellocchio S, microscopy. Biomaterials 2001;22(7):677±85.
Staffolani N, Guerra M, Becchetti E. In vitro cytotoxic effects of [30] Brown SA, Farnsworth LJ, Merritt K, Crowe TD. In vitro and in vivo
orthodontic appliances. J Biomed Mater Res (Appl Biomater) 2000; metal ion release. J Biomed Mater Res 1988;22(4):321±38.
53:560±7. [31] Gilbert HL, Smith SM, Lautenschlager EP. Scanning electrochemical
[7] Craig RG, Powers JM, Wataha JC. Dental materials, properties and microscopy of metallic biomaterials: reaction rate and ion release
manipulation. 7th ed. St Louis: Mosby-Year book Inc, 2000 p. 231±3. imaging modes. J Biomed Mater Res 1993;27(11):1357±66.
[8] Jacobs JJ, Gilbert JL, Urban RM. Corrosion of metal orthopaedic [32] Marek M. Measurement of metal ion release from biomedical implant
implants. J Bone Joint Surg 1998;80A:268±82. alloys. The Proceeding of Compatibility of Biomedical Implants, San
[9] Lei MK, Zhu XM. Corrosion resistance of plasma source ion nitrided Francisco, California, USA; 1994. p. 73±84.
austenitic stainless steels. Biomaterials 2001;22:641±7. [33] Jones LC, Hungerford DS, Kenna RV, Braem G, Grant V. Urinary
[10] Tracana RB, Sousa JP, Carvalho GS. Mouse in¯ammatory response to excretion levels of metal ions in patients undergoing total hip replace-
stainless steel corrosion products. J Mater Sci: Mater Med 1994; ment with a porous-coated prosthesis: preliminary results. The
5:596±600. Proceeding of Quantitative Characterization and Performance of
[11] Tracana RB, Pereira ML, Abreu AM, Sousa JP, Carvalho GS. Stain- Porous Implants for Hard Tissue Applications, Nashville, Tennessee,
less steel corrosion products cause alternations on mouse spleen cellu- USA, 1985. ASTM publications; 1987. p. 151±62.
lar. J Mater Sci: Mater Med 1995;6:56±61. [34] Traisnel M, Maguer D, Hildebrand HF, Lost A. Corrosion of surgical
[12] Dobbs HS, Minski MJ. Metal ion release after total hip replacement. implants. Clin Mater 1990;5(2-4):309±18.
Biomaterials 1980;1:193±8. [35] Matthew IR, Frame JW. Release of metal in vivo from stressed and
[13] Barrett RD, Bishara SE, Quinn JK. Biodegradation of orthodontic nonstressed maxillofacial fracture plates and screws. Oral Surg Oral
appliance. Part I. Biodegradation of nickel and chromium in vitro. Med Oral Pathol Oral Radiol Endod 2000;90(1):33±8.
Am J Orthod Dentofacial Orthop 1993;103:243±50. [36] Fathi MH, Feizi GH, Moosavi SB, Gahanshahi GH, Salehi M, Saatchi
[14] Staffolani N, Damiani F, Lilli C. Ion release from orthodontic appli- A, Mortazavi V. The role of metallic substrate of hydroxyapatite
ances. J Dent 1999;27(6):449±54. coated dental implants in clinical and pathological success, Esteghlal.
[15] Long M, Rack HJ. Titanium alloys in total joint replacementÐa IUT J Engng 2001;20(1):189±205.
materials science perspective. Biomaterials 1998;19:1621±39. [37] Nie X, Leyland A, Matthews A. Deposition of layered bioceramic
[16] Larsen RM, Patten JR, Wayman BE. Endodontic endosseous hydroxyapatite/TiO2 coatings on titanium alloys using a hybrid tech-
implants: case reports and update of materials. J Endod 1989; nique of micro-arc oxidation and electrophoresis. Surf Coat Technol
15(10):496±500. 2000;125:407±14.
[17] Parreira FR, Bramwell JD, Roahen JO, Giambarresi L. Histological [38] Leali Tranquilli P, Merolli A, Palmacci O. Evaluation of different
response to titanium endodontic endosseous implants in dogs. J Endod preparations of plasma-spray hydroxyapatite coating on titanium
1996;22(4):161±4. alloy and duplex stainless steel in the rabbit. J Mater Sci: Mater
[18] Locci P, Marinucci L, Lilli C, Belcastro S, Staffolani N, Bellocchio S, Med 1994;5(6±7):345±9.
Damiani F, Becchetti E. Biocompatibility of alloys used in orthodontics [39] Gross KA, Berndt CC, Goldschlag DD, Jacono VJ. In vitro changes of
evaluated by cell culture tests. J Biomed Mater Res 2000;51:561±8. hydroxyapatite coatings. Int J Oral MaxilloFac Implant 1997;12(5):
[19] Morais S, Sousa JP, Fernandes MH, Carvalho GS, deBruijn JD, 589±97.
vanBlitterswijk CA. Decreased consumption of Ca and P during in [40] Suchanek W, Yoshimura M. processing and properties of hydroxapa-
vitro biomineralization and biologically induced deposition of Ni and tite-based biomaterials for use as hard tissue replacement implants.
Cr in presence of stainless steel corrosion products. J Biomed Mater J Mater Res 1998;13(1):94±117.
Res 1998;42(2):199±212. [41] Hench LL. Bioceramics. J Am Ceram Soc 1998;81(7):1705±28.
[20] Morais S, Sousa JP, Fernandes MH, Carvalho GS, deBruijn JD, [42] Arciola CR, Mantanaro L, Moroni A, Giordano M, Pizzoferato A.
vanBlitterswijk CA. Effects of AISI 316L corrosion products in in Hydroxyapatite-coated orthopaedic screws as infection resistant
vitro bone formation. Biomaterials 1998;19(11±12):999±1007. materials: in vitro study. Biomaterials 1999;20(4):233±327.
[21] Fernandes MH. Effect of stainless steel corrosion products on in vitro [43] Hsu HC, Yen SK. Evaluation of metal ion release and corrosion
biomineralization. J Biomater Appl 1999;14(2):113±68. resistance of ZrO2 thin coating on the dental Co±Cr alloys. Dent
[22] Morais S, Dias N, Sousa JP, Fernandes MH, Carvalho GS. In vitro Mater 1998;14:339±46.
osteoblastic differentiation of human bone marrow cells in the [44] Silva PL, Santos JD, Monteiro FJ, Knowles JC. Adhesion and micro-
presence of metal ions. J Biomed Mater Res 1999;44(2):176±90. structural characterization of plasma-sprayed hydroxyapatite/glass
[23] Kim H, Johnson JW. Corrosion of stainless steel, nickel±titanium, ceramic coatings onto Ti±6A1±4V substrates. Surf Coat Technol
coated nickel±titanium, and titanium orthodontic wires. Angle 1998;102:191±6.
Orthod 1999;69(1):39±44. [45] Ferraz MP, Monteiro FJ, Santos JD. CaO±P2O5 glass hydroxyapatite
198 M.H. Fathi et al. / Dental Materials 19 (2003) 188±198

double-layer plasma-sprayed coating: in vitro bioactivity evaluation. J Al2O3 Plasma spraying for biomedical applications. Surf Engng
Biomed Mater Res 1999;45(4):376±83. 1990;6(3):209±12.
[46] Ferraz MP, Knowles JC, Olsen I, Monterio FJ, Santos JD. Flow cyto- [48] Gluszek J, Masalski J, Furman P. TiO2 ®lm as a protective coating on
metry analysis of effects of glass on response of osteosarcoma cells to stainless steel. The Proceeding of Fifth Polish Conference Corro-
plasma-sprayed hydroxyapatite/CaO±P2O5 coatings. J Biomed Mater sion'96ÐTheory and Practice, Gdansk, Poland, Vol. 3; 1996. p. 789±93.
Res 1999;47(4):603±11. [49] Yen SK. Characterization of electrolytic ZrO2 coating on AISI 316L
[47] Santose JDS, Monteiro FJ. Wear behavior of stainless steel after stainless steel. J Electroch Soc 1999;146(4):1392±6.

You might also like