You are on page 1of 7

doi:10.1111/iej.

12302

Deformation and fracture incidence of Reciproc


instruments: a clinical evaluation

G. Plotino1, N. M. Grande1 & P. F. Porciani2


1
Private Practice, Grande, Plotino & Torsello – Studio di Odontoiatria, Rome; and 2Tuscan School of Dental Medicine,
University of Siena, Siena, Italy

Abstract chi-square test with a level of significance set at


P < .05.
Plotino G, Grande NM, Porciani PF. Deformation and
Results A total of 8 Reciproc R25 instruments frac-
fracture incidence of Reciproc instruments: a clinical
tured during treatment, which represents 0.47% of
evaluation. International Endodontic Journal.
the number of instruments used and 0.21% of the
Aim To investigate the incidence of fracture and root canals treated. Five instruments fractured in
deformation observed on Reciproc instruments after primary treatment cases and three instruments dur-
clinical use. ing retreatments. Six Reciproc R25 deformed during
Methodology A total of 1696 Reciproc NiTi clinical use (0.35% of the number of instruments
instruments that were discarded after normal clinical used and 0.15% of the root canals treated), four dur-
usage over 30 months were collected. Of these 1580 ing retreatments and two during primary treatments.
instruments were R25 (93%), 76 were R40 (5%), and The deformation reported was unwinding of the
40 were R50 (2%). The treatments included 3780 blades, one at 1 mm from the tip and five between 2
root canal preparations, 3023 primary treatments and 3 mm from the tip. No deformations or fracture
and 757 retreatments. Root canal preparation was were registered for the Reciproc R40 and R50 instru-
standardized and followed the manufacturer’s direc- ments.
tions for the use of Reciproc files without creating Conclusions Root canal instrumentation following
prior glide path with hand instruments. In retreat- the manufacturer’s instructions can be performed
ment cases, Reciproc R25 instrument was used to with Reciproc instruments with a very low incidence
remove all filling material and to complete the prepa- of instrument fracture and deformation.
ration. Each instrument was used for shaping only
Keywords: deformation, fracture, nickel-titanium,
one tooth. All the instruments used were inspected
reciprocation.
under a microscope to detect signs of fracture or
deformation and, if any fracture existed, the length of Received 20 February 2014; accepted 20 April 2014
the files was measured. Data were analysed using a

that the angle of reciprocation was the same in both


Introduction
clockwise and counterclockwise directions. The recip-
Reciprocating movement has been used for stainless- rocation actually used with NiTi instruments is an
steel instruments for many years (Frank 1967, Klay- alternating movement in which the file rotates a cer-
man & Brilliant 1975). This original movement could tain angle in the cutting direction and shortly after in
be defined as ‘complete oscillating reciprocation’ in the opposite direction over a much smaller angle
(Kim et al. 2012). Consequently, these instruments
complete a full rotation over several cycles of recipro-
Correspondence: Gianluca Plotino, Via Calabria 25, 00187 Rome,
cating movements, and therefore, this type of recipro-
Italy (Tel.: +390642020595; e-mail: endo@gianlucaplotino. cation may be defined as ‘Partial or asymmetrical
com). reciprocation with a rotary effect’.

© 2014 International Endodontic Journal. Published by John Wiley & Sons Ltd International Endodontic Journal 1
Deformation and fracture incidence of Reciproc Plotino et al.

Several studies have shown that the actual recipro- made passively with a size 30 and 20 stainless-steel
cating movement of endodontic instruments is asso- K-file (VDW) to choose the correct size of the Reciproc
ciated with an extended cyclic fatigue instrument life file to be used: R25, R40 or R50. In cases in which
in comparison with conventional rotation (De-Deus R40 or R50 were used, a stainless-steel K-file size 20
et al. 2010, Pedulla et al. 2013b). This suggests that and 30, respectively, reached the apex passively. In
such instruments could operate for a longer period of these cases, working length was established using an
time before failure (Wan et al. 2011) and confirms electronic apex locator and patency was maintained
that alternating rotation appears to be a valid and with a size 10 stainless-steel hand K-file (VDW).
safe alternative to continuous rotary movement (Va- In cases in which R25 was used, the coronal and
no et al. 2010).
rela-Pati~ middle thirds were initially prepared based on a work-
The advantage of a single-use instrument has been ing length estimated from the preoperative radiograph,
reported by Arens et al. (2003) who stated that without any prior glide path with hand instruments.
defects could occur even with new files in the hands In all the cases, manufacturer’s instructions were fol-
of experienced endodontists, and therefore, they con- lowed in doing three up-and-down strokes advancing
cluded that for absolute safety, a single-use approach with the instrument, then removing it, cleaning the
should be followed. These observations were con- flutes and flushing the root canals with 5.25% sodium
firmed by Shen et al. (2009d) who recommended the hypochlorite. These steps were repeated until the esti-
single-use of endodontic NiTi instruments to reduce mated working length of the two coronal thirds was
instrument fatigue and the possibility of cross- reached by the instrument. When R40 or R50 instru-
contamination. ments were used, this procedure was repeated until the
Following these concepts, new single-file and sin- entire root canal length previously established with the
gle-use reciprocating NiTi instruments have been electronic apex locator was reached.
introduced. However, so far, no study has evaluated When preparation was performed with R25 instru-
the deformation and fracture of these instruments ment, a straight (not pre-bent) size 10 stainless-steel
during clinical use. The aim of this study was to hand K-file was used to check apical patency and to
investigate the incidence of fracture and deformation determine the final electronic working length as sug-
observed on Reciproc instruments (VDW, Munich, gested by the manufacturer (http://www.vdw-dental.
Germany) after clinical use. com/fileadmin/redaktion/downloads/presse/yared_reci
proc_concept_en.pdf). Preparation of the apical third
was completed with R25 with the same technique
Materials and methods
described previously, until the entire root canal lenght
A total of 1696 Reciproc NiTi instruments that were was reached. In cases in which a straight hand stain-
discarded after normal clinical usage over 30 months less-steel size 10 K-file did not reach the working
were collected. Three operators (GP, NMG and PFP) length, the R25 instrument was used to the point where
prepared 3780 root canals divided into 3023 primary the hand instrument had reached and the apex was pre-
treatments and 757 retreatments (Table 1). pared by hand with stainless-steel and NiTi hand files.
Root canal preparation was standardized strictly In retreatment cases, after the removal of restora-
following the manufacturer’s directions for the use of tions, the most coronal part of the root filling material
Reciproc files (http://www.vdw-dental.com/fileadmin/ was removed by Gates Glidden or ultrasonic tips.
redaktion/downloads/presse/yared_reciproc_concept_en. Reciproc R25 instruments were used to remove all
pdf). filling material and to complete the preparation as
In primary treatment cases, the pulp chamber was previously described once the root canal was made
opened using burs and ultrasonic tips and the coronal patent with a size 10 stainless-steel hand K-file. In
pulp tissue removed. Scouting of the root canals was those retreatment cases in which it was not possible
to reach the entire working length, Reciproc files were
Table 1 Number of teeth treated in this study used to the point of the canal where the hand file
reached. In all cases, a lateral brushing cutting action
Central Lateral
was performed to relocate the orifices and to prepare
incisors incisors Canines Premolars Molars
around the entire canal circumference.
Maxillary 80 75 53 298 509
All Reciproc files were used driven by an endodon-
Mandibular 18 47 32 142 394
tic motor (Silver Reciproc or Gold Reciproc; VDW)

2 International Endodontic Journal © 2014 International Endodontic Journal. Published by John Wiley & Sons Ltd
Plotino et al. Deformation and fracture incidence of Reciproc

with its specific reciprocating function (Reciproc ALL)


Results
and a 6:1 hand-piece (Sirona Dental Systems GmbH,
Bensheim, Germany). A total of 1696 Reciproc NiTi instruments were dis-
After canal preparation, irrigation using ultrasoni- carded within 30 months of clinical practice. Overall,
cally activated 5.25% sodium hypochlorite and 17% 1580 instruments were R25 (93%), 76 were R40
EDTA was used, and canals dried and then filled in (5%), and 40 were R50 (2%).
the same or a subsequent visit. No statistical significant differences were found
Each instrument was used for shaping the canals of between the number of fractured and deformed
only one tooth. All the instruments used were ultra- instruments (P = 0.44). A total of 8 Reciproc R25
sonically cleaned and sterilized by autoclaving before instruments were fractured during treatments, which
inspection under an optical stereomicroscope at sev- represents 0.47% of the number of instruments used
eral magnifications to assess for the signs of fracture and 0.21% of the root canals treated. Five instru-
or deformation. The discarded files were grouped ments fractured in primary treatment cases, which
according to the file size (R25, R40 and R50) and, if represents 0.29% of the number of instruments used
any fracture existed, the length of the files was mea- and 0.13% of the root canals treated, and three
sured from the shaft to the tip using a digital calliper, instruments during retreatments, which represents
to determine the length of the fractured fragment. A 0.18% of the number of instruments used and 0.08%
database of all cases involving Reciproc instruments of the root canals treated. No significant difference
was maintained and included data on instrument was found for fractured instruments between treat-
fracture, such as tooth and canal involved, type of ment and retreatment cases (P = 0.31).
the treatment (primary treatment or retreatment), size Five of the fractured fragments were removed
of fractured instrument, length of the fractured frag- (three during primary treatments and two during
ment and resolution of the case, that is, instrument retreatments) (Fig. 1), one of the instruments frac-
removed, bypassed or remained in the canal. tured during a primary treatment was bypassed, and
Data were analysed using a chi-square test with a only two instruments remained inside the root canal.
level of significance set at P < 0.05. In two cases, the length of the fractured fragment

Figure 1 (a) Radiograph of a mandibular second molar with a fragment of fractured instrument; (b) Radiograph in which the
fragment was partially removed; (c) Radiograph with gutta-percha cones after the complete removal of the fractured fragment;
(d) Postoperative radiograph after three-dimensional filling of the root canal and direct restoration with a fibre post in the distal
canal; (e) Photograph of the access cavity after root canal filling; (f) Photograph immediately after the coronal reconstruction;
(g) 1-year radiographic control; (h) 1-year control photograph.

© 2014 International Endodontic Journal. Published by John Wiley & Sons Ltd International Endodontic Journal 3
Deformation and fracture incidence of Reciproc Plotino et al.

was 16 mm from the tip of the instrument (one in ments, fracture can occur during clinical use (Paras-
the palatal canal of a maxillary first molar and one in hos & Messer 2006). Malentacca & Lalli (2002)
the single canal of a maxillary second premolar) and suggested that a reciprocating rotation might repre-
they were removed, whilst the other six fractures sent an advantage for the safer use of NiTi instru-
were reported between 2 and 3 mm from the tip. One ments, thus reducing fractures and deformations.
of these instruments fractured 3 mm from the apex in The results of the present study reveal a low per-
the buccal canal of a maxillary first premolar centage of fracture (<0.5%) and deformations (<0.4%)
(removed), one 5 mm from the apex in the disto-buc- for Reciproc NiTi instruments. This percentage is
cal canal of a maxillary second molar (removed) and lower than most of the previously published fracture
another one at the canal terminus of the mesio- rates on rotary instruments (Table 2) and is similar
buccal canal in a maxillary first molar (not bypasssed to that of De-Deus et al. (2013) in which only a single
nor removed). Three R25 instruments fractured in fracture (0.2% of 502 root canals instrumented with
second mandibular molars, one inside a curvature in Reciproc) occurred in a root canal with moderate cur-
the middle portion of the distal canal (bypassed), one vature. Therefore, the reciprocating movement may
at the orifice of the mesio-lingual canal (removed) play a role in preventing instrument fracture, as the
and the other one in the apical portion of the mesio- cyclical reversal of rotation seems to drastically
lingual canal in a C-shaped tooth (not bypasssed nor reduce torsional stress. Furthermore, the reciprocating
removed). rotation that is a characteristic of the Reciproc move-
Six Reciproc R25 instruments deformed during clin- ment may help the instrument in its advance towards
ical use (0.35% of the number of instruments used the canal terminus, following a mechanical balanced
and 0.15% of the root canals treated), four during force advance, as reported by De-Deus et al. (2013),
retreatments (0.23% of the number of instruments whilst reducing the risk of blockage for the tip of the
used and 0.10% of the root canals treated) and two instrument (Gavini et al. 2012). In fact, the actual
during primary treatments (0.11% of the number of reciprocating rotation aims to minimize the risk of
instruments used and 0.05% of the root canals trea- fracture of the instrument caused by torsional stress
ted). No significant difference was found for deformed as the angle of rotation in the direction of cutting is
instruments between treatments and retreatment
cases (P = 0.24). The deformation reported was
Table 2 Fracture and deformation rate of rotary instruments
unwinding of the blades, one at 1 mm from the tip reported in previous studies
and five between 2 and 3 mm from the tip.
No deformations or fracture were registered for the Fractures, Deformations,
% %
Reciproc R40 and R50 instruments.
Ramirez-Solomon et al. (1997) 3.7
Baumann & Roth (1999) 9.4
Discussion Sattapan et al. (2000) 21 28
Hulsmann et al. (2003) 6–10
The present study aimed to evaluate the deformation Parashos et al. (2004) 5 12
and fracture incidence of Reciproc NiTi instruments Ankrum et al. (2004) 1.7–6 2.4–15.3
discarded after normal clinical use. The use of NiTi Alapati et al. (2005) 3–23 7–22
instruments for shaping root canals represents the Peng et al. (2005) 23
Cheung et al. (2005) 23
gold standard of contemporary endodontic practice. It
Shen et al. (2006) 7–14 0.3–5
is well established that rotary NiTi files are faster and Wolcott et al. (2006) 2.4
produce equal or less canal transportation than stain- Di Fiore et al. (2006) 1.9
less-steel hand files whilst also being associated with Knowles et al. (2006) 1.3
reduced operator fatigue, fewer procedural errors and Cheung et al. (2007) 14
Wei et al. (2007) 12.9
shorter duration of canal preparation (Bergmans et al.
Tzanetakis et al. (2008) 1.8
2001, Peters 2004, Peters & Paque 2010). The Inan & Gonulol (2009) 16 10
mechanical properties of NiTi have enabled endodon- Shen et al. (2009a,b) 5 3
tic files to be more flexible, to be better able to con- Shen et al. (2009c) 0.3 1
form to canal curvature, to resist fracture and to Shen et al. (2009d) 0.26 2.9
Wu et al. (2011) 2.6
wear less than stainless-steel files (Plotino et al.
De-Deus et al. (2013) 0.2
2009). Despite the advantages of NiTi rotary instru-

4 International Endodontic Journal © 2014 International Endodontic Journal. Published by John Wiley & Sons Ltd
Plotino et al. Deformation and fracture incidence of Reciproc

designed to be less than the elastic limit of the instru- as the treatments were completed normally. This is of
ment (Kim et al. 2012). particular importance in cases with primary infection
The very low frequency of R25 instrument fracture (Ungerechts et al. 2014). Two R25 files fractured
reported in the present study can be possibly under- 16 mm from the tip of the instrument during the pas-
stood also as an improvement in cyclic fatigue resis- sive brushing action in the outward lateral move-
tance associated with the reciprocation kinematics ment. For this reason, both instruments were outside
(Pedulla et al. 2013a,b), the instrument cross-sec- the orifice, not engaged in the dentine and their
tional design (Grande et al. 2006, Plotino et al. 2012) retrieval was performed easily (Fig. 2).
and the superelastic M-wire alloy (Johnson et al. As previously reported, the reciprocating technique
2008). Cross-sectional ‘S-shape’ design may play a was the most rapid method to effectively remove
further role to provide great cutting ability to this gutta-percha and sealer from the root canal space
instrument (B€ urklein et al. 2012, Plotino et al. in during retreatment (Zuolo et al. 2013). The results of
press). Furthermore, the use of only one NiTi file in the present study reveal that removal of old filling
reciprocation to prepare the entire root canal, as pro- materials and re-preparation of the root canals was
vided by the Reciproc technique, may reduce the performed with a very low fracture rate using Reci-
accumulation of metal fatigue, costs and avoid cross- proc R25 instrument in retreatment cases. Deforma-
infection (Shen et al. 2009d, De-Deus et al. 2010). tions occurred more frequently in retreatments than
In the present study, six of the eight fractured files primary treatments (0.2% vs. 0.1%), as during the
were removed or bypassed. In these cases, file fracture removal of filling materials, too much pressure may
seems to have no impact from a clinical point of view, be exerted in some cases.

Figure 2 (a) Radiograph of a maxillary first molar with a 16-mm fragment of fractured instrument in the palatal canal; (b)
Intra-operative radiograph of the working length after the removal of the fractured instrument; (c) Photograph of the access
cavity with the fractured instrument protruding outside the orifice of the palatal canal; (d) Photograph of the fragment after
removal using the Instrument Removal System; (e) Photograph of the access cavity after the removal of the fractured instru-
ment; (f) Postoperative radiograph after three-dimensional filling of the root canal system; (g) Photograph of the indirect lith-
ium disilicate onlay immediately after cementation; (h) 1-year radiographic control.

© 2014 International Endodontic Journal. Published by John Wiley & Sons Ltd International Endodontic Journal 5
Deformation and fracture incidence of Reciproc Plotino et al.

Conclusions Frank AL (1967) An evaluation of the Giromatic endodontic


handpiece. Oral Surgery Oral Medicine Oral Pathology 24,
Root canal instrumentation can be performed with 419–21.
Reciproc system used as single instrument in all types Gavini G, Caldeira C, Akisue E, Candeiro G, Kawakami D
of root canals with a very low incidence of instru- (2012) Resistance to flexural fatigue of Reciproc R25 files
ment fracture and deformation. under continuous rotation and reciprocating movement.
Journal of Endodontics 38, 684–7.
Grande NM, Plotino G, Pecci R, Bedini R, Malagnino VA,
References
Somma F (2006) Cyclic fatigue resistance and three-
Alapati SB, Brantley WA, Svec TA, Powers JM, Nusstein JM, dimensional analysis of instruments from two nickel-tita-
Daehn GS (2005) SEM observations of nickel-titanium nium rotary systems. International Endodontic Journal 39,
rotary endodontic instruments that fractured during clini- 755–63.
cal use. Journal of Endodontics 31, 40–3. Hulsmann M, Herbst U, Schafers F (2003) Comparative
Ankrum MT, Hartwell GR, Truitt JE (2004) K3 Endo, ProTa- study of root-canal preparation using Lightspeed and
per and Profile systems: breakage and distortion in Quantec SC rotary NiTi instruments. International Endodon-
severely curved roots of molars. Journal of Endodontics 30, tic Journal 36, 748–56.
234–7. Inan U, Gonulol N (2009) Deformation and fracture of Mtwo
Arens FC, Hoen MM, Steiman HR, Dietz GC Jr (2003) Evalu- rotary nickel-titanium instruments after clinical use. Jour-
ation of single-use rotary nickel-titanium instruments. nal of Endodontics 35, 1396–9.
Journal of Endodontics 29, 664–6. Johnson E, Lloyd A, Kuttler S, Namerow K (2008) Compari-
Baumann MA, Roth A (1999) Effect of experience on quality son between a novel nickel-titanium alloy and 508 nitinol
of canal preparation with rotary nickel-titanium files. Oral on the cyclic fatigue life of ProFile 25/.04 rotary instru-
Surgery, Oral Medicine, Oral Pathology, Oral Radiology and ments. Journal of Endodontics 34, 1406–9.
Endodontics 88, 714–8. Klayman SM, Brilliant JD (1975) A comparison of the effi-
Bergmans L, Van Cleynenbreugel J, Wevers M, Lambrechts P cacy of serial preparation versus Giromatic preparation.
(2001) Mechanical root canal preparation with NiTi Journal of Endodontics 10, 334–7.
rotary instruments: rationale, performance and safety. Sta- Kim H, Kwak S, Cheung G, Ko D, Chung S, Lee W (2012)
tus report for the American Journal of Dentistry. American Cyclic fatigue and torsional resistance of two new nickel-
Journal of Dentistry 14, 324–33. titanium instruments used in reciprocation motion: Reci-
B€
urklein S, Hinschitza K, Dammaschke T, Sch€ afer E (2012) proc versus WaveOne. Journal of Endodontics 38, 541–4.
Shaping ability and cleaning effectiveness of two single-file Knowles KI, Hammond NB, Biggs SG, Ibarrola JL (2006)
systems in severely curved root canals of extracted teeth: Incidence of instrument separation using LightSpeed
Reciproc and WaveOne versus Mtwo and ProTaper. Inter- rotary instruments. Journal of Endodontics 32, 14–6.
national Endodontic Journal 45, 449–61. Malentacca A, Lalli F (2002) Use of nickel-titanium instru-
Cheung GS, Peng B, Bian Z, Shen Y, Darvell BW (2005) ments with reciprocating movement. Italian Journal of End-
Defects in ProTaper S1 instruments after clinical use: frac- odontics 16, 79–84.
tographic examination. International Endodontic Journal 38, Parashos P, Messer HH (2006) Rotary NiTi instrument frac-
802–9. ture and its consequences. Journal of Endodontics 32,
Cheung GS, Bian Z, Shen Y, Peng B, Darvell BW (2007) 1031–43.
Comparison of defects in ProTaper hand-operated and Parashos P, Gordon I, Messer HH (2004) Factors influencing
engine-driven instruments after clinical use. International defects of rotary nickel-titanium endodontic instruments
Endodontic Journal 40, 169–78. after clinical use. Journal of Endodontics 30, 722–5.
De-Deus G, Moreira E, Lopes H, Elias C (2010) Extended cyc- Pedulla E, Grande NM, Plotino G, Palermo F, Gambarini G,
lic fatigue life of F2 ProTaper instruments used in recipro- Rapisarda E (2013a) Cyclic fatigue resistance of two recip-
cating movement. International Endodontic Journal 43, rocating nickel-titanium instruments after immersion in
1063–8. sodium hypochlorite. International Endodontic Journal 46,
De-Deus G, Arruda TE, Souza EM et al. (2013) The ability of 155–9.
the Reciproc R25 instrument to reach the full root canal Pedulla E, Grande NM, Plotino G, Gambarini G, Rapisarda E
working length without a glide path. International Endodon- (2013b) Influence of continuous or reciprocating motion
tic Journal 46, 993–8. on cyclic fatigue resistance of 4 different nickel-titanium
Di Fiore PM, Genov KA, Komaroff E, Li Y, Lin L (2006) rotary instruments. Journal of Endodontics 39, 258–61.
Nickel-titanium rotary instrument fracture: a clinical Peng B, Shen Y, Cheung GSP, Xia TJ (2005) Defects in Pro-
practice assessment. International Endodontic Journal 39, Taper S1 instruments after clinical use: longitudinal exam-
700–8. ination. International Endodontic Journal 38, 550–7.

6 International Endodontic Journal © 2014 International Endodontic Journal. Published by John Wiley & Sons Ltd
Plotino et al. Deformation and fracture incidence of Reciproc

Peters OA (2004) Current challenges and concepts in the spective study from an undergraduate clinic. Journal of
preparation of root canal systems: a review. Journal of End- Endodontics 35, 193–6.
odontics 30, 559–67. Shen Y, Coil JM, McLean AG, Hemerling DL, Haapasalo M
Peters OA, Paque F (2010) Current developments in rotary (2009d) Defects in nickel-titanium instruments after
root canal instrument technology and clinical use: a clinical use. Part 5: single use from endodontic specialty
review. Quintessence International 41, 479–88. practices. Journal of Endodontics 35, 1363–7.
Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G Tzanetakis GN, Kontakiotis EG, Maurikou DV, Marzelou MP
(2009) A review of cyclic fatigue testing of nickel-titanium (2008) Prevalence and management of instrument frac-
rotary instruments. Journal of Endodontics 35, 1469–76. ture in the postgraduate endodontic program at the Dental
Plotino G, Grande NM, Testarelli L, Gambarini G (2012) Cyc- School of Athens: a five-year retrospective clinical study.
lic fatigue of Reciproc and WaveOne reciprocating instru- Journal of Endodontics 34, 675–8.
ments. International Endodontic Journal 45, 614–8. Ungerechts C, B ardsen A, Fristad I (2014) Instrument frac-
Plotino G, Giansiracusa Rubini A, Grande NM, Testarelli L, ture in root canals – where, why, when and what? A
Gambarini G (in press) Cutting efficiency of Reciproc and study from a student clinic. International Endodontic Journal
WaveOne reciprocating instruments. Journal of Endodontics. 47, 183–90.
doi: 10.1016/j.joen.2014.01.041. Varela-Pati~no P, Iba~
nez-Parraga A, Rivas-Mundi~ na B, Canta-
Ramirez-Solomon M, Soler-Bientz R, de la Garza-Gonzalez R, tore G, Otero XL, Martin-Biedma B (2010) Alternating
Palacios-Garza CM (1997) Incidence of Lightspeed separa- versus continuous rotation: a comparative study of the
tion and the potential for bypassing. Journal of Endodontics effect on instrument life. Journal of Endodontics 36, 157–9.
23, 586–7. Wan J, Rasimick BJ, Musikant BL, Deutsch AS (2011) A
Sattapan B, Nervo GJ, Palamara JE, Messer HH (2000) comparison of cyclic fatigue resistance in reciprocating
Defects in rotary nickel-titanium files after clinical use. and rotary nickel-titanium instruments. Australian End-
Journal of Endodontics 26, 161–5. odontic Journal 37, 122–7.
Shen Y, Cheung GS, Bian Z, Peng B (2006) Comparison of Wei X, Ling J, Huang X, Liu L (2007) Modes of failure of
defects in ProFile and ProTaper systems after clinical use. ProTaper nickel-titanium rotary instruments after clinical
Journal of Endodontics 32, 61–5. use. Journal of Endodontics 33, 276–9.
Shen Y, Haapasalo M, Cheung GS, Peng B (2009a) Defects Wolcott S, Wolcott J, Ishley D et al. (2006) Separation inci-
in nickel-titanium instruments after clinical use. Part 1: dence of ProTaper rotary instruments: a large cohort clini-
relationship between observed imperfections and factors cal evaluation. Journal of Endodontics 32, 1139–41.
leading to such defects in a cohort study. Journal of End- Wu J, Lei G, Yan M, Yu Y, Yu J, Zhang G (2011) Instrument
odontics 35, 129–32. separation analysis of multi-used ProTaper Universal
Shen Y, Cheung GS, Peng B, Haapasalo M (2009b) Defects rotary system during root canal therapy. Journal of End-
in nickel-titanium instruments after clinical use. Part 2: odontics 37, 758–63.
fractographic analysis of fractured surface in a cohort Zuolo AS, Mello JE Jr, Cunha RS, Zuolo ML, Bueno CE
study. Journal of Endodontics 35, 133–6. (2013) Efficacy of reciprocating and rotary techniques for
Shen Y, Coil JM, Haapasalo M (2009c) Defects in nickel-tita- removing filling material during root canal retreatment.
nium instruments after clinical use. Part 3: a 4-year retro- International Endodontic Journal 46, 947–53.

© 2014 International Endodontic Journal. Published by John Wiley & Sons Ltd International Endodontic Journal 7

You might also like