DOC NO.: MQS03 C01 Date Effective: 01/06/2017 Issue No.: 7 Page No. : 1 / 2
PREPARATION
CALIBRATION
9. The combustible gas Monitor is adjusted to read zero with inlet fresh air.
10. The applicable standard Calibration gas as given below is applied to the inlet of the instrument.
06. Butane 8% -
METROMAC MARINE AUTOMATION /METROMAC AUTOMATION
Title : CALIBRATION OF Prepared By: MR Approved By: MD / Joint M D
WORK INSTRUCTIONS COMBUSTIBLE GAS MONITORS
DOC NO.: MQS03 C01 Date Effective: 01/06/2017 Issue No.: 7 Page No. : 2 / 2
11. After 1 minute, the read out must be adjusted to the corresponding value using Span Control.
12. If Span is not adjustable check sensor or Control PCB and note that accordingly in the job
sheet.
13. Set high alarms as required (as per instruction manual).
14. Repeat the step 8 to 10 three to four times and ensure that the calibration is successful.
15. Repeat step 8 and reconfirm the calibration on purging with fresh air.
RECORDS
16. Note down the ‘As found’ and ‘As left’ readings carefully in the job sheet.
17. Provide ‘Calibration’ sticker with necessary details if the error is within limits.
18. Note down the Test method variation, if any, in the job sheet.
19. REFERENCE
1. BS EN 60079 -29-1: 2016
2. BS EN 60079-29-2-:2015
Manufacturer’s manual