 Home / Analyzers / Chromatograph Sample Valves

ANALYZERS

Chromatograph Sample Valves


 S Bharadwaj Reddy  Leave a comment

Ads
Like 0 Tweet Share 0 Share

Arguably, the component most critical to measurement accuracy in a gas chromatograph is the sample valve. Its purpose is to
inject the exact same sample quantity into the column at the beginning of each cycle. If the sample quantity is not repeatable, the
measured quantities exiting the column will change from cycle to cycle even if the sample composition does not change. If the
valve’s cycle time is not repeatable, species separation e ciency will vary from cycle to cycle. If the sample valve leaks such that a
small ow rate of sample continuously enters the column, the result will be an altered “baseline” signal at the detector (at best)
and total corruption of the analysis (at worst). Many process chromatograph problems are caused by irregularities in the sample
valve(s).

Cate
A photograph of the column (the coil of ne tubing about 6 inches in diameter, on the left) and sample valve (stainless-steel
cylinder with several tubes entering and exiting, on the right) for a gas chromatograph appears here: Sele

Ad

Ga


https://instrumentationtools.com/chromatograph-sample-valves/#.WoQNsKhuZPY 1/5
2/14/2018 Chromatograph Sample Valves Instrumentation Tools

ALL ANIMATION BASICS  MEASUREMENT  CONTROL SYSTEMS  Q&A  ELECTRONICS  ELECTRICAL  TOOLS MORE 

Like

A common form of sample valve uses a rotating element to switch port connections between the sample gas stream, carrier gas
stream, and column:

Late

Moh
The
whic

HAR
Hello
tech

tush
simp

Amit
It's g

Pert
It he

Jack
Prett
parti

hem
Three slots connect three pairs of ports together. When the rotary valve actuates, the port connections switch, redirecting gas sir h
more
ows.

Connected to a sample stream, carrier stream, and column, the rotary sample valve operates in two di erent modes. The rst
mode is a “loading” position where the sample stream ows through a short length of tubing (called a sample loop) and exits to a
waste discharge port, while the carrier gas ows through the column to wash the last sample through. The second mode is a
“sampling” position where the volume of sample gas held in the sample loop tubing gets injected into the column by a ow of
carrier gas behind it:

Inst


https://instrumentationtools.com/chromatograph-sample-valves/#.WoQNsKhuZPY 2/5
2/14/2018 Chromatograph Sample Valves Instrumentation Tools

Inst
ALL ANIMATION BASICS  MEASUREMENT  CONTROL SYSTEMS  Q&A  ELECTRONICS  ELECTRICAL  TOOLS MORE 

Lang

Sele

Write With Con dence


Improve grammar, word choice, and sentence structure in your writing. It's free!

Ad Grammarly Visit Site

The purpose of the sample loop tube is to act as a holding reservoir for a xed volume of sample gas. When the sample valve
switches to the sample position, the carrier gas will ush the contents of the sample loop toward the column. This valve
con guration guarantees that the injected sample volume cannot vary even if the sample valve’s actuation is not precise. The
sample valve need only remain in the “sampling” position long enough to completely ush the sample loop tube, and the proper
volume of injected sample gas is guaranteed. An analogy for the sample loop is that of a  measuring cup held underneath a
continuously-spilling stream of water: so long as the cup is held beneath the stream long enough to completely ll, it is guaranteed
to deliver a xed volume of water when removed from the stream and emptied. Like a measuring cup, the sample loop need only
be lled completely with sample gas to deliver a xed and unerring volume of gas to the chromatograph column when switched
from the ”loading” position to the “sampling” position.

While in the loading position, the stream of gas sampled from the process continuously lls the sample loop and then exits to a
waste port. This may seem wasteful but in fact is quite essential for practical sampling operation. The volume of process gas
injected into the chromatograph column during each cycle is so small (typically measured in units of microliters!) that a continuous
ow of sample gas to waste is necessary to purge the impulse line connecting the analyzer to the process and thereby ensure a
fresh sample, which in turn is necessary for the analyzer to obtain analyses of current conditions. If it were not for the continuous
ow of sample to waste, it would take a very long time for a sample of process gas to make its way through the long impulse tube
to the analyzer to be sampled, resulting in grossly delayed measurements of process conditions! 
https://instrumentationtools.com/chromatograph-sample-valves/#.WoQNsKhuZPY 3/5
2/14/2018 Chromatograph Sample Valves Instrumentation Tools

ALL ANIMATION BASICS  MEASUREMENT  CONTROL SYSTEMS  Q&A  ELECTRONICS  ELECTRICAL  TOOLS MORE 

Facebook Twitter Print Email More

Arduino Project Builder Oxygen Analyzer Power Cable What is calibration


Working Principle Manufanturer
Instrumentation Tools
Ad Cayenne instrumentationtools.com Ad Tinde Power Co., Ltd instrumentationtools.com

MS Deconvolution pH sensor calibration HART Communication Instrumentation


Software procedure Tutorial Part 1 Dictionary
Instrumentation Tools Instrumentation Tools Instrumentation Tools
Ad SpectralWorks instrumentationtools.com instrumentationtools.com instrumentationtools.com

Previous Post:

Industrial Applications of
 Chromatographs

Related Articles

ANALYZERS • GUEST ARTICLES ANALYZERS

Gas Analyzer Sampling System Online Gas Chromatograph Principle


 November 3, 2017  February 13, 2018

ANALYZERS ANALYZERS

Basics of Total Organic Carbon (TOC) Analyzers Thin-Layer Chromatography : Manual Method
 November 23, 2016  February 13, 2018

https://instrumentationtools.com/chromatograph-sample-valves/#.WoQNsKhuZPY 4/5
2/14/2018 Chromatograph Sample Valves Instrumentation Tools
ANALYZERS • CALIBRATION ANALYZERS
ALL ANIMATION BASICS  MEASUREMENT  CONTROL SYSTEMS Q&A ELECTRONICS
  ELECTRICAL  TOOLS MORE 
Speci c Conductance Calibration Procedure What is Chromatography ?
 May 14, 2015  February 13, 2018

ANALYZERS • CALIBRATION ANALYZERS

Oxidation-Reduction Potential (ORP) Sensor What is Gas Metering System ?


Calibration Procedure  September 24, 2016

 May 14, 2015

ANALYZERS • INTERVIEW QUESTIONS ANALYZERS • INTERVIEW QUESTIONS

Total Organic Carbon Analyzers Questions & Answers Gas Analyzers Interview Questions
 June 21, 2016  September 21, 2015

ANALYZERS • ANIMATION ANALYZERS

TOC Analyzer Working Principle Humidity Measurement Principle


 July 25, 2016  May 16, 2016

Copyright © 2015 to 2017. Created by S Bharadwaj Reddy.


https://instrumentationtools.com/chromatograph-sample-valves/#.WoQNsKhuZPY 5/5

You might also like