You are on page 1of 5

© Copyright AAMI 2011. Single user license only. Copying, networking, and distribution prohibited.

Columns and Departments

THE FUNDAMENTALS OF ...

Apnea Monitors
Robert M. Dondelinger

Apnea is a Greek word meaning “without during power outages. They can be thought of About the Author
wind.” Apnea, in the modern lexicon, refers to as being comprised of three modules—one that
Robert Dondelinger,
the cessation of breathing and is a reversible senses breathing, another that determines CBET-E, MS, is the
condition if caught early enough. Apneic events apneic events, and a third that reports and/or senior medical
fall under one of three classifications—central, records these events. logistician at
obstructive, or mixed. The event is classified as Sensing breathing: Because most apnea the U.S. Military
“central apnea” if the brain stops sending the monitors are used on infants, respiration is Entrance Processing
Command in
signals to the chest muscles to expand, causing often indirectly detected. Indirect detection
North Chicago, IL. E-mail: robert.
inhalation. This can occur even when the methods include mattress pressure pads, dondelinger@mepcom.army.mil
patient has normal muscle function in the impedance pneumography, and pneumatic
chest. An upper airway obstruction of the abdominal sensors. Mattress pressure pads
airflow is termed “obstructive apnea.” When detect changes in resistance or capacitance of a
elements of both are present, it is termed mattress transducer. Small pressure changes
“mixed apnea.” caused by respiratory motion are interpreted as
Prolonged apnea reduces blood and tissue breaths while large movements are essentially
oxygen levels, leading to permanent brain ignored by the electronics.
damage and, if not remediated, death. Apnea The popular impedance pneumography
occurring during otherwise normal sleep is method employs basically the same technology
called “sleep apnea” and is suspected to be a used in respiration modules found in inten-
factor in Sudden Infant Death Syndrome sive care unit (ICU)
(SIDS). Because the overall risk factors for cardiac monitors. They
apnea are reasonably well defined but the actual detect changes in
Apnea monitors basically monitor
onset is unpredictable, apnea monitors provide thoracic impedance by respirations and usually initiate both
constant monitoring of those at risk and set off passing a low-current visual and audible alarms when the time
an alarm when apnea is identified. high frequency signal between breaths, presumed to be an
through two electrodes
episode of apnea, exceeds a preset time.
Current Technology placed on either side of
Apnea monitors basically monitor respirations the chest. As the infant
and usually initiate both visual and audible breathes and the heart pumps blood, the
alarms when the time between breaths, transthoracic impedance changes and these
presumed to be an episode of apnea, exceeds a changes are interpreted as breaths. These
preset time. Many times, there is an integral same electrodes often perform double duty as
battery backup to provide continued operation electrocardiogram (ECG) electrodes to monitor
for portability, such as while traveling, and heart activity.

Biomedical Instrumentation & Technology Nov/Dec 2011 477


© Copyright AAMI 2011. Single user license only. Copying, networking, and distribution prohibited.
Columns and Departments

Carbon dioxide monitoring for apnea


detection works similar to end-tidal CO2
monitoring, employing a cannula placed in
front of the patient’s nose which is connected
by small-bore tubing to a small pump that
constantly samples gas from the cannula. By
constantly sampling the air, typically using an
infrared CO2 sensor, repetitive variations in CO2
levels are indicative of breathing.
Determining an apneic event: No matter the
method used, this is fairly straightforward.
Apnea monitors determine apneic events by
measuring the time between breaths and
comparing that to a preset standard. Typically
this standard is an adjustable interval as long as
60-seconds between breaths—the range of
adjustment varies between brands and mod-
els—with the actual setting determined by a
healthcare professional. Apnea monitors that
use their electrodes to also provide the heart rate
Apnea is believed to be a factor in Sudden Infant Death Syndrome (SIDS). Apnea monitors are
used on infants believed to be susceptible to SIDS. typically allow the user to set a lower rate limit
to provide an alarm if the heart rate is too slow
(bradycardia), probably caused by hypoxia (too
Pneumatic abdominal sensors are similar to little oxygen reaching body tissues and the
those found in the mattress pressure pad, but are brain), or too fast (tachycardia). Models that also
enclosed within a band worn around the monitor the ECG use a more complex algorithm
abdomen. The sensor detects the expansion and which takes both the time between breaths and
contraction of the abdomen caused by breathing the slower heart rate into account before
and interprets the pressure changes as breaths. indicating the patient has stopped breathing.
There are several other designs which High-end units provide a plethora of
utilize direct detection methods that either adjustments, including both upper and lower
sample or introduce some form of sensor into ECG alarm settings and amplitude adjust-
the breath stream of the individual being ments to better differentiate between muscle
monitored. These methods typically employ artifact (noise) and a breath or a heartbeat.
temperature, pressure, or carbon dioxide These additional adjustments are intended to
(CO2) sensors. One unusual design monitors reduce the incidence of false alarms while
tracheal breath sounds. allowing the user to precisely determine
Two direct methods of sensing apnea involve apneic alarm events.
placing a sensor under the nose or in front of Reporting/recording apneic events: Although
the mouth. Temperature monitors place a this varies tremendously between models, all
thermistor under the nose or in low-end units provide both audible and visual
front of the mouth. The ther- alarms. At a minimum, low-end units provide
Apnea monitors determine apneic the user with several choices of delay time for
mistor senses the temperature
events by measuring the time of the air being inhaled or determining apnea, and an audible alarm which
between breaths and comparing exhaled by the patient—typically resets either automatically or manually. On
that to a preset standard. cool air on inspiration and warm most units featuring self-resetting audible
on expiration. The change or alarms, there is usually a manual reset for the
difference in temperature indicates that the visual alarm indicator so that the light remains
patient is still breathing. Pressure monitors illuminated after the audible alarm resets.
work similarly, except that this method meas- Better monitors flash a light or make an audible
ures the small pressure fluctuations at the click each time a breath is detected and provide
nostril created by inhaling and exhaling, thus a visual indication (typically a digital readout) of
indicating respirations. respiration and heart rates.

478 Biomedical Instrumentation & Technology Nov/Dec 2011


© Copyright AAMI 2011. Single user license only. Copying, networking, and distribution prohibited.
Columns and Departments

Some apnea monitors also include optional similar to mistaking muscle artifact for
onboard blood pressure and pulse oximetry to breathing, is the possibility that other signals,
provide complete patient monitoring in a such as electrode vibrations, electronic signals
single device. Upper and lower alarm limits from other devices in the patient vicinity
can also be set for these vital signs. High-end (remember, apneas monitors are often used in
units often provide external documentation the patient’s home) will interfere with breathing
capabilities, such as the ability to “remember” detection and be counted as false respirations.
and download several hours of patient event Third, depending on the method of sensing
data, especially events leading up to an apneic breathing the monitor designer chose to use,
episode, to a microcomputer using a propri- the basic design may be more prone to misi-
etary software package. This software dentifying other events as breathing. For
normally provides tabular or waveform data example, impedance pneumography-based Origin and Evolution
(frequently both) of monitored parameters; the designs are more sensitive to cardiovascular
number, times, and triggering events of artifact than mattress pressure pads or pneu- Apnea monitors naturally
evolved from the monitors
alarms; equipment status such as alarm matic abdominal sensor-based designs. This
used in typical hospital and
settings; the times at which it was powered on sensing method may mistake gasps for air medical center intensive care
or off; battery condition, etc. (caused by spasms or an upper-airway obstruc- units (ICUs). In the late 1960s,
tion) to be sighs or large breaths when the solid state technology had
How to Manage the Device patient is actually in distress. Likewise, designs evolved to the point where the
It is strongly recommended that maintenance that purposely do not count concurrent size and the reduced number
of apnea monitors be closely tracked, given the inhalations and heartbeats can display low of controls and adjustments
allowed for more homecare
litigious nature of today’s society and the fact respiration rates and may cause an increase in
equipment to be developed
that these monitors deal with such a vital false apnea alarms.
and fielded. The apnea moni-
function, i.e., breathing, and are used on Risk management issues related to the use tor basically is a specialized
infants believed to be susceptible to SIDS. This of the apnea monitor include irritation to the respiration/cardiac monitor mi-
means individually, by unique equipment patient’s skin from the adhesive on the nus the video scope display, re-
identification or serial number, scheduling of electrodes, users inadvertently disabling the lying instead on a tachometer
preventive maintenance, functional testing, audible alarm, and poor user (parents or or other numerical reporting
electrical safety testing, and any calibration caregiver) training on the monitor’s operation of the monitored parameters.
As monitor designs evolved,
required. Additionally, remedial maintenance and control settings.
features such as heart rate de-
should be similarly tracked and repairs scrupu- tection, automatic sensitivity,
lously documented. Troubleshooting and noise-suppressing signal
Designs of basic apnea monitors are fairly well processing algorithms became
Regulations established, so they rarely are the source of standard as did the methods
The Food and Drug Administration (FDA) failures. More commonly it is the esoteric for determining apneic events.
considers apnea monitors to be medical devices features of high-end units that are more prone Although many methods of
detecting breathing have
and are regulated by them as such. Aside from to problems.
been used on legacy units,
this FDA regulation, there is no specific Sometimes, users complain that the apnea
impedance pneumography is
regulation of apnea monitors. monitor has too many false alarms, particularly the most widely used method
while monitoring an infant susceptible to SIDS, marketed today for both home
Risk Management Issues when the patient did not experience an apneic and hospital unit use.
Using apnea monitors involves several risk event. The truth of the matter is that there was
management issues, which can be divided into likely an apneic event and the sound of the
two general categories. The first category is audible alarm surprised the infant, causing him
related to the design of the device itself, while or her to gasp and spontaneously resume
the second is related to its use. breathing. If the unit is equipped with an event
Under the category of design issues, the most monitor printer or a way to download patient
important is the possibility the monitor will event data, this scenario will be borne out by
mistakenly recognize muscle artifact as a the evidence.
breath. If it does this often enough in a minute,
it will not sense that the monitored infant is Training and Equipment
experiencing an apneic episode and delay Biomedical electronics technicians having only
caregiver response. The second design issue, basic training can perform scheduled services

Biomedical Instrumentation & Technology Nov/Dec 2011 479


© Copyright AAMI 2011. Single user license only. Copying, networking, and distribution prohibited.
Columns and Departments

(preventive maintenance, calibration, and References


electrical safety testing), operational checkouts, 1. ECRI Institute. Healthcare Product Comparison
and minimal remedial maintenance on apnea System for Apnea Monitors. Available
monitors. Operation and service manuals are at: www.ecri.org/Products/Pages/hpcs.
necessary for more advanced remedial mainte- aspx?pagewanted=all. Accessed March 2011.
nance. A safety analyzer and a patient simulator
2. University of Iowa Hospital & Clinics. Home Apnea
capable of providing both respiration and
Monitor, Infants. Available at www.uihealthcare.
heartbeat signals are required to perform org/1/007237.htm. Accessed March 2011.
operational checkouts and remedial mainte-
With most models, nance. With most models, no special tools or 3. University of Maryland Medical Center. Home
no special tools or test equipment are required to service apnea Apnea Monitor, Infants, Overview. Available
monitors, but maintenance software may be at www.umm.edu/ency/article/007237.htm.
test equipment are Accessed March 2011.
required for more advanced models.
required to service
apnea monitors, but Future Development
maintenance software Apnea monitors have become increasingly
may be required for sophisticated since they were originally
developed and used in the early 1970s. Since the
more advanced models.
basic technology is considered mature, future
improvements are expected to include the
ability to record more information (additional
channels) for a longer duration. n

480 Biomedical Instrumentation & Technology Nov/Dec 2011


© Copyright AAMI 2011. Single user license only. Copying, networking, and distribution prohibited.

Expanded and Updated in 2011…


AAMI’s Top-Rated
Clinical Engineering
Resources on one CD!
This CD includes volumes of valuable guidance The Clinical
including how-to information, Joint Commission Engineering
compliance tips, troubleshooting advice, and practical Collection
resources to sharpen technical andAssociation
career skills.
for the Advancement
of Medical Instrumentation
4301 N. Fairfax Dr, Suite 301
Arlington, VA 22203-1633
Updated and expanded in 2011, this comprehensive www.aami.org
collection includes all of these AAMI publications on
one CD:
4 Medical Equipment Management Manual, 2009
For more information or to order additional copies, call
1-877-249-8226 or 1-240-646-7031

edition Any use of materials contained in this product—including reproduction, modification,


distribution, or republication—without the prior written consent of AAMI is strictly prohibited.

4 Career Moves CD, 2010 Includes Adobe Acrobat® Reader. © 1987-2011 Adobe Systems Incorporated. All rights reserved.
Adobe and Acrobat are trademarks of Adobe Systems Incorporated.

4 The Electrical Safety Manual, 2008 ed.


4 Computerized Maintenance Management Systems
Manual
4 ANSI/AAMI EQ56:1999/(R)2008, Recommended practice for a medical equipment
management program
4 ANSI/AAMI/IEC TIR62354:2009, General testing procedures for medical electrical
equipment
4 Guidance on clinical systems for invasive blood pressure monitoring (TIR9),
protective apparel and surgical drapes (TIR11), and forcasting pacemaker battery
service life (TIR21)
4 The Fundamental Collection, 2010
4 Clinical Engineering Management Guide: Lessons from the Field
4 IT World Reference CD
4 AAMI articles on imaging, IT, healthcare technology, and hemodialysis

Order Code: CEC


List $675 / AAMI member $475
Source Code: PB

To order call +1-877-249-8226 or visit


http://marketplace.aami.org

You might also like