Professional Documents
Culture Documents
2
Department of Chemical Engineering, Chungnam National University, Taejon 305-764, Korea
important methods for investigation and charac- Subsequently, it was washed to show neutral pH
terization of ion exchangers. At the same time, a with distilled water. PI-4 (1 g) was added in 100
great discrepancy between theory and practice of ml of 1 M NaCl aqueous solution with various
this method in application to ion exchangers ex- NaOH concentrations, and was allowed to stand
ists.6 Because the inflection points of the poten- for 3 days at 20 ⫾ 1 °C. From the supernatant
tiometric titration curve for strong acidic and ba- liquid in a flask, 10 ml of solution was titrated
sic ion exchangers are obvious, the capacity of the with 0.1 N HCl standard solution. In order to
ion exchanger can be accurately determined from investigate the effects of supporting electrolytes
potentiometric titration curves. However, week of NaCl and BaCl2, the concentration of electro-
acidic and basic type ion exchangers have an equi- lytes was varied in the range of 1.0 ⵒ 1.0⫺3 M of
librium constant, which is changed by the concen- NaCl and BaCl2.
tration of supporting electrolytes, and the inflec-
tion points of the potentiometric titration curve in
this case appear ambiguous.7–12 Determination of Ion Exchange Capacity
In this work, we studied a method for compu- An error in determination of the ion exchange
tation of the main characteristics of a phosphoric capacity from the titration could be unreasonably
cation exchanger by considering the effects of po- large if the titration conditions were not chosen
tentiometric titration curves in an ion exchanger/ properly. We evaluated the precision of the deter-
solution system. The acidity parameters of ion mination by analysis of the potentiometric titra-
exchangers, such as thermodynamic constant tion curves near the equivalent point. The preci-
(pK0), apparent equilibrium constant, and correc- sion depends on the number of experimental
tion for the apparent equilibrium constant (b), points in this region, and to imitate conditions for
were used for expressing the characteristics of a the real experiments we have chosen 10ⵒ12
carboxylic ion exchanger. For proving the theory, points around the inflection point. The curves
the phosphoric cation exchanger type was used were differentiated and the differential was fitted
for an experiment on potentiometric titration and to the Lorenzian distribution curve, which ap-
potentiometric titration curves at different con- peared to be the best fit. Then, the parameters of
centrations of the supporting electrolyte. From the maximum peak in ⭸pH/⭸g were found. The g
these results, characteristic equations related to value of the maximum peak was equal to the ion
the acidity parameters of ion exchanger systems exchange capacity.
were obtained. Also, the characteristic parame-
ters such as solution pH, supporting electrolyte
concentration (C), and degree of neutralization of Theoretical Discussion
ion exchanger () were calculated. Calculation of the Acidity Function
⫹ 冉 冊
1
z
⫺ 1 log共zC M ⫹ C H兲 ⫹ pK共 , C M兲 ⫽ 0 (6)
material balance condition to find the number of
the alkali equivalent (g) added to 1 g of the ion
exchanger with capacity E in a volume V.12,13 The
relationship between g and pH in the potentio-
where pK (, CM ) ⫽ ⫺log K(, CM ) and CM is the
metric titration curve is as follows:
concentration of the supporting electrolyte. This
equation relates to the most important properties
1
of our system: pH of the solution, degree of neu- g⫽ 共E ⫹ VC OH ⫺ VC H兲 (8)
tralization of the ion exchanger (i.e. the counte- z
rion sorption value), and the concentration of a
background electrolyte. Therefore, we refer to it The equations 6,7, and 8 can be used to express
as the function of acidity of ion exchangers. The the potentiometric titration curve of the ion ex-
acidity parameters can be used to describe all changer.
ionic equilibria in the system “ion exchanger-so-
lution” containing H⫹ and counter ions Mz⫹.
RESULTS AND DISCUSSION
log ⫺ pH ⫺ log a M ⫹ pK共 , C M ⫽ 1.0兲 ⫽ 0
1⫺
(13)
log ⫺ pH ⫺ log C M ⫹ pK共 , C M ⫽ 1.0兲 ⫽ 0
Figure 1. A plot of log K versus at CNaCl ⫽ 1.0 1⫺
mol/l. The data points were experimental data and
(14)
were fitted by different interpolation curves.
log ⫺ pH ⫹ pK共 , C M ⫽ 1.0兲 ⫽ 0 (15)
1⫺
supporting electrolyte. The dependency of log K
for experimental data showed the S curve [Fig. The pH for a certain can be calculated from eq
1(a)] of a third-order equation. For sufficient ac- 15. In order to figure out the calculated potentio-
curacy, log K from eq 7 needs four parameters as metric titration curve, the g values can be ob-
follows: tained by using eq 8. Because ion charge (z) is 1,
eq 8 can be rewritten as follows:
log K ⫽ y共3,0兲共1 ⫺ 兲 3 ⫹ 3y共2,1兲共1 ⫺ 兲 2
g ⫽ E ⫹ V共10 pH⫺14 ⫺ 10 ⫺pH兲 (16)
⫹ 3y共1,2兲共1 ⫺ 兲 2 ⫹ y共0,3兲 3 (9)
Therefore, the g values for a certain pH can be
where the constants of y(3,0), y(2,1), y(1,2), and calculated from eq 16.
y(0,3) were ⫺0.97, ⫺3.56, ⫺1.88, and ⫺4.14, re- The theoretical potentiometric titration curves
spectively, and their mean error was 0.08. It can fitted by using lines and the third-order equation
be rewritten as follows: in Figure 1 were presented in comparison with
the experimental potentiometric titration curve
log K ⫽ ⫺0.97 ⫺ 7.78 ⫹ 12.802 ⫺ 8.213 (10) in Figure 2. The (a) and (b) curves in Figure 2
were obtained by using the theoretical equation
The straight line [Fig. 1(b)] was obtained by the for the (a) and (b) lines in Figure 1, respectively.
first-order equation to minimize the parameters. As Figure 2 shows, it was found that the poten-
In order to obtain the straight line from eq 7, tiometric titration curves obtained from the third-
there needs to be at least two parameters. This order equation are more coincident with the po-
can be written as follows: tentiometric titration curves obtained from exper-
imental data, than those from the linear
log K ⫽ y共1,0兲共1 ⫺ 兲 ⫹ y共0,1兲 (11) equation. Even though the fit resulting from the
third-order equation showed a better result, the
where the constants of y(1,0) and y(0,1) were expression using one parameter was adopted be-
⫺1.29 and ⫺3.95, respectively, and their mean cause of its sufficient accuracy and simplicity.
error was 0.17. It can be rewritten as follows: Using the third-order equation can certainly im-
prove the accuracy in describing the characteris-
log K ⫽ ⫺1.29 ⫺ 2.66 (12) tic behavior of ion exchangers, but it leads to more
CALCULATION ON ION EXCHANGE CAPACITY 3185
Figure 2. Potentiometric titration curves computed Figure 3. A plot of pK versus for various concen-
from different approximations and experimental data. trations of NaCl.
pK 0 ⫽ 冕
0
1
pKdx (19)
Determination of Ion Exchange Capacity
The ion exchange capacity can be obtained only by
potentiometric titration as a position of the inflec-
This linear approximation is equal to the arith- tion point. The inflection point of the potentiomet-
metic average and can be rewritten as follows: ric titration curve for an ion exchanger is changed
by the concentration of supporting electrolytes,
1 the volume of equilibrium solution, and the
pK 0 ⫽ 2 共pK ⫽1 ⫹ pK ⫽0兲 ⫽ pK ⫽1/2 (20)
charge of titration ions. An error in determination
of the ion exchange capacity from the titration
The change of pK for the various concentrations of
NaCl and BaCl2 at specific is shown in Figure 5.
The CM dependency of pK can be expressed by the
following eq:
Figure 7. Potentiometric titration curves for phos- Figure 9. Curve fittings of a plot of ⭸pH/⭸g versus g
phoric acid ion exchanger PI-4 in various concentra- by Lorenzian distribution for NaCl and BaCl2 solutions
tions of BaCl2. The data points were experimental data. (0.002 M).
3188 SON, KIM, AND KIM
Table II. Parameters of Lorenzian Equation [⭸pH/ rials Research Center for Better Environment at Tae-
⭸g ⫽ f( g)] jeon National University of Technology.