NX 2.0.8600 / 3.0.8600
Type 4406/303
► Document History
► Referenced Documents
Document Title
DD+DIS055.11E NX 2.0.8600 / 3.0.8600 Service Manual
DD+DIS053.11M NX 2.0.8600 / 3.0.8600 Spare Parts List
DD+DIS012.06E Licensing Service Documentation
DD+DIS124.09E ID Scanner Service Manual
Edition 1, Revision 2
01-2012 printed in Germany Document Node ID: 32772656
sw_05_software-options_e_template_v05
Agfa Company Confidential Copyright © 2012 Agfa HealthCare N.V.
DD+DIS055.11E Software Options - Clean Installation Procedure
► Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D – 81539 München
Germany
Copyright 2012 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
1. Read the “Generic Safety Directions” document
(see the Agfa HealthCare Library => General Info => Agfa HealthCare =>
Publications => Service Manual) prior to attempting any operation, repair or
maintenance task on the equipment.
2. Strictly observe all safety directions within the “Generic Safety Directions” and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type ‘Order List’ in the Agfa HealthCare Library.
LIST OF CONTENTS
3.3.3 Adjusting BIOS Settings for HP 8000 Elite and 8200 Elite PCs (Windows 7) ........................15
3.8.2 Configuring the wireless Network Connection for the wireless ICU Workflow .......................36
5.2 Activation.................................................................................................................................44
8.5 For DX-M Mammography Systems only: Checking whether IP Gain Calibration Files are
present ....................................................................................................................................54
* Reference value. Required time depends on the complexity of the situation on site.
2 Preparation
2.1 Downloading / ordering the Clean Install Kit (if not available yet)
NOTE:
Starting with NX 2.0 staged on DELL Optiplex 755 the Clean Install Material is
delivered with the system.
NOTE
The NX 2.0.8600 / 3.0.8600 Service Manual chapter 3 Installation Planning, lists the
contents of the Starterkit DVDs.
In addition to the clean installation kit (mentioned in section 2.1 of this document)
following software is required for a clean installation procedure:
# Software Reference
1 Service updates or patches 2.2.3.1
2 HP Restore DVD (for Windows 7 only) 2.2.3.2
3 TWAIN digitizer software (if applicable) 2.2.3.3
4 Latest Hotfix 2.2.3.4
5 Configuration XML File 2.2.3.5
(1) Check whether service updates or patches are available on the Agfa Healthcare
Library for NX 2.0.8600 / NX 3.0.8600 (Filter setting: Computed Radiography >
CR Workstation Software > NX 2.0.8600 - 3.0.8600).
Status 10-2011: A batch file is available to set a dependency from the NX
Workstation service to Windows Management Instrumentation.
(2) Download the available service updates / patches.
NOTE:
In general it is possible to run NX 2.0.8600 / 3.0.8600 on Windows XP or Windows7 if
the PC supports both operating systems. Refer to CR/DR Interoperability Matrix,
Document ID 31333326 (Intranet Link / Extranet Link).
If the system was delivered with Windows 7, the HP Windows 7 Restore DVD was part
of delivery.
(3) Open the FTP Access document with information how to access the FTP server.
The document is available on the Agfa Healthcare Library, Document ID
32772588 (Intranet Link / Extranet Link).
(4) Connect to the FTP server.
(5) Download the PC model dependent HP Restore DVD* and burn it on DVD.
*depending on the PC model the restore DVD set comprises 1 or 2 DVDs.
(1) To find the compatible version number(s) refer to CR/DR Interoperability Matrix,
Document ID 31333326 (Intranet Link / Extranet Link).
(2) Go to the Agfa HealthCare Library to download the device software packages:
Computed Radiography CR Digitizers <Digitizer name> Software
During Clean Installation a Hotfix version is installed on the system that may not be the
latest available Hotfix version.
Run the FUSE updater on the Service PC to receive the latest available Hotfix version.
NOTE:
For instructions to obtain and install the Fast Update Service (FUSE) software refer to
the Service Manual FUSE - Fast Update Service (Automatic Software Update)
DD+DIS134.08E (Intranet Link / Extranet Link).
For external partners: Please contact your local Agfa representative.
To simplify the configuration of the NX workstation either use a backup of the previous
configuration or organize the default configuration XML file of the NX workstation of the
region or country.
IMPORTANT:
When performing a clean installation, the Application License File (ALF) should
normally still be available and can be used for the same software version after the
clean install.
This is not the case if the ALF got lost after a system crash. In this case fetch it from
the Electronic License Management System (ELMS).
For a detailed description of the procedures to obtain a License File (ALF) refer to the
Licensing Service Manual (DD+DIS012.06E) on the Agfa HealthCare Library
(Intranet Link / Extranet Link).
3.1 Prerequisites
ACTION:
Check the following items before starting the installation:
o NX Starterkit DVD 1
o NX Starterkit DVD 2
The system can be connected to a larger network like the hospital or company
network. In this case, if other NX systems are installed at the same time, an IP
conflict can occur.
REQUIRED TIME:
Approximately 10 minutes
NOTE:
These steps are only required at systems where Windows XP will be installed if RAID
was already activated before on the installed hard disks, even if RAID was manually
deactivated before .
If the installed hard disks have not been using RAID before, continue with section
3.3 "Adjust BIOS Settings".
(1) Press <CTRL-i> during system boot, when the screen is displaying the Intel
Matrix Storage Manager RAID overview.
On a HP PC it is necessary to first press <Enter> on the message:
Press any key for option ROM messages
(2) Choose option 3 to Reset Disks to Non-RAID in the Main Menu:
Figure 1
(3) Select both Member Disks with <Space> and press <Enter> to complete.
Figure 2
NOTE:
After changing the BIOS settings on HP desktop PCs, an additional screen is displayed
that informs about updated configuration options. Confirm this by pressing <F1>.
NOTE:
Be aware that the installation has to be cancelled in the next step!
Figure 3
(11) Access the CD-ROM drive by typing into the DOS prompt: K:\ or I:\
Remark: Depending on the SATA (Serial Advanced Technology Attachment)
emulation the CD-ROM will now be available on either K: or I:
(12) Type: Gdisk.exe 2 /i /DEL /ALL /Y
Remark: This will erase all parititons on the 2nd harddrive.
(13) Continue with section 3.4 Installing the Operating System (OS).
REQUIRED TIME:
Approximately 10 minutes
# Software Reference
1 BIOS settings on DELL PCs (Windows XP) 3.3.1
2 BIOS settings on HP PCs (Windows XP) 3.3.2
3 BIOS settings on HP PCs (Windows 7) 3.3.3
(6) For multiple core CPUs (e.g. Dell D830, Dell Optiplex 755 and Dell Optiplex 960):
Activate Multiple CPU core:
Set Multiple CPU core to <ON>.
NOTE:
Also check if a SCSI hard disk is installed. If yes, additional
configurations need to be performed later in section 3.7.
NOTE:
On HP 8530p and HP 8540p laptops the CD/DVD drive is referred to as "Notebook
Upgrade bay".
NOTE:
Also check if a SCSI hard disk is installed. If yes, additional configurations need to be
performed later in section 3.7.
(4) For HP 8000 Elite: Set Power > OS Power Management > Idle Power Savings to
Normal.
(5) Go to File Save Changes & Exit, press <Enter> and accept with <F10>.
NOTE:
After changing the BIOS settings on HP desktop PCs, an additional screen is displayed
that informs about updated configuration options.
Confirm the changes by pressing <F1>.
3.3.3 Adjusting BIOS Settings for HP 8000 Elite and 8200 Elite PCs (Windows 7)
NOTE:
At the HP 6200 Pro PC (Windows 7) and HP EliteBook 8560p notebook (Windows 7)
the BIOS settings do not have to be changed.
Figure 4
(4) Use the Up () and Down () keys for navigation. Use the Left () and right ()
arrow key for changes.
(5) For HP 8000 Elite and 8200 Elite: Set Power > OS Power Management > Idle
Power Savings to Normal.
(6) Select <File --> Save Changes and Exit>
(7) Again press <F10> during bootup.
Figure 5
(9) Ensure the disks are configured in RAID1 as shown in following figure:
Figure 6
(10) If required select <1. Create RAID Volume> and make following settings:
Leave default name for the RAID volume (may differ from Name shown in
screenshot above, can also be e.g. Volume0)
Select RAID Level: RAID1 (Mirror)
Accept all further default settings
(11) Ensure that both drives are member of the mirrored volume (In screenshot
above: Both drives are of type Member Disk(0).
(12) Select 4. Exit
NOTE:
Recommendation: Have the Acceptance Test Protocol document at hand and fill it in at
every installation and configuration step instead of filling it in afterwards.
The document is available in NX 2.0.8600 / 3.0.8600 Service Manual chapter 6
Acceptance Test.
IMPORTANT:
Installing the Operating System leads to a removal of all available data on the complete
hard disk, including all existing partitions!
The installation procedure depends on the operating system which has to be installed:
REQUIRED TIME:
Approximately 60 minutes
Figure 7
(5) Confirm the installation once more by pressing <Y>. Be aware that as soon as
<Y> is pressed, the (re-)installation is starting, and all data will be erased.
+-------------------------------------------------------------+
! AGFA NX on XP !
! Multi Language !
+-------------------------------------------------------------+
! !
! Installation will erase all current systems !
! !
! (Y) This machine will be totally reinstalled !
! (N) Exit to Operating system !
+-------------------------------------------------------------+
Figure 8
(6) After approximately 5 minutes, the RAID BIOS setting instruction is displayed.
Remove the DVD from the drive and press <CTRL+ALT+DEL> to reboot.
Figure 9
IMPORTANT:
Perform the next three steps only on an HP DC7800, HP DC7900, HP 8000 and
HP8530p / HP8540p system. On DELL systems continue with step (10).
(7) Enter the BIOS setup by pressing <F10> during booting.
(8) Set Storage Options/SATA emulation to: RAID
Exception:
On HP Elitebook 8540p: Set “SATA device mode” to: AHCI
On HP EliteBook 8530p: Do not activate the storage option RAID
(9) Save changes and exit the BIOS.
NOTE:
After changing the BIOS settings on HP desktop PCs, an additional screen is displayed
that informs about updated configuration options.
Confirm the changes by pressing <F1>.
IMPORTANT:
Perform the next steps (10) to (13) only on a DELL 755 or DELL 960 system. Also
perform them if installing a DELL 755 or DELL 960 system without RAID!
(10) Enter the BIOS setup by pressing <F2> during booting.
(11) Set ATA/SATA mode to: RAID ON
On DELL830: Do not modify the BIOS options.
(12) Save changes and exit the BIOS.
(13) Directly after reboot an extra screen is displayed showing automatically updated
configurations. Save the changes by pressing <F1>.
(14) Form now on the Operating System installs automatically: Wait until the login
screen appears. See also the IMPORTANT notes next page.
NOTE:
The RAID setting is applied for consistency between all hardware types and has no
effect on the functionality of a system without RAID hardware.
IMPORTANT:
On RAID systems, the installation may be interrupted by a failure of the system to boot
after an automatic reboot.
This can be solved by changing the boot order of the hard disks:
On an HP PC:
Press <F10> during booting to enter the BIOS setup.
Go to Storage and select Boot Order.
Set the other hard disk as primary boot media.
Press <F10> to accept changes.
Select Save Changes and Exit and press <F10> to accept.
On a DELL PC:
During booting press <F12> to enter the BIOS setup.
Select the other hard disk as primary boot media.
Result After changing the boot order, the installation will continue.
IMPORTANT:
Don’t click anything until the log on screen appears.
Key strokes might interrupt the installation.
The automatic installation consists of the Norton Ghost Restore Procedure, the
Windows Setup and post-installation customization.
Norton Ghost Restore Procedure:
A progress bar indicates the progress of the restore procedure. The PC will
automatically reboot afterwards.
Windows Setup:
During the installation a lot of screens of the Windows Setup are displayed.
The installation runs in the background.
Post-installation customization:
Automatic scripts run once the Windows Setup process has completed. Wait until
the log on screen appears. Do not interrupt these scripts.
REQUIRED TIME:
Approximately 60 minutes
IMPORTANT:
Each HP PC model has its own Windows 7
restore DVD (or DVD set).
Before inserting the DVD. compare the label
on the DVD with the model.
Figure 11
(7) If the DVD set comprises 2 DVDs*: Insert the 2nd DVD when prompted.
*depending on the PC model the restore DVD set comprises 1 or 2 DVDs.
(8) Wait until the Windows 7 desktop screen appears. User "crservice" is logged in
automatically.
In case the crservice login appears, enter the corresponding password to log-in.
NOTE:
The operating system is in following state now:
Language: English
Keyboard: English (US)
Time zone: UTC+1.00 (Brussels, Copenhagen, Madrid, Paris) or UTC-08:00
Pacific Time (US & Canada), depending on the PC model
User: crservice
Partitions: C: (System) with 60 GB and D: (Data) with remaining disk size
NOTE:
This step is required for Windows XP professional systems only.
REQUIRED TIME:
Approximately 5 minutes
The operating system has now been installed on the System & Configuration partition
(drive letter C:). For NX one more partition (Data & Backup partition) has to be defined
with the following parameters:
Drive letter: D:
Volume Label: DATA
Size: Complete Remaining rest of disk-space
IMPORTANT:
RAID Systems contain a second hard disk for RAID mirroring. On RAID systems
create both partitions on one disk!
NOTE:
The screenshots below are made on a two hard disk system. If the platform contains
only one hard disk, the previous OS installation step has allocated an amount of
Gbyte of the first hard disk found for the C partition. The D partition has to be made
manually.
Figure 12
Figure 14
Figure 15
(7) Leave the default maximum diskspace as partition size for the new partition click
<Next>.
212869
Figure 16 – Example
Figure 17
Figure 18
Figure 19
Result: The new partition and the remaining unpartitioned space is displayed in
the computer management window.
REQUIRED TIME:
Up to one hour, depending on the available hotfixes
(1) Install all the hotfixes from starterkit DVD2 directory XP\Hotfix or WIN7\Hotfix
(depending on the available operating system), starting with the hotfix with the
lowest revision number.
(2) Copy the latest hotfixes from the Service PC to the NX workstation.
The hotfixes are available in following directory on the service PC (if FUSE is
installed on the Service PC):
C:\Agfa\HealthCare\FUSE Updater\Patches.
If FUSE is not installed on the Service PC refer to section 2.2.3.4 for more
information about FUSE.
(3) Install the latest available hotfixes, starting with the lowest revision number.
REQUIRED TIME:
Approximately 10 minutes
NOTE:
This step is required for Windows 7 systems only.
Figure 20
Figure 21
Figure 22
(1) For Windows XP: Go to Start Settings Control Panel. If the Control
Panel is in Category View click Switch to classic view.
For Windows 7: Go to Start Control Panel. Select View by: Small icons
(3) Select the correct time zone and check the checkbox: Automatically adjust for
daylight saving changes
(4) Click <OK>.
(1) For Windows XP: In the Control Panel and choose Regional & Language
Settings.
For Windows 7: In the Control Panel and choose Region & Language.
(2) For Windows XP: Set the appropriate setting in the tabs Regional Options
and Languages.
For Windows 7: Set the appropriate setting in the tabs Formats and
Location.
(3) For Windows XP: Check the checkbox Apply all settings to the current user
account and to the default user profile.
For Windows 7: In the tab "Administrative", click on "Copy Settings" and
select checkboxes "Welcome screen and system accounts" as well as "New
user accounts".
Figure 25
NOTE:
This has only effect when users are not in a domain. For instructions to:
Change the language of single users
Change to non-standard string formats (e.g. Date & Time, figure
formats)
Configure the default language for new NX users
refer to NX 2.0.8600 / 3.0.8600 Service Manual chapter 4 Installation and
Configuration Appendices section Changing the regional settings of the
NX.
(5) For Windows XP: Go to the tab Languages and click <Details...>.
For Windows 7: Go to the tab Keyboards and Languages and click
<Change keyboard>.
(6) If the correct keyboard is not listed under Installed services, click <Add...>,
select the keyboard layout and click <OK>.
(7) Select in the Default Input Language dropdown list, the correct input language.
Figure 26
(10) Confirm that the files already installed on the hard disk can be used
(Only if a dialog is displayed asking for a decision).
(11) For Windows XP: Go to the tab Advanced and select the language to match
the language version of non-unicode programs.
For Windows 7: Go to the tab Administrative and select Change system
locale.... to match the language version of non-unicode programs with the
required language.
NOTE:
The following steps are only applicable for PCs with SCSI disk, not for SATA or IDE
disks.
Figure 27
REQUIRED TIME:
Approximately 10 minutes
NOTE:
The following actions need to be performed to configure the Windows network settings.
IMPORTANT:
If the wireless Intensive Care Unit (ICU) feature will be used on this NX workstation,
do not perform the steps described here (in section 3.8.1). Continue with section 3.8.2.
(3) For Windows XP: Right-click Local Area Connection and go to Properties.
The window Local Area Connection Properties opens.
For Windows 7: Select Connections: Local Area Connection: The Local
Area Connection Status opens. Select Properties. The window Local Area
Connection Properties opens.
Figure 30
(7) Click <OK> (on the Internet Protocol (TCP/IP) Properties dialog).
(8) Click <OK> (Windows XP) or <close> (Windows 7) to exit.
3.8.2 Configuring the wireless Network Connection for the wireless ICU Workflow
IMPORTANT:
This section is only applicable if the wireless Intensive Care Unit feature (technical
license: NX_ICU_CONFIGURATION) is used on this NX workstation.
The wireless NX must not be plugged into the Ethernet network: Only one operating
mode is possible, either wireless or fixed network.
The wireless NX must only be used for GenRad applications.
The images on the wireless NX are not available for Central Monitoring.
(1) For Windows XP: Go to: Start Settings Control Panel Network
Connections
For Windows 7: Go to Start Control Panel Network and Sharing
Center.
IMPORTANT:
Do not disable the Local Area Connection.
(2) For Windows XP: Right-click Wireless Network Connection and choose
Properties.
For Windows 7: Select Setup a new connection or network
(3) For Windows XP: Configure the wireless network connection according to
the settings of the wireless network on site.
For Windows 7: Select Set up a new network. Follow the installation wizard.
NOTE:
Gather information about network environment in the hospital (Domain or Workgroup)
or refer to the information collected during Installation Planning.
To change the hostname and or the workgroup of the NX PC, perform following steps:
(1) For Windows XP: Right-click the My Computer icon on the desktop and
select Properties.
For Windows 7: Go to Start Control Panel System. In section
Computer name, domain and workgroup settings select Change settings.
Figure 31
Figure 32
Figure 33
(8) Click <OK> (on the Computer Name Changes dialog box).
(9) Click <OK> (on the System Properties dialog box).
(10) Reboot the PC.
REQUIRED TIME:
Approximately 30 minutes (without additional Software Update)
Prerequisites:
Successful OS installation & partitioning (C and D)
NOTE:
The following actions need to be performed to install Application and Documentation.
IMPORTANT:
Before continuing, make sure the partition D (DATA) was created on disk 0 as
described in section 3.5.
Do not disconnect the network during installation of the NX software, otherwise the
installation will fail!
(5) The installation is guided by a setup wizard. Read carefully the messages and
instructions displayed in the wizard screens.
(6) Reboot the workstation after finishing the installation wizard.
(7) Log in using the crservice account.
NOTE:
The logging information of the process application installation can be found in the log
file: C:\Agfa\Healthcare\Log\ NX X.0.8600 Install.log
NOTE:
After first reboot of the system, Windows may detect the USB dongle as 'new
hardware' and the installation wizard pops up. Follow the instructions of the wizard to
install the WIBU drivers.
NOTE:
This is the moment when available NX software updates are installed. Check the NX
2.0.8600 / 3.0.8600 Service Bulletins to see if there is a SU released for NX 2.0.8600 /
3.0.8600. The most recent SU should be installed. The software is available on NX
Starterkit DVD 1 or on the Agfa HealthCare Library.
IMPORTANT:
Make sure that the ALF file to be activated, matches the version of the software which
is installed (including SU)!
(1) Select folder Service Software on Starterkit DVD1 to find the respective
installation package OR
Use the latest SU downloaded from the Agfa HealthCare Library.
(2) Stop the NX application (Shortcut: Stop NX)
(3) Run the SU installer.
(4) Restart the NX application (Shortcut: Restart NX Completely).
(1) Install additional patches. Follow the installation instructions provided with the
patch. Status 10-2011: A patch is available to set a dependency from the NX
Workstation service to Windows Management Instrumentation.
REQUIRED TIME:
Approximately 5 minutes
NOTE:
Installing the Online Help (any language) will also install a special Thai font on the NX
Workstation that is required to read the Thai Online Help. This implies that browsing
Thai Online Help directly from the CD will not work before the Online Help is actually
installed on the Workstation.
REQUIRED TIME:
Approximately 45 minutes
For the UPS installation and configuration procedure refer to NX 2.0.8600 / 3.0.8600
Service Manual chapter 4 Installation and Configuration Appendices.
REQUIRED TIME:
Approximately 30 minutes
NOTE:
This section is only applicable if a TWAIN digitizer is connected to the NX workstation:
(1) Install the digitizer software of the TWAIN digitizer (CR 30-X or DX-S or DX-G with
SW < NIM_2XXX*) as described in the installation instructions enclosed to the
software.
*Installing the DX-G Service Client on Windows 7 is not supported.
IMPORTANT:
When the configuration of the NX workstation is finished, and the digitizer is switched
on, it is possible, that the following error messages are displayed in the digitizer error
viewer on the NX workstation:
Symptom 1 Error indicating, that the registration of the serial number failed.
Reason Restore not performed: NX workstation does not recognize the serial
number of the connected device.
Solution Step 1: Switch off digitizer. Step 2: Restore the latest digitizer backup.
Step 3: Re-boot control PC. Step 4: Switch on digitizer.
Alternative solution:
Fill in site specific data (Digitizer service client menu Configuration – Site
specific data). When done, select buttons <update site specific
data> and <configure site with new EMS serial number>.
Symptom 2 Error indicating, that software versions loaded onto machine and
installed on PC are not identical.
Reason Software mismatch between digitizer software loaded on digitizer and
installed on NX workstation.
Solution Perform software upload to the digitizer as described in the instructions
in the digitizer device software package. If the software version on the
digitizer is higher than on the workstation, install the corresponding
digitizer software version on the workstation.
IMPORTANT:
First discuss with the hospital system administrator:
Whether it is allowed to install and activate FUSE (Internet connection
required).
At which date and time the weekly check for hotfixes shall be performed.
REQUIRED TIME:
Approximately 10 minutes
For instructions to obtain and install the Fast Update Service (FUSE) software refer to
the FUSE Service Manual FUSE - Fast Update Service (Automatic Software Update)
DD+DIS134.08E (Intranet Link / Extranet Link).
For external partners: Please contact your local Agfa representative.
CAUTION:
Due to wrong date & time and time zone settings license may become invalid.
Before activating a license key file, make sure the correct time zone, date and time
have been configured! For details refer to section 3.7.
REQUIRED TIME:
Approximately 10 minutes
NOTE:
If the installation wizard for the license dongle
has not appeared yet, unplug the dongle and
reconnect it. The wizard will pop up. Click
through it using the default settings.
The logging information of the Licensing
process can be found in the log file:
C:\Agfa\Healthcare\Log\LicensingService.log
Figure 34
5.2 Activation
REQUIRED TIME:
Approximately 15 minutes
IMPORTANT:
Activation will fail if the network is unplugged during activation of the NX Software.
(2) Start the NX Activation Utility using the shortcut on the desktop, or
start it via
Start Menu Agfa NX Service Install Tools.
Figure 35
Figure 36
(6) Select the language for the NX workstation and click <Next>.
(7) Check the overview of specified selections: To change options, click <Back>. To
cancel the activation, click <Cancel>. If all selections are correct, click
<Activate>. The activation starts.
NOTE:
The “Activation Helper” window opens. It compares the license with the available
compatible installers in directory C:\Agfa\Healthcare\NX\Installers\ for
additional installers. Some examples:
Micro Calcification Enhancement (MCE)
X-Ray Generator and DR Detector Interfacing Software (XRDI)
Sedecal Software Console (DX-D 300 U-arm)
(8) If one or more compatible installers are found, select them one after another and
click on Continue Activation: The corresponding installation wizard starts.
Figure 37
(9) If an installation wizard for compatible installers starts, click through the screens
until it is finished.
(10) Click the Close button when the message appears “Click the close
button below to continue with the Activation”.
NOTE:
The Activation Utility Helper reports following warnings when applicable:
Activation was performed while there was a version mismatch for the base
license.
Activation was performed while there was no base license loaded. (this is only
possible when using lab or demo dongle)
License found for <SW>, but no compatible installer found !
License found for <SW>, but no compatible installer selected by user !
<SW Installer> returns code <ExitCode> (if installation was NOT successful)
(11) Click <Next> after the activation has completed. This may take several minutes.
(12) Choose Reboot now and click <Finish>.
NOTE:
The logging information of the process NX activation can be found in the log file:
C:\Agfa\Healthcare\Log\Nx\NXactivate.log
NOTE:
After Activation, the NX workstation will be configured to use the default string formats
(e.g. date and time) for the selected language, and the selected language will be the
default language for new users.
For instructions how to tune the local string formats (e.g. date and time) or how to
change a user's language refer to NX 2.0.8600 / 3.0.8600 Service Manual chapter 4
Installation and Configuration Appendices.
REQUIRED TIME:
Approximately 5 minutes
(1) To identify and retrieve the required monitor software refer to the “CR/DR
Interoperability Matrix”, Document ID 31333326 (Intranet Link / Extranet Link)
(2) For instructions to install the software and calibrating the monitors, refer to
NX 2.0.8600 / 3.0.8600 Service Manual chapter 5 Monitor Options.
REQUIRED TIME:
Approximately 10 minutes
IMPORTANT:
If the ID tablet firmware version is 1.12.07 or higher this step is not required.
If a mobile ID Scanner is used with this NX workstation, updating the ID Tablet
firmware is not required.
For instructions refer to NX 2.0.8600 / 3.0.8600 Service Manual chapter 4 Installation
and Configuration Appendices section ‘Upgrading ID tablet firmware’. This section also
describes how to check the ID tablet firmware version.
REQUIRED TIME:
Approximately 10 minutes
IMPORTANT:
This section is only applicable if the ID Scanner is used in combination with this NX
workstation.
For detailed instructions refer to the ID Scanner Service Manual chapter 1 Controls,
Connections, and Setup Procedures DD+DIS124.09E (Intranet Link / Extranet Link).
A short overview of the instructions is:
(1) Install the ID Scanner drivers on the NX Workstation.
(2) Adjust the COM Port Properties.
REQUIRED TIME:
Approximately 5 – 15 minutes depending on the amount of configuration tasks
8 Completing Installation
NOTE:
The Altiris Client is a piece of software which supports software deployment in
production. It is mandatory to remove the Altiris Client as the hospital might also be
using Altiris for software deployment.
REQUIRED TIME:
Approximately 5 minutes.
Figure 38
Figure 39
Figure 40
IMPORTANT:
This step is required for Windows XP systems only. For Windows 7 systems the
mirroring already was done automatically during the installation of the operating
system.
RAID (Redundant Array of Independent Disks) is supported ONLY on hardware
that has been ORIGINALLY delivered as RAID system and equipped with a
second hard disk.
For specification of hardware types supporting RAID refer to NX 2.0.8600 /
3.0.8600 Service Manual chapter 3 Installation Planning.
It is not allowed to install RAID on systems not originally delivered as RAID
system, even if the hardware is compatible with RAID.
REQUIRED TIME:
Approximately 130 minutes (waiting time included)
Figure 41
Figure 42
R208364
Figure 44
Figure 45
(12) This step is required only, if the “Matrix Storage Manager setup wizard” does not
start automatically: Go to the folder created in step (9), e.g. C:\DRV\R173416
and run the Intel Matrix Storage Manager setup.exe..
(13) Confirm all screens of the Intel Matrix Storage Manager setup wizard.
Result: The PC will be automatically rebooted:
Figure 46: First dialog of Setup Wizard Figure 47: Last dialog of Setup Wizard
(16) Select: Protect data from a hard drive failure with RAID 1:
Figure 48
Figure 49
Figure 50
IMPORTANT:
Do not perform a reboot of the system until the migration process has finished
completely! Apart from that the migration process does not interfere with other actions
to be carried out on the PC.
For procedures on how to configure the supported Anti-virus software refer to Service
Manual chapter 4 Installation and Configuration Appendices.
REQUIRED TIME:
Approximately 10 minutes
After installation and after any change to the configuration of NX a backup is required:
(1) Open the Configuration Tool and load the active configuration.
(2) In the Configuration Tool menu select: File Save as…
(3) Choose a portable medium to save the configuration file on and store this file on a
safe place. Name the file that it can be identified later on: e.g.
<Site>_initial_configuration_<date>.xml
Keep the file outside the NX hard disk to have it available in case of a defect of
the NX hard disk.
(4) Run the script ‘Backup all data’:
Agfa NX Service Backup Restore Backup all data
For more information on backup scripts, see NX 2.0.8600 / 3.0.8600 Service
Manual chapter 4 Installation and Configuration Appendices .
NOTE:
Mammography detectors delivered as of Q III / 2011 have an IP Gain Calibration CD
ROM included. This CD ROM is loaded to the NX system in order to use the detector in
the system. On a clean installed NX workstation the information from the IP Gain
Calibration CD ROM(s) is not available anymore.
In a system with two or more NX workstations, where at least one additional workstation
is switched on, the IP Gain Calibration Files are automatically transmitted to the clean
installed NX workstation when the Mammography detector is entered in the ID tablet.
REQUIRED TIME:
Approximately 10 minutes
(1) Ensure that all NX workstations in the network are switched on and the NX
application is running.
(2) Enter each Mammography detector in the ID tablet.
(3) Select menu: Main Menu > Read & Initialize Cassette > Read
(4) Check parameter IP Gain Calibration:
If it is NO: No further action required
If it is YES, the IP Gain Calibration file for this detector is transferred now
automatically from another workstation.
NOTE:
If the IP Gain Calibration File is not available on the network (e.g. no other NX
workstation available) or cannot be transferred out of any reason (e.g. other NX
workstations switched off) this is indicated by error message:
“MESS446: Error, no valid image plate gain calibration file found"
IMPORTANT:
If the IP Gain Calibration CD ROM is lost, proceed as follows:
(1) Read the “Plate ID” in the read Plate ID on RF Tag Chip Card
cassette menu (Parameter
“Identification”) or on the RF Tag
Chip Card. Example: 302 - 1349TY0001 - 20102102
Figure 51
(2) Download the IP Gain Calibration file for this “Plate ID” from following Agfa
Intranet path: HealthCare ==> Services Info ==> Services Back Up Info
REQUIRED TIME:
Approximately 30 minutes
(1) Check if all required fields of the Acceptance Test (see NX 2.0.8600 / 3.0.8600
Service Manual chapter 6 Acceptance Test) have been filled out.
(2) Hand the filled-in acceptance test over to the customer.
(3) Refer to the respective Solution Manual (CR Mammography or DR Solution
Manual) for an additional Acceptance Test on Solution level.