Professional Documents
Culture Documents
In order for the test to be quantitative, the samples must be protected from
air to prevent the re-aeration as the dissolved oxygen level diminishes. Also,
since oxygen has limited solubility in water (9 mg/L @ 20 degrees Celsius ), it is
necessary to dilute the wastewater samples as the test is performed.
OBJECTIVES
The objective of this lab is to learn how to perform a BOD5 test of an influent
wastewater collected from the Northwest Mesa Wastewater Treatment Plant.
MATERIALS
Nutrient Solutions:
1. Phosphate buffer
2. Magnesium sulphate solution
3. Calcium chloride solution
4. Ferric Chloride solution
5. Ammonium chloride solution
PROCEDURE
1. Each BOD bottle was filled by slowly adding sufficient dilution water
with nutrient so that the stopper can be inserted without leaving an air
bubble but not so much that there is overflow
2. In addition to the sample bottles, triplicate bottles were completely
filled with dilution water to be incubated as blanks.
3. Each bottle was labelled carefully, identifying the sample and volume
used.
WW volume vs DO final
8
7
6
5
DO Final
4
3
2
1
0
2 2 3 3 4 4 5 5 7.5 7.5 10 12.5 12.5 15 15
WW volume(mL)
Calculation:
Sample calculation
Sample volume =2 mL
Initial DO =8.2
Final DO=4.1
DISCUSSION
From this experiment, the BOD of the wastewater sample analyzed was
determined to be 333.2 mg/L for the solution diluted with influent.
In the test, the observations made did not verify with the theory of the
dissolved oxygen’s depletion over a 5day period. In fact, it was observed that
more titrant was necessary after solutions sat for 5 days, for the influent
samples. As seen from the % depletion of our blank sample of dilution water &
influent sample, we see that oxygen concentration actually increases. This
suggests that an error was present in our experiment. Such possible sources of
error may have risen from the improper sealing of the crucibles, which would
have allowed for air to enter the samples. This causes error in the data since
outside sources of oxygen might contribute to reaeration in the solutions.
Other possible sources of error would include the loss of dissolved oxygen
during transfer from its package to the BOD bottles.
It was also observed that the stoppers used in this experiment were not
perfectly aligned to fit to the BOD bottles used, causing for an inevitable
bubble to be present in the solution. This might have allowed for air to enter
the solution, hence allowing for outside oxygen to add to the system. To
improve this error, one should have used extra precaution in closing the
system. Such measures of preventing this error include:
a) Sealing the flask with water and leaving the bottles in a closed
fumehood or oven (set at 20 degrees Celsius) for the duration of the 7-
day period.
b) Making sure the bottle stopper is aligned properly or using a vacuum
to leave the systems (bottles) in.
It should also be noted that the sample BOD concentration is not completely
precisely accurate in comparison to the United states wastewater standard of
BOD level. This is because the sample that was analyzed was left under a 7-day
incubation period, where as the standard is suggested to be studied after a 5-
day incubation period. For this reason, it should be noted that more time is
given for the sample to be depleted of oxygen, and thus giving a BOD result
that is likely to be .With this said, it is hard to say whether or not the results
obtained from this laboratory experiment provides sufficient data to make an
accurate conclusion for the wastewater sample analyzed.
APPENDICES