Professional Documents
Culture Documents
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
Radiography
Equipment for surveys [2.7] - (Pt A, Ch 2, Sec 2)
One or more of the following fracture detection methods may be required if
deemed necessary by the Surveyor : [2.7.1]
1. Radiography (X or γ rays)
2. Ultrasonic Test
3. Magnetic Particle Test
4. Dye Penetrant Test
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
welded fittings with outside diameter higher than 75 mm, at the Surveyor’s
discretion. [7.6.1]
For the rectification of defects on castings for crankshafts, see also [4.7.2]
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
Note 1:
- The absorption of X-rays and gamma-rays is stronger in copper based
alloys than in steel. For bronze propellers, 300 kV X rays can normally
be used up to 50 mm and Co60 gamma-rays up to 160 mm thickness.
Due to the limited thicknesses that can be radiographed as well as for
other practical reasons radiography is generally not a realistic method to
check the thickest parts of large propellers.
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
- The operating conditions for the re-test samples are to be agreed with the
Surveyor, as deemed appropriate.
- For the approval of the consumable, or for the continuation of the testing
program, the re-test samples are to produce satisfactory results. [1.7.1]
(General)
The test specimens shown in Fig 5 are to be taken for the following tests:
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
The test specimens, shown in Fig 12, are to be taken for the following tests:
The results of the tests are to comply with the requirements of [5.2.6], as
appropriate.
INTERNSHIP PROGRAM
DEPARTEMENT OF NAVAL ENGINEERING
ENGINEERING FACULTY OF HASANUDDIN UNIVERSITY
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
The test specimens, shown in Fig 14, are to be taken for the following tests:
The results of the tests are to comply with the requirements of [5.2.6], as
appropriate.
The chemical analysis of the weld metal of the second run in each test
assembly is to be supplied by the Manufacturer and is to include the content of
all significant alloying elements, if any.
The specimens shown in Fig 16 are to be taken for the following tests:
a) one transverse tensile test
b) three Charpy V-notch impact tests
c) one face and one root bend tests.
Three additional side bend tests [1.5.4] may be required by the Surveyor to be
carried out for samples welded in vertical, horizontal and overhead positions.
INTERNSHIP PROGRAM
DEPARTEMENT OF NAVAL ENGINEERING
ENGINEERING FACULTY OF HASANUDDIN UNIVERSITY
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
The weld tested shall reproduce the initial test conditions except for the
thickness. These tests revalidate the welder's qualifications for an
additional 2 years.
INTERNSHIP PROGRAM
DEPARTEMENT OF NAVAL ENGINEERING
ENGINEERING FACULTY OF HASANUDDIN UNIVERSITY
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
1. Radiography (X or γ rays)
This method uses X-rays to provide radiation to the object being tested. Later
the results of x-ray radiation are captured by the prepared film. Later, the film
will be printed to see the results of the density (light darkness) of the object
being tested. Defects in objects can be seen from changes in film density. The
radiographic examination is intended to be carried out by using X-ray. The use
of gamma-ray may be accepted provided that it is demonstrated to the Society's
satisfaction that this provides the same image quality as X-ray.
2. Ultrasonic test
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
welds or other weld joints uses the application of a low surface tension liquid at
one side of a compartment boundary or structural arrangement. If no liquid is
detected on the opposite sides of the boundaries after the expiration of a
defined period of time, this indicates tightness of the boundaries. In certain
cases, a developer solution may be painted or sprayed on the other side of the
weld to aid leak detection.
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
Problem Obtained
1. Found defect in film 5 side bottom startboardside
2. Found defect in film 3 side shell startboardside
Recommendation
1. Repair in joint 5 side bottom startboardside
2. Reshoot in joint 3 side shell startboardside
INTERNSHIP PROGRAM
DEPARTEMENT OF NAVAL ENGINEERING
ENGINEERING FACULTY OF HASANUDDIN UNIVERSITY
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
Problem obtained :
- Tank 10cp
1. Missing collar not installed
2. Not match between approval drawing with stiffner (150 x
150) on longitudinal main deck
Recommendation :
Tank 10cp
1. Install the missing collar
2. Drawing update to be as per actual on board
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
DAILY REPORT
NAME : DWIKI TIMIUR PRATAMA BAKRI DATE : February 19th, 2019
ID NUMBER : D31115502 RINA SURVEYOR : MR. LA ODE SAMRIN
Approved by,
Surveyor