Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

Determining Traces of Nitrogen Using Standard
Addition and Combustion Chemiluminescence
Analysis
Sponsored by PAC L.P. Feb 26 2014

Four ASTM standard methods (ASTM D4629, D5762, D6069 and D7184) are used to
determine the total nitrogen in naphtha range samples, and other trace nitrogen
containing liquid hydrocarbons can also be determined.

The standard methods D4629 and D5762 are performed at atmospheric pressure. ASTM
D6069 and D7184 are performed under reduced pressure and require a trace nitrogen kit
comprised of a vacuum pump, a needle valve for pressure control and a restrictor to
regulate the ozone concentration in the reaction chamber that would impact
measurement sensitivity.

The determination proceeds according to the combustion reaction in all cases as shown
in Equation 1, followed by chemiluminescence of the nitric oxide as shown in Equation 2
and its relaxation and emission of the characteristic infrared energy as shown in Equation
3.

R­N + O2 → CO2 + H2O + NO + MOx (1)

NO + O3 → NO* + O2 (2)

NO* → NO + hu (3)

The SO2 fluorescence signal is strongly quenched by water formed in the combustion
process in case of a simultaneous S detection as well as the chemiluminescence
reaction, so combustion gases are routed through a membrane drying system in order to
remove the water after the combustion reaction (1) takes place. Detection is equimolar
and linear with a dynamic range higher than 103.

Experimental Conditions

Instrumentation
®

P
Saved from URL: https://www.azom.com/article.aspx?ArticleID=10678
1
/9
Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

Figure 1 shows the AntekMultiTek® Vertical configuration with nitrogen detection and
liquid autosampler.

Figure 1. MultiTek VN with Model

Instrument Parameters
The operation parameters for the analysis are shown in the table below.

Table 1. Hardware parameters

Sample Volume (μL) 20
GFC 1­ Ar carrier (ml/min) 130
GFC 2­ Pyro O2 (ml/min) 450
GFC 3­ Ozone O2 (ml/min) 35
GFC 4­ Carrier O2 (ml/min) 25
GFC 5­ Auxiliary O2 (ml/min) 25
Furnace Temperature (ºC) 1050
Nitrogen PMT voltage (V) 700
If the system configuration does not have a vacuum system, the scope of the analysis
corresponds to D4629. At lower pressures very low trace levels can be reported as 0
mg/Kg if a calibration like the one shown in Figure 2 is done.

Calibration Curve
Based on ASTM D718, calibration was achieved with toluene standards with acridine as

P
Saved from URL: https://www.azom.com/article.aspx?ArticleID=10678
2
/9
Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

the nitrogen source at atmospheric pressure. Figure 2 shows the results.

Figure 2. N calibration at reduced pressure

Figure 3. Standard addition calibration

P
Saved from URL: https://www.azom.com/article.aspx?ArticleID=10678
3
/9
Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

The standard addition method was used to study two hydrodesulfurizednaphthas, one
reforming feed, one reformate and one hydrocarbon solvent. This sample set is
characterized by having very low amounts of nitrogen falling below the LOQ of D4629 in
the majority of cases. When samples were studied against the calibration curve of Figure
2, results were reported as zero mg/Kg of N.

Standard addition comprises spiking replicates with equal volumes of sample with a few
µL of a high concentration stock solution to obtain a calibration similar to Figure 3.

The non­spiked sample is known as the blank or zero concentration as analysis
proceeds.

The method was developed by spiking 10 mL of the sample with 5 and 10µL of a 1000
ng/µL stock solution of acridine in toluene to have 0, 0.5 and 1.0 ng/µL calibration points.

Discussion
Catalytic Reforming to produce high aromatic content components is a process that
converts hydrodesulfurizednaphthas into high octane streams for gasoline blending. In
order to prevent the irreversible deactivation of costly platinum containing catalysts, these
feeds should also contain the lowest possible nitrogen.

These feedstock normally contain part per trillion nitrogen levels so the quality control of
these materials is very challenging and the analytical instrumentation and method of
analysis should be stable, sensitive, stable and consistent enough to prevent serious
operational issues.

Figure 4 and 5 shows a satisfactory application of the standard addition analysis in a
MultiTek® VN to quality control these charges before processing. The method was also
applied to determine pure aromatic streams used in petrochemical plants.

P
Saved from URL: https://www.azom.com/article.aspx?ArticleID=10678
4
/9
Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

Figure 4. Hydrodesulfurized or Sweet Naphtha

Figure 5. Naphtha charge to reforming

It was determined that using ASTM D4629 in the MultiTek® VN without the Trace
Nitrogen vacuum kit, one can retrieve the values obtained in the standard addition
method for those values above the limit of detection calculated for that system.

P
Saved from URL: https://www.azom.com/article.aspx?ArticleID=10678
5
/9
Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

Equivalent results for the analysis of N in a reforming feed naphtha are shown in Figure
6.The plot on the left was obtained in a MultiTek® VN without the Trace N vacuum kit and
the one in the right was obtained in a VN instrument at reduced pressure.

Figure 6. Comparison of Standard Addition analysis a) without the vacuum pump. b) with
vacuum pump

Conclusion
The MultiTek® Analyzer shows the ability to precisely determine very low levels of
nitrogen even when the system is not configured with a vacuum pump. Standard addition
methods are highly suitable to very low concentration nitrogen samples and easy to set
up in a MultiTek® system offering a good alternative to determining trace level nitrogen.
Results show high stability with or without the vacuum system.

This is probably the only commercial instrument that combines sulfur, nitrogen, and
halides analysis all in one. Compact, powerful, automated, and can analyse gas, liquid,
or solid samples, it is the perfect solution to today's increasing demand worldwide for
quick, precise detection and analysis of contaminants, and corrosive elements.

Since MultiTek® offers precise results with high sensitivity and unmatched versatility, it's a
valuable process optimization tool that will offer faster ROI and a better bottom line.

P
Saved from URL: https://www.azom.com/article.aspx?ArticleID=10678
6
/9
Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

This information has been sourced, reviewed and adapted from materials provided by
PAC L.P.

For more information on this source, please visit PAC L.P.

P
Saved from URL: https://www.azom.com/article.aspx?ArticleID=10678
7
/9
Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

PAC L.P.

Address

8824 Fallbrook Drive
Houston
TX, 77064
United States

Phone: +1 (281) 940 1803

Fax: +1 (281) 580 0719

Email: sales. usa@ paclp. com

Visit Website

PAC is a leading global manufacturer of advanced analytical instruments for laboratories
and online process
applications in industries such as petrochemical, refinery, upstream, biofuels,
environmental, food & beverage, lubricants, automotive and pharmaceutical.

With a product portfolio of over 200 testing instruments, PAC serves its customers with
cutting­edge technology and significant research and development resources to support
its core technologies: chromatography, elemental analysis, physical properties, fuels
composition, and laboratory automation.

Our corporate strength is built upon the quality of our product lines, our powerful
technology team, and our commitment to customer support.

We're an industry leader in standards development, and we work tirelessly to establish
benchmark procedures that scientists and analysts around the world rely on every day to
P
Saved from URL: https://www.azom.com/article.aspx?ArticleID=10678
8
/9
Determining Traces of Nitrogen Using Standard Addition and Combustion Chemiluminescence Analysis

benchmark procedures that scientists and analysts around the world rely on every day to
build better products and a cleaner environment.

PAC operates as a unit of Roper Technologies, Inc., a diversified technology company
and a constituent of
S&P 500, Fortune 1000, and Russell 1000 indices.

PRODUCT LINES

PAC has combined the world’s most respected and long­established brands of analytical
and testing equipment into a single manufacturing, marketing and service organization.
Each of our brands have long histories of developing best­in­class analytical
instrumentation for lab and process applications. In fact, the dynamic synergy of the PAC
team and its unique technologies has led to the development of unique, cutting­edge
instruments.

In close cooperation with various standards organizations throughout the world, PAC
introduces innovative instruments & applications which adhere to various standards by
ASTM, CEN, DIN, GPA, IP, ISO, and UOP. PAC also complies with ISO 9001­2000
standards.

Primary Activity
provider of advanced analytical instruments