Professional Documents
Culture Documents
ADS
www.malvern.com Bohlin
ADS
Malvern Instruments makes every effort to ensure that this document is correct. However, due to
Malvern Instruments’ policy of continual product development we are unable to guarantee the
accuracy of this, or any other document after the date of publication. We therefore disclaim all
liability for any changes, errors or omissions after the date of publication. No reproduction or
transmission of any part of this publication is allowed without the express written permission of
Malvern Instruments Ltd.
Head office:
Printed in England
MAN0377-1.0 ADS User Guide.book Page i Thursday, March 8, 2007 2:42 PM
Table of Contents
Part 1 - Operator’s Guide
Hardware features
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Temperature Control Units (TCUs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Initial procedures
Checking the air supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Removing the immobilizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Levelling the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Installing the measuring systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Checking the fluid level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Checking the flow rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Powering on the components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Temperature calibration/verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Making a measurement
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
The Asphalt wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Loading the sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Making a manual measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Interpreting the results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Closing down the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Security
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
The User Configuration dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Setting up a user group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Creating users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Part 3 - Appendices
Specification
TCUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Minimum computer specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Chemical compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Test specifications and criteria. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Site requirements
Bench . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Space required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Power requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Operating environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
Air bearing supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
Cooling air/water supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
Fluids circulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
Air supply filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-6
Unpacking instructions
Installation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
Moving the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
Upgrading the computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
Changing the computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
Regulatory statements
CE Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-2
Index
Part 1 -
Operator’s Guide
MAN0377-1.0 ADS User Guide.book Page 2 Thursday, March 8, 2007 2:42 PM
MAN0377-1.0 ADS User Guide.book Page 1 Thursday, March 8, 2007 2:42 PM
Introduction
to this manual
Introduction
This manual covers the operation and maintenance of the Bohlin ADS (Automatic
Dynamic Shear) rheometer. In simple terms this is a dynamic shear rheometer
used with a software package for measurement of rheological properties. The
instrument is especially suited to analysing asphalt and meets the AASHTO stand-
ards.
The aims of this manual are to:
Explain what the ADS is and how it works.
Describe the ADS hardware components.
Describe the software and the tests it can run.
Explain how to set up the instrument before making a measurement.
Explain how to use the instrument to make a measurement, and how to inter-
pret the results.
Identify the user maintenance and troubleshooting procedures.
Show a supervisor how to set up users.
Explain all Health and Safety issues.
Part 3 - Appendices
Appendix A contains essential Health and Safety information which all users must
read. Appendices B and C give the specification and site requirements. Other
appendices contain supplementary information not necessary for the general opera-
tion of the system.
Malvern personnel
Malvern personnel (service engineers, representatives, etc.) have full access to the
instrument and are the only people authorised to perform service procedures that
may require the removal of the covers.
Warning!
Removal of the covers by unauthorised personnel will invalidate the war-
ranty of the instrument.
Supervisor
The supervisor is the person responsible for the management and safety of the
instrument and its operation. The supervisor is responsible for training the opera-
tors. The supervisor can perform all user maintenance routines identified in Chap-
ter 7.
Operator
An operator is a person trained in the use of the system. The operator can perform
all user maintenance routines identified in Chapter 7.
Warning!
Under no circumstances should the supervisor or an operator remove the
main cover of the instrument. Failure to follow these guidelines could
result in exposure to hazardous voltages.
Assumed information
Naming convention
The ADS is referred to either in full as the ADS, or as “the instrument”.
The combination of the ADS instrument, the computer and the software is
referred to as “the system”.
Menu commands
Menu commands in the software are always shown in bold text in this form:
main menu-menu item-submenu item.
For example, the command Tools-Options refers to selecting the Options
command from the Tools menu.
Help desk
Direct all queries about the system to the local Malvern representative initially.
Please quote the following information:
Model and serial number of the instrument (given on the back panel).
The software version (to find this, select Help-About in the software).
In the USA telephone 508-480-0200 x 350 or email support@malvern.com.
Contact the United Kingdom help desk if the local Malvern representative is not
available. Its direct line is +44 (0) 1684 891800.
i
Note
This help line is primarily English speaking.
Remote support
Malvern Instruments offers a remote support service, delivered by an Internet con-
nection. Benefits include fast and efficient fault diagnosis, reducing downtime and
costs.
On-line user training is also available, plus software updates. A high speed Internet
connection is recommended for making use of this facility.
Available configurations
The main options are:
Instrument – CVO-100-ADS, CVOR-200-ADS and Gemini-ADS.
Motor controller – the instrument can be configured as analog or digital.
Torque – the CV0-100-ADS is 100mNm, the C-VOR-200-ADS is 200mNm
and the Gemini-ADS is generally 150mNm, but can be 200mNm.
Gap setting
The instrument can automatically change the required gap setting as demanded by
testing specifications. The trim gap and testing gap are pre-programmed for each
test. The user can re-zero the gap using the front panel or the software.
Once zeroed, the instrument remembers the zero position so this does not need to
be reset. The instrument can be configured to correct automatically for changes
due to thermal expansion/contraction.
Software
The ADS combines with a computer running the Bohlin Windows software. Two
software packages are available:
TruGrade – the standard package for SuperPave Specification and grade
determination testing.
Research – a range of research grade software makes the unit into a full
research grade instrument.
Note
The ADS cannot be used without the computer connected.
TruGrade tests
Tests are categorised by type and mode. These are described briefly here and in
detail in Chapter 4.
Test types
There are three test types, each of which has a different application within the
AASHTO standard:
Original Binder (OB) – this is for testing a binder which has not had any
treatment.
Rolling Thin Film Oven (RTFO) – this is used for samples which have been
through an RTFO, simulating the aging that occurs during mix production and
pavement construction.
Pressure Aging Vessel (PAV) – this is used for samples which have been
through a PAV to simulate real-world aging conditions of the asphalt binder.
The PAV exposes the binder to the high pressures and temperatures that would
occur naturally during a pavement's life. Use 8mm plates for this type of test
because the aged sample is very stiff.
Test modes
There are four test modes:
Grade Determination and Pass /Fail – these are the common test modes.
Linearity – this is only run for qualifying other results.
The fourth test mode combines Grade Determination and Linearity.
Tests available
Taking the above combinations of test mode and type, the list of tests available to
an operator is as follows:
Linearity
Original binder
Original Grade Determination
PAV binder
PAV Grade Determination
RTFO binder
RTFO Grade Determination
Research software
The following tests are available if the research software is used:
Oscillation as a function of time, temperature, frequency and shear stress.
Viscosity as a function of shear rate, shear stress, temperature and time.
Creep and Creep Recovery (for compliance measurement and analysis).
Time Temperature Superposition and Viscoelastic Analysis.
These are described in detail in the online help, not in this manual.
Hardware features
Introduction
This chapter describes the hardware components of the instrument. It covers:
Main components – describes the main parts of the rheometer.
The control panel.
Connections, which are on the rear and base of the instrument.
Temperature Control Units (TCUs) – the Total Immersion Cell and Peltier
Thermal Chamber.
Main components
This diagram shows the main parts of the ADS:
4
5
6
ADS 3
8 5
6
7
8
ill 7716
The components are:
Caution!
Do not use the vertical position knob unless told to do so by a Malvern
representative or the help desk.
Vertical position knob – this adjusts the vertical position of the bearing for
servicing of the unit. It provides a manual override for moving the stage. (The
Gemini-ADS does not have this, though an optional handle is supplied.)
Control panel – this provides a manual override of all the software controls
for moving, gapping or zeroing the stage. There is a Normal Force indicator
which displays any axial load felt by the upper plate. This panel is described in
detail below.
Bearing – air bearing.
Chuck lock – the upper measuring system shaft has two flats which this chuck
lock, fitted to the lower end plate of the bearing tube, slides into. This locks the
spindle in place.
Caution!
Do not handle the chuck or rotate the motor without the air pressure
applied; it may damage the bearing.
Chuck – the end of the spindle is a high precision collet which is tightened and
loosened by hand. The chuck holds the upper measuring device (plate, cone,
etc.). It is used on all units and is backwards-compatible.
Caution!
During transport, even over short distances, the chuck must be secured to
prevent damage to the air bearing from vibration. For details on how to
secure the chuck, consult the Malvern Help Desk.
Base plates
The base plate fits below the lower measuring system. The base plates are inter-
changeable. The lower plates are keyed and set into the Total Immersion Cell
where they rest on the base plate’s outer edge and are secured with a locking ring.
Note
If a Peltier Thermal Chamber is used the base plates are different, featur-
ing an inbuilt PT100. This is described later in this chapter.
The base plates for the 25mm and 8mm lower plates differ as shown below:
Caution!
To ensure proper fit and operation, take care not to drop or scratch the
plate surfaces.
Control panel
The control panel on the front of the ADS looks like this:
ill 7713
The controls and indicators are as follows:
NORMAL FORCE display – displays the direction and normal (axial) force
acting on the bearing. Bars lighting up to the left of centre show a negative
normal force (i.e. pulling down on the measuring system). Bars to the right of
centre indicate a positive normal force (pushing up on the measuring system).
OVLD (Overload) LED – lights up when the normal force is outside its nor-
mal operating range (i.e. too large in either direction). When in the overload
state, the motor will not move up or down.
GAP LED – this is unlit in normal use. When a user presses the GAP button
the LED illuminates. See the description of the GAP button below.
GAP SIZE display – shows the gap between upper and lower measurement
systems in microns. When the OK LED is lit, the reading is of the actual gap
between the parallel plates. (It does not display above 9mm.) If the OK LED is
off, the reading shows the required gap. To set this gap press the GAP button.
GAP button – set the gap either by using this or the software. The button has
three modes of operation:
Mode 1 Set Gap – to set the gap press the GAP button until the OK LED
is off; using the S and T buttons, adjust the display to read the desired
gap. Once the desired gap is selected, press GAP until the OK LED ena-
bles. The gap is now set and can be activated by pressing the T button, at
which time the stage will move to the gap setting.
Mode 2 Live Stage Control – after the GAP button is held down for five
seconds, the OK button starts to blink. The S and T buttons can be used
to directly move the stage to any gap. Press the GAP button again to send
the gap measurement to the computer.
Mode 3 Gap Capture – this is the same as Mode 2 without using the S
and T buttons. It captures the current gap and sends it to the computer.
ZERO button – press this for a second to zero the instrument, bringing the
upper and lower measurement systems into contact.
SPEED display – shows the current speed of the upper measurement system
movement up or down in the lower stage of gap setting. The more bars lit, the
higher the speed.
Use the SLOW and FAST buttons to reduce or increase the speed. The speed
of the stage movement can be controlled during sample loading and unloading
while within the 9mm measuring range of the gap control.
Connections
This section describes the connections on the rheometer back panel. (The connec-
tions to TCUs are described in the following section.)
This diagram shows the connections on the back of the ADS:
7 6 5 4 3 1
0
l
3 2
ill 7717
Set these up as follows:
BEARING SUPPLY (only on an ADS with air bearing) – provides the air
supply for the bearing’s torque motor. Connect the blue 4mm air hose to this
inlet and switch the red open/closed valve on the air filter regulator to the open
position.
Caution!
Air must be 3 bar and must meet the specification given in Appendix B.
An alarm will sound if the air pressure falls below 2 bar.
GAP CONTROL – this is for the serial cable connected to a serial COM port
on the computer. Data relating to the stage, gap and temperature are communi-
cated via this connection. The digital fluid circulator is also controlled through
this cable.
INSTRUMENT PORT – connect one end of the cable marked computer
here and its 9 pin D-type connector to an available COM port on the compu-
ter. Note the COM port number, as this will be required by the software later.
Power cables
Connect the power cable to the socket on the back of the ADS and plug it into a
power outlet but do not switch it on yet.
Fuse
The fuse provides electrical protection for the rheometer. Chapter 7 gives details
of how to change fuses.
ill 7894
The Total Immersion Cell is EasySwap™ mounted in the base of the rheometer
housing. The fluid connections used for the chamber are under the unit (see
below). The white plastic cover, when fully seated, allows the water to cover both
the upper and lower plates.
Note
Ambient temperature measurements can be made without fluid circulating
in the chamber.
This graphic shows the connections where fluid from the circulator enters and
leaves the Total Immersion Cell:
2
2
ill 7895
The inlet is in the centre of the cell. There are three outlets . The hoses from
the three outlets are connected by a 3:1 “banjo” connector to form a single hose
inflow to the circulator. The connection procedure is described below.
Fluids circulator
The digital fluids circulator is from the Julabo F25-HE or FP35-HE series. When
configured and operated properly, this holds a temperature accurate to ±0.01°C.
Caution!
Be sure to provide at least 20cm of space in front of the vents for air to cir-
culate through the circulator. The circulator must not be installed inside
an enclosed space or near to a heat source or direct sunlight.
Making connections
There are three steps:
Connect the 9 pin RS232 cable from the control computer to the circulator
port marked SERIAL.
Connect the PT100 cable from the rheometer’s PT100 port to the circulator
port marked ext Pt100.
Connect the supplied power cable to the rear of the circulator and plug it into a
power socket.
Connect the 8mm tubing from the Total Immersion Cell ports to these, as follows:
Ensure that the outlet from the circulator is connected to the inlet in the centre
of the Total Immersion Cell.
Connect the three outlets from the Total Immersion Cell to the 3-in-1 “banjo”
back to the circulator inflow.
Caution!
Make sure no liquid enters the interior of the circulator. If the fluid level is
not maintained the heating and cooling coils may be exposed and become
damaged. Keep extra water available.
5 6 1 2 3 4
ill 7896
Its components are:
Upper heat exchanger thermal enclosure – the upper chamber slides up
for easy access to the lower plate for sample loading, trimming, cleaning and
plate changes.
Lower plate with embedded PT100 – there will be one of these with an
8mm plate and another with a 25mm plate.
Lower Peltier assembly – fits into the rheometer below the lower measure-
ment system, via the EasySwap™ base.
Heat exchanger cooling kit – this bottle contains the circulating fluid, which
is circulated by a small pump. It can simply be placed on the bench near the
rheometer.
Tubing and fittings.
The Peltier plate itself has two electrical connections:
PT100 (round plug) – connects to the PT100 port on the back of the ADS.
This is hard-wired to the base plate which holds the lower measuring system.
D type – connects to the Peltier Drive connection on the back of the Peltier
electronics box (shown below). This is not a removable connection at the Pel-
tier plate end; it’s hard-wired.
3 4
2 5
6
ill 7897
The connections, described in the following section, are:
PT100 connector – connect this to the TCU socket on the ADS.
PELTIER DRIVE connector – connect the serial cable from the Peltier Ther-
mal Chamber to this.
Fan – do not obstruct this.
On/Off switch.
Fuse – the fuse can be changed by the user.
Connections
This diagram shows the connections between the Peltier Thermal Chamber, the
electronics box and the rheometer back panel:
3 1
ill 7718
Software
features and tests
Introduction
This chapter describes:
The TruGrade software and the main differences between the supervisor and
operator levels of access. For full details on the TruGrade software, use its
online help. To open this, select Help-Contents or click the Help button on
any dialog.
The TruGrade tests which can be run; the test types and modes.
Note
The Bohlin research software is documented independently in its own
online help.
Logging on
Start the software by clicking the TruGrade icon on the desktop. This displays the
User Login dialog:
The supervisor may set up password changing; the software will remind users to
change their password if necessary.
TruGrade software
The TruGrade software is the standard package for SuperPave specification and
grade determination testing. It has two access levels:
Supervisor – gives access to all facilities. This includes setting up operator
users and their permissions, designing tests, modifying the Asphalt wizard,
etc.
Operator – gives access to a subset of the facilities. This allows the user to run
a test from the test list, but not to modify a test. Depending on their permis-
sions, an operator may be able to run calibration or validation. Tailoring the
user interface to specific operators makes testing easy for them.
Supervisor level
After initial installation only supervisor level access to the TruGrade software is
available. At this stage the supervisor should create the first user accounts.
At supervisor level the software has two main modes, selected using the list on the
toolbar. These are:
Test List View – this shows the available tests in its File Name list. The
result of a test appears on the right-hand side:
Test Parameter View – in this mode the supervisor can modify the test
parameters, using the parameter area of the window, shown below. This
replaces the File Name panel shown in the above dialog:
Operator level
The operator only has access to Test List View. This allows them to select a test
from the list the supervisor has set up. In this example the user has selected
Original Binder:
TruGrade tests
TruGrade tests are categorised by type and mode.
Test types
There are three test types, each of which has a different application within the
AASHTO standard:
Original Binder (OB) – this is used for the binder without any treatment.
Rolling Thin Film Oven (RTFO) – this is used for samples which have been
through an RTFO, simulating the aging that occurs during mix production and
pavement construction.
Pressure Aging Vessel (PAV) – this is used for samples which have been
through a PAV to simulate real world aging of the asphalt binder. The PAV
exposes the binder to the high pressures and temperatures that would occur
naturally during its life. Use 8mm plates for this type of test because the aged
sample is very stiff.
Test modes
The test modes are summarised here then described in detail below.
The Grade Determination and Pass /Fail modes are the common tests. Linear-
ity is only run for qualifying other results. The fourth test mode combines Grade
Determination and Linearity.
Pass/Fail test
This mode runs a single point test at the current set temperature. The pass/fail cri-
teria shown below are applied to the result. For a PAV sample, for example, a point
with G*.sin(delta) of 4.8MPa would have a Pass result.
OB – G*./sin(delta) >= 1kPa
RTFO – G*/sin(delta) >= 2.2kPa
PAV – G*.sin(delta) <= 5MPa
The default parameters for this test are given in the next section.
The Pass/Fail test generates just one point. It shows all the results from that point:
12.0
404.40
The data can also be displayed as a graph by selecting the Graph tab.
Linearity test
This test runs a series of single test points at one temperature, but across a range of
increasing amplitudes. This is called an Amplitude Sweep. On completion, a
linearity analysis is run on the data to generate a pass/fail result to indicate whether
the sample meets the linearity requirements of the standard.
Test parameters
The default parameters for Grade Determination and Pass /Fail modes for the
three test types are:
Thermal
Test equilibrium Initial Measuring
type time (mins) Strain stress system Gap
OB 10 12% 100Pa PP25 1.0mm
RTFO 10 10% 100Pa PP25 1.0mm
PAV 10 2% 5000Pa PP8 2.0mm
Temperatures
As specified by AASHTO T-315, the following temperatures are used for tests:
OB and RTFO – 46, 52, 58, 64, 70, 76, 82 or 88°C.
PAV – 40, 37, 34, 31, 28, 25, 22, 19, 16, 13, 10, 7 or 4°C.
Tests available
Taking the above combinations of test mode and type, the list of tests available to
an operator is as follows:
Linearity
Original binder
Original Grade Determination
PAV binder
PAV Grade Determination
RTFO binder
RTFO Grade Determination
Initial procedures
This chapter describes how to prepare the rheometer for use at the start of the day.
It covers the following:
Checking the air supply.
Removing the immobilizer.
Levelling the instrument.
Installing the measuring systems.
Checking the fluid level in the circulator bath.
Checking the flow rate.
Powering on the components.
Temperature calibration/verification – this needs to be carried out in line with
company procedures.
1 3
Caution!
Never use the supplied wrench for tightening. It is only for removing
plates.
i
Note
Filters should be large enough to adequately cover the drain holes. Avoid
using too small a filter as this may get sucked into the drain hole.
i
Note
Filters should be large enough to adequately cover the drain holes. Avoid
using too small a filter as this may get sucked into the drain hole.
Aim for a flow rate high enough to immerse the sample in the fluid and low
enough that the fluid does not overflow.
Temperature calibration/verification
This section describes the steps necessary for an operator with sufficient access to
verify/calibrate the rheometer using a TCD-05 meter. Each procedure runs as a
wizard, working through calibration/verification points defined by the supervisor.
i
Note
If routinely having to adjust the temperature more than two or three tenths
of a degree, or unable to hold a temperature, contact Malvern Instruments
for further in-depth calibration. (Please read the Troubleshooting sec-
tion in Chapter 7 first.)
2 4
ill 7751
The components are:
Power adapter.
Temperature meter.
Temperature sensor.
Modes of operation
There are two modes of TCD-05 operation:
Battery operation – the meter works using a standard 9V PP9 battery (pro-
vided). When running from the battery, the meter will work for up to 20 hours
on a single battery. The meter will automatically switch off if no buttons are
pressed for more than about 20 minutes. The meter will not automatically
shut down when connected to the software, so we recommend using the mains
adapter for unattended automatic calibrations with the software.
Power adapter operation – this is the recommended method of using the
meter with the software calibration procedures.
7. Read the two hex codes from the calibration certificate for the sensor.
8. Adjust the values in the first set of hex digits by clicking the S button to step
the value through the sequence from 0 - 9, A - F and back to 0. Continue until
the first digit is correct then click the T button.
The next digit will now flash. Change its value in the same way. Continue
until the four digits match the first code on the certificate.
9. Click the Menu key to accept the value. The second set of digits appears with a
small number 2 shown to the left of the code.
10. Enter the second code in the same way as described in step 8 for the first code.
11. Click the Menu key to accept the second four digit value.
12. Click the ESC key again to return to normal sensor readings.
13. The meter is now 'tuned' to the sensor.
14. From now on, when the meter is switched on, the meter should show the two
hex codes as being the calibration constants used for the sensor. This will clear
after a few seconds and the meter will then show the sensor readings.
This dialog controls the four main configuration options for calibration and verifi-
cation:
Calibration points.
Verification points – used to test the calibration.
Instrument temperature control.
Calibration device – the device to be used.
Calibration points
These are the points at which a calibrated measurement is taken. The results from
these are used to calibrate for all temperatures. Set the initial value to an accessible
temperature lower than the first required calibration point. This ensures that all
calibrations work in an ascending direction; we recommend that users follow this
rule.
Set up the list of calibration points using the Calibration Points temperature grid.
Points can be added to the end of the list with the Add button and removed with
the Delete button. Clear All removes all temperatures from the list.
Verification points
These are the points at which a calibrated measurement is verified. Each verifica-
tion point consists of a temperature and a tolerance. The results and the report
show pass or fail depending on whether the verified points are within the set toler-
ances. Set the tolerances to give a reasonable definition of whether a calibration is
suitable or not, typically 0.2°C.
As with calibration points, an initial value is required. This must be less than or
equal to the first verification point.
Verification points are shown in the points grid. Use the Add button to add new
points to the end of the list. Remove them using the Delete button. Clear All
removes all points from the list. Tolerances can be entered manually or alterna-
tively a fixed absolute or percentage value can be applied to all points in the list
using the Set Absolute and Set Percentage buttons in conjunction with the
Quick Set Tolerance value.
Parameter Function
Tolerance Defines the range of temperatures permitted while taking a
point. If the temperature measured by the instrument is outside
this range, the calibration or verification of the point will restart.
The permitted range of temperatures is defined as Target
Temperature ± Tolerance.
Equilibrium Defines the period of time (in seconds) to allow to ensure that
Buffer the temperature controller has stabilised and for the probe
connected to the external calibration meter to reach a state of
thermal equilibrium.
Calibration device
Malvern Instruments can provide a calibrated device that has been integrated with
the software application to provide an automated solution. Alternatively, any cali-
bration device can theoretically be used to take the calibrated temperature measure-
ments. This means, however, that the operator must manually type the calibrated
measurements into the software application when prompted to do so.
Note
To use a non-Malvern device, set the Calibration Measurement
Options parameter to Manual. This changes the behaviour of the calibra-
tion and verification wizard to prompt for measurements.
To use the Malvern Instruments integrated device, set the Calibration Measure-
ment Options parameter to Automatic. This enables the following options that
the user must configure to identify the location of the device and to define equilib-
rium criteria for taking a measurement:
Parameter Function
Device Type Set this to TCD-05.
Stability Buffer Defines stability of measurements from the TCD-05 device. A
measurement is taken if all measurements taken over the buffer
time are within the Stability Tolerance (see below). The
resulting measurement value is then an average of all
measurements during the buffer time.
COM Port Defines the serial port the TCD-05 device is connected to.
Stability Defines the stability of measurements from the TCD-05 device.
Tolerance A measurement is taken if all measurements within the stability
buffer time are within ± the stability tolerance of the midpoint of
the total range of measurements during this time.
Note
Changing these options does not affect the current calibration or verifica-
tion until the next time the Calibration/Verification wizard is run.
Software configuration
Total Immersion Cell
For this TCU set the parameters in the software up as follows:
Field Value
Calibration Measurement Options
Manual/automatic Automatic
Device type TCD-05
COM Port Select the correct COM port for the device
connection
Stability buffer * 600 seconds
Stability tolerance 0.1°C
Temperature control options
Tolerance 0.1°C
Equilibrium buffer * 300 seconds
Calibration Points
Initial Value 2.0°C (should be lower than 1st Cal Point)
Temperature Table See the following table
Verification Points
Initial Value 2.0°C (should be lower than 1st verification Point)
Temperature table See the following table
Note
Do not decrease in value the settings in the table marked with *.
Increasing the values will make the calibration/verification times longer.
Note
Do not decrease in value the setting in the table marked with *.
Increasing the value will make the calibration/verification times longer.
Recommended temperatures
Calibrate the specific temperatures regularly used.
The way the calibration software operates, each programmed temperature is
uniquely calibrated and the software interpolates between these calibrated tempera-
tures for any other temperature selected.
This table shows the default calibration and verification temperatures. The initial
value for each is shown in bold text:
Verification (°C)
Calibration (°C) (Initial value in bold)
Intermediate 13, 16, 19, 22, 25, 28, 31 10, 22, 34
High 46, 52, 58, 64, 70, 76, 82, 88 46, 52, 64, 76
Full 13, 16, 19, 22, 25, 28, 31, 10, 22, 34, 46, 52, 64, 76
46, 52, 58, 64, 70, 76, 82, 88
Calibration procedure
This procedure uses the standard Bohlin software not the TruGrade software.
To load the sensor:
1. Fit the upper Parallel Plate, PP25 DSR. Initialise and zero the instrument then
set a gap of approximately 4000 microns.
2. Set a temperature of 90°C in the Manual Settings box in the CREEP/
CREEP RECOVERY page of the main software:
ill 7899
5. Close the gap to secure the sensor between the upper and lower plates.
6. Check that the sensor and mains power are plugged into the TCD-05 and the
communication cable is plugged into the correct serial port on the computer.
8. Enable the Auto Tension check box in the Manual Settings section and set
it to 40%. This keeps a constant force on the probe throughout the calibration/
verification, compensating for thermal expansion.
9. Select Supervisor-Calibration-Options. Check that all the information is
entered from the table previously discussed, then click on OK.
10. Select Supervisor-Calibration-Calibration. This starts the wizard as shown
below; follow the on-screen instructions:
12. The wizard approaches the Initial Value set in the table (2.0°C) then contin-
ues to the first calibration point. Once within the tolerance band (± 0.1°C), it
holds for 600 seconds.
13. After taking the readings at the first point, it moves on to the next:
14. After all the calibration points have been taken, the instrument cools down to
the Initial Value temperature set in the Verification Points grid.
15. The wizard continues, verifying at the points selected.
Note
During the wizard’s Verification stage, the temperature shown on the main
page differs from that shown in the wizard. This is due to the correction
applied during the Calibration/Verification. Upon completion this will be
applied to the main temperature reading.
17. The Status row gives the pass or fail information. To view or print a report,
click the Report button.
18. Select Finish to complete the calibration/verification process.
To perform a verification only:
1. Select Supervisor-Temperature Calibration-Verification.
2. The procedure will follow the same sequence as for calibration but will only
verify the temperatures set in Verification Points on the Options page.
Note
When carrying out further verifications, the probe must be placed in the
same orientation as previously used.
Making a
measurement
Introduction
This chapter shows how to measure a sample. It assumes that the initial procedures
described in the previous chapter have been carried out.
This section covers:
Using the Asphalt wizard to make a measurement.
Loading the sample – this is a detailed description of this stage. Users may want
to read this at the appropriate stage in the Asphalt wizard section.
Making a manual measurement (i.e. without using the Asphalt wizard).
Interpreting the measurement results.
Closing down the system.
3. These were covered by the previous chapter so should not require any action.
Click Next to display the Sample dialog:
4. Type in the sample information, enough for other people to track the sample in
the future. The sample reference and test number must be entered before the
test begins. More general notes can be entered while the test is running.
Lock the air bearing assembly by sliding the chuck locking plate from the right
to the left as shown in the picture. The shaft has two flats which fit the slot on
the lock and these hold the shaft in place, stopping it turning. It may be neces-
sary to rotate the motor to align the flats on the shaft to fit within the locking
plate to prevent it rotating.
6. Click Next to display the Geometry dialog:
Set the appropriate plate or spindle in the Geometry list (PP = Parallel Plate):
PP 25mm (diameter) for Original Binder or RTFO aged material.
PP 8mm (diameter) for PAV aged material.
Check that the geometry in the ADS is the one shown in the dialog. If the
plates need zeroing, click the Zero button to do this.
7. Click Next to display the Load dialog:
Load sample by any approved method (the following section details these). The
wizard graphic shows use of a silicone mold for upper plate sample transfer.
8. After loading sample on one of the plates, click Next to display the Trim dia-
log:
The wizard automatically lowers the stage to the appropriate sample trim gap
for trimming the sample. Note the following:
For best results, use a heated tool for sample trimming. The tool should be
hot enough to cut through the sample but not so hot as to cause it to melt
away or burn from the plates. Excessive heat will impair the results.
Trim the edge for a clean uniform surface in one or two swipes. Avoid
excessive trimming, for it will remove too much sample, degrading the
quality of the results.
When trimming a high temperature sample, the tool does not need to be
heated unless the sample is modified material. If it is modified, heat the
tool until it’s just warm enough to cut away material.
For modified or PAV-aged material at intermediate temperatures, over-
heating the trimming tool can cause too much material to escape.
9. Once the sample is correctly trimmed, click Next to display the Gap dialog
(for PAV-aged material this dialog will show 2000 not 1000):
11. Click Next to start the test. Any notes for the test will be displayed in this dia-
log:
Note
Operators can enter comments in the Notes field or data in predefined
fields at any time while the test is running, or after the test completes.
12. When the test completes, save the results. Press the Save button on the toolbar
or use the File-Save command. The Save As dialog appears:
Choose the correct location and file name (if unsure, ask a supervisor about
naming conventions) then click the Save button.
13. Clean the plates as described in Chapter 7 before looking at the results.
Note
Always follow the industry’s Best Practices guidelines for proper testing
and handling of asphalt samples.
ill 7900
Inspect the sample in the mold for air bubbles or lines which indicate that
it is not homogenous. For most materials, one minute of cooling at room
temperature is sufficient for de-molding; check the sample by pulling
down on the corners of the mold as shown below:
ill 7901
If the sample does not release easily from the mold, put it in a refrigerator
for one to two minutes and try again.
Place the sample on the upper or lower plate by touching the center of the
plate with the sample, centering it, and slowly peeling back the silicone
mold. If the sample is not centered, use the corner of the mold to move it.
4. Lower the upper plate, slowly squeezing the sample between the two plates
until it reaches the gap setting (1.05mm or 2.05mm). Excess material must be
squeezed out between the plates.
5. Trim the excess material with a tool at least 4mm wide Here are some trim-
ming tips:
Make or purchase a good trimming tool, something like this:
ill 7902
When trimming a sample in the high temperature range, do not overheat
the tool. Heat it just enough to be able to cut away material.
When testing modified or PAV-aged material in the intermediate tempera-
ture range, overheating the trimming tool can cause too much material to
leak out from between the plates, ruining the sample.
On the other hand, not heating the tool enough can cause the binder mate-
rial to be pulled out from between the plates. Knowing the proper amount
of heating comes with experience.
i
Note
It is best to heat the trim tool on a temperature controlled hot plate set to a
suitable temperature.
ill 7921
8. Lower and secure the white plastic cover in place, allowing the water to cover
both plates.
9. Release the shaft lock by sliding it from left to right.
i
Note
Supervisors defining tests should use the online help for information on
the parameters to use, etc.
3. Select the test to run. Additional notes can be added to the Notes field during
the test.
4. Click the Start Measure button to run the test.
5. Two commands are available to control the test:
Test Control-Bypass current wait – to skip a temperature stage. This is
the same as pressing the Bypass button.
Test Control-Abort test – to terminate the test.
6. When the test completes, save the results. Press the Save button on the toolbar
or use the File-Save command. The Save As dialog appears:
Choose the correct location and file name (if unsure, ask the supervisor about
naming conventions) then click the Save button.
7. Look at the results for the test type as described below.
Pass/fail test
The Pass/Fail Results tab shows the result details, like this:
The grade is reported, along with the pass/fail temperature (the interpolated tem-
perature at which G*/Sin (Delta) crosses the limit value).
The Grade Determination graph page shows the pass/fail limit (the horizontal
line) and the data, like this:
The line and data points in the failure area below the last pass point will be col-
oured red.
Linearity test
The linearity calculation is used to check that over a set range of strain values the
data falls between the appropriate limits. The results are presented in table and
graph form. This example shows the graph:
Auto report
A report summarising the test data is available. Its data will be appropriate for the
test method used. To preview the report, press the Print Preview button on
the toolbar. The report format is shown below:
Caution!
Follow this procedure closely. Turning the air supply off too soon may
cause damage to the rheometer.
Maintenance and
Troubleshooting
Introduction
This chapter describes:
Setting up the fluids circulator.
Rheometer maintenance.
Circulator maintenance.
Changing fuses.
Troubleshooting – this section lists possible problems and their solutions.
Note
The ADS cannot be used without the computer connected.
Note
Before operating the circulator after transport, wait about one hour after
setting it up. This will allow any fluid that has accumulated laterally during
transport to flow back down, ensuring maximum cooling performance.
Caution!
Check that the line voltage matches the supply voltage specified on the
identification plate.
2. Use a 19mm (¾") wrench to unscrew the collar nuts on the pump connectors
labelled (the input) and (the output) and remove the sealing disks. Screw
the fittings back on.
3. Connect the tubing or hose from the circulator to the ADS (circulator output
to ADS input and vice versa). The three outlets from the ADS are fed into a
“banjo” which forms a single input back into the circulator.
4. Connect the mains power cable from the circulator’s control head to the mains
power outlet on the rear of the bath. Connect the mains power cable from the
rear of the bath to a grounded power outlet.
5. Connect the five pin control cable between the connectors. Secure it by screw-
ing until tight.
6. Connect one end of the PT100 cable to the output labelled ext Pt100 on the
rear of the control head (the top part of the circulator). Connect the other end
to the PT100 connector on the ADS.
7. Connect the circulator side of the GAP cable to the RS232 port on the back of
the circulator’s control head.
8. Fill the circulator with the fluid. (The recommended maximum filling level
with water as the bath liquid is 25mm below the tank rim.)
9. Ensure the water flow control valve on the instrument or line in is closed to
prevent overflow when the circulator is switched on.
Rheometer maintenance
To ensure accurate measurements, the instrument must be properly maintained.
Caution!
Samples left in place may attack the surfaces of the measuring systems.
Warning!
Take care near the hot plate; use proper tools to handle the heated base
plate.
7. Let the base plate cool before rinsing it with solvent, followed by a final rinsing
with acetone. Dry it with a clean paper towel or cloth.
8. Clean the upper plate by placing a paper towel on the hot plate and gently drag-
ging the plate across the paper towel until clean. Avoid scratching the plate.
9. Let the upper plate cool then rinse it with solvent, followed by a final rinsing
with acetone. Dry it with a clean paper towel or cloth. (Adjust the temperature
of the hot plate to avoid igniting the paper towel.)
Caution!
Do not contaminate the circulator bath fluid with solvents.
Filters
The ADS has both fluid and air filters.
Caution!
If any material or oil at all is seen reaching the right-hand cylinder below
the Festo filter, contact Malvern immediately.
If using the additional Domnick Hunter system, proceed according to its manufac-
turer’s documentation.
Circulator maintenance
The Julabo circulator manual gives full maintenance information. This section
briefly describes the main tasks.
Warning!
Always disconnect the power cord before starting.
Take care when emptying hot bath fluids. Check the temperature of the
bath fluid prior to draining (by switching the unit on for a moment, for
example).
ill 7903
3. Slide a short piece of tube onto the drain port and hold the end over a
bucket.
Changing fuses
Warning!
Fuses must only be changed by a qualified electrician.
Rheometer fuse
If the ADS power fails this is probably a fuse problem. The fuse is located below
the power cable inlet on the back of the rheometer (see Chapter 3).
To remove the fuse, remove the mains cable and pry the fuse cover out with a
screwdriver.
The following fuse ratings apply:
Troubleshooting
This section describes how to fix possible problems.
Part 2 -
Supervisor’s Guide
MAN0377-1.0 ADS User Guide.book Page 2 Thursday, March 8, 2007 2:42 PM
MAN0377-1.0 ADS User Guide.book Page 1 Thursday, March 8, 2007 2:42 PM
Security
Introduction
This chapter shows supervisors how to set up their users’ access permissions.
8 Chapter 8 Security
Security Chapter 8
Creating users
To create a user:
1. In the User Configuration dialog select Users-Add Users to open this dia-
log:
Chapter 8 Security
Part 3 -
Appendices
MAN0377-1.0 ADS User Guide.book Page 2 Thursday, March 8, 2007 2:42 PM
MAN0377-1.0 ADS User Guide.book Page 1 Thursday, March 8, 2007 2:42 PM
Warning!
Use of the system in a manner not specified by Malvern Instruments may
impair the protection provided by the system.
Warning!
This product must be connected to a protective earth. The metal parts of
the unit are earthed using a protective earth connection.
The ADS system components are mains powered devices and all power cables and
electrical sockets should be treated accordingly. Do not place cables in positions
where they are likely to become wet.
Crush hazard
Warning!
Keep hands out of the area between the plates when closing the gap.
Hot components
Warning!
The Total Immersion Cell, the Peltier Thermal Chamber and the lower
plate may become very hot. Take care when touching these, for example
for cleaning. Wear heat proof gloves if required.
Take care with all hoses and hose connections carrying hot fluids to/from
these components.
Fluids circulator
Follow all health and safety information in the circulator manufacturer’s documen-
tation. In particular observe the following warnings:
Warning!
Do not touch hot surfaces. Take care with liquids at high temperatures.
General requirements
The requirements listed below are applicable to all countries:
The power cord must be approved by an acceptable accredited agency respon-
sible for evaluation in the country where the power cord set will be installed.
The power cord set must have a minimum current capacity of 10A (7A in Japan
only) and a nominal voltage rating of 125 or 250 volts AC, as required by each
country's power system.
The area of the wire must be a minimum of 0.75mm2 or 18AWG, and the
length of the cord must be less than 3m.
Route the power cord so it is not likely to be walked on, or pinched by items
placed upon it or against it, or become wet. Pay special attention to the plug,
the electrical outlet, and the point where the cord exits the product.
Warning!
Do not operate this product with a damaged power cord set. If the power
cord set is damaged in any manner, replace it immediately.
Warning!
Do not use the power cord received with this product on any other
products.
Warning!
To prevent electric shock, plug the instrument or accessory into correctly
earthed electrical outlets.
Warning!
Do not use adapter plugs that bypass the grounding feature, or remove the
grounding feature from the plug or adapter.
Place the AC adapter in a ventilated area, such as a desk top or on the floor.
The AC adapter may become hot during normal operation of the instrument
or accessory. Use care when handling the adapter during or immediately after
operation.
Use only the Malvern-provided AC adapter approved for use with the instru-
ment and/or accessory. Use of another AC adapter may cause a fire or explo-
sion.
Warning!
Use proper lifting techniques to avoid back injury. Always lift the instru-
ment by holding it under its base.
ill 7610
This symbol on the product or on its packaging indicates that when the last user
wishes to discard this product it must not be treated as general waste. Instead it
shall be handed over to the appropriate facility for the recovery and recycling of
electrical and electronic equipment.
By not discarding this product along with other household-type waste, the volume
of waste sent to incinerators or landfills will be reduced and natural resources will
be conserved.
For more detailed information about recycling of this product, please contact the
local city office, a waste disposal service, or a Malvern Instruments representative.
Specification
This appendix gives general information on the system and also on test specifica-
tions and criteria.
Dimensions
Peltier
Gemini ADS CVO(R) ADS Circulator control box
Height 50cm 50cm 64cm 10.5cm
Width 34cm 30cm 23cm 24cm
Depth 34cm 30cm 42cm 22cm
Weight 28kg 28kg 32kg 0.7kg
Torque
The maximum torque for the CV0-100-ADS is 100mNm, the C-VOR-200-ADS
is 200mNm and the Gemini-ADS is generally 150mNm, but can be 200mNm.
B Appendix B Specification
TCUs
Digital circulator/Total Immersion Cell
Circulator JULABO F25-HE JULABO FP35-HE
Temperature range -28°C to 100°C -28°C to 100°C
Temperature stability ± 0.01°C ± 0.01°C
Specification Appendix B
Accessories
Solids clamping system
Maximum sample width 13mm with a free sample length of 50mm.
Working temperature range -15°C to 95°C depending upon the fluid
medium used
Chemical compatibility
Components of the instrument that may come into contact with the sample are
manufactured from materials that are considered to give the widest protection from
chemical attack. However, it is important to check that any sample used is chemi-
cally compatible with the materials mentioned.
Appendix B Specification
Parameter Values
Plate Diameter Range 8-15mm diameter
Gap range 1-3mm
Applied Shear Stress or 40 to 60kPa, 0.8 to 1.2%
Nominal Strain
Frequency 1.6Hz or 10rads-1
Temperature Range 5, 10, 15, 20, 25, 30, 35, 40°C ±0.2°C
Specification Appendix B
Parameter Value
Plate Diameter Range 15-25mm diameter
Gap range 1-3mm
Applied Shear Stress or 90 to 150Pa, 9 to 15%
Nominal Strain
Frequency 1.6Hz or 10rads-1
Temperature Range 45, 55, 65, 75°C ±0.2°C
Parameter Values
Plate Diameter Range 15-25mm diameter
Gap range 1-3mm
Applied Shear Stress or 80 to 260Pa, 8 to 12%
Nominal Strain
Frequency 1.6Hz or 10rads-1
Temperature Range 45, 55, 65, 75°C ±0.2°C
Appendix B Specification
Site requirements
Malvern Instruments makes every effort to ensure that this document is correct. However,
due to Malvern Instruments’ policy of continual product development we are unable to
guarantee the accuracy of this, or any other document after the date of publication. We
therefore disclaim all liability for any changes, errors or omissions after the date of
publication. No reproduction or transmission of any part of this publication is allowed
without the express written permission of Malvern Instruments Ltd.
Head office:
Printed in England
MAN0377-1.0 ADS User Guide.book Page 3 Thursday, March 8, 2007 2:42 PM
Bench
Seat the rheometer on a solid bench which will not be unduly affected by vibration.
If such a table is not available, the rheometer could be stood on top of a heavy slab
placed on a normal table; this helps to reduce the vibrations detected by the instru-
ment. If using a Peltier Chamber TCU, place its electronics box and fluid bottle on
the bench.
If using a fluids circulator, place this on the floor.
Space required
Provide enough space to allow easy access to all components of the system. If using
the Peltier Chamber allow 100cm (width) x 70cm (depth) of bench space. Allow at
least 800mm above the bench surface for access to the ADS.
Dimensions of the main components of the system are given below:
Power requirements
The mains power supply must be clean and filtered. If necessary, fit an Uninter-
ruptible Power Supply (UPS) to remove any spikes or noise. The power require-
ment for each component of the system is given below:
Power sockets
Component Power requirement required
ADS ~ 100-240V, 50-60Hz 1
Fluids circulator ~ 100-240V, 50-60Hz 1
Peltier control box ~ 100-240V, 50-60Hz 1
Computer See manufacturer’s documentation 1
Computer monitor See manufacturer’s documentation 1
Operating environment
Altitude Up to 2000m
Ambient temperature +5 ... +40°C (for storage and transportation)
Air humidity Max. relative. humidity 80% for temperatures up to +31°C
Linear decrease down to 50% relative humidity at a tem-
perature of +40°C
Protection class IP21 (IEC 60529)
Power supply Corresponds to Class I; according to VDE 0106 T1
Not for use in an explosive atmosphere
Max. mains fluctuation of ±10% are permissible
Overvoltage category II (IEC 60664)
Pollution degree 2 (IEC 60664)
Parameter Specification
Specification [ISO Standard 8573.1, quality class 121].
Maximum size for dirt 0.1 microns
Maximum size for oil: 0.01 microns
Maximum moisture dew point -40ºC
Air delivery at 25 litres/min [ISO Standard 8573.1, quality class 121 air]
Fluids circulator
Read the circulator manufacturer’s documentation for full details.
Parameter Specification
Altitude Up to 2000m - normal zero
Ambient temperature +5°C ... +40°C (for storage and transportation)
Air humidity: Max. relative humidity 80% for temperatures up to
+31°C, linear decrease down to 50% relative
humidity at a temperature of +40°C
Protection class IP21 (IEC 60 529)
Power supply Corresponds to Class I, according to VDE 0106 T1
Overvoltage category II
Pollution degree 2
In addition:
The circulator is for indoor use only.
It is not for use in an explosive atmosphere.
Maximum mains fluctuations of ±10% are permissible.
The Total Immersion Cell design meets or exceeds industry standards for
asphalt test equipment (AASHTO TP5 and SHRP B003).
pH at 25°C 7 to 8.5
Suspended matter <30mg/l
Size of suspended matter max. 0.1mm
Growth of algae not permissible
Filter setup
The filter setup involves two or three filters as shown below:
Rheometer.
The filters in the above components are user-replaceable; refer to the manufac-
turer’s documentation for each filter.
Note
If the ETC oven is to be supplied with the same air/Nitrogen source, an
additional regulator (part 250456) can be attached to the main filter/regula-
tor.
2
3
ill 7904
Air enters the regulator at and leaves through . Open/close the air passage
using the switch . To adjust the pressure, lift the upper blue knob and turn it.
Push it down to relock it in the new position.
Caution!
If any material or oil at all is seen reaching the right-hand cylinder below
the Festo filter, contact Malvern Instruments immediately.
Unpacking
instructions
Déballez l’instrument
Desempacotamento do instrumento
3. Carefully cut the plastic packing tape along the fold lines.
Découpez avec précaution le ruban d’emballage en
plastique, le long des pliures.
Die Kunststoffverpackungsbänder vorsichtig
entlang den Falzlinien durchschneiden.
Cortar con cuidado la cinta de embalaje de plástico
por la líneas de doblado.
Com cuidado, corte a fita de embalagem plástica
ao longo das dobras.
4. Open the wings of the carton and remove the top foam insert.
Ouvrez les rabats du carton et retirez
les blocs de mousse du haut.
Die Klappen des Kartons öffnen und
die oberen Schaumeinlagen ent-
fernen.
Abrir los costados de la caja y extraer
las piezas intercaladas de espuma de la
parte superior.
Abra as abas da caixa e retire o enchi-
mento de espuma que está em cima.
5. Remove all other packages from the carton and unwrap these.
Sortez tous les autres paquets du carton et déballez-les.
Sämtliche andere Pakete aus dem Karton nehmen und auspacken.
Retirar todos los demás paquetes de la caja y desenvolverlos.
Retire todas as outras embalagens da caixa e abra-as.
Installation
Introduction
The ADS system should initially be commissioned by Malvern trained personnel.
This appendix shows how to re-install the system, typically required under the
following circumstances:
Moving the instrument
If the system has to be moved from one laboratory to another, use the informa-
tion in this appendix to reconnect the system correctly.
Upgrading the computer
It may be company policy to upgrade computers periodically. Detail is given on
the actions required.
If required to change the computer, contact Malvern Instruments.
Only install the system in a suitable environment as detailed under Environmen-
tal conditions in Appendix C.
Warning!
Use proper lifting techniques to avoid back injury. Always lift the instru-
ment by holding it under its base.
E Appendix E Installation
Warning!
This product must be connected to a protective earth.
Caution!
The CVO-ADS can be moved by users.
If moving the system a significant distance or shipping it, pack all components, if
possible in their original packaging. Pack the fluids circulator, computer and moni-
tor according to their manufacturer’s instructions.
Installation Appendix E
Note
If the original software is subsequently updated with a new version, any
custom tests, user groups and users, etc. will be preserved.
Appendix E Installation
Regulatory
statements
This appendix presents regulatory information.
CE Declaration of Conformity
The CE badge on this product signifies conformance to European Commission
Directives.
ill 7982
ill 7983
Index
A Checking air supply 5-2
Checking flow rate 5-4
AASHTO T-315 2-1 Checking fluid level 5-4
AC adapter safety A-4 Chemical compatibility B-3
Access to the instrument 1-2 Chuck lock 3-3
Accessories B-3 Chuck overview 3-3
Adjusting flow rate 3-11 Circulator fluid filters 7-6
After a test 7-5 Circulator maintenance 7-7
Air bearing Cleaning 7-5
locking 6-4 Closing down system 6-17
position 3-2 Code input 5-6
Air bearing supply C-4 Comparison Strain 4-6
Air filter Computer
replacing 7-6 changing E-3
setup C-6 specification B-3
Air supply Connecting the hoses 3-11
checking 5-2 Connections 3-7, 3-14
Annealing binder 6-9 Connections (Peltier) 3-14
Application Support tab 8-3 Constants B-1
Asphalt wizard Control panel
running 6-3 details 3-5
stages 6-2 overview 3-2
ASTM specifications 2-1 Cooling air supply C-4
Auto report 6-16 COOLING connector 3-7
Auto-cal procedure 5-7 Cooling fluid supply 3-14
B Cooling water quality C-5
Cord set safety A-3
Base plate Creating users 8-3
cleaning 7-5 CREEP RECOVERY page 5-12
description 3-3 Crush hazard A-2
Bath temperature not changing 7-10 CVO-100-ADS 2-2
Bearing 3-2 CVOR-200-ADS 2-2
BEARING SUPPLY 3-7
Bench specification C-3 D
C Dimensions B-1