You are on page 1of 22

1

Description and Dosimetric Verification of the PEREGRINE


Monte Carlo Dose Calculation System for Photon Beams
Incident on a Water Phantom

C. L. Hartmann Siantar*, R. S. Walling*, T. P. Daly*, B. Faddegon#, Paul Bergstrom*, N.


Albright#, C. Chuang#, D. Garrett*, R. K. House*, D. Jong*, D. Knapp*, S. H. May*, E.
I. Moses*, R. W. Patterson*, C. Powell* D. J. Weiczorek#, and L. J. Verhey#

*Lawrence Livermore National Laboratory


#University of California San Francisco

Abstract
PEREGRINE is a 3D Monte Carlo dose calculation system written specifically for
radiotherapy. This paper describes the implementation and overall dosimetric
accuracy of PEREGRINE physics algorithms, beam model, and beam
commissioning procedure. Particle-interaction data, tracking geometries, scoring,
variance reduction, and statistical analysis are described. The program BEAM is
used to model the treatment-independent accelerator head resulting in the
identification of primary and scattered photon sauces and an electron
contaminant source. The magnitude of the electron source is increased to improve
agreement with measurements in the buildup region in the largest fields.
Published measurements provide an estimate of backscatter on monitor chamber
response. Commissioning consists of selecting the electron beam energy,
determining the scale factor that defines dose per monitor unit, and describing
treatment-dependent beam modifiers. We compare calculations with
measurements in a water phantom for open fields, wedges, blocks, and a multi-leaf
collimator for 6 and 18 MV Varian Clinac 2100C photon beams. All calculations are
reported as dose per monitor unit: no additional, field-specific normalization is
included in comparisons with measurements. Maximum discrepancies were less
than either 2% of the maximum dose or 1 mm in isodose position for all field sizes
and beam modifiers.

calculation in radiotherapy,8 9 10 11 12 13
I. Introduction and rapidly-increasing computer
speeds it is now possible to use 3D
Dose calculation accuracy is a
Monte Carlo methods for day-to-day
critical part of radiation therapy.
treatment planning.
Combining first-principles physics
Accurate dose calculation
with physical descriptions of the
requires accurate characterization of
radiation source and patient, Monte
the radiation source. This can be
Carlo transport methods have the
accomplished with high accuracy and
potential to calculate dose accurately
exceptional detail by simulating the
over a wide variety of treatment
transport of particles through the
delivery and patient conditions.1 2 3 4 5
6 7 accelerator head and beam delivery
Owing to development of faster
system.14 15 16 17 18 19 20 Recent work
codes designed specifically for dose
2

has described accelerator investigated. We describe a set of


simulations in a two-step process:5 21 experimental comparisons for 6 and
22
initial Monte Carlo simulation of the 18 MV x-ray beams (Varian Clinac
accelerator head to produce a file 2100C accelerator) incident on a
containing the phase space for a water phantom. Open field
large ensemble of representative comparisons are made for field sizes
particles, and condensation of the ranging from 2x2 to 38x38 cm2. We
phase space file into a photon beam also compare calculations with
model, i.e., a set of probability measurements for fields modified by
distributions which can be sampled to a representative wedge, block, and
obtain particles for transport through multileaf collimator.
the treatment-dependent parts of the
beam delivery system. Simulations II. Materials and
are based on generic manufacturer
descriptions, and can include Methods
assumptions about initial electron The first part of this section
beam characteristics (energy, spot describes the implementation of
size, divergence, etc.). A Monte Carlo Monte Carlo transport methods, a
treatment planning calculation beam model, and a beam
system must account for the details commissioning procedure for the
of each individual accelerator, PEREGRINE system. In the Monte
including beam tuning, to provide Carlo code description section, we
accurate dose calculations.23 describe the particle-interaction
This work summarizes the processes (physics), transport
PEREGRINE system,Error: Reference methods, tracking geometries,
source not found and its x-ray beam scoring, variance reduction, and
model. A method of determining the statistical analysis methods used for
beam representation, or instantiation, this study.
is demonstrated for a specific The beam model consists of a set
accelerator. Studies investigating the of photon sources representing
accuracy of photon beam simulations target, flattener, and primary
fall into two general categories: collimator, and a single extended
validation that beam models electron source. Published
accurately reproduce characteristics measurements provide an estimate of
of the initial phase space generated backscatter on monitor chamber
by Monte Carlo methods,Error: response. The magnitude of the
Reference source not found 24 and electron source is increased to
experimental verification of dose improve agreement with
distributions in a phantom. The x-ray measurements in the buildup region.
beam model used in PEREGRINE has Commissioning consists of three
already been shown to faithfully steps: (1) selecting the electron beam
reconstruct the phase space.Error: energy, (2) determining the scale
Reference source not found In this factor that defines dose per monitor
paper the accuracy of its dose unit, and (3) describing treatment-
predictions for x-ray therapy are dependent beam modifiers. To select
the electron beam energy, we use
 dose calculations made from full
Some of the features described in this article are
available as a commercial product from NOMOS treatment head simulations at
Corporation, Sewickley, PA. discrete electron energies to
3

calculate off-axis ratios (OARs) for a mesh (taken from a CT scan),


38x38 cm2 field at 10 cm depth. By recording dose deposition, defined as
comparing these OARs with energy/mass of collection volume, on
measurements, we choose an a user-specified grid. At each point of
intermediate beam energy which interaction, the phase space (energy,
gives the best match. trajectory, and position) descriptions
Representations of the beam model, for secondary photon and/or electron
pre-calculated from the phase space is stored in a set of arrays, which are
data, are then interpolated to this emptied before a new source particle
energy. The calibration scale factor is is selected. This process continues
set according to the Gy-to-MU for millions of histories, until the user-
calibration condition of the specific specified stopping condition (number
accelerator. Finally, collimator jaws, of histories or statistical figure of
wedges, wedge trays, block trays, and merit) is met.
multi-leaf collimators (MLCs) are
described in terms of density, 1. Physics
composition, shape, and location.
Block thickness, material, and Photons
density, and aperture shape and MLC
PEREGRINE determines the total
leaf positions are described by the
cross section for the photon
user at calculation time.
interacting in the medium from the
In the last part of the Materials
sum of the cross sections for
and Methods section, we provide a
Compton scattering, the photoelectric
description of the measurements
effect, pair and triplet production, and
used for dosimetric comparisons
Rayleigh scattering. Total cross
shown in the Results section. These
sections are taken from the Evaluated
include accelerator, detector, and
Photon Data Library (EPDL).Error:
water phantom characteristics.
Reference source not found Compton
scattering is treated in the incoherent
A. Monte Carlo Code scattering factor approximation.25 26
Description The photoelectric effect electron is
Monte Carlo particle (history) assumed ejected from the K shell of
simulations follow a three-step the atom with a direction determined
transport process. First, a particle is from Sauter’s K shell formula.27 The
selected from the radiation source, binding energy of the photoelectron is
described as set of energy, angular, deposited at the point of interaction.
and position distributions derived Pair production cross sections
from a particle phase-space file. The include both production of pairs in the
beam model is described in more field of the nucleus and their
detail in section IIB. The particle is production in the field of the atomic
then tracked through the treatment- electrons (triplet production).Error:
specific beam delivery components Reference source not found 28 Cross
(collimator jaws, wedges, blocks, sections for these processes are
multi-leaf collimator, etc.) and air added and treated as pair production
column until it reaches the boundary in the field of the nucleus. The energy
of the Cartesian grid that defines the sharing between the electron and the
patient (patient mesh). Finally, the positron is determined by the Bethe-
particle is tracked through the patient Heitler formula. Rayleigh scattering is
4

treated in the form factor threshold bremsstrahlung processes


approximation.Error: Reference from radiative stopping power.
source not found Error: Reference source not found The implementation for sampling
knock-on events of Møller scattering
Charged particles (for electrons) and Bhabha scattering
PEREGRINE uses unrestricted (for positrons) is the same as for the
stopping powers calculated from the EGS4 code.31 The bremsstrahlung
formulas described in ICRU report cross sections and emitted photon
37.29 The stopping powers that the spectra data were obtained from the
code produces are in agreement with Lawrence Livermore National
those tabulations. The density effect Laboratory’s Evaluated Electron Data
correction to the stopping power is Library (EEDL).32 For each
calculated using a standard, pre- bremsstrahlung event, the
specified material density. If bremsstrahlung photon energy is
PEREGRINE uses the material at a subtracted from the primary electron
non-standard density in the energy. The angle of the emitted
simulation, it does not recalculate the bremsstrahlung photon is determined
density effect correction. As a from a simplification of the method
simplification, PEREGRINE uses described and evaluated elsewhere.33
34
electron total stopping powers for The primary electron is not
both electrons and positrons. This is a deflected after a bremsstrahlung
reasonable approximation because of event.
the relatively small number of
positrons in photon teletherapy 2. Transport Methods
beams and the similar values of
stopping powers for positrons and Peregrine has two different sets of
electrons (for electron energies particle tracking algorithms, one for
between 0.1 and 10 MeV, the stopping the beam delivery system and one for
powers of electrons and positrons the patient.
differ by less than 5% for elements
ranging from carbon to lead30). Photons
PEREGRINE transports electrons
In the beam delivery system,
using Class II condensed history
photons are tracked using standard
methods (see Transport section), and
analog methods: given a particle
precalculates restricted collisional
energy, location and trajectory,
and radiative stopping powers, to
calculate distance to zone boundary
avoid double-counting processes that
and next collision; move particle to
are handled on an event-by-event
the minimum of these distances; if
basis. Restricted collisional stopping
the minimum distance is to a
powers are determined as described
collision, determine energy and angle
in ICRU 37, to subtract above-
of all daughter products. In this study,
threshold Møller (electrons) and
secondary photons created below 10
Bhabha (positrons) scattering
keV were not tracked and the
processes from collisional stopping
minimum photon tracking energy was
powers. Restricted radiative stopping
100 eV.
powers are determined by subtracting
a numerical integration of the above-
In the patient, photons are
tracked through the CT scan-derived
5

transport mesh using the delta electron tracking cut off. All energies
scattering method.35 This allows stated are kinetic energy.
PEREGRINE to avoid computationally In the patient mesh, termination
expensive distance-to-boundary of the electron trajectory is
calculations without biasing the determined by its kinetic energy and
resulting dose calculations. In this its location in the geometry. The
study, the minimum photon tracking track terminates when the electron’s
energy was 100 eV. energy falls below the energy needed
to traverse one-third of the voxel’s
Charged particles minimum dimension (approximately
PEREGRINE uses class-II 180 keV kinetic energy for 1
condensed history method for millimeter voxels in water). Electrons
charged particle transport,36 are never transported below 10 keV
modeling knock-on and kinetic energy. Once the particle
bremsstrahlung processes above reaches the minimum tracking
specified cut off energies as discrete energy, its residual energy is
events. The Moliere37 multiple deposited at a random location along
scattering method is employed, a straight-line trajectory of length
implemented as in the EGS4 code. equivalent to its residual range. The
The condensed history electron step termination of a positron trajectory
size taken is determined by the results in the emission of two 511 keV
minimum of the step size necessary kinetic energy annihilation photons.
to create a bremsstrahlung photon, to
generate a knock-on electron, to 3. Geometry
reach the next spatial boundary, to Beam modifiers are described as
reach the next energy bin boundary, collections of six-sided prisms, with
or 1 mm. The energy bins used in boundaries defined by the physical
PEREGRINE are such that the dimensions of the modifier. Collimator
fractional energy loss in crossing a jaws can move along a line or arc,
bin varies between about 8% and matching their physical
20%. The electron step is typically implementation in the accelerator
broken into two substeps by treatment head. Aperture blocks are
deflecting the particle by the multiple modeled as a diverging raster of voxel
scattering angle for the full step at a prisms with lateral dimensions
random point along the step. In the defined by the user. In multileaf
patient mesh, energy is deposited at collimators, the convex leaf ends
a random location along each (Varian design) and tongue-and-groove
substep. features are modeled explicitly.
In the beam delivery system, Wedges are defined as a series of
electrons can be tracked with contiguous trapezoidal prisms. All
different levels of approximation, block and wedge trays are modeled
depending on the beam modifier. as uniform slabs of material, with
While user-defined tracking and cutoff lateral extents corresponding to their
options can be modified, we used the physical dimensions. All space not
following options for this work: 0.01 occupied by solid, beam-modifying
MeV bremsstrahlung creation material is filled with air, in which
threshold, 0.1 MeV Møller electron particles are allowed to interact. The
creation threshold, and 0.1 MeV user assigns all beam-modifier
6

materials and densities in the device volume than the cube defined by the
description input file. grid points. This causes the
The patient mesh is taken directly calculation to reach a given
from the CT scan, with no reduction statistical noise level faster than it
in resolution. The user assigns would if the dose were collected in
material to each voxel by specifying the cubes. Overlapping spheres
pre-defined materials for ranges of CT causes neighboring points to be
number. The user specifies density correlated, and results in the
from the CT number with a resolution being slightly lower than
monotonically-increasing, piecewise- the grid spacing.
continuous linear function. Each
material can also be defined with a 5. Variance reduction
default density. Several variance reduction
techniques are used in PEREGRINE.
4. Scoring They are: source particle reuse, range
Dose is scored in PEREGRINE on a rejection, and splitting/Russian
grid that is independent from the CT rouletting in the patient.
derived Cartesian grid that is used for
particle transport. This permits Source particle reuse
PEREGRINE to speed up the Each particle that survives
calculation by using variance transport through the modifiers is
reduction techniques in the regions reused a set number of times. If the
that lie outside of the dose-scoring source photon is reused so many
grid. times that the ranges of the electrons
In PEREGRINE, dose is scored in a set created by different reuses overlap, it
of spheres that are centered on the will make non-independent
points of a grid with cubical unit depositions to the same dosel. This
cells. The spheres, referred to as amounts to expending computer time
dosels, fill all of space since they just without a reduction in statistical
touch along the cube diagonals. The noise. Photons are reused upon
spheres overlap elsewhere. The entering the CT grid. Electrons are
transport algorithms in PEREGRINE reused upon entering the air column
deposit energy at a random point below the last modifier. This permits
along the electron step, as opposed the reused electrons to spread out
to depositing energy inside a given CT before entering the CT grid, which
voxel. When energy is deposited at a prevents them from making non
point in space, the energy is added to independent depositions into the
every sphere that contains that point. same dosel. The number of times
To obtain dose, the energy is divided that source particles can be reused is
by the mass of the sphere, which is also limited by the requirement that
determined in a setup calculation that the source be adequately sampled
utilizes a closed form solution for the during the course of the run.
common volume of a sphere and a
rectangular parallelepiped. Spheres Range rejection
were selected as a way of An electron with CDSA range that
approximating the dose reported at a is less than 1/3 the smallest
point for a region of non-uniform dimension of a CT voxel is terminated
density. The spheres have a larger and its energy is dumped at a random
7

point along a straight line of length For dosels with similar density,
equal to its residual range. In a setup the variance tends to be proportional
calculation, each voxel is assigned a to the dose. If the variance were
range rejection energy using this exactly proportional to the dose for
criterion. Electrons are never every dosel in the problem, then the
transported below 10 keV, regardless dosel with the maximum dose would
of the range rejection energy. also be the dosel with the maximum
standard deviation. If this were true,
Russian rouletting and splitting the termination criterion based on the
Photons that are outside the standard deviation of the watch dosel
dose scoring region and are moving would be equivalent to the
away from the dose scoring region termination criterion based on the
are Russian rouletted. If descendents standard deviation of every dosel.
of rouletted photons move towards or The standard deviation that
enter the dose scoring region, they goes with a given dose tends to
are split. The number of times that decrease as the density of the dosel
roulletting and splitting can happen to increases. For this reason, the watch
the descendents of a photon is dosel is selected from dosels that
limited to avoid the generation of very have a mass that is between one
high and very low weight particles. fourth and 3 times the mass of a
water dosel.
6. Statistical analysis
The Monte Carlo calculation is B. Source Description
considered to have converged to a The non patient specific parts of
fractional error, F, when the standard the Varian high energy family of
deviation of the dose of every dosel is accelerators (Clinacs 2100C, 2100C/D,
less than F * Md where Md is the 2300C/D) are simulated for 6 and 18
largest dose in any dosel. Calculating MV using the BEAM Monte Carlo
the standard deviation in every dosel code.Error: Reference source not
results in a significant expenditure of found The physical dimensions and
memory and time due to the large materials of the accelerator were
size of the dosel array. For this obtained from the manufacturer.*
reason, we provide a statistical figure
of merit based only on the standard 1. BEAM Simulations
deviation calculated for a single The BEAM simulation extended
dosel, which we call the watch dosel. from the top of the bremsstrahlung
The watch dosel is selected target to the bottom of the monitor
during the first part of the run as the chamber. The electron beam incident
dosel with the maximum dose after a on the target was assumed to have no
fixed number of histories, H. H is divergence, to be mono-energetic, and
chosen to be large enough that the to have a uniform spatial distribution
watch dosel will, at the end of the with 1 mm radius. The delta ray and
run, have a dose that is close to Md. bremsstrahlung production cutoffs
The run is terminated when the were taken to be AE = 0.521 MeV
standard deviation of the watch dosel (kinetic + rest mass) and AP = 0.01
is less than F * Wd, where Wd is the
dose in the watch dosel. *
Varian Oncology Systems, Palo Alto,
CA
8

MeV. The electron and photon Sub-source 4: all electrons


transport cutoffs were taken to be To calculate histograms for each
ECUT = 0.611 MeV (kinetic + rest sub-source, particles are projected in
mass) and PCUT = 0.1 MeV. We used straight lines (assume no collisions,
the variance reduction technique of no collimator jaws, etc.) from the
splitting every bremsstrahlung photon bottom of the monitor unit chamber,
into 20 photons, and employed the which is the location of the phase
PRESTA38 electron step algorithm. The space file, to the isocenter plane,
number of histories for each case was which is defined to be 100 cm from
chosen to produce phase space files the target. Rmax is calculated for
containing about 30 million particles. each sub-source, where Rmax is an
This corresponded to 20 million and 6 estimate of the maximum radius
million electrons incident on the illuminated in the isocenter plane by
target at 6MV and 18 MV respectively. that sub-source when the jaws are set
to a 40 x 40 cm2 field. This radius lies
2. PEREGRINE Beam Model in the corner of the square field.
The phase space file resulting Phase space particles that cross the
from the BEAM simulation is used to isocenter plane beyond Rmax are
generate a set of histograms that can ignored. A histogram of the
be sampled to approximately recreate rotationally-symmetric planar energy
the phase space file. These fluence in the isocenter plane is
histograms form a part of the calculated for each sub-source
PEREGRINE device file, which extending to Rmax.
characterizes the accelerator The method used to calculate Rmax
according to the beam model, for photons does not work as well for
described elsewhere.Error: Reference the contaminant electron sub-source
source not found During a since electrons do not travel in
PEREGRINE calculation, the straight lines in the air. Values of Rmax
histograms are sampled to generate for electrons are checked to ensure
the particles that are then tracked that the simulated dose in the build
through treatment-specific beam up region of water phantoms is not
modifiers (collimator jaws, wedge, appreciably affected by increasing
block, multileaf collimator, etc.) and Rmax.
the patient. Annular tiles of equal width are
Particles in the phase space defined in the isocenter plane for
file are divided into 4 sub-sources each sub-source. A set of histograms
depending on the location of their last is calculated for the particles
interaction, which is determined by belonging to each tile of each sub-
the LATCH variable in BEAM: source. These histograms are
Sub-source 1: photons for calculated under the assumption that
which the target is the last the phase space file is rotationally
interaction. symmetric. If it is not, the histograms
Sub-source 2: photons for will still be rotationally symmetric.
which the primary collimator is the Histograms describing the
last interaction. energy spectrum and the directional
Sub-source 3: photons for spectrum are calculated for each tile
which the flattening filter is the last of each sub-source. The energy and
interaction directional spectra for particles in a
9

given tile are assumed to be 1. The radial dependence of the


independent. The propagation rotationally symmetric planar
direction of a particle is specified by energy fluence in the isocenter
its intersection with two planes, the plane for each sub-source.
isocenter plane and the virtual source 2. The energy spectrum for each tile
plane, VSP. The VSP is defined as of each sub-source.
follows (see Figure 1). Vectors 3. The radial and azimuthal angle
describing the particles in each tile, parts of the planar energy fluence
when projected back along their in the VSP for each tile of each
trajectory, converge to a minimum sub-source.
radius, or bottleneck, close to the The device file also contains the
central axis of the beam near the following additional information: (1)
point of their last scattering. For the fraction of the energy contained
example, particles in the flattening in each sub-source; (2) a factor that
filter sub-source form a bottleneck converts monitor units into dose in
close to the location of the flattening water, consistent with the calibration
filter. A plane perpendicular to the of the specific accelerator being
beam axis, called the VSP, is selected simulated; (3) a description of the
close to the bottleneck. The planar beam modifiers (material, density,
energy fluence in the VSP associated and dimensions); (4) coefficients that
with particles that cross the are used to correct for back scatter
isocenter plane in a given tile is a into the monitor unit chamber as a
function of polar coordinates about function of jaw setting (this
the beam axis, namely the radius and correction is required since
the azimuthal angle (see Figure 2). PEREGRINE does not simulate back
The azimuthal angle is the angle lying scatter into the monitor unit
in the VSP between the intersection chamber).
of the particle with the VSP and the Histograms from the device file
plane containing the central axis of are sampled during a PEREGRINE
the beam and the intersection of the simulation as follows:
particle with the isocenter plane. If 1. A particle is selected from one of
one rotates the particle intersection the four sub-sources.
in the isocenter plane about the beam 2. The (x,y) location is uniformly
axis, then the corresponding sampled, with particle weight
distribution in the VSP also rotates. modified to account for planar
The planar energy fluence in the VSP energy fluence.
is assumed to be separable in radius 3. The tile is randomly selected to be
and azimuthal angle. The exact one of the two tiles with center of
location of the VSP is automatically annulus closest to (x,y). This is
adjusted to make this a good equivalent to smoothly
approximation. Radial and azimuthal interpolating the tile probability
angle histograms of the planar energy distributions with the distance
fluence in the VSP are calculated for from the axis in the isocenter
each tile. plane.
The PEREGRINE device 4. The energy of the particle is
description then contains the sampled from an energy spectrum
following histograms: of the tile.
10

5. The intersection of the particle data the best. Backscatter from the
trajectory in the virtual source upper jaws, f1, is characterized by a
plane is sampled by treating the second-order polynomial fit to
virtual source plane energy measurements made by varying the
fluence as a probability position of the upper jaw while
distribution. The trajectory of the keeping the lower jaw fixed at 40 cm.
particle is determined by Backscatter from the lower jaws is
connecting the points in the characterized by another second-
virtual source plane and the order polynomial fit, f2, to lower-jaw
isocenter plane. measurements made with the upper
Figures 3 and 4 show the jaws open to 40 cm.
magnitude and distribution of dose Backscatter Correction = f1(y) [(y
distributions resulting from individual f2(x)/40) + 1 – y/40]
subsources for a 20x20 cm2 fields where x and y are the lower and
incident on a water phantom upper jaw openings, respectively, and
positioned at at 90 cm source-to- f1 and f2 are normalized such that
surface distance (SSD). Profiles were f1(40) = f2(40) = 1.
taken at the nominal depth of Finally, comparisons with
maximum dose, dmax, (1.5 cm at 6 MV, large-field measurements reveal a
3.2 cm at 18 MV), 10, and 20 cm. The deficit in dose calculations in the
primary photon subsource provides dose-buildup region for open fields.
the largest contribution to dose, This effect, shown for a 38x38 cm2
followed by the scattered-photon field in Figure 5, is evident in depth
subsource associated with the dose curves calculated with both
flattening filter. Scattered photon BEAM and PEREGRINE. These
subsources have depth dose simulations make use of the same
characteristics similar to the primary phase space data from BEAM
photon source. Depth dose simulation of the fixed components of
distributions in Figure 4 show the the accelerator, upstream of the jaws.
electron subsource contributing Results of BEAM and PEREGRINE
significantly to the dose for 6 and 18 agree within a statistical precision of
MV beams, from the surface down less than 1% demonstrating that the
through dmax. Scattered photon and discrepancy is not due to the beam
contaminant electron subsources model or radiation transport physics
play an increasingly important role for in PEREGRINE. The magnitude of the
increasing field size. dose deficit near the surface
Because PEREGRINE does not increases with increasing field size,
account for particles scattered back and goes away for fields blocked by
into the treatment head, a wedges or trays. Based on this
measurement-based method is used evidence, we hypothesize that it is
to estimate the effect of backscatter caused by a source of electrons in the
on the over-response of the monitor accelerator head that is not fully
chamber. While backscatter factor accounted for in the treatment head
measurements have been reported by simulation with BEAM. To account for
several authors, (Reference 39, for this, we increase the weight of the
example) we used measured electron subsource by 120% and 50%
backscatter factors reported by for 6 and 18 MV beams, respectively.
Dunzenli40 were used as these fit our While the 6 MV discrepancy is
11

smaller, it requires a greater beam energy for this study. The use of
proportion of added electron source, profiles to determine beam energy
because a smaller number the source relies on accurate modeling of the
electrons reach the central axis at flattener. Therefore, it is important to
100 cm SSD for 6 MV than for 18 MV. check that depth dose curves are
calculated accurately as well.
C. Beam Commissioning BEAM simulations were
The beam commissioning completed, and the corresponding
procedure consists of two parts: (1) device files generated for a set of
selecting/interpolating the initial beam energies around the nominal
electron energy incident on the energy of the machine. Using
bremsstrahlung target and (2) setting PEREGRINE, dose profiles were
the dose per monitor unit, based on calculated for each of these energies.
the specific calibration of the An interpolation was done between
accelerator. the calculated profiles to find beam
Field flatness is sensitive to beam energy. This energy was then used to
energy.Error: Reference source not generate a new device file by doing a
found Figures 6 and 7 show how the linear interpolation using the nearest
electron beam energy affects the field two device files, without rerunning
flatness and depth dose for 38x38 cm BEAM. For this study, voltages were
and 2x2 cm fields, respectively. The linearly interpolated from a library of
effect of electron beam energy on simulations at 6.0, 6.5, and 7.2 MeV
field flatness is most apparent for the and 17, 18, and 19 MeV for 6 and 18
largest field size. Figure 6 shows that MV beams, respectively. Voltages
a variation of 4% in the off-axis ratio selected were 6.2 and 18.5 MeV for 6
(defined at 10 cm from the axis) at 10 and 18 MV beams.
cm depth results from variation in Once the effective energy is
electron beam energy from 6 to 6.5 determined, we calibrate the internal
MeV, with a variation of 4% from 17 to particle fluence metric in terms of
19 MeV. Small fields have the most dose per monitor unit (MU). The user
sensitive depth dose curves to inputs the cGy/MU at 10 cm depth on
electron energy. Figure 7 shows that the central axis of a 10x10 cm2 field,
2x2 cm2 depth dose curves are most and PEREGRINE uses this number to
affected at shallower depths. For the determine the effective weight of
nominal 6 MV beam, variation in each history, so that dose is
electron beam energy from 6 to 7 MeV calculated in units of cGy/MU.
results in a difference of 7% at a For purposes of simulations, the
depth of 30 cm. A variation of 17 to 19 geometry of collimator jaws, wedges,
MeV results in a maximum difference wedge trays, block trays, and multi-
of less than 3% at a depth of 30 cm. leaf collimators (MLCs) are described
Because large-field flatness is more in terms of density, composition,
sensitive to electron energy than shape, and location. Block thickness,
depth dose for high energy x-ray material, and density, and aperture
beams and because profile shape and MLC leaf positions are
measurements are less subject to described by the user at calculation
systematic errors due to slight time. For the measurements shown in
misalignment of the gantry and beam this paper, wedges are composed of
scanner, it is used to estimate the stainless steel (density: 7.86 g/cm3,
12

composition: 0.5% Si, 18% Cr, 2.0% measurements, as justified below.


Mn, 69.9% Fe, 9.6% Ni by weight) and Profile data was shifted by up to 1
blocks are composed of Cerrobend mm to achieve the best match with
(density: 9.38 g/cm3, composition: calculation. The IC-10 was positioned
6.0% Cd, 8.4% Sn, 29.6% Pb, 56.0% Bi in depth with an accuracy of 1.5 mm,
by weight). All wedge dimension and including an estimate of systematic
position data were taken from Varian uncertainty in the point of
specifications, with the exception of measurement correction. This
the 60o wedge: we shifted the 60o correction, which was 1.8 mm at 6 MV
wedge by 2 mm to obtain good and 2.0 mm at 18 MV, was
agreement with measurements for automatically applied by Wellhofer
wedge factors. software during scanning. Separate
comparisons of IC-10 and
D. Measurements extrapolation ion chamber
measurements were done of depth
All measurements in this paper dose distributions in the build-up
were taken on the University of region at 100 cm SSD. IC-10 and
California San Francisco Varian Clinac extrapolation ion chamber
2100C using a Wellhofer water measurements agreed within 1.5 mm
phantom. Output, profile and depth at depths greater than 2.5 mm. Larger
dose measurements were made using discrepancies than this were
a Wellhofer IC-10 ionization chamber apparent within 2.5 mm of the
(0.147 cm3 active volume with a 6.0 surface, due to the finite diameter of
mm diameter and 6.3 mm active the IC-10. Therefore, comparisons
length, 0.4 cm wall thickness). Small- with measurement are limited to
field profile measurements were depths greater than 2.5 mm. The
made with a Scanditronix photon precision on the IC-10 ion chamber
diode (p-type silicon, chip thickness measurements is +/- 0.3%. Systematic
of 0.45 mm, 2.5 mm diameter). errors for relative dose
Measurements are reported as measurements beyond the build-up
dose per monitor unit. This was region are not expected to exceed 1%
determined as follows. All for measurements we report using
measurements were normalized to a either ion chamber or diode
single reference measurement taken measurements.41 42
at 10 cm depth for a 10x10 cm2 field For measurements with the
at 90 cm SSD. A dose rate was multileaf collimator, photon diode
assigned to measured current for the measurements are used because of
reference measurement according to their superior spatial resolution.
dose expected at that point based on Measurements for 5x5, 10x10, and
the calibration condition of the 20x20 cm fields demonstrated good
accelerator, 1 cGy / MU at 100 cm agreement between profile
SSD for a 10x10 cm2 field. No further measurements with a diode and IC-10
ion chamber corrections were used. ion chamber measurements in areas
of low dose gradient. Diode profile
We assign an overall experimental measurements were normalized to an
uncertainty of 1% in dose, 1.5 mm in ion chamber depth dose
depth position for the IC-10 measurement.
measurements and 1% in dose, 0.5
mm in depth position for the diode
13

III. Results respectively. We have three pieces of


evidence that support the presence of
We compare calculations with
additional electrons upstream of the
measurements for open fields ranging
jaws: (1) field-size and energy
from 2x2 to 38x38 cm, and for fields
dependence of the discrepancy, (2)
modified by wedges, blocks, and
excellent agreement between the
multi-leaf collimators. All
calculated and measured depth dose
comparisons are reported in dose per
for large fields below wedges
monitor unit, including a correction
(discussed below), and (3) reduction
for the variation in backscatter to the
of the discrepancy in the buildup for
monitor chamber with jaw position,
large fields in the presence of a 6-mm
with no further normalization done.
acrylic tray (data not shown).
That is, depth dose curves and
Increasing the weight of the electron
profiles reported include relative
subsource for both the 6MV and 18MV
output and wedge factors. All
beam models provides close
measurements had a source-to-
agreement in the build-up region
surface distance (SSD) of 90 cm. This
wherever discrepancies occur, but
distance is representative of typical
small discrepancies can remain
patient set-ups.
because increasing the weight is an
Open field comparisons were
approximation or because some
done for both square and rectangular
residual discrepancies may be
fields. Figures 8 and 9 show measured
experimental.
and calculated depth dose curves on
Figures 10 and 11 show profiles for
the central axis of the beam, at
several representative fields. The
depths greater than 5 cm, for 2x2,
38x38 cm2 field comparison (Figure
5x5, 10x10, 20x20, and 38x38 cm2
10) shows agreement to within a
fields. Calculated depth dose
maximum relative error of 1%
distributions are slightly steeper than
between calculations and
for measurements, with a maximum
measurements inside the field. Beam
deviation of 2%, evident for the 2x2
energies were chosen to achieve a
cm2 18 MV field. A possible
good match for these measurements.
explanation for this is that electron
In the penumbra region, the effects of
voltage, tuned by the procedure
the IC-10 chamber width cause up to
described above, is slightly low.
1 mm discrepancies with calculations
However, Figure 10 shows better than
for 6 MV. These effects are much
1% agreement between calculations
smaller for 18 MV, because of its
and measurements for a 38x38 cm2
broader penumbra. Comparisons
field profiles, which are also sensitive
outside the field (Figures 11) reveal
to beam energy.
that PEREGRINE agrees to within 2%
Figure 9 compares measured and
with measurements for 5x5 and 10x10
calculated depth dose curves on the
cm2 fields. However, outside the
central axis of the beam near the
20x20 cm2 field, PEREGRINE
phantom surface for 5x5, 10x10, and
underestimates the dose by as much
38x38 cm fields. With added
as 10% of the measured dose. This
electrons, calculations agree with
has little clinical significance, as
measurements to within 2% and 8%
these errors are less than 1% of the
(dose), or 0.4 mm and 1.2 mm
dose on the central axis. Calculations
(isodose displacement), of the dose
done with BEAM/DOSXYZ show the
measurement for 6 and 18 MV,
14

same discrepancy, agreeing with 10x10, and 40x40 cm2 fields,


PEREGRINE. BEAM/DOSXYZ show the respectively. For 18 MV, the
same discrepancy, agreeing with backscatter correction results in
PEREGRINE. This is also consistent factors of 0.9897, 1.0000, 1.0425 for
with the trends observed by other the same fields, respectively.
workers,Error: Reference source not PEREGRINE calculations include a
found 43 and demonstrates that the backscatter correction for all figures
difference is not due to deficiencies in this paper, unless otherwise stated.
in PEREGRINE or the beam model. Comparisons between
Measurement/calculation differences calculations and measurements for
may be due to a source of scattered 5x20, 20x5, 5x40, and 40x5 cm2
or leakage radiation that is not rectangular fields were also done in
currently being accounted for in the order to stress the backscatter
beam accelerator head simulation. correction factor and investigate the
Measurement/calculation effects of added electrons. Table 1
differences may be due to a source of summarizes measured and calculated
scattered or leakage radiation that is output factors (relative to a 10x10 cm2
not currently being accounted for in field at 10 cm depth), which agree to
the beam accelerator head within 1% with measurements. Depth
simulation. dose calculations (with added
electrons) in the build up region of
Figure 12 is a comparison between these rectangular fields agree with
calculated and measured output measurements to within 7% ( 6 MV)
factors on the central axis of the and 15% (18 MV), resulting in an
beam at a depth of 10 cm. isodose shift of less than 1 mm for
PEREGRINE includes a provision to each case. The added electrons have
account for backscatter as described little effect in the build up region,
in the Materials and Methods section. consistent with the small square field
Bea Field Meas Calc Δ results. Depth dose calculations
m Size . ROF . deeper than the maximum dose agree
ROF with measurements to within 2% and
1% relative dose for 6 and 18 MV,
6 MV 5x20 0.96 0.97 1.3%
respectively.
cm2 0 2
20x5 0.94 0.95 0.8% Table 1. Comparison between
cm2 9 0 measured and calculated relative
5x40 0.98 0.98 0.1% output factors (relative to a 10x10 cm2
cm2 4 3 field at 10 cm depth) for 6 and 18 MV
40x5 0.96 0.96 0.1% beams.
cm2 2 2
18 5x20 .977 .981 0.4%
With a backscatter correction,
MV cm2
PEREGRINE agrees with
20x5 .955 .956 0.03
measurements to within 1.2% and
cm2 %
1.6% for 2x2 and 38x38 cm2 fields,
respectively. For 6 MV, the 5x40 .994 .984 0.9%
cm2
backscatter correction used in
PEREGRINE (renormalized to 1 for a 40x5 .964 .955 1.0%
10x10 cm2 field) results in factors of cm2
0.9940, 1.0000, 1.0254 for a 2x2,
15

Figure 13 compares calculations The small residual discrepancy in the


with measurements for fields buildup observed for wedged fields
modified by a wedge. Profile may be due to experimental
calculations for a 20x40 cm2 field uncertainty, such as chamber
modified by a 45o wedge agree with positioning and changes in chamber
measurements to within 2% inside response at shallow depths.
the field. Both 6 and 18 MV show a Comparisons between
slightly greater slope to the calculations and measurements for 15
o
calculated profile than is measured. and 60 o wedges (20x40 and 15x40
This probably relates to a small error cm2 fields, respectively) show similar
in the composition or density used for results, with maximum discrepancies
the steel wedge. The 6 MV of 2% inside the beam at depths
calculations slightly underestimate greater than dmax, for both 6 and 18
dose, while 18-MV calculations MV, and 7% in the buildup region.
slightly overpredict dose, consistent summarizes measured and calculated
with the same trends shown for open wedge factors (defined as dose at 10
field output factors. Outside the cm with wedge relative to dose at 10
beam, calculations agree with cm for a 10x10 cm2 open field) for all
measurements to within 4% for 5, 10, wedges studied.
and 20 cm-depth profiles. For the
profile at dmax, calculations
underpredict the dose by up to 9%, Table 2. Comparison between
consistent with our observations for measured and calculated wedge
open fields. This difference amounts factors for 15 o, 45o, and 60 o wedges.
to 1% of the dose on the central axis. Wedge factor is defined as dose at 10
Calculated central axis depth cm depth with wedge in (20x40 cm2
dose curves shown in Figure 13 agree field for 15o and 45 o, 15x40 cm 2 for 60
with measurement to within 2% for o
) divided by dose at 10 cm depth for a
depths greater than dmax for both 6 10x10 cm2 open field.
and 18 MV beams. No additional
electrons need be added to the Bea Wedg Meas Calc. Δ
source to achieve this level of m e . WF WF
agreement, as the wedge absorbs
most electrons from the source. In

6 MV 15 o 0.916 0.912 0.4%


45o 0.59 0.588 1.3%
6
the buildup region, calculations agree 60o .497 .492 1.1%
with measurements to within 7%, o
resulting in isodose displacement of 18 15 0.911 0.922 1.2%
less than 1 mm. This is similar to MV 45 o
0.65 0.648 0.6%
differences found in open-field 2
buildup comparisons after an 60o .565 .564 0.1%
additional electron source has been
added. The wedge itself introduces a Calculations shown in Figure 14
new source of electrons, emanating are for a 7.5 cm-thick cerrobend block
from the bottom of the wedge tray. fixed on top of a 0.6 cm-thick acrylic
16

block tray. Results agree with ion model. The results of a set of
chamber measurements to better calculation/measurement
than 1% at 6 MV and better than 2% comparisons show the accuracy of
at 18 MV in unblocked areas, better the overall implementation of the
than 3% (6 and 18 MV) in blocked code, including the beam model and
areas. Calculated dose outside the commissioning procedure. The only
penumbra agrees with measurements normalization done was to use a
to better than 6% (0.5% of the single-point calibration. Results,
maximum dose) on the unblocked summarized in Table 3, indicate good
side and better than 3% (0.3% of the agreement between calculations and
maximum dose) on the blocked side. measurements in dose per monitor
A complex, comb pattern, with unit for full dose distributions under
blocked and open regions was used to open fields and for a variety of beam
compare PEREGRINE calculations to modifiers. In the low-dose gradient
photon diode measurements for the regions of the field, utilizing a
multileaf collimator (Figure 15). published correction curve for
Collimators were set to 20x26 cm2. On monitor chamber backscatter and an
the side of the beam close to the 5 empirical correction to the electron
cm wide-open region, the beam is source fluence, PEREGRINE agrees
collimated by two leaves, which with measurements to within 2% of
extend 2-cm beyond the collimator the dose at the measurement point.
jaw, which is set to the edge of the Calculated output factors and wedge
multileaf collimator. Leakage factors are good to within 2%. In the
radiation scattering around the penumbra region, PEREGRINE
collimator jaw and multileaf predictions result in spatial isodose
collimator is responsible for the small discrepancies of less than 1 mm.
peak at the edge of the field. On the 1 Outside the penumbra, discrepancies
cm side of the comb pattern, only the are larger: PEREGRINE systematically
collimator jaw blocks the field. predicts a lower dose than measured,
Because of the large number of high- with discrepancies as high as 15%.
dose-gradient areas, we compare While these differences are large
measurements with photon diode compared to the dose at the
measurements. Agreement between measurement point, they amount to
calculations and measurements is less than 1% discrepancies expressed
generally <1% of maximum dose for as a fraction of the maximum dose.
low-gradient areas of both blocked Where tested (open fields)
and unblocked areas for the 1-cm PEREGRINE agrees with EGS4
(one-leaf-in, one-leaf-out) area of the (BEAM / DOSEXYZ), with both codes
field. PEREGRINE dose reproduces underpredicting dose in the build-up
the small peaks outside the 4-cm side region of large fields and in the area
of the beam. blocked by the collimator jaws by over
2%. This, combined with the
systematic nature of the
IV. Summary and
discrepancies, suggest that the
Conclusions remaining discrepancies over 2%
In this paper, we provide an cannot be attributed to PEREGRINE
overview of the PEREGRINE code, radiation transport or to the beam
including descriptions of the radiation model, but rather to leakage or
transport physics and the x-ray beam
17

scatter radiation not accounted for in isodose position. This accuracy


the treatment head simulation. applies over the wide field size range
With the added electron source considered and for standard beam
and backscatter correction, the dose modifiers, including wedges, blocks,
calculation, relative to the and multi-leaf collimators. This result
accelerator output reference meets or exceeds the published
condition, is accurate to either 2% of results for other Monte Carlo codes.
maximum dose in the field or 1 mm in

Table 3. Summary of maximum discrepancies observed in


calculation/measurement comparisons. Profile discrepancies are quoted as
absolute dose difference between calculation and measurement when central axis
dose is set to 100%.

Field Added Results


Electrons
?
2-38 cm square ROF No effect 1% with backscatter correction; 3% (6
MV) and 5% (18 MV) if backscatter
correction not applied
Depth: < dmax (Buildup) Depth: > dmax
2-38 cm square PDD NO 3 mm (6 MV) 5 mm (18 2%
MV)
YES 1 mm (18 MV) 2%
5x20, 20x5, 5x40, No effect 1% with backscatter correction
40x5 cm rectangles
ROF
Depth: < dmax (Buildup) Depth: > dmax
5x20, 20x5, 5x40, NO 2 mm 2%
40x5 cm rectangles YES 1 mm 2%
PDD
Inside Penumbra Outside Field
Field
38 cm square No effect 1% 1 mm 1%
profiles, dmax-20 cm
2, 10, 20 cm square No effect 2% 1 mm 1%
profiles, dmax-20 cm
15 o, 45 o and 60 o No effect 2% 1 mm 1%
wedges
Cerrobend block No effect 2% 1 mm 0.5%
Multi-leaf collimator No effect - 1 mm 1%
in comb pattern
18

VI. References
1
K. R. Shortt, C. K. Ross, A. F. Bielajew and D. W. O. Rogers (1986) “Electron beam dose

distributions near standard inhomogeneities” Phys. Med. Biol. 31 235-249.

2
T. M. Jenkins and A. Rindi, ed (1988) “Monte Carlo Transport of Electrons and Photons”

Plenum Press, New York

3
T. R. Mackie (1990) “Applications of the Monte Carlo method in radiotherapy” Dosimetry of

Ionizing Radiation vol 3, ed K. Kase, B. Bjarngard and F. H. Attix (New York: Academic) pp

541-620.

4
P. Andreo (1991) “Monte Carlo techniques in medical physics,” Phys. Med. Biol. 26 861-920.

5
R. Mohan (1997) “Why Monte Carlo?” Proc. XIIth ICCR (Salt Lake City, USA) (Madison , WI:

Medical Physics Publishing) 16-18.

6
J. J. DeMarco T. D. Solberg, and J. B. Smathers, (1998) “A CT-based Monte Carlo simulation

tool for dosimetry planning and analysis,” Med. Phys. 25(1) 1-11.

7
M. R. Arnfield, C. Hartmann Siantar, J. Siebers, P. Garmon, L. Cox and R. Mohan 2000 “The

impact of electron transport on the accuracy of computed dose,” 27 (6) pp. 1266-1274.

8
H. Neuenschwander, T. R. Mackie, and P. J. Reckwerdt (1995) “MMC – a high-performance

Monte Carlo code for electron beam treatment planning,” Phys Med Biol 40: 543.

9
P. J. Keall and P. W. Hoban (1996) “Super-Monte Carlo: A 3D electron beam dose calculation algorithm,” Med Phys

23:2023.

10
C. L. Hartmann Siantar, P. M. Bergstrom, W. P. Chandler, L. Chase, L. J. Cox, T, et. al. (1997)

“Lawrence Livermore National Laboratory’s PEREGRINE Project,” Proc. XIIth ICCR (Salt

Lake City, USA) (Madison , WI: Medical Physics Publishing) 19-22.

11
M. Fippel, W. Labu, B. Huber and F. Nusslin (1999) “Fast Monte Carlo dose calculation for

photon beams based on the VMC elecgron algorithm” Med Phys 26: 1466-1475.

12
J. Sempau, S. J. Wilderman and A. F. Bielajew (2000) “DPM, a fast, accurate Monte Carlo

code optimized for photon and electron radiotherapy treatment planning dose

calculations” Phys Med Biol 24: 2263-2291.


13
C.-M. Ma, J. S. Li, T. Pawlicki, S. B. Jiang and J. Deng (2000) “MCDOSE – A Monte Carlo

dose calculation tool for radiation therapy treatment planning” XIIIth International Conf of

The Use of Computers in Radiation Therapy, published by Springer

14
R. Mohan, C. Chui and L. Lidofsky (1985) “Energy and angular distributions of photons

from medical linear accelerators,” Med. Phys. 12 592-597.

15
J. V. Siebers, P. J. Keall, B. Libby and R. Mohan (1999) “Comparison of EGS4 and MCNP4b

Monte Carlo codes for generation of photon phase space distributions for a Varian 2100C,”

Phys. Med. Biol. 44 3009-3026.

16
D. W. O. Rogers, B. A. Faddegon, G. X. Ding, C-M Ma, J. S. Wei and T. R. Mackie (1995)

“BEAM: a Monte Carlo code to simulate radiotherapy treatment units,” Med. Phys. 22 503-

524.

17
D. M. J. Lovelock, C. S. Chui and R. Mohan (1995) “A Monte Carlo model of photon beams

used in radiation therapy,” Med. Phys. 22 1387-1394.

18
G. Kuster, T. Bortfeld and W. Schlegel (1997) “Monte Carlo simulations of radiation beams

from radiotherapy units and beam limiting devices using the program GEANT,” Proc. XIIth

ICCR (Salt Lake City, USA) (Madison , WI: Medical Physics Publishing) 150-152.

19
B. Faddegon, J. Balogh, R. Mackenzie and D. Scora (1998) “Clinical considerations of
Monte Carlo for electron radiotherapy treatment planning,” Radiat Phys Chem 53: 217-227.
20
C.-M. Ma and S. B. Jiang (1999) “Monte Carlo modeling of electron beams from medical
accelerators,” Phys Med Biol 44 (1):R157-R189.
21
A. E. Schach von Wittenau, L. J. Cox, P.M. Bergstrom, W. P. Chandler, C.L. Hartmann

Siantar, and R. Mohan, "Correlated histogram representation of Monte Carlo derived

medical accelerator photon-output phase space," Medical Physics 26:7 1196-1211 (1999).

22
J. Deng, S. B. Jiang, A. Kapur, J. Li, T. Pawlicki and C-M. Ma (2000) “Photon beam

characterization and modeling for Monte Carlo treatment planning,” Phys. Med. Biol. 45

411-427.

23
B. A. Faddegon, P. O-Brien, and D. L. D. Mason (1999) “The flatness of Siemens linear

acclerator x-ray fields,” Med. Phys. 26 (2) 220-228.


24
C.-M. Ma, B.A. Faddegon, D. W. O. Rogers, and T. R. Mackie (1997) “Accurate
characterization of Monte Carlo calculated electron beams for radiotherapy” Med. Phys.
24(3): 401.
25
D. E. Cullen, M. H. Chen, J. H. Hubbell, S. T. Perkins, E. F. Plechaty, J. A. Rathkopf and J. H.

Scofield, Lawrence Livermore National Laboratory Report UCRL-50400, Volume 6 (LLNL,

Livermore, 1989)

26
J. H. Hubbell, W. J. Veigele, E.A. Briggs, R.T. Brown, D. T. Cromer and R. J. Howerton, J.

Phys. Chem. Ref. Data 4, 471 (1975)

27
C. M. Davisson and R. D. Evans, Rev Mod Phys 24, p.79 (1952)

28
J. H. Hubbell, H. A. Gimm, and I. Overbo, J. Phys. Chem. Ref. Data 9, 1023 (1980).

29
International Commission on Radiation Units and Measurements (ICRU), ICRU Report 37

(ICRU, Bethesda, 1984)

30
F. H. Attix (1986) Introduction to Radiological Physics, John Wiley & Sons, New York.

31
W.R. Nelson, H.Hirayama, and D.W.O. Rogers, “The EGS4 Code System,” Report SLAC--265,

Stanford Linear Accelerator Center, Stanford, Calif, 1985.

32
S. T. Perkins, D. E. Cullen and S. M. Seltzer, Lawrence Livermore National Laboratory

Report UCRL-50400, Volume 31 (LLNL, Livermore, 1991).

33
A. F. Bielajew, R. Mohan and C. Chui, “Improved Bremsstrahlung Photon Angular Sampling

in the EGS4 Code System”, National Research Council of Canada Report PIRS-0203 (1989).

34
B. A. Faddegon, C. K. Ross, D. W. O. Rodgers, “Angular distribution of bremsstrahlung from

15-MeV electrons incident on thick targets of Be, Al, and Pb”, Med. Phys. 18(4) p. 727-739

(1991).

35
I. Lux and L. Koblinger, Monte Carlo Particle Transport Methods: Neutron and Photon

Calculations, (CRC Press, Boca Raton, 1991) p. 40

36
M. J. Berger “Monte Carlo Calculation of the Penetration and Diffusion of Fast Charged

Particles” in Methods in Computational Physics, Vol1, (Academic Press, New York, 1963) p.

135.
37
G.~Z. Moliére, “Theorie der Streuung schneller geladener Teilchen. I. Einzelstreuung am

abgeschirmten Coulomb-Field,” Z. Naturforsch, 2a 133 -- 145 (1947).

38
A. F. Bielajew and D. W. O. Rogers (1987) “PRESTA: the parameter reduced electron-step

transport algorithm for electron Monte Carlo transport,” Nucl. Instrum. Methods B 18 165-

181.

39
K. L. Lam, M. S. Muthuswamy, and R. K. Ten Haken (1998) “Measurement of backscatter

to the monitor chamber of medical accelerators using target charge,” Med. Phys. 25 (3)

334-338.

40
C. Dunzenli, B. McClean, and C. Field (1993) “Backscatter into the beam monitor chamber:

Implications for dosimetry of asymmetric collimators,” Med. Phys. 20 (2:1) 363-367.

41
D. Sheikh-Bagheri, D. W. O. Rogers, C. K. Ross and J. P. Seuntjens (2000) “Comparison of

measured and Monte Carlo calculated dose distributions from the NRC linac,” submitted to

Med Phys.

42
M. Westermark, J. Arndt, B. Nilsson, and A. Brahme (2000) “Comparative dosimetry in

narrow high-energy photon beams,” Phys. Med. Biol. 45 685-702.

43
L. Wang, C-S. Chui, and M. Lovelock (1998) “A patient-specific Monte Carlo dose-

calculation method for photon beams,” Med. Phys. 25 (6) 867-878.

You might also like