Automatic Digital Refractometer RX-5000α

The RX-5000α ( alpha ) is an automatic digital refractometer which can set measurement
temperature internally and has the following features for measuring the refractive index, Brix or
concentration of various liquids accurately and speedily
• Since the RX-5000α ( alpha ) has a thermo-module to control temperature, a constant
temperature water bath is not necessary.
• The measurement starts automatically after the sample reaches your targeted temperature.
• The refractive index and Brix at your targeted temperature are speedily displayed.
• High Brix ± 0.03% and refractive index ± 0.00004 accuracy can be obtained.
• The RX-5000α ( alpha ) displays the top-and-bottom-limit-bar for your set control range.
• If the measurement value differs from that of your Standard Liquid or the other refractometer, it
can be adjusted in a certain range.
• 60 kinds of user scales can be input according to your sample.
• The RX-5000α ( alpha ) is capable of calling up 30 latest measurements.

* Special model of RX-5000α allows pre-programming of an additional scale (such as honey


moisture, inverted sugar% or salt concentration) in addition to Refractive Index and Brix (%) at
customer's request.
This product comes standard with a two (2) year limited warranty against manufacturer? defects
from the date of the original purchase.

FDA 21 CFR Part 11 Software Included in Standard Delivery.

QAQC LABS White Stone Va 22578 TEL (866) 244-1578


www.qclabequipment.com
Model RX-5000α Cat.No. 3261
Refractive index (nD) : Refractive index (nD) :
Measurement 1.32700 to 1.58000 Minimum 0.00001
Range Brix : 0.00 to 100.00% indication Brix : 0.01%
( 5 to 60°C ATC) Temperature: 0.01°C
Refractive index (nD) : Measurement
5 to 60°C
±0.00004 Temperature
Measurement
*±0.00002
Accuracy
Brix : ±0.03%
*Repeatability Power Supply AC100V to 240V, 50/60Hz
*±0.01%
[*1]
Power Dimensions & 37×26×14cm, 6.4kg
65VA
consumption weight (Main Unit only)
MODE-1:
Displays the measurement value once the sample reaches the target
temperature.

MODE-2:
Measures refractive index and temperature at fixed intervals and displays the
estimated measurement value at the target temperature.
Mode
MODE-3:
The thermo-module can be turn off. Without temperature control, the
measurement value is displayed in 4 seconds after the START key is pressed.

MODE-S:
Displays the measurement value once a certain level of sample stability is
achieved.
• Printer (for ATAGO digital printers)
Output terminals • Computer - RS-232C
Connection to a USB port requires a USB to RS-232 adapter (optional)