<<Contents>> <<Index>>

General PH400G
Specifications 4-Wire Type pH Meter

GS 12B7C1-E

GENERAL
The EXA PH intelligent pH conveter is a highly reliable
advanced function pH measurement system designed for
wide application in all industries such as for medium-scale
waste water treatment system, and general-purpose pH
measurement and control systems.
The EXA PH intelligent pH converter features highly
effective built-in functions for sensor diagnosis similar to
FLXA21 2-wire liquid analyzer. These functions monitor
sensor errors even while pH measurements are in
progress, enabling more highly reliable pH measurement.
The converter also offers an interactive system for setting
the various parameters, facilitating effective use of its
wealth of functions.
The EXA PH intelligent pH converter is available not only
in versions for general-purpose applications but also for
high-purity water service.

High-Purity Water pH Measurements


High-purity water today is widely used in numerous
process plants in a variety of chemical processes and PH400G
boilers, with heavy emphasis placed on pH measurement
and monitoring for water quality control and water
purification unit operational control.
However, the use of general-purpose pH meters for pH
measurements in high-purity water meets with the
following problems.
d Since measurement solutions are chemically pure, d Sensor characteristic deterioration is automatically
their pH values may vary owing to carbon dioxide in the checked during automatic calibration using the standard
atmosphere and / or impurities in measurement lines. solutions.
d The pH electrode output is likely to be affected by The sensor replacement time that previously depended
streaming potentials, externally induced potentials, etc. on intuition can now be decided automatically.
and thus likely to produce measurement error because d Daily operation can be performed without opening the
the liquid resistance between the electrodes in case.
measurement solution is high owing to low solution Even in rainy weather, since the routine maintenance
conductivity. operation such as calibration using standard solutions
Since the pH sensors and their holders for high-purity can be performed without opening the case lid, there is
water service have been designed considering no worry about insulation deterioration.
countermeasures against the above problems, they can d Ranges can be freely set in the field if the span is 1 pH
provide highly reliable pH measurements. Moreover, or more.
since the EXA PH intelligent pH converter is equipped d A timer function for cleaning is incorporated.
with a reference temperature conversion function to d Abundant built-in contact output function
compensate for pH variation owing to measured liquid d Operating panel has excellent operability.
temperature variation, any effect from sample liquid
Ryton pH Sensor Body
temperature fluctuation is slight, and devices such as
sampling system coolers, etc. can thus be simplified. d The sensor body uses Ryton which compares with
Teflon in corrosion resistance and heat resistance and
has excellent mechanical properties. This pH sensor can
FEATURES cover most applications.
d This pH sensor is easy to handle.
Microprocessor - Based Intelligent pH converter As this is a composite pH sensor, calibration using
d Sensor failure can be monitored even during pH standard solutions or maintenance can be easily
measurement. performed.
d Automatic calibration using standard buffer solutions Glass electrode and junction can also be simply
can be initiated by merely a single touch of a key. replaced.
Tables for three types of standard solution are
incorporated. This allows for simple automatic
calibration.

Yokogawa Electric Corporation GS 12B7C1-E


2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan ©Copyright Jan. 1991
Tel.: 81-422-52-5617 Fax.: 81-422-52-6792 6th Edition Nov. 2012
<<Contents>> <<Index>> 2
SYSTEM CONFIGURATION
For the pH sensors, see GS 12B07B02-E.
For the holders or cleaning devices, see GS 12J05C02-00E.

Fig. 1-a System Configuration (General Purpose)


pH sensors Holder, Holder with Cleaning system pH Converter

d KCL refillable type d Xerolyt d Guide Pipe d Suspension Type Holder d 4-wires pH Converter
PH8ERP HA405/HA485 PH8HG HH350G PH400G
<No Cleaning> <With Jet Cleaning>
pH

MEASURE
AUT CAL
MAN CAL
TCMP
DISP
HOLD

CONTACTS
YES NO MODE
S1

S2

S3
>

>
ENT
FAIL

d for Chemical
DPA405/DPA485
d Angled Floating Ball Holder
d Submersion Type Holder PB350G

d KCL filling type Float


PH8EFP
d Vertical Floating Ball Holder

d Hydrofluoric
Acid-resistive
HF405

No Cleaning Ultrasonic Cleaning


Jet Cleaning
Brush Cleaning
d Flow Type Holder
PH8HF
Accessories Cleaning Devices

d Sensor Stand d Ultrasonic Oscillator


PUS400G

No Cleaning Ultrasonic Cleaning


Jet Cleaning
Brush Cleaning

Fig. 1-b System Configuration (For Pure Water)


pH sensor Holder Accessories pH Converter
d For pure water PH8AX d 4-wires pH Converter
PH8HH PH400G
d Sensor Stand
d For pure water
PH8EHP d Calibration
pH

MEASURE
AUT CAL
MAN CAL
TCMP
DISP
HOLD

Reagent and YES

>
NO MODE
CONTACTS

S1

S2

S3
>

ENT

KCL Solution
FAIL

F1-a.b.c.EPS

CAUTION
Installation Location of Holders (Guide Pipe, Submersion Type, etc)
The holder should be used in a place that is as vibration free as possible.
Using the holder in a place where it is affected by vibration, may result in damage to the holder.

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 3
SPECIFICATIONS Notes:
(*1): The solution temperature should preferably be
below 808C when the pH value is in the range 11 to
EXA PH Series General Specifications 14. Even if the solution is below 808C, sensor
Measured object: Hydrogen ion concentration (pH) in service life may be shortened in store solutions. In
aqueous solutions such cases the KCl filling type is recommended.
(*2): Maximum temperature 508C when Hard PVC guide-
Measuring principle: Glass electrode method
pipe is used.
Measuring range: 0 to 14 pH (*3): For flow-through types, refer also to the solution
Measuring conditions: temperature and pressure graph (in notes following
Solution temperature: See Table 1, for high-purity Model and Suffix code table for flow-through type
water, 0 to 508C holders).
Solution pressure: See Table 2, for high-purity water, (*4): Solutions with normal pH ranges of 3 to 14 are
atmospheric pressure (measured liquid recommended for stainless steel (SUS316).
outlet side open to atmosphere)
Solution flow rate: Max. 2 m/s for submersion guide-
pipe type 3 to 11 l/min for flow-through Table 2. Process Pressure Range
type, for high-purity water, see Figure 2. pH
Solution conductivity: Min. 50 mS/cm, for high-purity Sensor PH8ERP PH8EFP
water, see Figure 1. Holder
Temperature compensation range: -10 to 1308C (for Submersion
Atmospheric pressure (Submersion depth: Max. 3 m)
pH converter)
Guide-pipe Atmospheric pressure
Suspension
Table 1. Process Temperature Range Float (Submersion depth: Max.3 m)

(a) pH and Liquid Temperature Ranges for pH sensors Atmospheric pressure to 10 kPa
when general purpose reserve
Atmospheric tank used.
Solution Solution Flow-through pressure
pH Sensor Holder Type Temperature to 50 kPa Atmospheric pressure to 500 kPa
pH Ranges Ranges (8C) when medium-pressure reserve
tank used.
Guide-pipe *3 *2
KCl refilable Solution 2 to 12 -5 to 80 (Note 1) For flow-through types, refer also to the solution temperature
type Flow-through and pressure diagram of Holder GS 12J5C2-E.
T03.EPS
*2
Guide-pipe 2 to 12 -5 to 80
KCl filling Submersion *3 *1 *1
0 to 14 -5 to 105 700
type
ml/min

Flow-through
600
T01.EPS
500
Solution flow rate

400
(b) Solution Temperature Ranges for Holders
300
Usable range
Liquid 200
Holder Type Holder Cleaner Temperature
Material 100
(8C)
0
PV None -5 to 50 0 0.1 0.2 0.3 0.4 0.5 0.6 50
Guide-pipe
PP None -5 to 80 Solution conductivity (mS/cm)
F02.EPS
None -5 to 100
PP
Provided -5 to 80 Figure 2 Solution flow rate and solution conductivity of sensor
Submersion and holder for high purity water
*4 None -5 to 100
S3
Provided -5 to 80
None -5 to 80
PP
Flow-through Provided -5 to 80
*3 *4 None -5 to 105
S3
Provided -5 to 80
T02.EPS
PV: Hard PVC
PP: Polypropylene
S3: Stainless steel (SUS 316)

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 4
1. pH Sensors: PH8ERP, PH8EFP, PH8EHP Table 4. Utilities (For Jet Cleaning or Brush Cleaning)
A compound sensor assembly consisting of glass Pressure (kPa) Flow Rate
electrode, reference electrode, temperature sensor, and
ground tip. Water jet 200 to 400 + Liquid pressure 5 to 20 l/min
Type: KCl filling and KCl refillable types Water brush 100 to 250 + Liquid pressure 20 to 30 l/min
Measuring range: 0 to 14 pH for KCl refillable type; normal Air jet 200 to 400 + Liquid pressure 100 to 300 Nl/min
pH =2 to 12 pH)
Air brush 150 to 250 + Liquid pressure 300 to 600 Nl/min
Refer to "Selection Criteria for pH Sensors and Holders"
T06.EPS
shown on the last page to select KCl filling or refillable
type sensors.
Temperature compensation sensor: Platinum RTD 2.2 Holder For High-purity Water: PH8HH
(Pt1000) Materials: Acrylic resin, SUS316, chloroprene rubber
Wetted part materials: Process connections: Rc1/4 (inlet) and R1/2 (outlet),
or 1/4 NPT (inlet) and 1/2 NPT (outlet)
pH Sensor Wetted parts materials Mounting: Mount on 50 A vertical or horizontal pipe, or
wall-mount (specify mounting hardware)
Ryton (PPS resin), glass, ceramics, titanium Weight:
or Hastelloy C, fluorocarbon rubber and Body: 1.7 kg
KCl refillable type
chlorinated polyethylen rubber (carbon sheath
Mounting hardware: 0.7 kg
only when suspended in guide-pipe)

KCl filling type


Same as above. Plus heat resistant soft PVC 3. 4-wires type pH Converters: PH400G
(KCl tube only when suspended in guide-pipe)
Measuring Range
T04.EPS
pH: - 2 to15 pH
Weight: KCl refillable type; Approx. 0.4 kg
Temperature : -10 to 1308C
KCl filling type;
Display Method: Digital display
Body: Approx. 0.4 kg
Display Range
General purpose: Approx. 0.3 kg pH : -2 to15 pH
Tank:
Medium-pressure: Approx. 1 kg
Temperature : -10 to 1308C
For high-purity water; Body: Approx. 0.4 kg Transmission Signal Output
Tank: Approx. 0.3 kg 4 to 20 mA DC, isolated transmission
Note: Although sensors other than the EXA PH pH sensor output, Max. load 600 V
(Pt1000 temperature sensor) can also be used, there 0 to 1 V DC, isolated transmission
are some restrictions on function. Also, since output, Minimum load 1 kV pH or
reference performance values will differ, please (However, the above current temperature can
consult your Yokogawa representative. be freely set.
(4 to 20 mA) signal and the voltage
(0 to 1 V) signal are not isolated from
each other.)
2. Holders
2.1 General-Purpose Holders: PH8HG, PH8HF, Transmission Signal Range
PH8HS pH: Freely adjustable to any desired range of 1 pH span
Materials: Polypropylene, Stainless steel, PVC (guide- or greater (set at 0 to14 pH when shipped).
pipe only) Temperature: Freely settable to any desired range of
Cleaning: Ultrasonic, jet, or brush cleaning 508C span or greater (set to 0 to 1008C
Cleaning Unit Wetted Materials: when shipped).
Ultrasonic unit: Stainless steel, PVC (guide-pipe only)
Jet unit: Polypropylene
Brush unit: Polypropylene, titanium (shaft),
Ryton(bearings)
Mounting: Mount on 50 A vertical or horizontal pipe
Mounting hardware: One or two sets (submersion type) or
one set (flow-through type)
Weight: See Table 3.
Table 3. Holder Weight

Materials Stainless steel


Polyethylen Hard PVC
Holder (SUS316)
Approx. 0.5 to Approx. 1.5
Submersion 2.2kg to 6kg
Flow-through Approx. 0.5 to Approx. 7 to
2kg 8.5kg
Guide-pipe Approx. 1kg Approx. 1.6kg
T05.EPS

Note: Table does not include mounting hardware weights.


Mounting hardware weight:
For submersion type: Approx. 1 kg / set
For flow-through type: Approx. 0.5 kg

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 5
Contact Output Contact ON / OFF
?Contact output function that can be set
S1, S2, S3 FAIL
Contact Function Power Power
Setting (free selectable) Operation (free selectable) ON OFF ON OFF
S1 OFF, low limit, high limit, ON/OFF, Proportional duty pulses
*1
ON Closed ON Closed
HOLD Proportional frequency pulses *2 Action Open Action Closed
OFF Open OFF Open
OFF, low limit, high limit, Ditto T08.EPS
S2
HOLD
OFF, low limit, high limit, Ditto When closed, each indicator lamp is lit corresponding to
S3 HOLD, Cleaning timer, each contact status.
Hi - Hi limit, Lo - Lo limit ?Contact Type: Relay contact output (dry contact)
?Contact Rating
FAIL Failure ON/OFF
250 V AC, 2 A, Max. 100 VA
T07.EPS
(*1) and (*2) are effective when the low high limit contact 220 V DC, 2 A, Max. 50 W
Contact Input: Dry contact for manual start for automatic
output is to be used.
cleaning
(*1) Proportional duty pulse Ambient Temperature: -10 to 558C
Example of high limit alarm (conrol) output (hood may be fitting as option)
Storage Temperature: -30 to 708C
% duty cycle control 10% Ambient Humidity: 10 to 90% RH
90% Construction: Watertight complying with JIS C0920
100 equivalent to NEMA 4 waterproof
construction
Material:
Case; Aluminum alloy casting
Cover and window; Polycarbonate
50% 50% Finish: Baked polyurethane resin coating
50 Colors:
t on t off Cover; Deep sea moss green (Munsell 0.6GY3.1/2.0)
Case; Frosty white (Munsell 2.5Y8.4/1.2)
Mounting: Pipe mounting, wall mounting or rack
mounting
10% 90% Panel mounting
0 Power Supply Voltage: 88 to 132 V AC, 50 / 60 Hz
Proportional pH Duty cycle 176 to 264 V AC, 50/60 Hz
Setpoint range Time
% of output range Power Consumption: Approx. 8.5 VA
F02.EPS
Electrical Connection:
pH sensor: Watertight plastic gland equivalent to JIS
(*2) Proportional frequency pulse
A8 attached to pH sensor side, f13.5 hole
Example of high limit alarm (conrol) output
Others: Watertight plastic gland equivalent to JIS A15
% controller output (for cable OD. 9 to 12 mm) attached, f21
hole
100 Weight:
Maximum pulse frequency Body: Approx. 2.5 kg
Mounting bracket: Approx. 0.4 kg
Dimensions: 144 (W) x 144 (H) x 135 (D) mm
0.3 s
Function Specifications
Input Impedance: 1012 V or more
50 Note: Use a sensor with solution earth sensor because of
50% pulse frequency differential amplifier (that uses two high-impedance
amplifiers).
Asymmetry Potential Adjustable Range: pH 7 6 2 pH
Slope Adjusting Range: Adjustable in the range from 70 %
to 110 % of the theoretical value.
Automatic Temperature Compensation Range:
0 -10 to 1308C (manual compensation is also available)
Proportional pH No pulses Time
Reference Temperature Conversion Coefficient
range
Setpoint % of output range (Reference Temperature 258C):
F03.EPS
On shipment: 0
Settable parameters Adjustable range: -1.00 to 1.00 pH / 108C
? pH value setpoint Conversion to the reference temperature is used only
? Control range for high-purity water or when the measured
? Pulse periode solution temperature coefficient is known.

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 6
Standard Performance Setting of reference electrode impedance high limit
(performance when combined with a pH sensor) value
Repeatability: 0.05 pH (sensor submerged 3 times in Setting of response time high limit value
the same buffer solution) One-point temperature calibration
Response Time: 10 sec (90 % response, using pH Sensor type selection (glass electrode or
sensor and buffer solution both at antimony electrode)
temperature equilibrium at 208C, with Error reset
adequate agitation) pH display value selection (0.1 pH / 0.01 pH)
Accuracy: 60.1 pH (using KCl filling type pH sensor Half value recovery time check ON / OFF
or high-purity water pH sensor ) Half value recovery time setting (0.1 to 10 min)
60.15 pH (using KCl refillable pH Burn-up or burn-down ON/OFF
sensor) Response stabilization judging parameter setting
Temperature Repeatability: 18C Auto-return (approx. 60 min) ON/OFF
pH standard solution temperature characteristics
setting
Operating Functions
Temperature sensor cable length correction
Display: 3 1/2 digit digital display (data display)
Signal output selection
Six digit alphanumerics (message or data display)
Display function: 4 to 20mA DC OFF, pH Temperature
pH value
Temperature value 0 to 1V DC OFF, pH Temperature
mA output T09.EPS

mV output Contact (S1, S2, or S3) output selection


Reference electrode impedance Setting of contact output delay time (0.1 to 20 sec)
The e.m.f. slope and hysteresis (0.01 pH to 0.2 pH)
Asymmetry potential Setting of proportional duty pulse contact output
mV (e.m.f.) period (5 to 100 sec)
90% response time Setting of maximum frequency (50 to 120 pulses /
Error display (at error occurrence) min) at the proportional frequency pulse contact
Hold display (when holding) output
Manual temperature compensation display (in Control range (0 to 10 pH)
manual temperature compensation setting) Details of failure detection by self-diagnosis function:
Interactive message Response time error during calibration (time until
Key operation requesting display pH value is settled)
Functions that can be set or executed at operation level: Asymmetry potential failure
One-touch calibration (Buffer selection is manual e.m.f. slope failure
but indication stability check is automatic.) Temperature range failure
Measurement of sensor 90 % response time pH range failure
Manual calibration Glass electrode impedance failure (measured
Selection of message area display contents solution must be (50 mS/cm or more and
Temperature coefficient setting temperature 608C or less)
Auto / manual temperature compensation selection Reference electrode impedance failure (measured
Manual temperature setting solution must be 50 mS/cm or more)
Hold set / reset Half value recovery time failure
Functions that can be set or executed at setting level: 90% response time failure
pH value setting for contacts (Sl, S2, and S3) Calibrating solution temperature failure
Output range setting (mA and mV outputs)
Hold parameter setting:
Hold provided / no selection
Selection of holding either the value
immediately before or the preset value
Preset value setting (mA and mV outputs)
Cleaning parameter setting:
Auto / manual cleaning selection
Manual cleaning start / stop selection
Automatic cleaning ON / OFF selection
Cleaning period: Setting range 0.1 to 36 hours
Relaxation time: Setting range 0.1 to 10 min
Cleaning time: setting range 0.1 to 10 min
Functions that can be set or executed at service level:
8C / 8F selection
Use / no use of conversion to reference
temperature
Temperature sensor selection
Checking item setting (asymmetry potential, e.m.f.
slope, reference electrode impedance)

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 7
4. Ultrasonic Oscillator (Model PUS400G) Cleaning water outlet flow rate: Max. 30 l/min,
Combination device: (which can be regulated by the bypass valve.)
Holder with ultrasonic cleaner (PH8HS, PH8HF) Connectable unit: Holder or sensor with the jet cleaning
Connection cable are provided with holders. system
Cleaning method: Continuous ultrasonic emission Tank: With a ball tap (ball float valve for level control)
(Frequency sweep method) Effective capacity of approx. 40 liters.
Oscillation frequency: Approx. 61 to 81 kHz Pump: Suction pump with capacitor-starting-type
Output voltage: Approx. 40 to 70V induction motor. Equipped with an electro-
Power supply: 100/110/115/200/220/240 V AC610% magnetic actuator with overcorrect protec-
50/60 Hz tion.
Power consumption: Approx. 15 VA Cleaning water inlet connection:
Ambient Temperature: -10 to 50 8C G 1/2
(hood may be fitting as option) Cleaning water outlet connections:
Storage Temperature: -25 to 70 8C G 1/2 (Optional Rc 1/2 or ANSI 1/2NPT adapters are
Construction: JIS C0920 Watertight (NEMA 4 equivalent available.)
waterproof construction) Electric connection:
Material: Case: Aluminum alloy casting Plastic watertight gland equivalent to JIS A15.
Window: Polycarbonate Applicable cable outside diameter: 9 to 12 mm
Mounting bracket: Stainless steel Conduit adapter: Optional
Finish: Baked polyurethane resin (Standard) Ambient temperature: 5 to 508C
Baked epoxy resin (Option) Power supply: 100 or 200 V AC 610%, 50/60 Hz, single
Color: Case; Frosty white (Munsell 2.5Y8.4/1.2 or phase
equivalent) (Must be equal to the converter’s power supply.)
Cover; Deep sea-moss green (Munsell 0.6GY3.1/ Power consumption: 0.4 kW
2.0 equivalent) Construction: For installation outdoors (in a place safe
Mounting: (2-inch) pipe mounting, wall or rack mounting from rain, wind, and direct sunlight)
or panel mounting Color: Light gray (Munsell 2.8GY6.4/0.9 equivalent)
Cable inlet: f22.7 x 2 Pg16 watertight plastic gland Dimensions: Approx. 442 (W) x 804 (D) x 800 (H) mm
Cable / Terminal: For 7 to 12 mm, M4 Finish: Baked polyurethane
Conduit adapter: Power Supply side only (Option) Weight: Approx. 55 kg
Connection: G 1/2 or 1/2 NPT
7. Terminal Box PH8TBG
Weight: Body; Approx. 2.5 kg
Mounting bracket; Approx. 0.7 kg Used when pH converter is installed remotely from the
(Note) Output of ultrasonic oscillator changes with power pH sensor.
supply voltage or connected cable length. Ambient temperature: -10 to 50 8C
Construction: JIS waterproof
5. Solenoid Valve for Jet/Brush Cleaning PH8MV Case material: Fiberglass reinforced polycarbonate resin
Pilot kick operated, 2-port valve. Open when energized. Case color: Grayish green (Munsell 2.5GY5.0/1.0)
Electrical connections:
Fluid: Normal tap water, industrial water, or air
pH Converter side:
Operating pressure: 0 to 1 MPa
f21 mm hole (With G1/2 plastic gland)
Forward (reverse) pressure resistance: 2 MPa
pH sensor side:
Fluid temperature: Water; 5 to 608C , Air; -10 to 608C
f13 mm hole (With G1/2 plastic gland)
Cv: 4.5
With Cable (Maximum length 20 m)
Process connection: Rc 1/2
Conduit adapter (optional)
Power supply: 100/110/200/220 V AC, 50/60 Hz
Construction: IP53 8. Accessories (Purchased Separately)
Material:
Body: Bronze See Model and Suffix Codes.
Sealing: Nitrile rubber
Ambient temperature: Maximum 508C
Electric connection: G 1/2
Weight: Approx. 0.9 kg

6. Cleaning Pump/Tank Assembly PH8PU1


Pump/tank assembly for water jet cleaning. Allows tap
water to be used for cleaning, using a level-controlled tank.
Cleaning water: Normal tap water or industrial water
(when isolation from other water lines is required);
maximum pressure of 500 kPa
Cleaning water outlet pressure: Max. 300 kPa,
(which can be regulated by the bypass valve.)

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 8
MODEL AND SUFFIX CODES 1. pH Sensors
When ordering, specify model and codes, product name For selection of KCl filling or refillable type pH sensors,
and / or part numbers. refer to "Selection Criteria For pH Sensors and Holders"
on page 24, and "Table of Corrosion-Resistant Materials"
on page 25 of this manual.
Item to be specified.
PH8ERP, d KCl Refillable Type pH Sensor
1. pH Sensor
PH8EFP
Model Suffix Code Option Specifications
PH8HG, PH8HS, Code
2. Holder KCl Refillable Type
PH8HF or PH8HH PH8ERP pH Sensor
3. pH Converter PH400G
Cable Length -03 3m
4. Ultrasonic oscillator for ultrasonic cleaning PUS400G -05 5m
-07 7m
5. Solenoid valve For jet or brush PH8MV -10 10m
6. Cleaning pump/tank cleaning only PH8PU1 -15 15m
-20 20m
Only for installation of pH converter Titanium
7. Terminal box PH8TBG Solution -TN
at distance from pH sensor. Ground Tip -HC Hastelloy C
8. Accessories PH8AX
-N Always -N
T10.EPS
-T For PH200/PH400 (*1)
pH Measuring System
d KCl Filling Type pH Sensor -E For PH202 (*2)
Option -B For PH100 (*3)
Model Suffix Code Specifications
Code -G For PH450G,PH202/TB (*5)
KCl Filling Type
PH8EFP pH Sensor Style *A Style A

-03 3m Option O-ring /PF Daielperfrow (*4)


Cable Length -05 5m
*1: Mark band is shown by mark and fork terminals are used.
and KCl Tube -07 7m *2: Mark band is shown by numeral and pin terminals are used.
Length -10 10m *3: The tag which indicated the color, the sign, and the number
-15 15m is attached to the cable of a sensor.
-20 20m *4: Choose Daielperfrow when this is used in organic solvent,
Solution -TN Titanium high alkali or high temperature solution. T11.eps
Ground Tip -HC Hastelloy C *5: Mark band is shown by numeral and M3 ring terminals are used.

-TT1 For general purpose


KCl Reserve Tank (250ml solution inlet) d pH Sensor for High-Purity Water
-TT2 For medium pressure,
(*1) Model Suffix Code Option Specifications
-TT3 Big volume tank
Code
(With 500 ml tank)
-TN1 For maintenance (for TT1, TT3) pH sensor
PH8EHP for high purity water
-TN2 For maintenance (for TT2)
Cable Length -03 3m
-N Always -N
-05 5m
-T For PH200/PH400 (*3)
pH Measuring System
-E For PH202 (*4) -07 7m
-B For PH100 (*5) -10 10m
-G For PH450G,PH202/TB (*9)
-15 15m
Style *A Style A
-20 20m
Option O-ring /PF Daielperfrow (*6)
Special glass sensor /HA Glass sensor for high alkali (*7) Solution
Ground Tip -TN Titanium
Special junction /TF Teflon junction (*8)
KCl Reserve Tank -TT1 For general purpose
*1: A 50A pipe mounting bracket is supplied with TT1, TT2 and TT3. (250 ml solution inlet)
(*1) -TT3 Big volume tank
Only a supply tube, but no KCl solution, is supplied with TN1 and TN2.
(With 500 ml tank)
Since a KCl solution is not supplied with TT2 and TT3, arrange it from
-TN1 For maintenance
among accessories or auxiliary parts. (for TT1)
*2: Prepare an air pressure regulator as shown in the diagram below
when the medium-pressure reserve tank is used. -N Always -N
To pH sensor, Regulator, (to be prepared separately)
Measuring System -H For PH200/PH400 (*2)
Air Sup.
-E For PH202 (*3)
Regulator
To pH sensor (to be prepared separately) -G For PH450G,PH202/TB (*4)
*3: Mark band is shown by mark and fork terminals are used. Style *A Style A
*4: Mark band is shown by numeral and pin terminals are used.
* 1: Since a KCl solution is not supplied with TT3, arrange it from
*5: The tag which indicated the color, the sign, and the number
among accessories or auxiliary parts.
is attached to the cable of a sensor. T13.eps
* 2: Mark band is shown by mark and fork terminals are used.
*6: Choose Daielperfrow when this is used in organic solvent,
* 3: Mark band is shown by numeral and pin terminals are used.
high alkali or high temperature solution. T12.eps * 4: Mark band is shown by numeral and M3 ring terminals are used.
*7: Choose when using in high alkali solution.
*8: Choose when using in the heavily contaminated application.
*9: Mark band is shown by numeral and M3 ring terminals are used.

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 9
2. Holders
d Guide-pipe Type
Model Suffix Code Option Code Description
PH8HG Guide-pipe (*1)

Material -PV Polyvinylchloride (fluid temperature 508C or less)


-PP Polypropylene (fluid temperature 808C or less)

Style Code *A Style A

Option, Mounting Bracket (*2) /MS5 Stainless steel mounting bracket (1 set)
*1 : Pipe length: 2 m
*2 : A set of 2-inch galvanized pipe mounting bracket is provided as standard.
T14.EPS

d Submersion Type
Model Suffix Code Option Code Description
PH8HS Submersion type holder
-PP Polypropylene
Material
-S3 Stainless steel
-10 1.0m

Pipe length -15 1.5m


-20 2.0m
-25 2.5m
-30 3.0m

pH Measuring -T Always -T
System

-NN None
-S3 For ultrasonic cleaning (Transducer: SUS316) (*1)
Cleaning -TN For ultrasonic cleaning (Transducer: Titanium) (*2)
System
-HC For ultrasonic cleaning (Transducer: Hastelloy C) (*3)
-JT For jet cleaning. The solenoid valve must be specified separately.
-BR For brush cleaning. The solenoid valve must be specified separately.
-NN None
-C3 3m
-C5 5m

Cable Length for -C6 7m


Ultrasonic Cleaning -C7 10m
-C8 15m
-C9 20m

Connector for Jet -JP Rc1/2


or Brush Cleaning -NP 1/2 NPT

Style Code *A Style A


/MS1 Mounting bracket : 1 set
/MS2 Mounting bracket : 2 sets
Mounting
Bracket /MS3 Stainless steel mounting bracket : 1 set

Options (*5) /MS4 Stainless steel mounting bracket : 2 sets

Special /F1 With flange (Without Cleaning System)


Mounting /F2 With flange (With Cleaning System)
O-ring /PF Daielper (*4)

*1: General purpose (Normal pH3 to 14) *2: For salt water *3: For acid (Normal pH0 to 4)
*4: Choose Daielperfrow when this holder is used in organic solvent, high alkali or high temperature alkali.
*5: The required number of mounting bracket sets depends on the installation location and flow rate. In general,
one set is sufficient for pipe lengths of 1 meter, and otherwise two sets are required.
T15.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 10
d Flow-Through Type
Model Suffix Code Option Code Description
PH8HF Flow-through type holder
-PP Polypropylene Refer to note below
Material
-S3 Stainless steel for selection
-JPT Rc1
-NPT 1 NPT female thread

Process Connection -J10 JIS 10K 25 FF flange

-A15 ANSI Class 150 1 FF flange equivalent


(for polypropylene holder -PP)
ANSI Class 150 1 RF flange with serration
(for Stainless steel holder -S3)
pH Measuring System -T Always -T

-NN None

-S3 For ultrasonic cleaning (Transducer: SUS316) (*1)

-TN For ultrasonic cleaning (Transducer: Titanium) (*2)


Cleaning System
-HC For ultrasonic cleaning (Transducer: Hastelloy C) (*3)

-JT For jet cleaning. The solenoid valve must be specified separately

-BR For brush cleaning. The solenoid valve must be specified separately

-NN None

-C1 1m

-C3 3m

-C5 5m
Cable Length for
-C6 7m
Ultrasonic Cleaning
-C7 10m

-C8 15m

-C9 20m

-JP Rc1/2
Connector for Jet or
Brush Cleaning -NP 1/2 NPT

Style Code *A Style A

Options O-ring /MF1 Mounting bracket (stainless steel)

/PF Daielperfrow (*4)

*1: General purpose (Normal pH 3 to 14) *2: For salt water *3: For acid (Normal pH 0 to 4)
*4: Choose Daielperfrow when this holder is used in organic solvent, high alkali or high temperature alkali.
Notes:(1) Mounting bracket is generally not required when the stainless steel holder is installed in-line in a pipe
It is required where the holder is installed in a sampling rack (in which case the U-bolt included in /MF1 in not used).
(2) Criteria for material selection (-PP or -S3)
In general, polypropylene is recommended from the viewpoint of chemical resistance. 105
100
However stainless steel is recommend in any of the following cases: *
(a) The liquid contains organic reagent, oxidizing agents, etc., which can attack Stainless Steel
80
polypropylene.
* (b) The temperature/pressure correlation of the process condition falls within the hatched 60
area of the diagram shown right.
(c) The use of polypropylene is not reasonable from a viewpoint of strength or past (8C)
40
experience.
Temperature

(3) For stainless steel, normally a 3 to 14 pH value is recommended. 20 Polypropylene


(or Stainless Steel)
0 100 200 300 400 500
Pressure (kPa)
T16.EPS

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 11
d pH Holder for High Purity Water 5. Solenoid Valve
Model Suffix Code Option Specifications
Code Model Suffix Code Option Code Description
pH holder for high purity water, PH8MV Solenoid valve
PH8HH wall-mount type
Fluid -A Air
Connection -JPT Inlet; Rc1/4, Outlet; Rc1/2
ports -W Water
-NPT Inlet; 1/4NPT, Outlet; 1/2NPT
Power Supply -100 100V AC
-H Always -H -110 110V AC
voltage
Style *A Style A -200 200V AC
-220 220V AC
Option Mounting Bracket /P Pipe mounting bracket
Power Supply -50 50Hz
T17.EPS
Frequency -60 60Hz
3. Interigent pH Converter Style Code *D Style D
T20.EPS
d 4-Wire Type pH Converter
Option
Model Suffix Code Code Description
PH400G 4-wire pH Converter
Power -1 100/110 V AC, 50/60 Hz
Supply -2 200/220 V AC, 50/60 Hz
Language for -J Japanese
warning, etc -E English
A Always A
*B Style B
Mounting /U Pipe, wall mounting bracket (stainless steel)
hardware /PM Panel mounting bracket (stainless steel)
Hood /H3 Awning hood (carbon steel)
Options /H4 Awning hood (stainless steel)
/X1 Baked epoxy resin
Tag plate /SCT Stainless steel tag plate
Conduit Adapter /AFTG G1/2
/ANSI 1/2 NPT
/SPS Teflon coated SUS steel screws
T18.EPS

4. Ultrasonic Oscillator (PUS400G)


Model Suffix Code Option Code Description
PUS400G Ultrasonic oscillator for pH meter
-NN Always -NN
Application -NN General purpose
-1 100V AC 50/60Hz
-2 110V AC 50/60Hz
Power Supply -3 115V AC 50/60Hz
-4 200V AC 50/60Hz
-5 220V AC 50/60Hz
-6 240V AC 50/60Hz
Language -J Japanese
-E English
/PS Pipe mounting bracket (stainless steel)
Mounting Bracket /W Wall mounting bracket (stainless steel)
/PA Panel mounting bracket (stainless steel)
Hood /H Awning hood (carbon steel)
Options
/H2 Awning hood (stainless steel)
Special Finish /X1 Baked epoxy resin
Tag Plate /SCT Stainless steel tag plate
/AFTG G1/2
Conduit Adapter
/ANSI 1/2 NPT
/SPS Teflon coated SUS steel screws
T19.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.22-00
<<Contents>> <<Index>> 12
6. Cleaning Pump / Tank
Model Suffix Code Option Code Description
PH8PU1 Cleaning tank / pump assembly
Power Supply -3 200V AC, 50/60 Hz
-5 100V AC, 50/60 Hz
Style code *B Style B
Options /PT With Rc 1/2 adapter
Cleaning Water Inlet/Outlet /NP With 1/2 NPT adapter
Anchor volts /AN 4 L-shape M12 x 160 (SS41) bolts
Conduit adapter /APUG G 1/2
/ANSI 1/2NPT
T21.EPS

7. Terminal Box
Option Option
Model Suffix Code Specifications Model Suffix Code Specifications
Code Code
PH8TBG Terminal Box PH8TBS Terminal Box
pA Style A -A Always -A
/P Pipe mounting hardware /P Pipe mounting hardware
Mounting hardware Mounting hardware
/W Wall mounting hardware /W Wall mounting hardware
Option Length of special cable u; Standard 03 (3 m), 07 (7 m), Option Length of special cable u; Standard 03 (3 m), 07 (7 m),
(max. 20 m) / C uu (max. 20 m) / C uu 10 (10 m)
10 (10 m)
Adapter for conduit / ATBG Rc 1/2 screw thread Adapter for conduit / ATBG Rc 1/2 screw thread
construction construction
T22.eps T22-1.eps

8. Accessories
9. Consumables
Model Suffix Code Option Specifications
Code Part Name Part Remarks
Number
PH8AX Accessories for pH meter (*1)
General K9142TN One for PH8ERP, PH8EFP, PH8EHP
Calibration -L Two bottles, each containing
250ml solution (pH7 and pH4) purpose K9319NA One for PH8ERP/PF, PH8EFP/PF
Reagents -P 24 bags, each bag containing
Glass High K9142TU One for PH8EFP/HA
powder for 500ml solution
(pH7 X 12 bags and pH4 X 12 electrode alkali K9319NC One for PH8EFP/HA, /PF
bags) and two 500ml
polyethylene bottles. General K9142TH One for PH8ERP, PH8EFP
purpose K9319QA One for PH8ERP, PH8EFP/PF
Style *A Style A
High purity K9142TK One for PH8EHP
water
Option (*2) /STD Sensor stand (with mounting
Junction K9142HW One for PH8ERP/TF, PH8EFP/TF
bracket for 50A 2-inch pipe) Teflon
/KCLL KCl solution (one 250ml K9319QB One for PH8ERP/TF, /PF,PH8EFP/TF, /PF
polyethylene bottle) KCl solution
(3.3mol/l)
K9084LP Six 250 ml polyethylene bottles
/KCLP KCl powder (three bags,
250ml solution each) Buffer solution for
calibration (pH 4) K9084LL Six 250 ml polyethylene bottles
/TMP Thermometer (0 to 100 8C)
Buffer solution for
* 1: Including the following: calibration (pH 7) K9084LM Six 250 ml polyethylene bottles
Two 200ml polyethylene cups Buffer solution for
One cleaning bottle K9084LN Six 250 ml polyethylene bottles
calibration (pH 9)
* 2: Either /KCLL or /KCLP is required for PH8EFP-u-u-TT2. Powder for buffer
K9020XA 12 bags,each for preparation
solution (pH 4) of 500 ml
T23.EPS
Powder for buffer
K9020XB 12 bags, each for preparation
solution (pH 7) of 500 ml
Powder for buffer 12 bags, each for preparation
K9020XC of 500 ml
solution (pH 9)
KCl powder 8 bags, each for preparation
(for PH8EFP, K9020XU of 250ml
PH8EHP)
KCl powder 2 bags, 1 bottle of 3.3 mol/l KCl,
(for PH8ERP) K9142UT 1 syringe

(Note) The pH value of the calibrating buffer solution may vary


depending on storage conditions.
Prepare a new solution from powder for accurate instrument
calibration T24.EPS

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 13
WIRING DIAGRAMS
(1) Wiring for pH conveter
PH400G
*3 pH converter
*5
Output signal + mA L1
(4 to 20mA DC) - Power supply
L2
Output signal +
mV
(0 to 1V DC) - *5
Contact
S1
Contact input output S1 High and low
R1 alarms
PH8TBG *1 (cleaning start R2 S2
Contact
Terminal box command) output S2
*5
*2 Contact output S3 (cleaning
PH8ERP S RE S3
T1 or high and low alarms)
PH8EFP GE
PH8EHP T1 T2 FAIL Contact output FAIL (failure)
pH sensor T2 SE
RE GE
SE S

Note : External wiring connection terminal size is for M3 screw.


*4

Grounding terminal (M4 screw)


Grounding (100 or less)
*1: Terminal box is used only where pH converter is installed some distance from pH sensor (ordinary not needed).
*2: This cable is specified in the option code for the PH8TBG.
*3: Use only shielded cable with an outside diameter of 9 to 12 mm. F04.EPS
*4: Be sure to ground pH converter case grounding terminal (grounding resistance; 100 or less).
Only when the above grounding is impossible, ground at power cable side. But be sure to avoid two point-grounding.
*5: Be sure to use cable with the outside diameter of 9 to 12 mm.

Wiring to use S3 contact output for cleaning

PH400G Converter

51
S3 Power supply Solenoid valve
52

(2) Wiring for ultrasonic cleaning system

pH Meter Holder PUS400G


Ultrasonic Oscillator
Ultrasonic Transducer Cable
U1

U2

Ultrasonic L1
Vibrator
L2

Power Supply Grounding (100V or less)

F05.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 14
WIRING USING CONDUIT INSTALLATION
Notes for Conduit Installation

pH converter PH400G
Ultrasonic oscillator Terminal box
PUS400G PH8TBG

(*1): Hanging (for drip)

(*2): Bend (for water


proofing)
(*2) (*2) (*3): Conduit use is
prohibited for sensor
(*2) cable and ultrasonic
(*3) (*3) cleaning cable
(*3)
(*1) (*1)

(*1)
(a) Recommended conduit installation (adapter not needed)

(*3): Conduit use is


prohibited for sensor
cable and ultrasonic
(*4)
cleaning cable
(*4)
(*5) (*4): Conduit adapter
(*4) (*3) (*5)
(*3) (for installation)
(*6) (*5): Flexible conduit
(*6)
(*5) (*6): Union

(*3)
(*6)
F06.EPS
(b) Conduit installation using a conduit adapter

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 15
DIMENSIONS Unit : mm
KCl refillable KCl filling pH sensor for
pH/ sensor pH sensor pure water
PH8ERP-u PH8EFP-u PH8EHP-u
Medium pressure KCl reserve tank
for KCl filling pH sensor
(with mounting bracket)
(PH8EFP-u)-TT2

Rc 1 / 4
Coupling
with valve

60
Cable length (Standard

Cable length (Standard


(Standard:3m,5m,7m,
10m,15m,20m)

10m,15m,20m)

10m,15m,20m)
Cable length

;3m,5m,7m,

;3m,5m,7m,
(Standard;3m,5m,7m,
Tube length

Tube length

10m,15m,20m)
;3m,5m,7m,
(Standard
10m,15m,20m)
f60.5

263
308
88
JIS 2-inch
64

64

64
stanchion
103

103

115
f27.5 f27.5
f38 f38 f27.5
f38
Guide-pipe for submersion type (with mounting hardware)
PH8HG
General purpose KCl reserve tank
Min :110 for KCl filling pH sensor and
Max:515 pH sensor for high purity water
(with mounting bracket)
(PH8EFP-u)-TT1
(PH8EHP-u)-TT1
F07.EPS

85 95

f48.6 JIS 2-inch stanchion


Approx.115

f63

600

JIS 2-inch
stanchion (vertical)

Approx.75 Approx.45

Big volume tank KCl reserve tank


for KCl filling pH sensor and
pH sensor for high purity water.
(with mounting bracket)
f48.6 54 (PH8EFP-u)-TT3
(PH8EHP-u)-TT3

2000 Approx.105
f105
Max:1900
Min :50

Approx.151

Sensor JIS 2-inch Pipe


(f60.5) 2-f(M8 screw) holes

7060.2
60

Hole requirements
25

for wall mounting


Approx. 20 F08.EPS 60

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 16
Submersion holder — Polypropylene — (See separate drawing for mounting hardware)
Unit:: mm
PH8HS-PP

<Without Cleaner> <With Ultrasonic Cleaner> <With Jet Cleaner> <With Brush Cleaner>
(L: Holder length; 1, 1.5, 2, 2.5, 3 m)
Cable for ultrasonic cleaning
83 (Cable length: 3, 5, 7, 10, 15, 20 m) u 1/2NPT u 1/2NPT
62 42 35 (Code : -NP) 35 (Code : -NP)

44 64

105 105 u Rc1/2 105 u Rc1/2


(Code : -JP) (Code : -JP)
f34 55 55 55

f22 f22 f22

L+31

L-40 L-323 L-40 L-323 L-40 L-323

L+145 L+165 L+165

f51 55 55 55

138.5

f50 Approx.155 Approx.155 Approx.155

75 75 75 F10.EPS

Submersion holder — Stainless steel — (See separate drawing for mounting hardware)
PH8HS-S3

<Without Cleaner> <With Ultrasonic Cleaner> <With Jet Cleaner> <With Brush Cleaner>
(L: Holder length; 1, 1.5, 2, 2.5, 3 m)
Cable for ultrasonic cleaning
u 1/2NPT u 1/2NPT
83 (Cable length: 3, 5, 7, 10, 15, 20 m) 35 35
62 42 (Code : -NP) (Code : -NP)

44 64
u Rc1/2 u Rc1/2
176 176 176
(Code : -JP) (Code : -JP)

42 42 42

f 34 f 22 f 22 f 22

L+31 L+31 L+31 L+31


L-323 L-323 L-323

L+56 L+56 L+56

49 49 49

141.0

f 48.6 Approx. 200 Approx. 200 Approx. 200

75 75 75 F11.EPS

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 17
Mounting hardware for submersion type holder
UNIT: mm
PH8HS.../MS1 1 set
PH8HS.../MS2 2 sets
JIS 2-inch (OD 60.5)
<For Holder without cleaner> 68 64 Stanchion

94 90

54

Max : 515
Sensor Holder
Min : 110
600

Max : 515
Min : 110

f48.6
<For Holder with Cleaner> 95
62

600

Cleaner Holder JIS 2-inch (OD 60.5)


Stanchion

130 90

54
100
F12.EPS
Stainless steel hardware for mounting submersion type holder
PH8HS.../MS3 1 set
PH8HS.../MS4 2 sets Max : 500

Holder with cleaner Min : 236

Holder without cleaner

51 44 48.5
28

55
22 44 44 44 44 44 44 44 22

55 65

600

Cleaner holder JIS 2-inch (OD 60.5)


Stanchion
Sensor holder F13.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 18
Mounting hardware for submersion type holder Mounting hardware for submersion type holder
— Flange mounting hardware for holder without cleaner — Flange mounting hardware for holder with cleaner
— Material: Polypropylene or Stainless steel — Material: Polypropylene or Stainless steel

f210 f330 Unit: mm

JIS 10K 100 FF flange


JIS 10K 200 FF flange
(Note) (Note)
8-f19 holes 12-f23 holes

L: Holder length; 1, 1.5, 2, 2.5, 3 m

74 f290
18

f175 78
L+31
22
Min : 150
Max : L-115
Min : 405
Max : L-130

Note: Only mating dimensions are


Approx. 155
according to flange standard.

F14.EPS
Flow-through type holder with mounting hardware PH8HF

fa
(pH Sensor) 4-fD holes
b

c
e

Outlet
O-ring
fB fA
fd
165

h
11061

E
Inlet C

g
f

15061
O-ring

Model and Code A B C D E a b c d e f g h (Approx.)


PH8HF-PP-J10-T-NN-NN 125 90 14 19 - 80 9 11 60 25 30 70 250
PH8HF-PP-A15-T-NN-NN 108 79.2 14.2 15.7 - 80 9 11 60 25 30 70 250
PH8HF-S3-J10-T-NN-NN 125 90 14 19 - 70 10 6 60.5 26 26 69 245
PH8HF-S3-A15-T-NN-NN 108 79.2 12 15.7 2 70 10 6 60.5 26 26 69 245 F15.EPS

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 19

Unit : mm
Holder for high purity water PH8HH
Hole dimensions for Holder mounting

Sensor holding screw 1. Pipe mounting (2-inch pipe)

60
102 64

20

70
SAMPLE IN
180
171

14 70
Ground connection
SAMPLE OUT terminal M5

6- f10 holes 2. Wall mounting


28Max
110

70

50
50
35

Rc 1/4 73
27 56 100 100
Sample inlet
15

70 Rc 1/2 148
100 Sample outlet
1/4NPT adapter
120
1/2 NPT adapter F16.EPS

Interigent pH Converter PH400G


Hood (optional)
Option code : /Hh Four M6 screws, 8 deep

184 220
80

72

144 20
80

144 23 112

Adaptor for conduit work


(option code : /AFTG, /ANSI)
Cable inlet port (13.5 dia. holes)
36 For JIS A8 cable gland

Cable inlet port (21 dia. holes)


equivalent to JIS A15 cable gland
A B A: For output signal / contact input
B: For sensor cable
C: For power supply
Approx. 55

36
D: For contact output (S1 and S2)
E: For contact output (S3 and FAIL) Adaptor
49
C D E 38

36 36

Ground terminal G 1/2 female ( / AFTG)


(M4 screw) 1/2 NPT female ( / ANSI)
F17.eps

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 20
Pipe Mounting Bracket / Wall Mounting Bracket (option code : /U)

d
Example of bracket used for pipe mounting

188
M6, 4 screws Unit : mm

174

200

50

Nominal 50A (O.D 60.5mm) 100


mounting pipe

d
Example of bracket used for wall mounting

135 13 M6, 4 screws

224
200

35

15
70
10 mm dia., 3 holes 100 F18.eps

Panel Mounting Bracket (option code : / PM) Unit: mm

23 M6, 4 screws
12 max.(panel thickness)

100 137 +20

137 +20
178
Panel cutout dimensions
F19.eps

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 21
Ultrasonic oscillator (PUS400G)
Pipe mounting Panel/, Wall mounting
Unit : mm
202 155
162 115 48 162 1 to 10
Panel mounting bracket

50
Hood

110
(optional)

130
180

180
30

30
60
288

151
38 80 3.9 185
206

68
96
Pipe mounting bracket
(optional)
Power supply cable inlet
42 Cable gland DIN PG16 equivalent
68

Ground terminal
96

(M4 screw)
Ultrasonic transducer cable inlet
Cable gland DIN PG16 equivalent
Power supply cable inlet 2-f5.5 holes 100
42 Cable gland DIN PG16 equivalent +1.1 (for hood mounting) 7060.2
Ground terminal 156 0
(M4 screw) 5660.2
Ultrasonic transducer cable inlet 121

8
Cable gland DIN PG16 or equivalent
100

27
2-f6.6 holes 13
+1.1
173 0

7060.2 (for hood mounting)

185
138

138
5660.2
8
27

27
JIS 2-inch (OD f60.5mm) 2-f6.6 holes

15
stanchion (vertical) (for oscillator
121 mounting) 7060.2
203

JIS 2-inch pipe


stanchion (horizontal) 3-f10 holes
2-f6.6 holes
(for hood
mounting) <Panel cutout> <Wall mounting bracket>

48
F20.EPS

<Pipe mounting bracket>

General Purpose Solenoid Valve


PH8MV*D
103
72 46
102
14.5

71 2-Rc1/2 G1/2

Cautions on Installation of Solenoid Valve for Jet / Brush Cleaning


1. Do not allow a sample solution to flow backward into the solenoid valve or to be replaced with the driving fluid.
For this take relevant measures; e.g. install a check valve to prevent inverse pressure between the inlet and outlet of the solenoid
valve, or install the solenoid valve higher than the holder, especially when using the air jet/brush cleaning system.
2. Make sure to avoid the risk of corrosion of the solenoid body (bronze) and seal (nitrilel rubber) by vapor or gaseous components
generated from a sample solution, especially when using the air jet/brush cleaning system.

F21.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 22
Terminal Box PH8TBG

126 65 48 Unit: mm

30
178
75
254
60.2

Waterproof gland

Pipe mounting hardware


(example)
4-M5 screws, 9mm deep
7060.2 (for mounting) For conduit construction
Adapter (option)

<Pipe mounting hardware> <Wall mounting hardware>


100 4 - [ 5.5 hole Nominal 50A 100 4 - [ 5.5 hole
7060.2 pipe mounting 7060.2

8
8

7560.2 75
60.2
13
203 185

27
15

48 7060.2 3 - [ 10 hole

F22.eps

Sensor stand
(PH8AX-u)/STD

Sensor (Example)

JIS 50A 2-inch stanchion


(64)
115

108
F24.EPS

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 23
Cleaning Pump/Tank Assembly PH8PU1 Unit: mm
(450)
4-f15 holes

T : Tank 50
V1 : Float valve 1/2 B
V2 : Check valve 1/2 B
V3 : Ball valve
G : Pressure gauge
P : Pump
(442)

(408)
Cleaning
water inlet
Cleaning
G 1/2B water outlet
male(std.) G 1/2B
female(std.)
1/2 PVC
Pipe Teflon braided
horse
Flow diagram

Motor

Cleaning water
Electromagnetic outlet
actuator Cleaning water
inlet
Pump
(800)

(504) Height to outlet


Cable
connection Tank

(290) Height to inlet


WATER IN OVER FLOW

Overflow outlet
(16 ATS type)

(804) (138)

F25e.eps

Notes on installation of cleaning pump / tank assembly


(1) The cleaning pump / tank assemble may be installed outdoors. However, the place must be sheltered from rain, wind, and direct sunlight.
(2) Since the noise level will increase to a maximum of 100 dB (A), utmost attention should be paid to the installation environment.
(3) Connect a wired soft vinyl-chloride tube to the cleaning water outlet. Do not connect a rigid PVC pipe directly.

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 24
SELECTION CRITERIA FOR pH SENSORS AND HOLDERS

<General Overall Criteria>


(1) When any of the two conditions listed below are applicable, select a KCl filling type pH sensor and either the submersion
or flow-through type holder.
d The solution is out of the range 2 < pH < 12.
d The solution contains organic or oil in the order of a few percent.
(2) When any of the two conditions listed below are applicable, consult our salesperson.
d Strong oxidizing solutions such as aqua regia, chromic acid , hypochloric acid, perchloric acid.
d The solution contains corrosive gases (ammonia, chlorine, hydrogen sulfide).

<Individual Criteria>
s: Can be used, n:Shortens useful life, x:Cannot be used

Concentration pH * Holder
Chemical
W/V (%) (25 8C) Flow-through,
Guide-pipe
Submersion
Sulfuric acid 0.5 1.0
0.05 2.0
Hydrochloric acid 0.4 1.0
Inorganic acid

0.04 2.0
Nitric acid 0.6 1.0
0.06 2.0
Phosphoric acid 1.0 1.5
Boric acid 0.6 5.0
Carbonic acid 0.6 3.6
Chromic acid 1.2 0.8
Sulfurous acid 0.8 1.4
Acetic acid 0.6 2.8
Organic acid

Formic acid 0.5 2.3


Oxalic acid 0.9 1.0
Lactic acid 0.9 2.4
Phenol acid 0.9 5.4
Monochloroacetic acid 0.9 1.8
Calcium hydroxide 0.2 12.4
Alkali

Potassium hydroxide 0.5 12.7


Sodium hydroxide 0.4 12.9
Ammonium chloride 5
Acid salts

Aluminous water 5
Zinc chloride 5
Ferric chloride 5
Ferric nitrate 5 1.3
Basic salts

Sodium sulfite 5
Sodium carbonate 5 11.8
Sodium phosphate 5
Potassium chloride 5
Neutral salts

Sodium sulfate 5
Calcium chlorine 5
Sodium nitrate 5 8.2
Aluminum chloride 5
Hydrogen peroxide 1
Oxdizing

1 12.5
agents

Sodium hypochlorite solution


Chlorinated lime 1
Potassium bichromate 5 4.5
10
solvents

Alcohol
Organic

Organic solvent or oil


(excluding alcohol)

Note: pH values in table are calculated from dissociation constant (including measured value).
T25.EPS
CAUTION
Select the material of wetted parts with careful consideration of process characteristics. Inappropriate
selection may cause leakage of process fluids, which greatly affects facilities. Considerable care must be
taken particularly in the case of strongly corrosive process fluid such as hydrochloric acid, sulfuric acid,
hydrogen sulfide, and sodium hypochlorite. If you have any questions about the wetted part construction
of the product, be sure to contact Yokogawa.

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 25
Table of Corrosion-Resistant Materials (The data should be used for reference only)
Note: This table shows corrosion resistance for each single substance alone. If a sample contains two or more substances, then the
corrosion resistance may differ from that given in this table.

: Excellent
: Good
: Not so good
X : Unusable Sensor
Holder material Ultrasonic transducer material Seal O-ring
Sensor solution ground tip body
material material
Polypropylene SUS 316 Hastelloy C Titanium Viton Ryton Remarks
Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge

Sulfurous acid 100 20 6 30 6 30 6 30


90
5 20 5 30 X 5 30 5 30 5 30
Hydrochloric acid 5 80 5 b X 37 60
37 90 X
20 20 10 b 20 30 10 b 20 20
Inorganic acid

Chromic acid
20 40 X
10 20 14 30 X 15 43 20 40 5 20
Hypochlorous Acid
10 40 40 X
Hydrobromic acid 40 30 Strong acid
10 20 10 30 10 30 10 100 Weak acid 5 20
Nitric acid 60 X
10 80 10
57 20 57 25 X 57 30
Hydroiodic acid
57 70
3 20 6 30 5 30 5 30 90 20
Sulfuric acid
3 100 5 100 X 5 70 5 100 X 30 90
30 60 15 30 5 30 5 30 85 90
Phosphoric acid
30 100 5 b 5 b 5 60
15 80 10 b 10 b 10 b 15 30
Ammonia water
15 100 28 65 20 65 20 65
10 b 10 b 10 b 10 20
Potassium hydroxide
25 b 25 b 25 b 10 90
Sodium 20 80 20 30 20 30 20 30 10 20
Alkali

hydroxide 20 100 20 b 20 b 20 b Strong alkaliX 10 90


Sodium hydroxide, 100 93 Weak alkali 90
Sodium hydroxide9 to 11%
+Sodium chloride 15%
Potassium 5 b 5 b 5 b 5 b
carbonate 35 b 35 b 35 b 35 b
Sodium carbonate sat. 100 25 b 25 b 25 b 25 90
Zinc chloride 20 b 20 b 20 b
25 25 X 10 b
Aluminum chloride
25 25 X 25 b X
Ammonium chloride 35 40 25 b 25 b 25 b 25 90
Potassium chloride sat. 60 sat. 60 sat. 60 sat. 60 20 90
Chlorides

sat. 80 25 b 25 b 25 b 25 90
Calcium chloride
sat. 100
20 40 30 b X 30 b X 30 b 20 60
Ferric chloride
20 60
Sodium chloride, 20% 100 90 X 90 X 90 20
+ Saturated Cl2
(Electrolysis solution)
Seawater, Magnesium 24 24 24 24
chloride sat. 80 42 b 42 b 40 b 80
5 60 20 b 20 b 20 b 10 90 Polypropylene may
Nitrates Sulfates

Ammonium sulfate sometimes be eroded


sat. 30 sat. 30 sat. 30 by ammonium sulfate
crystals
Potassium sulfate 10 b 10 b 10 b 10 90
Sodium sulfate 20 b 20 b 20 b 10 90
Corrsion 90
20 b 20 b 20 b 10
Ammonium nitrate resistance
Sodium nitrate is good for 50 b 50 b
Sodium sulfite usual salts. 20 b 20 b
Hydrogen peroxide 10 30 10 30 10 30
30 90
Others

Sodium sulfide 2 60-90 X 2 60-90 15 30 5 90


20 80
Potassium bichromate 10 b 10 b 10 b
Sodium sulfide 60 80 10 b 10 b 10 90
Sodium bisulfate 10 b 10 b
20 30 X 30 30 20 X
Wet chlorine gas 40
60 X
Gases

Sea water + Saturated Cl2 95 X 95 95


Bromine gas 30 30 30 X
Hydrogen sulfide 20 20
80 30-90 80
Sulfurous acid gas
100
Note: "b" refers to the boiling point. T26.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 26

Seal O-ring Sensor


Holder material Ultrasonic transducer material body
Sensor solution ground tip material
material
Polypropylene SUS 316 Hastelloy C Titanium Viton Ryton Remarks
Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge

Acetaldehyde 20 100 30 100 20


100 20 50 25
Acetone 100 25 X 100 b
100 110
100 20
Aniline 100 70 100 25 100 90
100 100
Ether 100 20 100 25 100 20
100 70 100 25
Ethylene glycol
100 100
96 70 100 b 100 90
Ethyl alcohol
Methyl chloride 100 20 X 100 25
100 70 100 24 X 100 20
Glacial acetic acid
100 100
Glycerin 100 70 100 25
100 100
100 20 100 20
Chlorophenol 100 70
100 100 X
Xylene 100 20 X 100 20
100 20 X
Chlorobezene
100 100 X
Chloroform 100 20 X 100 b 100 b 100 b 100 90
100 20
100 70
Dioxane 100 90
100 100 X
Organic substances

Dichloroethare 100 70 X
100 20
Ethyl nitrate 100 105 100 90
100
Carbon tetrachloride 100 20 X 90 b 90 b 100 24 X
Trichloroethylene 100 20 X 100 b 100 b 100 b 100 90 X
Toluene 100 20 X 145 100 90
Benzophenone
100 20
Benzaldehyde 100 70 100 20
100 100 X 100 90 X
Benzyl alcohol
100 20 100 30 100 30 100 25 100 90
benzene
10 70 37 b 37 b 37 b
Fomaldehyde
10 100
Methylnaphthelen
100 20
Methyl ethyl ketone 100 90
70
Methyl alcohol 100 20 100 25 100 25
100 20 100 90 X
Nitrobenzene 100 70
100 100 X
100 20 10 b 10 b
Acetic acid 100 70
100 100 X
100 20 95 30 95 30 95 30 100 90
Phenol
100 100
Benzonic acid 100
100 20 100 20
Motor oil 100 70
100 100
Petroleum ether 100 20 100 20
100 20 101 100 20
Kerosene
100 70 X
10 40 50 100 50 100 50 100
Tartaric acid 10 60
10 80
100 70 100 25 100 180 100 180
Oil and fats
Carbon sulfide 100 20 X 100 25 100 25
Note: "b" refers to the boiling point. T27.EPS

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
<<Contents>> <<Index>> 27
Enquiry Specifications Sheet for PH400G pH Converter System
For enquires on the Yokogawa pH converter system, please tick (v) the appropriate box u and write down the relevant
information in the blanks.

1 General Information
Company name;
Contact Person; Department;
Plant name;
Measurement location;
Purpose of use; u Indication, u Recording, u Alarm, u Control
Power supply; V AC Hz

2 Measurement Conditions
(1) Process temperature; to Normally [8C]
(2) Process pressure; to Normally [kPa]
(3) Flow rate; to Normally [l/min]
(4) Flow speed; to Normally [m/s]
(5) Slurry or contaminants; u No, u Yes
(6) Name of process fluid;
(7) Components of process fluid;
(8) Others;

3 Installation Site
(1) Ambient temperature;
(2) Location; u Outdoors, u I ndoors
(3) Others;

4 Requirements
(1) Measuring range; u pH 0 to 14 u
(2) Transmission output; u 4 to 20 mA DC u
(3) System configuration selection; u Sensor, u Holder, u pH Converter, u Cleaning system, u Terminal box,
u Accessories
(4) Sensor cable length; u 3m, u 5m, u 7m, u 10m, u 15m, u 20m, u m
(5) Sensor operating pressure; u10 kPa or less, u Greater than 10kPa
(6) Type of holder; u Guide pipe, u Submersion, u Flow-through, u Suspension, u Angled floating ball,
u Vertical floating ball
(7) Cleaning method; u No cleaning, u Ultrasonic cleaning, u Jet cleaning, u Brush cleaning
(8) Sample temperature; u -5 to 1058C, u -5 to 1008C, u -5 to 808C
(9) Others;

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation GS 12B7C1-E 6th Edition 2012.11.10-00
Subject to change without notice