TABLE 1 Dimensions of the Proposed Antenna

EARLY DETECTION OF BREAST


TUMORS USING UWB MICROSTRIP Elements Dimensions
ANTENNA IMAGING Substrate FR4 er 5 4.4, h 5 0.8 mm
1 1 2 L 5 25 mm, W 5 16 mm
Radouane Karli, Hassan Ammor, and Bal S. Virdee Microstrip line LMl 5 13 mm
1
Electronic and Communication Laboratory EMI, Mohammed V
WMl 5 1.6 mm
University in Rabat, Morocco; Corresponding author:
radouane.karli@gmail.com Slots S15 1.5 3 1.5 mm2
2
Faculty of Life Sciences & Computing, Center for Communications S2 5 S3 5 0.5 3 6 mm2
Technology, London Metropolitan University, United Kingdom S4 5 1.6 3 4 mm2
Steps St1 5 10 3 1 mm2
St2 5 7 3 0.75 mm2
Received 4 February 2016
Patch 12 3 8 mm2
Ground Lp 5 11.5 mm
Wp 5 16 mm
ABSTRACT: This paper describes the detection of tumors in human
breast tissues using an UWB planar antenna operating at a frequency
recommended by FCC. The antenna has a measured bandwidth from tered from different locations. As the dielectric properties of
3.92 GHz to 11.32 GHz for |S11| < 10 dB. The detection technique is tumor differ from healthy breast tissue, this suggests that the
based on using the fact that the tumor’s electrical properties, in particu- reflected and the scattered signals will be different for both
lar conduction and specific absorption rate (SAR), change significantly types of tissues. The cause of this difference in electrical proper-
from those of healthy biological tissue when exposed to microwave radi-
ties in terms of relative permittivity and conductivity is attrib-
ation. The hemispherical breast model used comprised breast skin, fat
uted to the fact that normal breast tissue, although, a fatty and
and the tumor tissue. Investigation was done with the tumor located at
several positions inside the heterogeneous breast model. The patch lossy medium, is largely transparent to microwave radiation
antenna was operated in its near field by placed it in contact with the compared to lesions, containing more fluids, in terms of water
breast to enhance its sensitivity. High current density has been observed and blood, providing a higher lossy medium to microwave prop-
for various positions and sizes of the deep seated tumor while keeping agation [3].
SAR within safe limits specified by FCC of 1.6 W/Kg. A comparison The breast imaging now allows not only the detection of
technique is used to identify whether or not tumor (benign or malignant) sub-centimeter lesions, but also their characterization. The anal-
is present in the breast. The results reveal the validity of the proposed ysis of breast lesions using magnetic resonance imaging (MRI)
simple and effective technique. V C 2016 Wiley Periodicals, Inc.
is based primarily on morphological criteria (size, shape, con-
Microwave Opt Technol Lett 58:2101–2106, 2016; View this article
tour, internal decision-contrast enhancement), but also kinetic
online at wileyonlinelibrary.com. DOI 10.1002/mop.29980
(speed, current, dynamic contrast enhancement). This non-
Key words: microwave imaging; breast cancer; UWB antenna; tumor
invasive radiological technique is based on the physical phe-
tissues nomenon of nuclear magnetic resonance (NMR) [4,5]. Although
MRI is a powerful medical diagnostic technique that provides
three-dimensional images with high anatomical precision, how-
1. INTRODUCTION
ever it is very expensive to buy and maintain especially by third
Breast cancer is the most common cancer among women partic- world developing countries.
ularly in developed countries [1]. Early detection of breast can- Over the last two decades, medical imaging applications
cer is known to be a key factor in the successful treatment of based on microwaves has been developed hugely. The emerging
the disease. Since the birth of radiological techniques several microwave imaging techniques gives better results with the
methods have been developed, including microwave imaging advantages of non-ionizing, comfortable, sensitive to tumors and
[2]. This technique involves transmitting microwave signals
specific to malignancies [6]. In addition, the development of
through the breast tissue and recording the received signals scat-
UWB microstrip patch antennas is shown to reduce cluttered
data to produce highly localized images [6–8].

Figure 1 Geometry of the proposed antenna (a) Top view, and (b)
Bottom view. [Color figure can be viewed in the online issue, which is Figure 2 The photograph of the UWB antenna. [Color figure can be
available at wileyonlinelibrary.com] viewed in the online issue, which is available at wileyonlinelibrary.com]

DOI 10.1002/mop MICROWAVE AND OPTICAL TECHNOLOGY LETTERS / Vol. 58, No. 9, September 2016 2101
Figure 3 Comparison between simulation and measurements of: (a)
S11, and (b) VSWR. [Color figure can be viewed in the online issue,
which is available at wileyonlinelibrary.com]

This paper describes a compact UWB microstrip antenna for


medical imaging. In particular, the proposed antenna is used to
detect and characterize hidden tumors of various sizes inside the
breasts. The detection is based on comparing the electrical prop-
erties of healthy and malignant breast tissues when radiated with
microwave energy. The simulation results presented here reveal
the effectiveness and sensitivity of the simple antenna structure
in the detection of tumors.

2. UWB PLANAR ANTENNA


In this study a typical low-profile microstrip antenna design that
is based on rectangular radiator with a 50X feedline was chosen
because the single layer antenna structure can be accurately
modeled, and is cost effective to manufacture.

TABLE 2 Comparison Between Simulated and Measured


Reflection Coefficient
Figure 4 (a) Radiation patterns, and (b) Gain performance. [Color figure
Bandwidth for Bandwidth for Resonant
can be viewed in the online issue, which is available at wileyonlinelibrary.
S11  26 dB S11  210 dB frequency Level |S11|
com]
Simulated 3.59–13.51 GHz 3.92–11.06 GHz 4.55 GHz 218.8 dB
8.07 GHz 234 dB
9.39 GHz 225.16 dB The rectangular patch design, in Figure 1, was designed as
Measured 3.59–11.75 GHz 3.92–11.32 GHz 5.43 GHz 221.39 dB
described in Ref. [9] with slots to increase its bandwidth. A
7.24 GHz 229.13 dB
50X feedline was used, and the antenna was excited through the
10.55 GHz 225.34 dB
SMA connector. The antenna design was implemented on FR4

2102 MICROWAVE AND OPTICAL TECHNOLOGY LETTERS / Vol. 58, No. 9, September 2016 DOI 10.1002/mop
TABLE 3 Comparison of Previous Designs With the Proposed
Antenna

Bandwidth (GHz) for Antenna


Antenna S11 210 dB or VSWR<2 size (mm2) Gain (dBi)
[12] 3–10.26 34 3 36 Not reported
[14] 4–9 30 3 30 2–6
[15] 1.15–4.4 75 3 75 2–8
This work 3.92–11.32 16 3 25 2.25–4.56

Figure 6 Current density distribution in the tissue without tumor, (a)


skin, and (b) breast fat. [Color figure can be viewed in the online issue,
which is available at wileyonlinelibrary.com]

nuity in the SMA connector to the antenna feedline introduced


Figure 5 (a) Breast model with tumor, and (b) Close-up of antenna
by the solder and fabrication tolerance.
placed in contact with the breast. [Color figure can be viewed in the
online issue, which is available at wileyonlinelibrary.com]
Figure 4 shows the antenna’s radiation patterns at various
spot frequencies and gain performance. The antenna radiates
omni-directionally in the E- and H-planes, and is stable over its
epoxy substrate, and the optimized dimensions of the antenna
operating frequency range. The antenna gain varies from 2.25
are given in Table 1. Photograph of the antenna is shown in
dBi to 4.55 dBi over the operating band with an average radia-
Figure 2.
tion efficiency of 96%. This antenna structure was employed to
The simulated and measured return-loss and VSWR of the
detect tumors by microwave imaging.
optimized antenna is shown in Figure 3. The antenna has an
Table 3 presents a comparison between the performance of some
operating frequency band of 7.6 GHz for S11 better than 10 dB,
recently developed UWB antennas and the proposed antenna. The
which covers the entire UWB frequency range as specified by
proposed antenna has advantages of a wider impedance bandwidth,
FCC. Figure 3 shows across its operating frequency the VSWR
is 2:1 indicating good impedance match. compact size, and good gain performance.
Table 2 compares the simulated and measurement reflection
3. TUMOR DETECTION
coefficient results. It shows the bandwidth results are compara-
ble. The difference in performance is attributed to any disconti- Several models representing breasts have been reported by
researchers [10,11]. In this study we used a simple hemispherical
model to represent the breast, as shown in Figure 5. The most
common size of the breast was used here. The dielectric proper-
TABLE 4 Dielectric Properties of the Breast Model at 6GHz ties of the breast are shown in Table 4, where r is the conductiv-
Conductivity r (S/m) Permittivity er ity of tissue and er is the dielectric permittivity [12]. The typical
size of the tumor is in the range from 0.2 cm to about 1.5 cm
Skin 1.1 39 [11–14]. In this study, we have modeled a spherical tumor with a
Breast tissues 0.59 4.49
radius of 0.3 cm and 0.5 cm in order to compare our results with
Tumor 4 50
other researchers work reported in Refs. 12,14,15.

DOI 10.1002/mop MICROWAVE AND OPTICAL TECHNOLOGY LETTERS / Vol. 58, No. 9, September 2016 2103
Figure 7 SAR distribution in tissue without tumor, (a) Skin, and (b)
Breast fat. [Color figure can be viewed in the online issue, which is
available at wileyonlinelibrary.com]

The hemispherical breast model has a radius of 5.5 cm. The


breast is composed of outer skin of 0.2 mm thickness, and inner
fatty breast tissue of 5.3 cm radius. The tumor is located inside
the breast. The proposed antenna is placed in contact with the
breast, as shown in Figure 5, to increase its sensitivity. This is
because the breast skin has electrical properties comparable to
the breast tumors and when the antenna is placed away from the
breast most of the signal is scattered from the breast skin. This
results in masking of the reflections from the tumor. The breast
was modeled with Ansys HFSS, which is a 3D electromagnetic
simulation tool.
The signal reflected by a biological tissue depends on the
density of induced currents in and over it. We can use this
information to analyze whether or not a breast has a tumor by
comparing it with a healthy breast scenario. The breast model is
exposed in the near field of the antenna, radiating a permissible
125 mW at 6 GHz. The current density and specific absorption
rate (SAR) in the skin and fatty tissue of the breast with no Figure 8 Current density distribution in tissue with 3 mm tumor, (a)
tumor are shown in Figures 6 and 7. The SAR value refers to Skin, (b) Breast fat, and (c) Tumor. [Color figure can be viewed in the
power averaged over 10 g of tissue. Figures 8 and 9, show the online issue, which is available at wileyonlinelibrary.com]
current density and SAR in the case of a breast with a tumor.
Figures 8 and 9 show higher current density on the skin and From Figure 8 it is clear that peak volume current density on
tumor with safe peak SAR values of microwave radiation on the tumor of 3 mm radius is as high as 9.94 A/m2 at position (4, 2,
breast fat and skin. The tumor was positioned at location 0.75). However, the current density is 3.65 A/m2 at position (4,
defined by co-ordination x 5 4, y 5 2, and z 5 0.75. The spheri- 4.5, 0.75), and 0.84 A/m2 at (4, 4, 4.75). SAR for tumor of 3 mm
cally modeled tumors of radii 3 mm and 5 mm were analyzed radius are 9.05 3 1023, 2.84 3 1023 and 1.64 3 1024 W/Kg,
in the model at 6 GHz. respectively, at positions (4, 2, 0.75), (4, 4.5, 0.75) and (4, 4, 4.75),

2104 MICROWAVE AND OPTICAL TECHNOLOGY LETTERS / Vol. 58, No. 9, September 2016 DOI 10.1002/mop
Figure 9 SAR distribution in the tissue with 3 mm tumor, (a) Skin, (b) Breast fat, and (c) Tumor. [Color figure can be viewed in the online issue,
which is available at wileyonlinelibrary.com]

TABLE 5 Peak Current Densities and Peak SAR Values on the Tumor of Different Radii at Different Locations at 6 GHz

Tumor Radius

0.3 cm 0.5 cm
2 2
Tumor x,y,z position in cm r (A/m ) SAR (W/Kg) r (A/m ) SAR (W/Kg)
23
4, 2, 0.75 9.94 9.05 3 10 9.46 7.89 3 1023
4, 4.5, 0.75 3.65 2.84 3 1023 3.23 1.23 3 1023
4, 4, 4.75 0.845 1.64 3 1024 2.84 8.47 3 1024

TABLE 6 Peak Current Density and Peak SAR Value on Skin and Breast Fat at Different Locations at 6 GHz for a Tumor Radius of
0.3 cm

Tumor radius 5 0.3 cm

Skin Breast fat

Tumor x,y,z position in cm r (A/m2) SAR (W/Kg) r (A/m2) SAR (W/Kg)


22
4, 2, 0.75 19.13 5.71 3 10 1.15 4.8 3 1023
4, 4.5, 0.75 19.55 5.84 3 1022 1.15 4.81 3 1023
4, 4, 4.75 19.82 5.60 3 1022 1.19 4.9 3 1023

which is quite small and within safe limit defined by FCC of 1.6 the position of the tumor inside the breast has an effect on its
W/Kg). detection. In particular, the further the antenna is from the tumor
The peak current densities and peak SAR values on the the lower the detection, which justifies the need for the use of
tumor of radii 0.3 mm and 0.5 mm at various positions is given an array antenna.
in Table 5. The antenna is placed in contact with the breast skin Tables 6 and 7 show the values of peak volume current den-
as explained above to increase its sensitivity. The table shows sity, r, and peak SAR value on skin and breast fatty tissue for

TABLE 7 Peak Current Density and Peak SAR Value on Skin and Breast fat at Different Locations at 6 GHz for a Tumor Radius of
0.5 cm

Tumor radius 5 0.5 cm

Skin Breast fat


2 2
Tumor x,y,z position in cm r (A/m ) SAR (W/Kg) r (A/m ) SAR (W/Kg)
22
4, 2, 0.75 21.12 6.37 3 10 1.21 5.27 3 1023
4, 4.5, 0.75 20.69 5.37 3 1022 1.24 5.38 3 1023
4, 4, 4.75 20.65 6.76 3 1022 1.213 5.57 3 1023

DOI 10.1002/mop MICROWAVE AND OPTICAL TECHNOLOGY LETTERS / Vol. 58, No. 9, September 2016 2105
TABLE 8 Peak Current Density and Peak SAR Value of Breast No Tumor Case at 6 GHz

No Tumor

Skin Breast fat


2 2
Tumor x,y,z position in cm r (A/m ) SAR (W/Kg) r (A/m ) SAR (W/Kg)
22
4, 2, 0.75 11.67 2.36 3 10 0.729 1.94 3 1023
4, 4.5, 0.75 9.96 1.67 3 1022 0.734 1.6 3 1023
4, 4, 4.75 9.96 1.67 3 1022 0.733 1.6 3 1023

different sizes of tumors in the breast and different position. 7. R. Karli and H. Ammor, Miniaturized UWB microstrip antenna with
The tables confirm the larger the tumor the higher the magni- sub-centimeter lot for detecting malignant tumors by microwave
tude of the current density and SAR value. In both cases the imaging, Int J Microw Opt Technol 9 (2014), 214–220.
8. J.C.Y. Lai, C.B. Soh, E. Gunawan, and K.S. Low, UWB microwave
SAR values are well within safe limits defined by FCC (1.6 W/
imaging for breast cancer detection- experiments with heterogeneous
Kg). breast phantoms, Prog Electromagn Res M 16 (2011), 19–29.
Table 8 shows the values of the peak volume current density 9. R. Karli, H. Ammor, J. Therzaz, M. Cha€ıbi, and A. Mediavilla
and peak SAR value on skin and breast fatty tissue for the sce- Sanchez, Design and construction of miniaturized UWB microstrip
nario of the breast with no tumor. It is evident that for a non- antenna with slots for UWB applications, Microw Opt Technol Lett
malignant case the current densities and SAR values over the 57 (2015), 460–463.
skin and breast fat are much lower. 10. M.J.A. Klemm, D. Leendertz, I.J. Gibbins, A. Craddock, A. Preece,
This study shows the proposed antenna can be used to effec- and R. Benjamin, Microwave radar-based breast cancer detection:
Imaging in inhomogeneous breast phantoms, IEEE Antenna Wireless
tively detect malignant tissue inside breasts. This study is proof
Propag Lett 8 (2009), 1349–1352.
of concept which needs to be verified experimentally. We 11. M.A.S. Banu, S. Vanaja, and S. Poonguzhali, UWB microwave
believe with an array antenna the sensitivity for detection can detection of breast cancer using SAR, Int J Adv Electric Electron
be greatly enhanced. Eng 2 (2013), 87–92.
12. J. Valvano, Tissue thermal properties and perfusion, In: optical-ther-
mal response of laser-irradiated tissue, Chapter 14, Plenum, New
York, 1995.
4. CONCLUSION
13. M. Lazebnik, et al. A large-scale study of the ultra wide band micro-
The study presented in this paper confirms that a planar UWB wave dielectric properties of normal, benign and malignant breast
antenna can be used in a medical imaging system to detect tissues obtained from cancer surgeries, IOP Publishing, Phys Med
tumor tissues growing inside the human breast. The study is Biol 52 (2007), 6093–6115.
based on current densities and SAR in the various tissues consti- 14. H.B. Lim, N.T. Nhung, E. Li, and N.D. Thang, Confocal microwave
tuting the breast tissue including breast skin and fat. The hetero- imaging for breast cancer detection: Delay-multiplyand- sum image
reconstruction algorithm, IEEE Trans Biomed Eng 55 (2008), 1697–
geneous breast model was radiated with 125 mW in the near
1704.
field. The SAR in the breast tissues are well within safe limits 15. G. Bindu, A. Lonappan, V. Thomas, C.K. Ananadanand, and K.T.
specified by FCC. The results demonstrate a single miniature Mathew, Active microwave imaging for breast cancer detection,
planar antenna has sufficient sensitivity to detect tumors when Prog Electromagn Res PIER 58 (2006), 149–169.
the antenna is in contact with the breast skin. We believe that
with an array antenna the detection sensitivity will be greatly C 2016 Wiley Periodicals, Inc.
V
enhanced for accurate detection and location of tumors in breast
tissues. Compared with other sophisticated imaging techniques
the proposed technique is effective and relatively easy to HIGH SENSITIVE MICROWAVE SENSOR
implement. BASED ON SYMMETRICAL SPLIT RING
RESONATOR FOR MATERIAL
REFERENCES CHARACTERIZATION
1. S. Raghavan and M. Ramaraj, An overview of microwave imaging Rammah A. Alahnomi,1 Z. Zakaria,1 E. Ruslan,2
towards for breast cancer diagnosis, Progress In Electromagnetics Amyrul Azuan Mohd Bahar.,1 and S. R. Ab Rashid1
1
Research Symposium Proceedings, Moscow, Russia, August 2012, Centre for Telecommunication Research and Innovation (CeTRI),
pp. 627–630. Universiti TeKnikal Malaysia Melaka (UTeM), Durian Tunggal, Melaka
2. S.A. AlShehri and S. Khatun, UWB imaging for breast cancer detec- 76100, Malaysia; Corresponding author: zahriladha@utem.edu.my
2
tion using neural network, Prog Electromagn Res C 7 (2009), 79–93. Faculty of Technology Engineering (FTK), Universiti TeKnikal
3. M. Kahar, A. Ray, D. Sarkar, and P.P. Sarkar, An UWB microstrip Malaysia Melaka (UTeM), Durian Tunggal, Melaka 76100, Malaysia
monopole antenna for breast tumor detection, Microw Opt Technol
Lett 57 (2015), 49–54. Received 14 February 2016
4. A. Shahira Banu, S. Vanaja, and S. Poonguzhali, UWB microwave
breast cancer detection using SAR, Int J Adv Electric Electron Eng ABSTRACT: A simple novel microwave planar sensor based on sym-
2 (2013), 87–91. metrical split ring resonator (SSRR) has been applied for the first time
5. P.K. Priya and S. Poonguzhali, Detection of breast cancer using to measure the dielectric permittivity of materials under test (MUT). The
microwave absorption loss, International Conference on Electronics results achieve narrow resonance with high Q sensitivity of 407 which
and Communication Engineering, 20th May 2012, pp. 161–164. makes it great promising applications for biosensors in the future. Three
6. E.C. Fear, S.C. Hagness, P.M. Meaney, M. Okoniewski, and M.A. practical materials have been used as standards to validate the sensor.