1. Selecting the Correct Instrument
hand instrument ISO 30 goes hand instrument ISO 20 goes
No
passively* to working length passively* to working length
Yes Yes No
R50 R40 R25 R25
*Passively means that the instrument goes directly to working length with a gentle watch winding
movement (small right left rotations) but without filing action.
1 2 3 4 5 6 7
2. Electronic Length Determination
R40 Medium or Wide Canals Before Starting Preparation with R40 or R50:
Before preparation, check the working length by using a C-PILOT® File or a
K-File and an apex locator. Set the stopper at that length. After the coronal
and middle thirds of the canal have been prepared, the working length
R50 should be re-checked.
1 2 3 4 5 6 7
3. Preparation Step by Step
Ensure you have achieved a straight line access to the root canal entrance.
3x
1. Place irrigant in the 3. Move the instrument in a slow in- 4. Clean the debris 5. Irrigate the canal. 6. Make sure the canal
access cavity of the and-out pecking motion. The ampli- from the flutes in is free to approx. 3 mm
root canal. tude of the in-and-out movements the Interim Stand. beyond the prepared
2. Introduce the should not exceed 3 mm. Only very canal section with an
RECIPROC® Instru- light pressure should be applied. The ISO size 10 C-PILOT®
ment into the canal. instrument will advance easily in the File.
Press the motor foot canal. One in-and-out movement =
pedal when orifice is 1 peck. Remove the instrument In this way, continue root canal preparation with RECIPROC®
reached. from the canal after 3 pecks. until approx. 2/3 of the working length has been reached.
1 2 3 4 5 6 7
3. Preparation Step by Step
After approx. 2/3 of working length (WL) has been reached with R25
Nein
ISO 10 hand instrument used for length deter- ISO 10 hand instrument used for length
No
mination goes to WL without being pre-curved determination only goes to WL pre-curved
Yes
Glide path up to ISO 15
Hand instrument ISO 15 goes to WL
without being pre-curved
R25 to full WL R25 to full WL
1 2 3 4 5 6 7
4. Retreatment with R25 – Step by Step
Retreatment of gutta-percha and carrier-based fillings
1. Removal of gutta-percha in the 2. Use R25 as described in the 3. Use a brushing motion against
coronal third e. g. with a Gates step by step instructions (steps lateral walls to remove residual
Glidden drill, an ultrasonic instru- 1-5) until working length has obturation material.
ment such as VDW.ULTRA® or been reached. If resistance is
depending on the consistency encountered, do not apply
of the gutta-percha directly pressure. Remove the instru-
After reaching working length with R25,
with R25. Use a drop of solvent ment from the canal, re-apply use R40 or R50 for an increased apical
(e. g. eucalyptus oil) as required. solvent and try again. enlargement, as necessary.
1 2 3 4 5 6 7
5. Tips
by
or to achieve an easier advancement of the instrument.
cation
2. Clean the instrument, irrigate the canal and check that the canal is not
blocked after 3 pecks.
cipro
Re
Δ Never apply pressure if resistance is encountered, instead repeat point 2 above!
Δ Pull the RECIPROC® Instrument out of the canal after 3 pecks or when
resistance is encountered!
Δ Use the instruments in reciprocation with the VDW RECIPROC® motor only!
1 2 3 4 5 6 7
6. Product Overview
Instruments
Paper Points
VW000270 Rev.3/15.07.2015
Gutta-Percha
GUTTAFUSION®
for RECIPROC®
Obturators
VDW GmbH
www.RECIPROC.com Bayerwaldstr. 15 • 81737 Munich • Germany