You are on page 1of 4

Analysis of Infrared Spectrum Characteristic and

Variation Trend of SF6 PD Decomposition


Fan LIU, Degang GAN, Siyu ZHOU, Can HU, Xiaoxing ZHANG
Ping LIU State Key Laboratory of Power Transmission Equipment &
Sichuan Electirc Power Test & Research Institute System Security and New Technology,
Chengdu, P.R.China Chongqing University,Chongqing, P.R.China
Liufan2003@yahoo.com.cn

Abstract—SF6 decomposition components have a close method at home and abroad were HF, SO2, SOF2 test tube,
relationship with the insulation defects in GIS. And the inner test accuracy was up to ul / l level, but its stability was
insulation state could be evaluated by monitoring the internal
gas decomposition components. Fourier transform infrared
vulnerable to temperature/humidity, existed cross-
spectroscopy (FTIR) and gas chromatography (GC) were used interference, HF had corrosion to test tube,so it can not be
to detect the SF6 partial discharge decomposition products. used for partial discharge on-line monitoring; Fourier
The sample interval was 12 hours, total discharge time was 96 transform infrared absorption spectrometry had fast
hours. It’s found that the components were varied in detection speed, multi-component detection and high
discharge time. The gas were detected by FTIR , such as SOF2 precwasion, strong anti-interference ability.Absorbence and
、 SO2F2 、 SO2 、 SOF4 、 SF4 , different gas kinds and the component concentration were in a linear relationship etc,
trend of concentration changes can be detected in different so it can be effectively applied to real-time on-line
periods. The gas kinds which the FTIR detected were more monitoring [5].
than that of GC. Then, Quantitative analysis was finished by
using the absorbance value of each absorption peak. The In this paper, the main components of SF6
paper found that there were different components and gas decomposition gas were SOF2, SO2F2, SF4, SOF4, SO2, SF6
formation rates in different discharge time, so the method gas discharge decomposition test device can be used for
could be used for real-time online monitoring of GIS. partial discharge decomposition, simulated metal
protrusions defects generated by PD, made the
I. INTRODUCTION decomposition of components for infrared absorption
Gas-insulated switchgear had compact structure, high spectrum. Through each component gas infrared spectral
reliability, so it has been widely used. In the manufacturing absorption characteristics of different periods, analyzed the
and assembly process left some potential insulation defects characteristics of SOF2 concentrations and the association
which difficult to discover,these defects caused the different between discharge time, made a foundation for SF6
levels of partial discharge within GIS.The energy of partial discharge decomposition of infrared real-time on-line
discharge can make SF6 gas decomposed,produced SF4, SF3, monitoring system.
SF2, S2F10 Low-fluoride sulfide, which reacted with the
trace moisture, oxygen in the SF6 gas and generated SOF2, II. INFRARED DETECTION PRINCIPLE
SOF4, SO2F2, SO2, HF, H2S and other compounds[1-2]. The All molecules were moving, the molecules obey the
active gas PD (partial discharge, PD) generated can rules of quantum mechanics[6]. In accordance with quantum
accelerate insulation aging and corrosion of metal surfaces, mechanics, Born-Oppenheimer approximation of molecular
which may eventually lead to failure of GIS. Therefore, in energy was:
international gas component monitoring was proposed to
control the content of corrosive gas, determined the overall E = E f + Er + Ev + Ee
level of PD and inferred causes of partial discharge [3]. (1)
Currently ,both at home and abroad for SF6 gas In which translational motion was continuous, while the
discharge detection methods were mainly used gas rotation, vibration, and electronic movement have been
chromatography, test tube method and infrared absorption quantized. When infrared light through the sample, the
spectroscopy [4]: gas chromatography can detect SOF2, sample selected absorption of infrared light, the selection
SO2F2, SO2, and CF4, detection accuracy can reach ul / l was:
level, but the chromatographic characteristics of injection ΔE = E2 − E1 = =ν (2)
determined the detection time was longer, so it can not do
In which = =6.624 × 10
continuous on-line monitoring, Chromatographic column -34
J .s , v was the photons’
which used for a period of time need clean, such a feature frequency.
determine chromatography technology with high
requirements of the environment,it was not suitable for on- According to the election laws of quantum mechanics
line monitoring applications;, the only practical test tube laws, if and only if the molecular dipole moment produced a

978-1-4244-8286-3/10/$26.00 ©2010 IEEE 409


change in infrared absorption, such a molecule known as the Figure 1. Test device of SF6 discharge decomposition Uu
infrared active molecules. Molecular rotational absorption
of infrared light section was far-infrared, the vibration The author used SF6 gas decomposition test device to
absorption of infrared light section was mid-infrared, multi- make SF6 decomposition of partial discharge testing[7], test
atom molecule existed pure rotational absorption spectra, connection shown in Figure 1.
but the pure vibrational absorption spectrum was not The experiment used Needle - plate electrode defect
available, the vibration absorption accompanied by rotation model, the electrode spacing was 10mm, making the
absorption. In accordance with the Maxwell - Boltzmann experimental device vacuumized and filling three
dwastribution law, most of the vibrational level transitions atmospheric pressure SF6 gas, the applied voltage was
were from the electronic ground state n=0 to n=1 transitions, 24kV, discharge time was 96 hours, gas production interval
it can be drawn pairs of atomic and molecular vibration of was 12 hours.
the classical equation. B. Infrared Detection and Analysis
1 k (3)
ν =
1
2π c μ TABLE I. CLASSICAL IMBIBITION WAVE NUMBER OF SF6 AND ITS
DECOMPOSITION PRODUCTS
In which v1 was the wave number, c was the speed of
Gas species Wave number/ cm-1cm-1
light, k was the vibration force constants, u was the SOF2 530,1330,1340
equivalent quality. SO2F2 539,544,552
SOF4 752
Many of the atom molecule infrared absorption formula
SO2 1167,1360
was: SF4 746
CF4 1280,1283
V v − r = V v + B [( J ' + 1) − J ( J + 1)] (4) HF 3644
H2S 2625
In which vv-r was the vibration - rotation wave number, SF6 610,946,1270,1595,1720
B was the rotational constant, J was the rotational quantum
number. When the group vibrated, if the dipole moment Using Nicolet6700 Fourier transform infrared
change paralleled to the axis of symmetry group, spectrometer, Specac 20-meter-long optical path for
then ΔJ = J ' − J = ±1 .if the dipole moment change infrared analysis of gas pool. H2O and CO2 had a great
perpendiculared to the axis of symmetry group, then, impact to infrared absorption spectra ,the mathod to avoid
ΔJ = 0, ±1 . Due to ΔJ had different values, a few their effects in the measure was: the first making the tank
absorption lines occured around the same absorption, the cacuumized and then passing into the vacuum with
spectra obtained was not an absorption line, but rather an nitrogen, applied again and again , after the above operation
absorption band. repeated then led to the decomposition of SF6 component.
Gas tank vacuum limit was 4kpa, residual gas was the
Infrared spectra of gas species detection can take nitrogen, the gas tank sampling atmospheric pressure was
advantage of Lambert - Beer law to achieve quantitative 100kpa, resolution was 0.5cm-1, scanning frequency was
terms, the expression was as follows: 32, the room temperature was 19oC, humidity was 27%.
Since SF6 gas has high concentration, and its had
A = a×b×c (5)
similar properties with its decomposition , so spectral
Where A was the absorbance value, a was the absorption overlap was more serious. To be able to detect small
coefficient, b was a measure of the optical path, c was the amounts of decomposition of gas components, the position
gas concentration. The measuring principle of Fourier must be selected in the absence of SF6 absorption bands for
transform infrared spectrometer was analysis. SF6 and its decomposition products were mainly
dwastributed in the absorption peak of 2000cm-1 ~ 480cm-1,
I0 (6) the specific circumstances was shown in Table I[8] .
A = lg
I
From the Table 1 we can see SF6 had a lot of absorption
In which I0 was the background light intensity, I was the peaks, multi-infrared absorption spectra was the result of
sampling intensity. joint action by molecular rotation and vibration, from the
formula (4) we can know that the multi-cellular spectra was
III. EXPERIMENT AND ANALYSIS an absorption band rather than an absorption line, in the
A. Experimental device above experiment, intensive absorption bands occured at a
few wave numbers, the basic analysis of other components
can not be used, it must avoid the detection of SF6
absorption band.

C. FTIR spectra and concentration trend analysis


In the experiment, first using infrared absorption
spectroscopy analyzed pre-discharge decomposition of SF6
gas, as compared spectra; then collected post-discharge
decomposition of SF6 gas for various time periods, made
infrared spectrum analysis of the components.Made
comparative analysis of the two spectra overlap,the result
was shown in Figure 2, what we can find in the spectra of
the two peaks in the situation was different. Specifically

410
expressed in SF6 decomposition of the infrared absorption and the decomposition of component species also
spectra of the absorption peak increased. increased, new SOF4 and SO2 were the major components
of SF6 discharge decomposition, according to the
absorption peak of the peak of SO2 we can determine the
concentration of the gas.
Using OMNIC softwwere can read the absorbance peak
of the absorption peak ,then we can make table 2.Currently
the decomposition components of SF6 in the absorption peak
of the absorption’s coefficient was unknown , therefore
according to the absorption peak of each decompositon
Figure 2. The infrared spectroscopy of pure SF6 and its decomposition
component we can not use Lambert - Beer law for accurate
products quantitative calculation. However, infrared absorption
spectroscopy had good linear relationship , the absorption
In order to overcome the impact of SF6, this paper used coefficient of gas molecules was only relevant to pressure,
the spectrum of geographical segment analysis method, that temperature, In order to maintain the above conditions the
was: extracted the clear difference between the absorption same in this experiment , so in equation (5), the absorption
peaks of absorption bands from the several sub-plans. Each coefficient a was a constant volume, so the decomposition
sub-graph corresponded to 1~2 kinds of characteristics of of component concentration trend analysis was feasible.
decomposition of gas composition. From Table II we can find with different discharge time ,
Removed the SF6 absorption band, made decomposition the components and concentration were also different. After
of components compared with the infrared absorption 12 hours discharge we can detect SOF2, SO2F2, SF4, with 36
spectra of pure SF6, obtained the decomposition products hours discharge we can detect SO2, while the SOF4 only
absorption spectra of the five sections:they were:529 cm-1 ~ after 72 hours discharge can be detected. Although after 36
560 cm-1; 720 cm-1 ~ 760 cm-1; 1096 cm-1 ~ 1104 cm-1; 1156 hours discharge we can detect SO2, its concentration change
cm-1 ~ 1186 cm-1; 1320 cm-1 ~ 1360 cm-1, as shown in little after 60 hours discharge, but after 84 hours discharge,
Figure 3,4. Detected gases were SOF2, SO2F2, SF4, SOF4, its concentration increased rapidly. The total trend SO2F2
SO2. peak was to increase, but compared with SOF2 and SO2 ,its
trend change gently. Based on the above phenomenon we
can use the gas production rate as discharge time, gas
production rate per unit time was defined as the increasing
amount of gas species concentration, in this paper c was
used to indicate the quantity:

c =
∂A
=
A i 2 − A i1 (7)
∂t Δt

a: SOF2,b: SO2F2,c: SF4 In which A was the absorbance value, Ai 2 was the second
Figure 3. The infrared spectroscopy of conponents after 12 hours
discharge
measured absorbance value, Ai1 was the first time measured
absorbance value Δt was the sampling time interval.
Observed in Table II we can easily find after 96 hours
discharge ,the absorption peaks were not saturate, some
components of the gas production rate continues to increase,
which indicates that the longer discharge,the more intense
decomposition ,the greater impact of the insulation.
According to IEC 60480 standards available we can find
:SOF2, SO2F2, SO2 were stable decomposition product of
PD .For the experiment we knew:the absorption peak of
SOF2 was 530 cm-1, SO2F2 was 544 cm-1, SO2 was 1167 cm-
1
.the absorption peak affected by other peaks was small, the
peak was sensitive to the concentration changes of
substances . the peak concentration can be characterized by
peak change, so selected them to make a preliminary
quantitative analysis. With curve-fitting we can obtain more
a: SOF2,b: SO2F2,c: SF4,d: SOF4,e: SO2 than three kinds of gas component concentration with time-
Figure 4. The infrared spectroscopy of conponents after 96 hours varying trends in discharge.The trend was shown in Figure
discharge 5-7, SOF2 increased slowly before 60 hours dicharge, then
the absorption peak increased rapidly as the the discharge
Figure 3 was the infrared spectroscopy of conponents time raised. The SO2F2 production rate was high within 24
after 12 hours discharge, this time the component which hours discharge, followed by up to 84 hours discharge, gas
we can detect only SOF2, SO2F2, SF4 three kinds. Figure 4 production rate changed little, after discharge 84 hours
was the infrared spectroscopy of conponents after 96 hours discharge the gas production rate increased once again; SO2
discharge, this time we can detect five kinds of discharged to 84 hours, gas production rate changed little,
components. From the diagram we can see :as the discharge component concentration was low, but after 84 hours
time increased, the absorption peak of the peak increased, discharge gas production rate increased suddenly ,thus

411
according to the law we can use SO2 concentration and gas Figure 8 The infrared spectroscopy of SF6 decomposition products by
production rate to determine the defects’ discharge time. experiment and IEC standard drawing

IV. CONCLUSION
In this paper, using Fourier transform infrared absorption
spectrometry and gas chromatography analyed SF6
decomposition of component species , the concentration of
each component trends and made the following findings:
(1)The infrared absorption spectrometry and gas
chromatography test results were similar, but the
Figure 5 Figure 6
components which the former detected was more than the
latter ,FTIR can detect more intuitive and judge under the
spectrum of qualitative changed in component
concentration, its also had strong anti-interference ability, so
this mathod can be used for real-time online monitoring.
(2)SF6 with different discharge time had different
components, each component was not the same
concentration change rate ,based on these characteristics we
can determine the extent of discharge within GIS.
(3)Detected SOF2, SO2F2, SF4, SOF4, SO2 five kinds of
Figure 7 the variation of SO2’s absorbance with time gas components,SOF2, SO2F2, SO2 had strong regularity
,which can be used as discharge decomposition of the
Figure 5 shows the variation of SOF2’s absorbance with characteristics of SF6 gas, according to their changes which
time. Figure 6 shows the variation of SO2F2’s absorbance can be diagnosed GIS within the state of the insulation.
with time. Figure 8 shows the infrared spectroscopy of SF6
decomposition products by experiment and IEC standard
drawing, we can easily find the infrared spectroscopy of SF6 REFERENCES
decomposition products obtained by experiment was fully [1] GB/T18867-2002 Gases for electronic industry-sulfur
conswastent to the peak position of each component which hexafluoride[S],National Bureau of standards promulgate, 2002.
IEC standard layout, there were a number of small [2] Zhang Xiaoxing,Yao Yao,Tang Ju etal. Actuamity and perspective of
interfering peaks which affected by micro-water actions, proximate analysis of SF6 decomposed products under partial
discharge[J]. High Voltage Engineering,2008, 34(4):664-669.
they were also conswastent to standard drawing.
[3] F.Y.Chu. SF6 Decomposition in gas-insulated equipment. IEEE
trans. Electrical Insulation[J],1986,21(5):693-725.
[4] Huang Xiaohui. Diagnoswas defect of 500kV TA by analysis SF6
gas[J]. High Voltage Engineering, 2007,33(10):224-226.
[5] Xiao Yan,Yu Weiyong.present status and prospect of research of on-
line partial discharge monitoring system in GIS[J]. High Voltage
Engineering,2005,31(1):47-49.
[6] IEC 60480-2004 Guidelines for the checking and treatment of sulfur
hexafluoride(SF6) taken from electrical equipment and specification
for its re-use[S].2004 Yang Haiping,Cheng Hanping,Wang Jing etal.
influence of temperature on coal pyrolyswas and char gasification[J].
Proceedings of the CSEE, 2008,28(8):42-47.
[7] Sun Baizhong,Wang Qing,Sun Baomin etal. experiment study on
combustion performance of oil shale and semi-coke blends[J] .
Proceedings of the CSEE, 2006,26(10):110-114.
[8] Birtwhwastle.D,Blackmore.P,Krivda.A etal.monitoring the condition
of insulator shed materials in overhead dwastribution networks[J].
IEEE Trans. Dielectrics and Electrical Insulation , 1999, 6(5):612-
619.
a: SOF2,b: SO2F2

TABLE II. EACH PEAKS’ PEAK VALUES AT EACH MOMENT

Corresonding Discharge time/h


Wave number of peak/ cm-1
components 12 24 36 48 60 72 84 96
530 SOF2 0.211 0.215 0.274 0.282 0.325 0.399 0.560 0.845
539 SO2F2 0.188 0.197 0.205 0.208 0.173 0.208 0.225 0.333
544 SO2F2 0.108 0.224 0.232 0.249 0.277 0.239 0.289 0.390
552 SO2F2 0.134 0.175 0.169 0.182 0.172 0.182 0.212 0.264
746 SF4 0.118 0.140 0.114 0.148 0.239 0.175 0.330 0.553
752 SOF4 0 0 0 0 0 0.026 0.082 0.188
1103 unknown 0.066 0 0 0.140 0.264 0.164 0.189 0.698
1167 SO2 0 0 0.065 0.133 0.188 0.208 0.342 3.406
1330 SOF2 0 0 0 0.133 0.168 0.342 0.131 0.342
1340 SOF2 0.140 0.145 0.175 0.211 0.167 0.290 0.335 0.431

412

You might also like