You are on page 1of 10

Effects of Multiple Coatings of Two All-in-one Adhesives

on Dentin Bonding
Shuichi Itoa/Franklin R. Tayb/Masanori Hashimotoc/Masahiro Yoshiyamad/Takashi Saitoe/
William W. Brackettf/Jennifer L. Wallerg/David H. Pashleyh

Purpose: Simple changes to bonding technique can improve resin-dentin bond strengths. The objective of this study
was to evaluate the effects of multiple coatings of two all-in-one adhesive resins on both microtensile bond strength
(µTBS) and nanoleakage.

Materials and Methods: The mid-coronal occlusal dentin of extracted human molars was used. Two all-in-one
adhesives – iBond (Heraeus Kulzer) and Xeno III (Caulk Dentsply) – were applied to 320-grit abraded dentin sur-
faces. In groups 1 and 3 during bonding, monomer application and solvent evaporation were done 1, 2, 3, 4 or 5
times on the dentin surface before light curing. In groups 2 and 4 after light curing the first layer, the adhesive was
re-applied, the solvent evaporated, and the layer light cured. This was repeated from 2 to 5 times, followed by
creation of composite buildups. After 24 h storage in 37ºC water, the teeth were sectioned perpendicular to the
adhesive interface to produce multiple beams of composite-bonded dentin, approximately 0.9 mm2 in area. These
were tested to failure in tension. Data were evaluated by three-way ANOVA (material vs coatings vs light curing)
followed by multiple comparisons at α = 0.05. Additionally, nanoleakage of silver uptake and adhesive layer
thickness were evaluated using transmission electron microscopy (TEM).

Results: The results indicated that bond strengths increased with the number of coatings (p < 0.0001) with both
adhesives, up to 3 layers, especially if each layer was light cured. Nanoleakage of silver tended to decrease with
each coat in both adhesive systems.

Conclusion: By simply applying more coats of adhesive, the strength and quality of dentin adhesion can be
improved.

Key words: all-in-one adhesives, dentin bond strength, nanoleakage, multiple coatings.

J Adhes Dent 2005; 7: 133–141. Submitted for publication: 06.02.04; accepted for publication: 05.08.04.

a Instructor, Department of Operative Dentistry and Endodontology, School of Reprint requests: Dr. David H. Pashley, Department of Oral Biology and Maxillo-
Dentistry, Health Sciences University of Hokkaido, Hokkaido, Japan. facial Pathology, School of Dentistry, Medical College of Georgia, Augusta, GA
b Assistant Professor, Department of Conservative Dentistry, Faculty of Den- 30912-1129, USA. Tel: +1-706-721-2033, Fax: +1-706-721-6252. e-mail:
tistry, The University of Hong Kong, Hong Kong, China. dpashley@mail.mcg.edu
c Instructor, Division of Pediatric Dentistry, Hokkaido University, Graduate
School of Dental Medicine, Hokkaido, Japan.
t is well known that bond strength is affected by the
d Professor, Department of Operative Dentistry, School of Dentistry, Okayama
University, Okayama, Japan. I extent of resin infiltration into the exposed collagen net-
work.3,8 Ideally, adhesive monomers should occupy the
e Professor, Department of Operative Dentistry and Endodontology, School of
Dentistry, Health Sciences University of Hokkaido, Hokkaido, Japan. interfibrillar spaces around exposed collagen fibrils.3,14
f Associate Professor, Department of Oral Rehabilitation, School of Dentistry, Dentin adhesives are currently available as three-step,
Medical College of Georgia, Augusta, Georgia, USA.
two-step, and single-step systems, depending on how the
g Associate Professor, Division of Biostatistics, Medical College of Georgia,
Augusta, Georgia, USA. three cardinal steps of etching, priming, and bonding to
h Regents’ Professor, Department of Oral Biology and Maxillofacial Pathology, tooth substrates are accomplished or simplified.12 Re-
School of Dentistry, Medical College of Georgia, Augusta, Georgia, USA. cently introduced all-in-one adhesives further combined

Vol 7, No 2, 2005 133


Ito et al

Table 1 Composition of the all-in-one adhesives used

Material Manufacturer Composition Application procedure

i-Bond Heraeus Kulzer, Hanau, Germany 4-META, UDMA, glutaraldehyde, acetone, Apply three layers with scrubbing, allow
water 30 s of dwell time, evaporate solvent,
light cure for 10 s.
SBMP 3M-ESPE, St Paul, MN, USA Bis-GMA, HEMA, CQ
Xeno III Dentsply Degussa, Hanau, Germany A. Universal: HEMA, aerosil R-947 (fumed 1. Dispense liquid from Univeral and
A. Universal – 360-02 silica), BHT (stabilizer), ethanol, water Catalyst bottles and mix thoroughly.
B. Catalyst – 360-05 B. Catalyst: Pyro-EMA-SK, PEM-F, UDMA, 2. Leave on dentin undisturbed for 20 s.
BHT, CQ, p-dimethyl amine ethyl benzoate Gently air dry.
(co-initiator) 3. Light activate for 10 s.

Abbreviations: 4-META = 4-methacryloxyethyltrimellitic acid; SBMP = Scotchbond Multi-Purpose Plus adhesive; BHT = 2,6-di-tert-butyl-p-cresol; bis-GMA = (1-methylethylidene)bis[4,1-phe-
nyleneoxy(2-hydroxy-3,1-propanediyl)] bismethacrylate; CQ = camphorquinone (2,3-bornanedione); HEMA = 2-hydroxylethyl methacrylate; PEM-F = Penta-methacryl-oxy-ethyl-cyclo-phos-
phazen-monofluoride; pyro-EMA-SK = tetra-methacryl-ethyl-pyrophosphate; UDMA = urethane dimethacrylate (1,6-dimethacryl-ethyl-oxy-carbonylamino-2,4,4-trimethyl hexane).

these three bonding procedures into a single step. How- Buehler, Lake Bluff, IL, USA) to remove occlusal enamel
ever, recent studies have shown that some all-in-one and superficial dentin. Each surface of mid-coronal dentin
adhesives exhibited relatively low bond strengths when was ground with 320-grit silicon carbide paper under
compared with two-step self-etching systems.2,6,7 As running water for 30 s just prior to bonding.
these products create very thin coatings,6,13 they may be
oxygen inhibited and hence poorly polymerized.19 The Experimental Design
manufacturer’s instructions for iBond recommend appli- Two single-step adhesives were used in this study. The
cation of three coats of adhesive prior to evaporation of compositions of these adhesives, Xeno III (Dentsply
solvent and light curing. A recent study also reported that Degussa, Hanau, Germany) and iBond (Heraeus Kulzer,
bonding of Prompt L-Pop may similarly be improved with Hanau, Germany) are shown in Table 1.
an additional layer of adhesive.6,13 There were four groups. Group 1 was composed of
Evaluation of silver uptake by transmission electron Xeno III bonded specimens in which the solvent in each
microscopy (TEM) can provide clues on the extent of resin coating was evaporated but not light cured until the final
monomer infiltration15 or degree of water uptake.18 Re- coating was applied. Group 2 specimens were bonded
cently, several TEM studies reported incomplete resin in- with Xeno III, but the solvent was evaporated from each
filtration into exposed collagen fibrils.10,17 Although there layer and the coating was light cured before applying the
are two reports showing that multiple coats of Prompt next layer. Group 3 specimens were bonded with iBond,
L-Pop can improve bond strengths to dentin,6,13 this and the solvent was evaporated, but only light cured after
effect using other all-in-one adhesives has not yet been application of the final coat. Group 4 specimens were
determined. bonded with iBond but each layer was both evaporated
Using the microtensile test, the purpose of this study and light cured before application of the next layer.
was to evaluate the effects on bond strength of multiple
applications of two all-in-one adhesives to dentin with and Group 1 (evap)
without light curing each layer. Additionally, the effects of For Xeno III, equal volumes of liquids A and B were mixed
multiple coats of all-in-one adhesives on nanoleakage of for approximately 5 s. A generous amount was applied to
the hybrid and adhesive layers were evaluated using TEM. the dentin surface. After leaving the adhesive undisturbed
The null hypotheses tested were that 1 to 5 applications for 20 s, the adhesive was uniformly spread with a gentle
of all-in-one adhesives have no effect on either bond air stream, and the solvent evaporated for 5 s. For
strength or nanoleakage, and that there was no difference multiple consecutive applications (2, 3, 4, or 5 coats),
between curing each layer vs curing only the last layer. adhesive application and solvent evaporation steps were
done repeatedly (Fig 1). In one subgroup, after applying
one coat of A + B liquid and evaporating the solvent,
MATERIALS AND METHODS another coat of liquid B only was applied, the solvent
evaporated and the layer light cured. In another subgroup,
Fifty-six extracted noncarious human third molars were after the first layer was applied and the solvent evaporat-
used in this study. The teeth were collected after obtain- ed, an additional layer of Scotchbond Multi-Purpose Plus
ing the patients’ informed consent under a protocol (3M-ESPE, St Paul, MN, USA) adhesive was applied and
approved by the Medical College of Georgia Institutional light cured for 10 s.
Review Board. Flat dentin surfaces were created in All adhesives were subsequently light cured for 10 s
mid-coronal dentin perpendicular to the tooth’s longitudi- with the light-curing unit (Bisco VIP halogen light, Bisco,
nal axis using a slow-speed diamond saw (Isomet, Schaumburg, IL, USA) operated at 600 mW/cm2.

134 The Journal of Adhesive Dentistry


Ito et al

Group 2 (evap + lc)


Each layer of Xeno III was applied, the solvent evaporated Tooth preparation
and then the layer was light cured before any further treat-
Bonding resin application These procedures were prepared
ment. Thus, each layer was light cured before receiving
the next application. 1,2,3,4,5 times in groups 1 and 3
Air dry for 5 seconds
with gentle air stream
Group 3 (evap)
For iBond, the bonding resin was applied on the dentin sur-
Light irradiation for 10 seconds These procedures were prepared
face and left undisturbed for 30 s. The surface was sub-
1,2,3,4,5 times in groups 2 and 4
sequently gently air dried to evaporate the solvent to form
a slightly shiny adhesive surface. For multiple consecutive
applications of adhesive (2, 3, 4, or 5 coats), the solvent Incremental resin composite buildup and
evaporation steps were done repeatedly but light curing light irradiation of each layer for 60 sec.
(10 s) did not occur until the last coat was applied (Fig 1).
Fig 1 Bonding, evaporation, light-curing steps in the experimental
Group 4 (evap + lc) groups.
After light curing the first layer, a second layer of iBond
was applied, the solvent evaporated, and the adhesive
light cured for 10 s in a similar manner. These steps were
repeated for 3, 4, and 5 coatings. Transmission Electron Microscopy (TEM) of Nanoleakage
Following bonding, five 1 mm increments of TPH resin Four additional teeth with flat dentin surfaces produced in
composite (Caulk/Dentsply for Xeno III) and Venus (Her- the same manner as for microtensile bond testing were
aeus-Kulzer for iBond) were used to create resin compos- employed for each of the subgroups in Groups 1 to 4
ite buildups. Each layer was light cured for 40 s. (Table 1). After treating the dentin surfaces with the
respective bonding protocol, a microfilled resin compos-
Statistics ite containing pre-polymerized fillers (EPIC-TMPT, Parkell,
A three-factor ANOVA was used to examine differences in Farmingdale, NY, USA) was coupled to the bonded dentin
the mean microtensile bond strength (µTBS). The three in order to facilitate ultramicrotomy. The bonded speci-
factors were type of adhesive (Xeno III or iBond), light mens were stored in distilled water at 37°C for 24 h. A
curing (after each layer or after last layer only), and layers 1-mm-thick slab was sectioned from each specimen using
(1 to 5 layers, 1 layer + B, 1 layer +MP). All two- and the slow-speed saw under water cooling. The slabs were
three-factor interactions between the three variables were immersed in 50 wt% ammoniacal silver nitrate for 24 h,
included in the model. A Bonferroni correction to the according to a TEM silver tracer technique.16 Following
alpha level was used to determine post-hoc differences. the reduction of the diamine silver ions into metallic silver
Statistical significance was assessed using an alpha grains, the slabs were dehydrated in an ascending etha-
level of 0.01, and all statistical analyses were performed nol series and embedded in epoxy resin in preparation for
using SAS 8.2 (SAS Institute, Cary, NC, USA). ultramicrotomy. Ninety- to 100-nm-thick undemineralized
sections containing the resin-dentin interfaces were col-
Microtensile Bond Testing lected on single-slot, carbon- and formvar-coated copper
After the bonded specimens had been stored in water at grids. They were examined without further staining using
37ºC for 24 h, the built-up teeth were sectioned perpen- a transmission electron microscope (Philips EM208S,
dicular to the adhesive interface to remove the peripheral Eindhoven, The Netherlands) operating at 80 kV.
3 mm from each side of the tooth, yielding a cube-like
specimen that represents the center of the tooth. This
central cube was cut into 0.9-mm-thick slabs. Each slab, RESULTS
in turn, was sectioned into 1.0-mm-wide beams (adhesive
area: approximately 0.9 mm2) with a diamond saw under Effects on Bond Strength of Multiple Coats Without
water cooling/lubrication. Although 9 beams were pro- Curing each Layer
duced, only the 8 with the greatest dentin thicknesses
were selected for testing. Two teeth were used for each of Groups 1 and 3: Xeno III (evap) and iBond (evap)
the 7 subgroups (14 teeth) per material. Beams prepared Single applications of Xeno III or iBond produced low µTBS
were attached to a testing apparatus with a cyanoacrylate (7.2 ± 6.3 and 12.2 ± 7.5 MPa, respectively) that were
adhesive (Zapit, Dental Ventures of America, Corona, CA, not statistically different. Four of the 16 beams bonded
USA). A tensile load was applied with a material tester with one coat of Xeno III in group 1-1 (evap) exhibited pre-
(Vitrodyne V1000, John Chatillon & Sons, Greensboro, test failures (Table 2). No other subgroups in groups 1 and
NC, USA) at a crosshead speed of 1.0 mm/min until 3 showed any pretest failures. The second layer of Xeno
failure, to obtain the ultimate tensile strength, calculated III increased the bond strength, although the increase was
in MPa. Pretest failures were recorded as zero values in not statistically significant (Table 2). The effect of multiple
the calculation of mean bond strengths in each group. applications of Xeno III reached significance after three

Vol 7, No 2, 2005 135


Ito et al

Table 2 Effects of multiple XIII and iBond adhesive coatings on microtensile bond strengths to dentin

Layers
Groups 1 2 3 4 5 1+B 1 + SBMP

1 X (evap) 7.2 ± 6.3a 22.6 ± 9.2a,b 30.0 ± 9.9b 43.5 ± 7.7c 41.4 ± 8.0c 28.4 ± 9.3b 27.8 ± 5.9b
4/16 0/16 0/16 0/16 0/16 0/16 0/16

2 X (evap + lc) 7.2 ± 6.3a 47.3 ± 19.2b 65.4 ± 20.4d 81.8 ± 20.8d 68.9 ± 22.6d 24.4 ± 15.2b 64.7 ± 21.2d
4/16 0/16 0/16 0/16 0/16 0/16 0/16

3 iB (evap) 12.2 ± 7.5A 18.5 ± 6.2A 30.6 ± 7.0b 34.2 ± 6.0b,c 51.6 ± 14.8b,c – 62.3 ± 23.0c
0/16 0/16 0/16 0/16 0/16 0/16

4 iB (evap + lc) 12.2 ± 7.5A 39.6 ± 15.2B 42.7 ± 12.3b,c 36.6 ± 7.7b 35.8 ± 13.4b – 58.0 ± 23.5c
0/16 0/16 0/16 0/16 0/16 0/16

Values are mean ± SD (n = 16) MPa. In groups 1 and 2, 4 out of the 16 failed prematurely. Zero values were used for those 4 specimens in each group. (evap) = solvent evaporation
between coatings but no light curing; (evap + lc) = solvent evaporation and light curing between each coating. Superscript letters designate statistically different (p < 0.01) groups.
1 + B = after applying the first coating, a second coating of solution B was applied. 1 + MP = after applying the first layer, an additional layer of Scotchbond Multi-Purpose Plus adhesive
was applied and light cured.

layers (p < 0.001) and a plateau after four coats, since by one coat of Xeno III alone. However, when one coat of
there was no statistically significant difference between light-cured Xeno III was covered with one coat of
four and five coats (Table 2). When one coat of Xeno III light-cured SBMP adhesive, the bond strengths increased
was covered with a second layer of solution B and then significantly to 64.7 MPa.
light cured, the µTBS was not different than if a second When the second layer of adhesive was applied to the
layer of A + B had been applied (Table 2). Similarly, when already light-cured first layer in Group 4 specimens
one coat of Xeno III was covered with one layer of Scotch- bonded with iBond, the bond strength increased and was
bond Multi-Purpose Plus (SMBP) adhesive and then light as high as three layers of iBond where only the last layer
cured, there was no significant improvement in bond was light cured (Table 2). Addition of a third, fourth, or
strength over using a second coat of Xeno III. fifth light-cured layers produced no further increase in
Using iBond (Group 3), application of two coats did not bond strength. When one layer of light-cured iBond was
significantly increase µTBS, although three coats (the covered with one layer of light-cured SBMP adhesive, the
manufacturer’s instruction) significantly (p < 0.002) in- bond strength was significantly (p < 0.002) higher than a
creased bond strength (Table 2). There was no significant second, third, fourth, and fifth coat of light-cured iBond.
effect of a fourth or fifth coat on improving bond strength.
Covering one coat of light-cured iBond with SBMP adhe- Effects on Nanoleakage of Multiple Coatings not Cured
sive significantly improved bond strength (p < 0.002) by a Individually
factor of five as compared to a one-coat application. TEM of specimens treated with one layer of iBond (not
Using SBMP adhesive increased the µTBS 3.4 times used according to the manufacturer’s directions of apply-
(p < 0.002) over that produced by a second coating of ing three layers) revealed direct contact of the resin
iBond (Table 2). composite with the surface of dentin (Fig 2a). That is,
there was no layer of polymerized adhesive between the
Effects on Bond Strength of Individually-cured Multiple top of the hybrid layer and the bottom of the composite.
Coats Nanoleakage in the form of heavy silver deposits could be
seen within the 1-µm-thick, partially demineralized hybrid
Groups 2 and 4: Xeno III (evap + lc) and layer (Fig 2a). When one layer of Xeno III was used, the
iBond (evap + lc) hybrid layer was covered by 2 to 3 µm of adhesive
When a second layer of adhesive was applied to the al- (Fig 2b). The amount of nanoleakage was similar when a
ready light-cured first layer in Group 2 specimens bonded second coat of adhesive was applied before light curing,
with Xeno III, the microtensile bond strength increased which created a 4-µm-thick adhesive layer above a
6.6-fold (p < 0.002) over the single coat value (compare 1-µm-thick hybrid layer (not shown). Both all-in-one adhe-
7.2 ± 6.3 to 47.3 ± 19.2, Table 2). Adding a third coating sives removed most of the smear layer, although some
significantly (p < 0.002) increased bond strength again to smear plugs remained in some tubule orifices.
65.4 MPa. Addition of a fourth and fifth coat did not With the application of three consecutive coats of
significantly increase bond strength over that produced by iBond prior to light curing, the thickness of the adhesive
three coats. When one coat of light-cured Xeno III was layer increased to 7 to 8 µm (Fig 2c). The increase in
covered with another coat of bottle B, there was no thickness of the adhesive layer was also accompanied by
significant increase in bond strength over that produced an increase in the propensity to form linear and spherical

136 The Journal of Adhesive Dentistry


Ito et al

Fig 2a TEM micrographs of specimens bonded with one coat of Fig 2b TEM micrographs of specimens bonded with one coat of
iBond that had the solvent evaporated prior to application of the next Xeno III that had the solvent evaporated prior to application of the
layer but no light-curing until application of the final layer. All speci- next layer but no light-curing until application of the final layer. All
mens were undemineralized and were immersed in AgNO3 prior to specimens were undemineralized and were immersed in AgNO3 prior
embedding to reveal nanoleakage. C: composite; D: intertubular to embedding to reveal nanoleakage. C: composite; D: intertubular
dentin; P: peritubular dentin. One-coat of iBond showing direct dentin; P: peritubular dentin. An adhesive (A) layer 2 µm thick is
contact of composite with the dentin surface, with an absence of an seen over a 2 µm thick hybrid layer (between open arrows). Silver
adhesive layer. Silver deposits (arrow) were seen within the 1 µm uptake was seen sporadically within the hybrid layer, surrounded by
thick hybrid layer and between the dentin and composite and within large areas of hybrid layer free of silver.
blisters (pointers).

Fig 2c TEM micrographs of specimens bonded with three coats of Fig 2d TEM micrographs of specimens bonded with three coats of
iBond that had the solvent evaporated prior to application of the next Xeno III that had the solvent evaporated prior to application of the
layer but no light-curing until application of the final layer. All speci- next layer but no light-curing until application of the final layer. All
mens were undemineralized and were immersed in AgNO3 prior to specimens were undemineralized and were immersed in AgNO3 prior
embedding to reveal nanoleakage. C: composite; D: intertubular to embedding to reveal nanoleakage. C: composite; D: intertubular
dentin; P: peritubular dentin. Sporadic silver uptake was seen in the dentin; P: peritubular dentin. Three-coats of Xeno III produced a
1-2 µm thick hybrid layer (H). In the adjacent 7-8 µm thick adhesive 2-3 µm thick hybrid (H) layer that showed scattered silver uptake and
layer (A), large accumulations of silver were seen (2-3 µm in diame- a relatively thick 8-9 µm layer of silica filled adhesive (A) that
ter globular structures adjacent to the hybrid layer with occasional exhibited only individual spot-like silver uptake.
1 x 8-10 µm linear deposits of silver extending across the adhesive
layer).

silver-filled structures within the iBond adhesive layer iBond (compare Figs 2c vs 2d), although the Xeno III
(Fig 2c). In addition, numerous isolated silver grains adhesive showed similar isolated spot-like silver grains.
could be identified throughout the entire adhesive layer A further increase in the number of adhesive coats to
(Fig 2c). Extensive nanoleakage could be seen within the five layers did not necessarily result in an overall increase
hybrid layer. Three coats of Xeno III exhibited less silver in the thickness of the hybrid layer or the polymerized
uptake into the hybrid layer compared to three coats of adhesive layers (Figs 2e and 2f), but it did decrease the

Vol 7, No 2, 2005 137


Ito et al

Fig 2e TEM micrographs of specimen bonded with five coats of Fig 2f TEM micrographs of specimen bonded with five coats of
iBond that had the solvent evaporated prior to application of the next Xeno III that had the solvent evaporated prior to application of the
layer but no light-curing until application of the final layer. All speci- next layer but no light-curing until application of the final layer. All
mens were undemineralized and were immersed in AgNO3 prior to specimens were undemineralized and were immersed in AgNO3 prior
embedding to reveal nanoleakage. C: composite; D: intertubular to embedding to reveal nanoleakage. C: composite; D: intertubular
dentin; P: peritubular dentin. Five coats of iBond produced a 1-2 µm dentin. Five coats of Xeno III produced a 1 µm thick hybrid layer
thick hybrid layer (H) with only sporatic silver uptake. The adhesive (H between two open arrows) that showed minimal silver uptake,
layer (A) was only 7-8 µm thick but was almost free of silver and a 10 µm thick adhesive layer (A) that showed only spot-like
nanoleakage that was limited to spot-like deposits (open arrow- silver deposits.
heads). Silver deposits were seen both in the bottom of the hybrid
layer (white arrow) and in the top of the hybrid layer (pointer).

amount of nanoleakage in the hybrid layers produced by shown in Figs 3d and 3f. Although the thickness of the hy-
both adhesives. brid layers and the nanoleakage within the hybrid layer
were not very different from those previously described,
Effect on Nanoleakage of Multiple Coatings Cured there was a complete absence of silver uptake within the
Individually top of the 15-µm-thick adhesive layer (Figs 3d and 3f). The
Figure 3a is an overview of the resin-dentin interface that most distinctive features, however, were the dramatic re-
was bonded using three coats of iBond that were light duction in the size of the isolated silver grains within the
cured separately, creating an adhesive layer that was intermixed zone between the all-in-one adhesives and the
about 17 to 20 µm thick. The amount of nanoleakage in Scotchbond Multi-Purpose resin (Fig 3e), followed by the
the adhesive layer was less than that observed in the total elimination of these silver grains in the latter (Fig 3f).
comparable 3-coat group, where the layers were not light
cured (compare Fig 2c vs Fig 3a). Although the thickness
of the adhesive layer was similar to that achieved previ- DISCUSSION
ously with the use of three consecutive coats of adhesive
(Fig 2c), there was a complete absence of linear distribu- The results of this study require rejection of the null
tions of silver within with the adhesive when these coats hypothesis that multiple coatings of these two all-in-one
were light cured separately (Fig 3a). This did not, however, adhesives do not increase µTBS. The results also require
eliminate the heavy silver staining within the hybrid layer. rejection of the second null hypothesis that there is no
The absence of linear silver-filled channels in the adhe- difference in µTBS regardless of whether the multiple
sive of iBond (Fig 3a) and the gradual reduction in the size layers were light cured.
of the isolated silver grains (Fig 3c) became more promi- The TEM observations indicate that one layer of iBond
nent with the use of five coats of separately light-cured (not used according to the manufacturer’s recommenda-
iBond adhesives, which generated a 35-µm-thick adhe- tion) is insufficient to adequately create a hybrid layer
sive layer between the composite and dentin, although covered by an adhesive layer that will provide sufficient
nanoleakage in the hybrid layer was severe. Similar coupling to subsequently applied resin composite. How-
results were obtained using three vs five coats of Xeno III ever, one coating of Xeno III which contains nanofillers
(Figs 3b and 3d), although the amount of nanoleakage in was enough to create a hybrid layer covered by an adhe-
the hybrid layer was much less than was seen in hybrid sive layer. Although two or more coatings with either
layers created by iBond. material dramatically increased µTBS when both layers
The difference between the use of multiple coatings of were light cured, there was a further significant increase
iBond or Xeno III and the use of the Scotchbond Multi-Pur- in µTBS (Table 2). This was especially evident using Xeno
pose resin over a single layer of these cured adhesives is III. Using three layers of Xeno III without light curing

138 The Journal of Adhesive Dentistry


Ito et al

Fig 3a TEM micrographs of specimen bonded with three coats of Fig 3b TEM micrographs of specimen bonded with three coats of
iBond that had the solvent evaporated from each layer and in which Xeno III that had the solvent evaporated from each layer and in which
each layer was light-cured before application of the next layer. All each layer was light-cured before application of the next layer. All
specimens were undemineralized and had been immersed in AgNO3 specimens were undemineralized and had been immersed in AgNO3
to reveal nanoleakage. One-coat specimens are not shown since to reveal nanoleakage. One-coat specimens are not shown since
they were the same as Figs 2a and 2b. Three-coats of iBond they were the same as Figs 2a and 2b. Three-coats of Xeno III
produced a 2 µm thick hybrid layer (H) and a 17-20 µm thick adhe- produced a 2 µm thick hybrid layer (H) and a 12-13 µm thick adhe-
sive layer. Although there was considerable silver uptake by the sive (A) layer that contained little silver. C: composite; D: intertubular
hybrid layer, the overlying adhesive (A) layer was devoid of major dentin.
silver uptake, showing only tiny silver grains scattered throughout
the layer. C: composite; D: intertubular dentin.

Fig 3c TEM micrographs of specimen bonded with five coats of Fig 3d TEM micrographs of specimen bonded with five coats of
iBond that had the solvent evaporated from each layer and in which Xeno III that had the solvent evaporated from each layer and in which
each layer was light-cured before application of the next layer. All each layer was light-cured before application of the next layer. All
specimens were undemineralized and had been immersed in AgNO3 specimens were undemineralized and had been immersed in AgNO3
to reveal nanoleakage. One-coat specimens are not shown since to reveal nanoleakage. One-coat specimens are not shown since
they were the same as Figs 2a and 2b. Five-coats of iBond produced they were the same as Figs 2a and 2b. Five-coats of Xeno III
an adhesive (A) layer 34 µm thick that was free of silver uptake, produced a 22-24 µm thick adhesive (A) layer that took up little
although the hybrid layer (H) showed severe silver nanoleakage. silver. The hybrid layer (H) was almost free of silver. C: composite;
C: composite; D: intertubular dentin. D: intertubular dentin.

between layers (but with solvent evaporation at each albeit at a much lower vapor pressure. Thus, multiple
step) produced a mean µTBS of 30.0 MPa. However, if applications may tend to permit the accumulation of
each layer was light cured, three layers gave a µTBS of water on the surface. Such water should be evaporated to
65.4 MPa (p < 0.05 compared to three coats, of which avoid low bond strengths.11
only the last one was light cured). Solution B in Xeno III contains all of the comonomers,
iBond contains water and acetone as solvents. The initiators, and accelerators except HEMA (Table 1). It
evaporation of acetone is much faster than that of water. contains no solvent. Solution A contains HEMA and a
After acetone is gone, water must evaporate by itself, water-ethanol mixture. It was thought that adding an

Vol 7, No 2, 2005 139


Ito et al

Fig 3e TEM micrographs of specimen bonded with one coat of Fig 3f TEM micrographs of specimen bonded with one coat of
iBond that had the solvent evaporated and then was light-cured Xeno III that had the solvent evaporated from each layer and then
before application of Scotchbond Multi-Purpose (SBMP) adhesive was light-cured before application of Scotchbond Multi-Purpose
that was subsequently light-cured. All specimens were undemineral- (SBMP) adhesive that was subsequently light-cured. All specimens
ized and had been immersed in AgNO3 to reveal nanoleakage. were undemineralized and had been immersed in AgNO3 to reveal
One-coat specimens are not shown since they were the same as nanoleakage. One-coat specimens are not shown since they were
Figs 2a and 2b. One layer of iBond (light-cured) followed by the the same as Figs 2a and 2b. One layer of Xeno III (evaporated and
application of one layer of Scotchbond Multi-Purpose (SBMP) adhe- light-cured) followed by the application of one layer of SBMP to
sive to dentin. This treatment produced a 15 µm thick adhesive layer dentin produced a 14-15 µm thick adhesive layer that consisted of
that consisted of 9 µm of SBMP (solid line R) and 8 µm of iBond 12-13 µm of SBMP (line R) and a 5 µm thick layer of Xeno III (A) with
adhesive (dotted line A) with a 2 µm thick intermixed zone (IZ). There a 2-3 µm thick intermixed zone (IZ). Although there was moderate
was much less silver uptake into the hybrid layer created by a single silver uptake by the hybrid layer (H), there was no silver uptake in the
coat of iBond compared to its control (Fig 2a). C: composite; adhesive layer (A). C: composite; D: intertubular dentin.
D: intertubular dentin; H: hybrid layer.

additional coating of B (without A) to the first coating of adhesives, there are several techniques that can be
A + B would enhance the µTBS of Xeno III because it employed. If the solvent is evaporated between each
would be less hydrophilic. This was unsuccessful coat, the concentration of comonomers that exists after
(Table 2), regardless of whether the first layer was light each coating should increase, thereby facilitating
cured or not prior to the application of material from comonomer infiltration with minimal increase in the
solution B (that was subsequently light cured). thickness of the adhesive layer.
When one layer of iBond or Xeno III was covered with These solvated adhesives only contain 25 to 30 wt%
SBMP adhesive, there was a significant increase comonomers and 70 to 85% solvents (preliminary gravi-
(p < 0.05) in µTBS (Table 2) regardless of whether the metric analysis). The presence of the solvent facilitates
first coat was light cured or not. The adhesive of SBMP is wetting and spreading of the solvated comonomers and
basically a blend of bis-GMA/HEMA without any volatile may assist in preserving the width of interfibrillar spaces
solvent. This adhesive layer may have increased the during infiltration.3,5 This may facilitate inward diffusion
mechanical properties of the bond due to improved of the comonomers into the matrix. When most of the
polymerization. The more hydrophobic SBMP adhesive solvent is evaporated, the concentration of the adhesives
may have also sealed the surface of the more hydrophilic must increase dramatically, although the thickness of the
iBond. Indeed, Figs 3e and 3f shows that almost no silver layer may be extremely thin.6,13 The cumulative concen-
uptake was seen in the SBMP adhesive layer (R), unlike tration of these constituents must rise in the adhesive
the substantial silver uptake seen in the underlying iBond layer with each application. We speculate that this con-
and Xeno III adhesives. tributes to improved polymerization of the final mixture.19
It is likely that the increased bond strength seen in Although the thickness of the resin layer increases
both all-in-one adhesives with multiple coatings is due to enough to prevent oxygen inhibition of its entire thick-
several mechanisms operating simultaneously. As the ness,6,13 we believe that the final adhesive layer has an
first layer of adhesive begins to etch dentin, it is probably improved quality (ie, better cross linking, improved inter-
rapidly buffered,9 so that additional layers of unpolymer- penetration of polymer chains, improved percent conver-
ized comonomers may improve the etching ability of sion). We speculate that the bond strength reached a
adhesives. Although the hybrid layer thickness of iBond plateau following sequential coatings when a limiting
specimens increased from 0.8 to 1.5 µm when one vs critical variable had been reached (eg, film thickness,
three layers was applied, the values were not statistically cross-linking density, concentration of photoinitiators,
significantly different. When using multiple coatings of accelerators, etc).

140 The Journal of Adhesive Dentistry


Ito et al

In contrast, when each layer was light cured following REFERENCES


solvent evaporation, the improved bond strengths may
have been due to mechanical rather than chemical 1. Bonilla ED, Stevenson RG, Yashar M, Caputo AA. Effect of application
technique and dentin bonding agent interaction on shear bond strength.
factors. After the first layer was cured, it would be unlikely Oper Dent 2003;28:568-573.
that there could be any more resin infiltration into the ma- 2. Bouillaguet S, Gysi P, Wataha JC, Ciucchi B, Cattani M, Godin C, Meyer
JM. Bond strength of composite to dentin using conventional, one-step,
trix. However, it might be possible to improve the quality and self-etching adhesive system. Dent Mater 2001;17:373-380.
of each layer of adhesive. That is, since each layer was 3. Carvalho RM, Mendonca JL, Santiago SL, Silveira RR, Garcia FCP, Tay FR,
polymerized, only the oxygen-inhibited portion of the layer Pashley DH. Effects of HEMA/solvent combinations on bond strength to
dentin. J Dent Res 2003;82:597-601.
would be available to mix with the newly applied layer. 4. Choi KK, Condon JR, Ferracane JL. The effects of adhesive thickness on
However, the oxygen-inhibited layer in iBond can be 8 to polymerization contraction stress of composite. J Dent Res 2000;79:
10 µm thick and can easily mix with subsequently applied 812-817.
5. Eddleston CL, Hindle AR, Agee KA, Carvalho RM, Tay FR, Rueggeberg FA,
layers. This might improve its quality. One layer of iBond Pashley DH. Dimensional changes in acid-demineralized dentin matrices
produced an adhesive layer thickness that ranged from following the use of HEMA-water versus HEMA-alcohol primers. J Biomed
0 to 3 µm. When three layers were applied and each one Mater Res 2003;67A:900-907.
6. Frankenberger R, Perdigão J, Rosa BT, Lopes M. ’No bottle’ vs ’multi-bottle’
was light cured, the adhesive layer was about 15 to dentin adhesives – a microtensile bond strength and morphological study.
17 µm thick. Five layers of iBond (each cured individually) Dent Mater 2001;17:373-380.
produced a 34-µm-thick adhesive layer (Fig 3c), while 7. Fritz UB, Finger WJ. Bonding efficiency of single-bottle enamel/dentin
adhesive. Amer J Dent 1999;12:277-282.
five layers of individually cured Xeno III produced a 8. Gwinnett AJ. Moist versus dry dentin: Its effect on shear bond strength.
26-µm-thick adhesive layer. We speculate that the major J Dent 1992;5:127-129.
improvement in bond strength with successive layers in 9. Hiraishi N, Kitasako Y, Nikaido T, Nomura S, Burrow MF, Tagami J. Effect
of artificial saliva contamination on pH value change and dentin bond
group 2 and 4 specimens was due to increased adhesive strength. Dent Mater 2003;19:429-434.
layer thickness.4 Multiple layers of adhesive tend to cre- 10. Hashimoto M, Ohno H, Sano H, Kaga M, Oguchi H. Degradation patterns
of different adhesives and bonding procedures. J Biomed Mater Res Part
ate thicker adhesive layers that increase the compliance B: Appl Biomaterials 2003;66B:324-330.
of the joint. This may lower local stress concentrations 11. Jacobson T, Söderholm K-J. Some effects of water on dentin bonding. Dent
during tensile testing, resulting in higher yield strengths. Mater 1995;11:132-136.
12. Kugel G, Ferrari M. The science of bonding: from first to sixth generation.
Regardless of the mechanism(s) responsible for J Am Dent Assoc 2000;131(suppl):20S-25S.
increased bond strength, the need for multiple applica- 13. Pashley EL, Agee KA, Pashley DH, Tay FR. Effects of one versus two
tions increases the bonding time of these 'simplified' applications of an unfilled, all-in-one adhesive on dentin bonding. J Dent
2002;30:83-90.
adhesives. Although we did not attempt to optimize the 14. Pashley DH, Ciucchi B, Sano H, Homer JA. Permeability of dentin to
solvent evaporation time, we used 5 s. Five seconds for adhesive agents. Quintessence Int 1993;24:618-631.
each of 4 layers adds 20 s to the procedure. However, 15. Sano H, Yoshiyama M, Ebisu S, Burrow MF, Takatsu T, Ciucchi B, Carvalho
RM, Pashley DH. Comparative SEM and TEM observations of nanoleakage
adding these 20 s, plus light curing each of the four within the hybrid layer. Oper Dent 1995;20:160-167.
layers for 10 s (40 s cumulative curing time plus 20 s 16. Tay FR, Carvalho RM, Yiu CKY, King NM, Zhang Y, Agee K, Bouillaguet S,
evaporation time), requires more time than is stated Pashley DH. Mechanical disruption of dentin collagen fibrils during
resin-dentin bond testing. J Adhes Dent 2000;2:175-192.
current manufacturer’s instructions. 17. Tay FR, Pashley DH, Yoshiyama M. Two modes of nanoleakage expression
These results confirm how important bonding tech- in single-step adhesives. J Dent Res 2002a;81:472-476.
nique is to producing optimal resin-dentin bonds.1,20 18. Tay FR, King NM, Chan K-M, Pashley DH. How can nanoleakage occur in
self-etching adhesive systems that demineralize and infiltrate simulta-
Simple changes in bonding technique, such as applying neously. J Adhes Dent 2002;4:255-269.
more layers of all-in-one adhesives, can lead to large 19. Wang Y, McMann TR, Spencer P. Morphologic and chemical characterization
of the dentin/self-etch interface [abstract 1715]. J Dent Res 2001;80
increases in initial bond strength. Whether this improves (special issue).
the long-term durability of resin-dentin bonds remains to 20. Swift EJ Jr, Wilder AD Jr, May KN Jr, Waddell SL. Shear bond strength of
be determined. one-bottle dentin adhesives using multiple applications. Oper Dent 1997;
22:194-199.

ACKNOWLEDGMENTS
We thank Amy Wong of the Electron Microscopy Unit, The University of Hong
Kong for technical assistance and Michelle Barnes of the Medical College of
Georgia for secretarial support. This work was supported, in part, by grants DE Clinical relevance: By simply applying more coats of
014911 and DE 015306 (to DHP) from the National Institute of Dental and
Craniofacial Research. The materials (iBond/Heraeus Kulzer, Hanau, Germany adhesive, the strength and quality of dentin adhesion
and Xeno III/Caulk Dentsply, Germany) used in this study were generously can be improved.
supplied by each manufacturer. The testing device used for microtensile bond
testing (Ciucchi’s jig) was supplied by Kuraray Co, Ltd, Osaka, Japan.

Vol 7, No 2, 2005 141

You might also like