You are on page 1of 10

Clin Oral Impl Res 2000: 11: 448–457 Copyright C Munksgaard 2000

Printed in Denmark ¡ All rights reserved

ISSN 0905-7161

An in vitro study on compensation of mismatch


of screw versus cement-retained implant
supported fixed prostheses
Pietrabissa R, Gionso L, Quaglini V, Di Martino E, Simion M. An in Riccardo Pietrabissa1,
vitro study on compensation of mismatch of screw versus cement-retain- Luca Gionso2,
ed implant supported fixed prostheses. Virginio Quaglini3,
Clin Oral Impl Res 2000: 11: 448–457. C Munksgaard 2000. Elena Di Martino4,
Massimo Simion2
In common practice a perfect fit of the prosthetic framework with the 1
Laboratory of Biological Structure
implant abutments is almost impossible to achieve. The mismatch, which Mechanics, Dipartimento di
is principally induced by the technological process adopted to manufac- Bioingegneria, Politecnico di Milano,
ture the fixed prostheses, strains the framework thus generating constraint Milano; 2Università degli Studi di Milano,
reactions. These are static forces that load the implant components and Servizio di Odontostomatologia I,
the bone at the implant–bone interface and may cause the bone remodel- Ospedale San Raffaele, Milano;
3
Laboratory of Biological Structure
ling. Depending on the magnitude of such forces, i.e. depending on the
Mechanics, Dipartimento di Ingegneria
magnitude of the mismatch, the bone remodelling may lead to the Strutturale, Politecnico di Milano, Milano;
loosening of the screws and of the implant–bone interface and hence cause 4
CeBITeC, Dipartimento di
implant failure. The present study shows an in vitro comparison of 3 differ- Bioingegneria, Politecnico di Milano and
ent connecting abutments (standard, EsthetiCone and CerAdapt, Nobel IRCCS San Raffaele, Milano, Italy
Biocare AB, Göteborg, Sweden) with relation to the mechanical stresses
induced by geometrical mismatches (technology induced errors). Two ex-
perimental devices were purposely realized and used to assess the ability
of the different abutments to compensate errors. One was designed for Key words: dental implants – fixed
translation errors and the other for rotation errors. The experimental prosthesis mismatching – position
apparatus set-up includes 2 freestanding implants supporting a prosthetic errors – experimental study –
structure and the connecting abutments. The implants and the abutments biomechanics
were used as delivered by the manufacturer, while the prostheses were
purposely realized and instrumented with strain gauges. The data obtained Prof. Riccardo Pietrabissa, Laboratory of
Biological Structure Mechanics,
with the error devices do not give quantitative information on what hap-
Dipartimento di Bioingegneria,
pens in clinical applications where the implants are connected to living Politecnico di Milano, Piazza L. da Vinci,
bone, which is a tissue much more deformable than the steel used for the 32, 20133 Milano, Italy
error devices. Results allow direct comparisons of the behaviour of the Tel.: π39 02 2399 3349
different investigated abutments with respect to position errors. The Cer- Fax: π39 02 2399 3360
Adapt system (cement retained ceramic abutments) showed the least e-mail pietrabissa/biomed.polimi.it
strain in presence of translation errors. The standard system (screw retain-
ed abutments) showed the least strain in presence of rotation errors. Accepted for publication 23 August 1999

The use of osseointegrated implants as a support oral implants in totally edentulous jaws (Adell et
for dental prostheses has obtained general consent al. 1981; Albrektsson 1988; Albrektsson & Senner-
by most clinicians worldwide. The Brånemark im- by 1991) has encouraged several clinicians to ex-
plant technique was originally introduced for tend the application of dental implant treatment
treatment of completely edentulous jaws and the to partial edentulism. In the last few years, several
prosthetic components consequently have been de- studies have shown promising results in the treat-
signed in order to satisfy this purpose. Successful ment of partially edentulous patients with both
long-term prognosis of osseointegrated Brånemark fixed bridge restorations and single tooth replace-

448
Compensation of mismatch of implant supported fixed prostheses
ments (Jemt 1986; Jemt et al. 1989; Jemt et al. abutment construction, preparation of the master
1992; van Steenberghe 1989; van Steenberghe et al. model, waxing, casting, welding and ceramic pro-
1991; Henry et al. 1993). cessing. Each of these steps may add errors result-
Recently, the application of dental implants to ing in a mismatch between the position of the abut-
partial fixed prosthetic restorations has increased ments with respect to the fixed prosthesis gold cyl-
the esthetic requirements of the final restoration. inders. Such errors can be related to the horizontal
Therefore, new abutment designs have been intro- and vertical position and to the rotation between
duced in clinical practice as alternatives to the the different structures. These errors are defined as
standard abutment cylinder. A conical abutment position errors.
(EsthetiCone, Nobel Biocare AB, Göteborg, Clinicians experience the mismatch due to posi-
Sweden) is available since 1991 for multiple unit tion errors during the fixed prosthesis application
reconstruction. The advantage of this abutment as the mismatch does not allow a smooth assembly
design is that it is available in a 1 mm height con- of the fixed prosthesis on the abutments. The mis-
figuration, which allows the prosthetic finishing match strains the assembled framework thus gen-
line to be positioned under the gingival margins. erating reactive forces at the implant–bone inter-
This characteristic prevents non-esthetic titanium face. These static forces load the bone, and bone
abutment exposure over the gingiva even in cases remodelling may occur resulting in implant site
in which thin mucosal tissues are present. In 1989 displacement toward the direction which produces
the traditional concept of prosthetic superstruc- the decrease of the framework strain. The greater
tures connected to implants through transmucosal the mismatch, the greater the force applied to the
abutments and retained with screws was modified implant–bone interface, the more pronounced the
with a new anchoring method. This method, using bone remodelling process, the higher the risk of
the so called ‘‘UCLA’’ abutment (Lewis et al. loosening of the implant–bone interface mechan-
1988) provides the fabrication of customized gold ical stability and hence the risk of implant failure.
alloy abutments directly connected to the implant Lack of data exists about the ability of cement
heads. The prosthetic superstructure is temporarily or screw-retained prosthesis to transfer or compen-
cemented on the gold abutments with the same sate the stresses induced by position errors.
technique used for cementation on copings in This paper focuses on an in vitro study com-
natural tooth abutments. Cement-retained pros- paring 3 different abutment systems with respect to
theses benefit from the absence of the non-esthetic their compliance when increasing mismatches are
screw access through the crown surface, with the applied.
possibility to adapt the shape of the crown and the
emerging profiles to the specific anatomic situation
Materials and methods
and from the positioning of the finishing lines ac-
cording to the shape and location of the mucosal Assumptions
margins. The work takes into consideration fixed prosthetic
Recently, Prestipino & Ingber (1993) have pro- units bonded and supported by two dental im-
posed a new ceramic abutment (CerAdapt, Nobel plants. Position errors that cause the mismatch be-
Biocare AB, Göteborg, Sweden). This particular tween the two fixed prosthesis gold cylinders and
abutment consists of a sinterized alumina cylinder the two abutments are six in an absolute Cartesian
that has to be directly connected to the implant coordinate system X, Y, Z: translation errors AX,
head prepared like a natural tooth. The prosthetic AY, AZ and rotation errors qX, qY, qZ. In our case
superstructure is retained with temporary cement. it is convenient to adopt the fixed prosthesis Car-
The advantages of this abutment are the high bioc- tesian coordinate system in which the Z axis is the
ompatibility of the material, the improved esthetics axis of one of the two implants and the X axis is
due to the brightness through the mucosal tissues, the fixed prosthesis longitudinal axis so that the
the finishing lines and the emerging profile cus- XZ plane contains the axis of the two implants if
tomization, and the elimination of the screw holes they are parallel each other. Using this coordinate
through the crown surface. However, the drawback system no translation position error may be re-
of all the cement-retained prostheses is the trouble- ferred to the Y axis as shown in Fig. 1a. The Y
some retrievability of the superstructure. error AY means the fixed prosthesis reorientation
Another issue to be considered in fixed implant and hence the coordinate system reorientation (Xø,
supported prostheses is the possible mechanical Yø, Z) with only an Xø error (AXø).
stresses introduced in the system by small dimen- Looking at the rotation errors considered as ro-
sional errors in the construction of the superstruc- tations of the Cartesian coordinate system of one
ture. The final prosthetic superstructure is manu- fixed prosthesis gold cylinder with respect to the
factured through different steps: impression of the other (Fig. 1b), the qZ error may be neglected as

449
Pietrabissa et al.

this paper: two translation errors – AX, called fixed


prosthesis length error, and AZ, called fixed pros-
thesis shear error – and two rotation errors – qX,
called fixed prosthesis torsion error, and qY, called
fixed prosthesis bending error.
All the errors are fixed prosthesis errors under
the hypothesis that they depend on the technologi-
cal process but, as the errors are mismatches be-
tween the fixed prosthesis gold cylinders and the
abutments, hence relative errors, they have been
represented as abutment errors (ĀX, ĀZ, q̄X, q̄Y)
for the experimental reasons explained in the fol-
lowing. Each abutment position error is in the op-
posite direction of the corresponding fixed pros-
thesis position error. It holds: ĀXΩªAX, ĀZΩ
ªAZ, q̄XΩªqX, q̄YΩªqY. The four abutment
errors considered in this work are represented in
Fig. 2.

Fig. 1. (a) The adopted Cartesian coordinate system used to


identify the translation errors. The X axis is the fixed prosthesis
longitudinal axis and the Z axis is the abutment axis. The errors
in Y direction are converted in Xø errors as a consequence of
the fixed prosthesis and coordinate system rotation (angle a)
around the Z axis. (b) Rotation errors are defined as relative
rotations of the two fixed prosthesis gold cylinder coordinate
systems (X1, Y1, Z1 and X2, Y2, Z2). In this occurrence rotation
around X2 or Y2 axis causes qX or qY errors while rotation
around the Z2 axis does not cause errors as the connection site
is supposed to have an axial symmetry.

Fig. 2. Schematic representation of a mandible with two im-


plants supporting a fixed prosthesis. In detail the possible abut-
ment position errors are (a) length error, (b) shear error, (c)
torsion error and (d) bending error.

the rotation of a gold cylinder around its axis of


symmetry does not cause mismatches.
As a consequence of the previously mentioned Fig. 3. (a) Sketch of the device used to impose translation
considerations only four errors are considered in errors. (b) Sketch of the device used to impose rotation errors.

450
Compensation of mismatch of implant supported fixed prostheses

Experimental procedure
The adopted experimental procedure is intended
to apply the desired abutment position errors and
measure the correspondent fixed prosthesis strains.
Two purposely designed devices have been fabri-
cated and used, a translation error device (Fig. 3a)
and a rotation error device (Fig. 3b). Both the de-
vices have been built in steel.
The translation error device has two slides that
may be translated using two micrometer screws,
0.5 mm pitch. One enables to set ĀX errors and
the other ĀZ errors. The rotation error device has
two shafts that may be rotated to the desired angle
using a goniometer. One shaft enables to set q̄X
errors and the other q̄Y errors. Each slide/shaft of
both devices houses one implant with its abutment.
The fixed prosthesis is equipped with two strain
gauges on its upper surface in order to measure
the strains induced by position errors. The fixed
prosthesis is mounted on the abutments of the
translation (rotation) error device and the slides
(shafts) are translated (rotated) to obtain the mini-
mum fixed prosthesis strain which corresponds to
the minimum position error. In general a global
strain equal to zero cannot be obtained because Fig. 4. The fixed prosthesis instrumented for measuring strains
the translation (rotation) error device cannot com- due to (a) length errors, (b) shear errors, (c) bending errors and
(d) torsion errors.
pensate rotation (translation) errors. Hence the
minimum fixed prosthesis strain is assumed as ref-
erence for abutment position errors equal to zero. for the different error type. As a consequence the
When the minimum position error is reached the strain gauges have been applied in different ways
fixed prosthesis is removed from the abutments and depending on the error type to measure as follows.
the abutments are moved one at a time to set the For fixed prosthesis length and shear errors the
desired error. The fixed prosthesis is mounted strain gauges are both parallel to the fixed pros-
again on the abutments and the strain gauges allow thesis longitudinal axis (X) and they measure
measuring the strains induced by the set position strains in X direction (eX). With reference to Fig.
error. The fixed prosthesis is then removed again 4a and 4b, where the positive direction of the abut-
and the procedure is repeated using translation ment errors are indicated, one notes that:
steps of 20 mm and rotation steps of 1æ. The maxi-
eX1±0 and eX2±0 for ĀX±0
mum investigated translation errors have been set
eX1∞0 and eX2±0 for ĀZ±0.
to 0.1 mm while the maximum investigated ro-
tation errors depend on the feasibility of assemb- For fixed prosthesis bending and torsion errors
ling the fixed prosthesis on the rotated abutments. one strain-gauge is parallel to the fixed prosthesis
longitudinal axis (X) and it measures strains in X
direction (eX) while the other is set to 45æ with re-
Measurements spect to X axis and it measures strains in that di-
Position errors are set moving the translation slides rection (e45). With reference to Fig. 4c and 4d,
or the rotation shafts and measuring the new posi- where the positive directions of the abutment
tion of the moved abutment using a thread gauge errors are indicated, one notes that:
or a goniometer.
eX±0 and e45∑0 for q̄Y±0
The fixed prosthesis strain induced by the posi-
eX∑0 and e45±0 for q̄X±0.
tion error is measured using two strain gauges ap-
plied on the upper surface of the fixed prosthesis. The strain-gauges fixed as previously shown en-
On the basis of the theory of elasticity and con- able to compare the effects due to the same posi-
sidering the fixed prosthesis, as a first approxi- tion errors on fixed prostheses mounted using dif-
mation, similar to an elastic beam, it is possible to ferent connecting abutments as well as the effects
predict the deformed shape of the fixed prosthesis due to increasing position errors.

451
Pietrabissa et al.

Fig. 5. Typical strains measured by each strain gauge as functions of the different position errors.

The strain-gauges used for measurements are fixed prosthesis fabrication has been the V Classic
LY13–0.6/120 (Hottinger Baldwin Messtechnik gold alloy (Metaux Precieux SA Metalor, Neuch-
GmbH, Darmstadt, Germany) with 0.6 mm refer- atel, Switzerland). The fixed prostheses have 20
ence length in the measurement direction and 120 mm distance between the two gold cylinder axes.
ohm reference resistance. The strain gauges have Two Brånemark standard fixtures SDCA 001
been connected to a conditioner-amplifier (3.75¿10 mm) have been screwed into the slides/
DMCPLUS12 (Hottinger Baldwin Messtechnik shafts of the error devices to support the abut-
GmbH, Darmstadt, Germany). ments.
Fig. 5 shows typical strains measured by the Three different pairs of Brånemark abutments
strain gauges as functions of the position errors. have been used to connect the fixed prosthesis to
the implants. The first pair has been the standard
abutment 3 mm high and the second the Estheti-
Fixed prostheses and implant systems Cone abutment 3 mm high. In both cases the fixed
Six fixed prostheses have been prepared identical prosthesis has been connected using gold screws
in shape and dimensions. The material used for the according to the indications of the manufacturer.

452
Compensation of mismatch of implant supported fixed prostheses
Table 1. Types and dimensions of the tested abutments

Abutment type Junction Height Implant diameter


(mm) (mm)

Standard gold screws 3 3.75


EsthetiCone gold screws 4.75 3.75
CerAdapt cement 11.9 3.75

The screws have been tightened with a Torque


Controller handpiece (Nobel Biocare AB, Göte-
borg, Sweden) using a torque of 20 N cm. The
third type has been the CerAdapt abutment pre-
pared with a high-speed diamond burr. The fixed
prosthesis shoulder has been located 3 mm over
the abutment base and the axial walls have been
gently grided to achieve 6æ of inclination. Before
waxing, a die-spacer layer 20 mm thick has been
applied to each abutment.
The fixed prosthesis has been retained with a
temporary cement (Temp Bond, Kerr Ltd,
Peterborough, United Kingdom) applying an axial Fig. 6. Behaviour of the 3 tested abutments in terms of fixed
force of 100 N for 60 s. The characteristics of the prosthesis strains measured with the translation error device
3 types of abutment are shown in Table 1. imposing abutment length errors.

Results
Figs 6 to 9 show the strains measured on the fixed
prosthesis versus the position errors for the 3
tested connecting abutments. The test procedure
has been evaluated, in a series of repetitions, to
give results with a measuring error lower than 5%
of the measured value.

Abutment length error


In Fig. 6 the fixed prosthesis strains have been cal-
eX1πeX2
culated as . The minimum strains have
2
been obtained using the CerAdapt abutments (the
strain is always under 100 me). In comparison with
the standard abutments, the EsthetiCone abut-
ments induced lower strains for ĀXÆ60 mm and
higher strains for ĀXØ80 mm. The maximum
strain has been 200 me reached using the Estheti-
Cone abutments for ĀXΩ100 mm.

Abutment shear error Fig. 7. Behaviour of the 3 tested abutments in terms of fixed
prosthesis strains measured with the translation error device
In Fig. 7 the strains have been calculated as imposing abutment shear errors.
eX2ªeX1
. The minimum strains have been ob-
2
tained using the CerAdapt abutments (the strain is induced lower strains for ĀZÆ60 mm and higher
always under 10 me except for ĀZΩ40 mm which strains for greater errors (Ø80 mm). The maximum
scored a strain of 18 me). In comparison with the absolute strain has been 68 me reached using the
EsthetiCone abutments, the standard abutments standard abutments for ĀZΩ100 mm.

453
Pietrabissa et al.

or even lower than using the EsthetiCone and Cer-


Adapt abutments. The EsthetiCone abutments in-
duced lower strains than the CerAdapt abutments
except for 3æ bending error. The maximum strain
has been 160 me reached using the EsthetiCone
abutments for q̄YΩ3æ.

Abutment torsion errors


In Fig. 9 the fixed prosthesis strain is e45. The stan-
dard abutments induced the minimum strain for
every torsion error investigated, except for 0 and 1æ.
The strains obtained using these abutments have al-
ways been lower than 70 me. The EsthetiCone abut-
ments induced always the greatest strain with a
maximum of 170 me for q̄XΩ3æ. The CerAdapt abut-
ments induced intermediate values of strain.
It should be noted that the zero strain has not
been achieved with the angular device, i.e. a non-
zero strain is measured by the strain gauges when
Fig. 8. Behaviour of the 3 tested abutments in terms of fixed
prosthesis strains measured with the rotation error device-im- no error is set on the device. This is due to the
posing abutment bending errors. presence of either fixed prosthesis length or shear
error that cannot be compensated by the rotation
error device.

Discussion
This experimental study was designed in order to
compare 3 different implant–prosthesis connection
systems (one cemented and two screw-retained)
with respect to their capability to compensate fixed
prosthesis dimensional and/or shape errors.
The results from the study showed that the ce-
mented abutments (CerAdapt) have the best capa-
bility of compensating translation (axial and trans-
versal) errors. In fact, at the maximum translation
error (0.1 mm) CerAdapt connected prosthesis in-
duced only half of the strain as compared to both
the prostheses connected with standard or Estheti-
Cone abutments (90 versus 200 me). A possible ex-
planation of this finding could be that the presence
of a deformable cement layer might compensate
the translation errors produced during the manu-
facturing process. The cement could act as an ad-
sorber for the deformation caused by the mis-
Fig. 9. Behaviour of the 3 tested abutments in terms of fixed match and therefore it could be able to preserve
prosthesis strains measured with the rotation error device im- the structure (implant–abutment–prosthesis) from
posing abutment torsion errors.
greater strains. The gap between the abutments
and the prosthesis introduced by the die-spacer,
during waxing and casting, could allow the forma-
tion of a cement layer with different thickness at
Abutment bending error the abutment–prosthesis interface. This might
In Fig. 8 the fixed prosthesis strain is eX. The stan- compensate the translation error and reduce the
dard abutments induced the minimum strain for strain.
every imposed bending error and the strain value The standard and EsthetiCone abutments
(which was always lower than 40 me) has been half showed approximately the same capability of com-

454
Compensation of mismatch of implant supported fixed prostheses
pensating translation errors and the strains of the ture et provoque donc des contraintes. Ces forces statiques qui
prosthesis displayed an almost linear increasing chargent les composants de l’implant et de l’os au niveau de
l’interface os-implant peuvent donc occasionner le remodelage
with error. However, shear errors were more easily osseux. Suivant l’amplitude de ces forces, c.-à-d. suivant la ma-
compensated (strain lower than 70 me for 0.1 mm gnitude de cette disparité, le remodelage osseux peut provoquer
error) in comparison with axial errors (200 me for le relâchement des vis et celui de l’interface os-implant provo-
0.1 mm error). quant alors l’échec. L’étude compare in vitro trois connections
différentes (standard, EsthetiCone et CerAdapt, Nobel Biocare
With rotation errors, the EsthetiCone abutments AB, Göteborg, Suède) en relation avec les stress mécaniques
induced lower capability of compensating both induits par des disparités géométriques (erreurs induites
bending and torsion errors. In fact, they caused technologiquement). Deux systèmes expérimentaux ont été réa-
greater strains in most of the tests as compared to lisés et utilisés pour vérifier l’habilité des différents piliers à
both the other two systems. The lower compensat- compenser les erreurs. Un système a été conçu pour les erreurs
de translation et l’autre pour les erreurs de rotation. L’appareil
ing ability of the EsthetiCone abutments in pres- expérimental comprenait deux implants libres portant une
ence of rotation errors could be due to their larger structure prothétique ainsi que les piliers de connection. Les
height as compared to the standard abutments. implants et les piliers ont été utilisés suivant les indications don-
Major height could cause a greater horizontal dis- nées par le fabriquant tandis que les prothèses ont été réalisées
et accompagnées de jauges de force. Les données obtenues avec
placement of the top of the abutment at the same les systèmes d’erreur ne donnent pas d’information quantitative
degree of rotation. sur ce qui arrive dans les applications cliniques où les implants
For rotation errors the behaviour of the CerAd- sont reliés à l’os vivant qui est un tissu beaucoup plus déforma-
apt abutments stands between those of the stan- ble que l’acier utilisé pour les systèmes d’erreur. Les résultats
dard and EsthetiCone abutments. This behaviour permettent des comparaisons directes dans le comportement
des piliers étudiés vis-à-vis des erreurs de position. Le système
could be correlated to the combination of two fac- CerAdapt accusait la contrainte la plus faible en présence des
tors: 1) the overall height of the tested CerAdapt erreurs de translation. Le système standard, c.-à-d. celui dont
abutments was the highest and this causes a les piliers étaient retenus par des vis, possédait la force la plus
stronger lever arm and 2) the compensating effect faible en présence des erreurs de rotation.
of the cement layer. However, the results concern-
ing the CerAdapt abutments should be regarded Zusammenfassung
with caution as their error transfer capability could
In der Privatpraxis ist es fast unmöglich, eine perfekte Passge-
vary depending on the shape and the height of the nauigkeit zwischen dem prothetischen Gerüst und den Implan-
final abutment preparation in the different clinical tataufbauteilen zu erreichen. Diese Passungenauigkeit, welche
situations. im Prinzip durch die individuelle Anfertigung der festsitzenden
The results from this experimental study are Rekonstruktion bedingt ist, setzt das Gerüst unter Spannung
suitable for performance of direct comparisons of und bewirkt zwangsweise Reaktionen. Es handelt sich um stati-
sche Kräfte, welche die Implantatkomponenten und den Kno-
the behaviour of the 3 considered abutments when chen an der Implantat-Knochen-Kontaktfläche belasten und zu
connected to prosthetic superstructures having po- Umbauvorgängen im Knochen führen können. Abhängig von
sition errors. The data obtained with the error de- der grösse dieser Kräfte, d.h. abhängig vom Ausmass der Pas-
vices do not give quantitative information on what sungenauigkeit, können die Umbauvorgänge im Knochen zur
Lockerung von Schrauben und zur Lockerung im Bereich der
happens in clinical applications where the implants Knochen-Implantat-Berührungsfläche führen und dadurch ei-
are connected to living bone which is a tissue much nen Implantatmisserfolg bewirken. Die vorliegende Studie zeigt
more deformable than the steel used for the error einen in vitro Vergleich von 3 verschiedenen Prothetikaufbauten
devices. (Standard, EsthetiCone und CerAdapt, Nobel Biocare AB, Gö-
The results show that system CerAdapt abut- teborg, Sweden) in Relation zu mechanischem Stress, welcher
durch geometrische Passungenauigkeiten induziert wird (tech-
ments/cemented prosthesis transfers the least nisch bedingte Fehler). Es wurden zwei experimentelle Vorrich-
strain in presence of translation errors, the system tungen entwickelt und angewendet, um zu testen, inwiefern die
standard abutments/screw retained prosthesis verschiedenen Prothetikteile Fehler kompensieren können. Eine
showed the highest capability of compensating ro- Vorrichtung wurde für die Messung von Translationsfehlern
und die andere für die Registrierung von Rotationsfehlern ent-
tation errors, whereas the system EsthetiCone wickelt. Der experimentelle Aufbau bestand aus zwei freistehen-
abutments showed the least capability. den Implantaten, welche eine prothetische Rekonstruktion und
Further studies are necessary to quantify the in- prothetische Aufbauteile trugen. Die Implantate und die Auf-
fluence of position errors on long-term success of bauteile wurden so verwendet, wie sie vom Hersteller geliefert
the implant-supported restorations in clinical prac- wurden, während die Rekonstruktionen individuell angefertigt
und mit Dehnmessstreifen versehen wurden. Die Daten, wel-
tice. sche mit den Fehlervorrichtungen aufgezeichnet wurden, erge-
ben keine quantitative Information über das, was bei der klini-
schen Anwendungen bei Implantaten, die mit vitalem Knochen
Résumé verbunden sind, passiert. Bei Knochen handelt es sich um vita-
En pratique privée, une adaptation parfaite de la prothèse sur les Gewebe, welches leichter deformierbar ist als der Stahl, der
l’implant est pratiquement impossible. Cette difficulté qui est für die Fehlervorrichtungen verarbeitet wurde. Die Resutate er-
principalement provoquée par le processus technologique utilisé lauben den direkten Vergleich des Verhaltens von den verschie-
pour fabriquer les prothèses fixes, met à l’épreuve la superstruc- denen untersuchten Prothetikteilen in Bezug auf Positionie-

455
Pietrabissa et al.
rungsfehler. Das CerAdapt System (zementierte keramische
Prothetikteile) zeigte die kleinste Spannung in Bezug auf Trans-
lationsfehler. Das Standardsystem (verschraubte Prothetikteile)
zeigte die kleinste Spannung in Bezug auf Rotationsfehler.

Resumen
En la practica, el ajuste perfecto de la estructura protésica con
los pilares de los implantes es casi imposible de conseguir. El
desajuste, que es inducido principalmente por los procesos tec-
nológicos utilizados para fabricar la prótesis fija, contrae la
estructura generando por ello reacciones a la contracción. Estas
son fuerzas estáticas que cargan los componentes del implante
y el hueso en la interfase hueso-implante y pueden causar remo-
delación ósea. Dependiendo de la magnitud de dicha fuerza,
i.e. dependiendo de la magnitud del desajuste, la remodelación
ósea puede conducir al aflojamiento de los tornillos y de la
interfase hueso-implante y por tanto causar fracaso del implan-
te. El presente estudio muestra la comparación in vitro de 3
pilares de conexión (standard, EsthetiCone y CerAdapt, Nobel
Biocare AB, Goteborg, Sweden) en relación con los estrés mecá-
nicos inducidos por desajustes geométricos (errores inducidos
tecnológicamente). Se realizaron dos dispositivos experimenta-
les a propósito y utilizados para valorar la habilidad de los
diferentes pilares para compensar los errores. Se diseñó uno
para compensar los errores de traslación y otro para los errores
de rotación. El montaje del aparato experimental incluye dos
implantes libres soportando una estructura protésica y sus pila-
res de conexión. Los implantes y sus pilares se usaron tal y
como se distribuyeron por el fabricante, mientras que las próte-
sis fueron fabricadas a propósito e instrumentadas con medido-
res de tensión. Los datos obtenidos con los dispositivos de error
no dan una información cuantitativa de los que ocurre en las
aplicaciones clı́nicas cuando los implantes se conectan a hueso
vivo que es un tejido mucho más deformable que el acero usado
para los dispositivos de error. Los resultados permiten compa-
raciones directas del comportamiento de los diferentes implan-
tes investigados respecto a los errores de posición. El sistema
CerAdapt (pilares cerámicos retenidos por cemento) mostraron
la menor tensión en presencia de errores de traslación. El siste-
References
ma estándar (pilares retenidos por tornillos) mostró la menor Adell, R., Lekholm, U., Rockler, B. & Brånemark, P.I. (1981)
tensión en presencia de errores de rotación. A 15-year study of osseointegrated implants in the treatment
of the edentulous jaw. International Journal of Oral Surgery
6: 387–416.
Albrektsson, T. (1988) A multicenter report on osseointegrated
oral implants. Journal of Prosthetic Dentistry 60: 75–84.
Albrektsson, T. & Sennerby, L. (1991) State of the art in oral
implants. Journal of Clinical Periodontology 18: 474–481.
Henry, P., Tolman, D. & Bolender, C. (1993) The applicability
of osseointegrated implants in the treatment of partially
edentulous patients: three-year results of a prospective multi-
center study. Quintessence International 24: 123–129.
Jemt, T. (1986) Modified single and short-span restorations
supported by osseointegrated fixtures in the partially eden-
tulous jaw. Journal of Prosthetic Dentistry 55: 243–247.
Jemt, T., Lekholm, U. & Adell, R. (1989) Osseointegrated im-
plants in the treatment of partially edentulous patients: a
preliminary study on 876 consecutively placed fixtures. In-
ternational Journal of Oral and Maxillofacial Implants 4:
211–217.
Jemt, T., Lindèn, B. & Lekholm, U. (1992) Failures and compli-
cations in 127 consecutively placed fixed partial prostheses
supported by Brånemark implants: from prosthetic treat-
ment to first annual check-up. International Journal of Oral
and Maxillofacial Implants 7: 40–44.
Lewis, S., Beumer, J.3d, Hornburg, W. & Moy, P. (1988) The

456
Compensation of mismatch of implant supported fixed prostheses
‘‘UCLA’’ abutment. International Journal of Oral and Max- porting fixed partial prostheses in the treatment of partial
illofacial Implants 3: 183–189. edentulism. Journal of Prosthetic Dentistry 61: 217–223.
Prestipino, V. & Ingber, A. (1993) Esthetic high-strength im- van Steenberghe, D., Brånemark, P.I., Quirynen, M., De Mars,
plant abutments. Part I. Journal of Esthetic Dentistry 5: 29– G. & Naert, I. (1991) The rehabilitation of oral defects by
36. osseointegrated implants. Journal of Clinical Periodontology
van Steenberghe, D. (1989) A retrospective multicenter evalu- 18: 488–493.
ation of the survival rate of osseointegrated fixtures sup-

457

You might also like