Professional Documents
Culture Documents
Helical Springs
Section : Bench #6
Rachel Chaggaris & Quinn Trussell
Professor Jerome Rifkin
Colorado School of Mines
1
● MiniTini mechanical test frame (#6)
● Controller
● Programmable power supply PSP-210
● BK Precision DC Power Supply
● Digital Multimeter
● Steel Ruler
● Twisted pair wires
● Red and black single wires
● Wire Cutters
● Red and black grabber leads
● Safety glasses
2
Figure 2.2 Wire connections for load cell [2].
3. Procedure
1. Acquire equipment and materials listed.
2. Use Figure 2.1 and Figure 2.2 to set up and wire equipment.
3. Calibrate the MiniTini
a. Test and record the range of motion of the MiniTini.
b. Record the voltage of the linear motion sensor when fully compressed.
c. Within the range limits, move the actuator to at least five different locations,
recording voltage readings and distance from compressed at each location.
d. Produce a graph of voltage vs. displacement to obtain the relationship of
volts/inches. This relationship is given with the equation of the line of best fit.
4. Attach the first spring (unstretched) to the MiniTini using the machine bolts and washers.
5. Record initial readings for the linear motion sensor and force sensor
6. Within the range limits, move the actuator to several different locations, (the more
readings, the more accurate the data) and record voltage readings and the distance from
the initial unstretched coil for each location.
7. Using the equation from step 3, part d, calculate the position at each location using
voltage readings of the linear motion sensor.
8. Using the calibration equation available on Blackboard, calculate the force applied using
voltage readings of the force sensor [3].
9. Repeat steps 4-8 for the second spring, as well as both springs in parallel.
10. Create a graph relating force and displacement. The slope of the line of best fit of this
data is the experimental spring constant (k) for each respective spring.
11. Compare the results with the calculated spring constant.
4. Results
3
4.1 Equations
4
Parallel Springs Force vs. Displacement Equation: (9)
𝑅 = 13.588𝑅 + 68.647
derived from the line of best fit of the graph of force (lb) vs. displacement (in) for spring 1 and
spring 2 in parallel. Refer to Figure 4.2.
5
Figure 4.1 Linear potentiometer calibration for MiniTini #6 with voltage output as a function of position.
The line of best fit equation is used to convert the voltage readings to the displacement of the springs to
observe their force vs. displacement relationship. The voltage increases as the position increases because
the positive and negative leads were accidently switched. This problem was fixed through a simple
calculation.
6
2.88 2.63 2.08 98.3
3.22 2.83 2.32 103.8
3.62 3.07 2.40 105.7
3.91 3.24 2.45 106.8
4.19 3.41 2.51 108.2
4.55 3.63 2.57 109.6
4.83 3.79 2.62 110.7
5.11 3.96 2.67 111.9
5.43 4.15 2.72 113.0
5.73 4.33 2.78 114.4
6.01 4.5 2.83 115.6
6.26 4.65 2.87 116.5
7
Figure 4.2 The linear relationship of force (lb) vs. displacement (in) observed in spring 1, spring 2, and
spring 1 & 2 in parallel. The slope of the line of best fit for each data set gives the spring constant (k) in
lb/in for the respective springs.
8
was 12.4%, which is comparatively high. This relatively higher error is most likely due to the
fact that there were two springs in the MiniTini apparatus, therefore increasing the chances of
error as there is more equipment that may contribute to the error. Despite the anticipated errors,
the experimentally determined spring constants can be accepted when comparing these constants
to the theoretical spring constants.
5. Error Analysis
5.2 Uncertainty
The digital multimeter and MiniTini provided measurements that were wired to be give readings
using the DC power supply, thus the uncertainty of these instruments are dependent on these
instruments ability to perform accurately and precisely. It is likely that the MEL laboratory
equipment provided could induce slight errors to the experiment. The ruler used to measure the
position for the linear potentiometer calibration for the MiniTini was accurate to 0.05 in.
6. Conclusion
6.1 Discussion
After completion of calculation and analysis of these calculations, this experiment proved to be
successful. All errors were within 5% representing precise and accurate measurements. The
experimental average of spring 1 and spring 2’s spring constant had a 9.92% error and spring 2’s
spring constant had a 14.2% error; these are relatively large experimental errors, but still confirm
the results of the lab. The larger error of 12.4% for the parallel springs was anticipated as there is
more room for error when using two springs at once in the MiniTini apparatus. This lab was an
effective way to measure the spring constants both as a single spring and in parallel.
6.2 Recommendations
The only recommendation for this lab would be the use of the DAQ and VI. This would give us a
more accurate reading of voltages at exponentially more points. It would be a constant, linear
axial force instead of stopping at different locations to take measurements. Other than this , there
are no recommendations for the lab since everything went smoothly.
6.3 Comments
9
This was an interesting lab that showed students how to properly use the lab equipment. The
instructions were clear and sufficient materials were posted beforehand to prepare students for
the lab.
References
[1] “Springs Experiment.” MEL Department. Spring 2018. https://elearning.mines.edu
[2] MEL Miner. “An introduction to Helical Springs.” YouTube. MEL Department - Colorado
School of Mines, 6 Aug. 2014. Web. 25 Feb 2018
[3] “Fall 2017 MiniTini Calibrations.” MEL Department. Spring 2018.
https://elearning.mines.edu
[4] “Springs Experiment Pre-Experiment Worksheet.” MEL Department. Spring 2018.
https://elearning.mines.edu
10