Radiography X-ray System

DM-5125
Operation Manual

509-ho, Unitechvil, 1141-2, Baegsuk-dong, Ilsandong-ku, Goyang-si,


Gyeonggi-do, Republic of Korea
Tel : 82-31-907-7222, Fax : 82-31-904-2231,
E-mail : dongmun@dong-mun.co.kr
Operation Manual: DM-5125

TABLE OF CONTENTS

PART 1. Generator
1. DM-100[12.5kW] ----------------------------------------------------------------
2. DM-200, 300[15kW] ----------------------------------------------------------------
3. 20kW/32kW/40kW/50kW/80kW ----------------------------------------------------------------
4. DIG-650RF[50kW 2-Tube] ----------------------------------------------------------------

PART 2. X-ray Tube


1. E7132X ----------------------------------------------------------------
2. E7239X ----------------------------------------------------------------
3. E7240X ----------------------------------------------------------------
4. E7252X ----------------------------------------------------------------

PART 3. Bucky Table


1. DM-FT-1 [Fixed Table-Top Table] ---------------------------------------------------------------
2. DM-BT-1 [4-way Table] ---------------------------------------------------------------
3. DM-RFT-5[RF Tilting Table] ---------------------------------------------------------------

PART 4. Tube Stand


1. DM-FCT-1[Floor to Ceiling Mounted] ----------------------------------------------------------------

PART 5. Bucky Stand


1. VB-1[Bucky Stand] ----------------------------------------------------------------

PART 6. High Tension Cable


1. High Tension Cable ----------------------------------------------------------------

PART 7. Collimator
1. LB-50[Collimator] ----------------------------------------------------------------

2/146
Operation Manual: DM-5125

PART 1. Generator

1. DM-100[12.5kW Generator]

3/146
Operation Manual: DM-5125

NOTICE 1

This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines
thoroughly before your using the generator.

The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No
responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which
many results from the use of this manual.

Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and
shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein.

Always keep the manual at hand for your reference.

NOTICE 2

1. This equipment should be used only by the legally qualified persons and practitioners.

2. For the safe installation


a. Please avoid wet and humid locations.
b. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust,
sulfur and other air pollutants.
c. Protect the product from slope, vibration, and shock.
d. Do not install the product where chemicals are stored or gases are generated.
e. Check the frequency, voltage, and current of power supply.
f. Check the earth ground power about the electric discharge and polarity.

3. Before the operation of the machine


a. Examine contact of switch, polarity, dial setting, and display, and check if the product properly operates.
b. Check the safety of the ground power connections.
c. Check the correctness and safety of the cable and plug connections.
d. Be careful that mixed use of the machine may give wrong diagnostic results and damages.
e. Check the exterior circuits, which occur direct physical contact to the patient.
f. Check the earth ground power about the electric discharge and polarity.

4. During the operation of the machine


a. Do not exceed the designated time for diagnosis and cures.
b. Monitor the conditions of device and influence on the patient.
c. When you find any troubles about generators or patients, stop the operation of the machine and take
appropriate measures for the safety of the patient.
d. Be careful that the patient should not touch the device.

4/146
Operation Manual: DM-5125

5. After the operation of the machine


a. Turn off the power of the machine after returning the operation switches and scales to the original position in
accordance with the prescribed regulation.
b. Do not pull out the plugs and lines with compelling forces and it can hurt the connections.
c. For the safe storage and installation
d. Please avoid wet and humid locations.
e. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust,
sulfur and other air pollutants.
f. Protect the product from slope, vibration, and shock.
g. Do not install the product where chemicals are stored or gases are generated.

6. If the device is broken, mark proper description and ask repairs to the engineer.

7. Do not modify the machine.

8. Inspections and maintenance


a. All device and parts should be checked on the regular basis.
b. If you operate the device after long period of non-use, check the safety and conditions of the machine.

9. Operate the machine in correct ways according to the manuals.

10. Be careful about following cautions


a. Wear the lead apron in order to prevent the x-ray irradiation.
b. Regular inspection should be executed about leakage dose and irradiation meter.
c. Be careful about the x-ray exposures.
d. Do not exceed the maximum weight for the x-ray tables.

NOTICE FOR THE OPERATION 1

1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate
measures.

2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off
the power during operation.

3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.

5/146
Operation Manual: DM-5125

NOTICE FOR THE OPERATION 2

1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic
drugs, oxygen and hydrogen near the working locations.

2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must
be checked after the occurrence of the earthquake.

3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use
this machine” and ask repairs to the engineers.

4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.

OTHER NOTICE AND ADVICE

1. Power on and off


Do not turn off the machine for 30 seconds after turning on the power. Please wait at least 30 seconds for the
resupply of the power.

2. Regarding the operation environment


a. X-ray room
Practitioners should take enough cares for the release of the x-rays out of the x-ray room. This device is
designed to be protective against humidity and dust. However, if you use the machine at the place with high
humidity and dust, it can shorten the life of electronic insult and grow rust on the metal parts.
b. Do not leave the machine exposed to the strong direct sunlight.

3. Exposure of the x-rays


a. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.
b. Take enough cares in order not to make excessive radiation on the same body locations.

4. Each cable has enough length and flexibility but do not bend the cable forcibly.

DANGEROUS

Do not use the machine with following equipment.


*The equipment, which release electronic waves such as mobile phone, radios, wireless toys, and etc. The
officials in hospital ought to give enough advises and education for the operator of the machine, doctors,
patient and other the third parties.

6/146
Operation Manual: DM-5125

SYMBOLS AND TERMS

The following symbols will be used in this manual: their meanings and applications are described further below.

It warns of the conditions and situations, if not taken over or avoided, that
DANGER cause heavy damages to people and even the death.

It warns of the conditions and situations, if not taken over or avoided, that
WARNING could cause heavy damages to people or cause unrepeatable damages to the
equipment.

It warns of the conditions and situations, if not taken over or avoided, that
CAUTION could cause damages to people, or to the equipment.

WAIVER CLAUSE

Dongmun do not take responsibility for the following causes for damage and repairs.

1. Any installation, movement, modification and repairs by other engineering services, which was not designated
by Dongmun.

2. Troubles and causes to our machine incurred by other equipment.

3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun.

4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual.

5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the
powers, installation and working places.

6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders.

7. Accident and damage happened by the use beyond the original purpose of the machine.

8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the
conditions and results of the medical prescriptions.

7/146
Operation Manual: DM-5125

SCOPE AND PERIOD OF WARRENTY

1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from
normal use and operation of product.

2. Service fee is charged for the following cases.


- Malfunctioning after the period of warranty
- Disorder caused by fire, earthquake, or lightening
- Inappropriate movement or negligence of product
- Disorder from repair or reconstruction by other than Dongmun service agent or other designated ones.

3. Note that defects not related to the main functions of the product are not eligible for this warranty.

4. Requirements for Service


- Stop using the product when a problem is found, and check the manual for a solution.
- Turn off the product before service request, and check the model number, serial number, the date of purchase
before calling the service center.

5. Defects or depreciation in appearance is not eligible for refund or change.

6. Dongmun is not responsible for indirect damage related to the use of products.

7. Dongmun is not responsible for injury or damage after the end of warranty period.

8. All detailed adaptability warranties and other warranties are subjected to this statement.

8/146
Operation Manual: DM-5125

SECTION 1. COMPOSITION / SPCECIFICATION

1-1. COMPOSITION

The electric composition block diagram of DIG-Series is shown in Fig 1-1.

EMI AD/DC HF
AC Input HV Tank
Filter Converter DC/AC Inverter
X-Ray
Tube

IGBT
Driver

Starter

Operation
Panel
HT-Controller

Transformer Power Collimator

Filament

Interface

Fig 1-1 COMPOSITION OF THE HFG

AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is
inverted in high-frequency voltage of about 20kHz by IGBT inverter.
The high-frequency voltage is converted into the 40kV to 125kV DC high voltage by a high-voltage transformer
and the multi-voltage rectifier circuit. This DC high voltage is supplied to an X-ray tube through a high-voltage
cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage controls the tube current.

The operation panel can operate selection in each mode of operation and control, selection and setting of
conditions. These operation conditions are transmitted by the communication system from OP-CONT to HT-
CONT. HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.

9/146
Operation Manual: DM-5125

The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.

Operation Panel

High Voltage Generator

Main CPU

A/D Converter D/A Converter

kV F/B
PWM
PWM

Filament mA F/B
kV & mA Feedback Amplifier

Drive Circuit RMS DC


Converter

Filament
IGBT
Inverter

HV Transformer
kV
mA
Filament Transformer

A C
FILAMENT
X-RAY Tube

Fig 1-2 CONTROL SIGNAL OF X-RAY EXPOSURE

10/146
Operation Manual: DM-5125

1-2. SPECIFICATION

1 Rating Out-put 12.5 kW


2 Tube Voltage kV 40 ~ 125kV (1kV step incr./Decr.)
3 Tube Current mA 20S, 40S, 75S,100S mA
4 ‘kV Accuracy ±10 %
5 ‘mA Accuracy ±15 %
6 mAs Range 0.4 ~ 320 mAs
7 Digital Display Digital display (LED & 7 - Segment)
8 Setting Method Sheet Push Switch
9 Self-Diagnosis Code display in LED digital
10 Selection Manual / APR
11 APR 15 Region
Display (OP panel)
‘kV 3 Digital LED
12 ‘mAs 3 Digital LED
Selection / Setting LED (12)
Alarm LED (1)
Weight
13 HFG 14 kg
Op panel 1.5 kg

1-3. ELECTRICAL REQUIREMENT

Model DIG SERIES


Line Nominal Voltage Single Phase 200V / 220V / 240V
Automatic Line Compensation Range ±10 %
Frequency 50 Hz / 60 Hz
Line Impedance Less Than 0.12 Ω
Grounding conditions Based on the local regulation or 3 independent ground circuit

1-4. PROPER ENVIRONMENT

1. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.

11/146
Operation Manual: DM-5125

g. Where the equipment is exposed to chemicals or gas.

2. Keep away from the place with strong vibration and maintain proper environment and conditions.

OPERATION ENVIRONMENT
Temperature range + 10°C ~ + 40°C,
Relative Humidity Range 30% ~ 75% RH (No dews)
Others No Direct Sunlight
No combustible and corrosive gases

3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere.

STORAGE AND TRANSPORTATION


Temperature range -25°C ~ +60°C
Relative Humidity Range 10% ~ 95% RH(No dews)
Air Pressure 700hPa ~ 1060hPa
Others No corrosive gases

12/146
Operation Manual: DM-5125

1-5. WIRE CONNECTION

1) FRONT PART

① (CN204) ② (TB502) ③ (TB3) ④ (TB1)


PIN No. Connect to External Main Power Connect to Main
Connect to OP
Device Selection Power
1 Rx+ AC 12V AC 240V AC INPUT
2 Rx- AC 12V RTN AC 220V F.G
3 Tx+ AC 100V AC 200V AC INPUT
4 Tx- AC 100V RTN
5 EXP BUCKY
6 GND BUCKY RTN
7 GND EXT
8 DC 24V EXT RTN
9 DC 24V
10 DC 24V RTN
11 SWITCH
12 SWITCH RTN

13/146
Operation Manual: DM-5125

2) REAR PART

⑤ (CN107) ⑥ (CN108) ⑦ (TB102) ⑧ (TB2)


PIN No. Connect to Connect to Connect to TUBE
Connect to TANK
TANK TANK (ROTOR)
1 EP F/B COM ROTOR1 OUTPUT
2 IP F/B L1 ROTOR2 OUTPUT
3 GND L2 ROTOR3
4 GND S1
5 INTERLOCK S2
6 NC

14/146
Operation Manual: DM-5125

1-6. ADDITIONAL FUNCTIONS

DIP S/W FUNCTION

1) REMOTE HT-CONTROLLER

DIP S/W No. Function Remark

DIP S/W 1 Not defined


Focus Selection
DIP S/W 2 On Large Focus
Off Small Focus
Apply to mass chest x-ray
DIP S/W 3 On Use mass chest x-ray
Off Not use mass chest x-ray
Bucky Type Selection
DIP S/W 4 On Motor Type Bucky
Off Solenoid Type Bucky

2) OPERATION PANEL

DIP S/W No. Function Remark

DIP S/W 1 Current Data Calibration Mode

DIP S/W 2 Preheat Data Calibration Mode

DIP S/W 3 Serial EEPROM Initialize Mode

DIP S/W 4 RAM Monitor Mode

15/146
Operation Manual: DM-5125

SECTION 2. DISCRIPTIONS / FUNCTIONS

2-1. OUTLINE AND DIMENSIONS

FIG 2-1. DIMENSION OF HFG

FIG 2-2. DISPLAY OF OP PANEL

16/146
Operation Manual: DM-5125

2-2. OPERATION PANEL DISPLAY

FIG. 2-3 OP PANEL DISCRIPTION

1 Size Display / Selection Switch (Large) 12 Exposure wait display (Yellow)

2 Size Display / Selection Switch (Middle) 13 Error Display (Red)

3 Size Display / Selection Switch (Small) 14 Tube kV Display

4 APR-1 Display / Selection Switch 15 mAs Display

5 APR-2 Display / Selection Switch 16 Tube kV Down Switch

6 APR-3 Display / Selection Switch 17 Tube kV Up Switch

7 APR-4 Display / Selection Switch 18 mAs Down Switch

8 APR-5 Display / Selection Switch 19 mAs Up Switch

9 APR Storage Switch 20 Bucky Display/Selection Switch

10 X-ray exposure ready display (Green) 21 X-ray exposure ready Switch

11 X-ray exposure Display (Yellow) 22 X-ray exposure Switch

17/146
Operation Manual: DM-5125

1) SIZE DISPLAY/SELECTION SWITCH (LARGE)

Make selection for body size “large” under APR radiography.


Choose the APR No. 1 to 5 after the selection of Body size.

2) SIZE DISPLAY/SELECTION SIWTCH (MIDDLE)

Make selection for body size “Middle” under APR radiography.


Choose the APR No. 1 to 5 after the selection of Body size.

3) SIZE DISPLAY/SELECTION SIWTCH (SMALL)

Make selection for body size “Small” under APR radiography.


Choose the APR No. 1 to 5 after the selection of Body size.

4) APR-1 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.

5) APR-2 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.

6) APR-3 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.

18/146
Operation Manual: DM-5125

7) APR-4 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.

8) APR-5 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.

9) APR STORAGE SWITCH

Save the changed value of kV and mAs for APR.


If this switch is pushed longer than 0.5 sec after the selection of kV and mAs, the data will be saved with the
buzzer sound.

10) X-RAY EXPOSURE READY LED (GREEN LED)


This LED is lighted when x-ray exposure is ready.

11) X-RAY EXPOSURE LED (YELLOW LED)


This LED is lighted during the x-ray exposure.

12) EXPOSURE WAIT LED (YELLOW LED)


No function.

13) ERROR LED


This LED is lighted during the occurrence of Error.

14) TUBE kV DISPLAY


Indicate the kV of radiography.

15) mAs DISPLAY


Indicate the mAs of radiography.

19/146
Operation Manual: DM-5125

16) TUBE kV DOWN SWITCH


Operate the setting of tube kV. Move 1 kV by pressing the switch one time and will be changed consecutively
by keep pressing the button. Setting range is 40~125kV.

17) TUBE kV UP SWITCH


Operate the setting of tube kV. Move 1 kV by pressing the switch one time and will be changed consecutively
by keep pressing the button. Setting range is 40~125kV.

18) mAs DOWN SWITCH


Operate the setting of mAs for radiography. Setting range is 0.4~320 mAs.

19) mAs UP SWITCH


Operate the setting of mAs for radiography.

20) BUCKY DISPLAY / SELECTION SWITCH

Select and display the sort of Bucky. The LED will be turned on during Bucky 1 and turned off at non-bucky.

21) X-RAY EXPOSURE READY SWITCH

When the ready switch is pushed, it rotate the rotor, control the heating of filament and make the exposure at
ready condition.

22) X-RAY EXPOSURE SWITCH (X-RAY SWITCH)

Make the exposure of x-rays by pressing this siwtch after the ready LED is displayed.
Touch off from the button after the radiography is completed.

20/146
Operation Manual: DM-5125

SECTION 3. OPERATION / EXPOSURE

3-1. SELECTION OF THE RADIOGRAPHY METHOD

There are radiography modes shown below in this equipment.

1. Manual radiography mode


Radiography conditions are performed by manual setup.

2. APR radiography mode


You can select the radiography region; direction and patient size on the control panel and make x-ray exposure
with automatically set radiography conditions.
Each radiography conditions can be changed and saved.

3-2. SETTING MANUAL RADIOGRAPHY

1. Select the manual radiography mode


Check if the APR mode is chosen. If they are selected in APR mode, please press APR setting switch, and
release the setup conditions.

2. Setting tube voltage "kV"


Set up the tube voltage with “kV” setting switch.
Setting range: 40kV – 125 kV (1kv step)

3. Setting tube current “mA”


Set up the tube current with “mA” setting switch.
Setting range: 0.4~320 mAs
0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.4, 8.0, 10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125,
160, 200, 250, 320 mAs

3-3. SETTING APR RADIOGRAPHY

1. Selection of APR mode


APR mode is selected by size and number of APR.

2. Selection of Radiography Regions


Choose the [L], [M], [S] radiography with setting switch.

3. Selection of APR numbers


Choose the number out of No. 1~5 after the selection of the size

21/146
Operation Manual: DM-5125

4. Change and storage


Change the settings with tube kV down and up switch after the selection of APR according to the size and APR no.

3-4. RADIOGRAPHY

1. PREPARATION
After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table. If the
X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the lamp will
be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography preparation is
ended.

2. X-RAY RADIOGRAPHY BY RADIOGRAPHY SWITCH ON THE PANEL


1) Press the X-ray exposure preparation switch. Ready LED will be lighted on when all preparation work is
completed.
2) If you press the x-ray preparation switch and exposure switch at the same time, the x-ray shooting will be
started.
3) X-rays exposure LED will be lighted on and the buzzer sounds during the exposure.

3. X-ray Radiography Using Hand-Switch


1) Press the hand switch one time and make it at ready condition. The x-ray ready LED will be lighted on when
it is prepared to make exposure.
2) When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started.
3) During exposure, the exposure LED will be lighted on and the buzzer sounds.
4) If the x-ray exposure LED is lighted or the buzzer is off, touch off from the switch and close the radiography
procedures.

If you touch out from the hand switch during exposure, the x-ray
shooting is stopped and you cannot get clear pictures. Please touch
CAUTION off your hand after the x-ray exposure signs and the buzzer sound are
turned off except the case you want to stop the x-ray exposure
urgently

22/146
Operation Manual: DM-5125

PART 1. Generator

2. DM-200[15kW Generator]

23/146
Operation Manual: DM-5125

NOTICE 1

This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines
thoroughly before your using the generator.

The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No
responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which
many results from the use of this manual.

Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and
shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein.

Always keep the manual at hand for your reference.

NOTICE 2

1. This equipment should be used only by the legally qualified persons and practitioners.

2. For the safe installation


a. Please avoid wet and humid locations.
b. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust,
sulfur and other air pollutants.
c. Protect the product from slope, vibration, and shock.
d. Do not install the product where chemicals are stored or gases are generated.
e. Check the frequency, voltage, and current of power supply.
f. Check the earth ground power about the electric discharge and polarity.

3. Before the operation of the machine


a. Examine contact of switch, polarity, dial setting, and display, and check if the product properly operates.
b. Check the safety of the ground power connections.
c. Check the correctness and safety of the cable and plug connections.
d. Be careful that mixed use of the machine may give wrong diagnostic results and damages.
e. Check the exterior circuits, which occur direct physical contact to the patient.
f. Check the earth ground power about the electric discharge and polarity.

4. During the operation of the machine


a. Do not exceed the designated time for diagnosis and cures.
b. Monitor the conditions of device and influence on the patient.
c. When you find any troubles about generators or patients, stop the operation of the machine and take
appropriate measures for the safety of the patient.
d. Be careful that the patient should not touch the device.

24/146
Operation Manual: DM-5125

5. After the operation of the machine


a. Turn off the power of the machine after returning the operation switches and scales to the original position in
accordance with the prescribed regulation.
b. Do not pull out the plugs and lines with compelling forces and it can hurt the connections.
c. For the safe storage and installation
d. Please avoid wet and humid locations.
e. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust,
sulfur and other air pollutants.
f. Protect the product from slope, vibration, and shock.
g. Do not install the product where chemicals are stored or gases are generated.

6. If the device is broken, mark proper description and ask repairs to the engineer.

7. Do not modify the machine.

8. Inspections and maintenance


a. All device and parts should be checked on the regular basis.
b. If you operate the device after long period of non-use, check the safety and conditions of the machine.

9. Operate the machine in correct ways according to the manuals.

10. Be careful about following cautions


a. Wear the lead apron in order to prevent the x-ray irradiation.
b. Regular inspection should be executed about leakage dose and irradiation meter.
c. Be careful about the x-ray exposures.
d. Do not exceed the maximum weight for the x-ray tables.

NOTICE FOR THE OPERATION 1

1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate
measures.

2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off
the power during operation.

3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.

25/146
Operation Manual: DM-5125

NOTICE FOR THE OPERATION 2

1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic
drugs, oxygen and hydrogen near the working locations.

2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must
be checked after the occurrence of the earthquake.

3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use
this machine” and ask repairs to the engineers.

4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.

OTHER NOTICE AND ADVICE

1. Power on and off


Do not turn off the machine for 30 seconds after turning on the power. Please wait at least 30 seconds for the
resupply of the power.

2. Regarding the operation environment


a. X-ray room
Practitioners should take enough cares for the release of the x-rays out of the x-ray room. This device is
designed to be protective against humidity and dust. However, if you use the machine at the place with high
humidity and dust, it can shorten the life of electronic insult and grow rust on the metal parts.
b. Do not leave the machine exposed to the strong direct sunlight.

3. Exposure of the x-rays


a. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.
b. Take enough cares in order not to make excessive radiation on the same body locations.

4. Each cable has enough length and flexibility but do not bend the cable forcibly.

DANGEROUS

Do not use the machine with following equipment.


*The equipment, which release electronic waves such as mobile phone, radios, wireless toys, and etc. The
officials in hospital ought to give enough advises and education for the operator of the machine, doctors,
patient and other the third parties.

26/146
Operation Manual: DM-5125

SYMBOLS AND TERMS

The following symbols will be used in this manual: their meanings and applications are described further below.

It warns of the conditions and situations, if not taken over or avoided, that
DANGER cause heavy damages to people and even the death.

It warns of the conditions and situations, if not taken over or avoided, that
WARNING could cause heavy damages to people or cause unrepeatable damages to the
equipment.

It warns of the conditions and situations, if not taken over or avoided, that
CAUTION could cause damages to people, or to the equipment.

WAIVER CLAUSE

Dongmun do not take responsibility for the following causes for damage and repairs.

1. Any installation, movement, modification and repairs by other engineering services, which was not designated
by Dongmun.

2. Troubles and causes to our machine incurred by other equipment.

3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun.

4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual.

5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the
powers, installation and working places.

6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders.

7. Accident and damage happened by the use beyond the original purpose of the machine.

8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the
conditions and results of the medical prescriptions.

27/146
Operation Manual: DM-5125

SCOPE AND PERIOD OF WARRENTY

1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from
normal use and operation of product.

2. Service fee is charged for the following cases.


- Malfunctioning after the period of warranty
- Disorder caused by fire, earthquake, or lightening
- Inappropriate movement or negligence of product
- Disorder from repair or reconstruction by other than Dongmun service agent or other designated ones.

3. Note that defects not related to the main functions of the product are not eligible for this warranty.

4. Requirements for Service


- Stop using the product when a problem is found, and check the manual for a solution.
- Turn off the product before service request, and check the model number, serial number, the date of purchase
before calling the service center.

5. Defects or depreciation in appearance is not eligible for refund or change.

6. Dongmun is not responsible for indirect damage related to the use of products.

7. Dongmun is not responsible for injury or damage after the end of warranty period.

8. All detailed adaptability warranties and other warranties are subjected to this statement.

28/146
Operation Manual: DM-5125

SECTION 1. COMPOSITION / SPCECIFICATION

1-1. COMPOSITION

The electric composition block diagram of DIG-Series is shown in Fig 1-1.

EMI AD/DC HF
AC Input HV Tank
Filter Converter DC/AC Inverter
X-Ray
Tube

IGBT
Driver

Starter

Operation
Panel
HT-Controller

Transformer Power Collimator

Filament

Interface

Fig 1-1 COMPOSITION OF THE HFG

AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is
inverted in high-frequency voltage of about 20kHz by IGBT inverter.
The high-frequency voltage is converted into the 40kV to 125kV DC high voltage by a high-voltage transformer
and the multi-voltage rectifier circuit. This DC high voltage is supplied to an X-ray tube through a high-voltage
cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage controls the tube current.

The operation panel can operate selection in each mode of operation and control, selection and setting of
conditions. These operation conditions are transmitted by the communication system from OP-CONT to HT-
CONT. HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.

29/146
Operation Manual: DM-5125

The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.

Operation Panel

High Voltage Generator

Main CPU

A/D Converter D/A Converter

kV F/B
PWM
PWM

Filament mA F/B
kV & mA Feedback Amplifier

Drive Circuit RMS DC


Converter

Filament
IGBT
Inverter

HV Transformer
kV
mA
Filament Transformer

A C
FILAMENT
X-RAY Tube

Fig 1-2 CONTROL SIGNAL OF X-RAY EXPOSURE

30/146
Operation Manual: DM-5125

1-2. SPECIFICATION

Output Power 15 kW
kV Range 40 ~ 125kV (1kV step incr./Decr.)
Small Focus 50, 100, 150
mA Range
Large Focus 200
Time Range 0.004 ~ 6.0sec in 33 Steps
mAs Range 0.4 ~ 320mAs in 30 Steps
kV Accuracy ±10 %
mA Accuracy ±15 %
Digital Display Digital Display (LED & 7 - Segment)
Self-Diagnosis Code Display in LED
APR 30 Memories
kV 3 Digital LED
Display mA 3 Digital LED
Sec/mAs 4 Digital LED
Generator 280W X 370H X 452D
Dimension
Operator Panel 161W X 53H X 367D
Generator 14Kg
Weight
Operator Panel 2Kg

1-3. ELECTRICAL REQUIREMENT

Model DIG SERIES


Line Nominal Voltage Single Phase 200V / 220V / 240V
Automatic Line Compensation Range ±10 %
Frequency 50 Hz / 60 Hz
Line Impedance Less Than 0.12 Ω
Grounding conditions Based on the local regulation or 3 independent ground circuit

1-4. PROPER ENVIRONMENT

1. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.

31/146
Operation Manual: DM-5125

g. Where the equipment is exposed to chemicals or gas.

2. Keep away from the place with strong vibration and maintain proper environment and conditions.

OPERATION ENVIRONMENT
Temperature range + 10°C ~ + 40°C,
Relative Humidity Range 30% ~ 75% RH (No dews)
Others No Direct Sunlight
No combustible and corrosive gases

3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere.

STORAGE AND TRANSPORTATION


Temperature range -25°C ~ +60°C
Relative Humidity Range 10% ~ 95% RH(No dews)
Air Pressure 700hPa ~ 1060hPa
Others No corrosive gases

32/146
Operation Manual: DM-5125

1-5. WIRE CONNECTION

1) FRONT PART

12
CN204

1
TB502
TB1
TB3

1 1

3 3

① (CN204) ② (TB502) ③ (TB3) ④ (TB1)


PIN No. Connect to External Main Power Connect to Main
Connect to OP
Device Selection Power
1 Rx+ AC 12V AC 240V AC INPUT
2 Rx- AC 0V AC 220V F.G
3 Tx+ SWITCH AC 200V AC INPUT
4 Tx- SWITCH RTN
5 EXP AC 100V(Bucky 2)
6 GND AC 0V(Bucky 2)
7 GND AC 100V(Bucky 1)
8 DC 24V AC 0V(Bucky 1)
9 EXP 100V
10 EXP 0V
11 DC 24V
12 DC 0V

33/146
Operation Manual: DM-5125

2) REAR PART

3
TB102
1

CN108

TB2 CN107

2
1

⑤ (CN107) ⑥ (CN108) ⑦ (TB102) ⑧ (TB2)


PIN No. Connect to Connect to Connect to TUBE
Connect to TANK
TANK TANK (ROTOR)
1 EP F/B COM ROTOR1 OUTPUT
2 IP F/B L1 ROTOR2 OUTPUT
3 GND L2 ROTOR3
4 GND S1
5 INTERLOCK S2
6 NC

34/146
Operation Manual: DM-5125

1-6. ADDITIONAL FUNCTIONS

DIP S/W FUNCTION

1) REMOTE HT-CONTROLLER
DIP S/W No. Function Remark
DIP S/W 1 Not Defined
DIP S/W 2 Not Defined
Apply to mass chest x-ray
DIP S/W 3 ON Use mass chest x-ray
OFF Not use mass chest x-ray
Bucky Type Selection
DIP S/W 4 ON Motor Type Bucky
OFF Solenoid Type Bucky

2) OPERATION PANEL
DIP S/W No. Function Remark
ON Calibration Mode
DIP S/W 1
OFF Normal Mode
ON -
DIP S/W 2
OFF -
ON -
DIP S/W 3
OFF -
ON RAM Monitor Mode
DIP S/W 4
OFF Normal Mode
ON -
DIP S/W 5
OFF -
ON -
DIP S/W 6
OFF -
ON -
DIP S/W 7
OFF -
ON X-ray END Buzzer
DIP S/W 8
OFF X-ray ON Buzzer

35/146
Operation Manual: DM-5125

SECTION 2. DISCRIPTIONS / FUNCTIONS

2-1. OUTLINE AND DIMENSIONS

FIG 2-1. DIMENSION OF HFG

FIG 2-2. DISPLAY OF OP PANEL

36/146
Operation Manual: DM-5125

2-2. OPERATION PANEL DISPLAY

1 12 2 15 3 4 16 5 6 7

9 10 11 13 14 17 18

FIG. 2-3 OP PANEL DISCRIPTION

1 Patient Size Selection Switch 10 APR Region Selection Switch & Lamp

2 “Over” Lamp 11 APR Direct Selection Switch

3 “Error” Lamp 12 kV Setting Switch & Display

4 Small Focus Lamp 13 Backup Release Switch

5 Large Focus Lamp 14 Bucky Selection Switch & Display

6 X-ray Ready Lamp 15 mA Selection Switch & Display

7 X-ray Exposure Lamp 16 mAs/Sec Setting Switch & Display

8 mA/Sec Selection Switch & Lamp 17 X-ray Ready Switch

9 APR Mode Selection Switch 18 X-ray Exposure Switch

37/146
Operation Manual: DM-5125

1) Patient Size Selection Switch


It operates the selections of patient size. The selected size will be displayed on the segment. Save the changed
APR conditions by pressing the size switches.
You can set the exposure conditions (kV, mA and mAs/Sec) by selecting the exposure regions, direction and
patient size. Each condition can be changed and stored by the exposure size switch.

2) “Over” Lamp
The over lamp is lighted on when the exposure conditions exceeds the normal standards. You cannot make
exposure while the over lamp is lighted on.

3) “Error” Lamp
The lamp is lighted in case of errors.

4) Small Focus Lamp


The lamp is lighted on when the small focus is selected.

5) Large Focus Lamp


The lamp is lighted on when the large focus is selected.

6) X-ray Ready Lamp


The X-ray ready lamp is lighted on when it is at the ready condition.

7) X-ray Exposure Lamp


The X-ray exposure lamp is lighted on during the X-ray exposure.

8) mAs/Sec Selection Switch & Lamp


The mAs or Sec radiography mode is selected.
The selected radiography mode is displayed with lamp.

9) APR Mode Selection Switch


It operates the on/off of the APR mode.

10) APR Region Selection Switch & Lamp


It operates the selection of APR regions and the selected location will be displayed on the lamp.

11) APR Direction Selection Switch


It operates the selection of ARP directions the selected direction will be displayed in the lamp.

12) kV Setting Switch & Display


Up-button increases the kV value of X-ray tube whereas down-button decreases kV. If you keep pressing
buttons, the value of kV will increase or decrease continuously.

38/146
Operation Manual: DM-5125

13) Backup Release Switch


Release the back-up conditions.

14) Bucky Selection Switch & Display


Select and display Bucky 1 (Table) or Bucky 2 (Stand).

15) mA Setting Switch & Display


Up-button increases the mA value of X-ray tube whereas down-button decreases mA. If you keep pressing
buttons, the value of mA will increase or decrease continuously.

16) mAs/Sec Setting Switch & Display


Up-button increases the mAs/Sec value of X-ray tube whereas down-button decreases mAs/Sec. If you keep
pressing buttons, the value of mAs/Sec will increase or decrease continuously.

17) X-ray Ready Switch


It circulates the rotor and controls the heat of the filament that makes ready actions for the X-ray exposure.

18) X-ray Exposure Switch


Press the X-ray exposure switch after the [6] X-ray ready lamp is lighted on.

39/146
Operation Manual: DM-5125

SECTION 3. OPERATION / EXPOSURE

3-1. SELECTION OF THE RADIOGRAPHY METHOD

There are radiography modes shown below in this equipment.

1. Manual radiography mode


Radiography conditions are performed by manual setup.

2. APR radiography mode


You can select the radiography region; direction and patient size on the control panel and make x-ray exposure
with automatically set radiography conditions.
Each radiography conditions can be changed and saved.

3-2. SETTING MANUAL RADIOGRAPHY

1. Select the manual radiography mode


Check if the APR mode is chosen. If they are selected in APR mode, please press APR setting switch, and
release the setup conditions.

2. Setting tube voltage "kV"


Set up the tube voltage with “kV” setting switch.
Setting range: 40kV – 125 kV (1kv step)

3. Setting tube current “mA”


Set up the tube current with “mA” setting switch.
Setting range: 50, 100, 150mA (Small Focus), 200, 300mA (Large Focus)

4. Setting the tube current-time “mAs” and exposure-time “sec”


Select the radiography mode with mAs/Sec selection switch.

The “Sec” radiography makes X-ray exposure by setting time.


The “mAs” radiography makes X-ray exposure by setting current-time.

A setting of mAs values in mAs mode.


mAs value is chosen with the mAs/Sec setting switch.
Setting range: 0.4 ~ 320mAs in 30 Steps
Step: 0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10.0, 12.5, 16.0, 20.0, 25.0, 32.0,
40.0, 50.0, 63.0, 80.0, 100, 125, 160, 200, 250, 320mAs

40/146
Operation Manual: DM-5125

A setting of Sec values in Sec mode.


Sec value is chosen with the mAs/Sec setting switch.
Setting range: 0.004 ~ 6.0sec in 33 Steps
Step: 0.004, 0.005, 0.006, 0.008, 0.010, 0.013, 0.016, 0.020, 0.025, 0.032, 0.040, 0.050, 0.063, 0.080,
0.100, 0.125, 0.160, 0.200, 0.250, 0.320, 0.400, 0.500, 0.630, 0.800, 1.00, 1.25, 1.60, 2.00, 2.50,
3.20, 4.00, 5.00, 6.00sec

NOTICE
About “OVER” display of radiographic conditions
When set-up radiographic condition “kV”, “mA” and “mAs” or “sec” exceeds the permissible value of
equipment, OVER sign lights up.
X-ray exposure is prohibited when the over sign is turned on. Please reset the appropriate radiographic
conditions.

3-3. SETTING APR RADIOGRAPHY

1. Selection of APR mode


Choose APR mode with the [9] APR selection switch and the lmap displayed APR mode will be lighted on the
upper left of the OP Panel.

2. Selection of Radiography Regions


Choose the radiographic regions with [10] APR region selection switch and you can move the lamp display
with up/down button. Available option is 5.

3. Selection of Radiographic Directions


Choose the radiographic directions with [11] APR direction selection switch out of 2 options.

4. Selection of Radiographic Size


Choose the radiographic size with [1] Patient Size selection switch out of 3 options, [L]. [M], [S].

5. Change of the Radiographic Conditions


You can change the selected radiographic conditions with the same method of the 3-2 manual radiographic
condition setting. The over lamp can be lighted from time to time when the radiographic conditions change.
Please be careful in setting of the radiographic conditions since the X-ray exposure is prohibited under the case.

6. Memory of the Changed Radiographic Conditions


You can save the changed conditions above. How to store is pressing the selected switch for 2 second until the
buzzer beeps.

Memory of the changed radiographic condition


Be careful that other radiographic conditions can be selection if you
CAUTION choose the selected radiographic size switch. You cannot save the
memory while the over lamp is lighted on.

41/146
Operation Manual: DM-5125

3-4. RADIOGRAPHY

1. PREPARATION
After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table. If the
X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the lamp will
be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography preparation is
ended.

2. X-RAY RADIOGRAPHY BY RADIOGRAPHY SWITCH ON THE PANEL


1) Press the X-ray exposure preparation switch. Ready LED will be lighted on when all preparation work is
completed.
2) If you press the x-ray preparation switch and exposure switch at the same time, the x-ray shooting will be
started.
3) X-rays exposure LED will be lighted on and the buzzer sounds during the exposure.

3. X-ray Radiography Using Hand-Switch


1) Press the hand switch one time and make it at ready condition. The x-ray ready LED will be lighted on when
it is prepared to make exposure.
2) When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started.
3) During exposure, the exposure lamp will be lighted on and the buzzer sounds.
4) If the X-ray exposure lamp is lighted or the buzzer is off, it means the radiography work is completed.
Then, you can stop pressing the hand switch.
5) If you get your hands off and stopped X-ray shooting during exposure, the check code will be displayed.

If you touch out from the hand switch during exposure, the x-ray
shooting is stopped and you cannot get clear pictures. Please touch
CAUTION off your hand after the x-ray exposure signs and the buzzer sound are
turned off except the case you want to stop the x-ray exposure
urgently

You cannot make radiography while the over lamp is lighted on.
CAUTION Change the conditions for the radiography in those cases.

42/146
Operation Manual: DM-5125

PART 1. Generator

3. DIG-325/525/650/1150
[20/32/40/50/80kW Generator]

43/146
Operation Manual: DM-5125

NOTICE 1

This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines
thoroughly before your using the generator.

The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No
responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which
many results from the use of this manual.

Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and
shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein.

Always keep the manual at hand for your reference.

NOTICE 2

1. This equipment should be used only by the legally qualified persons and practitioners.

2. For the safe installation


a. Please avoid wet and humid locations.
b. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust,
sulfur and other air pollutants.
c. Protect the product from slope, vibration, and shock.
d. Do not install the product where chemicals are stored or gases are generated.
e. Check the frequency, voltage, and current of power supply.
f. Check the earth ground power about the electric discharge and polarity.

3. Before the operation of the machine


a. Examine contact of switch, polarity, dial setting, and display, and check if the product properly operates.
b. Check the safety of the ground power connections.
c. Check the correctness and safety of the cable and plug connections.
d. Be careful that mixed use of the machine may give wrong diagnostic results and damages.
e. Check the exterior circuits, which occur direct physical contact to the patient.
f. Check the earth ground power about the electric discharge and polarity.

4. During the operation of the machine


a. Do not exceed the designated time for diagnosis and cures.
b. Monitor the conditions of device and influence on the patient.
c. When you find any troubles about generators or patients, stop the operation of the machine and take
appropriate measures for the safety of the patient.
d. Be careful that the patient should not touch the device.

44/146
Operation Manual: DM-5125

5. After the operation of the machine


a. Turn off the power of the machine after returning the operation switches and scales to the original position in
accordance with the prescribed regulation.
b. Do not pull out the plugs and lines with compelling forces and it can hurt the connections.
c. For the safe storage and installation
d. Please avoid wet and humid locations.
e. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust,
sulfur and other air pollutants.
f. Protect the product from slope, vibration, and shock.
g. Do not install the product where chemicals are stored or gases are generated.

6. If the device is broken, mark proper description and ask repairs to the engineer.

7. Do not modify the machine.

8. Inspections and maintenance


a. All device and parts should be checked on the regular basis.
b. If you operate the device after long period of non-use, check the safety and conditions of the machine.

9. Operate the machine in correct ways according to the manuals.

10. Be careful about following cautions


a. Wear the lead apron in order to prevent the x-ray irradiation.
b. Regular inspection should be executed about leakage dose and irradiation meter.
c. Be careful about the x-ray exposures.
d. Do not exceed the maximum weight for the x-ray tables.

NOTICE FOR THE OPERATION 1

1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate
measures.

2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off
the power during operation.

3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.

45/146
Operation Manual: DM-5125

NOTICE FOR THE OPERATION 2

1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic
drugs, oxygen and hydrogen near the working locations.

2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must
be checked after the occurrence of the earthquake.

3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use
this machine” and ask repairs to the engineers.

4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.

OTHER NOTICE AND ADVICE

1. Power on and off


Do not turn off the machine for 30 seconds after turning on the power. Please wait at least 30 seconds for the
resupply of the power.

2. Regarding the operation environment


a. X-ray room
Practitioners should take enough cares for the release of the x-rays out of the x-ray room. This device is
designed to be protective against humidity and dust. However, if you use the machine at the place with high
humidity and dust, it can shorten the life of electronic insult and grow rust on the metal parts.
b. Do not leave the machine exposed to the strong direct sunlight.

3. Exposure of the x-rays


a. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.
b. Take enough cares in order not to make excessive radiation on the same body locations.

4. Each cable has enough length and flexibility but do not bend the cable forcibly.

DANGEROUS

Do not use the machine with following equipment.


*The equipment, which release electronic waves such as mobile phone, radios, wireless toys, and etc. The
officials in hospital ought to give enough advises and education for the operator of the machine, doctors,
patient and other the third parties.

46/146
Operation Manual: DM-5125

SYMBOLS AND TERMS

The following symbols will be used in this manual: their meanings and applications are described further below.

It warns of the conditions and situations, if not taken over or avoided, that
DANGER cause heavy damages to people and even the death.

It warns of the conditions and situations, if not taken over or avoided, that
WARNING could cause heavy damages to people or cause unrepeatable damages to the
equipment.

It warns of the conditions and situations, if not taken over or avoided, that
CAUTION could cause damages to people, or to the equipment.

WAIVER CLAUSE

Dongmun do not take responsibility for the following causes for damage and repairs.

1. Any installation, movement, modification and repairs by other engineering services, which was not designated
by Dongmun.

2. Troubles and causes to our machine incurred by other equipment.

3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun.

4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual.

5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the
powers, installation and working places.

6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders.

7. Accident and damage happened by the use beyond the original purpose of the machine.

8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the
conditions and results of the medical prescriptions.

47/146
Operation Manual: DM-5125

SCOPE AND PERIOD OF WARRENTY

1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from
normal use and operation of product.

2. Service fee is charged for the following cases.


- Malfunctioning after the period of warranty
- Disorder caused by fire, earthquake, or lightening
- Inappropriate movement or negligence of product
- Disorder from repair or reconstruction by other than Dongmun service agent or other designated ones.

3. Note that defects not related to the main functions of the product are not eligible for this warranty.

4. Requirements for Service


- Stop using the product when a problem is found, and check the manual for a solution.
- Turn off the product before service request, and check the model number, serial number, the date of purchase
before calling the service center.

5. Defects or depreciation in appearance is not eligible for refund or change.

6. Dongmun is not responsible for indirect damage related to the use of products.

7. Dongmun is not responsible for injury or damage after the end of warranty period.

8. All detailed adaptability warranties and other warranties are subjected to this statement.

48/146
Operation Manual: DM-5125

SECTION 1. COMPOSITION / SPCECIFICATION

1-1. COMPOSITION

The electric composition block diagram of DIG-Series is shown in Fig 1-1.

EM swich HF
Power AD/DC HV TANK
Supply Delay Convertor DC/AC Inverter
EM swich X-ray
TUBE

IGBT
Driver

Stater

OP
PANEL
HT CONT
Collimator
PWR

FILAMENT

IF

AEC ION
Chamber

Fig 1-1 COMPOSITION OF THE HFG

AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is
inverted in high-frequency voltage of about 20kHz by IGBT inverter.
The high-frequency voltage is converted into the 40 (+20kV/-20kV) to 125kV (+62.5kV/-62.5kV) DC high
voltage by a high-voltage transformer and the multi-voltage rectifier circuit. This DC high voltage is supplied to
an X-ray tube through a high-voltage cable, and PWM INV controls this X-ray tube voltage. Modulating the
Filament voltage controls the tube current.
.
The operation panel can operate selection in each mode of operation and control, selection and setting of
conditions. These operation conditions are transmitted by the communication system from OP-CONT to HT-
CONT.
HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.

49/146
Operation Manual: DM-5125

The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.

Operation Panel

High Voltage Generator

Main CPU

A/D CONV D/A CONV

PWM kV FB PWM

kV, mA FB Amp

Filament
mA FB
Drive
circuit rms/DC CONV

IGBT AEC
FIL. INV.

HV Trans kV
FIL. Trans. mA

A C
FIL.
X-ray TUBE

ION Chamber

Fig 1-2 CONTROL SIGNAL OF X-RAY EXPOSURE

50/146
Operation Manual: DM-5125

1-2. SPECIFICATION

DIG-325 20kW
DIG-325 32kW
1 Rating Output DIG-525 40KW
DIG-650 50kW
DIG-1150 80KW
Radiography 40-150kV (1kV STEP)
2 Tube Voltage kV
Fluoroscopy 40-125kV (1kV STEP)
10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160mA
Radiography (Small Focus)
3 Tube Current mA
200, 250, 320, 400, 500, 630, 800mA (Large Focus)
Fluoroscopy 0.1~4.0mA (0.1mA Step)
4 kV Accuracy ±10%
5 mA Accuracy ±15%
6 Timer Range 0.001 sec – 10 sec (38 step), 0.003 sec ~ 6 sec (PXR-161 Only)
DIG-325 0.1mAs - 630mAs
DIG-525 0.1mAs - 630mAs
7 mAs Range
DIG-650 0.1mAs - 800mAs
DIG-1150 0.1mAs - 999mAs
20mAs or more: Less than 15%
8 mAs Accuracy
Less than 20mAs: Less than 20%
9 Display Method LED digital indication
10 Setting Method Sheet Push Switch
11 Self-diagnostic Function Error Code Indication
12 Technique Selection Manual, APR, AEC
13 APR 216 exam techniques
14 AEC (Option) Ionization Chamber 5 (Option)
Display (OP Panel)
kV 3 Digit LED
mA 3 Digit LED
15 mAs 4 Digit LED (change into sec)
Time 4 Digit LED (change into mAs)
Selection/setting LED (36)
Alarm LED (2)
Weight 62 kg
16
Operation Panel 2 kg

51/146
Operation Manual: DM-5125

1-3. ELECTRICAL REQUIREMENT

Model DIG SERIES


Line Nominal Voltage Single Phase 200V / 220V / 240V
Automatic Line Compensation Range ±10 %
Frequency 50 Hz / 60 Hz
Line Impedance Less Than 0.12 Ω
Grounding conditions Based on the local regulation or 3 independent ground circuit

1-4. PROPER ENVIRONMENT

1. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.
g. Where the equipment is exposed to chemicals or gas.

2. Keep away from the place with strong vibration and maintain proper environment and conditions.

OPERATION ENVIRONMENT
Temperature range + 10°C ~ + 40°C,
Relative Humidity Range 30% ~ 75% RH (No dews)
Others No Direct Sunlight
No combustible and corrosive gases

3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere.

STORAGE AND TRANSPORTATION


Temperature range -25°C ~ +60°C
Relative Humidity Range 10% ~ 95% RH(No dews)
Air Pressure 700hPa ~ 1060hPa
Others No corrosive gases

52/146
Operation Manual: DM-5125

SECTION 2. DISPLAY / FUNCTIONS

2-1. STRUCTURES AND SIZE

FIG 2-1. STRUCTURE AND FIGURES (16kW ~ 50kW)

FIG 2-2. STRUCTURE AND FIGURES (50kW ~ 80kW 3Phase)

53/146
Operation Manual: DM-5125

400. 0 m
m

200 m
.0m
36. 0 m
m

42. 0 m
m

FIG 2-3. CONFIGURATION OF OPERATION PANEL

54/146
Operation Manual: DM-5125

2-2. OPERATION PANEL

22

10 12
16 5
9 3 4 11 1 2 6

23 24

13 14 15 18 19 20 17 21 25 7 8

FIG. 2-4 OPERATION PANEL

1 Power on switch 14 APR exposure region selection/display


2 Power on switch 15 APR exposure direction selection
3 kV Display Segment 16 Patient size selection and memory
4 kV Setting Switch 17 AEC Reset/ mode display
5 mAs/sec Setting Switch 18 Backup release switch
6 mAs/sec Setting Switch 19 Density setting/display switch
7 X-ray Exposure Ready Switch 20 Bucky selection/display
8 X-ray Exposure Switch 21 AEC mode and field selection switch
9 “OVER” Display (Red) 22 X-ray exposure hand switch
10 “ERROR” Display (Red) 23 Small focus display
11 X-ray Exposure ready sign 24 Large focus display
12 X-ray Exposure Display (Yellow) 25 AEC film/screen combination selection
13 APR Menu Selection Switch

55/146
Operation Manual: DM-5125

1) Power ON Switch
Turn on the power of generator.

2) Power OFF Switch


Turn off the power of generator.

3) kV Setting/Display Segment
Up-Button increases the kV value of tube voltage whereas Down-button decreases kV. If you keep buttons
pressed, the value of kV will be increased or decreased continuously.

4) mA Setting/Display Segment
Up-Button increases the mA value of tube voltage whereas Down-button decreases mA. If you keep buttons
pressed, the value of mA will be increased or decreased continuously.

5) mAs/Sec Setting/Display Segment


Up-Button increases the mAs/Sec value of tube voltage whereas Down-button decreases mAs/Sec. If you keep
buttons pressed, the value of mAs/Sec will be increased or decreased continuously.

6) mAs/Sec Selection Switch


The mAs or sec radiography mode is selected.
The selected radiography mode is displayed with LED.

7) X-ray Exposure Ready Switch


It circulates the rotor and controls the heat of the filament that makes ready actions for the x-ray exposure.

8) X-ray Exposure switch


Press the exposure switch after the ready sign is lighted on.

9) “Over” LED (Red)


The over LED is lighted on when the exposure conditions exceeds the normal standards. You cannot make
exposure while the over LED is lighted on.

10) “Error” Indication (Red)


The LED is lighted in case of errors.

11) X-ray Exposure Ready LED (Green)


The exposure ready sign is lighted on when it is at the ready condition.

12) X-ray Exposure LED (Yellow)


The X-ray exposure sign is lighted on during the X-ray exposure.

56/146
Operation Manual: DM-5125

13) APR Mode Setting Switch


It operates the on/off of the APR mode.

14) APR Exposure Region Selection/Display switch


It operates the selection of exposure regions and the selected location will be displayed on the LED.

15) APR Exposure Direction Setting Switch


It operates the selection of exposure directions the selected direction will be displayed on the LED.

16) Patient Size Selection and Memory Switch


It operates the selections of patient size. The selected size will be displayed on the segment. Save the changed
APR conditions by pressing the size switches.

17) AEC Reset/Mode Display Switch


If the x-ray exposure is closed early by the AEC backup timer, the AEC mode LED will blink off and the
buzzer sounds. The next exposure is forbidden until the reset switch is pushed.

18) Backup Release Switch


Release the back-up conditions.

19) Density Setting/Display Switch (AEC/APR mode)


It selects the density level of the exposure and increases or decreases the mAs/sec under the APR mode.

20) Bucky Selection/Display


Select and display BUCKY1 (table) and BUCKY 2(stand).

21) AEC Mode and Field Selection Switch


It selects the field on the AEC mode. Each LED represents the physical location of irradiation detector and the
combination of designated field is available.

22) X-ray Exposure Hand Switch


It circulates the rotor and controls the heat of the filament that makes ready actions for the x-ray exposure.
Press the exposure switch after the ready sign is lighted on. Please touch off from the button after the LED is
lighted on.

23) Small Focus Display (Green LED)


It is displayed when the small focus is selected.

24) Large Focus Display (Green LED)


It is displayed when the large focus is selected.

25) Film/Screen Combination Selection


It operates the adjustment of kVp and divided by low speed, medium speed and high speed (200, 400, 800)

57/146
Operation Manual: DM-5125

SECTION 3. OPERATION / EXPOSURE

3-1. POWER ON & OFF

1. Power-on Switch
a. Press the Power on switch.
b. If the power is turned on, initial condition of equipment will be checked.
c. During the checking actions, "-" mark will be displayed on kV, mA, and mAs/sec display segment.
d. After the check, it displays radiographic conditions.
e. Radiographic conditions display the final status of the previous operation.

2. Power-off Switch
a. Press the Power off switch.
b. Please turn off the power supply except the time of use.

Do not turn on and off the power of the equipment continuously except
emergent cases.
CAUTION Have the interval of 30 seconds or more between ON-OFF operations of a
power supply.

3-2. SELECTION OF THE RADIOGRAPHY METHOD

There are radiography modes shown below in this equipment.

1. Manual radiography mode


Radiography conditions are performed by manual setup.

2. APR radiography mode


You can select the radiography region; direction and patient size on the control panel and make x-ray exposure
with automatically set radiography conditions.
Each radiography conditions can be changed and saved.

3. AEC radiography mode


It helps to produce the optimal images by AEC (Automatic Exposure Control Option) Menu selection.

58/146
Operation Manual: DM-5125

3-3. SETTING MANUAL RADIOGRAPHY

1. Select the manual radiography mode


Check if the APR mode or AEC mode are chosen. If they are selected in APR or AEC mode, please press APR
or AEC setting switch, and release the setup conditions.

2. Setting tube voltage "kV"


Set up the tube voltage with “kV” setting switch.
Setting range: 40kV – 125 or 150kV (1kv step)

3. Setting tube current “mA”


Set up the tube current with “mA” setting switch.
Setting range:
10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160 : Small Focus
200, 250, 320, 400, 500, 630, 800, 999 m: Large Focus

Model mA Range Remark


DIG-325 10~400mA
DIG-525 10~500mA
DIG-650 10~630(800)mA
DIG-1150 10~999mA

4. Setting the radiography mode (mAs/sec radiography)

Select the radiography mode with mAs/sec selection switch (display).

"sec" radiography makes x-ray exposure by set-up time


"mAs" radiography makes x-ray exposure by the setup time of current.

- A setup of mAs value in “mAs" mode


- mAs value is chosen with the mAs/sec setting switch.
Setting range: 0.1mAs ~ 800mAs
Step :
0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 12.5, 16, 20, 25, 32, 40, 50, 63,
80, 100, 125, 160, 200, 250, 320, 400, 500, 630, 800, 999mAs

Model mAs Range Remark


DIG-325 0.1~630mAs
DIG-525 0.1~630mAs
DIG-650 0.1~800mAs
DIG-1150 0.1~999mAs

59/146
Operation Manual: DM-5125

- A setup of sec value in “sec” mode


- Sec Value is chosen with the mAs/sec [5] setting switch.
Setting range: 0.001sec ~ 10sec
Step :
0.001, 0.002, 0.003, 0.004, 0.005, 0.006, 0.008, 0.010, 0.013, 0.016, 0.020, 0.025, 0.032, 0.040, 0.050, 0.063,
0.080, 0.100, 0.125, 0.160, 0.200, 0.250, 0.320, 0.400, 0.500, 0.630, 0.800, 1.00, 1.25, 1.60, 2.00, 2.50, 3.20,
4 .00, 5.00, 6.30, 8.00, 10.00sec
About "OVER" display of radiography conditions
When set-up radiography condition "kV", "mA", and "mAs or sec" exceeds
NOTICE the permissible value of equipment, OVER sign lights up.
X-ray exposure is prohibited when the over sign is turned on. Please reset
the appropriate radiography conditions.

Model sec Range Remark


DIG-325 0.001~10 sec
DIG-525 0.001~10 sec
DIG-650 0.001~10 sec
DIG-1150 0.001~10 sec

3-4. SETTING APR

1. Selection of APR mode


Choose APR mode with the (13) APR selection switch and the LED displaying APR mode will be lighted on
the upper left of the OP panel.

2. Selection of Radiography Regions


Choose the radiography regions with (14) radiography region selection switch and you can move the LED
display with up/down button. The available option is 9.

3. Selection of Radiography Directions


Choose the radiography directions with (15) radiography direction switch out of 4 options.

4. Selection of the Radiograph Size


Choose the radiography size with (16) patient size selection switch out of 6 options, [LL] [L] [M] [S] [SS]
[BABY].

The radiography condition will be selected after the selection of the radiography symbol, radiography region,
and radiography method.
As well the order for the selection will not be changed.

60/146
Operation Manual: DM-5125

5. Selection of the Density


Adjust the mAs/sec settings with the density switch.
Increase the conditions of mAs/sec if the Density is +1 ~ +3.
Increase the conditions of mAs/sec if the Density is -1 ~ -3.

6. Change of the Radiography Conditions


You can change the selected radiography conditions with the same method of the 3-3 manual radiography
condition setting. The over LED can be lighted from time to time when the radiography conditions are changed.
Please be careful about the setting of the radiography conditions since the x-ray exposure is prohibited under
the case.

7. Memory of the Changed Radiography Conditions


You can save the conditions changed above. The method for store is pressing the selected switch for 2 seconds
until the buzzer sounds.

Memory of the changed Radiography condition


Be careful that other radiography conditions can be selected if you choose
CAUTION the selected radiography size switch. You cannot save the memory while the
over LED is lighted and when the density setting is not 0.

3-5. AEC RADIOGRAPHY

1. Selection of AEC radiography mode


In order to use AEC mode, the DIP S/W 5 of operation panel should be turned on.
When DIP S/W 5 is turned off: AEC mode is not available.

2. Switch [17] is the AEC mode selection switch.


Ion-chamber is chosen with switch [18] after you select the AEC mode with switch [17].

3. The selection of density in the AEC mode can adjust the level of x-ray exposure levels.

4. “Ready” action is not operated if the bucky and field are not chosen after the AEC mode selection.

3-6. RADIOGRAPHY

1. Photography Preparation
After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table.
If the X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the
lamp will be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography
preparation is ended.

61/146
Operation Manual: DM-5125

2. X-ray Radiography Using Operation Panel


a. Press the X-ray exposure preparation switch. Ready LED will be lighted on when all preparation work is
completed.
b. If you press the x-ray preparation switch and exposure switch at the same time, the x-ray shooting will be
started.
c. X-rays exposure LED will be lighted on and the buzzer sounds during the exposure.

3. X-ray Radiography Using Hand-Switch


a. Press the hand switch one time and make it at ready condition. The x-ray ready LED will be lighted on when
it is prepared to make exposure.
b. When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started.
c. During exposure, the exposure LED will be lighted on and the buzzer sounds.
d. If the X-ray exposure LED is turned off or buzzer does not sound, it means the radiography work is
completed. Then, you can stop pressing the hand switch.
e. If you touched out and stopped x-ray shooting during exposure, the check code will be displayed.

If you touch out from the hand switch during exposure, the x-ray shooting is
stopped and you cannot get clear pictures. Please touch off your hand after
CAUTION
the x-ray exposure signs and the buzzer sound are turned off except the case
you want to stop the x-ray exposure urgently.

62/146
Operation Manual: DM-5125

SECTION 4. TROUBLES

4-1. ERROR CODE AND CHECK CODE

In case any problems happen during the operation of the device, they will be indicated on the kV/mA segment.

1. Error Code
The error code is displayed when problems, such as abnormalities, occur to the equipment. While the error code
is displayed, radiography operation is forbidden for safety. A power supply should be switched off and on again
and an error code will be cleared. If the problem of equipment is not solved at the time of the power-supply re-
switched on, the error code will be displayed again.
Since it is the failure of equipment when an unusual state is not recovered, please
inform the service section of our company or our agent.

2. Check Code
Unlike an error code, a check code is the warning sign when wrong operations are applied by the user. X-ray
radiography operation is forbidden for the safety during the check code. The way to release a check code is
pressing the WR switch for 2 seconds. CH01 will be released if thermal switch is off during the re-supply of the
power.

63/146
Operation Manual: DM-5125

ERROR CODE LIST

Error Code DESCRIPTION ACTION


Err 01 CHARGE MONITOR ERROR (HT-CONT)
Err 02 OVER TIME ERROR (OP PANEL) AT STAND BY RESUPPLY OF
Err 03 READY-OUT SIGNAL ERROR AT SAND BY THE POWER
Err 04 X-RAY OUT SIGNAL ERROR AT STAND BY
Err 05 READY SWITCH INPUT ERROR AT STAND BY
Err 06 X-RAY SWITCH INPUT ERROR AT STAND BY
Err 07 HAND SWITCH INPUT ERROR AT STAND BY
Err 08 HAND SWITCH X-RAY INPUT ERROR AT STAND BY
Err 09 ROTOR DETECTER SIGNAL AT STANDBY
Err 10 FILLAMENT CURRENT DETECTER ERROR AT STANDBY
Err 11 ‘kV FEEDBACK ERROR AT STANDBY
Err 12 ‘mA FEEDBACK ERROR AT STANDBY
Err 13 ROTOR DETECTER ERROR AT EXPOSURE
Err 14 FILLAMENT CUREENT DETECTER ERROR AT EXPOSURE
Err 15 ‘kV FEEDBACK ERROR DETECTER ERROR AT EXPOSURE
Err 16 ‘mA FEEDBACK DETECTIVE ERROR EXPOSURE
Err 17 NOT DEFINED
Err 18 X-RAY TIME OVER IN SEC MODE
Err 19 X-RAY TIME OVER IN mAs MODE
Err 20 X-RAY TIME OVER IN AEC MODE
Err 21 HV INTERLOCK ERROR
Err 22 IGBT FEEDBACK ERROR
Err 23 ‘kV FEEDBACK OVER DETECTER ERROR AT EXPOSURE
Err 24 ‘mA FEEDBACK OVER DETECTER ERROR AT EXPOSURE
Err 25 NO ZERO-CROSS OR NO INPUT
Err 26 ZERO-CROSS OVER OR WRONG INPUT FREQUENCY
Err 27 BUCKY 1 OK SIGNAL ERROR (HT-CONT)
Err 28 BUCKY 2 OK SIGNAL ERROR (HT-CONT)
Err 29 AEC A/D DATA-IN ERROR (HT-CONT)
Err 30 AEC A/D DATA-IN ERROR (HT-CONT) AT NON-EXPOSURE
Err 31 I/F BOARD CONDITION ERROR
Err 32 AEC BOARD CONDITION ERROR
Err 33 SMALL-FOCUS SELECTION ERROR (HT-CONT)
Err 34 LARGE-FOCUS SELECTION ERROR (HT-CONT)

64/146
Operation Manual: DM-5125

CHECK CODE LIST

Check Code DESCRIPTION Measure


Ch 01 TUBE THERMOSTAT OPERATION Reset
Ch 02 When another switch is pushed during exposure Release by “WR” switch
Ch 03 When ready switch on OP PANEL is detached during exposure
Ch 04 When X-RAY switch on OP PANEL is detached during exposure
Ch 05 When a hand switch (ready) is detached during exposure
Ch 06 When a hand switch (X-RAY) is detached during exposure
Ch 07 Shortage of power-supply capacity for exposure Power-supply re-switch
Ch 08 Simultaneous exposure prohibition Reset
Ch 09 Warning to X-rays room door opening
Ch 10 AEC Level Error

# CH01 is will be released if the TUBE terminal switch is off, in case resupply of the power.

65/146
Operation Manual: DM-5125

SECTION 5. MAINTANENCE

5-1. MAINTANANCE AND PERIODIC OPERATOR

Maintenance has check menus carried out daily by the user and periodic inspection by technical service engineers
who completed special training programs on the repair of the generator.

User checks items based on " Daily Checking Items by the operator " and should keep data records.

The following checks and maintenance procedures, together with the suggested intervals, are the manufacturer’s
recommendation for most effective Periodical Maintenance schedule for this generator.

Exclusively service personnel specifically trained on medical X-ray generator must perform service tasks here
described.

The first periodic maintenance service should be performed six (6) months after installation, and the subsequent
service every twelve (12) months.

When a trouble and a poor part are discovered, please stop use of equipments, turn off a power supply, and contact a
near service station.

5-2. DAILY CHECKING ITEMS BY THE OPERATOR

Check following items according to the regular inspection plans.


Please refer to the daily inspection method regarding the way of checking.

In order to avoid the danger of an electroshock, please do not remove the


cover of equipment by any means.
CAUTION
Only the expertise person who received training performs the check and
repair accompanied by removal of a cover.

INTERVAL
Check item
Daily Weekly Monthly
Air Conditioner ○
Temperature and Humidity ○
Operation Panel ○
Radiographic Condition ○
Radiation field lamp ○
Warning light in use ○

66/146
Operation Manual: DM-5125

5-3. WAYS OF DAILY CHECKING (CHEKING MENUS)

Please check clear environment inside the operation room and x-ray room.
The system is designed to operate with normal conditions under following environment. Please keep following
temperature and humidity.

The recommended operating environment is follow.


Temperature : +10 - +35 ℃
Relative humidity : 45 – 85 %, 45~65 % (X-ray room)

In the environment with a rapid change of temperature and humidity, it is easy


to generate dew condensation.
CAUTION
Installation of an air conditioner is recommended

The ingredient that makes silver etc. sulfurate may be contained in the exhaust

CAUTION gas of an automatic development machine. Please install the independent air
conditioner.

Operation Panel
Please check whether each switch of an operation panel lights up.
When you clean the surface of an operation panel, it should be wiped with the soft and fine cloth. When using a
cleaner, it is made to contain in cloth and is used.

Don't blow a liquid upon the operation panel directly.


CAUTION It goes into the inside of equipment from a crevice, and becomes the cause of
failure.

(The check of radiographic condition


Before the commencement of work, trial radiography is performed and film density, unevenness, and etc. are
checked.

Radiation field lamp


It confirms whether a radiation field lamp lights up before the commencement of work.

Warning light in use


It confirms whether warning light in use switch on before the commencement of work.

67/146
Operation Manual: DM-5125

PART 1. Generator

4. DIG-650RF
[50kW 2-TUBE Generator]

68/146
Operation Manual: DM-5125

NOTICE 1

This manual is designed to ensure correct use and operation of DIG-Series HFG (2-TUBE).
Please read all the lines thoroughly before your using the generator.

The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No
responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which
many results from the use of this manual.

Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and
shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein.

Always keep the manual at hand for your reference.

NOTICE 2

1. This equipment should be used only by the legally qualified persons and practitioners.

2. For the safe installation


a. Please avoid wet and humid locations.
b. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust,
sulfur and other air pollutants.
c. Protect the product from slope, vibration, and shock.
d. Do not install the product where chemicals are stored or gases are generated.
e. Check the frequency, voltage, and current of power supply.
f. Check the earth ground power about the electric discharge and polarity.

3. Before the operation of the machine


a. Examine contact of switch, polarity, dial setting, and display, and check if the product properly operates.
b. Check the safety of the ground power connections.
c. Check the correctness and safety of the cable and plug connections.
d. Be careful that mixed use of the machine may give wrong diagnostic results and damages.
e. Check the exterior circuits, which occur direct physical contact to the patient.
f. Check the earth ground power about the electric discharge and polarity.

4. During the operation of the machine


a. Do not exceed the designated time for diagnosis and cures.
b. Monitor the conditions of device and influence on the patient.
c. When you find any troubles about generators or patients, stop the operation of the machine and take
appropriate measures for the safety of the patient.
d. Be careful that the patient should not touch the device.

69/146
Operation Manual: DM-5125

5. After the operation of the machine


a. Turn off the power of the machine after returning the operation switches and scales to the original position in
accordance with the prescribed regulation.
b. Do not pull out the plugs and lines with compelling forces and it can hurt the connections.
c. For the safe storage and installation
d. Please avoid wet and humid locations.
e. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust,
sulfur and other air pollutants.
f. Protect the product from slope, vibration, and shock.
g. Do not install the product where chemicals are stored or gases are generated.

6. If the device is broken, mark proper description and ask repairs to the engineer.

7. Do not modify the machine.

8. Inspections and maintenance


a. All device and parts should be checked on the regular basis.
b. If you operate the device after long period of non-use, check the safety and conditions of the machine.

9. Operate the machine in correct ways according to the manuals.

10. Be careful about following cautions


a. Wear the lead apron in order to prevent the x-ray irradiation.
b. Regular inspection should be executed about leakage dose and irradiation meter.
c. Be careful about the x-ray exposures.
d. Do not exceed the maximum weight for the x-ray tables.

NOTICE FOR THE OPERATION 1

1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate
measures.

2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off
the power during operation.

3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.

70/146
Operation Manual: DM-5125

NOTICE FOR THE OPERATION 2

1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic
drugs, oxygen and hydrogen near the working locations.

2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must
be checked after the occurrence of the earthquake.

3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use
this machine” and ask repairs to the engineers.

4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.

OTHER NOTICE AND ADVICE

1. Power on and off


Do not turn off the machine for 30 seconds after turning on the power. Please wait at least 30 seconds for the
resupply of the power.

2. Regarding the operation environment


a. X-ray room
Practitioners should take enough cares for the release of the x-rays out of the x-ray room. This device is
designed to be protective against humidity and dust. However, if you use the machine at the place with high
humidity and dust, it can shorten the life of electronic insult and grow rust on the metal parts.
b. Do not leave the machine exposed to the strong direct sunlight.

3. Exposure of the x-rays


a. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.
b. Take enough cares in order not to make excessive radiation on the same body locations.

4. Each cable has enough length and flexibility but do not bend the cable forcibly.

DANGEROUS

Do not use the machine with following equipment.


*The equipment, which release electronic waves such as mobile phone, radios, wireless toys, and etc. The
officials in hospital ought to give enough advises and education for the operator of the machine, doctors,
patient and other the third parties.

71/146
Operation Manual: DM-5125

SYMBOLS AND TERMS

The following symbols will be used in this manual: their meanings and applications are described further below.

It warns of the conditions and situations, if not taken over or avoided, that
DANGER cause heavy damages to people and even the death.

It warns of the conditions and situations, if not taken over or avoided, that
WARNING could cause heavy damages to people or cause unrepeatable damages to the
equipment.

It warns of the conditions and situations, if not taken over or avoided, that
CAUTION could cause damages to people, or to the equipment.

WAIVER CLAUSE

Dongmun do not take responsibility for the following causes for damage and repairs.

1. Any installation, movement, modification and repairs by other engineering services, which was not designated
by Dongmun.

2. Troubles and causes to our machine incurred by other equipment.

3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun.

4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual.

5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the
powers, installation and working places.

6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders.

7. Accident and damage happened by the use beyond the original purpose of the machine.

8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the
conditions and results of the medical prescriptions.

72/146
Operation Manual: DM-5125

SCOPE AND PERIOD OF WARRENTY

1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from
normal use and operation of product.

2. Service fee is charged for the following cases.


- Malfunctioning after the period of warranty
- Disorder caused by fire, earthquake, or lightening
- Inappropriate movement or negligence of product
- Disorder from repair or reconstruction by other than Dongmun service agent or other designated ones.

3. Note that defects not related to the main functions of the product are not eligible for this warranty.

4. Requirements for Service


- Stop using the product when a problem is found, and check the manual for a solution.
- Turn off the product before service request, and check the model number, serial number, the date of purchase
before calling the service center.

5. Defects or depreciation in appearance is not eligible for refund or change.

6. Dongmun is not responsible for indirect damage related to the use of products.

7. Dongmun is not responsible for injury or damage after the end of warranty period.

8. All detailed adaptability warranties and other warranties are subjected to this statement.

73/146
Operation Manual: DM-5125

SECTION 1. COMPOSITION / SPCECIFICATION

1-1. COMPOSITION

The electric composition block diagram of DIG-Series is shown in Fig 1-1.

EM swich HF
Power AD/DC HV TANK
Supply Delay Convertor DC/AC Inverter
EM swich X-ray
TUBE

IGBT
Driver

Stater

OP
PANEL
HT CONT
Collimator
PWR

FILAMENT

IF

AEC ION
Chamber

Fig 1-1 COMPOSITION OF THE HFG

AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is
inverted in high-frequency voltage of about 20kHz by IGBT inverter.
The high-frequency voltage is converted into the 40 (+20kV/-20kV) to 125kV (+62.5kV/-62.5kV) DC high
voltage by a high-voltage transformer and the multi-voltage rectifier circuit. This DC high voltage is supplied to
an X-ray tube through a high-voltage cable, and PWM INV controls this X-ray tube voltage. Modulating the
Filament voltage controls the tube current.
.
The operation panel can operate selection in each mode of operation and control, selection and setting of
conditions. These operation conditions are transmitted by the communication system from OP-CONT to HT-
CONT.
HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CPU.

74/146
Operation Manual: DM-5125

The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.

Operation Panel

High Voltage Generator

Main CPU

A/D CONV D/A CONV

PWM kV FB PWM

kV, mA FB Amp

Filament
mA FB
Drive
circuit rms/DC CONV

IGBT AEC
FIL. INV.

HV Trans kV
FIL. Trans. mA

A C
FIL.
X-ray TUBE

ION Chamber

Fig 1-2 CONTROL SIGNAL OF X-RAY EXPOSURE

75/146
Operation Manual: DM-5125

1-2. SPECIFICATION

1 Rating Output DIG-650RF 50kW


2 Tube Voltage kV DIG-650RF 40-150kV (1kV STEP)
10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160mA
3 Tube Current mA Radiography (Small Focus)
200, 250, 320, 400, 500, 630, 800mA (Large Focus)
4 kV Accuracy ±10%
5 mA Accuracy ±15%
6 Timer Range 0.001 sec – 10 sec (38 Step)
7 mAs Range DIG-650 0.1mAs - 800mAs (37 Step)
20mAs or more: Less than 15%
8 mAs Accuracy
Less than 20mAs: Less than 20%
9 Display Method LED digital indication
10 Setting Method Sheet Push Switch
11 Self-diagnostic Function Error Code Indication
12 Technique Selection Manual, APR, AEC, ABC
13 APR 240 Exam Techniques
14 AEC (Option) Ionization Chamber 5 (Option)
Display (OP Panel) Radiography OP Fluoroscopy OP
kV 3 Digit LED 3 Digit LED
mA 3 Digit LED 2 Digit LED
15 mAs 4 Digit LED (change into sec) -
Time 4 Digit LED (change into mAs) 2 Digit LED
Selection/setting LED (36) LED (6)
Alarm LED (2) LED (1)
Weight 62 kg
16
Operation Panel 2 kg

1-3. ELECTRICAL REQUIREMENT

Model DIG SERIES


Line Nominal Voltage Single Phase 200V / 220V / 240V
Automatic Line Compensation Range ±10 %
Frequency 50 Hz / 60 Hz
Line Impedance Less Than 0.12 Ω
Grounding conditions Based on the local regulation or 3 independent ground circuit

76/146
Operation Manual: DM-5125

1-4. PROPER ENVIRONMENT

1. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.
g. Where the equipment is exposed to chemicals or gas.

2. Keep away from the place with strong vibration and maintain proper environment and conditions.

OPERATION ENVIRONMENT
Temperature range + 10°C ~ + 40°C,
Relative Humidity Range 30% ~ 75% RH (No dews)
Others No Direct Sunlight
No combustible and corrosive gases

3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere.

STORAGE AND TRANSPORTATION


Temperature range -25°C ~ +60°C
Relative Humidity Range 10% ~ 95% RH(No dews)
Air Pressure 700hPa ~ 1060hPa
Others No corrosive gases

77/146
Operation Manual: DM-5125

SECTION 2. DISPLAY / FUNCTIONS

2-1. STRUCTURES AND SIZE

FIG 2-1. STRUCTURE AND FIGURES

ERROR

Sec Sec
mAs mAs

kV Size WR

mA Non 1 2 3 Bucky

Field Film/
Density AEC Select Screen

FIG 2-2. CONFIGURATION OF OPERATION PANEL

78/146
Operation Manual: DM-5125

ERROR

kV

mA

min Reset

Density

R- R/F
ABC
Auto Table

FIG 2-3. CONFIGURATION OF FLUOROSCOPIC OPERATION PANEL

79/146
Operation Manual: DM-5125

2-2. RADIOGRAPHIC OPERATION PANEL DESCRIPTION

18 19 20 21 22 23 24 25 17

ERROR

Sec Sec
mAs mAs

kV Size WR

mA Non 1 2 3 Bucky

Field Film/
Density AEC Select Screen

1 6 2 3 4 5 7 9 8 10 11 12 13 14 15 16

FIG. 2-4 RADIOGRAPHIC OPERATION PANEL

1 Power on switch 14 X-ray exposure ready switch


2 Power off switch 15 X-ray exposure switch
3 mAs/sec display/setting switch 16 APR mode store switch
4 kV display/setting switch 17 APR menu selection switch
5 mA display/setting switch 18 “Over” LED (Red)
6 mAs/sec selection switch 19 Small focus LED (Green)
7 Density display/setting switch 20 Large focus LED (Green)
8 APR patient size setting switch 21 High speed rotor LED
9 Bucky selection switch 22 X-ray exposure ready LED (Green)
10 AEC display/setting 23 X-ray exposure LED (Yellow)
11 Field display/setting switch 24 LCD 1
12 Film/Screen selection switch 25 LCD 2
13 Page up/down & write switch

80/146
Operation Manual: DM-5125

1) Power ON Switch
Turn on the power of generator.

2) Power OFF Switch


Turn off the power of generator.

3) mAs/Sec Display/Setting Switch


It operates the setting of mAs or Sec. Up-Button increases and Down-button decreases the values. When you
keep pushing the buttons, the values will be changed consecutively. The selected [mAs] [Sec] will be displayed
on the [mAs/Sec] segment.

4) kV Display /Setting Switch


It operates the setting of kV. Up-Button increases and Down-button decreases the values. When you keep
pushing the buttons, the values will be changed consecutively. The selected kV will be displayed on the kV
segment.

5) kV Display /Setting Switch


It operates the setting of mA. Up-Button increases and Down-button decreases the values. When you keep
pushing the buttons, the values will be changed consecutively. The selected mA will be displayed on the mA
segment.

6) mAs/Sec Selection Switch


Select mAs/Sec exposure mode.
- mAs/Sec selection [mAs] mode.
- [Sec] mode for the exposure time.
Indicate the selected exposure mode on the menu.

7) Density Display/Setting Switch


Set the select density level of exposure.

8) APR Size Selection and Memory Switch


Select the patient size for radiography.

9) Bucky Selection Switch


Select BUCKY1, BUCKY2, BUCKY3 or Non-BUCKY.

10) AEC Display/Setting Switch


Set or release the AEC mode.

11) Field Display/Setting Switch


Select the field in AEC mode. Each LED indicate the physical location of AEC, and combination of designated
fields are possible.

81/146
Operation Manual: DM-5125

12) Film/Screen Display/Selection Switch


Adjust the kVp, which is the combination of programmed film and screen and display as low, middle, high
speed (200, 400, 800)

13) Page Up/Down Switch & Write Switch


Operate the page up/down of APR menu displayed on LCD 2(25).

14) X-ray Exposure Ready Switch


It circulates the rotor and controls the heat of the filament that makes ready actions for the x-ray exposure.

15) X-ray Exposure switch


After the ready LED is turned on, the exposure is executed by pressing X-ray exposure switch.

16) Write Switch


Store the changed conditions on the selected APR menu.

17) APR Menu Selection Switch


Select the APR Menu using the switch located below LCD 2.

18) “Over” LED (Red)


The over LED is lighted on when the exposure conditions exceeds the normal standards. You cannot make
exposure while the over LED is lighted on.

19) Small Focus Display (Green LED)


Indicate that the small focus is selected.

20) Large Focus Display (Green LED)


Indicate that the large focus is selected.

21) High-Speed Rotor LED(Green)


Indicate that high-speed rotor is selected.

22) X-ray Exposure Ready LED (Green)


Lighted when the exposure for radiography is ready.

23) X-ray Exposure LED (Yellow)


Lighted during the exposure of X-ray.

24) LCD 1
- Display the error no and description under the error conditions
- Display over when the conditions are over the standards.
- Display the selected menu in APR Mode.

25) LCD 2
- Display the APR Menu.
- Select the displayed menu by using switch 17.

82/146
Operation Manual: DM-5125

2-3. FLUOROSCOPIC OPERATION PANEL DESCRIPTION

1 3 8 2 4 9 10

ERROR

kV

mA

min Reset

Density

R- R/F
ABC
Auto Table

5 6 7

FIG. 2-5 FLUOROSCOPIC OPERATION PANEL

1 kV display/setting switch 6 ABC Mode display / selection


2 mA display/setting switch 7 R/F table mode display / selection
3 MAs/Sec display/reset switch 8 Error display (Red)
4 Density display/setting switch 9 X-ray exposure ready LED (Green)
5 R-Auto Mode display/selection 10 X-ray exposure LED (Yellow)

1) kV display/setting Switch
It operates the setting of kV. Up-button increases and Down-button decreases the values. When you keep
pushing the buttons, the values will be changed consecutively. The selected kV value will be displayed on the
segment.

83/146
Operation Manual: DM-5125

2) mA display/setting Segment
It operates the setting of mA. Up-button increases and Down-button decreases the values. When you keep
pushing the buttons, the values will be changed consecutively. The selected mA value will be displayed on the
segment.

3) mAs/Sec display/reset switch


It displays the accumulated time under fluoroscopic mode. You can reset the accumulated time with reset switch.

(4) Density display/setting Switch


It operates the selections of density under the ABC mode and controls the brightness of monitor.

5) R-auto mode display/selection switch


Selects or release R-auto mode.

(6) ABC mode display/selection switch


It selects or release ABC mode.

7) R/F table mode display/selection switch


It selects or release R/F table mode.

8) Error LED (Red)


The error LED is lighted on, when errors occur.

9) X-ray Exposure Ready LED (Green)


The exposure ready sign is lighted on when it is at the ready condition.

10) X-ray exposure LED (Yellow)


The X-ray exposure sign is lighted on during the X-ray exposure.

84/146
Operation Manual: DM-5125

SECTION 3. OPERATION / EXPOSURE

3-1. POWER ON & OFF

1. Power-on Switch
a. Press the Power on switch.
b. If the power is turned on, initial condition of equipment will be checked.
c. During the checking actions, "-" mark will be displayed on kV, mA, and mAs/sec display segment.
d. After the check, it displays radiographic conditions.
e. Radiographic conditions display the final status of the previous operation.

2. Power-off Switch
a. Press the Power off switch.
b. Please turn off the power supply except the time of use.

Do not turn on and off the power of the equipment continuously except
emergent cases.
CAUTION Have the interval of 30 seconds or more between ON-OFF operations of a
power supply.

3-2. SELECTION OF THE RADIOGRAPHY METHOD

There are radiography modes shown below in this equipment.

1. Manual radiography mode


Radiography conditions are performed by manual setup.

2. APR radiography mode


You can select the radiography region; direction and patient size on the control panel and make x-ray exposure
with automatically set radiography conditions.
Each radiography conditions can be changed and saved.

3. AEC radiography mode


It helps to produce the optimal images by AEC (Automatic Exposure Control Option) Menu selection.

4. Fluoroscopy mode

5. ABC mode
This is the control mode to gain the best image under the fluoroscopy mode.

85/146
Operation Manual: DM-5125

3-3. SETTING MANUAL RADIOGRAPHY

1. Select the manual radiography mode


Clear up the conditions by pushing bucky switch [13], and [16] switch when the APR mode and AEC mode are
selected.

2. A set up of tube voltage "kV"


Set up the tube voltage with “kV” setting switch.
Setting range: 40kV – 150kV (1kv step)

Model kV Range Remark


DIG-525 40~125kV
DIG-650 40~150kV

3. A set up of tube current “mA”


Set up the tube current with “mA” [5] setting switch.
Setting range:
10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160 mA: Small Focus
200, 250, 320, 400, 500, 630, 800, mA: Large Focus

4. A set up of mAs/Sec
mAs/sec value is chosen with the mAs/sec setting switch.
Setting range: 0.1mAs ~ 800mAs
Step :
0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 12.5, 16, 20, 25, 32, 40, 50, 63,
80, 100, 125, 160, 200, 250, 320, 400, 500, 630, 800mAs

Model mAs Range Remark


DIG-650 10~800mAs

Sec Value is chosen with the mAs/sec [5] setting switch.


Setting range: 0.001sec ~ 10sec
Step :
0.001, 0.002, 0.003, 0.004, 0.005, 0.006, 0.008, 0.010, 0.013, 0.016, 0.020, 0.025, 0.032, 0.040, 0.050, 0.063,
0.080, 0.100, 0.125, 0.160, 0.200, 0.250, 0.320, 0.400, 0.500, 0.630, 0.800, 1.00, 1.25, 1.60, 2.00, 2.50, 3.20,
4 .00, 5.00, 6.30, 8.00, 10.00sec

About "OVER" display of radiography conditions


When set-up radiography condition "kV", "mA", and "mAs or sec" exceeds
NOTICE the permissible value of equipment, OVER sign lights up.
X-ray exposure is prohibited when the over sign is turned on. Please reset
the appropriate radiography conditions.

86/146
Operation Manual: DM-5125

3-4. SETTING APR MODE

1. APR mode selection


APR mode is set up with the selection of Bucky, symbol on LCD 2 and patient size.

2. Radiography symbol selection


Select the radiography symbol by switch [19].
Options are 7 parts and selected symbol is displayed on the LCD.

3. Radiography Region Selection


Select the radiography region by switch [19] after the symbol is chosen.
Selectable region is different according to each symbol.
The selected region is displayed on the LCD 1.

4. Radiography method Selection


Select the radiography region by switch [19] after the region is chosen.
Selectable region is different according to each region.
The selected method is displayed on the LCD 1.

5. Page up/down
Selectable by page up/down key when the displayed menu on the LCD 2 is more than 7.

6. Density Selection
Increase or decrease the mAs/sec selection by density selection switch.
The selected mAs/sec will be increased when the density is +1~+5.
The selected mAs/sec will be increased when the density is -1~-5.

7. Change of radiography condition


The radiography conditions can be changed with the same method for “manual radiography”
The over LED can be lighted when the conditions are changed. The x-ray exposure is prohibited in this case and
special cautions are required.

8. Memory of the Changed Radiography Conditions


The changed radiography conditions can be saved by pushing the “WR” switch for 2 sec until the buzzer
sounds.

Memory of the changed Radiography condition


Be careful that other radiography conditions can be selected if you choose
CAUTION the selected radiography size switch. You cannot save the memory while the
over LED is lighted and when the density setting is not 0.

87/146
Operation Manual: DM-5125

9. Method to save the APR data (example)


When you save the head (symbol), skull (region), LAT (method).

a. Choose the Bucky. (Select the bucky out of four)

b. If you push SW27 (Page Up), SW28 (Page Down), SW29 (WR) at the same time
APR store menu will be displayed on the LCD2 as following.

c.

SW19 SW21 SW23 SW25


1. Modify 2. Write 3. Delete
LCD2

SW20 SW22 SW24 SW26

If you select the “Write”(SW21), the numbers from 1 to 8 will be displayed on the LCD2.

SW19 SW21 SW23 SW25


1 2 3 4
LCD2
5 6 7 8
SW20 SW22 SW24 SW26

d. If you select SW19, SW21, and SW23 in order, the numbers on the LCD1 will be displayed as following.
The saving places for SW19, SW21, SW23 will be allotted by order.

e. The following symbols will be displayed on the LCD2.

1 2 3
LCD1

SW19 SW21 SW23 SW25


HEAD CHEST ABDOMEN SPINE
LCD2
HIP UPPER EXT LOWER EXT
SW20 SW22 SW24 SW26

Select the “HEAD”(SW19) of the LCD2. The head will be displayed on the number of the LCD1 and the region
will be indicated on the LCD2.

88/146
Operation Manual: DM-5125

1 HEAD 2 3
LCD1

SW19 SW21 SW23 SW25


SKULL SINUS MANDIBLE MAXILLA
LCD2
NASAL CAV TM JOINTT DENTAL JAW /→
SW20 SW22 SW24 SW26

[ /→] mark on the right bottom of LCD2 means that it has the next page. Move the previous and next page with
page up/down switch.

f. If you select “SKULL”(SW19) on the LCD 2, skull will be indicated on the No. 2 of LCD1 and method will
be displayed on the LCD2.

1 HEAD 2 SKULL 3
LCD1

SW19 SW21 SW23 SW25


AP PA LAT AXI
LCD2
OBL TOWNE WATERS /→
SW20 SW22 SW24 SW26

g. If you select “LAT”(SW23) on the LCD2, the LAT will be displayed on the No. 3 of LCD1 and “wirte ok?”
will be screened when all the selection work of the menu is completed.

1 HEAD 2 SKULL 3 LAT


LCD1

SW19 SW21 SW23 SW25

LCD2
Write OK?
SW20 SW22 SW24 SW26

h. If you push “WR”(SW29) after you select the kV, mA, sec, Body Size, they will be stored in the locations
above.

89/146
Operation Manual: DM-5125

i. If you push the SW27 (Page Up), SW28(Page Down) at the same out of APR menus, they will be returned to
the conditions before selection.

10. The change of the selected APR data


When you modify the APR settings where the HEAD (Symbol), SKULL (Region), LAT (Method) are stored
on the location 1,2,3.

a. Push the SW27(Page Up), SW28(Page Down), SW29(WR) at the same time.
Then, the LCD 2 will display as following.

SW19 SW21 SW23 SW25


1 Modify 2 Write 3 Delete
LCD2

SW20 SW22 SW24 SW26

b. Press SW19 and select “modify”.


If you select the “modify” function, stored APR menus will be displayed according to the selected bucky.

SW19 SW21 SW23 SW25


HEAD
LCD2

SW20 SW22 SW24 SW26

If only the word “head” is displayed, it means there is only “head” on the APR menu.

c. Press SW19 and select the “HEAD”(Symbol).


If you select the “head” it will be displayed on the LCD1 and the skull (region) will be displayed on the
LCD2.

HEAD
LCD1

SW19 SW21 SW23 SW25


SKULL
LCD2

SW20 SW22 SW24 SW26

d. Push the SW21 and select the skull (region). If you selelct the skull it will be displayed on the LCD1 and
LCD2 as following.

90/146
Operation Manual: DM-5125

HEAD SKULL
LCD1

SW19 SW21 SW23 SW25


LAT
LCD2

SW20 SW22 SW24 SW26

e. Press SW23 and select “LAT”(Method).


After the selection of the LAT, change the value of kV, mA, sec and store the changed menus by pressing the
WR button.

HEAD SKULL LAT


LCD1

SW19 SW21 SW23 SW25


Modify OK?
LCD2

SW20 SW22 SW24 SW26

11. The deletion of the stored APR setting (example)


You use this key when you intend to delete the stored APR data.

a. If you want to delete the stored APR data above.

b. If you push the SW 27,28, 29 at the same time, they will be displayed as following and select the “delete”.

SW19 SW21 SW23 SW25


1 Modify 2 Write 3 Delete
LCD2

SW20 SW22 SW24 SW26

c. If you select the “delete”, the stored APR menus will be displayed on the LCD2 according to the selected
Bucky.

91/146
Operation Manual: DM-5125

SW19 SW21 SW23 SW25


HEAD
LCD2

SW20 SW22 SW24 SW26

d. If you push the SW19 and select the Head (symbol), the “head” will be demonstrated on the LCD1 and skull
(region) will be on the LCD2.

HEAD
LCD1

SW19 SW21 SW23 SW25


SKULL
LCD2

SW20 SW22 SW24 SW26

e. If you select the “skull” by pressing SW21, the LCD1 and LCD2 will be displayed as following.

HEAD SKULL
LCD1

SW19 SW21 SW23 SW25


LAT
LCD2

SW20 SW22 SW24 SW26

f. If you push the SW23 and choose the LAT, the question “Delete OK?” will be screened on the LCD2 to ask
weather you want to delete.

HEAD SKULL LAT


LCD1

SW19 SW21 SW23 SW25


Delete OK?
LCD2

SW20 SW22 SW24 SW26

92/146
Operation Manual: DM-5125

g. If you want to delete the selected APR menu, press the WR (SW29).
If you select the wrong menu on the APR selections, you can move to the previous steps by pressing the
SW27 and SW28 at the same time.

3-5. RADIOGRAPHY

1. PHOTOGRAPHY PREPARATION
After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table.
If the X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the
lamp will be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography
preparation is ended.

2. X-ray radiography using operation panel


a. Press the X-ray exposure preparation switch. Ready LED will be lighted on when all preparation work is
completed.
b. If you press the x-ray preparation switch and exposure switch at the same time, the x-ray shooting will be
started.
c. X-rays exposure LED will be lighted on and the buzzer sounds during the exposure.

3. X-ray radiography using hand-switch


a. Press the hand switch one time and make it at ready condition. The x-ray ready LED will be lighted on when
it is prepared to make exposure.

b. When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started.

c. During exposure, the exposure LED will be lighted on and the buzzer sounds.

d. If the X-ray exposure LED is turned off or buzzer does not sound, it means the radiography work is
completed. Then, you can stop pressing the hand switch.

e. If you touched out and stopped x-ray shooting during exposure, the check code will be displayed.

If you touch out from the hand switch during exposure, the x-ray shooting is
stopped and you cannot get clear pictures. Please touch off your hand after
CAUTION
the x-ray exposure signs and the buzzer sound are turned off except the case
you want to stop the x-ray exposure urgently.

You cannot make radiography while the over LED is lighted on. Change the
CAUTION conditions for the radiography in those cases.

93/146
Operation Manual: DM-5125

SECTION 4. TROUBLES

4-1. ERROR CODE AND CHECK CODE

In case any problems happen during the operation of the device, they will be indicated on the kV/mA segment.

1. Error Code
The error code is displayed when problems, such as abnormalities, occur to the equipment. While the error code
is displayed, radiography operation is forbidden for safety. A power supply should be switched off and on again
and an error code will be cleared. If the problem of equipment is not solved at the time of the power-supply re-
switched on, the error code will be displayed again.
Since it is the failure of equipment when an unusual state is not recovered, please
inform the service section of our company or our agent.

2. Check Code
Unlike an error code, a check code is the warning sign when wrong operations are applied by the user. X-ray
radiography operation is forbidden for the safety during the check code. The way to release a check code is
pressing the WR switch for 2 seconds. CH01 will be released if thermal switch is off during the re-supply of the
power.

94/146
Operation Manual: DM-5125

ERROR CODE LIST

Error Code DESCRIPTION ACTION


Err 01 CHARGE MONITOR ERROR (HT-CONT)
Err 02 OVER TIME ERROR (OP PANEL) AT STAND BY RESUPPLY OF
Err 03 READY-OUT SIGNAL ERROR AT SAND BY THE POWER
Err 04 X-RAY OUT SIGNAL ERROR AT STAND BY
Err 05 READY SWITCH INPUT ERROR AT STAND BY
Err 06 X-RAY SWITCH INPUT ERROR AT STAND BY
Err 07 HAND SWITCH INPUT ERROR AT STAND BY
Err 08 HAND SWITCH X-RAY INPUT ERROR AT STAND BY
Err 09 ROTOR DETECTER SIGNAL AT STANDBY
Err 10 FILLAMENT CURRENT DETECTER ERROR AT STANDBY
Err 11 ‘kV FEEDBACK ERROR AT STANDBY
Err 12 ‘mA FEEDBACK ERROR AT STANDBY
Err 13 ROTOR DETECTER ERROR AT EXPOSURE
Err 14 FILLAMENT CUREENT DETECTER ERROR AT EXPOSURE
Err 15 ‘kV FEEDBACK ERROR DETECTER ERROR AT EXPOSURE
Err 16 ‘mA FEEDBACK DETECTIVE ERROR EXPOSURE
Err 17 NOT DEFINED
Err 18 X-RAY TIME OVER IN SEC MODE
Err 19 X-RAY TIME OVER IN mAs MODE
Err 20 X-RAY TIME OVER IN AEC MODE
Err 21 HV INTERLOCK ERROR
Err 22 IGBT FEEDBACK ERROR
Err 23 ‘kV FEEDBACK OVER DETECTER ERROR AT EXPOSURE
Err 24 ‘mA FEEDBACK OVER DETECTER ERROR AT EXPOSURE
Err 25 NO ZERO-CROSS OR NO INPUT
Err 26 ZERO-CROSS OVER OR WRONG INPUT FREQUENCY
Err 27 BUCKY 1 OK SIGNAL ERROR (HT-CONT)
Err 28 BUCKY 2 OK SIGNAL ERROR (HT-CONT)
Err 29 AEC A/D DATA-IN ERROR (HT-CONT)
Err 30 AEC A/D DATA-IN ERROR (HT-CONT) AT NON-EXPOSURE
Err 31 I/F BOARD CONDITION ERROR
Err 32 AEC BOARD CONDITION ERROR
Err 33 SMALL-FOCUS SELECTION ERROR (HT-CONT)
Err 34 LARGE-FOCUS SELECTION ERROR (HT-CONT)

95/146
Operation Manual: DM-5125

CHECK CODE LIST

Check Code DESCRIPTION Measure


Ch 01 TUBE THERMOSTAT OPERATION Reset
Ch 02 When another switch is pushed during exposure Release by “WR” switch
Ch 03 When ready switch on OP PANEL is detached during exposure
Ch 04 When X-RAY switch on OP PANEL is detached during exposure
Ch 05 When a hand switch (ready) is detached during exposure
Ch 06 When a hand switch (X-RAY) is detached during exposure
Ch 07 Shortage of power-supply capacity for exposure Power-supply re-switch
Ch 08 Simultaneous exposure prohibition
Ch 09 Warning to X-rays room door opening Reset
Ch 10 AEC Level Error

# CH01 is will be released if the TUBE terminal switch is off, in case resupply of the power.

96/146
Operation Manual: DM-5125

SECTION 5. MAINTANENCE

5-1. MAINTANANCE AND PERIODIC OPERATOR

Maintenance has check menus carried out daily by the user and periodic inspection by technical service engineers
who completed special training programs on the repair of the generator.

User checks items based on " Daily Checking Items by the operator " and should keep data records.

The following checks and maintenance procedures, together with the suggested intervals, are the manufacturer’s
recommendation for most effective Periodical Maintenance schedule for this generator.

Exclusively service personnel specifically trained on medical X-ray generator must perform service tasks here
described.

The first periodic maintenance service should be performed six (6) months after installation, and the subsequent
service every twelve (12) months.

When a trouble and a poor part are discovered, please stop use of equipments, turn off a power supply, and contact a
near service station.

5-2. DAILY CHECKING ITEMS BY THE OPERATOR

Check following items according to the regular inspection plans.


Please refer to the daily inspection method regarding the way of checking.

In order to avoid the danger of an electroshock, please do not remove the


cover of equipment by any means.
CAUTION
Only the expertise person who received training performs the check and
repair accompanied by removal of a cover.

INTERVAL
Check item
Daily Weekly Monthly
Air Conditioner ○
Temperature and Humidity ○
Operation Panel ○
Radiographic Condition ○
Radiation field lamp ○
Warning light in use ○

97/146
Operation Manual: DM-5125

5-3. WAYS OF DAILY CHECKING (CHEKING MENUS)

Please check clear environment inside the operation room and x-ray room.
The system is designed to operate with normal conditions under following environment. Please keep following
temperature and humidity.

The recommended operating environment is follow.


Temperature : +10 - +35 ℃
Relative humidity : 45 – 85 %, 45~65 % (X-ray room)

In the environment with a rapid change of temperature and humidity, it is easy


to generate dew condensation.
CAUTION
Installation of an air conditioner is recommended

The ingredient that makes silver etc. sulfurate may be contained in the exhaust

CAUTION gas of an automatic development machine. Please install the independent air
conditioner.

Operation Panel
Please check whether each switch of an operation panel lights up.
When you clean the surface of an operation panel, it should be wiped with the soft and fine cloth. When using a
cleaner, it is made to contain in cloth and is used.

Don't blow a liquid upon the operation panel directly.


CAUTION It goes into the inside of equipment from a crevice, and becomes the cause of
failure.

The check of radiographic condition


Before the commencement of work, trial radiography is performed and film density, unevenness, and etc. are
checked.

Radiation field lamp


It confirms whether a radiation field lamp lights up before the commencement of work.

Warning light in use


It confirms whether warning light in use switch on before the commencement of work.

98/146
Operation Manual: DM-5125

PART 2. X-ray Tube

1. E7132X

Refer to the Attached E7132X Data Sheet.

99/146
Operation Manual: DM-5125

PART 2. X-ray Tube

2. E7239X

Refer to the Attached E7239X Data Sheet.

100/146
Operation Manual: DM-5125

PART 2. X-ray Tube

3. E7240X

Refer to the Attached E7240X Data Sheet.

101/146
Operation Manual: DM-5125

PART 2. X-ray Tube

4. E7252X

Refer to the Attached E7252X Data Sheet.

102/146
Operation Manual: DM-5125

PART 3. Bucky Table

1. DM-FT-1[Fixed Table-Top Table]

103/146
Operation Manual: DM-5125

RADIATION WARNING INFORMATION

X-ray and Gamma-ray and dangerous to both patient and operator unless Established safe exposure procedures are
strictly observed.
The useful beam can produce serious or fatal bodily injuries to any persons in the surrounding area if used by an
unskilled operator.
Adequate precautions must always taken to avoid exposure to useful beam, as well as to leakage radiation from
within the source housing or to scattered radiation resulting from the passage through matter.
The authorized to operator, participate in or supervise the operation of the equipment must be thoroughly familiar
and comply completely with the currently established safe exposure factors and procedures described in the
National Concil on Radiation Protection and Measurement ( NCRP ) “ Medical X-ray and Gamma-ray protection
for energies up to 10 Mev- Equipment design and Use “NCRP Report-33 as revised or replaced in the future.
Those responsible for the planning of X-ray and Gamma-ray equipment installation must be thoroughly familiar
and comply completely with the structural shielding requirements outlined in NCRP-34 as revised or replaced in
the further.
Failure to observe these warnings may cause serious or fatal bodily injuries to the operator, patient or those in
attendance.

OUTLINE DATA

LEGAL OBLIGATION
You have legal obligation that failure to follow the manufacturer’s instructions or manufacturer’s will affect
radiation of any component by user or installer which well affect radiation safety, causes the user or installer to
assume full responsibility for that product.
The manufacturer is required to prescribe maintenance information and a schedule of performance to ensure
equipment complies with specified para-meters After installation the installer must supply manufacturer’s data
including this manual to the equipment purchaser or user.
The equipment purchaser or user must follow maintenance instructions.

104/146
Operation Manual: DM-5125

SECTION 1. INTRODUCTION

This manual provides information relative to the fixed table-top table.


The table system, hereinafter to as the DM-FT-1.
The apparatus describe in this a heavy-duty, fixed table-top table is designed for general radiographic
examinations.

SECTION 2. DESCRIPTION

This table system is designed for general diagnostic radiography.


The table base fixed the table top.
The bucky is electromagnetically locked.

2-1. EQUIPMENT SUPPLIED


1. Table Leg
2. Table Top
3. Bucky

SECTION 3. SPECIFICATION

Table top : 2000 mm ( 78.74” ) X 740 mm ( 29.13” )


Table top height : 620 mm ( 24.4” )
Table top – to – film distance : 65 mm ( 3.34” )
Table top maximum load : 136Kg
Table top inherent filtration : 1.0 mm Aluminum equivalent at 100 KVP
Bucky travel : 230mm ( 20.66” )
Weight : 40 Kg
Power : Voltage : 24 VAC
Frequency : 50 or 60 Hz

105/146
Operation Manual: DM-5125

SECTION 4. INSTALLATION

4-1. PRELIMINARY CONSIDERATION

The following factors must be consider, in order to planning an efficient Installation site layout.
- Location of tube stand.
- Location of auxiliary equipment (bucky stand, cassette holder, etc.).
- Incoming power for X-ray generator and auxiliary equipments.

The tube stand, floor-ceiling tube stand should be installed before installation of DM-FT-1 table system, and tube
stand must be aligned to longitudinal travel.

The floor should be leveled the area where the system is being installed.

This table shoule not be used in the presence of explosive or flammable gases
WARNING as might be found in an operating room.

4-2. Electrical Wiring

Interconnecting cable is supplied with this system, it is highly recommended to use adequate size of floor trough
from control console to high tension transformer and rear of the DM-BT-1 table.

The trough should have removable cover flush with the finished surface of the floor.

If the power is being supplied from wall disconnect box, #12 AWG stranded wire should be used for table power
and #10 AWG for equipment ground.

4-3. System Calibration

1. Set the table into position under the tube stand.

2. A plum line can be used from the X-ray, align the table center line along the longitudinal travel of the
movement.
Note : To align the table bucky verses tube stand, the table-top should be removed from the DM-FT-1 table.

3. The table-top should be level (Longitudinally and transversely) at this time, if necessary the table should be
shimmed.
Note : Shims are supplied by manufacturer.

106/146
Operation Manual: DM-5125

4. Secure the table to the floor.


Note : Use adequate hardwares for that particular installation.
See location of the table anchor bracket in the major component layout.

5. Check all connections between each components before apply the power to the table.

6. Check bucky lock release function by pulling the cassette tray toward to the front direction.

7. Check the table-top center at the center of the table-top is aligned to the table bucky.

8. The table bucky function should be checked, make an exposure and verify that bucky grid starts oscillating
during the X-ray generator exposure cycle.

107/146
Operation Manual: DM-5125

SECTION 5. MAINTENANCE

At least once a month external surfaces and exposed tracks and bearing surfaces should be wiped a cloth slightly
moistened with light oil to provide sufficient lubrication for smooth operation and protect against rust.

The following checks for proper operation should be made:


- Bucky travel free and smooth, and grid operating properly.
- All locks should be operating properly.
- Spacing between locks and bearings should be checked.
- All electrical connections and coil cords in table should be proper condition.

DM-FT-1 DIAGRAM

Bucky
DC24V Solenoid
(Generator)

108/146
Operation Manual: DM-5125

PART 3. Bucky Table

2. DM-BT-1[4-way Table]

109/146
Operation Manual: DM-5125

RADIATION WARNING INFORMATION

X-ray and Gamma-ray and dangerous to both patient and operator unless Established safe exposure procedures are
strictly observed.
The useful beam can produce serious or fatal bodily injuries to any persons in the surrounding area if used by an
unskilled operator.
Adequate precautions must always taken to avoid exposure to useful beam, as well as to leakage radiation from
within the source housing or to scattered radiation resulting from the passage through matter.
The authorized to operator, participate in or supervise the operation of the equipment must be thoroughly familiar
and comply completely with the currently established safe exposure factors and procedures described in the
National Concil on Radiation Protection and Measurement ( NCRP ) “ Medical X-ray and Gamma-ray protection
for energies up to 10 Mev- Equipment design and Use “NCRP Report-33 as revised or replaced in the future.
Those responsible for the planning of X-ray and Gamma-ray equipment installation must be thoroughly familiar
and comply completely with the structural shielding requirements outlined in NCRP-34 as revised or replaced in
the further.
Failure to observe these warnings may cause serious or fatal bodily injuries to the operator, patient or those in
attendance.

OUTLINE DATA

LEGAL OBLIGATION
You have legal obligation that failure to follow the manufacturer’s instructions or manufacturer’s will affect
radiation of any component by user or installer which well affect radiation safety, causes the user or installer to
assume full responsibility for that product.
The manufacturer is required to prescribe maintenance information and a schedule of performance to ensure
equipment complies with specified para-meters After installation the installer must supply manufacturer’s data
including this manual to the equipment purchaser or user.
The equipment purchaser or user must follow maintenance instructions.

110/146
Operation Manual: DM-5125

SECTION 1. INTRODUCTION

This manual provides information relative to the fixed table-top table.


The table system, hereinafter to as the DM-BT-1.
The apparatus describe in this a heavy-duty, fixed table-top table is designed for general radiographic
examinations.

SECTION 2. DESCRIPTION

This table system is designed for general horizontal diagnostic radiograpy.


The table base support the table top.
The transverse and longitudinal table-top locks are controlled by a foot SW at the front the table base.
The table motious and bucky are electromagnetically locked.
The bucky station AEC (Automatic Exposure Control) are option.
A single station AEC is available as an potion on the bucky by using chamber.

2-1. EQUIPMENT SUPPLIED


1. Table Leg
2. Table Top
3. Bucky (Speed is adjustable to the max. 5 Step)
4. Compression device (Option)

SECTION 3. SPECIFICATION

Table base : 1262 mm ( 49.68” ) X 650 mm ( 25.59” )


Table top : 2000 mm ( 78.74” ) X 740 mm ( 29.13” )
Table top height : 700 mm ( 27.55” )
Table top longitudinal : 350 mm ( 13.77” )
Table top transverse travel : 150 mm ( 5.9” )
Table top – to – film distance : 65 mm ( 2.55” )
Table top maximum load : 136Kg
Table top inherent filtration : 1.0 mm Aluminum equivalent at 100 KVP
Bucky travel : 230 mm ( 9.05” )
Weight : 70 Kg
Power : Voltage : 220 VAC
Frequency : 50 or 60 Hz
Current : 3A

111/146
Operation Manual: DM-5125

See Outline Data

112/146
Operation Manual: DM-5125

SECTION 4. INSTALLATION

4-1. PRELIMINARY CONSIDERATION

The following factors must be consider, in order to planning an efficient Installation site layout.
- Location of tube stand.
- Location of auxiliary equipment (bucky stand, cassette holder, etc.).
- Incoming power for X-ray generator and auxiliary equipments.

The tube stand, floor-ceiling tube stand should be installed before installation of DM-FT-1 table system, and tube
stand must be aligned to longitudinal travel.

The floor should be leveled the area where the system is being installed.

This table shoule not be used in the presence of explosive or flammable gases
WARNING as might be found in an operating room.

4-2. Electrical Wiring

Interconnecting cable is supplied with this system, it is highly recommended to use adequate size of floor trough
from control console to high tension transformer and rear of the DM-BT-1 table.

The trough should have removable cover flush with the finished surface of the floor.

If the power is being supplied from wall disconnect box, #12 AWG stranded wire should be used for table power
and #10 AWG for equipment ground.

4-3. System Calibration

1. Set the table into position under the tube stand.

2. A plum line can be used from the X-ray, align the table center line along the longitudinal travel of the
movement.
Note : To align the table bucky verses tube stand, the table-top should be removed from the DM-FT-1 table.

3. The table-top should be level (Longitudinally and transversely) at this time, if necessary the table should be
shimmed.
Note : Shims are supplied by manufacturer.

113/146
Operation Manual: DM-5125

4. Secure the table to the floor.


Note : Use adequate hardwares for that particular installation.
See location of the table anchor bracket in the major component layout.

5. Check all connections between each components before apply the power to the table.

6. Turn SW 1 (ON & OFF) on.

7. All locks should functioning.

8. Verify that power on & off indication green lamp should be on.

9. Depress the foot-SW, the longitudinal and transverse locks should be released.

10. Check bucky lock release function by pulling the cassette tray toward to the front direction.

11. Check the table-top transverse center, depress the foot-SW and move the table-top is aligned to the table bucky,
the BZ1 (table center indication Buzzer) should be sounded

12. The table bucky function should be checked, make an exposure and verify that bucky grid starts oscillating
during the X-ray generator exposure cycle.

114/146
Operation Manual: DM-5125

SECTION 5. MAINTENANCE

At least once a month external surfaces and exposed tracks and bearing surfaces should be wiped a cloth slightly
moistened with light oil to provide sufficient lubrication for smooth operation and protect against rust.

The following checks for proper operation should be made:


- Bucky travel free and smooth, and grid operating properly.
- All locks should be operating properly.
- Spacing between locks and bearings should be checked.
- All electrical connections and coil cords in table should be proper condition.

115/146
Operation Manual: DM-5125

DM-BT-1 SCHEMATIC

116/146
Operation Manual: DM-5125

DM-BT-1 BUCKY SCHEMATIC

117/146
Operation Manual: DM-5125

PART 3. Bucky Table

3. DM-RFT-5[RF Tilting Table]

118/146
Operation Manual: DM-5125

SECTION 1. APPLICATIONS

Used in combination with an X-Ray high tension apparatus, an X-Ray tube unit,
An X-ray tube support unit , this table is applied for the following X-Ray examinations:

1. Fluoroscopy

2. Spot Filming

3. Bucky Radiography

119/146
Operation Manual: DM-5125

SECTION 2. CONSTITUTIONS

Dimensions : See the following drawing

1. DM RFT-1 consists of Radiography table, bucky device unit, spot film device, Image intensifier, Camera,
Monitor with trolley

2. The relative accessories : compression band, hand grips and foot rest.

3. Standard Accessories
- Built-in Bucky Device 1Set
- Foot Rest 1Set
- Hand Grips 1Set
- Foot Switch 1Set

120/146
Operation Manual: DM-5125

SECTION 3. SPECIFICAIONS

Parts DM-RFT-5
type motor driven
Table tilting range standard : +90˚ ∼ -15˚ option : +90˚ ∼ -20˚
speed 90˚/20 sec., 50/60Hz
type motor driven
horizontal : ±50cm
movement range upward(tilting position) : 50cm
Tabletop sliding
lateral : ±11cm(option)
3.6cm/sec., 60Hz
speed
5cm/sec., 50Hz
material Lminated 60mmT
focal distance 41cmH
Tabletop
filtration less than 1.0mm AL. Eq. at 100KV
tabletop to film 6.5cm
type solenoid reciprocating
grid standard : 40LPC 10:1, option : 60LPC, 70LPC
Bucky device cassette range 13x18cm ∼ 35x43cm(5x7" ∼ 14x17")
tarvel range 100cm
AEC mode Ion chamber(option)
parking slide parking from left to right
locking electromagnetic lock
longitudinal movement 60cm
Spot film device transversal movement ±12cm
vertical movement 14cm ∼ 50cm
grid standard : 40LPC 10:1, option : 60LPC, 70LPC
AEC mode Ion chamber(option)
Compression cone manual operation
Field collimation under couch motor driven
Image intensifier 12", 9", 6" selectable
TV Camera CCD
Image assembly
Monitor 23", 17", 15", 14" selectable
Trolley four wheels, two shelves, 51cmLx44cmWx138cmH
Power requirement 220VAC, 3KVA, 50/60Hz, single phase
Weight net weight 650kg except image assembly
Dimensions approx. 200cmL x 72cmW x 85cmH
Installation space minimum 310cmL x 158cmW x 260cmH

121/146
Operation Manual: DM-5125

SECTION 4. BASIC TECHNICAL CONCEPT

Model DM-RFT-5 is designed for performing Radiography & Fluoroscopy as well as Castro Intestinal
examinations. This unit not only makes your patients feel comfortable and safe during the operation, but also
makes your diagnosis extremely efficient and reliable.

4-1. Safe and High Operability of Table

1. Table-top sliding switch and table tilting switch are provided on the tower, near fluoroscopy screen or I/I unit.

2. 81cm high table-top is suitable for treating decubit us patients.

122/146
Operation Manual: DM-5125

SECTION 5. INSTALLATION

123/146
Operation Manual: DM-5125

SECTION 6. OPERATION SWITCHES

The table tilting switches are provided on the tower, near the fluoroscopy screen, the tilting angle can be read by
the indicator provided on the front cover. The table tilting switch is provided on the spot film device.

6-1. Table Switches

124/146
Operation Manual: DM-5125

6-2. Switches of Fluoroscopic Operation Panel (Spot Film Device)

SPOT Collimator long Collimator long

X-RAY : Tube open : Tube close


FLOURO. Collimator lat

ON : Tube close

TILT

DOWN
Collimator lat
TILT : Tube open
TILT SLIDE SLIDE
DOWN
UP UP DOWN
-15°

VERTICAL

LOCK ALL LOCK

REALSE

125/146
Operation Manual: DM-5125

PART 4. Tube Stand

1. DM-FCT-1 [Floor to Ceiling Mounted]

126/146
Operation Manual: DM-5125

SECTION 1. INTRODUCTION

Congratulations on your purchasing our floor – to – ceiling tube support.


This manual will guide you how to use our tube support effectively
Please read the contents thoroughly.

SECTION 2. FEATURES

Dongmun Tube support Units series are designed for heavy duty usage with all tables and erect bucky stands,
assuring a wide and precise tube positions with stability, rigidity of tube arm and, support capacity.
Its fully counter balance providing all positions and angulations.

SECTION 3. SPECIFICATIONS

Model DM-FCT-1 : Floor To Ceiling


Rails Floor 2,440mmL
Ceiling 2,440mmL
Stand vertical 650-1,900mm
Transversal 880-1,230mm
Longitudinal Min. 1,400mm
Rotating of Vertical Column ±180º manual detents at each 90º positions
Rotating of Horizontal X-ray Beam ±180º, manual lock
Lock Release System Electromagnetic Lock
Lock Release System Vertical, Transverse,
Longitudinal, Rotating
Components Tube Stand,
Tube Arm & Ring
Base Assembly
Floor/Ceiling Rails
Angle Indicator
Power requirements DC 24V
Dimensions 2100x1,020x2,150Hmm

127/146
Operation Manual: DM-5125

SECTION 4. CONSTRUCTIONS

128/146
Operation Manual: DM-5125

129/146
Operation Manual: DM-5125

SECTION 5. OPERATION

This tube support unit is operated by operation to move longitudinally, Transversally, and vertically, and locked
electromagnetically.

1. Special care must be taken not to let the X-Ray tube unit come in contact with the table or other objects during
the traveling.

2. Avoid rough handling and giving a jerk.

3. The functions of Operation switches:

Angle Indicator

Rotating Lock S/W All Lock S/W Vertical Lock S/W

Longitudinal Lock S/W Horizontal Lock S/W

130/146
Operation Manual: DM-5125

SECTION 6. MAINTENANCE AND INSPECTION

1. The rails should be kept clean from dust

2. Check the parts regularly.

3. External inspection
a. Any defect in the unit
b. Lack of lubrication
c. Any deformation or corrosion
d. Connecting parts (Pins and screws): Wear, deformation, breakage, loosening or missing of those parts.
e. Bearings: Wear, rusting or lack of lubrication
f. Other check Points
Any failure or defects in the section related with the X-Ray Tube support.

131/146
Operation Manual: DM-5125

PART 5. Bucky Stand

1. VB-1 [Bucky Stand]

132/146
Operation Manual: DM-5125

SECTION 1. DESCRIPTION

Dongmun Vertical Bucky Stand , DM-VB-1 is designed for space efficiency and adopt free standing mounting and
counter weight balance system, It is engineered for use of radiography of Chest Skull, Thoracic, spinal. The height
is widely adjusted from 50 cm to 165 cm above the floor level.

SECTION 2. SPECIFICATIONS

Vertical Column Counter weight balance system, Floor mounted


Vertical Travel Approximate 120cm,
Center to floor distance 50cm – 165 cm above floor level.
Front Plate to Film distance Approximate 3 cm
Inherent Filteration of Front plate Less than 0.5 mm Al. eq. at 100kVp
Vertical Locking System Electromagnetic lock. DC 24V
Cassette Loading Left or Right loading
Reciprocating Standard : Solenoid
Option : Motor
Bucky Device Unit Film Formats Up to 14 X 17 “ ( 35 X 43 cm )
Grid Standard Bucky 103LP1 10;1 F.D = 40-72”
Option Bucky 150LPI 10;1 F.D = 40-72”
Dimensions 590W x 570D x 1800mmH
Power requirements 110/ 220 VAC , 50/60Hz
Options Bucky device for 15angle, upward tilting

133/146
Operation Manual: DM-5125

SECTION 3. OPEARTION

3-1. Components

Stand

Jaw Rest
Vertical Movement
Bucky Device

Lock Switch

Base

3-2. How to Operator

1. Power ON:
1. Tube Stand power will be supplied automatically by Power ON of console.
2. Turn ON the Console Power.
a. To operator the system, push the power ON switch of Console.
b. After power on, system will diagnosing automatically during 15 seconds.

2. How to Use
a. Movement of Electromagnetic lock
- This product uses electromagnetic lock method.
- Electromagnetic lock will be released when push the lock switch right side of Cassette tray.
Electromagnetic lock will be locked when released the lock switch.

134/146
Operation Manual: DM-5125

- Bucky device will be release if pull the hand grip of cassette tray and push the cassette tray and then bucky
device will be locked.

b. Movement of Bucky device


- When turn ON the system, bucky device will be locked by electromagnetic and if push the right side of lock
switch, electromagnetic lock will be released and if release the lock switch, electromagnetic lock will be
locked again.
- Do not move bucky device in locked condition.

c. How to use Bucky device


- Bucky device controls (reduce) X-ray scattering and grid was installed.
- When turn ON the system, bucky device will be locked by electromagnetic, when pull cassette tray hand
grip, electromagnetic lock will be released.
- Pull out cassette tray hand grip and make a position with moving left or right.

d. How to install cassette on cassette tray


- Insert cassette to the cassette tray as below image.
- Take cassette out from cassette tray as reverse procedure.

1 2 3

135/146
Operation Manual: DM-5125

Circuit Diagram

136/146
Operation Manual: DM-5125

PART 6. High Tension Cable

1. High Tension Cable

Refer to the Attached Specifications Data Sheet.

137/146
Operation Manual: DM-5125

PART 7. Collimator

1. LB-50 [Collimator]

138/146
Operation Manual: DM-5125

SECTION 1. PRODUCT DESCRIPTION

1-1. General Description

This X-ray collimator is designed for such applications as shown in the specifications of this manual and must be
sued with the X-ray tube to restrict and define the area covered by the X-ray beam and also to reduce stray
radiation.

The beam that is collimated by the two control knobs shall over the smallest area of the patient necessary to
produce the required radiograph. The collimator that allows a step less adjustment of a field size utilizes a light
localizer to project a visible beam with coverage equivalent to the coverage of the X-ray beam.

1-2. Features and Performance

1. Provided with a push button switch timer to automatically switch off the lamp of light localizer in 30 seconds.

2. The lamp of light localizer easily replaceable with minimum error of the same lamp position and the lamp
filament position can be easily adjusted.

3. Maximum shutter opening of collimator is 48cm x 48cm at SID 100cm

4. Minimum X-ray field size at SID 100cm is 0cm x 0cm (blades overlapped)

5. The average illuminance of light field at SID 100cm is 160Lux or more.

6. The contrast ratio on the light field edge is 4 : 1 or more, which allows an easy identification of light field and
its size.

7. Leakage radiation is within 35mR/h at a position 100cm from the focal spot.

139/146
Operation Manual: DM-5125

1-3. Specification

Applications : Manual-driven adjustable collimator for radiography


Maximum kVp : 150kVp
Maximum field size : 48cm x 48cm at SID 100cm
Minimum field size : 0cm x 0cm at SID 100cm
Project lamp : Halogen lamp, 24VAC, 150W
Lamp timer : Push button type 30sec, electronic timer
Light projection : More than 160Lux at SID 100cm
Inherent filtration : 1.2mm Al equivalent
Tube focal distance : 60mm as the standard tube focal distance from tube focus to the bottom surface of
range
Field coverage scale : 13, 15, 18, 20, 23, 25, 28, 30, 36 and 40cm
SID indicate scale : 90, 100, 120 and 180cm
Leakage radiation : Not more than 35mR/h
Rotation of radiation field : +180
Cone rail tracks : Mounted (148W x 164L cm)
Dimension : 183W x 188D x 178H mm
Weight : 5Kg

1-4. Dimension and Function

1 X-ray field adjust knobs 7 Installation plate of flange


2 Push button switch of lamp 8 Stoper knob
3 X-ray field coverage scale 9 Front cover
4 Back cover 10 Cone rails
5 Body case 11 Center cross marked acrylic plate
6 Flange ring

140/146
Operation Manual: DM-5125

SECTION 2. INSTGALLATION AND OPERATION

2-1. Responsibility of Installer


It is the responsibility of the installer to correctly install this collimator as instructed in this manual. The installer
must certify that the X-ray equipment on which the housing assembly the collimator is installed is in compliance
with KFDA regulations.

2-2. Electrical Supply Requirement


It is imperative that a sufficient power is supplied to the collimator lamp to obtain the specified voltage under on-
load condition being measured at the lamp socket.

2-3. Installation Method


The following instruction should be read and thoroughly understood attemption to installed collimator. It is
assumed that all persons installing, adjusting and maintaining this collimator have the knowledge and experience
necessary to achieve the objective outlined.

1. Unpacking
Carefully unpacking the collimator and check each piece for visible damage incurred during shipment. Any
damage should be referred to the agency that delivered the collimator.

2. Collimator mounting procedures


Collimator mounting consists of removing the flange ring from the collimator, fastening the flange ring to the
X-ray tube port bottom and then fastening the collimator to the flange ring assembly.

a. Remove the flange ring (A) from the top of the collimator by remove the two installation plates of ring (C).

b. Use the proper screw length to insure that the mounting screw engage at least 5 threads into the X-ray tube
port bottom, or mounting, surface tapped holes if adaptor plate is used.

c. Attach the flange ring assembly to the X-ray tube (or tube support) with mounting spacer (no supply) inserted
between the ring and the X-ray tube support.

d. Thread the side safety screw (B2) in about 6mm, leaving the flange screw (B1) flush with the inside of the
ring.

e. Lift the collimator up to the installed flange ring.

f. While the supporting the collimator, thread the side safety screw (B1) in, then tighten both safety screw
(B1 and B2).

g. Check the installation thoroughly for correct and safe mounting.

141/146
Operation Manual: DM-5125

2-4. Operation

1. Push lamp timer and confirm that light field is illuminated.

2. Adjust collimator position so that the centering cross mark of collimator is aligned with the center of an image
receptor.

3. Collimator light field to the image receptor size operating the 2 manual control knobs.

2-5. Cares to be taken on Operation

1. Do not fail to use within the specified maximum tube voltage.

2. Do not use installing on those X-ray tube which focal spot distance are different from the ones specified.

3. If the lamp replacement become necessary, do it after the inside temperature decreased sufficicently.

4. Never fail to use the same type of lamp as the specified.

5. Do not touch on the projection lamp, glass mirror and front acryil plate except it become necessary to do so.

142/146
Operation Manual: DM-5125

6. In case that it is illuminated successively 5 times or so, keep it off for about 2 minutes till next operation for
cooling purpose.

2-6. Identification of Illuminance.

1. Set up
Mount the collimator on the tube housing assembly.
Locate the photometer on the measurement field, and adjust so that the distance between the tube focal spot and
input surface (image receptor) of photometer is 100cm.

2. Measurement
Adjust the blade to light field of 35cm x 35cm. Switch off the room light, and switch on the collimator lamp at
the adjusted supply voltage.
Measure the illuminance in the center of the 4 quadrants of the light field. The average illuminance of the center
of the 4 quadrants shall not be less than 160Lux.

143/146
Operation Manual: DM-5125

SECTION 3. INSTRUCTION FOR MAIN ADJUSTMENT

3-1. Lamp replacement and adjustment method of lamp filament position.

1. For replacement of lamp, first remove the back cover (4), but never move the front acrylic plate (11).

2. Mark sure first that the lamp surface temperature sufficiently decreased, and pull lamp out of the lamp socket.

3. Wrap over the replacement lamp surface with a soft cloth or the like so as not to allow a finger to touch on the
surface, and insert the lamp into the lamp socket so that the lamp pins will reach to the utmost depth of lamp
socket.

4. After the replacement work, illuminate the lamp by pushing lamp timer, and measure if the centering cross mark
is in the center of light field.

5. If the centering cross mark is deviated from the center, adjust the lamp filament position as follows.
(See the drawing below.)

a. Measure A size B size


Increase A size is larger, loosen the adjustment screw and move the adjustment plate (Z) op-down adjustment
then fasten the screw.

b. Measure C size D size


Increase C size is larger, loosen the adjustment screw and move the adjustment plate (X, Y) to right side then
fasten the screws. In case D is larger, move the adjustment plate to left side.

144/146
Operation Manual: DM-5125

3-2. Glass Mirror Replacement

For replacement of glass mirror, first remove the back cover and body cover, and open the collimator shutter to the
maximum positions in both longitudinal and transverse.

145/146