Operation Manual
TABLE OF CONTENTS
PART 1. Generator
1. DM-100[12.5kW] ----------------------------------------------------------------
2. DM-200, 300[15kW] ----------------------------------------------------------------
3. 20kW/32kW/40kW/50kW/80kW ----------------------------------------------------------------
4. DIG-650RF[50kW 2-Tube] ----------------------------------------------------------------
PART 7. Collimator
1. LB-50[Collimator] ----------------------------------------------------------------
2/146
Operation Manual: DM-5125
PART 1. Generator
1. DM-100[12.5kW Generator]
3/146
Operation Manual: DM-5125
NOTICE 1
This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines
thoroughly before your using the generator.
The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No
responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which
many results from the use of this manual.
Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and
shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein.
NOTICE 2
1. This equipment should be used only by the legally qualified persons and practitioners.
4/146
Operation Manual: DM-5125
6. If the device is broken, mark proper description and ask repairs to the engineer.
1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate
measures.
2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off
the power during operation.
3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.
5/146
Operation Manual: DM-5125
1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic
drugs, oxygen and hydrogen near the working locations.
2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must
be checked after the occurrence of the earthquake.
3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use
this machine” and ask repairs to the engineers.
4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.
4. Each cable has enough length and flexibility but do not bend the cable forcibly.
DANGEROUS
6/146
Operation Manual: DM-5125
The following symbols will be used in this manual: their meanings and applications are described further below.
It warns of the conditions and situations, if not taken over or avoided, that
DANGER cause heavy damages to people and even the death.
It warns of the conditions and situations, if not taken over or avoided, that
WARNING could cause heavy damages to people or cause unrepeatable damages to the
equipment.
It warns of the conditions and situations, if not taken over or avoided, that
CAUTION could cause damages to people, or to the equipment.
WAIVER CLAUSE
Dongmun do not take responsibility for the following causes for damage and repairs.
1. Any installation, movement, modification and repairs by other engineering services, which was not designated
by Dongmun.
3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun.
4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual.
5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the
powers, installation and working places.
6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders.
7. Accident and damage happened by the use beyond the original purpose of the machine.
8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the
conditions and results of the medical prescriptions.
7/146
Operation Manual: DM-5125
1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from
normal use and operation of product.
3. Note that defects not related to the main functions of the product are not eligible for this warranty.
6. Dongmun is not responsible for indirect damage related to the use of products.
7. Dongmun is not responsible for injury or damage after the end of warranty period.
8. All detailed adaptability warranties and other warranties are subjected to this statement.
8/146
Operation Manual: DM-5125
1-1. COMPOSITION
EMI AD/DC HF
AC Input HV Tank
Filter Converter DC/AC Inverter
X-Ray
Tube
IGBT
Driver
Starter
Operation
Panel
HT-Controller
Filament
Interface
AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is
inverted in high-frequency voltage of about 20kHz by IGBT inverter.
The high-frequency voltage is converted into the 40kV to 125kV DC high voltage by a high-voltage transformer
and the multi-voltage rectifier circuit. This DC high voltage is supplied to an X-ray tube through a high-voltage
cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage controls the tube current.
The operation panel can operate selection in each mode of operation and control, selection and setting of
conditions. These operation conditions are transmitted by the communication system from OP-CONT to HT-
CONT. HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.
9/146
Operation Manual: DM-5125
The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.
Operation Panel
Main CPU
kV F/B
PWM
PWM
Filament mA F/B
kV & mA Feedback Amplifier
Filament
IGBT
Inverter
HV Transformer
kV
mA
Filament Transformer
A C
FILAMENT
X-RAY Tube
10/146
Operation Manual: DM-5125
1-2. SPECIFICATION
1. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.
11/146
Operation Manual: DM-5125
2. Keep away from the place with strong vibration and maintain proper environment and conditions.
OPERATION ENVIRONMENT
Temperature range + 10°C ~ + 40°C,
Relative Humidity Range 30% ~ 75% RH (No dews)
Others No Direct Sunlight
No combustible and corrosive gases
3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere.
12/146
Operation Manual: DM-5125
1) FRONT PART
13/146
Operation Manual: DM-5125
2) REAR PART
14/146
Operation Manual: DM-5125
1) REMOTE HT-CONTROLLER
2) OPERATION PANEL
15/146
Operation Manual: DM-5125
16/146
Operation Manual: DM-5125
17/146
Operation Manual: DM-5125
Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.
Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.
Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.
18/146
Operation Manual: DM-5125
Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.
Choose after the selection of body size under the APR radiography.
The saved kV and mAs will be displayed on each segment.
19/146
Operation Manual: DM-5125
Select and display the sort of Bucky. The LED will be turned on during Bucky 1 and turned off at non-bucky.
When the ready switch is pushed, it rotate the rotor, control the heating of filament and make the exposure at
ready condition.
Make the exposure of x-rays by pressing this siwtch after the ready LED is displayed.
Touch off from the button after the radiography is completed.
20/146
Operation Manual: DM-5125
21/146
Operation Manual: DM-5125
3-4. RADIOGRAPHY
1. PREPARATION
After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table. If the
X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the lamp will
be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography preparation is
ended.
If you touch out from the hand switch during exposure, the x-ray
shooting is stopped and you cannot get clear pictures. Please touch
CAUTION off your hand after the x-ray exposure signs and the buzzer sound are
turned off except the case you want to stop the x-ray exposure
urgently
22/146
Operation Manual: DM-5125
PART 1. Generator
2. DM-200[15kW Generator]
23/146
Operation Manual: DM-5125
NOTICE 1
This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines
thoroughly before your using the generator.
The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No
responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which
many results from the use of this manual.
Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and
shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein.
NOTICE 2
1. This equipment should be used only by the legally qualified persons and practitioners.
24/146
Operation Manual: DM-5125
6. If the device is broken, mark proper description and ask repairs to the engineer.
1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate
measures.
2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off
the power during operation.
3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.
25/146
Operation Manual: DM-5125
1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic
drugs, oxygen and hydrogen near the working locations.
2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must
be checked after the occurrence of the earthquake.
3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use
this machine” and ask repairs to the engineers.
4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.
4. Each cable has enough length and flexibility but do not bend the cable forcibly.
DANGEROUS
26/146
Operation Manual: DM-5125
The following symbols will be used in this manual: their meanings and applications are described further below.
It warns of the conditions and situations, if not taken over or avoided, that
DANGER cause heavy damages to people and even the death.
It warns of the conditions and situations, if not taken over or avoided, that
WARNING could cause heavy damages to people or cause unrepeatable damages to the
equipment.
It warns of the conditions and situations, if not taken over or avoided, that
CAUTION could cause damages to people, or to the equipment.
WAIVER CLAUSE
Dongmun do not take responsibility for the following causes for damage and repairs.
1. Any installation, movement, modification and repairs by other engineering services, which was not designated
by Dongmun.
3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun.
4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual.
5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the
powers, installation and working places.
6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders.
7. Accident and damage happened by the use beyond the original purpose of the machine.
8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the
conditions and results of the medical prescriptions.
27/146
Operation Manual: DM-5125
1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from
normal use and operation of product.
3. Note that defects not related to the main functions of the product are not eligible for this warranty.
6. Dongmun is not responsible for indirect damage related to the use of products.
7. Dongmun is not responsible for injury or damage after the end of warranty period.
8. All detailed adaptability warranties and other warranties are subjected to this statement.
28/146
Operation Manual: DM-5125
1-1. COMPOSITION
EMI AD/DC HF
AC Input HV Tank
Filter Converter DC/AC Inverter
X-Ray
Tube
IGBT
Driver
Starter
Operation
Panel
HT-Controller
Filament
Interface
AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is
inverted in high-frequency voltage of about 20kHz by IGBT inverter.
The high-frequency voltage is converted into the 40kV to 125kV DC high voltage by a high-voltage transformer
and the multi-voltage rectifier circuit. This DC high voltage is supplied to an X-ray tube through a high-voltage
cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage controls the tube current.
The operation panel can operate selection in each mode of operation and control, selection and setting of
conditions. These operation conditions are transmitted by the communication system from OP-CONT to HT-
CONT. HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.
29/146
Operation Manual: DM-5125
The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.
Operation Panel
Main CPU
kV F/B
PWM
PWM
Filament mA F/B
kV & mA Feedback Amplifier
Filament
IGBT
Inverter
HV Transformer
kV
mA
Filament Transformer
A C
FILAMENT
X-RAY Tube
30/146
Operation Manual: DM-5125
1-2. SPECIFICATION
Output Power 15 kW
kV Range 40 ~ 125kV (1kV step incr./Decr.)
Small Focus 50, 100, 150
mA Range
Large Focus 200
Time Range 0.004 ~ 6.0sec in 33 Steps
mAs Range 0.4 ~ 320mAs in 30 Steps
kV Accuracy ±10 %
mA Accuracy ±15 %
Digital Display Digital Display (LED & 7 - Segment)
Self-Diagnosis Code Display in LED
APR 30 Memories
kV 3 Digital LED
Display mA 3 Digital LED
Sec/mAs 4 Digital LED
Generator 280W X 370H X 452D
Dimension
Operator Panel 161W X 53H X 367D
Generator 14Kg
Weight
Operator Panel 2Kg
1. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.
31/146
Operation Manual: DM-5125
2. Keep away from the place with strong vibration and maintain proper environment and conditions.
OPERATION ENVIRONMENT
Temperature range + 10°C ~ + 40°C,
Relative Humidity Range 30% ~ 75% RH (No dews)
Others No Direct Sunlight
No combustible and corrosive gases
3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere.
32/146
Operation Manual: DM-5125
1) FRONT PART
12
CN204
1
TB502
TB1
TB3
1 1
3 3
33/146
Operation Manual: DM-5125
2) REAR PART
3
TB102
1
CN108
TB2 CN107
2
1
34/146
Operation Manual: DM-5125
1) REMOTE HT-CONTROLLER
DIP S/W No. Function Remark
DIP S/W 1 Not Defined
DIP S/W 2 Not Defined
Apply to mass chest x-ray
DIP S/W 3 ON Use mass chest x-ray
OFF Not use mass chest x-ray
Bucky Type Selection
DIP S/W 4 ON Motor Type Bucky
OFF Solenoid Type Bucky
2) OPERATION PANEL
DIP S/W No. Function Remark
ON Calibration Mode
DIP S/W 1
OFF Normal Mode
ON -
DIP S/W 2
OFF -
ON -
DIP S/W 3
OFF -
ON RAM Monitor Mode
DIP S/W 4
OFF Normal Mode
ON -
DIP S/W 5
OFF -
ON -
DIP S/W 6
OFF -
ON -
DIP S/W 7
OFF -
ON X-ray END Buzzer
DIP S/W 8
OFF X-ray ON Buzzer
35/146
Operation Manual: DM-5125
36/146
Operation Manual: DM-5125
1 12 2 15 3 4 16 5 6 7
9 10 11 13 14 17 18
1 Patient Size Selection Switch 10 APR Region Selection Switch & Lamp
37/146
Operation Manual: DM-5125
2) “Over” Lamp
The over lamp is lighted on when the exposure conditions exceeds the normal standards. You cannot make
exposure while the over lamp is lighted on.
3) “Error” Lamp
The lamp is lighted in case of errors.
38/146
Operation Manual: DM-5125
39/146
Operation Manual: DM-5125
40/146
Operation Manual: DM-5125
NOTICE
About “OVER” display of radiographic conditions
When set-up radiographic condition “kV”, “mA” and “mAs” or “sec” exceeds the permissible value of
equipment, OVER sign lights up.
X-ray exposure is prohibited when the over sign is turned on. Please reset the appropriate radiographic
conditions.
41/146
Operation Manual: DM-5125
3-4. RADIOGRAPHY
1. PREPARATION
After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table. If the
X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the lamp will
be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography preparation is
ended.
If you touch out from the hand switch during exposure, the x-ray
shooting is stopped and you cannot get clear pictures. Please touch
CAUTION off your hand after the x-ray exposure signs and the buzzer sound are
turned off except the case you want to stop the x-ray exposure
urgently
You cannot make radiography while the over lamp is lighted on.
CAUTION Change the conditions for the radiography in those cases.
42/146
Operation Manual: DM-5125
PART 1. Generator
3. DIG-325/525/650/1150
[20/32/40/50/80kW Generator]
43/146
Operation Manual: DM-5125
NOTICE 1
This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines
thoroughly before your using the generator.
The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No
responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which
many results from the use of this manual.
Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and
shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein.
NOTICE 2
1. This equipment should be used only by the legally qualified persons and practitioners.
44/146
Operation Manual: DM-5125
6. If the device is broken, mark proper description and ask repairs to the engineer.
1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate
measures.
2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off
the power during operation.
3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.
45/146
Operation Manual: DM-5125
1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic
drugs, oxygen and hydrogen near the working locations.
2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must
be checked after the occurrence of the earthquake.
3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use
this machine” and ask repairs to the engineers.
4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.
4. Each cable has enough length and flexibility but do not bend the cable forcibly.
DANGEROUS
46/146
Operation Manual: DM-5125
The following symbols will be used in this manual: their meanings and applications are described further below.
It warns of the conditions and situations, if not taken over or avoided, that
DANGER cause heavy damages to people and even the death.
It warns of the conditions and situations, if not taken over or avoided, that
WARNING could cause heavy damages to people or cause unrepeatable damages to the
equipment.
It warns of the conditions and situations, if not taken over or avoided, that
CAUTION could cause damages to people, or to the equipment.
WAIVER CLAUSE
Dongmun do not take responsibility for the following causes for damage and repairs.
1. Any installation, movement, modification and repairs by other engineering services, which was not designated
by Dongmun.
3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun.
4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual.
5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the
powers, installation and working places.
6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders.
7. Accident and damage happened by the use beyond the original purpose of the machine.
8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the
conditions and results of the medical prescriptions.
47/146
Operation Manual: DM-5125
1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from
normal use and operation of product.
3. Note that defects not related to the main functions of the product are not eligible for this warranty.
6. Dongmun is not responsible for indirect damage related to the use of products.
7. Dongmun is not responsible for injury or damage after the end of warranty period.
8. All detailed adaptability warranties and other warranties are subjected to this statement.
48/146
Operation Manual: DM-5125
1-1. COMPOSITION
EM swich HF
Power AD/DC HV TANK
Supply Delay Convertor DC/AC Inverter
EM swich X-ray
TUBE
IGBT
Driver
Stater
OP
PANEL
HT CONT
Collimator
PWR
FILAMENT
IF
AEC ION
Chamber
AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is
inverted in high-frequency voltage of about 20kHz by IGBT inverter.
The high-frequency voltage is converted into the 40 (+20kV/-20kV) to 125kV (+62.5kV/-62.5kV) DC high
voltage by a high-voltage transformer and the multi-voltage rectifier circuit. This DC high voltage is supplied to
an X-ray tube through a high-voltage cable, and PWM INV controls this X-ray tube voltage. Modulating the
Filament voltage controls the tube current.
.
The operation panel can operate selection in each mode of operation and control, selection and setting of
conditions. These operation conditions are transmitted by the communication system from OP-CONT to HT-
CONT.
HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.
49/146
Operation Manual: DM-5125
The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.
Operation Panel
Main CPU
PWM kV FB PWM
kV, mA FB Amp
Filament
mA FB
Drive
circuit rms/DC CONV
IGBT AEC
FIL. INV.
HV Trans kV
FIL. Trans. mA
A C
FIL.
X-ray TUBE
ION Chamber
50/146
Operation Manual: DM-5125
1-2. SPECIFICATION
DIG-325 20kW
DIG-325 32kW
1 Rating Output DIG-525 40KW
DIG-650 50kW
DIG-1150 80KW
Radiography 40-150kV (1kV STEP)
2 Tube Voltage kV
Fluoroscopy 40-125kV (1kV STEP)
10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160mA
Radiography (Small Focus)
3 Tube Current mA
200, 250, 320, 400, 500, 630, 800mA (Large Focus)
Fluoroscopy 0.1~4.0mA (0.1mA Step)
4 kV Accuracy ±10%
5 mA Accuracy ±15%
6 Timer Range 0.001 sec – 10 sec (38 step), 0.003 sec ~ 6 sec (PXR-161 Only)
DIG-325 0.1mAs - 630mAs
DIG-525 0.1mAs - 630mAs
7 mAs Range
DIG-650 0.1mAs - 800mAs
DIG-1150 0.1mAs - 999mAs
20mAs or more: Less than 15%
8 mAs Accuracy
Less than 20mAs: Less than 20%
9 Display Method LED digital indication
10 Setting Method Sheet Push Switch
11 Self-diagnostic Function Error Code Indication
12 Technique Selection Manual, APR, AEC
13 APR 216 exam techniques
14 AEC (Option) Ionization Chamber 5 (Option)
Display (OP Panel)
kV 3 Digit LED
mA 3 Digit LED
15 mAs 4 Digit LED (change into sec)
Time 4 Digit LED (change into mAs)
Selection/setting LED (36)
Alarm LED (2)
Weight 62 kg
16
Operation Panel 2 kg
51/146
Operation Manual: DM-5125
1. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.
g. Where the equipment is exposed to chemicals or gas.
2. Keep away from the place with strong vibration and maintain proper environment and conditions.
OPERATION ENVIRONMENT
Temperature range + 10°C ~ + 40°C,
Relative Humidity Range 30% ~ 75% RH (No dews)
Others No Direct Sunlight
No combustible and corrosive gases
3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere.
52/146
Operation Manual: DM-5125
53/146
Operation Manual: DM-5125
400. 0 m
m
200 m
.0m
36. 0 m
m
42. 0 m
m
54/146
Operation Manual: DM-5125
22
10 12
16 5
9 3 4 11 1 2 6
23 24
13 14 15 18 19 20 17 21 25 7 8
55/146
Operation Manual: DM-5125
1) Power ON Switch
Turn on the power of generator.
3) kV Setting/Display Segment
Up-Button increases the kV value of tube voltage whereas Down-button decreases kV. If you keep buttons
pressed, the value of kV will be increased or decreased continuously.
4) mA Setting/Display Segment
Up-Button increases the mA value of tube voltage whereas Down-button decreases mA. If you keep buttons
pressed, the value of mA will be increased or decreased continuously.
56/146
Operation Manual: DM-5125
57/146
Operation Manual: DM-5125
1. Power-on Switch
a. Press the Power on switch.
b. If the power is turned on, initial condition of equipment will be checked.
c. During the checking actions, "-" mark will be displayed on kV, mA, and mAs/sec display segment.
d. After the check, it displays radiographic conditions.
e. Radiographic conditions display the final status of the previous operation.
2. Power-off Switch
a. Press the Power off switch.
b. Please turn off the power supply except the time of use.
Do not turn on and off the power of the equipment continuously except
emergent cases.
CAUTION Have the interval of 30 seconds or more between ON-OFF operations of a
power supply.
58/146
Operation Manual: DM-5125
59/146
Operation Manual: DM-5125
The radiography condition will be selected after the selection of the radiography symbol, radiography region,
and radiography method.
As well the order for the selection will not be changed.
60/146
Operation Manual: DM-5125
3. The selection of density in the AEC mode can adjust the level of x-ray exposure levels.
4. “Ready” action is not operated if the bucky and field are not chosen after the AEC mode selection.
3-6. RADIOGRAPHY
1. Photography Preparation
After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table.
If the X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the
lamp will be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography
preparation is ended.
61/146
Operation Manual: DM-5125
If you touch out from the hand switch during exposure, the x-ray shooting is
stopped and you cannot get clear pictures. Please touch off your hand after
CAUTION
the x-ray exposure signs and the buzzer sound are turned off except the case
you want to stop the x-ray exposure urgently.
62/146
Operation Manual: DM-5125
SECTION 4. TROUBLES
In case any problems happen during the operation of the device, they will be indicated on the kV/mA segment.
1. Error Code
The error code is displayed when problems, such as abnormalities, occur to the equipment. While the error code
is displayed, radiography operation is forbidden for safety. A power supply should be switched off and on again
and an error code will be cleared. If the problem of equipment is not solved at the time of the power-supply re-
switched on, the error code will be displayed again.
Since it is the failure of equipment when an unusual state is not recovered, please
inform the service section of our company or our agent.
2. Check Code
Unlike an error code, a check code is the warning sign when wrong operations are applied by the user. X-ray
radiography operation is forbidden for the safety during the check code. The way to release a check code is
pressing the WR switch for 2 seconds. CH01 will be released if thermal switch is off during the re-supply of the
power.
63/146
Operation Manual: DM-5125
64/146
Operation Manual: DM-5125
# CH01 is will be released if the TUBE terminal switch is off, in case resupply of the power.
65/146
Operation Manual: DM-5125
SECTION 5. MAINTANENCE
Maintenance has check menus carried out daily by the user and periodic inspection by technical service engineers
who completed special training programs on the repair of the generator.
User checks items based on " Daily Checking Items by the operator " and should keep data records.
The following checks and maintenance procedures, together with the suggested intervals, are the manufacturer’s
recommendation for most effective Periodical Maintenance schedule for this generator.
Exclusively service personnel specifically trained on medical X-ray generator must perform service tasks here
described.
The first periodic maintenance service should be performed six (6) months after installation, and the subsequent
service every twelve (12) months.
When a trouble and a poor part are discovered, please stop use of equipments, turn off a power supply, and contact a
near service station.
INTERVAL
Check item
Daily Weekly Monthly
Air Conditioner ○
Temperature and Humidity ○
Operation Panel ○
Radiographic Condition ○
Radiation field lamp ○
Warning light in use ○
66/146
Operation Manual: DM-5125
Please check clear environment inside the operation room and x-ray room.
The system is designed to operate with normal conditions under following environment. Please keep following
temperature and humidity.
The ingredient that makes silver etc. sulfurate may be contained in the exhaust
CAUTION gas of an automatic development machine. Please install the independent air
conditioner.
Operation Panel
Please check whether each switch of an operation panel lights up.
When you clean the surface of an operation panel, it should be wiped with the soft and fine cloth. When using a
cleaner, it is made to contain in cloth and is used.
67/146
Operation Manual: DM-5125
PART 1. Generator
4. DIG-650RF
[50kW 2-TUBE Generator]
68/146
Operation Manual: DM-5125
NOTICE 1
This manual is designed to ensure correct use and operation of DIG-Series HFG (2-TUBE).
Please read all the lines thoroughly before your using the generator.
The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No
responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which
many results from the use of this manual.
Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and
shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein.
NOTICE 2
1. This equipment should be used only by the legally qualified persons and practitioners.
69/146
Operation Manual: DM-5125
6. If the device is broken, mark proper description and ask repairs to the engineer.
1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate
measures.
2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off
the power during operation.
3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures
anytime under normal conditions.
70/146
Operation Manual: DM-5125
1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic
drugs, oxygen and hydrogen near the working locations.
2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must
be checked after the occurrence of the earthquake.
3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use
this machine” and ask repairs to the engineers.
4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.
4. Each cable has enough length and flexibility but do not bend the cable forcibly.
DANGEROUS
71/146
Operation Manual: DM-5125
The following symbols will be used in this manual: their meanings and applications are described further below.
It warns of the conditions and situations, if not taken over or avoided, that
DANGER cause heavy damages to people and even the death.
It warns of the conditions and situations, if not taken over or avoided, that
WARNING could cause heavy damages to people or cause unrepeatable damages to the
equipment.
It warns of the conditions and situations, if not taken over or avoided, that
CAUTION could cause damages to people, or to the equipment.
WAIVER CLAUSE
Dongmun do not take responsibility for the following causes for damage and repairs.
1. Any installation, movement, modification and repairs by other engineering services, which was not designated
by Dongmun.
3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun.
4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual.
5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the
powers, installation and working places.
6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders.
7. Accident and damage happened by the use beyond the original purpose of the machine.
8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the
conditions and results of the medical prescriptions.
72/146
Operation Manual: DM-5125
1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from
normal use and operation of product.
3. Note that defects not related to the main functions of the product are not eligible for this warranty.
6. Dongmun is not responsible for indirect damage related to the use of products.
7. Dongmun is not responsible for injury or damage after the end of warranty period.
8. All detailed adaptability warranties and other warranties are subjected to this statement.
73/146
Operation Manual: DM-5125
1-1. COMPOSITION
EM swich HF
Power AD/DC HV TANK
Supply Delay Convertor DC/AC Inverter
EM swich X-ray
TUBE
IGBT
Driver
Stater
OP
PANEL
HT CONT
Collimator
PWR
FILAMENT
IF
AEC ION
Chamber
AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is
inverted in high-frequency voltage of about 20kHz by IGBT inverter.
The high-frequency voltage is converted into the 40 (+20kV/-20kV) to 125kV (+62.5kV/-62.5kV) DC high
voltage by a high-voltage transformer and the multi-voltage rectifier circuit. This DC high voltage is supplied to
an X-ray tube through a high-voltage cable, and PWM INV controls this X-ray tube voltage. Modulating the
Filament voltage controls the tube current.
.
The operation panel can operate selection in each mode of operation and control, selection and setting of
conditions. These operation conditions are transmitted by the communication system from OP-CONT to HT-
CONT.
HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CPU.
74/146
Operation Manual: DM-5125
The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.
Operation Panel
Main CPU
PWM kV FB PWM
kV, mA FB Amp
Filament
mA FB
Drive
circuit rms/DC CONV
IGBT AEC
FIL. INV.
HV Trans kV
FIL. Trans. mA
A C
FIL.
X-ray TUBE
ION Chamber
75/146
Operation Manual: DM-5125
1-2. SPECIFICATION
76/146
Operation Manual: DM-5125
1. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.
g. Where the equipment is exposed to chemicals or gas.
2. Keep away from the place with strong vibration and maintain proper environment and conditions.
OPERATION ENVIRONMENT
Temperature range + 10°C ~ + 40°C,
Relative Humidity Range 30% ~ 75% RH (No dews)
Others No Direct Sunlight
No combustible and corrosive gases
3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere.
77/146
Operation Manual: DM-5125
ERROR
Sec Sec
mAs mAs
kV Size WR
mA Non 1 2 3 Bucky
Field Film/
Density AEC Select Screen
78/146
Operation Manual: DM-5125
ERROR
kV
mA
min Reset
Density
R- R/F
ABC
Auto Table
79/146
Operation Manual: DM-5125
18 19 20 21 22 23 24 25 17
ERROR
Sec Sec
mAs mAs
kV Size WR
mA Non 1 2 3 Bucky
Field Film/
Density AEC Select Screen
1 6 2 3 4 5 7 9 8 10 11 12 13 14 15 16
80/146
Operation Manual: DM-5125
1) Power ON Switch
Turn on the power of generator.
81/146
Operation Manual: DM-5125
24) LCD 1
- Display the error no and description under the error conditions
- Display over when the conditions are over the standards.
- Display the selected menu in APR Mode.
25) LCD 2
- Display the APR Menu.
- Select the displayed menu by using switch 17.
82/146
Operation Manual: DM-5125
1 3 8 2 4 9 10
ERROR
kV
mA
min Reset
Density
R- R/F
ABC
Auto Table
5 6 7
1) kV display/setting Switch
It operates the setting of kV. Up-button increases and Down-button decreases the values. When you keep
pushing the buttons, the values will be changed consecutively. The selected kV value will be displayed on the
segment.
83/146
Operation Manual: DM-5125
2) mA display/setting Segment
It operates the setting of mA. Up-button increases and Down-button decreases the values. When you keep
pushing the buttons, the values will be changed consecutively. The selected mA value will be displayed on the
segment.
84/146
Operation Manual: DM-5125
1. Power-on Switch
a. Press the Power on switch.
b. If the power is turned on, initial condition of equipment will be checked.
c. During the checking actions, "-" mark will be displayed on kV, mA, and mAs/sec display segment.
d. After the check, it displays radiographic conditions.
e. Radiographic conditions display the final status of the previous operation.
2. Power-off Switch
a. Press the Power off switch.
b. Please turn off the power supply except the time of use.
Do not turn on and off the power of the equipment continuously except
emergent cases.
CAUTION Have the interval of 30 seconds or more between ON-OFF operations of a
power supply.
4. Fluoroscopy mode
5. ABC mode
This is the control mode to gain the best image under the fluoroscopy mode.
85/146
Operation Manual: DM-5125
4. A set up of mAs/Sec
mAs/sec value is chosen with the mAs/sec setting switch.
Setting range: 0.1mAs ~ 800mAs
Step :
0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 12.5, 16, 20, 25, 32, 40, 50, 63,
80, 100, 125, 160, 200, 250, 320, 400, 500, 630, 800mAs
86/146
Operation Manual: DM-5125
5. Page up/down
Selectable by page up/down key when the displayed menu on the LCD 2 is more than 7.
6. Density Selection
Increase or decrease the mAs/sec selection by density selection switch.
The selected mAs/sec will be increased when the density is +1~+5.
The selected mAs/sec will be increased when the density is -1~-5.
87/146
Operation Manual: DM-5125
b. If you push SW27 (Page Up), SW28 (Page Down), SW29 (WR) at the same time
APR store menu will be displayed on the LCD2 as following.
c.
If you select the “Write”(SW21), the numbers from 1 to 8 will be displayed on the LCD2.
d. If you select SW19, SW21, and SW23 in order, the numbers on the LCD1 will be displayed as following.
The saving places for SW19, SW21, SW23 will be allotted by order.
1 2 3
LCD1
Select the “HEAD”(SW19) of the LCD2. The head will be displayed on the number of the LCD1 and the region
will be indicated on the LCD2.
88/146
Operation Manual: DM-5125
1 HEAD 2 3
LCD1
[ /→] mark on the right bottom of LCD2 means that it has the next page. Move the previous and next page with
page up/down switch.
f. If you select “SKULL”(SW19) on the LCD 2, skull will be indicated on the No. 2 of LCD1 and method will
be displayed on the LCD2.
1 HEAD 2 SKULL 3
LCD1
g. If you select “LAT”(SW23) on the LCD2, the LAT will be displayed on the No. 3 of LCD1 and “wirte ok?”
will be screened when all the selection work of the menu is completed.
LCD2
Write OK?
SW20 SW22 SW24 SW26
h. If you push “WR”(SW29) after you select the kV, mA, sec, Body Size, they will be stored in the locations
above.
89/146
Operation Manual: DM-5125
i. If you push the SW27 (Page Up), SW28(Page Down) at the same out of APR menus, they will be returned to
the conditions before selection.
a. Push the SW27(Page Up), SW28(Page Down), SW29(WR) at the same time.
Then, the LCD 2 will display as following.
If only the word “head” is displayed, it means there is only “head” on the APR menu.
HEAD
LCD1
d. Push the SW21 and select the skull (region). If you selelct the skull it will be displayed on the LCD1 and
LCD2 as following.
90/146
Operation Manual: DM-5125
HEAD SKULL
LCD1
b. If you push the SW 27,28, 29 at the same time, they will be displayed as following and select the “delete”.
c. If you select the “delete”, the stored APR menus will be displayed on the LCD2 according to the selected
Bucky.
91/146
Operation Manual: DM-5125
d. If you push the SW19 and select the Head (symbol), the “head” will be demonstrated on the LCD1 and skull
(region) will be on the LCD2.
HEAD
LCD1
e. If you select the “skull” by pressing SW21, the LCD1 and LCD2 will be displayed as following.
HEAD SKULL
LCD1
f. If you push the SW23 and choose the LAT, the question “Delete OK?” will be screened on the LCD2 to ask
weather you want to delete.
92/146
Operation Manual: DM-5125
g. If you want to delete the selected APR menu, press the WR (SW29).
If you select the wrong menu on the APR selections, you can move to the previous steps by pressing the
SW27 and SW28 at the same time.
3-5. RADIOGRAPHY
1. PHOTOGRAPHY PREPARATION
After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table.
If the X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the
lamp will be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography
preparation is ended.
b. When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started.
c. During exposure, the exposure LED will be lighted on and the buzzer sounds.
d. If the X-ray exposure LED is turned off or buzzer does not sound, it means the radiography work is
completed. Then, you can stop pressing the hand switch.
e. If you touched out and stopped x-ray shooting during exposure, the check code will be displayed.
If you touch out from the hand switch during exposure, the x-ray shooting is
stopped and you cannot get clear pictures. Please touch off your hand after
CAUTION
the x-ray exposure signs and the buzzer sound are turned off except the case
you want to stop the x-ray exposure urgently.
You cannot make radiography while the over LED is lighted on. Change the
CAUTION conditions for the radiography in those cases.
93/146
Operation Manual: DM-5125
SECTION 4. TROUBLES
In case any problems happen during the operation of the device, they will be indicated on the kV/mA segment.
1. Error Code
The error code is displayed when problems, such as abnormalities, occur to the equipment. While the error code
is displayed, radiography operation is forbidden for safety. A power supply should be switched off and on again
and an error code will be cleared. If the problem of equipment is not solved at the time of the power-supply re-
switched on, the error code will be displayed again.
Since it is the failure of equipment when an unusual state is not recovered, please
inform the service section of our company or our agent.
2. Check Code
Unlike an error code, a check code is the warning sign when wrong operations are applied by the user. X-ray
radiography operation is forbidden for the safety during the check code. The way to release a check code is
pressing the WR switch for 2 seconds. CH01 will be released if thermal switch is off during the re-supply of the
power.
94/146
Operation Manual: DM-5125
95/146
Operation Manual: DM-5125
# CH01 is will be released if the TUBE terminal switch is off, in case resupply of the power.
96/146
Operation Manual: DM-5125
SECTION 5. MAINTANENCE
Maintenance has check menus carried out daily by the user and periodic inspection by technical service engineers
who completed special training programs on the repair of the generator.
User checks items based on " Daily Checking Items by the operator " and should keep data records.
The following checks and maintenance procedures, together with the suggested intervals, are the manufacturer’s
recommendation for most effective Periodical Maintenance schedule for this generator.
Exclusively service personnel specifically trained on medical X-ray generator must perform service tasks here
described.
The first periodic maintenance service should be performed six (6) months after installation, and the subsequent
service every twelve (12) months.
When a trouble and a poor part are discovered, please stop use of equipments, turn off a power supply, and contact a
near service station.
INTERVAL
Check item
Daily Weekly Monthly
Air Conditioner ○
Temperature and Humidity ○
Operation Panel ○
Radiographic Condition ○
Radiation field lamp ○
Warning light in use ○
97/146
Operation Manual: DM-5125
Please check clear environment inside the operation room and x-ray room.
The system is designed to operate with normal conditions under following environment. Please keep following
temperature and humidity.
The ingredient that makes silver etc. sulfurate may be contained in the exhaust
CAUTION gas of an automatic development machine. Please install the independent air
conditioner.
Operation Panel
Please check whether each switch of an operation panel lights up.
When you clean the surface of an operation panel, it should be wiped with the soft and fine cloth. When using a
cleaner, it is made to contain in cloth and is used.
98/146
Operation Manual: DM-5125
1. E7132X
99/146
Operation Manual: DM-5125
2. E7239X
100/146
Operation Manual: DM-5125
3. E7240X
101/146
Operation Manual: DM-5125
4. E7252X
102/146
Operation Manual: DM-5125
103/146
Operation Manual: DM-5125
X-ray and Gamma-ray and dangerous to both patient and operator unless Established safe exposure procedures are
strictly observed.
The useful beam can produce serious or fatal bodily injuries to any persons in the surrounding area if used by an
unskilled operator.
Adequate precautions must always taken to avoid exposure to useful beam, as well as to leakage radiation from
within the source housing or to scattered radiation resulting from the passage through matter.
The authorized to operator, participate in or supervise the operation of the equipment must be thoroughly familiar
and comply completely with the currently established safe exposure factors and procedures described in the
National Concil on Radiation Protection and Measurement ( NCRP ) “ Medical X-ray and Gamma-ray protection
for energies up to 10 Mev- Equipment design and Use “NCRP Report-33 as revised or replaced in the future.
Those responsible for the planning of X-ray and Gamma-ray equipment installation must be thoroughly familiar
and comply completely with the structural shielding requirements outlined in NCRP-34 as revised or replaced in
the further.
Failure to observe these warnings may cause serious or fatal bodily injuries to the operator, patient or those in
attendance.
OUTLINE DATA
LEGAL OBLIGATION
You have legal obligation that failure to follow the manufacturer’s instructions or manufacturer’s will affect
radiation of any component by user or installer which well affect radiation safety, causes the user or installer to
assume full responsibility for that product.
The manufacturer is required to prescribe maintenance information and a schedule of performance to ensure
equipment complies with specified para-meters After installation the installer must supply manufacturer’s data
including this manual to the equipment purchaser or user.
The equipment purchaser or user must follow maintenance instructions.
104/146
Operation Manual: DM-5125
SECTION 1. INTRODUCTION
SECTION 2. DESCRIPTION
SECTION 3. SPECIFICATION
105/146
Operation Manual: DM-5125
SECTION 4. INSTALLATION
The following factors must be consider, in order to planning an efficient Installation site layout.
- Location of tube stand.
- Location of auxiliary equipment (bucky stand, cassette holder, etc.).
- Incoming power for X-ray generator and auxiliary equipments.
The tube stand, floor-ceiling tube stand should be installed before installation of DM-FT-1 table system, and tube
stand must be aligned to longitudinal travel.
The floor should be leveled the area where the system is being installed.
This table shoule not be used in the presence of explosive or flammable gases
WARNING as might be found in an operating room.
Interconnecting cable is supplied with this system, it is highly recommended to use adequate size of floor trough
from control console to high tension transformer and rear of the DM-BT-1 table.
The trough should have removable cover flush with the finished surface of the floor.
If the power is being supplied from wall disconnect box, #12 AWG stranded wire should be used for table power
and #10 AWG for equipment ground.
2. A plum line can be used from the X-ray, align the table center line along the longitudinal travel of the
movement.
Note : To align the table bucky verses tube stand, the table-top should be removed from the DM-FT-1 table.
3. The table-top should be level (Longitudinally and transversely) at this time, if necessary the table should be
shimmed.
Note : Shims are supplied by manufacturer.
106/146
Operation Manual: DM-5125
5. Check all connections between each components before apply the power to the table.
6. Check bucky lock release function by pulling the cassette tray toward to the front direction.
7. Check the table-top center at the center of the table-top is aligned to the table bucky.
8. The table bucky function should be checked, make an exposure and verify that bucky grid starts oscillating
during the X-ray generator exposure cycle.
107/146
Operation Manual: DM-5125
SECTION 5. MAINTENANCE
At least once a month external surfaces and exposed tracks and bearing surfaces should be wiped a cloth slightly
moistened with light oil to provide sufficient lubrication for smooth operation and protect against rust.
DM-FT-1 DIAGRAM
Bucky
DC24V Solenoid
(Generator)
108/146
Operation Manual: DM-5125
2. DM-BT-1[4-way Table]
109/146
Operation Manual: DM-5125
X-ray and Gamma-ray and dangerous to both patient and operator unless Established safe exposure procedures are
strictly observed.
The useful beam can produce serious or fatal bodily injuries to any persons in the surrounding area if used by an
unskilled operator.
Adequate precautions must always taken to avoid exposure to useful beam, as well as to leakage radiation from
within the source housing or to scattered radiation resulting from the passage through matter.
The authorized to operator, participate in or supervise the operation of the equipment must be thoroughly familiar
and comply completely with the currently established safe exposure factors and procedures described in the
National Concil on Radiation Protection and Measurement ( NCRP ) “ Medical X-ray and Gamma-ray protection
for energies up to 10 Mev- Equipment design and Use “NCRP Report-33 as revised or replaced in the future.
Those responsible for the planning of X-ray and Gamma-ray equipment installation must be thoroughly familiar
and comply completely with the structural shielding requirements outlined in NCRP-34 as revised or replaced in
the further.
Failure to observe these warnings may cause serious or fatal bodily injuries to the operator, patient or those in
attendance.
OUTLINE DATA
LEGAL OBLIGATION
You have legal obligation that failure to follow the manufacturer’s instructions or manufacturer’s will affect
radiation of any component by user or installer which well affect radiation safety, causes the user or installer to
assume full responsibility for that product.
The manufacturer is required to prescribe maintenance information and a schedule of performance to ensure
equipment complies with specified para-meters After installation the installer must supply manufacturer’s data
including this manual to the equipment purchaser or user.
The equipment purchaser or user must follow maintenance instructions.
110/146
Operation Manual: DM-5125
SECTION 1. INTRODUCTION
SECTION 2. DESCRIPTION
SECTION 3. SPECIFICATION
111/146
Operation Manual: DM-5125
112/146
Operation Manual: DM-5125
SECTION 4. INSTALLATION
The following factors must be consider, in order to planning an efficient Installation site layout.
- Location of tube stand.
- Location of auxiliary equipment (bucky stand, cassette holder, etc.).
- Incoming power for X-ray generator and auxiliary equipments.
The tube stand, floor-ceiling tube stand should be installed before installation of DM-FT-1 table system, and tube
stand must be aligned to longitudinal travel.
The floor should be leveled the area where the system is being installed.
This table shoule not be used in the presence of explosive or flammable gases
WARNING as might be found in an operating room.
Interconnecting cable is supplied with this system, it is highly recommended to use adequate size of floor trough
from control console to high tension transformer and rear of the DM-BT-1 table.
The trough should have removable cover flush with the finished surface of the floor.
If the power is being supplied from wall disconnect box, #12 AWG stranded wire should be used for table power
and #10 AWG for equipment ground.
2. A plum line can be used from the X-ray, align the table center line along the longitudinal travel of the
movement.
Note : To align the table bucky verses tube stand, the table-top should be removed from the DM-FT-1 table.
3. The table-top should be level (Longitudinally and transversely) at this time, if necessary the table should be
shimmed.
Note : Shims are supplied by manufacturer.
113/146
Operation Manual: DM-5125
5. Check all connections between each components before apply the power to the table.
8. Verify that power on & off indication green lamp should be on.
9. Depress the foot-SW, the longitudinal and transverse locks should be released.
10. Check bucky lock release function by pulling the cassette tray toward to the front direction.
11. Check the table-top transverse center, depress the foot-SW and move the table-top is aligned to the table bucky,
the BZ1 (table center indication Buzzer) should be sounded
12. The table bucky function should be checked, make an exposure and verify that bucky grid starts oscillating
during the X-ray generator exposure cycle.
114/146
Operation Manual: DM-5125
SECTION 5. MAINTENANCE
At least once a month external surfaces and exposed tracks and bearing surfaces should be wiped a cloth slightly
moistened with light oil to provide sufficient lubrication for smooth operation and protect against rust.
115/146
Operation Manual: DM-5125
DM-BT-1 SCHEMATIC
116/146
Operation Manual: DM-5125
117/146
Operation Manual: DM-5125
118/146
Operation Manual: DM-5125
SECTION 1. APPLICATIONS
Used in combination with an X-Ray high tension apparatus, an X-Ray tube unit,
An X-ray tube support unit , this table is applied for the following X-Ray examinations:
1. Fluoroscopy
2. Spot Filming
3. Bucky Radiography
119/146
Operation Manual: DM-5125
SECTION 2. CONSTITUTIONS
1. DM RFT-1 consists of Radiography table, bucky device unit, spot film device, Image intensifier, Camera,
Monitor with trolley
2. The relative accessories : compression band, hand grips and foot rest.
3. Standard Accessories
- Built-in Bucky Device 1Set
- Foot Rest 1Set
- Hand Grips 1Set
- Foot Switch 1Set
120/146
Operation Manual: DM-5125
SECTION 3. SPECIFICAIONS
Parts DM-RFT-5
type motor driven
Table tilting range standard : +90˚ ∼ -15˚ option : +90˚ ∼ -20˚
speed 90˚/20 sec., 50/60Hz
type motor driven
horizontal : ±50cm
movement range upward(tilting position) : 50cm
Tabletop sliding
lateral : ±11cm(option)
3.6cm/sec., 60Hz
speed
5cm/sec., 50Hz
material Lminated 60mmT
focal distance 41cmH
Tabletop
filtration less than 1.0mm AL. Eq. at 100KV
tabletop to film 6.5cm
type solenoid reciprocating
grid standard : 40LPC 10:1, option : 60LPC, 70LPC
Bucky device cassette range 13x18cm ∼ 35x43cm(5x7" ∼ 14x17")
tarvel range 100cm
AEC mode Ion chamber(option)
parking slide parking from left to right
locking electromagnetic lock
longitudinal movement 60cm
Spot film device transversal movement ±12cm
vertical movement 14cm ∼ 50cm
grid standard : 40LPC 10:1, option : 60LPC, 70LPC
AEC mode Ion chamber(option)
Compression cone manual operation
Field collimation under couch motor driven
Image intensifier 12", 9", 6" selectable
TV Camera CCD
Image assembly
Monitor 23", 17", 15", 14" selectable
Trolley four wheels, two shelves, 51cmLx44cmWx138cmH
Power requirement 220VAC, 3KVA, 50/60Hz, single phase
Weight net weight 650kg except image assembly
Dimensions approx. 200cmL x 72cmW x 85cmH
Installation space minimum 310cmL x 158cmW x 260cmH
121/146
Operation Manual: DM-5125
Model DM-RFT-5 is designed for performing Radiography & Fluoroscopy as well as Castro Intestinal
examinations. This unit not only makes your patients feel comfortable and safe during the operation, but also
makes your diagnosis extremely efficient and reliable.
1. Table-top sliding switch and table tilting switch are provided on the tower, near fluoroscopy screen or I/I unit.
122/146
Operation Manual: DM-5125
SECTION 5. INSTALLATION
123/146
Operation Manual: DM-5125
The table tilting switches are provided on the tower, near the fluoroscopy screen, the tilting angle can be read by
the indicator provided on the front cover. The table tilting switch is provided on the spot film device.
124/146
Operation Manual: DM-5125
ON : Tube close
TILT
DOWN
Collimator lat
TILT : Tube open
TILT SLIDE SLIDE
DOWN
UP UP DOWN
-15°
VERTICAL
REALSE
125/146
Operation Manual: DM-5125
126/146
Operation Manual: DM-5125
SECTION 1. INTRODUCTION
SECTION 2. FEATURES
Dongmun Tube support Units series are designed for heavy duty usage with all tables and erect bucky stands,
assuring a wide and precise tube positions with stability, rigidity of tube arm and, support capacity.
Its fully counter balance providing all positions and angulations.
SECTION 3. SPECIFICATIONS
127/146
Operation Manual: DM-5125
SECTION 4. CONSTRUCTIONS
128/146
Operation Manual: DM-5125
129/146
Operation Manual: DM-5125
SECTION 5. OPERATION
This tube support unit is operated by operation to move longitudinally, Transversally, and vertically, and locked
electromagnetically.
1. Special care must be taken not to let the X-Ray tube unit come in contact with the table or other objects during
the traveling.
Angle Indicator
130/146
Operation Manual: DM-5125
3. External inspection
a. Any defect in the unit
b. Lack of lubrication
c. Any deformation or corrosion
d. Connecting parts (Pins and screws): Wear, deformation, breakage, loosening or missing of those parts.
e. Bearings: Wear, rusting or lack of lubrication
f. Other check Points
Any failure or defects in the section related with the X-Ray Tube support.
131/146
Operation Manual: DM-5125
132/146
Operation Manual: DM-5125
SECTION 1. DESCRIPTION
Dongmun Vertical Bucky Stand , DM-VB-1 is designed for space efficiency and adopt free standing mounting and
counter weight balance system, It is engineered for use of radiography of Chest Skull, Thoracic, spinal. The height
is widely adjusted from 50 cm to 165 cm above the floor level.
SECTION 2. SPECIFICATIONS
133/146
Operation Manual: DM-5125
SECTION 3. OPEARTION
3-1. Components
Stand
Jaw Rest
Vertical Movement
Bucky Device
Lock Switch
Base
1. Power ON:
1. Tube Stand power will be supplied automatically by Power ON of console.
2. Turn ON the Console Power.
a. To operator the system, push the power ON switch of Console.
b. After power on, system will diagnosing automatically during 15 seconds.
2. How to Use
a. Movement of Electromagnetic lock
- This product uses electromagnetic lock method.
- Electromagnetic lock will be released when push the lock switch right side of Cassette tray.
Electromagnetic lock will be locked when released the lock switch.
134/146
Operation Manual: DM-5125
- Bucky device will be release if pull the hand grip of cassette tray and push the cassette tray and then bucky
device will be locked.
1 2 3
135/146
Operation Manual: DM-5125
Circuit Diagram
136/146
Operation Manual: DM-5125
137/146
Operation Manual: DM-5125
PART 7. Collimator
1. LB-50 [Collimator]
138/146
Operation Manual: DM-5125
This X-ray collimator is designed for such applications as shown in the specifications of this manual and must be
sued with the X-ray tube to restrict and define the area covered by the X-ray beam and also to reduce stray
radiation.
The beam that is collimated by the two control knobs shall over the smallest area of the patient necessary to
produce the required radiograph. The collimator that allows a step less adjustment of a field size utilizes a light
localizer to project a visible beam with coverage equivalent to the coverage of the X-ray beam.
1. Provided with a push button switch timer to automatically switch off the lamp of light localizer in 30 seconds.
2. The lamp of light localizer easily replaceable with minimum error of the same lamp position and the lamp
filament position can be easily adjusted.
4. Minimum X-ray field size at SID 100cm is 0cm x 0cm (blades overlapped)
6. The contrast ratio on the light field edge is 4 : 1 or more, which allows an easy identification of light field and
its size.
7. Leakage radiation is within 35mR/h at a position 100cm from the focal spot.
139/146
Operation Manual: DM-5125
1-3. Specification
140/146
Operation Manual: DM-5125
1. Unpacking
Carefully unpacking the collimator and check each piece for visible damage incurred during shipment. Any
damage should be referred to the agency that delivered the collimator.
a. Remove the flange ring (A) from the top of the collimator by remove the two installation plates of ring (C).
b. Use the proper screw length to insure that the mounting screw engage at least 5 threads into the X-ray tube
port bottom, or mounting, surface tapped holes if adaptor plate is used.
c. Attach the flange ring assembly to the X-ray tube (or tube support) with mounting spacer (no supply) inserted
between the ring and the X-ray tube support.
d. Thread the side safety screw (B2) in about 6mm, leaving the flange screw (B1) flush with the inside of the
ring.
f. While the supporting the collimator, thread the side safety screw (B1) in, then tighten both safety screw
(B1 and B2).
141/146
Operation Manual: DM-5125
2-4. Operation
2. Adjust collimator position so that the centering cross mark of collimator is aligned with the center of an image
receptor.
3. Collimator light field to the image receptor size operating the 2 manual control knobs.
2. Do not use installing on those X-ray tube which focal spot distance are different from the ones specified.
3. If the lamp replacement become necessary, do it after the inside temperature decreased sufficicently.
5. Do not touch on the projection lamp, glass mirror and front acryil plate except it become necessary to do so.
142/146
Operation Manual: DM-5125
6. In case that it is illuminated successively 5 times or so, keep it off for about 2 minutes till next operation for
cooling purpose.
1. Set up
Mount the collimator on the tube housing assembly.
Locate the photometer on the measurement field, and adjust so that the distance between the tube focal spot and
input surface (image receptor) of photometer is 100cm.
2. Measurement
Adjust the blade to light field of 35cm x 35cm. Switch off the room light, and switch on the collimator lamp at
the adjusted supply voltage.
Measure the illuminance in the center of the 4 quadrants of the light field. The average illuminance of the center
of the 4 quadrants shall not be less than 160Lux.
143/146
Operation Manual: DM-5125
1. For replacement of lamp, first remove the back cover (4), but never move the front acrylic plate (11).
2. Mark sure first that the lamp surface temperature sufficiently decreased, and pull lamp out of the lamp socket.
3. Wrap over the replacement lamp surface with a soft cloth or the like so as not to allow a finger to touch on the
surface, and insert the lamp into the lamp socket so that the lamp pins will reach to the utmost depth of lamp
socket.
4. After the replacement work, illuminate the lamp by pushing lamp timer, and measure if the centering cross mark
is in the center of light field.
5. If the centering cross mark is deviated from the center, adjust the lamp filament position as follows.
(See the drawing below.)
144/146
Operation Manual: DM-5125
For replacement of glass mirror, first remove the back cover and body cover, and open the collimator shutter to the
maximum positions in both longitudinal and transverse.
145/146