Professional Documents
Culture Documents
1 Purpose .................................................................................................................................................... 3
3 Procedure ................................................................................................................................................ 5
2.1 Equipment
1) SartoJet Pump
4) Silicon Tubing
5) Torque wrench
6) Weighing Balance
7) pH indicator
8) Measuring Cylinder
9) Hose barb adapters
10) ¾” Tri-clamps and EPDM gaskets
2.2 Material
1) Sample
2) WFI
3) 5 L of 1M Sodium hydroxide
4) 0.1M Sodium hydroxide
5) 20% Ethanol
6) IPA
Retentate side
PI-1
TFF
V-1
PI-2 PI-3
Sartorius
SartoJet®
V-2
Feed Side Permeate Side
0.000g
Sartorius Balance WB-01
P-01
3 Procedure
3.1 Setting up
1) Remove filter cassette from its packaging and handle the filter by the outer edges only.
2) Check and record cassette from its packaging.
3) Check and record cassette labelling.
4) Place the filter cassette with the semi-circular cut-outs facing downward on the two
guide rods between the two holding plates.
5) Tighten alternately the nuts until the filter cassette is secured in place by the holding
plates.
6) Use a torque wrench to tighten the nuts in opposite diagonal pairs to the torque
specified below
i. Sartocon Slice holding device: 20 Nm for Hydrosart cassettes
25 Nm for PES cassettes
ii. Do not exceed, even briefly, the permissible torque, check the hydraulic pressure
or torque before use.
3.3 Determining the water flux rate for filter cassettes (CWF)
The purpose of comparing the initial flux rate with the flux rate determined after every cleaning
procedure will enable the evaluation of how well the cassette has been cleaned.
CWF is determined by measuring the permeate flow rate when flushed with water at a particular
operating condition.
18) Use either buffer solution or air to recover the hold-up volume in the system.
4 Integrity Test
Before conducting Integrity test, ensure that the filter cassettes are properly wetted by following
the steps outlined in section 3.2. The air diffusion rates given in Integrity Test Values are
reference values only.
* It does not matter if the air source is connected to the feed or retentate side. If the air source is
connected to the retentate side, the feed side will be closed and vice versa.
* It does not matter if the integrity tester is connected to the feed or retentate side. If the
integrity tester is connected to the feed side, the retentate side will be closed and vice versa.
If the measured air diffusion rates exceed those values, the membrane might be damaged.
High diffusion values can be caused by incomplete wetting of membranes or insufficient
torque pressure.
5 Additional Support
For more information and application support, do not hesitate to contact your local Sartorius
Stedim Biotech application specialists.