Professional Documents
Culture Documents
Table of Contents
Overview ........................................................................................................................ 2
Connecting the Smart Uploader Cable............................................................................. 4
Installing the Smart Programmer V3.0 Tool .................................................................... 5
Getting the Software File ................................................................................................ 7
Getting the Calibration File (PC board replacement only) ............................................... 8
Replacing the PC Board: Software and Calibration Upload ........................................... 10
Upgrading the Software Version: Software Upload ....................................................... 16
System Type Verification.............................................................................................. 21
Troubleshooting ............................................................................................................ 22
Catalogue numbers for Smart Systems .......................................................................... 27
Introduction The uploader tool Smart Programmer V3.0 is a computer program that must
be used when performing one of the following cases on a Smart range
Direct-Q, RiOs-DI, RiOs, Synergy, Simplicity or ICW-3000 system:
Case 1: Case 2:
PC board replacement Software version upgrade
NOTE:
A new PC board does not have any software or calibration data on it.
*The software file is obtained from the Field Service Support Database.
**The calibration file is obtained from the LWD Test Bench Manager
Database. This database CANNOT be replicated, therefore you need a
Millipore network connection to download the file.
Item Go to
1 A computer Page 4
(with Windows XP or Windows 7)
2 An RS232 serial port on the computer Page 4
(or a validated RS232/USB adaptor, refer to the
troubleshooting section)
2 The Smart Uploader Cable Page 4
(Millipore reference FTPF07908)
3 The Smart Programmer V3.0 Page 5
(from the Field Service Support Database)
4 The software file Page 7
(from the Field Service Support Database)
5 The calibration file (PC board replacement only) Page 8
(from the LWD Test Bench Manager Database)
Millipore The Smart Cable is needed to connect a computer to a system. The Millipore
reference reference for the Smart Uploader Cable is FTPF07908. The cable ends in a
4-pin Molex connector and in a 9-pin RS232 connector. If your computer
does not have a RS232 port refer to the troubleshooting section.
NOTE:
If your computer does not
have a RS232 port refer to
the troubleshooting section.
What is it The Smart Programmer V3.0 tool is the computer program used to upload a
software and calibration file onto the PC board.
The installation setup file SmartUploaderSetupV3.exe needs to be installed on
your computer in order to use the Smart Programmer V3.0 tool.
Procedure Before installing the new software Smart Programmer V3.0 tool, make sure
that the software Smart Programmer V2.4 is removed (if previously installed
on your computer).
Step Action
1 Go to the Control Panel Select Add or Remove Programs
2 • Select the Smart Programmer file Click on Remove
• If the Smart Programmer file is not shown Delete the
Smart Programmer V2.4 icon from the desktop
To install the Smart Programmer V3.0 on your computer, follow the procedure
below:
Step Action
1 Go to the Field Service Support Database O&M manuals
Others Uploader Software Smart Platform
2 Save the SmartUploaderSetupV3.exe file from the database onto
your computer.
3 Double-click the
SmartUploaderSetupV3.exe
file.
Procedure
(continued) Step Action
5 The Smart Programmer V3.0 can be accessed by clicking on the
desktop icon.
Where to find it The software file (file type .s19) is obtained from the Field Service Support
Database.
Procedure To obtain the software file from the database, follow the procedure below:
Step Action
2 Select Software
Select Smart system
Example shown:
Software version V3.80
for Direct-Q or RiOs-DI
systems only
What is it The calibration file contains the default calibration values specific to the
system.
Where to find it The calibration file (file type .cal) is obtained from the LWD Test Bench
Manager Database.
Note:
A network connection to Strasbourg server is needed to generate the
calibration file. A replicated copy of the database does not work.
Procedure To obtain the file from the database, follow the procedure below:
Step Action
1 Go to the LWD Test
Bench Manager
Database.
Procedure
(continued)
Step Action
3 • Click on
1- Choose system
− Select system by selecting the catalogue number then the lot
number then the serial number (click on the arrows)
Refer to the reference list at the end of this document if needed
• Click on
2- Create the file .cal
The calibration file is automatically saved on your C:\ drive. The
filename contains the catalogue, lot and serial numbers.
Exemple: ZRQSVP030 – F5EN91285 – F5EN91285I.cal.
Before you When a new PC board is installed on a Smart system, the software and
upload calibration files need to be uploaded onto the PC board.
Before performing the upload procedure, you must have:
• obtained the software file from the Field Service Support Database
• obtained the calibration file from the LWD Test Bench Manager
Database (server connection required)
• connected the Smart Uploader Cable between the system and the
computer
Upload To upload the software and calibration files, follow the procedure below:
procedure
Step Action
The system
display looks like
this.
Step Action
2 Click on
Connect
Device.
The system
display now
looks like this.
3 Go to the File
Identification
box.
Click on Load
File.
4 Select the
system software
file (file type
.s19).
Click on Open.
Step Action
5 Click on Device
Configuration.
6 Scroll the
Selected Model
menu and select
the system type.
If your system
has a UV lamp,
then check the
UV present box.
7 Click on Prog.
Device.
8 Click on OK.
ATTENTION
Do not click on Disconnect device after this step. If it is pressed here, then the system
display will be in alarm mode (“err cal” will be displayed).
9 Go to the
Calibration file
box.
Click on Load
File.
10 Select the
calibration file
(file type .cal).
Click on Open.
11 Click on Upload
Config.
Step Action
12 Click on OK.
13 Click on
Disconnect
Device.
Before you When a new software version is released, the new software file needs to be
upload uploaded onto the PC board.
• obtained the software file from the Field Service Support Database
• connected the Smart Cable between the system and the computer
Step Action
1 Unplug and
re-plug the power
cord into the power
outlet.
Step Action
Click on Connect
Device anytime
during these 10
seconds.
3 Go to the File
Identification box.
Step Action
4 Select the system
software file (file
type .s19).
Click on Open.
5 Click on Device
Configuration.
6 Scroll the
Selected Model
menu and select
the system type.
If your system
has a UV lamp,
then check the
UV present box.
Step Action
7 Click on Prog.
Device.
8 Click on OK.
9 Click on
Disconnect
device.
System type The system will have an incorrect behaviour if the system type is wrong.
verification The System Type is the second Display Screen shown after powering ON the
system.
110 Direct-Q
121 RiOs-DI UV
120 RiOs-DI
130 RiOs
141 AFS UV
140 AFS
151 Synergy UV
or ICW-3000
150 Synergy
161 Simplicity UV
160 Simplicity
No RS232 If your laptop does not include an RS232 port, then you should use the following
port on PC accessories to add a RS232 port to your laptop:
• Europe only
Ordering information:
Both items are available from Oracle iProcurement (DELL B2B Store), make
sure you have access and follow the procedure below:
Step Action
1 Type the following address in your browser:
http://millitwister.millipore.com/oracle/oracle11i.nsf
2
Click on:
3 Enter Username and Password
4 Select XX Internet Procurement (XX = your location)
5 Click on:
6 Click on :
Ordering information:
Both items are available from Oracle iProcurement (DELL B2B Store)
The ordering procedure is similar to the European one except Dell
Corporation becomes Dell Inc. (US).
The extension cable is located in “Accessories” and the E-Legacy Extender
can be found in "Accessories”/”E Series Docking Solutions”.
Step Action
1 Install adaptor’s driver as per instructions
2 Reboot laptop
3 Click on the start menu button
4 Open “Control panel”
5 Open “System”
6 Open the “Hardware” page
Open “Device manager”
Time Out error A time out error can be caused by a faulty cable connection and also by a
communication port issue.
If a standard RS232/USB adaptor is used, it has to be associated to COM1.
NOTE: the E-Legacy port ALWAYS uses COM1.
If the RS232/USB adaptor is associated with a different COM port, then the
following procedure should be followed:
Step Action
1 In “Ports” right click on the RS232/USB port
2 Select “Properties”
3 Select “Port Settings” tab
NOTE: This is where the speed settings can be modified
4 Click on “Advanced”
5 Select COM1 in the “COM Port number” list, even if COM1 is
already used
6 Click “OK” twice
7 Close the Device Manager window
8 Click “OK” in the “system properties” window
9 Close the “control panel” window
Adaptor not If a RS232/USB adaptor is used and the RS232/USB adaptor is not visible in
visible in the “Ports” section, then the following procedure should be followed:
“Ports”
Step Action
1 Check the name of the adaptor on the installation CD or on the
adaptor’s box, compare with names in “Ports”
2 If the adaptor still cannot be found then re-install the adaptor’s
driver