Service Manual

Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

© Copyright 2014 Boditech Med. INCOPERATION

The information contained herein is subject to change without


prior notice. The only warranties for the ichroma™ II and services
are set forth in the express warranty statement accompanying such
products and services. Nothing herein should be construed as
constituting an additional warranty. Boditech Med Inc. shall not be
liable for technical or editorial errors or omissions contained
herein.
Additionally, any malfunction in elements including the optical
sensor assembly, the system board and the drive train should be
consulted with and serviced by Boditech Med Inc. Any
unauthorized operations on elements not suggested in this
document will void the existing warranty.

1/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Contents

1. Product Description 3

1.1 Product Features 4

1.2 Identifying front parts 5

2. Troubleshooting 7

Flowchart 2.1 – Initial Troubleshooting 8

Flowchart 2.2 – No Power 9

Flowchart 2.3 – No LCD display 10

Flowchart 2.4- Non-responding Touchscreen 11

Flowchart 2.5- Cartridge holder does not show 12

Flowchart 2.6- No ID chip error 13

Flowchart 2.7- Barcode Error 14

ichroma™ II error message description 15

3. Parts Catalog 17

3.1 Serial Number Location 17

3.2 The ichroma™ II Major Components 18

3.3 The ichroma™ II Moving Part Components 20

4. Removal and Replacement Preliminaries 21

4.1 Tools Required 21

4.2 Service Considerations 21

4.3 Plastic Parts 21

4.4 Cables and Connectors 21

4.5 Preventing Electrostatic Damage 22

5. Removal and Replacement procedures 22

5.1 Disassembly Sequence chart 22

5.2 Preparing the ichroma™ II for disassembly 23

5.3 System 23

5.4 Bottom case 25

5.5 Application Module 26

5.6 Top Case 27

6. Specifications 28

7. Connector Listing 29

8. Screw information 32

9. Software Update 32

2/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

1. Product Description

The Fluorescence reader, the ichroma™ II, is a one-dimensional scanner configured to


measure and quantity the fluorescence signal. The system consists of a controller, a test
device carrier and a detection module as well as the user interface elements such as
Touchscreen, a display window and output ports. The ichroma™ II runs on DC12V 5A
power source, which is external to the main body. The AC adapter requires 100~240V AC,
50/60Hz power source

3/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

1.1 Product Features


Features
 Dimension
 276 X 220 X 91 mm ( W X L X H )
 Weight
 1.3 kilograms
 Processor/Chipset
 Main Processor : Android / Sub Processor : MSP430XXX
 LCD Module
 TFT 7 inch 800x480
 7 inch (800x480) Touchscreen
 Light source
 Laser Diode and LED
 Sensor
 Si-Detector
 Power
 AC adaptor
 AC input : 100~240V AC, 50/60 Hz
 DC output 12V and 5A
 Accessories
Default
 Power Cable & AC Adapter[Included]
 User Manual [Included]
 System check cartridge set [Included]
Optional
 Barcode Scanner
 Thermal Printer [Built-in]
 Keyboard
 USB OTG Cable
 WIFI Dongle
 Bluetooth Dongle
 SD card
 AA battery 4ea

4/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

1.2 Identifying front parts

The front face of the ichroma™ II is shown below and its parts are identified in Table 1-1

Overal View

Table 1-1
Front
Item Components Function
1 LCD display & Touchscreen Result and operation Control, Display
2 ID chip socket Connects an ID Chip
3 Device carrier port Accept/Ejects device carrier
4 Temperature Sensor It measures room temperature around ichroma™ Ⅱ.
5 Built-in Printer(Optional) Result and operation Control, Printing

5/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

The parts on the back side of the ichroma™ II are shown below and described in Table 1-2

Bottom & Back side

Table 1-2
Back side
Item Components Function
When the battery cover is opened, you can install/set
1 Battery Cover
batteries and SD card.
2 Power Switch Main power switch
3 Power Connector For connecting the Power Cable
4 Mini USB Connector For connecting other device
5 USB Port For connecting Barcode Scanner / Keyboard/ Mouse
6 LAN Port For connecting LAN cable

6/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

2. Troubleshooting

WARNING: Only authorized technicians trained by Boditech Med Inc.


should repair this equipment. All troubleshooting and repair procedures are
detailed to allow only subassembly-/module-level repair. Because of the
complexity of the individual boards and subassemblies, do not attempt to
make repairs at the component level or modifications to any printed wiring
board. Improper repairs can create a safety hazard. Any indication of
component replacement or printed wiring board modification may void any
warranty or exchange allowances.

Table 2-1
Troubleshooting Flowcharts Overview(continued)
Flowchart Description
2.1 “Flowchart 2.1- Initial Troubleshooting”
2.2 “Flowchart 2.2- No Power”
2.3 “Flowchart 2.3- No LCD display”
2.4 “Flowchart 2.4- Non-responding Touchscreen ”
2.5 “Flowchart 2.5- Cartridge holder does not Show”
2.6 “Flowchart 2.6- No ID chip error”
2.7 “Flowchart 2.7- Barcode Error”

7/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Flowchart 2.1 – Initial Troubleshooting

Begin
troubleshooting

Power Go to “Flowchart 2.2 - No


On? N
Power“

LCD
Windows N
Go to “Flowchart 2.3 - No
On? LCD display“

Error Go to “Error message


Messages? N
discription“

Touchscreen
not responding N Go to “Flowchart 2.4 - Not
working Responding Touchscreen“

Cartridge holder Go to “Flowchart 2.5 –


in and out?(when N Cartridge holder does not
selected In/Out) show“

Operating Go to “Flowchart 2.6 - No


N
ID chip
ID chip error“

Scanning
without N Go to “Flowchart 2.7 - Barcode
“barcode Error“ Error“

END

8/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Flowchart 2.2 – No Power

No Power

Connected
to the power Reconnect the cable Power on? Done
N to the power outlet
Y
line?

N
Y

Power cable
Connected to N Connect Power on? Y Done
Power cable
ichroma™ II

N
Y

Power on? Y Done

Change
Power Cable

Check Change AC
Power on? N power switch Abnormal
adapter

Power
On? Y
Y
Replace the main board
Nomal -> Call customer service

Done Done

9/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Flowchart 2.3 – No LCD display

No LCD display

Power on? Go to
N “Flowchart 2.2 – No Power

Re-start ichroma™ II

LCD display is
OK ? Y Done

Connected to the
FFC cable and N Reconnect the cables.
Lcd display cable?

LCD display
Y Is ok ? Y Done

N
Change FFC cable.

LCD display
Is ok ? Y Done

LCD display N
Is ok ? Y Done

N Main board failure


Call customer service

Change LCD module

10/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Flowchart 2.4- Non-responding Touchscreen

Touchpad
Not responding
properly

Connected to the
FFC cable and N Reconnect the cables.
Lcd display

Is it possible to
select button? Y Done

Change LCD module

Is it possible to
select button? Y Done

Main board failure


Call customer service

11/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Flowchart 2.5- Cartridge holder does not show

Holder does not work in


spite of touching In/Out
button.

Is the motor cable


connected?
N Reconnect the cables.

Does the holder Done


work?
Y

Change a motor.

Does the holder


work? Y Done

Main board failure


Call customer service

12/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Flowchart 2.6- No ID chip error

No ID Chip

Insert ID Chip? N Insert ID Chip

“Insert ID Chip“
Message on LCD Y Insert another ID Chip

Done OK? Y Done

Reseat ID Chip
Socket connector

Replace the ID Chip


OK? N
Socket connector

Done OK?

OK? N Replace the ID Chip


Socket module

Call customer service

13/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Flowchart 2.7- Barcode Error

Barcode Error

Reseat test device

OK ? Y Done

Insert other
test divice

Check for foreign materials


OK ? N In the holder

OK ? N Call customer service


Done

Done

14/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

ichroma™ II error message description

Recommended Remedial / Corrective


Error code Probable causes
measures

Technical problems in test


Error 25 - 31 Try again the test with new test cartridge.
cartridge

Turn the power off and contact regional


Error S1, Error
Optical failure distributor or Boditech Med Inc.’s
S2
technical support team.

Improper sample or buffer Try again the test with new test cartridge
Value Error 1 - 10
is used. and test sample.

Technical problems in test


Caution Error Try again the test with new test cartridge.
cartridge

Check barcode on test cartridge is


damaged or contaminated by foreign
materials..
Problem in identification of Insert test cartridge into the cartridge
Barcode Error holder to the end.
barcode on test cartridge
If problems are not solved, contact
regional distributor or Boditech Med Inc.’s
technical support team.

ID chip and test cartridge Match ID chip and test cartridge. Then try
Mismatch error
do not match. again the test.

Try again after checking user ID and


Password error Incorrect password
password.

ID chip cartridge LOT number of ID chip and


Match LOT number of ID chip and test
Lot Number test cartridge do not
cartridge. Then try again the test.
Mismatch match.
Change ID chip and try again with new ID
chip.
Invalid Item ID chip problem If problems are not solved, contact
regional distributor or Boditech Med Inc.’s
technical support team.

SW & Test batch Test item is not included in Contact regional distributor or Boditech
not compatible software. Med Inc.’s technical support team.

15/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Optical failure during self- Contact regional distributor or Boditech


Optics part Error !
testing procedure Med Inc.’s technical support team.

Moving part Moving part failure during Contact regional distributor or Boditech
Error ! self-testing procedure Med Inc.’s technical support team.

Internal storage failure Contact regional distributor or Boditech


Storage disk
during self-testing
Error ! Med Inc.’s technical support team.
procedure
It is possible to proceed test. But test
result may not correct due to operation
Temperature sensor failure temperature is out of suggested
Temperature
during self-testing condition(15-35℃).
sensor Error !
procedure
Contact regional distributor or Boditech
Med Inc.’s technical support team.
It is possible to proceed test. But it is
Built-in printer failure impossible to print the test result out.
Printer initializing
during self-testing Contact regional distributor or Boditech
Error !
procedure
Med Inc.’s technical support team.
It is possible to proceed test. But test
result may not correct due to operation
Environment temperature
Temperature out temperature is out of suggested
is out of operation
of recommended condition(15-35℃).
temperature range during
range. Proceed?
self-testing procedure Please set environment temperature as
15-35℃.

16/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

3. Parts Catalog

This chapter provides an illustrated parts breakdown and a reference for spare part numbers
and option part numbers.

3.1 Serial Number Location

When ordering parts or requesting information, provide the ichroma™ II serial number
and model number located on the bottom of the ichroma™ II

Serial Number location

17/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

3.2 The ichroma™ II Major Components

Table 3-1

Spare Parts: The ichroma™ II Major Components


18/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Item Description Spare Part Number


1 ichroma II Top Case
2 7” touch LCD
3 ichorma II LCD Bracket
4 ichroma II LCD B/D
5 ichorma II ID Chip B/D
6 ichroma II Printer option Cover
7 ichroma II Bottom Case
8 ichroma II Battery terminal_02
9 ichroma II Battery terminal_01
10 ichroma II Main B/D
11 ichroma II Moving Assembly
12 ichroma II Holder Hinge bracket
13 ichroma II Holder
14 ichroma II Front Case
15 ichroma II Temp. Cover & Temp. B/D
16 ichroma Rubber foot
17 ichroma II Battery Cover
18 ichroma II AA Battery
19 ichroma II Printer Cover
20 TP29FP6RZE Roller(Printer Roller)
21 TP29FP6RZE Module(Printer Module)
22 Afias6 Printer Case
23 TP27FX Control B/D(Printer Control B/D)

19/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

3.3 The ichroma™ II Moving Part Components

Table 3-2

Spare Parts: The ichroma™ II Moving Part Components


Item Description Spare Part Number
1 ichroma II Moving Base
2 ichorma II Motor Block
3 Stepper Motor(ichorma II Motor)
4 ichroma II Optic
5 ichorma II Cartridge Holder
6 ichroma II Barcode B/D
7 ichroma II Barcode PCB Bracket
8 Interrupter PCB

20/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

4. Removal and Replacement Preliminaries


This chapter provides essential information for proper and safe removal and replacement service.

4.1 Tools Required

You will need the following tools to complete the removal and replacement
procedures:
 Magnetic screwdriver
 Flat-blade screwdriver

4.2 Service Considerations


The following sections include some of the considerations that you should keep in
mind during disassembly and assembly procedures.

As you remove each subassembly from the ichroma™ II, place the
subassembly (and all accompanying screws) away from the work area to
prevent damage.

4.3 Plastic Parts

Using excessive force during disassembly and reassembly can damage plastic parts. Use
care when handling the plastic parts. Apply pressure only at the points designated in
the maintenance instructions.

4.4 Cables and Connectors

CAUTION: When servicing the computer, ensure that cables are placed in
their proper locations during the reassembly process.Improper
cable placement can damage the ichroma™ II

Cables must be handled with extreme care to avoid damage. Apply only the tension
required to unseat or seat the cables during removal and insertion. Handle cables by
the connector whenever possible. In all cases, avoid bending, twisting, or tearing
21/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

cables. Ensure that cables are routed in such a way that they cannot be caught or
snagged by parts being removed or replaced.

4.5 Preventing Electrostatic Damage


Many electronic components are sensitive to electrostatic discharge (ESD). Circuitry
design and structure determine the degree of sensitivity. Networks built into many
integrated circuits provide some protection, but in many cases, the discharge contains
enough power to alter device parameters or melt silicon junctions.
A sudden discharge of static electricity from a finger or other conductor can destroy
static-sensitive devices or microcircuitry. Often the spark is neither felt nor heard, but
damage occurs. An electronic device exposed to electrostatic discharge might not be
affected at all and can work perfectly throughout a normal cycle. Or the device might
function normally for a while, then degrade in the internal layers, reducing its life
expectancy.

5. Removal and Replacement procedures

This chapter provides removal and replacement procedures.

5.1 Disassembly Sequence chart


Use the chart below to determine the section number to be referenced when removing
the ichroma™ II components.

Disassembly Sequence Chart


Section Description # of Screws Removed
5.2 Preparing the ichroma™ II for disassembly 6
5.3 System 6

5.4 Bottom case 13

5.5 Application Module 0


5.6 Top Case 10

22/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

5.2 Preparing the ichroma™ II for disassembly

Before you begin any removal or installation procedures:


1) Power off the ichroma™ II . (Turn off the System.)
2) Disconnect all external devices connected to the ichroma™ II
3) Disconnect the AC power cable.

5.3 System
1) Loosen the 6 screws that secure the lower case to the base board

Loosens the screws


2) Turn the ichroma™ II front side up.
3) Lift the front side of the Top Case and swing to the lift gently.
4) Remove the Front Case.

Lifting the Top Case and remove the fornt case


23/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

5) Disconnects the ID chip socket connector①, LCD connecter②, Printer Power connecter③,
Printer data connecter④ From the Main board
 Disconnects the ID chip socket connector ①

 Disconnects LCD connecter②

 Disconnects Printer Power connecter③ & Printer Data Connector④

24/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

5.4 Bottom case


1) Disconnects the Battery terminal connector①, Motor connector②, Interrupter
connecter③, Barcode connecter④, Optic connecter⑤ From the Main board

2) Remove the Moving Part and Main B/D(Remove the eleven screws).

25/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

3) Remove the Temp. Cover (Remove the two screws).

5.5 Application Module


1) Remove the application Module

26/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

5.6 Top Case


1) Remove the LCD.

2) Remove the ID Chip B/D.

3) Remove the Printer Option Cover(option: Printer Module)

27/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

6. Specifications

Table 6-1
ichroma™ II
Dimensions 276 (L) x 220 (W) x 91 (H) mm
Weight 1.3 kg (±0.1)
Power requirements 12V DC @5A
Temperature 15℃~35℃
Relative humidity Max 70%
Dry, clean, flat, horizontal surface away from direct
Operating location
sunlight and mechanical vibration.
Laser diode, 2.5mW,
Light source
LED source
Detector Silicon photo diode
Communication USB 4 port, LAN Port, USB OTG port
Display TFT 7 inch LCD. Printer(Optional)
Keypad Touchscreen
Patient result : 1000
Result saved System check test result: 500
Control material test result: 1000
Item saved 500(5 LOTs/item)
- This device meets the EMI guideline as per EN61326-2-6

28/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

7. Connector Listing

This section provides specification and refernce information for the conncetor.

interrupter PCB to Main B/D Cable

Table 7-1
Left pin Right pin
1 to 1
2 to 2
3 to 3

ID Chip B/D to Main B/D Cable

Table 7-2
Left pin Right pin
1 to 1
2 to 2
3 to 3
4 to 4
5 to 5

29/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Barcode B/D & Temp. B/D to Main B/D Cable

Table 7-3
Left pin Right pin
1 to 1
2 to 2
3 to 3
4 to 4

Printer Data Cable

Table 7-4
Left pin Right pin
1 to 1
2 to 2
3 to 3
4 to 4
5 to 5
6 to 6

30/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

Printer Power Cable

Printer module connector Main B/D connector

Table 7-5
Left pin Right pin
1 to 1
2 to 2

LCD B/D to Main B/D(FPC cable (60 PIN))

Optics B/D to Main B/D Cable(FPC cable(40 PIN))

31/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

8. Screw information

Tapping 2 M3x8 Machin M3x8 Tapping 2 M2x6 Tapping 2 M3x10

9. Software Update

1) Application updating
Setup > System > Update > Application
① Open the battery cover in the bottom of ichroma™ Ⅱ.
② Insert a SD card which includes updating information into the SD card slot as following
picture.

③ Tap “Application” in the “Update” screen.

32/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

④ Tap “OK” button. ⑤ Tap “Install” button.

⑥ When updating is completed, tap “Done” button. Then ichroma™ Ⅱ will restart
automatically.
If there is no SD card in the SD card slot or there is no updating
Note information in a SD card, ichroma™ Ⅱ will display a message for
checking SD card.

2) Firmware updating
Setup > System > Update > Firmware

 Firmware updating using SD Card


① Open the battery cover in the bottom of ichroma™ Ⅱ.
② Insert a SD card which includes updating information into the SD card slot as following
picture.

33/34
Doc. No SVM-IR2-EN
ichroma™ II Rev.No 00
Service Guide Rev. Date 2016. 02. 29

③ Tap “Firmware” in the “Update” screen.

④ Tap “Yes” button to proceed firmware updating.

⑤ When updating is completed, ichroma™ Ⅱ will restart automatically.

34/34
Respect for life

Boditech Med Inc.


43, Geodudanji 1-gil, Dongnae-myeon,
Chuncheon-si, Gang-won-do, 200-883, Korea
Tel: +82-33-243-1400 Fax: +82-33-243-9373