Professional Documents
Culture Documents
Experimantal work
1
1- Introduction :
Pressure is one of the broadest and most complex areas in the field of
physical measurements. Its large scope results from the numerous and
different types of instruments which are used for pressure measurement,
while Its complexity stems from the many-sided nature of pressure itself .
The second type of pressure one must consider is that which is artificially
produced to transfer or amplify force to produce work . This is the type of
pressure which is used in a service station to lift a car or in a machine shop
to operate a hydraulic pressure. These lectures will deal primarily with
static fluid (liquid and gas) pressure and its measurement. The study of
fluids at rest is called hydrostatics .
2
2 - Concepts :
P=F/A (1)
dynes/cm 2 PPPP
PPPP
F=AP (2)
3
Types of Pressures
Atmospheric pressure P atm BBBB
The units, psi and atm, often carry a trailing “a” or “g” to indicate that the
pressure is absolute or gauge pressure. Thus, by psig, we me gauge
pressure in psi and by psia, we me absolute pressure in psi. So is the
meaning of atma or atmg if nothing is noted otherwise.
Example:
A pressure gauge on a tank reads 1.5 bar. What is the gauge pressure?
Solution:
4
The gauge reads the gauge pressure directly. So, we must interpret the
reading as gauge pressure and to avoid confusion, we should report the
pressure as 1.5 barg instead of 1.5 barg.
Example:
A gauge on a tank reads 2.0 bar. What is the absolute pressure in the tank?
Solution:
Pabs = Patm + Pgauge
Since atmospheric pressure was not given, we must assume it is 14.7psi.
Thus, we find the absolute pressure in the tank as:
P abs = 1.01325+2.0
BBBB BBBB
= 3.01325 bara
Example:
The pressure gauge on a tank reads 20. cm Hg vacuum. What is the
absolute pressure in the tank?
Solution:
The gauge reads a vacuum gauge pressure directly. The pressure is below
atmospheric, or is –20 cm Hg relative to atmospheric pressure. Thus, the
absolute pressure is:
Pabs = Patm + Pgauge = 76 + (- 20)
= 56 cm Hg
5
Pressure Sensing Devices
6
Example:
What is the pressure (in kPa) caused by a 25.0 cm column of fluid (SG =
6.43) at sea level and ambient conditions?
Solution:
P gh
kg m
h 25 cm 0.250 m; ref 1000 ; g 9.8 In SI units:
m3 s2
P= ρgh = (ρ water *SG)* 9.8 * 0.25
BBBB BBBB
Example:
What is the pressure (in psi), caused by a 6.34 ft column of mercury at
ft
ambient temperature in a location where g 29.7 2 ?
s
Solution:
From Table B1, we find the specific gravity for Hg: SG 13.546 .
lb
The density of water (reference fluid): ref 62.4 m3
ft
Result 34.4
7
Figure 4: Arrangement of Manometers
8
Figure 5: Manometer variables
The line ‘ab’ is at the interface between the manometer fluid and the
higher-pressure fluid. The hydrostatic pressure on each leg is the same at
that point. It becomes our reference point. We do a pressure balance by
equating the pressures on each leg.
Applying the above allows us to develop the general manometer equation
as:
P1 1 gd1 P2 2 gd2 f gh
Example:
Determine the pressure drop across the orifice meter as shown below
Fluid: water
r 1 = 1000 kg/m3
32 mm
Manometer Fluid d
r f = 1100 kg/m 3 10 mm
Solution:
This is a differential manometer. Note that the hydrostatic pressure above
the 32 mark is the same on both sides, it cancels out. The reference line to
select is at the 10 mark. The manometer equation is
P1 1 g d l P2 f gd 1 gl
which simplifies to
P1 1 gd P2 f gd
We now can write the pressure difference across the orifice from the above
equation as:
P1 P2 P gd f 1
Substituting the appropriate known quantities, we get:
9
m kg 1 N
P 9.8 2 0.022 m 1100 1000 3
s m 1 m kg
s2
N
21.6 2 21.6 Pa
m
This expression states that the total force is equal to the product of the
pressure and the total area over which it is acting.
Example 1
A 1000 kg weight is placed on the top of a vertical steel column which has
a cross sectional area of 50 cm 2 . What is the pressure on the end of the
PPPP
PPPP
column?
Solution
P = 20 kgf/cm 2PPPP
3 - Fluid columns :
P=W/A (6)
11
If we do not know the weight of the fluid, but we do know its density ( )
and its height, we can determine the pressure it will generate. First,
remembering that weight is equal to the product of mass and gravitational
acceleration, we may say :
W = mg
Substituting for the weight in equation (6) we get :
mg
P = ------ (7)
A
The density of the fluid () is defined in terms of its mass per unit volume
as :
m
= ---- (8)
V
Solving equation (8) for m we get :
m=V
gv
P = --------- (9)
A
Since the volume of any column is equal to the product of its cross
sectional area and height, we may rewrite equation (9) as :
g(hA)
P = ----------------
A
or P=gh ( 10 )
11
Equation (10) says only that the force per unit area is the same. Another
illustration that the fluid pressures at any given level are equal. Equation
(10) is used quite often in the measurement of pressure ; however, its
form, as presented here, makes it somewhat cumbersome to use. Let us,
therefore, at this time, introduce the concept of weight density (D).
Weight density is the weight of a substance per unit volume and is defined
mathematically as :
D=w/v ( 11 )
Since : W=mg
D = mg / V
But : m/V=
So : D=g (12 )
The only difference between density ( ) and weight density (D) is that
the latter will vary with the gravitational acceleration, since it contains
weight as a term.
Since the pressure is only dependent upon the height of the column and the
type of fluid, the height of the column is sometimes used as a shorthand
method to describe a pressure. If someone said that a certain pressure was
equivalent to a 10 inch column of mercury, anyone who desired the actual
12
pressure in psi could simply multiply the given height by the weight
density of mercury.
Vacuum, is another area where in the units which have been adopted are .
based on the height of a column of mercury. Since vacuum concerns itself
with relatively low pressures, the units which have been adopted represent
small linear distances.
millionth high .
4- Types of Pressure : -
Actually, there are three types of pressure, each of which requires slightly
different handling. These are absolute, gauge and atmospheric pressure.
This pressure is due to the weight of the atmosphere acting over the
surface of the earth, or if you will, a column of air which extends several
hundred thousand meters above the earth's surface. This atmospheric
pressure is part of our environment. It is not noticed because the pressure
inside the human body is equalized, and thus no differential exists .
If the atmosphere was removed from the earth, there would be no pressure.
The absence of all pressure we shall call true zero pressure . Any pressure
which is referred back to true zero pressure is called absolute pressure.
13
When atmospheric pressure is described, it is always considered as an
absolute pressure, since it is referred to true zero pressure. Many times a
pressure is generated or measured which is not referred to true zero
pressure, but rather to the natural environment.
P BBBB
abs
BBBB =P BBBB
gauge
BBBB +P BBBB
ATMOSPHERIC PRESSURE :
Since most pressures are measured as gauge, generally all pressures are
considered to be gauge, unless otherwise specified. Either the abbreviation
kg or kgf for gauge pressure is acceptable .
The measurements are referred to true zero and indicate the pressure above
zero, as illustrated in
14
Pressure Gauges
15
The basic Bourdon tube is a hollow tube with an oval cross section and
shaped in a circular arc of about 180°.
The size, shape, and material of a Bourdon tube depend upon the
pressure range and the type of gauge desired. Low pressure Bourdon
tubes ( pressures up to 150 bar) are often made of stainless steel or other
high pressure materials. high pressure Bourdon tubes tend to have more
circular cross sections than their lower range more oval counterparts .
As a pressure is applied to the Bourdon tube, its circular shape and oval
cross section force it to straighten, causing a displacement. This
displacement, although not . linear with respect to pressure, is a function
of pressure applied .
16
When selecting a pressure gauge consideration should be given to a
number of parameters which have an effect on gauge accuracy, safety,
and utility such as:
Accuracy required
Selection Criteria
Dial size
17
Second, by a requirement that the pressure gauge user/operator be able
to read the scale from a distance that physical circumstances may
dictate .Pressure Gauge Range Selection - The pressure gauge selected
should have a dial size which permits the desired readable scale
increments from the operator required viewing distance.
Many companies have a wide range of dial sizes from 1½", 2", 2½", 4",
4.5", 6", 8", 10", 12", 16" up to 20 inch dial size available.
Also a wide varieties of accuracy class are available starting from 0.06%
through 0.1%, 0.25%, 0.6%, 1%, 1.6%, 2.5%, 4% and ended by 5%
Gauges with diaphragms are also available for separation between the
gauge and the media of operating.
Selection Criteria
18
This will give the operator a visual indication of normal, ideal operating
conditions. This means a "green light" condition and everything is
operating the way it should be. If however ,the "visual" of the pointer is
way below or way above this 12 o'clock position, then we have a
situation that needs attention
Pressure transducers
19
· digital processing module,
· digital indication (in the unit(s) specified by the manufacturer),
· electrical power supply unit (generally incorporated).
21
Liquid manometer
Evangelisti Torricelli inversion of a liquid-filled tube to generate a
vacuum was not unique - he was but one of a group centered on Galileo
that was attempting to refute the accepted wisdom that "nature abhors a
vacuum". However, he is generally credited with the discovery that the
height of the mercury in the inverted tube could be used as a measure of
atmospheric pressure.
This discovery can be fairly called the beginning of pressure metrology,
was followed by a number of replications of his experiments, and the
use of Torricellian barometers in important experiments
During the last 350 years the mercury manometers had various forms
which incorporated most of the means used in U- tube manometers for
both detecting a liquid and measuring the height of a liquid column. The
different types of mechanisms for detecting the liquid surface and
measuring the height of a liquid column for the most important types of
these mechanisms were:
1. Index Point Detectors
2. Float Detectors
3. Visual and Optical Detectors
Column height measured by scale
Column height measured by precision screw
Photocell Detectors
4. Optical Interferometer Designs
5. White-Light Interferometer Manometers
6. Laser Interferometer Manometers
7. Capacitance/Gauge Block Manometer
8. Capacitance/Interferometer Manometer
9. Ultrasonic Manometers
21
22
5. Dead Weight Testers (Pressure Balances) : -
P= F/A (1)
Consider the circular piston and cylinder of Fig. 7. The piston has some
cross sectional area (A). The total weight of the piston assembly is made
23
up of the weight of the piston ( W p ) and any weight (W) placed on top of
BBBB BBBB
the piston .
When the force exerted on the bottom of the piston is slightly greater than
the total weight of the piston assembly (W + W p ) , the piston will be
BBBB BBBB
moved up. If this upward force is exactly the same as the total weight of
the piston assembly, it will neither rise nor fall, but rather it will be in
equilibrium .
Used
Weights
Piston
Cylinder
Applied pressure
The pressure of the fluid in the cylinder generates a force over the bottom
of the piston .The force acting on the bottom of the piston is a function of
the fluid pressure and piston area, or from equation (2) :
F=PA
24
determined to be in equilibrium, the magnitude of the pressure can be
measured by using the expression :
P=W/A
Normally a piston area is selected which will allow weights to be used that
are convenient, with respect to the pressure range desired .
As the piston area gets smaller, it requires less weight to generate a given
pressure; however, there is a minimum diameter which can be specified for
the piston before the tolerance in machining becomes a' problem .
A=W/P
100 kg
A = ----------------
2,000 kg/cm 2 PPPP
25
A= 0.05 cm 2
PPPP PPPP
The piston diameter, if desired, can be found by using the formula found :
D2 PPPP
A= --------
4
or, solving for the diameter, we get :
4A PPPP
D= -------
Substituting for the required area we get :
4 x 0.05 PPPP
D = ----------------
3.1416
= 0.0637
or : D = 0.252 cm
Since a diameter of one-quarter cm is easily machined to close tolerance,
this piston area is suitable. The complete tester would be as illustrated in
Fig. 8. It would be made up of10 weights of10 kg each .
Since the piston assembly has weight, it would be desirable if its weight
also 10 kg and thus could act as one of the load weights .
The weight of the piston itself would define and produce the WEIGHTS
minimum pressure, 200 kg/cm 2 . As each weight was added, it would
PPPP PPPP
generate an additional 200kg/cm 2 , until with nine weights plus the piston
PPPP
PPPP
the pressure would be 2,000 kg/cm 2 . Although simplified, this is the PPPP
PPPP
Since each weight operates with a fixed piston area, the weights can be
marked to indicate the nominal pressure in Kg/cm 2 that they will generate PPPP PPPP
26
Let us illustrate the weights which might be supplied with a typical tester.
Consider 10,000 kg/cm 2 unit with a piston area of 0.01cm 2 .
PPPP PPPP PPPP PPPP
Table (2) lists the weights available, their serial number, and their nominal
pressure (considering the piston area) .
Notice that the total of all the weights is 100 kg, as which is required when
10000 kg/cm 2 is generated . The minmum pressure is 100 kg/cm 2 , as
PPPP
PPPP
PPPP
PPPP
limited by the piston. The resolution is also 100 kg/cm 2 , as limited by the
PPPP PPPP
1 1,000 10
2 1,000 10
3 1,000 10
4 1,000 10
5 1,000 10
6 1,000 10
7 1,000 10
8 1,000 10
9 1,000 10
- 400 4
- 200 2
1 100 2
2 100 1
piston 100 1
27
The weights are serialized because in precision work neither the nominal
area of the piston nor nominal pressure, as indicated on the weights, is
sufficient. If pressures of +0.1% of indicated value ,or better are to be
generated, the actual mass of each weight used and the piston area must be
known to a minimum of five significant figures .
One more word about the basic concept of a dead weight tester. Dead
weight testers, almost without exception, use air, oil, or water as a pressure
media. Since friction cannot be tolerated, there must be some clearance
between the piston and cylinder. This clearance allows some of the fluid to
leak, by the piston.
If the leakage rate of the fluid by the piston is not closely controlled, the
piston may tend to fall too rapidly; thus, equilibrium cannot be assured.
The piston is generally rotated in the cylinder either manually or
mechanically to decrease friction and the fluid leakage which restricts the
piston fall rate. A rotation rate which has proved adequate for most testers
is one (1) rev/sec. .
Due to the fact that some leakage does exist, the effective area of the
tester is not that of the piston alone. It is taken as the mean area of the
piston and cylinder. If a piston has an area of 0.02000 cm 2 , and its
PPPP
PPPP
0.02000 + 0.0210
A = ---------------------
2
or
A = 0.02005 cm 2 PPPP
P=W/A (14)
28
If the indicated 'pressure (P) is to be accurate to +0. I% of indicated value
or better, the weights and effective area must be known to a level, which
is compatible with the accuracy required, assuming that the tester in
question has the precision and stability required .
effective area is 0.01cm 2 . Notice that the calibration value given is mass
PPPP
PPPP
weight in kgf. This means the value listed is for standard gravity.
P = -------- -------- ( 15 )
A gS BBBB BBBB
Where: W = m * g L BBBB
29
TABLE 3. PARTIAL DEAD WEIGHT CALIBRATION REPORT
Serial No. Nominal Pressure Weight ( Kg )
( Kg/cm 2 ) PPPP PPPP
1 1,000 10.0063
2 1,000 10.0058
3 1,000 10.0072
4 1,000 10.0028
5 1,000 10.0037
6 1,000 10.0041
7 1,000 10.0035
8 1,000 10.0062
9 1,000 10.0018
- 400 4.006
1 200 2.005
2 200 2.001
- 100 1.0006
piston 100 1.0003
* Values given are for " apparent mass versus brass “ in air
The values given in Table 3 are further specified as " apparent mass versus
brass " in air. Even though the weights may be made out of cast iron, they
must be corrected for air buoyancy as if they were made of density 8.0
g/cm 3 . Thus, equation (15) now becomes :
PPPP
PPPP
The density of air may be taken as 0.0012 g/cm 3 for almost all air PPPP
PPPP
31
buoyancy corrections concerned with dead weights .
Is it necessary to use the actual mass values and correction for air
buoyancy and gravity?
Even though the weights may be class F 1 or better individually, their BBBB BBBB
Thus far we have treated the mean area of the piston/cylinder as the
effective area. This is not strictly true. When the mean piston area was
measured it was at some temperature ( t c ) which is listed on the BBBB BBBB
calibration report. The mean area of the piston is listed at this temperature
and designated A o . This is the mean area of the piston at the temperature
BBBB BBBB
The problem, as it relates to the mean piston area , is two-fold . First , the
piston is generally made of tungsten carbide or stainless steel. The
'cylinder Is made of a similar material. If the tester is not used at the same
temperature as that of its calibration, the mean piston area will not be the
same, due to thermal expansion.
c " .
BBBB BBBB
The value of " k " and " c " will vary between testers, depending upon
BBBB BBBB BBBB BBBB
31
The mean area correction ( K ) for temperature is generally written in the
form :
K = [ 1 + ( K + c ) ( t – 20 )]
BBBB BBBB BBBB BBBB ( 17 )
The mean area (zero pressure) corrected for temperature thus becomes :
A = A o [ 1 + ( K + c ) ( t – 20 )]
BBBB BBBB BBBB BBBB BBBB BBBB ( 18 )
However, the area (A), as derived in equation (18), is not the effective area
of the tester at the point of use. Let us define. the effective area (A e ) as the BBBB BBBB
mean piston area at the temperature and pressure of use ; whereas A o was BBBB BBBB
The mean area of the piston must be established at the applied pressure
because pressure causes deformation of the piston .
A = Ao + B P BBBB BBBB ( 19 )
32
The effective area (A e ) may now be written :
BBBB BBBB
A e = ( A o + BP )[ 1 + ( K + c ) ( t – 20 )]
BBBB BBBB BBBB BBBB BBBB BBBB BBBB BBB B (
20 )
b=B/ A BBBB
o BBBB ( 21 )
A e = ( 1 + bP )[ 1 + ( K + c ) ( t – 20 )]
BBBB BBBB ( 23 )
BBBB BBBB BBBB BBBB BBBB BBBB
use equation ( 23 ) .
Example :
A dead weight tester has a mean area of 0.010000 cm 2 . The pressure PPPP PPPP
b = ----------------------------
0.01 cm 2 PPPP
b = 5 x 10 -8 /Kg PPPP
PPPP
33
This form of “ b " says that the piston area will change five parts per 100
million/kg. Not very significant on the surface; however, at 1000 Kg/cm 2 , PPPP
PPPP
the piston area will have changed +0.05%. The numerical value
of "b" is a function of the piston cylinder configuration and its materials.
It should always be considered, since its magnitude may be much larger
than illustrated here. .
We may now write the expression for the true pressure generated by
a dead weight tester as :
Wt BBBB BBBB
Pt = ------- (24)
At BBBB BBBB
P t = ------ -------------------------------------
BBBB BBBB (25)
A o (1+ bP )[ 1 +(k +c ) (t - 20)]
BBBB BBBB
Example :
A o = 0.010001 cm 2
BBBB BBBB PPPP
PPPP
b= 7x10 -8
PPPP
PPPP
/Kg
34
t = 20°C k+c=1.8x10 -4 /°C PPPP
PPPP
K 1 = 1 - -----
BBBB BBBB
b BBBB BBBB
Using the densities of. 0012 g / cm 3 and 8.4 g / cm 3 for air and brass, PPPP
PPPP
PPPP
PPPP
respectively, we get:
0.0012
K 1 = 1 - ----------
BBBB BBBB
8.4
= (1 - 0.00014 )
K 1 = 0.99986
BBBB BBBB
gL BBBB BBBB
The entire numerator of equation (25) which describes the true force thus
becomes: .
W t = 90.008 Kg x 0.99986 x 1.0006
BBBB BBBB
K 2 = [ 1 + ( k + c ) ( t - 20 ) ]
BBBB BBBB BBBB BBBB BBBB BBBB
= 1 + ( 9.0 x 10 -5 ) PPPP
PPPP
K 2 = 1.00009
BBBB BBBB
The correction for the mean area at the pressure of use may be found
using the expression:
K 3 = ( 1 + bP )
BBBB BBBB
35
The problem at this point is that the pressure (P) is still unknown.
90 kg
P = -----------
0.01 cm 2 PPPP
P = 9000 kg/cm2
K 3 =1.00063
BBBB BBBB
The total expression for the true Pressure generated is, therefore;
90.049396
= --------------
0.0100064
P t = 8997.56
BBBB BBBB
Since the weights would have nominally been marked to indicate 9,000
kg/cm 2 , the true pressure was 0.027 % lower than the indicated pressure.
PPPP
PPPP
However, this is not to imply that the error is always of this small
36
magnitude. It might have been greater , dependent upon the magnitude of
the factors discussed. ,in another situation.
There are certain other corrections which could be considered with regard
to the true pressure generated by a dead weight tester; however , the
magnitude of the corrections which they represent is so small that they are
seldom applied in general work and, therefore, will not be discussed in this
text .
One correction, however, which has not been discussed and should always
be considered when using a dead weight tester, is that of a fluid head.
Consider the test configuration illustrated in Fig.8. The pressure sensed by
the piston and weights is sensed at the bottom of the piston - level 0.
If the test gauge is above the dead weight tester it will sense a pressure
which is less than that sensed by the tester by an amount of the center of
the Bourdon tube gauge , find the true pressure applied to the gauge and
the error in the gauge, if the test media is oil .
Referring to Table I to find the weight density of oil and then substituting
it and the vertical height into the equation for pressure we get :
P= 9 x 0.03
P= 0.27
Since the gauge is above the tester, the pressure applied to the gauge is the
true pressure, as measured by the dead weight tester, minus the fluid head,
or :
P= 100 - 0.27
P= 99. 73 kg/cm 2 PPPP
measured - true
e = ------------------
true
37
or, substituting our values:
99.8 - 99.73
e = --------------
99.73
+ 0.07
e = ---------
99.73
e = + 0.0007 = + 0.7 %
the fluid head was considered, whereupon the actual error was found to be
+.07 kg/cm 2 (+0.07 %) . The magnitude of a fluid head should be
PPPP
PPPP
38
Fig (10-1) DHI pressure balance
39
Some low range units operate with gas (generally pure nitrogen) as the test
media. These instruments, known as air dead weight testers, are generally
restricted in their pressure to 60 bar or less .
Air dead weight testers are operated in exactly the same manner as any
other dead weight tester, and all corrections, as previously discussed, are
applicable except the ones concerning piston distortion and fluid head.
Since the instruments operate over relatively low pressures, the change in
area of the piston is generally negligible .
The weight and piston assembly of either type tester may be enclosed if
desired. This allows the reference pressure surrounding the piston and
weights to be varied. If the space surrounding the piston and weights is
evacuated, the tester will read absolute pressure .
Some testers overcome this problem in that the piston tilts from a vertical
position through 90°, thus, any component of the piston weight may be
used to generate a pressure.
If the piston is tipped at some angle ( ), then the total weight still acts
perpendicular to the earth, but only a component of the weight acts over
the area. The weight component W c which acts over the area is :
BBBB BBBB
W c = W ( sin )
BBBB BBBB
41
This feature allows this type of tester to generate any pressure from zero
up to the nominal piston pressure. The accuracy of the pressure determined
in this manner is restricted only by the accuracy, with which the angle ()
is measured.
41
Liquid filled Dead Weight Testers (DWT), almost without exception, use
either water or light oil for a test media. One such tester is illustrated in
Fig. 11. The instrument shown has a range of 1000 bar and a basic
accuracy, oft 0.1 % of indicated pressure. The interesting fact concerning
this tester is that it employs a by fluid system .
Oil is generally considered as the better test media for the tester since it
provides a good seal and excellent lubrication. Water, however, is
considered as a better test media for a test gauge, as it will not contaminate
the test gauge. The tester illustrated in Fig.11 circumvents this problem by
using a closed oil system for the piston and cylinder while it uses water for
the pressurized test fluid. The pressure from the water is transmitted to the
oil through a thin diaphragm, as illustrated in Fig, 11 .
One of the more interesting features of this tester is that it is dual range.
Two complete piston cylinder assemblies are provided. One low range and
one high range cylinder.
The effective area of the two piston/ cylinders is in a ratio of 5:1, each
weight supplied, therefore, will be used to generate two pressures, the high
range piston/cylinder has a capacity of 1000 bar while the low range
piston/cylinder has m operating range 200 bar using the same weights.
Each weight is marked with two nominal pressures, one if the weight is
used witl1the low range and the other if it is used with high range .
42
Fig. 12 Dual range DWT.
All of the DWT discussed to this point have employed a simple piston
which is considered as one where no rigorous employed is made to control
its distortion but instead a correction is determined and applied.
The jacket pressure is usually applied and increased until there is zero
clearance between tile piston and cylinder. It is then reduced by an amount
that will provide a known clearance as the test pressure desired. The
mathematics of determining the piston/cylinder clearance which is desired
at a given operated pressure point of a little consequence, since the-
control clearance tester is considered as a primary standard and thus, is
encountered only in labs of the highest level.
43
The simple piston/cylinder arrangement shows an increase in effective
area with increase pressure. That is not necessarily true with other
modified designs. If the effective area of the piston decreases with
increased pressure, 'B' or 'b', as a case may be, will be noted with a minus
sign. The discussion of DWT and specifically piston deformation
presented in this section was intended to explain the basic concepts and
corrections required to achieve meaning full quantitative results.
A piston gauge is essentially made up of a piston-and-cylinder assembly (PCA) and
of a system for the application of a known vertical load on the piston. The PCA is the
essential part of the apparatus, as it is used to define the effective area on which
pressure is applied. There are basically three designs of piston cylinder assembly
(PCA) .
The most commonly used design, known as the 'simple' type, is shown
diagrammatically in figure (13). The action of the pressure-transmitting
fluid on the base of the piston causes mechanical distortion leading to
excessive leakage of fluid at high. Such pressure balances can thus be used
only over a limited pressure range, although with appropriate design they
may be used successfully up to pressures of about 800 MPa.
44
Firstly, the radial gap between the piston and cylinder is substantially
small, often with an interference fit at lower pressures.
This external pressure contracts the gap between the piston and the
cylinder over part of the engagement length thus reducing the fluid
leakage; a typical design is shown in figure (14). The upper pressure limit
is set by the reduction of the gap to an interference fit.
45
In the third design, so called controlled clearance piston cylinder assembly,
a variable fluid pressure, p j , in an independent pressure system, is applied
BBBB BBBB
to the external surface of the cylinder, thus providing control of the gap
between the piston and cylinder; such an arrangement is shown in figure
(15).
P
Fig. (15) Controlled-clearance piston-cylinder assembly
Although in principle this design will cover a very wide range of pressure
using only one piston-cylinder assembly, in practice it is usually better to
use a series of assemblies to achieve the best sensitivity for a particular
pressure range. This type of pressure balance mostly used in very high
pressure applications, is not commonly available commercially, and is
generally more difficult to operate than the two other designs.
46
1-Preparation for calibration
The calibration should only be carried out when the pressure balance is in
good working order. The operation of the pressure balance under
calibration and the pressure reference standard should be carried out
according to the laboratory's calibration procedure prepared from this
guideline, and the manufacturer's technical manual.
2-Calibration room
3-Devices installation
47
Install appropriate drain to control the nature of the fluid in the
tubings.
Attach a suitable temperature measurement system.
4- Pressure generation
5- Pressure reference
The pressure reference instrument in general use for the calibration
of a pressure balance is another pressure balance. For the ranges
lower than 300 kPa, the standard instrument may be a mercury
column manometer. Other instruments may be used as an alternative
for specific cases (low gauge pressure for example).
The calibration of an absolute pressure balance may be carried out in
gauge mode, with an added uncertainty in Ao.
48
In all cases, the reference instrument used for the calibration has to
meet the following conditions:
(a) to be traceable to a National Standard with a recognised calibration
certificate.
(b) to have an uncertainty better than the presupposed uncertainty of the
balance to be calibrated. Complete the uncertainty budget on the
reference standard pressure balance to check this condition.
(a) Check that the oil is free from impurities. If not, drain all the tubings
and replace the oil in the tank.
(b) With the pressure circuit closed and half the set of weights placed on
the piston, the piston shall be moved upwards and downwards by
means of the spindle pump. Thus, the mobility of the piston is
examined over the total range of displacement.
(c) If necessary, and using the technical manual, remove the piston-
cylinder assembly, and clean the surfaces of the two pieces with a
suitable solvent or pure soap, and with dry soft cloth according to
the manufacturer's recommendations. Inspect the piston and the
cylinder for surface scratches and corrosion. Relubricate the piston
with clean liquid if the piston-cylinder operates in liquid, or if the
balance operates in gas but with an oil-lubricated piston-cylinder
assembly.
(d) Examine the free rotation time (for the hand-rotating pressure
balances only). Weights corresponding to 2/10 of maximum
pressure are placed upon the piston. The initial rotation rate should
be approximately 30 rpm. Measure the elapsed time until the piston
is stationary. This time should be at least 3 min.
(e) Examine the descent rate of the piston. The piston descent rate is
observed at maximum pressure when the piston is rotating. Measure
49
the time interval in which the piston drops from top to bottom
position. This time should be at least 3 min.
Note: For these two last parameters, the stated values should be related
to the technical instructions of the manufacturer.
(f ) Connect the pressure balance to the standard instrument.
(g) Identify the reference level for both pressure balances. The reference
level should be defined by the manufacturer at the bottom surface of
the piston when the balance is in equilibrium. In the absence of this,
and when the bottom surface of the piston is not accessible, the
reference level is generally defined at the outlet pipe connection
level. The difference in height between the reference level of the
standard and the reference level of the balance to be calibrated shall
be reduced as much as possible and measured. In any case the
difference in height between the reference levels of both standard
and balance under calibration will need to be measured in order to
apply the appropriate head correction.
(h) For absolute pressure, pump for 30 min. at the beginning of the
calibration to eliminate the water vapor in the belljar. Use dry
nitrogen as working gas.
(j) Rotate the piston or cylinder while respecting the manufacturer's
recommendation.
(k) For hand rotating balances, check the clockwise and anticlockwise
direction influence (if any), or indicate the rotation direction in the
certificate.
The scope of this method is to determine the bias error and the
repeatability of the calibrated pressure balance. This is done by
determining the generated pressure corresponding to well identified
51
weights. In that method the weighing of the masses of the
instrument under calibration is optional.
Method A procedure
Three measuring series are carried out, each of them with at least five
pressure points regularly spaced over the whole range of the
pressure balance. For twin range balances, at least five pressure
points should be carried out for each range. The pressure points
should be selected evenly spaced across the range of the instrument
under test.
Method B procedure
(a) The value of the mass of each weight (including the floating
elements when removable) of the pressure balance shall be
determined by a laboratory accredited for such mass measurements.
The relative uncertainty of the mass determination should not
usually exceed 20 % of the likely total measurement uncertainty of
the pressure balance to be calibrated. For example, if the supposed
expanded uncertainty of the pressure balance is 5x10 -5 x p, the
relative uncertainty of mass determination should be within l x10 -5
x m.
51
the pressure comparison measurements by using a least-squares
analysis: in this case a tare value in pressure units should be given.
The p-method mentioned letter allows the determination of this
initial value.
52
equilibrium should be considered as reached when the proper falling
rate of both pistons is found (no flow of fluid in the tubing between
the two pressure balances). Both pistons have to rotate during the
adjustment. In the case of hand rotating units, the influence of the
clockwise/ anticlockwise rotation, and of the spin rate will be
checked.
(e) Note the reference number of each of the weights applied on both
balances.
(f) Note the temperature of the piston-cylinder assembly of both
balances.
If the balance is not equipped with a temperature probe, note the
surrounding air temperature using an electronic thermometer
attached to some suitable point of the balance. This information
shall be included in the certificate.
53
I- Method A procedure
Usually the results will be presented for the standard value of the
gravity 9,80665 m s-2 (unless customer requests his own local
gravity) and the reference temperature 20 °C in the form of a table.
It will include:
(a) the pressure indicated by the balance under calibration (p m )
(b) the reference pressure measured by the standard instrument (mean of
the three determinations), in Pa and in the unit of the pressure
delivered by the balance if different (p r);
(c) the standard deviation of the reference pressure p r.
(d) the difference between the indicated pressure and the reference
pressure (pm-pr);
(e) the uncertainty of this difference, in the conditions of the calibration.
A table that lists all weights applied on the unit to be calibrated for
each pressure point of the calibration shall be included in the
Calibration Certificate.
54
(c) linear thermal expansion coefficients of the piston-cylinder assembly
under calibration (if not determined experimentally, e.g. by using
literature data, this shall be stated)
(d) position of the pressure reference level
(a) The computing method in more general use can be used to calculate
step by step the effective area of the pressure balance to be
calibrated from the mass applied on its piston and the pressure
delivered by the standard instrument.
(b) From this method, the effective area is calculated by reversing the
equation of definition of the pressure presented before
55
(iii) the corresponding temperature of the measuring assembly during the
calibration;
(iv) the individual value of the effective area Ap calculated at the
reference temperature and at reference pressure
(v) the mean value of the effective area Ap;
(vi) the experimental standard deviation of the mean.
56
and the uncertainty of the pressure distortion coefficient of the
standard.
57
Finally for simple and re-entrant piston cylinder assemblies the complete pressure
generated equation is given by:
gL air
Mg L g
1
W
S M
P
Ao [1 ( c p )(t Tr )](1 po )
Where:
M Total mass applied on the piston
Ao Nominal effective area for used piston cylinder
αp+αc Thermal expansion coefficient for piston and cylinder materials
t Temperature of the piston cylinder assembly
Tr Reference temperature
ρair Density of air
ρM Density of the weights
g Local gravity at place of standard PCA calibration
gL Local gravitational acceleration
gS standard gravitational acceleration
γ Surface tension of the used oil
χ Circumference of the piston
Γw Tare force
Po Nominal pressure
λ Distortion coefficient of piston cylinder assembly
58
2. Those in which the effective area of a piston cylinder assembly is obtained
by a process of comparison with some other fundamental pressure standard.
The methods of group (1) for effective area of a piston cylinder assembly are mainly
restricted to the evaluation of primary or perhaps national standards and are dealt
with in two ways:
a) Calculation from measured dimensions by analytical method.
b) Characterization of controlled clearance pressure balance using Heydmann-
Welch model for effective area of a piston cylinder assembly calculation.
The dimensional measurements (group(a)) required for the calculation of effective
areas are the diameters of both the piston and cylinder at suitably spaced positions
along the axis of the assembly. These data may usefully be supplemented by
measurements of the degree of departure from true circularity and straightness, which
will contribute to the uncertainties attaching to the final result.
In the controlled-clearance system group (b) only the dimensions of the piston are, in
principle, required. These measurements will usually be entrusted to a laboratory
specializing in dimensional metrology and we shall therefore make no attempt to
describe them. The methods of group (2) are of particular significance in the
calibration of balances used as secondary standards and working standards.
SI Unit of
Pressure
Fig (6) Traceability to the basic physical units of pressure through dimensional and mass measurements
59
According to Heydemann and Welch model for controlled clearance pressure balance
characterization, five quantities are required for the characterization of controlled
clearance pressure balance. These quantities are Ao, αP+αc , λ, d and Pz as described
by equation
Mg 1 air W
M
P
Ao [1 ( c p )(t Tr )](1 po )[1 ( D EW FW 2 )( pzo szW qzW 2 p j )]
or Pz = PZ0 + sZ W + qZ W2 Pz = PZ0 + sZ W
or d = D + EW + FW2 d = D + EW
Where first or second order equation selection depends on the shape of the obtained
data. Pj is the pressure in the applied to the outside of the cylinder to control the
piston/cylinder gap. The values of these parameters will be calculated from two
measuring experiments, the first is the falling rate measurements where the values
Pzo, sZ and qZ will be calculated. The second is the “d” measurements, in which the
effective area is measured by a crossfloat with a second pressure balance as the jacket
pressure is varied at constant system pressure. From those measurements, the
parameters D, E and F will be calculated.
61
Calibration of pressure balance using cross floating method
The effective areas of working instruments and, indeed, of many reference standards,
being usually determined by comparison with some form of independent standard. In
this part we shall describe the procedures for such comparisons and finally refer to
some points of interest in the expression of the resulting data.
Fig. (7) Schematic diagram illustrating the comparison of two pressure balance
(cross floating experiment
We have already noted that there are commonly available two independent forms of
primary standard in the field of pressure measurement, viz. the pressure balance itself
and the manometric column. The calibration of an individual pressure balance may,
in principle, be carried out by comparison with either of these primary devices, and
we shall therefore consider both techniques. It must be emphasized, however, that the
intercomparison of a pressure balance and a manometric column is generally a more
complicated and difficult operation than the comparison of two pressure balances and
the difficulty of such 'mixed' comparisons is greatly accentuated at the higher
pressures. The direct comparison with a standard balance is therefore the method
used in the great majority of pressure-balance calibrations, while the mixed
comparisons are normally reserved for applications where there are special reasons
for their adoption.
For the above reason, the determination of the ratio of the effective areas of two
piston-cylinder assemblies is a particularly important aspect of calibration, for which
the above figure gives a generalized schematic diagram. The two assemblies, of
61
effective areas SA and SB, are shown mounted on a common pressure system and are
'in equilibrium', at the applied pressure P, reckoned at an appropriate reference level,
when the loads are WA and WB respectively.
The calibration should only be carried out when the pressure balance is in good
working order. The operation of the pressure balance under calibration and the
pressure reference standard should be carried out according to the laboratory's
calibration procedure prepared from this guideline, and the manufacturer's technical
manual.
Calibration room
Devices installation
Install the devices out of the air disturbances such as ventilation and air-
conditioning.
Install the balance to be calibrated as near the standard instrument as
possible.
Use a rigid, stable table supporting the full load, checked in with a spirit
level.
Minimise the height difference between the reference levels of the two
instruments to be compared.
Respect the verticality of the piston as recommended by the manufacturer
use the built-in spirit level, or a laboratory spirit level on the top of the piston to
minimise the tilt. This should be checked also at full mass load.
Use short, wide bore pipework. This is more critical at low pressure.
Insure the cleanliness and the tightness of the tubings.
Install appropriate drain to control the nature of the fluid in the tubings.
Attach a suitable temperature measurement system.
Pressure generation
62
For gas pressure:
(a) Use a clean and dry gas (nitrogen for example), at a temperature near ambient.
(b) Adjust the pressure input to the range of the intercompared instruments.
(c) Clean the tubings of any liquid (for the oil-lubricated type).
Pressure reference
The pressure reference instrument in general use for the calibration of a
pressure balance is another pressure balance. For the ranges lower than 300 kPa,
the standard instrument may be a mercury column manometer. Other instruments
may be used as an alternative for specific cases (low gauge pressure for
example).
The calibration of an absolute pressure balance may be carried out in gauge
mode, with an added uncertainty in Ao.
In all cases, the reference instrument used for the calibration has to meet the
following conditions:
(a) to be traceable to a National Standard with a recognised calibration
certificate.
(b) to have an uncertainty better than the presupposed uncertainty of the
balance to be calibrated. Complete the uncertainty budget on the reference
standard pressure balance to check this condition.
63
spindle pump. Thus, the mobility of the piston is examined over the total
range of displacement.
(c) If necessary, and using the technical manual, remove the piston-cylinder
assembly, and clean the surfaces of the two pieces with a suitable solvent or
pure soap, and with dry soft cloth according to the manufacturer's
recommendations. Inspect the piston and the cylinder for surface scratches
and corrosion. Relubricate the piston with clean liquid if the piston-cylinder
operates in liquid, or if the balance operates in gas but with an oil-lubricated
piston-cylinder assembly.
(d) Examine the free rotation time (for the hand-rotating pressure balances only).
Weights corresponding to 2/10 of maximum pressure are placed upon the
piston. The initial rotation rate should be approximately 30 rpm. Measure the
elapsed time until the piston is stationary. This time should be at least 3 min.
(e) Examine the descent rate of the piston. The piston descent rate is observed at
maximum pressure when the piston is rotating. Measure the time interval in
which the piston drops from top to bottom position. This time should be at
least 3 min.
Note: For these two last parameters, the stated values should be related to the
technical instructions of the manufacturer.
(f ) Connect the pressure balance to the standard instrument.
(g) Identify the reference level for both pressure balances. The reference level
should be defined by the manufacturer at the bottom surface of the piston
when the balance is in equilibrium. In the absence of this, and when the
bottom surface of the piston is not accessible, the reference level is generally
defined at the outlet pipe connection level. The difference in height between
the reference level of the standard and the reference level of the balance to be
calibrated shall be reduced as much as possible and measured. In any case the
difference in height between the reference levels of both standard and balance
under calibration will need to be measured in order to apply the appropriate
head correction.
(h) For absolute pressure, pump for 30 min. at the beginning of the calibration to
eliminate the water vapor in the belljar. Use dry nitrogen as working gas.
(j) Rotate the piston or cylinder while respecting the manufacturer's
recommendation.
(k) For hand rotating balances, check the clockwise and anticlockwise direction
influence (if any), or indicate the rotation direction in the certificate.
Methods to apply
Both methods that follow are comparative ones, consisting of comparing the balance
to be calibrated and the standard instrument when both are submitted to the same
pressure and the same environmental conditions. However, dependent on the
presupposed accuracy of the balance to be calibrated, and according to customer
64
requirement alternative methods may be used:
The scope of this method is to determine the bias error and the repeatability of
the calibrated pressure balance. This is done by determining the generated pressure
corresponding to well identified weights. In that method the weighing of the masses
of the instrument under calibration is optional.
Method A procedure
Three measuring series are carried out, each of them with at least five pressure
points regularly spaced over the whole range of the pressure balance. For twin range
balances, at least five pressure points should be carried out for each range. The
pressure points should be selected evenly spaced across the range of the instrument
under test.
Method B procedure
(d) The value of the mass of each weight (including the floating
elements when removable) of the pressure balance shall be
determined by a laboratory accredited for such mass
measurements. The relative uncertainty of the mass
determination should not usually exceed 20 % of the likely total
measurement uncertainty of the pressure balance to be
calibrated. For example, if the supposed expanded uncertainty
of the pressure balance is 5x10-5 x p, the relative uncertainty of
mass determination should be within l x10-5 x m.
65
(b) If the floatbase weight cannot be determined by weighing, the
corresponding base pressure may be determined from the results of the
pressure comparison measurements by using a least-squares analysis: in this
case a tare value in pressure units should be given. The p-method
mentioned letter allows the determination of this initial value.
Note (valid for both methods): Ascending measuring series can be considered to be
identical to descending measuring series, as the balances used for pressure
measurements usually have no significant hysteresis effect.
66
the influence of the clockwise/ anticlockwise rotation, and of the spin rate will
be checked.
(e) Note the reference number of each of the weights applied on both balances.
(f ) Note the temperature of the piston-cylinder assembly of both balances.
If the balance is not equipped with a temperature probe, note the surrounding
air temperature using an electronic thermometer attached to some suitable point of
the balance. This information shall be included in the certificate.
(a) When using a pressure balance as standard instrument, the cross floating
method cannot be used. In this case a differential pressure transducer
equipped with a by-pass is used to measure the difference between the
pressures measured by both balances. For each pressure point:
(b) Place the corresponding weights on both pressure balances.
(c) Adjust the pressure to equilibrate the standard balance.
(d) Read the zero of the transducer.
(e) Close the by-pass.
(f) Adjust the pressure on both sides to equilibrate both balances.
(g) Record the reading of the transducer. If the differential pressure is so high that
the needed uncertainty cannot be reached from the calibration of the
transducer, adjust the mass on the reference balance and repeat the three last
operations.
(h) Note the reference number of each of the weights applied on both balances.
(j) Note the temperature of the piston-cylinder assembly of both balances. If the
balance is not equipped with a temperature probe, note the surrounding air
temperature.
(k) Note the residual pressure in the belljar of both balances.
Usually the results will be presented for the standard value of the gravity
9,80665 m s-2 (unless customer requests his own local gravity) and the reference
temperature 20 °C in the form of a table. It will include:
67
(a) the pressure indicated by the balance under calibration (p m )
(b) the reference pressure measured by the standard instrument (mean of the three
determinations), in Pa and in the unit of the pressure delivered by the balance
if different (pr);
(c) the standard deviation of the reference pressure p r.
(d) the difference between the indicated pressure and the reference pressure (pm-
pr);
(e) the uncertainty of this difference, in the conditions of the calibration.
The method used to estimate this uncertainty shall be reported in the Certificate.
A table that lists all weights applied on the unit to be calibrated for each
pressure point of the calibration shall be included in the Calibration Certificate.
Method B procedure
(a) The computing method in more general use can be used to calculate step by
step the effective area of the pressure balance to be calibrated from the mass
applied on its piston and the pressure delivered by the standard instrument.
(b) From this method, the effective area is calculated by reversing the equation of
definition of the pressure presented before
(c) The use of other methods, such as differential method (p-method) to
eliminate potential zero-errors is not excluded, but requires some experience in
the analysis of the results. Particularly, the p-method may be the only one
available if method B is used for the determination of effective area of pressure
balances with an unknown initial weight that cannot be determined by weighing.
68
(d) The step-by-step determination of the effective area as a function of pressure
allows a modelling of effective area. In any case all the experimental data and
the residuals of the modelling shall be reported in the certificate to make
apparent the validity of the used method.
(e) The results may be presented in the form of a table, including:
(i) the reference pressure measured by the reference standard instrument in
each pressure point, in Pa and in the unit of the pressure delivered by the
balance if different;
(ii) the corresponding mass applied on the floating element of the balance to
be calibrated;
(iii) the corresponding temperature of the measuring assembly during the
calibration;
(iv) the individual value of the effective area Ap calculated at the reference
temperature and at reference pressure
(v) the mean value of the effective area Ap;
(vi) the experimental standard deviation of the mean.
(f) Then, the effective area as a function of pressure is analysed using a least-
squares method. Three cases can be observed:
(i) the dependence upon pressure is not significant relating to the standard
deviation (this is always the case for the low-range pressure balances). The
effective area at null pressure Ao is calculated as the mean value of all the
determinations. If the theoretical pressure distortion coefficient is known, it
shall be used for calculating the effective area. The type A standard
uncertainty is estimated from the experimental standard deviation of the
mean Ao.
(ii) the dependence upon pressure can be considered to be linear. The
effective area at null pressure Ao and the pressure distortion coefficient I
are calculated by analogy with the best least-squares straight line. The type
A combined standard uncertainty is estimated using the standard
uncertainties of Ao and .
(iii) the dependence upon pressure cannot be considered to be linear. The
effective area at null pressure Ao and the pressure distortion coefficient
(first order) and ’ (second order) are calculated by analogy with the least-
squares second-order best fit. The type A combined standard uncertainty is
estimated using the standard uncertainties of Ao, and ’.
(g) The standard uncertainties of each of the parameters shall be estimated using
literature on statistics.
(h) The certificate shall report:
(i) the calculated value of the effective area under reference conditions A o and
the corresponding uncertainty, estimated from the standard deviation of Ao,
and the contribution of the standard, the measurement of the mass applied
to the moving element and the temperature.
69
(ii) when relevant, the pressure distortion coefficient(s) and the corresponding
uncertainty estimated from the standard deviation of and the uncertainty
of the pressure distortion coefficient of the standard.
71