CHAPTER 1
INTRODUCTION
1.0 Introduction
A.A. Michelson invented in the 1880s a simple interferometer, which basically consists of
a beamsplitter and two plane mirror. Initially, the interferometer was used to study the
speed of light and to fix the standard the meter with the wavelength of a spectra line. The
use of the Michelson interferometer as a spectrometer was started more than half a
century later. In 1949, astrophysicist Peter Fellgett applied the interferometer to measure
the light from celestial bodies, and computed the first Fourier transform spectrum.
4
CHAPTER 2
LITERATURE REVIEW
One of the most basic tasks in spectroscopy is to characterize the spectrum of a light
source: How much light is emitted at each different wavelength. The most straightforward
way to measure a spectrum is to pass the light through a monochromator, an instrument
that blocks all of the light except the light at a certain wavelength (the un-blocked
7
wavelength is set by a knob on the monochromator). Then the intensity of this remaining
(single-wavelength) light is measured. The measured intensity directly indicates how
much light is emitted at that wavelength. By varying the monochromator's wavelength
setting, the full spectrum can be measured. This simple scheme in fact describes how
some spectrometers work.
Fourier transform spectroscopy is a less intuitive way to get the same information.
Rather than allowing only one wavelength at a time to pass through to the detector, this
technique lets through a beam containing many different wavelengths of light at once, and
measures the total beam intensity. Next, the beam is modified to contain a different
combination of wavelengths, giving a second data point. This process is repeated many
times. Afterwards, a computer takes all this data and works backwards to infer how much
light there is at each wavelength.
To be more specific, between the light source and the detector, there is a certain
configuration of mirrors that allows some wavelengths to pass through but blocks others
(due to wave interference). The beam is modified for each new data point by moving one
of the mirrors; this changes the set of wavelengths that can pass through.
As mentioned, computer processing is required to turn the raw data (light intensity
for each mirror position) into the desired result (light intensity for each wavelength). The
processing required turns out to be a common algorithm called the Fourier transform
(hence the name, "Fourier transform spectroscopy"). The raw data is sometimes called an
"interferogram".
8
Figure 2.1: The spectrum of light emitted by the blue flame of a butane torch. The
horizontal axis is the wavelength of light, and the vertical axis represents how much light
is emitted by the torch at that wavelength.
The method of Fourier transform spectroscopy can also be used for absorption
spectroscopy. The primary example is "FTIR Spectroscopy", a common technique in
chemistry. In general, the goal of absorption spectroscopy is to measure how well a
sample absorbs or transmits light at each different wavelength. However, any technique
for emission spectroscopy can also be used for absorption spectroscopy as follows:
Assume you have a working spectrometer that can measure the spectrum of any light
source shining into it (as described above). Shine a broadband light source into the
spectrometer, then shine the same light source through the sample into the same
spectrometer. Comparing the two spectra, it will be obvious which wavelengths get
absorbed by the sample and which wavelengths can pass right through the sample. (More
precisely, the spectrum with the sample, divided by the "background" spectrum without
the sample, equals the fraction of light that the sample can transmit at each wavelength.)
9
Figure 2.2: An interferogram from a Fourier transform spectrometer. The horizontal axis
is the position of the mirror, and the vertical axis is the amount of light detected. This is
the "raw data" which can be transformed into an actual spectrum
2.3 History
A.A. Michelson invented in the 1880s a simple interferometer, which basically consists of
a beamsplitter and two plane mirror. Initially, the interferometer was used to study the
speed of light and to fix the standard the meter with the wavelength of a spectra line. The
use of the Michelson interferometer as a spectrometer was started more than half a
century later. In 1949, astrophysicist Peter Fellgett applied the interferometer to measure
the light from celestial bodies, and computed the first Fourier transform spectrum.
10
Infrared spectroscopy (IR spectroscopy) is the subset of spectroscopy that deals with the
infrared region of the electromagnetic spectrum. It covers a range of techniques, the most
common being a form of absorption spectroscopy. As with all spectroscopic techniques, it
can be used to identify compounds and investigate sample composition. A common
laboratory instrument that uses this technique is an infrared spectrophotometer.
The infrared portion of the electromagnetic spectrum is usually divided into three
regions; the near-, mid- and far- infrared, named for their relation to the visible spectrum.
The far-infrared, approximately 400–10 cm−1 (1000–30 μm), lying adjacent to the
microwave region, has low energy and may be used for rotational spectroscopy. The mid-
infrared, approximately 4000–400 cm−1 (30–2.5 μm) may be used to study the
fundamental vibrations and associated rotational-vibrational structure. The higher energy
near-IR, approximately 14000–4000 cm−1 (2.5–0.8 μm) can excite overtone or harmonic
vibrations. The names and classifications of these subregions are merely conventions.
They are neither strict divisions nor based on exact molecular or electromagnetic
properties.
11
CHAPTER 3
METODOLOGY
3 Infrared spectroscopy
3.1 Theory
Infrared spectroscopy exploits the fact that molecules absorb specific frequencies that are
characteristic of their structure. These absorptions are resonant frequencies, i.e. the
frequency of the absorbed radiation matches the frequency of the bond or group that
12
vibrates. The energies are determined by the shape of the molecular potential energy
surfaces, the masses of the atoms, and the associated vibronic coupling.
In order for a vibrational mode in a molecule to be "IR active," it must be associated with
changes in the permanent dipole.
A molecule can vibrate in many ways, and each way is called a vibrational mode.
Linear molecules have 3N-5 degrees of vibrational modes whereas nonlinear molecules
have 3N-6 degrees of vibrational modes (also called vibrational degrees of freedom). As
an example H2O, a non-linear molecule, will have 3*3-6 = 3 degrees of vibrational
freedom, or modes.
Simple diatomic molecules have only one bond and only one vibrational band. If
the molecule is symmetrical, e.g. N2, the band is not observed in the IR spectrum, but
only in the Raman spectrum. Unsymmetrical diatomic molecules, e.g. CO, absorb in the
IR spectrum. More complex molecules have many bonds, and their vibrational spectra are
correspondingly more complex, i.e. big molecules have many peaks in their IR spectra.
13
The atoms in a CH2 group, commonly found in organic compounds can vibrate in
six different ways: symmetrical and antisymmetrical stretching, scissoring, rocking,
wagging and twisting:
The simplest and most important IR bands arise from the "normal modes," the simplest
distortions of the molecule. In some cases, "overtone bands" are observed. These bands
arise from the absorption of a photon that leads to a doubly excited vibrational state. Such
bands appear at approximately twice the energy of the normal mode. Some vibrations, so-
called 'combination modes," involve more than one normal mode. The phenomenon of
Fermi resonance can arise when two modes are similar in energy, Fermi resonance results
in an unexpected shift in energy and intensity of the bands.
The infrared spectrum of a sample is recorded by passing a beam of infrared light through
the sample. Examination of the transmitted light reveals how much energy was absorbed
at each wavelength. This can be done with a monochromatic beam, which changes in
wavelength over time, or by using a Fourier transform instrument to measure all
wavelengths at once. From this, a transmittance or absorbance spectrum can be produced,
showing at which IR wavelengths the sample absorbs. Analysis of these absorption
characteristics reveals details about the molecular structure of the sample. When the
frequency of the IR is the same as the vibrational frequency of a bond, absorption occurs.
This technique works almost exclusively on samples with covalent bonds. Simple
spectra are obtained from samples with few IR active bonds and high levels of purity.
14
More complex molecular structures lead to more absorption bands and more complex
spectra. The technique has been used for the characterization of very complex mixtures.
Gaseous samples require a sample cell with a long pathlength (typically 5–10 cm), to
compensate for the diluteness.
A beam of infrared light is produced and split into two separate beams. One is passed
through the sample, the other passed through a reference which is often the substance the
sample is dissolved in. The beams are both reflected back towards a detector, however
first they pass through a splitter which quickly alternates which of the two beams enters
the detector. The two signals are then compared and a printout is obtained.
A reference prevents fluctuations in the output of the source affecting the data. The
reference also allows the effects of the solvent to be cancelled out (the reference is
usually a pure form of the solvent the sample is in)
Fourier transform spectroscopy is in frequent use in many various practical and scientific
applications. Here, we shall review some of these applications.
3.6.1.1 Methods
3.6.1.2 Spectroscopy
As with the GCMS the FTIR spectrum can be compared to that of a known
sample, thus providing evidence for the identification of a compound. Spectroscopy can
also help to identify materials used in failed products, especially polymers, additives and
fillers. Samples can be taken by dissolution, or by cutting a thin slice using a microtome
from the specimen under examination. Surfaces can be examined using Attenuated total
reflectance spectroscopy, and the method has also been adapted to the optical microscope
with infra-red microspectroscopy.
18
Forensic chemists usually perform their analytical work in a sterile laboratory decreasing
the risk of sample contamination. In order to prevent tampering, forensic chemists must
keep track of a chain of custody for each sample. A chain of custody is a document which
stays with the evidence at all times. Among other information, contains signatures and
identification of all the people involved in transport, storage and analysis of the evidence.
This makes it much more difficult for intentional tampering to occur, it also acts
as a detailed record of the location of the evidence at all times for record keeping
purposes. It increases the reliability of a forensic chemist's work and increases the
strength of the evidence in court.
3.6.1.4 Examples
Polymers for example, can be attacked by aggressive chemicals, and if under load, then
cracks will grow by the mechanism of stress corrosion cracking. Perhaps the oldest
known example is the ozone cracking of rubbers, where traces of ozone in the atmosphere
attack double bonds in the chains of the materials. Elastomers with double bonds in their
chains include natural rubber, nitrile rubber and styrene-butadiene rubber. They are all
highly susceptible to ozone attack, and can cause problems like car fires (from rubber fuel
lines) and tire blow-outs. Nowadays, anti-ozonants are widely added to these polymers,
so the incidence of cracking has dropped. However, not all safety-critical rubber products
are protected, and since only ppb of ozone will start attack, failures are still occurring.
Another highly reactive gas is chlorine, which will attack susceptible polymers
such as acetal resin and polybutylene pipework. There have been many examples of such
pipes and acetal fittings failing in properties in the USA as a result of chlorine-induced
cracking. Essentially the gas attacks sensitive parts of the chain molecules (especially
secondary, tertiary or allylic carbon atoms), oxidizing the chains and ultimately causing
chain cleavage. The root cause is traces of chlorine in the water supply, added for its anti-
bacterial action, attack occurring even at parts per million traces of the dissolved gas.
Most step-growth polymers can suffer hydrolysis in the presence of water, often a
reaction catalysed by acid or alkali. Nylon for example, will degrade and crack rapidly if
exposed to strong acids, a phenomenon well known to those who accidentally spill acid
onto their shirts or tights. Polycarbonate is susceptible to alkali hydrolysis, the reaction
20
simply depolymerising the material. Polyesters are prone to degrade when treated with
strong acids, and in all these cases, care must be taken to dry the raw materials for
processing at high temperatures to prevent the problem occurring.
CHAPTER 4
CONCLUSIONS
4.0 Conclusion
commercial instrument with its maximum resolution. Then, looking at the spectrum with
the naked eye, or using some commercial search program based on the correlation
between the observed and a library spectrum, the user tries to identify the unknown
compound. The multicomponent analysis of gas mixture has usually aken place by the so
called subtraction technique: the library spectrum multiplied by a constant is subtracted
from the observed mixture spectrum. The constant, which gives the desired concentration
of the component, is determined in such a way that the corresponding component just
disappears after subtraction. This analysis method, however, requires high resolution (
0.1cm −1 − 0.5cm −1 ) and is slow and laborious to carry out. It requires an experienced
user, and it is very difficult to automatize.
An important scientific application of FTS is molecular spectroscpy. In this
application, infrared FT spectrometers are used to study geometric structures of
molecules, and the behavior of molecules in rotation and vibrations. IR spectra reveal
quantum mechanical phenomena and interactions in an effective way. In this type of
research, high resolution and accuracy are very important, and FT-IR spectrometry is the
most commonly used method.