Professional Documents
Culture Documents
Easytemp TMR31
Compact thermometer with M12 plug-in connector,
Pt100 or 4...20mA
www.uk.endress.com/e-direct
• PC programmable transmitter
with 4...20mA output or Pt100
• Highly accurate sensor and electronics
• M12 plug-in connector with IP66
• Compact thermometer from stainless steel
• Sensor length selectable
Application
The Easytemp TMR31 is used in tanks and pipes of -50...150°C
(200°C with extension neck). Due to its flexible adjustment to the
process, TMR31 is the solution for simple measuring tasks in all
industries.
Function
The compact thermometer consists of a complete sensor with
Pt100, class A and a housing with various process connections.
There are also other process connections available as accessories.
The built-in electronics can be programmed via a PC using the
M12 plug-in connector and convert the Pt100 input signal into
a temperature linear 4...20mA signal.
Dimensions (mm)
M12
Ø 18
Ø6
L
61*
Mounting location b)
M12 SW 27 G 1⁄2”
Ø 18
Pipe installation
Ø6
a) At angle sections,
against the direction
of flow 65* L
b) In smaller pipes,
c) L = selectable length from 30 to 300mm
turned against the
direction of flow * Version with extension neck add 35mm
c) Perpendicular to the
direction of flow Electrical connection
4...20mA Pt100
Measuring range red
3
red
Code Measuring range 4 0V
(4... 20 mA)
AA 4...20mA, 0...+100°C
BA 4...20mA, 0...+150°C
CA 4...20mA, -50...+100°C
DA 4...20mA, -50...+150°C
EB 4...20mA, -50...+200°C with extension neck white
XB 4...20mA, add. spec. with extension neck 1 2 white
10...35VDC
please add to order code (4...20mA)
TD 052/DE/en/08.05
LL/INDD 2.0
Technical data Power supply
Sensor Supply voltage Ub = 10...35V
Sensing element 1 x Pt100 Residual ripple Permitted residual ripple Uss ≤ 3 V at
Tolerance Class A to IEC 751 Ub ≥ 13V, fmax = 1kHz
Process temperature -50...+150°C (200°C with extension neck) Accuracy
Material 316L Response time 1s
Sensor diameter 6mm Reference operating
Sensor length 30...300mm selectable *1 conditions 0°C (Icebath)
Response time t50 = ≤ 2.0s; t90 = ≤ 4.0s Measuring error 0.1K or 0.08%
Max. Pressure 30 bar (at 100mm and 20°C, depending Influence of
on process connection) supply voltage ≤ ±0.01%/V deviation from 24V
Process connection Influence of ambient
Version G 1⁄2”, 316L; without ((flange welding boss temperature Td = ±(15ppm/K * (full scale value
or compression fitting needed) (temperature drift) + 200) + 50ppm/K * of set measuring
Transmitter housing range) * ∆ϑ
Degree of protection IP 66 ∆ϑ = deviation of ambient temperature
Electrical connection M12-plug-in connector from the reference operating condition
Material 304H Long term stability ≤ 0.1 K/year or 0.05%/year
Electronic/output values Influence of load ≤ ± 0.02%/100Ω
Output signal analogue 4...20mA, 20...4mA Environmental conditions
Signal on alarm - Downscale measuring range: Ambient temperature -40...+85°C
linear decrease to 3.8mA Storage temperature -40...+100°C
- Upscale measuring range: Climate class as per EN 60 654-1, Class C
linear increase to 20.5mA Condensation permitted
- Sensor break; Sensor short-circuit: Shock resistance 4 g/2...150Hz as per IEC 60 068-2-6
≤ 3.6mA or ≥ 21.0mA Electromagnetic
Load max. (Vpower supply -10 V )/0.022 A compatibility (EMC) Shock resistance and interference emission
Induced current as per EN 61 326-1 (IEC 1326) and
requirement ≤ 3.5mA NAMUR NE 21
Current limitation ≤ 23mA
Switch-on delay 2s