Flow measurement

Selection guide
How to use this selection guide
Selecting the measuring principle Example

➊ D
 efine all the primary requirements for your measuring • Fluid: drinking water (urban area)
point: fluid type, nominal diameter, accuracy, pressure,  • Pipe diameter: DN 1600 
temperature, etc. • Required accuracy: <0.3% o.r.
• Measured variable: Volume flow

➋ I n the table on Page 4, select the category for the


type of fluid in question: Liquid – Gas – Steam –
 • Liquids 
Special applications.

➌ I n the table, specify all the requirements (see Point ➊) • Liquids in general
you defined for your measuring point.

 • Volume flow
• Nominal diameter DN > 500

➍  ow select the measuring principles to which all the
N • Electromagnetic
requirements apply (•).
 • Ultrasonic
• Differential pressure

Selecting the flowmeter


➎ N
 ow open out this brochure completely. The overview • Electromagnetic
contains all the measuring principles you selected  • Ultrasonic
• Differential pressure

under Point ➍ and the corresponding Endress+Hauser
flowmeters.

➏ Now mark the sensors that can be installed in the pipe • Promag W
of your choice (DN = 1600 mm).  • Prosonic Flow W (clamp-on)
• Prosonic Flow W (insertion)

• Prosonic Flow P
• Prosonic Flow C
• Deltatop DP
• Deltatop DV

➐M
 ark those transmitters that can be combined with the • 50 (W)
sensors in question and that meet the required accuracy  • 53 (W) 
specifications (<0.3% o.r.)

➑ You have found the two flowmeters that suit your • Promag 50W (standard)
measuring point.  • Promag 53W (high-end) 

2
➋ ➌ ➍
Requirements Fluids (examples) Measuring principle

Differential
Ultrasonic
magnetic

pressure
Thermal
Electro-

Coriolis

Vortex
Liquids in general Water • • • • •
Flow rates < 2 l/h Additives, flavors • •
Non-conductive liquids Oil, hydrocarbons • • • •
Viscous liquids > 500 mPa�s Molasse, shower gel • •
Cryogenic fluids Liquid O2 , Ar, N2 • • •
Liquids

Volume flow Water, milk, solvents • • • • •


Mass flow Chemicals • • •
Accuracy ≤ 0.2% • •
Nominal pipe diameter DN > 500 • • •
Operable flow range > 1:30 • • •
Low pressure loss • • •
Density, concentration, viscosity metering Foods, shower gel •

➎ ➑ ➏ ➑ ➐ ➐
Diameter Combinable Accuracy
with (best possible accuracy)

Promag D
Water/Waste water
DN 25 to 100
(1" to 4")
10 Sensor
D, L, W, P, H
±0.5%
Basic

Promag L
Water/Waste water
DN 50 to 1200
(2" to 48")
50 Sensor
D, L, W, P, H
±0.5%
Optional: ±0.2%
Electromagnetic (EMF)

Standard
Standard

53
Proline Promag

Promag W
Promag W DN 25 to 2400 Sensor
±0.2%
Wasser/Abwasser
Water/Waste water (1" to 96")
High-End W, P, H
High-end

Promag P
Chemical/Process
DN 15 to 600
(½" to 24")
51
Custody
Sensor
W, P
±0.5%
Optional: ±0.2%
applications transfer

Promag H
Hygiene
DN 2 to 100
(1/12" to 4")
23 Sensor
P, H
±0.5%
Two-wire

Promag S
Inhomogeneous fluids
DN 15 to 600
(½" to 24")
55
Special
Sensor
S, H
±0.2%
High solid content applications

Prosonic Flow W
Clamp-on sensor
DN 15 to 4000
(½" to 156")
91 Sensor
W
Onsite: <2.0%
Under reference conditions: ±0.5%
Standard applications Basic

Prosonic Flow W DN 200 to 4000 93 Sensor Onsite: <2.0%


Proline Prosonic Flow

Insertion sensor (8" to 156") W, P, C Under reference conditions: ±0.5%


Standard applications High-end
Ultrasonic

Prosonic Flow P
Clamp-on sensor
DN 15 to 4000
(½" to 156")
92 F Sensor
F
Liquids
For 2 or 3 beams: ±0.5%
Process applications Inline For 4 beams: ±0.3%

Prosonic Flow C
Inline sensor
DN 300 to 2000
(12" to 78")
93 T
Portable
Sensor
P
Onsite: <2.0%
Under reference conditions: ±0.5%
Water/Waste water

Prosonic Flow F DN 25 to 300


Inline sensor (1" to 12")
Chemical industry

Orifice plate DO: ±0.5 to 0.8%


Deltatop DO DN 10 to 1000 Deltabar S
Deltatop
Pitot tube DP: ±1.0%
Orifice plate (3/8" to 40") PMD 70/75 Nozzle DN: ±0.7 to 1.0%
Differential pressure

Venturi tube DV: ±0.7 to 1.0%


Orifice plate DO: ±0.5 to 0.8%
Deltatop DP DN 40 to 12,000 Deltabar M
Deltatop
Pitot tube DP: ±1.0%
Pitot tube (1½" to 480") PMD 55 Nozzle DN: ±0.7 to 1.0%
Venturi tube DV: ±0.7 to 1.0%

Deltatop DN DN 50 to 500
Nozzle (2" to 20")

Deltatop DV DN 50 to 2000
Venturi tube (2" to 78")

3
Selection table “Measuring principles”
Requirements Fluids (examples) Measuring principle

Differential
Ultrasonic
magnetic

pressure
Thermal
Electro-

Coriolis

Vortex
Liquids in general Water • • • • •
Flow rates < 2 l/h Additives, flavors • •
Non-conductive liquids Oil, hydrocarbons • • • •
Viscous liquids > 500 mPa�s Molasse, shower gel • •
Cryogenic fluids Liquid O2 , Ar, N2 • • •
Liquids

Volume flow Water, milk, solvents • • • • •


Mass flow Chemicals • • •
Accuracy ≤ 0.2% • •
Nominal pipe diameter DN > 500 • • •
Operable flow range > 1:30 • • •
Low pressure loss • • •
Density, concentration, viscosity metering Foods, shower gel •
Process temperature > 200 °C • • •
Process pressure > 100 bar Crude oil • • • •
Gas flow in general Natural gas, air • • • •
Contaminated gas Biogas, exhaust gas • • •
Volume flow Exhaust gas • •
Mass flow CO2 • • • •
Gas

Accuracy ≤ 0.5% Rare gas •


Nominal pipe diameter DN > 300 • •
Operable flow range > 1:30 • •
Low pressure loss Air • •
Process temperature > 200 °C Exhaust gas • •
Process pressure > 100 bar CNG • • •
Volume flow • •
Mass flow • •
Steam

Nominal pipe diameter DN > 300 •


Operable flow range > 1:10 •
Low pressure loss •
Process temperature > 300 °C Superheated steam • •
Process pressure > 100 bar Superheated steam • •
Hygienic applications (SIP/CIP) Milk, beer • • •
Slurries, suspended solids Thick sludge • •
Liquid/liquid mixtures Oil/water, emulsions •
applications

Highly abrasive fluids (e.g. in mining) Ore slurry •


Special

Bidirectional metering (forward/reverse) • • • •


Measurement from outside without process
interruption/retrofitting/temporary metering

Filling/bottling applications
see Flow measurement OEM (inside)
Gas refueling applications

Flow monitoring see Flow monitoring (inside)

Please note:
The selection table above provides a general overview and is used for quick clarification of suitable measuring principles. The table is not exhaustive.
4 If further clarification is required, feel free to contact Endress+Hauser any time. We’ll be glad to be of assistance.
Supplementary documentation

Flow handbook Flow measurement Metering flow the cost-


A practical guide: Measurement For liquids, gases and steam optimised way
technologies – applications – Products and services at a glance Comprehensive metering point
solutions FA00005D/06/EN solutions for planning, commissio-
ISBN 3-9520220-4-7 (Order number 71024542) ning and maintenance
(Order number 50104766) CP006D/06/en
(Order number 50109047)

Flow measurement in utilities Precision for the real world Reduce energy costs
Energy management in your hands – Our calibration facilities at your Tailor-made energy monitoring –
customized solutions for compressed service rapid return on investment
air, gas, steam and water We calibrate your flowmeters on For steam, air, gas, oil, water,
the world’s most advanced rigs – electricity, heating and cooling
CP007D/06/en
in our factory or onsite
(Order number 71009419) CP014D/06/en
CP010D/06/en (Order number 71109850)
(Order number 71091218)

5
Applicator ® –
Optimal planning
and sizing
The first steps are always the most important when it comes
to planning metering points: An optimally dimensioned
flowmeter at the right location provides many advantages.
• Consistent process quality
• Product savings due to error-free, precise measurement
• Diminished risk of instrument failure at a later time

Proven over 20 years of use


Applicator is a software for selecting and specifing flow-
meters that has proven its abilities over many years. A wealth
of industry experience and specialized know-how is embed-
ded in Applicator. It enables people with limited technical
knowledge to specify metering points quickly and easily.
The most recent version of Applicator has a raft of new
features, including selection functions for all the other
areas covered by Endress+Hauser – namely level, pressure,
temperature, liquid analysis, etc. You’ll find that it opens up
an enormous range of application possibilities in metrology
and control.

Selection – straight to your objective


When starting out to select the ideal flowmeter, Appli-
cator guides you through all the important parameters.
Just enter the type of fluid, required process data or
metering point attributes and Applicator already displays
the measuring principles and families of meters best
suited to the case in hand.
The Selection Tool contains over 600 products from all
of the Endress+Hauser fields of operation.

Sizing – safe specification


By entering further application-specific values, such as
process pressure, process temperature or measuring
range, Applicator immediately determines the ideal
flowmeter type – and along with the suitable nomi-
nal diameter, working range, pressure loss, accuracy,
material loads, conformity with the Pressure Equipment
Directives (PED) and corrosion resistance.
Messages and warnings make it easy to select the right
meter, help avoid common mistakes, and enable optimi-
zation of the measuring point to suit specific require-
ments. Additional Sizing Tools exist for level, pressure
and energy measuring devices.

6
W@M W@M – Life cycle management from Endress+
Hauser is an information system which provides tech-
nical and operative management with complete data
to maintenance and service within a plant, at any time
or location. W@M is an open system based on Inter-
net/intranet technology, and incorporating all software
flow and archival – from commissioning on through programs, products and services from Endress+Hauser.

Fieldcheck®
• Simulation and
testing of alarm/
process conditions
directly in the plant
(without flow)
• Verification of flow- Applicator®
meters directly in • Optimal planning and
the piping (without sizing of flowmeters
removal) • Administration and
documentation for
entire plant projects

FieldCare®
• Easy and dependable commissioning and • Process-optimization through the visualization of
configuring of measuring devices via PC measurement data in realtime (line plotting function)
• Complete documentation of device settings • Analysis of diagnostic data for maintenance purposes
• Storing device data (upload/download) • Administration and analysis of verification results
• Simplified trouble-shooting from the “Fieldcheck” test instrument

Project documentation – data


collection, archival, export
The use of additional expansion mod-
ules permits cost-saving administration
and documentation of data for entire
plant projects:
• Clear presentation in a tree structure
of any desired number of metering
points (e.g. by building, floor, meter-
ing point), according to ISA 88.
• Uncomplicated inclusion of measur-
ing devices directly from the Selection
and Sizing functions. Also possible for
instruments from other manufactur-
ers or projects.
• Detailed listing of project manage-
ment data (responsible person, cost-
ing and budget data, etc.)
• Multiple data exchange possibilities
in Excel or XML format: data export
to other programs, data import from
other projects, etc.

The Applicator newsletter, which


can be subscribed to online, informs
about all the latest updates:
www.endress.com >> Corporate
Websites >> Global website >> Device
support: Select and size products

7
Flow measurement Sensors Transmitters
Diameter Measuring range Pressure PN Process Materials Process connection Protection Hygienic Combinable Housings Display/Operation Ambient Power supply Galvanic isolation Inputs Outputs Communication Accuracy Ex-approvals Protection
temperature (wetted parts) certification with temperature (best possible accuracy)

Promag D
Water/Waste water
DN 25 to 100
(1" to 4")
0 to 10 m/s
0 to 600 m³/h
PN 16
Cl 150 0 to +60 °C Polyamide liner Wafer: DIN, ANSI, JIS
IP67
(NEMA 4X)
– 10 Sensor
D, L, W, P, H
Compact Alu
Remote Alu
Two-line LCD/
push buttons –20 to +60 °C
85 to 250 V AC (45 to 65 Hz)
20 to 28 V AC (45 to 65 Hz) or 11 to 40 V DC
Electronical connections are
galvanically isolated from each other
– 4 to 20 mA, pulse/status HART® ±0.5% FM, CSA Div. 2
IP67
(Nema 4X)
JIS 10 K Basic Operation via FieldCare

Promag L
Water/Waste water
DN 50 to 1200
(2" to 48")
0 to 10 m/s
0 to 40,000 m³/h
PN 6 to 16
Cl 150
–20 to +90 °C
Polyurethane liner
PTFE liner
Lap-joint flanges (DN < 350): DIN, ANSI
Welded flanges (DN > 300): DIN, ANSI,
IP67/IP68
(NEMA 4X/6P)
– 50 Sensor
D, L, W, P, H
Compact SS
Compact Alu
Two-line, backlit LCD/
push buttons
–20 to +60 °C
Optional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz)
20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are
galvanically isolated from each other
Status
0/4 to 20 mA, pulse/frequency
(1 kHz passive), status
HART®, PROFIBUS DP,
PROFIBUS PA
±0.5%
Optional: ±0.2%
ATEX, IEC, FM, CSA, TIIS, NEPSI
IP67
(Nema 4X)
Electromagnetic (EMF)

Hard rubber liner AWWA, AS Standard Remote Alu Operation via FieldCare
PN 6 to 40, Cl 150 to 300 HART®, PROFIBUS DP,
53 Compact SS Four-line, backlit LCD/
Proline Promag

Promag W DN 25 to 2400 0 to 10 m/s JIS 10 to 20 K


–20 to +80 °C
Polyurethane liner
Flange: DIN, ANSI, JIS, AWWA, AS
IP67/IP68

Sensor
Compact Alu touch control
–20 to +60 °C 85 to 260 V AC (45 to 65 Hz) Electronical connections are Status, 0/4 to 20 mA, pulse/frequency PROFIBUS PA, FOUNDATION
±0.2% ATEX, IEC, FM, CSA, TIIS, NEPSI
IP67
Water/Waste water (1" to 96") 0 to 163,000 m³/h AWWA Class D Hard rubber liner (NEMA 4X/6P) W, P, H Optional: –40 to +60 °C 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC galvanically isolated from each other current input (10 kHz, active/passive), relays, status Fieldbus, MODBUS® RS485, (Nema 4X)
High-end Remote Alu Operation via FieldCare
AS Table E, PN 16 EtherNet/IP

51
PN 10 to 40
Promag P DN 15 to 600 0 to 10 m/s Cl 150 to 300 PTFE liner IP67/IP68 Sensor Compact Alu
Two-line, backlit LCD/
85 to 260 V AC (45 to 65 Hz) Electronical connections are 0/4 to 20 mA, pulse/frequency ±0.5% IP67
Chemical/Process –40 to +180 °C Flange: DIN, ANSI, JIS, AS – Custody push buttons –20 to +60 °C Status HART®  ATEX, IEC
(½" to 24") 0 to 9600 m³/h JIS 10 to 20 K PFA liner (NEMA 4X/6P) W, P Remote Alu 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC galvanically isolated from each other (1 kHz passive), status Optional: ±0.2% (Nema 4X)
applications transfer Operation via FieldCare
AS Table E, PN 16
Flange: DIN, ANSI, JIS, Two-wire
Promag H
Hygiene
DN 2 to 100
(1/12" to 4")
0 to 10 m/s
0 to 282 m³/h
PN 16 to 40
Cl 150 –20 to +150 °C PFA liner
external thread, internal thread,
weld connection, clamp connection,
IP67
(NEMA 4X)
3-A, EHEDG 23 Sensor
P, H
Compact Alu
Four-line LCD/
touch control –20 to +60 °C
12 to 30 V DC
13.9 to 30 V DC (Ex)
Electronical connections are
galvanically isolated from each other

4 to 20 mA,
pulse/frequency, status
HART® ±0.5% ATEX, FM, CSA
IP67
(Nema 4X)
JIS 10 to 20 K Two-wire Operation via FieldCare
couplings (45 to 65 Hz)

55
PN 10 to 40
Promag S DN 15 to 600 0 to 10 m/s Cl 150 to 300
Polyurethane, PTFE, PFA,
IP67/IP68 Sensor
Compact SS Four-line, backlit LCD/
–20 to +50 °C Electronical connections are Status, 0/4 to 20 mA, pulse/frequency
HART®, PROFIBUS DP, FM, CSA Div. 2
IP67
Inhomogeneous fluids –40 to +180 °C natural rubber liner, Flange: DIN, ANSI, JIS, AS – Special Compact Alu touch control 20 to 260 V AC/DC (45 to 65 Hz) PROFIBUS PA, ±0.2% In preparation: ATEX II 3G,
(½" to 24") 0 to 9600 m³/h JIS 10 to 20 K (NEMA 4X/6P) S, H Optional: –40 to +50 °C galvanically isolated from each other current input (10 kHz, active/passive), relays, status (Nema 4X)
High solid content hard rubber liner applications Remote Alu Operation via FieldCare FOUNDATION Fieldbus IEC Zone 2
AS Table E, PN 16

Promass F
Universal flowmeter
DN 8 to 250
(3/8" to 10") 0 to 2,200,000 kg/h
PN 16 to 100
Cl 150 to 600
–50 to +200 °C
–50 to +350 °C (HT)
904L (1.4539)/
316L (1.4404)
Flange: DIN, ANSI, JIS,
Tri-Clamp, couplings, VCO
IP67
(NEMA 4X)
3-A, EHEDG 40 Sensor
E
Compact Alu
Two-line, backlit LCD/
Operation via HART® resp.
–20 to +60 °C
Optional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz)
20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are
galvanically isolated from each other
Status 0/4 to 20 mA, frequency, status HART®
Liquids: ±0.5%
Gases: ±1.0%
ATEX, IEC, FM, CSA, TIIS, NEPSI
IP67
(Nema 4X)
DN 25, 50, 80 (HT) JIS 10 to 63 K Alloy C-22 (2.4602) Basic FieldCare
Liquids: ±0.15%
Promass A
Low flow rates
DN 1 to 4
(1/24" to 5/32")
0 to 450 kg/h
PN 16 to 400
Cl 150 to 600 –50 to +200 °C
904L (1.4539)
Alloy C-22 (2.4602)
Tri-Clamp, VCO
IP67
(NEMA 4X)
3-A, EHEDG 80 Sensor
F, A, I,
Compact SS
Compact Alu
Two-line, backlit LCD/
push buttons
–20 to +60 °C
Optional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz)
20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are
galvanically isolated from each other
Status
0/4 to 20 mA, pulse/frequency
(1 kHz passive), status
HART , PROFIBUS PA
® Optional: ±0.1%
ATEX, IEC, FM, CSA, TIIS, NEPSI
IP67
(Nema 4X)
JIS 10 to 63 K Standard H, E, S, P Remote Alu Operation via FieldCare
Gases: ±0.35%
HART®, PROFIBUS DP, Liquids: ±0.1%
Promass I
Hygiene, viscosity
DN 8 to 80
(3/8" to 3")
0 to 180,000 kg/h
PN 40 to 100
Cl 150 to 600 –50 to +150 °C
Measuring tube: Titanium grade 9
Flange: Titanium grade 2
Flange: DIN, ANSI, JIS,
Tri-Clamp, couplings
IP67
(NEMA 4X)
3-A, EHEDG 83 Sensor
F, A, I,
Compact SS
Compact Alu
Four-line, backlit LCD/
touch control
–20 to +60 °C
Optional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz)
20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are
galvanically isolated from each other
Status,
current input
0/4 to 20 mA, pulse/frequency
(10 kHz, active/passive), relays, status
PROFIBUS PA, FOUNDATION
Fieldbus, MODBUS® RS485,
Optional: ±0.05%
ATEX, IEC, FM, CSA, TIIS, NEPSI
IP67
(Nema 4X)
JIS 10 to 63 K High-end H, E, S, P Remote Alu Operation via FieldCare
Proline Promass

EtherNet/IP Gases: ±0.35 %

84
Coriolis

Promass H DN 8 to 50
PN 40
–50 to +200 °C (Zr) Zirconium 702/R 60702 IP67 Sensor
Compact SS Four-line, backlit LCD/
–20 to +60 °C 85 to 260 V AC (45 to 65 Hz) Electronical connections are
0/4 to 20 mA, pulse/frequency
HART®  Liquids: ±0.1% IP67
Chemically aggressive 0 to 70,000 kg/h Cl 150 to 300 Flange: DIN, ANSI, JIS – Custody Compact Alu touch control Status (10 kHz, active/passive), ATEX, IEC, FM, CSA, TIIS, NEPSI
(3/8" to 2") –50 to +150 °C (Ta) Tantalum 2.5W (NEMA 4X) F, A Optional: –40 to +60 °C 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC galvanically isolated from each other MODBUS® RS485 Gases: ±0.35 % (Nema 4X)
fluids JIS 10 to 20 K transfer Remote Alu Operation via FieldCare 2 × pulse (90°/180°), relays, status

Promass E
Basic applications
DN 8 to 80
(3/8" to 3")
0 to 180,000 kg/h
PN 40 to 100
Cl 150 to 600 –40 to +140 °C
904L (1.4539)
316L (1.4404)
Flange: DIN, ANSI, JIS,
Tri-Clamp, couplings, VCO
IP67
(NEMA 4X)
– 200 Sensor
E
Compact SS
Compact Alu
Four-line LCD/
push buttons –40 to +60 °C
Two-wire
18 to 30 V DC
Electronical connections are
galvanically isolated from each other
Status
4 to 20 mA,
pulse/frequency, status
HART®, PROFIBUS PA
Liquids: ±0.25%
Gases: ±0.75 %
ATEX, IEC, cCSAus
IP66/IP67
(NEMA 4X)
JIS 10 to 63 K Two-wire Operation via FieldCare

PN 40 to 63
Promass S DN 8 to 50
0 to 70,000 kg/h Cl 150 to 300 –50 to +150 °C
904L (1.4539) Flange: DIN, ANSI, JIS, IP67
3-A, EHEDG
Food & Beverages (3/8" to 2") 316L (1.4435) Tri-Clamp, couplings (NEMA 4X)
JIS 10 to 40 K

Promass P DN 8 to 50
PN 40 to 63 Flange: DIN, ANSI, JIS,
IP67 3-A, EHEDG,
Life Sciences Industries 0 to 70,000 kg/h Cl 150 to 300 –50 to +200 °C 316L (1.4435) Hygienic clamps, couplings and
(3/8" to 2") (NEMA 4X) ASME BPE
(Pharma, biotechnology) JIS 20 to 40 K flanges

Prosonic Flow W
Clamp-on sensor
DN 15 to 4000
(½" to 156")
0 to 15 m/s
0 to 678,550 m³/h
No limit
Non-invasive
–20 to +80 °C
Optional:
not wetted
(Clamp-on sensor)
Clamp-on sensor
IP67/IP68
(NEMA 4X/6P)
– 91 Sensor
W
Remote Alu
Two-line LCD/
push buttons –20 to +60 °C
85 to 250 V AC (45 to 65 Hz)
20 to 28 V AC (45 to 65 Hz) or 11 to 40 V DC
Electronical connections are
galvanically isolated from each other
– 0/4 to 20 mA, pulse/status HART®
Onsite: <2.0%
Under reference conditions: ±0.5%
FM, CSA Div. 2
IP67
(Nema 4X)
Standard applications 0 to 130 °C Basic Operation via FieldCare

Prosonic Flow W DN 200 to 4000 0 to 15 m/s IP68 93 Sensor


Four-line, backlit LCD/
–20 to +60 °C 85 to 260 V AC (45 to 65 Hz) Electronical connections are Status,
0/4 to 20 mA, pulse/frequency HART®, PROFIBUS DP,
Onsite: <2.0% IP67
Proline Prosonic Flow

Insertion sensor PN 16 –20 to +80 °C 316L (1.4404) Welding connector – Remote Alu touch control (10 kHz, active/passive), PROFIBUS PA, ATEX, IEC, FM, CSA, TIIS, NEPSI
(8" to 156") 0 to 678,550 m³/h (NEMA 6P) W, P, C Optional: –40 to +60 °C 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC galvanically isolated from each other current input Under reference conditions: ±0.5% (Nema 4X)
Standard applications High-end Operation via FieldCare relays, status FOUNDATION Fieldbus
Ultrasonic

Prosonic Flow P
Clamp-on sensor
DN 15 to 4000
(½" to 156")
0 to 15 m/s
0 to 678,550 m³/h
No limit
Non-invasive
–40 to +170 °C
not wetted
(Clamp-on sensor)
Clamp-on sensor
IP67/IP68
(NEMA 4X/6P)
– 92 F Sensor
F
Compact Alu
Remote Alu
Two-line LCD/
push buttons
Compact: –40 to +60 °C
Remote: –40 to +80 °C
12 to 35 V DC,
EEx i: 12 to 30 V DC,
Electronical connections are
galvanically isolated from each other
– 4 to 20 mA, pulse/status
HART®, PFM, Profibus PA,
FOUNDATION Fieldbus
Liquids
For 2 or 3 beams: ±0.5%
ATEX, IEC, FM, CSA, TIIS
In preparation: NEPSI
IP67
(Nema 4X)
Process applications Inline Operation via FieldCare EEx d: 15 to 35 V DC For 4 beams: ±0.3%

Prosonic Flow C
Inline sensor
DN 300 to 2000
(12" to 78")
0 to 10 m/s
0 to 110,000 m³/h
PN 16
CI 150
–20 to +60 °C Carbon steel, epoxy coated Flange: DIN, ANSI, AWWA
IP68
(NEMA 6P)
– 93 T
Portable
Sensor
P
Portable Plastic
Four-line, backlit LCD/
touch control 0 to +60 °C
Rechargeable battery + charger: 100 to 240 V AC
(47 to 63 Hz)
Electronical connections are
galvanically isolated from each other
4 to 20 mA Data logger –
Onsite: <2.0%
Under reference conditions: ±0.5%
– –
Water/Waste water Operation via FieldCare

Prosonic Flow F DN 25 to 300 0 to 10 m/s PN 40


–40 to +150 °C
CF3M/316/316L IP67/IP68
Inline sensor Optional: Flange: DIN, ANSI, JIS –
(1" to 12") 0 to 2400 m³/h CI 300 A105/A106/316/316L (NEMA 4X/6P)
Chemical industry –40 to +200 °C
Meter body: CF3M (1.4404), Two-wire
PN 10 to 250 –200 to +400 °C Flange: DIN, ANSI, JIS
72 Two-line LCD/
Proline Prowirl

Prowirl F DN 15 to 300 Liquids: 0.16 to 2360 m³/h


Cl 150 to 1500 Optional:
Alloy C-22 (2.4602)
Optional: with integrated reduction
IP67

Sensor Compact Alu
push buttons
Compact: –40 to +70 °C 12 to 36 V DC, Electronical connections are

4 to 20 mA, HART®, PFM, Profibus PA, Liquids: ±0.75% Volume
ATEX, IEC, FM, CSA, TIIS, NEPSI
IP67
Flanged version (½" to 12") Gases: 3 to 19,700 m³/h DSC sensor: 316L (1.4435), (NEMA 4X) F, W Remote Alu Remote: –40 to +80 °C EEx i: 12 to 30 V DC, galvanically isolated from each other pulse/status Foundation Fieldbus Gases/steam: ±1.0% Volume (Nema 4X)
Vortex

JIS 10 to 40 K –200 to +450 °C of line size Standard Operation via FieldCare


Inconel 718, Alloy C-22 (2.4602) EEx d: 15 to 36 V DC

73
Two-wire
PN 10 to 40 –200 to +400 °C Two-line LCD/ Pressure, temperature, 4 to 20 mA, Liquids: ±0.75% Volume, ±1% Mass
Prowirl W DN 15 to 150 Liquids: 0.19 to 625 m³/h
Cl 150 to 300 Optional:
Meter body: CF3M (1.4404)
Wafer: DIN, ANSI, JIS
IP67

Sensor Compact Alu
push buttons
Compact: –40 to +70 °C 12 to 36 V DC, Electronical connections are
density using HART®, pulse/status,
HART®, PFM, Profibus PA,
Gases/steam: ±1.0% Volume ATEX, IEC, FM, CSA, TIIS, NEPSI
IP67
Wafer version (½" to 6") Gases: 4 to 5210 m³/h DSC sensor: 316L (1.4435) (NEMA 4X) With flow F, W Remote Alu Remote: –40 to +80 °C EEx i: 12 to 30 V DC, galvanically isolated from each other Foundation Fieldbus (Nema 4X)
JIS 10 to 20 K –200 to +450 °C computer Operation via FieldCare PROFIBUS PA, FF frequency Gases/steam: ±1.7% Mass
EEx d: 15 to 36 V DC
HART®, Profibus DP,
t-mass I 65 Two-line, backlit LCD/ 0/4 to 20 mA, pulse/frequency Gases:
Proline t-mass

DN 80 to 1500 290 psi g 316L (1.4404) Sensor Compact Alu –20 to +60 °C 85 to 260 V AC (45 to 65 Hz) Electronical connections are PROFIBUS PA,
Gases: 0 to 120 m/s –40 to +130 °C Thread: NPT, G1A IP67 – push buttons 4 to 20 mA (1 kHz, active/passive), ±1.5% o.r. (10% to 100% o.f.s.) ATEX, IEC, FM, CSA, TIIS, NEPSI IP67
Thermal

Insertion sensor (3" to 60") 20 bar rel. Alloy C-22 I, F Remote Alu Optional: –40 to +60 °C 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC galvanically isolated from each other MODBUS® RS485,
Standard Operation via FieldCare relays, status ±0.15% o.f.s. (1% to 10% o.f.s.)
Foundation Fieldbus

t-mass F DN 15 to 100
Gases: 0 to 120 m/s
PN 16 to 40
–40 to +100 °C
316L (1.4404)
Flange: DIN, ANSI, JIS IP67 –
Flanged version (½" to 4") Cl 150 to 300 Alloy C-22

Compact SS Orifice plate DO: ±0.5 to 0.8%


Liquids: 0 to 12 m/s C22.8 Wafer (corner tapping)
Deltatop DO DN 10 to 1000
Gases: 0 to 65 m/s
PN 6 to 420
–200 to +1000 °C 316Ti (1.4571) Flange (flange tapping)
IP68

Deltabar S
Deltatop
Remote SS Four-line LCD/push buttons
–40 to +85 °C
Two-wire
No galvanic isolation – 4 to 20 mA
HART®, PROFIBUS PA, Pitot tube DP: ±1.0%
ATEX, IEC, FM, CSA, TIIS, NEPSI IP65
Orifice plate (3/8" to 40")
Steam: 0 to 65 m/s
Cl 150 to 2500
15Mo3 (1.5415) Pipe unit (meter run)
(NEMA 6P) PMD 70/75 Compact Alu Operation via FieldCare 11.5 to 45 V DC FOUNDATION Fieldbus Nozzle DN: ±0.7 to 1.0%
Differential pressure

Remote Alu Venturi tube DV: ±0.7 to 1.0%


Orifice plate DO: ±0.5 to 0.8%
Liquids: 0 to 12 m/s Four-line LCD/push buttons
Deltatop DP DN 40 to 12,000
Gases: 0 to 65 m/s
PN 6 to 420
–200 to +1000 °C
C22.8 Cutting ring, IP68

Deltabar M
Deltatop
Compact Alu
DIP switch –40 to +85 °C
Two-wire
No galvanic isolation – 4 to 20 mA
HART®, PROFIBUS PA, Pitot tube DP: ±1.0%
ATEX, IEC, FM, CSA, TIIS, NEPSI IP65
Pitot tube (1½" to 480")
Steam: 0 to 65 m/s
Cl 150 to 2500 316Ti (1.4571) flange, flow tap (NEMA 6P) PMD 55 Remote Alu
Operation via FieldCare
11.5 to 45 V DC FOUNDATION Fieldbus Nozzle DN: ±0.7 to 1.0%
Venturi tube DV: ±0.7 to 1.0%
e.g. C22.8
Liquids: 0 to 12 m/s Flange: DIN, ANSI, JIS
Deltatop DN DN 50 to 500
Gases: 0 to 65 m/s
PN 6 to 420
–200 to +1000 °C
316Ti (1.4571)
Wafer: DIN, ANSI, JIS
IP68

Nozzle (2" to 20") Cl 150 to 2500 15Mo3 (1.5415) (NEMA 6P)
Steam: 0 to 65 m/s Butt-weld version
13CrMo45
e.g. C22.8
Liquids: 0 to 12 m/s Flange: DIN, ANSI, JIS
Deltatop DV DN 50 to 2000
Gases: 0 to 65 m/s
PN 6 to 420
–200 to +1000 °C
316Ti (1.4571)
Wafer: DIN, ANSI, JIS
IP68

Venturi tube (2" to 78") Cl 150 to 2500 15Mo3 (1.5415) (NEMA 6P)
Steam: 0 to 65 m/s Butt-weld version
13CrMo45

Flow measurement OEM


Dosimag DN 4, DN 8,
EMF

0 to 10 m/s IP67 Electronical connections are ±0.5% IP67


Filling/bottling DN 15K, DN 15
0 to 1.66 l/s
PN 16 –40 to +130 °C PFA liner Tri-Clamp, weld connection
(NEMA 4X)
3-A, EHEDG Compact version Compact SS Operation via FieldCare –20 to +60 °C 20 to 30 V DC
galvanically isolated from each other
– Pulse, status –
Optional: ±0.25%

(NEMA 4X)
Conductive liquids (5/32", 3/8", ½")

Dosimass DN 8, DN 15,
0 to 5 kg/s 904L (1.4539) IP67 Electronical connections are IP67
Filling/bottling DN 25 PN 40 –40 to +125 °C
316L (1.4404)
Tri-Clamp, couplings
(NEMA 4X)
3-A Compact version Compact SS Operation via FieldCare –20 to +60 °C 20 to 30 V DC
galvanically isolated from each other
– Pulse, status – ±0.15% ATEX II 3G
(NEMA 4X)
Liquids (3/8", ½", 1")
Internal thread:
CNGmass DN 8, DN 15,
316L (1.4404) G½" (DN 8) IP67 Compact version 20 to 28 V AC Electronical connections are ATEX, IEC, FM, CSA IP67
Fueling DN 25 0 to 150 kg/min max. 350 bar –50 to +125 °C – Compact Alu Operation via FieldCare –40 to +60 °C – Pulse/frequency, status MODBUS® RS485 ±0.5%
316L (1.4435) G¾" (DN 15) (NEMA 4X) Custody transfer 20 to 30 V DC galvanically isolated from each other In preparation: TIIS, NEPSI (NEMA 4X)
Compressed Natural Gas (3/8", ½", 1")
G1" (DN 25)
Internal thread:
CNGmass DCI DN 8, DN 15,
316L (1.4404) G½" (DN 8) IP67 Compact/remote version Compact SS Four-line, backlit LCD/
85 to 260 V AC (45 to 65 Hz) Electronical connections are
0/4 to 20 mA, pulse/frequency
ATEX, CEC, NEC IP67
Fueling DN 25 0 to 150 kg/min max. 350 bar –50 to +150 °C – Compact Alu touch control –40 to +60 °C Status (10 kHz, active/passive), HART®, MODBUS® RS485 ±0,5%
316L (1.4435) G¾" (DN 15) (NEMA 4X) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC galvanically isolated from each other In preparation: NEPSI (NEMA 4X)
Coriolis

(3/8", ½", 1") Custody transfer Remote Alu Operation via FieldCare 2 × pulse (90°/180°), relays, status
Compressed Natural Gas G1" (DN 25)

LPGmass DN 8, DN 15, PN 40
316L (1.4404) Flange: DIN, ANSI, JIS, IP67 Compact version 20 to 28 V AC Electronical connections are ATEX, IEC, FM, CSA IP67
Fueling DN 25, DN 40 0 to 70,000 kg/h Cl 150 to 300 –40 to +125 °C – Compact Alu Operation via FieldCare –40 to +60 °C – Pulse/frequency, status MODBUS® RS485 ±0.2%
904L (1.4539) VCO (NEMA 4X) Custody transfer 20 to 30 V DC galvanically isolated from each other In preparation: TIIS, NEPSI (NEMA 4X)
Liquified Petroleum Gas (3/8", ½", 1", 1½") JIS 10 to 63 K

DN 1, DN 2,
Cubemass DN 4, DN 6 0 to 1000 kg/h 160/400 bar –50 to +200 °C 904L (1.4539) VCO
IP67
– Compact version Compact Alu Operation via FieldCare –40 to +60 °C
20 to 28 V AC Electronical connections are
– 4 to 20 mA, pulse/frequency MODBUS® RS485 ±0,1%
ATEX, IEC, CEC, NEC IP67
Low flow rates (NEMA 4X) 10 to 30 V DC galvanically isolated from each other In preparation: TIIS, NEPSI (NEMA 4X)
(1/24", 1/12", 5/32", ¼")

DN 1, DN 2, Remote SS Two-line, backlit LCD/


Cubemass DCI DN 4, DN 6 0 to 1000 kg/h 160/400 bar –50 to +200 °C 904L (1.4539) VCO
IP67
– Compact/remote version Compact Alu touch control –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) Electronical connections are
Status 4 to 20 mA, pulse/frequency MODBUS® RS485 ±0,1%
ATEX, IEC, CEC, NEC IP67
Low flow rates (NEMA 4X) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC galvanically isolated from each other In preparation: TIIS, NEPSI (NEMA 4X)
(1/24", 1/12", 5/32", ¼") Remote Alu Operation via FieldCare

Flow monitoring
Magphant
EMF

DN 1 to 2000 IP66 IP66


Limit switch/Trend display (½" to 78")
0 to 5 m/s PN 16 –20 to +120 °C PVDF Weld nipple
(NEMA 4X)
– Compact version Compact Alu LED –20 to +60 °C 20 to 30 V DC No galvanic isolation – 4 to 20 mA, relays – ±2.0% FM, CSA Div. 2
(NEMA 4X)
Conductive liquids

Flowphant DTT31 DN 20 to 1000


0.03 to 3 m/s
Weld nipple, compression fitting,
One-line, backlit LCD –
18 to 30 V DC
1× or 2 × PNP switching outputs,
Programmable by a PC
Standard applications (3/4" to 40")
as a relative value between max. PN 100 –20 to +85 °C AISI 316L
threaded connection: ANSI, ISO
IP65 – Compact version Compact SS Color change on faults, LED’s for –40 to +85 °C
(reverse polarity protection)
No galvanic isolation – optional: 1× PNP switching output
(ReadWin® 2000 and FieldCare)
±2.0% – IP65
Liquids 0 and 100% status indication, push buttons and 1× 4 to 20 mA output, active

Flowphant DTT35 DN 20 to 1000


0.03 to 3 m/s
AISI 316L and FKM Viton, Conical metal-metal G½",
One-line, backlit LCD –
18 to 30 V DC
1× or 2 × PNP switching outputs,
Programmable by a PC
Hygienic applications as a relative value between max. PN 100 –20 to +85 °C IP65 3-A Compact version Compact SS Color change on faults, LED’s for –40 to +85 °C No galvanic isolation – optional: 1× PNP switching output ±2.0% – IP65
Thermal

(3/4" to 40") EPDM (sealings) Tri-Clamp (reverse polarity protection) (ReadWin® 2000 and FieldCare)
Liquids 0 and 100% status indication, push buttons and 1× 4 to 20 mA output, active

t-switch ATT11 DN 40 to 1000 Liquids: 0 to 3 m/s Threaded connection: BSP, NPT, IP66 Compact SS LED, 4-digits, LCD (option) –10 to +65 °C

Subject to modification
Limit switch (1½" to 40") Gases: 0 to 50 Nm/s
PN 25 –10 to +80 °C 316L (1.4404/1.4435)
couplings, Tri-Clamp (NEMA 4X)
3-A, EHEDG Compact version Compact Alu DIL switches without LCD
18 to 30 V AC/DC No galvanic isolation – Relays – ±5.0% – IP66
Liquids, gases

t-trend ATT12 DN 40 to 1000 Liquids: 0 to 3 m/s Threaded connection: BSP, NPT, IP66 Compact SS LED, 4-digits, LCD (option) –10 to +65 °C
Trend display (1½" to 40") Gases: 0 to 50 Nm/s
PN 25 –10 to +80 °C 316L (1.4404/1.4435)
couplings, Tri-Clamp (NEMA 4X)
– Compact version Compact Alu push buttons without LCD
18 to 30 V AC/DC No galvanic isolation – 4 to 20 mA, open collector – ±5.0% – IP66
Liquids, gases
Instruments International

Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
Tel. +41 61 715 81 00
Fax +41 61 715 25 00
http://www.endress.com
info@ii.endress.com

CP00001D/06/EN/13.11