Operational Principles for Cone-Beam

Computed Tomography
David C. Hatcher
J Am Dent Assoc 2010;141;3S-6S

The following resources related to this article are available online at


jada.ada.org ( this information is current as of October 19, 2010 ):

Updated information and services including high-resolution figures, can be found


in the online version of this article at:
http://jada.ada.org/cgi/content/full/141/suppl_3/3S

Downloaded from jada.ada.org on October 19, 2010


Information about obtaining reprints of this article or about permission to reproduce
this article in whole or in part can be found at:
http://www.ada.org/prof/resources/pubs/jada/permissions.asp

© 2010 American Dental Association. The sponsor and its products are not endorsed by the ADA.
Operational principles for cone-beam
computed tomography
David C. Hatcher, DDS, MSc, MRCD(c)

any of the diagnostic chal-

M lenges that routinely arise in


clinical practices can benefit
from imaging investigations.
Until recently, dentists were
ABSTRACT
Background. Cone-beam computed tomography (CBCT)
was introduced into the U.S. market in 2001. Today, there are

Downloaded from jada.ada.org on October 19, 2010


limited to using two-dimensional radio- more than 3,000 installed units in the United States. There are
graphs to assess three-dimensional (3-D) numerous CBCT manufacturers and types of units. To produce
anatomy. Cone-beam computed tomography the best imaging results, clinicians need to be knowledgeable
(CBCT) was introduced into the European about the CBCT unit, the clinical issue being investigated and
market in 1998. In May 2001, QR (Verona, how to optimize the unit’s operational parameters. The author
Italy) introduced CBCT, a 3-D digital identifies the variables that should be considered for each
imaging technology, into the U.S. market. It imaging session and addresses the building blocks required to
has been estimated that more than 3,000 design the appropriate imaging strategy. The remaining arti-
CBCT units consisting of more than 30 dif- cles in this supplement address imaging for orthodontics, the
ferent machines have been installed in the investigation and localization of impacted teeth and implant
United States (oral communication, Chris planning, and customized imaging protocols designed to solve
Scharff, vice president of sales, Imaging the clinical issues being presented.
Sciences International, Hatfield, Pa., Methods. The author addresses CBCT from an operational
July 2010). point of view. An ideal imaging examination answers the clin-
During a CBCT scan, the scanner (x-ray ical question while maintaining an acceptable radiation dose
source and a rigidly coupled sensor) rotates, and cost. The quality and value of each imaging study is pro-
usually 360 degrees, around the head to portional to the protocol being used. The author also addresses
obtain multiple images (ranging from imaging protocol variables (raw data frames, scan time, voxel
approximately 150 to 599 unique radio- size, field of view and milliampere settings) and their effects on
graphic views). The scanning software col- the final image quality and radiation dose, as well as CBCT
lects the raw image data and reconstructs accuracy and the radiation dose.
them into viewable formats. The scan time Results. CBCT can provide image volumes of the maxillofa-
can range between five and 40 seconds, cial region and can be useful in clinical dentistry.
depending on the unit and protocol setting. Conclusion. CBCT has been shown to be a precise imaging
The x-ray source emits a low milliampere modality and is a valuable tool for use in dental applications.
and a shaped or divergent beam. The beam Clinical Implications. CBCT can be used for diagnosis
size is constrained by a circular or rectan- and treatment planning for all of the dental specialties.
gular collimator to match the sensor size, Key Words. Cone-beam computed tomography; cone-beam
but in some cases it can be constrained (colli- imaging; radiation risk.
mated) further to match the anatomical JADA 2010;141(10 suppl):3S-6S.
region of interest. After the scan, the resul-
tant image set or raw data are subjected to a
reconstruction process that results in the
production of a digital volume (a cylindrical Dr. Hatcher is a clinical professor, College of Dental Medicine, University of Southern
or spherical shape that is composed of Nevada, Henderson, and an adjunct associate clinical professor, Arthur A. Dugoni School
of Dentistry, University of the Pacific, San Francisco, and he maintains a private practice
volume elements called “voxels” that are in oral and maxillofacial radiology. Address reprint requests to Dr. Hatcher at 99 Scripps
stacked in rows and columns) of anatomical Drive, Suite 101, Sacramento, Calif. 95825, e-mail “David@ddicenters.com”.

JADA 141(10 suppl) http://jada.ada.org October 2010 3S


Copyright © 2010 American Dental Association. All rights reserved. Reprinted by permission.
A C
Figure. Visualization options for a cone-beam computed tomography volume. A. Visualization of the full data volume by means of a
shaded surface display method with thresholds set to show the soft tissues. B. Visualization of the full data volume by means of a shaded
surface display with the threshold set to show hard tissues (bone and teeth) only. C. The volume rendering method. The data attenuation

Downloaded from jada.ada.org on October 19, 2010


values corresponding to the soft tissues were made partially transparent, allowing for visualization of the underlying skeleton and teeth.

data that can be visualized with specialized soft- using CBCT for investigations in orthodontics,
ware. Voxels are the smallest subunit of a digital impactions and implants. The authors of these
volume. CBCT voxels generally are isotropic (that articles discuss the coordinated effort required by
is, X, Y and Z dimensions are equal) and range in clinicians in the investigative process to achieve
size from approximately 0.07 to 0.40 millimeters successful imaging outcomes that meet the clin-
per side. Each voxel is assigned a gray-scale value ical objectives. The authors also define clinical
that approximates the attenuation value of the rep- objectives and associated imaging goals, imaging
resented tissue or space. The latest generation of protocols, diagnostic portfolios and value proposi-
CBCT units produces 12- or 14-bit images in which tion. When possible, the authors attempt to blend
12 bits is 212 (4,096) shades of gray and 14 bits is 214 clinical applications using CBCT technology with
(16,384) shades of gray. Computer monitors used to the appropriate scholarly validation.
visualize the 12- or 14-bit digital or voxel volume CBCT imaging sessions are based on a process
can display only eight bits (256 shades) of gray at a designed to answer specific clinical questions.
time. The software uses a technique called “win- This design process results in the development of
dowing and leveling” that allows the operator to imaging goals, a customized or case-specific
access and visualize all of the data. Windowing imaging protocol and an imaging portfolio. It may
allows the data to be scrolled through, thus visual- require input from several people, including the
izing eight bits at a time with air and soft tissues clinician ordering the scan, the radiology technol-
(low-attenuation structures) at one end of the spec- ogist and the reviewing radiologist. An ideal
trum and bone and teeth (high-attenuation struc- imaging session provides answers to the clinical
tures) at the other end of the spectrum. Once the question via a precise display of relevant anatom-
optimum window level has been achieved, the con- ical information that can aid the clinician, the
trast and brightness (leveling) are adjusted by the radiologist and the patient. An illustration of the
clinician for optimal viewing. The small isotropic levels of decision making in the imaging process
voxel size along with the large number of gray are provided as online supplemental data to this
levels have contributed to accuracy and precision article (available at “http://jada.ada.org”).
when clinicians measure the dimensions of anatom-
ical structures and visualize anatomical form.1-4
ABBREVIATION KEY. CBCT: Cone beam computed
The diagnostic value of an imaging study needs tomography. E1990: Effective dose calculations using 1990
to be balanced against the risk to the patient and tissue-weighting factors. E2007: Effective dose calculations
the cost of service. The trend for CBCT use has using 2007 tissue-weighting factors. FOV: Field of view.
been to apply it clinically and then validate its FOVMand: 6-centimeter field of view, mandible. FOVMax:
utility scientifically. CBCT technology has several 6-cm field of view, maxilla. PSP: Photostimulable phos-
beneficial applications in the clinical setting. The phor. TLDs: Thermoluminescent dosimeters. 3-D:
articles in this supplement address the benefits of Three-dimensional.

4S JADA 141(10 suppl) http://jada.ada.org October 2010


Copyright © 2010 American Dental Association. All rights reserved. Reprinted by permission.
Imaging protocol variables include field TABLE 1
of view (FOV), voxel size, scan time, Effective dose for i-CAT Classic* cone-beam
mA setting and patient immobilization.
Clinicians select imaging protocol vari- computed tomography.†‡
ables to investigate a patient’s condi- CONE-BEAM E1990§ E2007¶
tion on the basis of their knowledge COMPUTED
Standard High Resolution Standard High Resolution
TOMOGRAPHY
about image sensor responses or con-
22cm FOV 92.8 µSv# NA** 206.2 µSv NA
version of the analog signal to a digital
signal. There are many options clini- 13cm FOV 39.5 µSv NA 133.9 µSv NA

cians can use to visualize the digital 6cm FOVMand†† 23.9 µSv 47.2 µSv 96.2 µSv 118.5 µSv
volume. The diagnostic portfolio 6cm FOVMax‡‡ 9.7 µSv 18.5 µSv 58.9 µSv 93.3 µSv
includes the visualization options clini- * i-CAT Classic is manufactured by Imaging Sciences International (Hatfield, Pa.).
cians can use to provide a customized † Source: Roberts and colleagues.6
‡ The data shown in this table are calculations of the effective dose using the 1990 and
image set that is relevant to the clinical 2007 International Commission on Radiological Protection tissue-weighting factors for
issue being addressed. the i-CAT Classic cone-beam computed tomography machine with a flat panel sensor
(amorphous silicon)7 using a field of view (FOV) of 22, 13 and 6 centimeters and time set-
tings of 20 and 40 seconds. This unit captures angular images (raw data) at 15 images

Downloaded from jada.ada.org on October 19, 2010


IMAGE QUALITY per second. A 20-second rotation around the head (standard resolution) creates 300
images. High-resolution scans are produced by increasing scan time to 40 seconds at 15
Images acquired by CBCT are con- images per second for 599 or 600 images. The 22-cm scan requires two 20-second rota-
verted to a voxel volume and stored tions. The effective dose for Next Generation i-CAT (Imaging Sciences International) in
landscape mode (standard resolution) was 87 microsieverts.8
digitally on computers. The image § E1990: Effective dose calculations using 1990 tissue-weighting factors.
quality or feature detection ability can ¶ E2007: Effective dose calculations using 2007 tissue-weighting factors.
# µSv: Microsieverts.
be related to physical characteristics of ** NA: Not available.
digital images represented by Nyquist- †† FOVMand: 6-centimeter field of view, mandible.
‡‡ FOVMax: 6-cm field of view, maxilla.
Shannon sampling theorem, volume
averaging, dynamic range, modulation TABLE 2
transfer function and signal-to-noise
ratio. The effects of these variables are
Effective dose for two-dimensional
provided as online supplemental data techniques.*†
to this article (available at “http:// TWO- ROUND COLLIMATION RECTANGULAR
DIMENSIONAL COLLIMATION
jada.ada.org”). IMAGING
Data visualization. The recon- PSP/ D-Speed Film PSP/
F-Speed Film F-Speed Film
structed volumes can be viewed using
Complete Series of 170.7 µSv‡ 388.0 µSv 34.9 µSv
specialized software. The smallest sub- Radiographs
unit in a CBCT volume is the voxel. The * Source: Ludlow and colleagues. 9

voxels created from a CBCT generally † The data shown in this table are calculations of the effective dose using the 2007
International Commission on Radiological Protection tissue-weighting factors for common
are isotropic. The voxels are stacked in two-dimensional maxillofacial intraoral and extraoral techniques. For intraoral imaging,
rows and columns. Information (for the sensors were F- and D-speed film and photostimulable phosphor (PSP) storage using
round or rectangular collimation.
example, dimension, 3-D location and ‡ µSv: Microsieverts.
value) for each voxel is stored in the
computer. supplemental data to this article (available at
The voxel volume can be retrieved and viewed “http://jada.ada.org”). Shaded surface display is a
by using a range of viewing options. Visualization software technique that allows the user to set a
options include multiplanar or orthogonal (that is, threshold range for the data on the basis of an
coronal, axial, sagittal) viewing angles. The data attenuation value. The data with an attenuation
can be sliced into a single row or column of voxels. value outside the selected range will not be visible.
Multiple voxel layers can be combined to create a The shaded surface display creates a 3-D object of
slab. Clinicians can produce and visualize oblique the anatomy that can be visualized from any
and curved slices or slabs, and they can render and selected angle. It is common to use shaded surface
visualize the entire volume from any angle. display technique to visualize soft tissue or bone
There are several techniques for visualizing a surfaces. Volume rendering is a method that uses
volume, including shaded surface display and all of the voxels but allows the operator to assign
volume rendering (Figure). Illustrations of addi- transparency values to voxels on the basis of their
tional visualization options are provided as online attenuation values. For example, if the superficial

JADA 141(10 suppl) http://jada.ada.org October 2010 5S


Copyright © 2010 American Dental Association. All rights reserved. Reprinted by permission.
TABLE 3 to the area of interest can optimize the effective
Effective dose for two- dose. Having shorter scan times, reducing the mA
setting or both can reduce the dose, but doing so
dimensional extraoral imaging.* also can decrease the signal and therefore image
TWO-DIMENSIONAL EFFECTIVE DOSE quality (Tables 1,6-8 29 and 39).
IMAGING
Panoramic (Charge- 14.2-24.3 µSv† CONCLUSIONS
Coupled Device)
Diagnostic imaging investigations are processes
Cephalometric 5.1-5.6 µSv
(Photostimulable that begin with designing an imaging protocol to
Phosphor) address specific individual clinical goals.
* Source: Ludlow and colleagues.9 Selecting the optimum technique begins by deter-
† µSv: Microsieverts.
mining the imaging goals. Clinicians need to
determine precisely what information needs to be
soft tissues were assigned a 70 percent trans- revealed during the imaging study; this will allow
parency value, then the underlying skeleton could them to determine what imaging modalities can
be visualized through the soft tissues. All CBCT fulfill the imaging goals. The optimum imaging

Downloaded from jada.ada.org on October 19, 2010


units include viewing software, but third-party modality fulfills the imaging goals, has the lowest
software also is available for general viewing or radiation dose and has an acceptable cost.
specialized applications, such as implant planning The articles in this supplement outline the
and assessment for orthodontic treatment. utility of CBCT for specific clinical investigations.
Dose. Radiation dose studies allow clinicians to The authors discuss specific clinical objectives,
assess risk and use a calculation method that imaging goals, imaging protocols and appropriate
helps clinicians compare risk between disparate image portfolios. ■
imaging devices. The dose calculations are
Disclosure. Dr. Hatcher did not report any disclosures.
expressed as an effective dose in microsieverts.5,6
Dose studies generally use a dosimetry phantom 1. Honey OB, Scarfe WC, Hilgers M, et al. Accuracy of cone-beam
computed tomography imaging of the temporomandibular joint: com-
(human skull encased in a soft tissue–equivalent parisons with panoramic radiology and linear tomography. Am J
material). The phantoms are sliced into multiple Orthod Dentofacial Orthop 2007;132(4):429-438.
2. Ludlow JB, Gubler M, Cevidanes L, Mol A. Precision of cephalo-
layers along the axial plane to allow access to metric landmark identification: cone-beam computed tomography vs
internal anatomy. For each study session, cali- conventional cephalometric views. Am J Orthod Dentofacial Orthop
2009;136(3):312.e1-e10.
brated thermoluminescent dosimeters (TLDs) are 3. Lagravere MO, Gordon JM, Guedes IH, et al. Reliability of tradi-
placed on the radiosensitive anatomy to be tested tional cephalometric landmarks as seen in three-dimensional analysis
in maxillary expansion treatments. Angle Orthod 2009;79(6):
(for example, ramus, thyroid gland, salivary 1047-1056.
gland, bone marrow). The phantoms are imaged 4. Chung RR, Lagravere MO, Flores-Mir C, Heo G, Carey JP, Major
PW. A comparative analysis of angular cephalometric values between
with selected variables including unit type, FOV, CBCT generated lateral cephalograms versus digitized conventional
scan time, mA setting and voxel size. To obtain lateral cephalograms. Int Orthod 2009;7(4):308-321.
5. Yamashina A, Tanimoto K, Sutthiprapaporn P, Hayakawa Y. The
the effective dose, the absorbed dose from the reliability of computed tomography (CT) values and dimensional mea-
TLDs is weighted by the 1990 or 2007 Interna- surements of the oropharyngeal region using cone beam CT: com-
parison with multidetector CT. Dentomaxillofac Radiol 2008;37(5):
tional Commission on Radiological Protection 245-251.
tissue-weighting factors. There are distinct effec- 6. Roberts JA, Drage NA, Davies J, Thomas DW. Effective dose from
cone beam CT examinations in dentistry. Br J Radiol 2009;82(973):
tive dose variations among CBCT units, which can 35-40.
be attributed to factors including FOV, mA set- 7. Baba R, Ueda K, Okabe M. Using a flat-panel detector in high res-
olution cone beam CT for dental imaging. Dentomaxillofac Radiol
ting, kilovolt (peak), scan time (including pulsed 2004;33(5):285-290.
versus continuous dose), sensor sensitivity and the 8. Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT
devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg
number of image captures. Oral Med Oral Pathol Oral Radiol Endod 2008;106(1):106-114.
The operator can control the FOV, the mA set- 9. Ludlow JB, Davies-Ludlow LE, White SC. Patient risk related to
common dental radiographic examinations: the impact of 2007 Interna-
ting and the scan time settings, which relate tional Commission on Radiological Protection recommendations
directly to the effective dose. Matching the FOV regarding dose calculation. JADA 2008;139(9):1237-1243.

6S JADA 141(10 suppl) http://jada.ada.org October 2010


Copyright © 2010 American Dental Association. All rights reserved. Reprinted by permission.