Professional Documents
Culture Documents
Presented By:
Mohammad Umar Rehman
Electrical Engineering Department
Aligarh Muslim University, Aligarh
Outline
• Introduction
• Historical Review
• The Basic Problem of Tomography
• CT Setup
• CT Projections
P j i
• Tomographic Images
• I
Image Re-construction:
R t ti The
Th Radon
R d Transform
T f
• Projection Slice Theorem
• CT Generations
G ti
• Pros & Cons
• Conclusion & Future Prospects
EE-655/Tomography/Umar 2
Introduction
• Tomography is derived from the Greek word “Tomos”
meaning a slice or a cutting
• Tomography
T h refers
f to t the
th cross-sectional
ti l imaging
i i off an object
bj t
from either transmission or reflection data collected by
illuminating the object from many different directions.
• Part of a larger field called Modern Medical Imaging.
• Popularly known as Computed (Axial) Tomography
(CT/CAT) & uses X-rays
X as radiation
di i source
• Mathematical foundations by Johann Radon, Austria, 1917
• Rediscovered by Allan Cormack,1963,
Cormack 1963 USA
• First Scanning Machine by an Engineer, Godfrey Hounsfield,
1971, UK. Nobel Prize in Medicine 1979 to Cormack &
Hounsfield
EE-655/Tomography/Umar 3
Historical Review
Imaging Modality Year Inventor Wavelength Physical principle
Energy
X-Ray 1895 Röntgen Measures variable tissue
(Nobel 1901) 3-100 keV absorption of X-Rays
EE-655/Tomography/Umar 5
CT Setup
EE-655/Tomography/Umar 6
CT Setup
Setup…contd
contd
X-Ray Source
Linear Translation
Parallel Beam
Detector Rotation
EE-655/Tomography/Umar 7
A Real One
Control Panels When to breathe indicator
Gantry
Bed (moves in
& out of gantry)
EE-655/Tomography/Umar 8
CT Projections
• A fine
fi X-ray
X b
beam i usedd to irradiate
is i di the h slice
li off interest.
i
• Once the measurement is made, the beam is moved over and
another measurement is collected until entire slice of interest
has been exposed.
• This set of measurements taken at the numerous beam
positions, is known as a projection
• It is desirable to collect as many samples per projection as
possible However,
possible. However in practical systems,
systems there are limitations
on sampling due to:
i. The physical size of the X-ray detectors,
ii. Desire to minimize radiation exposure to the patient, and
iii. The desire to minimize the amount of data that must be
collected and manipulated.
manipulated
EE-655/Tomography/Umar 9
Tomographic Images
• The tomographic
g p image
g is a p picture of a slab of the
patient’s anatomy
• The 2D CT image corresponds to a 3D section of the
patient
• CT slice thickness is very thin (1 to 10 mm) and is
approximately
pp y uniform
• The 2D array of pixels in the CT image corresponds to an
equal number of 3D voxels (volume elements) in the
patient
• Each pixel on the CT image displays the average X-ray
attenuation pproperties
p of the tissue in the corresponding
p g
voxel
EE-655/Tomography/Umar 10
Fig. Depiction of how a slice of an object can be viewed
as a sett off voxels
l
EE-655/Tomography/Umar 11
Image Re-construction:
Th Radon
The R d T Transform
f
• Let the slice of the image be represented by a 2D function
μ (x,
( y)) then
th its
it Radon
R d Transform
T f i given
is i b
by:
EE-655/Tomography/Umar 12
Radon Transform …contd
contd
EE-655/Tomography/Umar 14
The Radon Transform
Transform…contd
contd
• In most signal processing applications, a signal is
measured (typically a voltage or current), then some type
of transform is applied to abstract the desired useful
information.
information
• In CT, the Radon Transform of an image is measured.
• Our goal is to obtain the original function μ (x,
(x y) .
• The field of CT falls under the broad mathematical
category of inverse problems.
• Next we define the Inverse Radon Transform
EE-655/Tomography/Umar 15
The Inverse Radon Transform
∂g(s,θ )
1 π ∞
∂s
μ(x, y) = ℜ−1{g(s,θ )} = 2 ∫ ∫ dsdθ
2π 0 −∞x cosθ + y sinθ − s
• But inverse Radon Transform is valid only when g(s, θ) is
known for all possible s and θ .
• i.e. infinite projections…impossible
• In a typical medical imaging system sampled version of the
Radon Transform is measured.
• Here comes into pplayy that is known as Filtered Back
Projection (FBP)
• Widely adopted technique for image reconstruction
EE-655/Tomography/Umar 16
Projection Slice Theorem
• Based on Fourier Transform
• 1D FT of the projection of an object is same as the
2D FT of the object along the line drawn through the
centre
t off the
th 2D FT plane
l
• Thus FBP algorithm allows the image μ(x, y) to be
recovered from its sampled Radon Transform given
by:
EE-655/Tomography/Umar 17
Projection Slice Theorem…contd
Theorem contd
• Multiplication of projection Pθ(ω) with the ramp filter
|ω|
• Multiplication in frequency domain is equivalent to
convolution in time domain ÆFiltering
• Integration of the filtered projection w.r.t. θ ÆBack
Projection
• Implementation issue: what is the best range of values
of ω?
• Practical implementation by “windowing” |ω|
EE-655/Tomography/Umar 18
X-Ray
X Ray
Imaging
Inverse Filter
Back
Project
EE-655/Tomography/Umar 19
CT:A comparison
1971 2000
• Original axial CT image from the dedicated Axial CT image of a
Siretom CT scanner. Ability to see the soft normal brain usingg a state-
tissue structures of the brain, including the of-the-art CT system.
black ventricles for the first time.
• 80×80
80 80 pixel • 512×512 pixel
• 0.35 sec acquisition time
• 1-4 hours acquisition time
• 1 sec computation
• 1-5 days computation
EE-655/Tomography/Umar 20
CT Generations
EE-655/Tomography/Umar 21
First Generation (1972)
• Rotate/translate,, ppencil beam
• Only two X-ray detectors used (two different slices)
• Parallel ray geometry
• Translated linearly to acquire 160 rays across a 24 cm FOV
• Rotated slightly between translations to acquire 180
projections
j ti att 1-degree
1d iintervals
t l
• About 4.5 minutes/scan with 1.5 minutes to reconstruct slice
• Pencil beam geometry allowed very efficient scatter reduction,
reduction
best of all scanner generations
EE-655/Tomography/Umar 22
EE-655/Tomography/Umar 23
Second Generation (1975)
( )
• rotate/translate, narrow fan beam
• Incorporated linear array of 30 detectors
• More data acquired to improve image quality
(600 rays × 540 views)
• Shortest scan time was 18 seconds/slice
• Narrow fan beam allows more scattered radiation to be
detected
Source
Object
Detector
EE-655/Tomography/Umar 24
Third Generation (1976)
• Rotate/rotate, wide fan beam
• Number of detectors increased substantially (to more than 800
d t t )
detectors)
• Angle of fan beam increased to cover entire patient
– Eliminated need for translational motion
• Mechanically joined X-ray tube and detector array rotate
together
• Newer systems have scan times of ½ second
EE-655/Tomography/Umar 25
Fourth Generation (1977)
• Rotate/stationary
• Designed to overcome the problem of ring artifacts
• Stationary ring of about 4,800
4 800 detectors
EE-655/Tomography/Umar 26
Fifth generation (1980s)
• Stationary/stationary
• Developed specifically for cardiac tomographic
imaging
• No conventional X-ray tube; large arc of tungsten
encircles patient and lies directly opposite to the
detector ring
• Electron beam steered around the patient to strike the
annular tungsten target
• Capable of 50-msec scan times; can produce fast-
frame-rate CT movies of the beating heart
EE-655/Tomography/Umar 27
EE-655/Tomography/Umar 28
Sixth Generation (1990)
• Spiral/helical
p CT scanners acquire
q data while the
table is moving
• By avoiding the time required to translate the patient
table,
bl theh totall scan time
i required
i d to image
i the
h patient
i
can be much shorter
• Allows the use of less contrast agent and increases
patient throughput
• In so
somee instances
s ces thee eentiree sc
scan be do
donee w
within a
single breath-hold of the patient
EE-655/Tomography/Umar 29
EE-655/Tomography/Umar 30
Seventh(& last!) Generation (2000)
• Multiple detector array & cone beam
• When using multiple detector arrays, the collimator
spacing
p g is wider and more of the X-rays y that are
produced by the tube are used in producing image
data
– Opening up the collimator in a single array scanner
increases the slice thickness, reducing spatial resolution in
the slice thickness dimension
– With multiple detector array scanners, slice thickness is
determined by detector size, not by the collimator
EE-655/Tomography/Umar 31
EE-655/Tomography/Umar 32
Pros & Cons
• CT scanning is painless, noninvasive and accurate.
• A major advantage
ad antage of CT is its ability
abilit to image bone,
bone
soft tissue and blood vessels all at the same time.
• Unlike conventional X-rays,
X rays CT scanning provides
very detailed images of many types of tissue as well
ass thee lungs,
u gs, bo
bones,
es, andd bblood
ood vesse
vessels.
s.
• CT examinations are fast and simple; in emergency
cases,, theyy can reveal internal injuries
j and bleedingg
quickly enough to help save lives.
• No radiation remains in a patient's body after a CT
examination.
EE-655/Tomography/Umar 33
Pros & Cons…contd
Cons contd
• There is always a slight chance of cancer from excessive
exposure to radiation.
di i However, theh benefit
b fi off an accurate
diagnosis far outweighs the risk.
• The effective radiation dose for this procedure varies.
varies
• The risk of serious allergic reaction to contrast materials
that contain iodine is extremely rare, and radiology
departments are well-equipped to deal with them.
• Finite resolution causes some blurring of material
boundaries
• Large data volumes (GB) can require considerable
computer
t resources for
f visualization
i li ti and d analysis
l i
EE-655/Tomography/Umar 34
Conclusion & Future Prospects
• Successful, modern medical imaging is possible thanks to
high-speed digital computers and efficient, signal processing
algorithms.
• The introduction of the first CT scanner in 1972 revolutionized
medical imaging and has improved contemporary health care.
care
• The emerging localized tomography promises to provide
solutions to problems containing only a partial set of
projection data. Localized tomography will also help in cases
where only a portion of the object is of interest.
• Efforts are being made to further reduce the ionizing radiation
dose
• Development of image reconstruction algorithms is an open
area of research
EE-655/Tomography/Umar 35
References
• Many!!!
• To name a few…
few
1. Webb, S., “The Physics of Medical Imaging”, Institute of
Physics
y Publishing:
g Bristol, U.K., 1988.
2. Smith, B. J. & R. R. Adhami, “Medical Imaging:
Computerized Tomography”, IEEE Potentials Magazine,
Dec1998/Jan 1999
1999,Vol:17,
Vol:17 Issue:5.
Issue:5
3. Parra, Lucas C., Lecture notes on Biomedical Engineering,
Cityy College
g of New York.
4. www.radiologyinfo.org
EE-655/Tomography/Umar 36
The Presentation
• Could have been better if
• I had described little bit more mathematics
• Shown you some more figures
• Discussed more on CT instrumentation
• But still…
still
EE-655/Tomography/Umar 37
Thank You! for a “patient”
Listening
EE-655/Tomography/Umar 38
EE-655/Tomography/Umar 39