IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO.

5, OCTOBER 2009

2925

Measurements of Charge Transfer Inefciency in a CCD With High-Speed Column Parallel Readout
Andr Sopczak, Member, IEEE, Khaled Bekhouche, Chris Damerell, Tim Greenshaw, Michal Koziel, Konstantin Stefanov, Tuomo Tikkanen, Tim Woolliscroft, and Steve Worm

AbstractCharge Coupled Devices (CCDs) have been successfully used in several high energy physics experiments over the past two decades. Their high spatial resolution and thin sensitive layers make them an excellent tool for studying short-lived particles. The Linear Collider Flavour Identication (LCFI) collaboration is developing Column-Parallel CCDs (CPCCDs) for the vertex detector of a future Linear Collider. The CPCCDs can be read out many times faster than standard CCDs, signicantly increasing their operating speed. A test stand for measuring the charge transfer inefciency (CTI) of a prototype CPCCD has been set up. Studies of the CTI have been performed at a range of readout frequencies and operating temperatures. Index TermsCharge traps, data analysis, detectors, particle detectors, particle tracking, radiation damage, radiation effects in devices, semiconductor device radiation effects, signal analysis.

I. INTRODUCTION HE study of radiation hardness is crucial for the application of CCD detectors in High Energy Physics experiments [1][4]. The LCFI collaboration has been developing and testing new CCD detectors for about 10 years [1][5]. Previous experimental results on CCD radiation hardness were reported for example in [6][10]. Several models increased the understanding of radiation damage effects in CCDs [11][13]. The measurements and analyses reported in this paper have been carried out in the LCFI collaboration [3]. Simulation and modeling of CCD radiation hardness effects for a CCD prototype with sequential readout was reported at IEEE2005 [14]; comparing full TCAD simulations with analytic models was reported at IEEE2006 [15] and in [16]; simulation and modeling of a CCD prototype with column parallel readout (CPCCD) was reported at IEEE2007 [17] and in [18]. This work focuses on experimental measurements and a method to determine the charge transfer inefciency (CTI) performed with the CPCCD at a test stand at Liverpool University.

Fig. 1. Schematic view of how radiation damage is created in phosphorusdoped silicon. Two main types of vacancy defect are created: Vacancy-Oxygen (electron trap at 0.17 eV below the conduction band) and Vacancy-Phosphorus (electron trap at 0.44 eV below the conduction band).

Manuscript received November 15, 2008; revised March 19, 2009, and May 29, 2009. Current version published October 07, 2009. Presented on behalf of the LCFI Collaboration. A. Sopczak, K. Bekhouche, and M. Koziel are with Lancaster University, Lancaster, U.K. (e-mail: andre.sopczak@cern.ch; khaledbekhouche@yahoo. com; michal.koziel@cern.ch). C. Damerell, K. Stefanov, and S. Worm are with the STFC Rutherford Appleton Laboratory, Rutherford, U.K. (e-mail: c.damerell@rl.ac.uk; k.d.stefanov@rl.ac.uk; steven.worm@cern.ch). T. Greenshaw and T. Woolliscroft are with Liverpool University, Liverpool, U.K. (e-mail: green@liv.ac.uk; woolliscroft@gmail.com). T. Tikkanen is with Liverpool University, Liverpool, U.K., and also with Leicester University, Leicester, U.K. (e-mail: tvt1@star.le.ac.uk). Color versions of one or more of the gures in this paper are available online at http://ieeexplore.ieee.org. Digital Object Identier 10.1109/TNS.2009.2026648

The environment with high radiation near the interaction point at a future Linear Collider creates damage to the CCD material which leads to defects acting as electron traps in the silicon, as indicated in Fig. 1. The radiation level at a e/cm per year and Linear Collider is estimated to be n/cm per year at the inner vertex detector layer (14 mm radius) [16], [19], [20]. The mechanism of creating traps has been discussed in the literature, for example in [21][23]. These traps result in charge transfer inefciency. In a phosphorus-doped device, two main types of traps are created. The rst one is relatively shallow with energy level 0.17 eV below the conduction band and the second is deep with energy level 0.44 eV below the conduction band. The column parallel technology is in development to cope with the required readout rate. CPC1 is a two-phase CCD prototype capable of 50 MHz readout frequency. In this paper we demonstrate the method to determine the CTI value with an un-irradiated CPCCD (CPC1). II. TEST STAND FOR CCD OPERATION AND CRYOSTAT UNIT A test stand has been set up with readout electronics and a cryostat unit. The temperature range of the cryostat is from room C. This temperature has been temperature down to about achieved with cold nitrogen gas by boiling liquid nitrogen. The

0018-9499/$26.00 2009 IEEE


Authorized licensed use limited to: Andre Sopczak. Downloaded on October 12, 2009 at 10:27 from IEEE Xplore. Restrictions apply.

2926

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009

Fig. 2. CPC1 (in the center of the picture) with external electronics.

Fig. 3. On-chip amplier with a oating diffusion detection node.

very low operating temperature is required to obtain sensitivity to the CTI peak structure for 0.17 eV traps where the peak posiC [18]. This is for particular tion is expected to be near clocking conditions used in the setup. The CPC1 is a 2-phase 750(H) pixels, m size each CCD with 400(V) pixel. Initial measurements have been performed on an un-irradiated device in standalone mode, where the signals from four columns of the CCD were amplied and connected to external 14-bit ADCs. A Fe source emitting 6 keV X-rays was attached to a holder at a distance of 5 cm from the CCD to provide the signal charge. Fig. 2 shows a picture of CPC1 and the associated control and front-end readout electronics, which were also placed in the cryostat, wherefrom the four outputs were fed to rack-mounted ampliers and ADCs. Fig. 3 illustrates the on-chip amplier. The main purpose of the on-chip amplier in a CCD is the conversion of a charge packet into a voltage or a current. The change in the detection when a charge packet enters the node voltage node is sensed by the on-chip amplier. In order to achieve a large bandwidth a two-stage source follower amplier is used. The rst stage (First SF) consists of a source follower and a current sink for biasing purposes. The second stage (Second SF) drives the external load. The reset FET, controlled by RG, is connected to the oating diffusion region. In the off-state it can collect the next charge packet and in the on-state it resets the

detection node to a reference voltage (RD). The bias voltage CS determines the bias current of the rst stage. The schematic diagram in Fig. 4 illustrates the electronics used to drive and read out the CPCCD. The apparatus is controlled by a LabView program through interface crates. The sequencer receives the master clock from the signal generator to provide a clock for the ADC and Sine Module and to trigger a pulse generator, which produces the reset signal RG for the CCD output gate. The sine module produces two clock signals for the two-phase CCD. In addition, an AC signal for output gate OG is generated. It was observed that the CTI increased sharply when the amplitude of the sinusoidal clock pulses applied to read out the CCD was reduced to less than 2 V peak-to-peak. In order to suppress any effect from the clock amplitude, settings for each clock pulses. data acquisition were tuned to produce 3.00 The timing settings (delays) depend on the readout frequency and operating temperature and had to be adjusted for each measurement. The performance of the drivers limited the maximum clocking frequency to about 22 MHz. The X-ray hit rate (occupancy) is kept constant. We expect that the errors on the CTI calculation decrease when increasing the occupancy. If we consider that 10 points are sufcient to t the X-ray peak with a , Gaussian function, the occupancy would be where 5000 is the number of frames (4 400 pixels). In these measurements the occupancy was small (about 0.1%), resulting from the rather low activity of the Fe source. Such a low occupancy is expected in Linear Collider operations, and has also been used in recent simulations [18]. Our cryostat did not allow placing the source very far from the detector. This may have caused some non-uniformity in the irradiation across the CCD. This non-uniformity is estimated to be about a 1% reduction at the edges of the sensitive area compared to the center. Our method is based on the typical methods used for serially read out CCDs [24], [25], where CTI is determined by tting a line to the read out charge signal as function of read (pixel) number, which is equal to the number of charge transfers needed to transfer the charge collected in that pixel across the CCD into the readout electrode. A linear function can be expected when the CTI is small.

III. SIGNAL MEASUREMENT AND REMOVING CHARGE SHARING BY USING A 3 3 CLUSTER METHOD The fast ADCs convert the signal charge after amplication with a wideband preamplier. The four columns are read out in 4 channels by ADCs. Three columns (channels 1 to 3) are adjacent whereas channel 4 is separate. The signal charges 10 overclocks were acquired in of 4 400 pixels plus 4 30005000 frames per measurement. This leads to a sufcient ). The collected statistical precision (around few times data have been analysed using the ROOT package [26]. First, we begin by applying correlated double sampling, where the difference between the signals of two consecutive pixels is taken to be the signal charge collected by the latter pixel. This decreases electronic noise. Common mode noise is reduced with the help of the overclocks, which are charges read out

Authorized licensed use limited to: Andre Sopczak. Downloaded on October 12, 2009 at 10:27 from IEEE Xplore. Restrictions apply.

SOPCZAK et al.: MEASUREMENTS OF CHARGE TRANSFER INEFFICIENCY IN A CCD WITH HIGH-SPEED COLUMN PARALLEL READOUT

2927

Fig. 4. Schematic diagram of CPC1s readout.

Fig. 6. Illustration of the 3 3 cluster method. A signal charge (from the X-ray) is only accepted if no charge from X-rays is present in the neighboring pixels. The X-ray charge is dened by its threshold determined after a rst t of the X-ray signal peak with a Gaussian function.

Fig. 5. Distribution of ADC codes for channel (column) two. Three regions are observed: (I) the high peak region which represents the noise, (II) the region separating the two peaks which represents the charge sharing between pixels, and (III) the X-ray peak region which represents the collected charge in a single pixel. The noise peak is tted by a Gaussian function to determine the noise threshold. The standard deviation of the noise is about 60 electrons. The data was taken at 80 C with 8 MHz readout frequency.

a tail remains (Fig. 7). In order to separate the X-ray signal region from the charge-sharing region we exclude all ADC codes below an X-ray threshold which is determined by tting the below the tted peak X-ray peak. We set the threshold to be centroid. In this way we obtain the measured X-ray signal data in the pixels with full X-ray charge (Fig. 8). IV. BASELINE CORRECTION FOR DIFFERENT OPERATING TEMPERATURES AND READOUT FREQUENCIES Owing to an RC coupling at the preamplier input, the output voltage decays exponentially with a time constant equal to ( is of the order of 100 s).1 This is expected to give rise to an exponential baseline that varies during the clocking sequence, adding a contribution to the readout signal that depends on the pixel number. The shape of the baseline can be determined by comparing the average output level in the pixels to each other. For each pixel an average of the ADC codes was computed excluding the frames where the code exceeded the noise threshold. The average ADC code can usually be expected to be an exponential function of the pixel number with the RC decay. Figs. 9, 10, 11 and 12 show the distributions of the average ADC codes for all frames of a measurement as a function
1For a readout frequency of 4 MHz reading out the whole column of 400 pixels takes a time comparable to  .

after the last pixel of the column. The average level of these is subtracted from all the pixels in the column. As an example, Fig. 5 shows the pulse-height distribution for channel 2 summing up all reads in all the 400 pixels of the column. CTI is always calculated for channel 2 because it is physically adjacent to channels 1 and 3. This allows to apply a 3 3 cluster method (Fig. 6), in order to remove hits with charges from X-ray shared between pixels. This method is only applied to the 3 adjacent columns (1, 2 and 3). Column 2 is considered, where the signal charge of a pixel is accepted only if no charges from X-ray are present in the neighboring pixels. The noise threshold is used to separate noise from charges shared between pixels. It is determined by tting the noise peak with a Gaussian function. The threshold is usually set to be above the tted peak centroid. Even after the 3-by-3 clustering,

Authorized licensed use limited to: Andre Sopczak. Downloaded on October 12, 2009 at 10:27 from IEEE Xplore. Restrictions apply.

2928

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009

Fig. 7. X-ray signal peak with a tail from some charge sharing even after applying the 3 3 cluster method. The distribution is tted with a Gaussian function to determine the threshold to be used to completely remove the charge sharing. This is a different dataset than that in Fig. 5, which was obtained using a higher preamplier gain.

Fig. 9. Distribution of average ADC code in the noise region versus the pixel number at 2 MHz and 60 C with an exponential t. The noise region was delimited by a 3 noise threshold. Fit: amplitude A = 0:117 0:056 and exponent B = ( 5:29 2:85) 10 .

0 6

Fig. 8. X-ray peak after the removal of charge sharing.

of the pixel number for different readout frequencies and operating temperatures. These distributions are tted by an exponen, where corresponds to the tial function given by signal charge at the rst pixel, is the slope and is the pixel number. This function will be subtracted from the X-ray data. Figs. 9 and 12 show approximately zero baseline level, whereas Figs. 10 and 11 show non-zero baseline ts. The baseline level C is much smaller at low temperatures, for example at C (Fig. 12). Figs. 10 and 11 show base(Fig. 9) and at lines for higher temperatures C and C, respectively. These observed baselines show different frequency dependence from what is expected from the RC effect. This indicates that other electronic effects also inuence the baseline. For the same readout frequency (8 MHz) Figs. 11 and 12 show the temperature dependence explicitly. V. DETERMINATION OF CHARGE TRANSFER INEFFICIENCY The charge transfer inefciency (CTI) in one pixel is dened as the ratio of signal lost during transfer (captured by traps) to the initial signal charge. In order to determine the CTI we rst make the overclock correction and apply the 3 3 cluster method. Then, the X-ray peak was tted with the Gaussian function to determine the X-ray threshold. This was used to construct

Fig. 10. Distribution of average ADC code in the noise region versus the pixel number at 4 MHz and 6 C with an exponential t. The noise region was delimited by a 3 noise threshold. Fit: amplitude A = 15:64 0:44 and exponent B = ( 16:24 1:33) 10 .

Fig. 11. Distribution of average ADC code in the noise region versus the pixel number at 8 MHz and 17 C with an exponential t. The noise region was delimited by a 3 noise threshold. Fit: amplitude A = 8:31 0:31 and exponent B = ( 11:86 1:74) 10 .

0 2

the distribution with a t of average ADC codes as shown in Fig. 13. The distribution is tted with the rst-order polynomial , where corresponds to the charge at the function

Authorized licensed use limited to: Andre Sopczak. Downloaded on October 12, 2009 at 10:27 from IEEE Xplore. Restrictions apply.

SOPCZAK et al.: MEASUREMENTS OF CHARGE TRANSFER INEFFICIENCY IN A CCD WITH HIGH-SPEED COLUMN PARALLEL READOUT

2929

Fig. 12. Distribution of average ADC code in the noise region versus the pixel number at 8 MHz and 109 C with an exponential t. The noise region was delimited by a 3 noise threshold. Fit: amplitude A = (0:082 5:035) 10 and exponent B = (1:073 16:853) 10 .

Fig. 15. CTI versus temperature at 2 MHz readout frequency. The CTI is shown resulting from ts to average ADC codes, with and without baseline removal. The shown error bars result from the precision of the ts.

Fig. 13. Fit to average ADC codes without baseline removal at 2 MHz and 31 C. CT I = (2:99 2:69) 10 .

Fig. 16. CTI versus temperature at 4 MHz readout frequency. The CTI is shown resulting from ts to average ADC codes, with and without baseline removal. The shown error bars result from the precision of the ts.

VI. CTI RESULTS PRE-IRRADIATION FOR DIFFERENT READOUT FREQUENCIES Figs. 15, 16 and 17 show the CTI values as a function of temperature for an un-irradiated CPC1 at different readout frequencies. The CTI has been calculated using a linear t of average ADC codes versus pixel number. For this CCD with 400 pixels means that only 0.4% of signal charge is a CTI value of lost, which is acceptable in normal operation. In the temperaC no frequency dependence of the CTI ture region below measurements is observed. One may note that in the region of C) the effect of baserather high temperatures (above line removal is large and the determined CTI values indicate that the baseline effect is overestimated. The effect of the baseline removal decreases with increasing frequency as the readout (as visible clocking sequence becomes shorter compared to in Fig. 17 for 8 MHz readout frequency). The apparent trend of the CTI at high temperatures in the operating range used is probably due to the increase in background charge level as it is related to the dark current. The negative CTI

Fig. 14. Fit to average ADC codes with baseline removal at 2 MHz and 31 C. CT I = ( 0:09 2:70) 10 .

rst pixel, is the slope and is the pixel number. This distribution is plotted without baseline removal. Fig. 14 shows the distribution of the averages as a function of pixel number with baseline removal. The CTI is determined for the two cases using .

Authorized licensed use limited to: Andre Sopczak. Downloaded on October 12, 2009 at 10:27 from IEEE Xplore. Restrictions apply.

2930

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009

support and the organizers of the IEEE 2008 conference for their hospitality. REFERENCES
[1] C. J. S. Damerell, Radiation damage in CCDs used as particle detectors, ICFA Instrum. Bull., vol. 14, p. 1, 1997. [2] K. Stefanov, Radiation Damage Effects in CCD Sensors for Tracking Applications in High Energy Physics, Ph.D. dissertation, Saga Univ., Saga, Japan, 2001, and references therein; K. Stefanov, et al., Electron and neutron radiation damage effects on a two phase CCD, IEEE Trans. Nucl. Sci., vol. 47, p. 1280, 2000. [3] LCFI Collaboration Homepage [Online]. Available: http://hepwww.rl.ac.uk/lc/ [4] S. D. Worm, Recent CCD development for the vertex detector of the ILCIncluding ISIS (in-situ storage information sensors), presented at the 10th Topical Seminar on Innovative Particle and Radiation Detectors. Siena, Italy, Oct. 15, 2006. [5] T. J. Greenshaw, Column parallel CCDs and in-situ storage image sensors for the vertex detector of the international linear collider, in Nuclear Science Symp., San Diego, CA, Oct. 29Nov. 4 2006. [6] M. S. Robbins, The Radiation Damage Performance of Marconi CCDs, Marconi Tech. Note S&C 906/424 2000 (unpublished). [7] J. E. Brau and N. B. Sinev, Operation of a CCD particle detector in the presence of bulk neutron damage, IEEE Trans. Nucl. Sci., vol. 47, p. 1898, 2000. [8] J. E. Brau, O. Igonkina, C. T. Potter, and N. B. Sinev, Investigation of radiation damage in the SLD CCD vertex detector, IEEE Trans. Nucl. Sci., vol. 51, p. 1742, 2004. [9] J. E. Brau, O. Igonkina, C. T. Potter, and N. B. Sinev, Investigation of radiation damage effects in neutron irradiated CCD, Nucl. Instrum. Meth. A, vol. 549, p. 117, 2005. [10] J. E. Brau, O. Igonkina, N. B. Sinev, and J. Strube, Investigations into properties of charge traps created in CCDs by neutron and electron irradiation, Pramana-J. Phys., vol. 69, p. 1093, 2007. [11] A. M. Mohsen and M. F. Tompsett, The effect of bulk traps on the performance of bulk channel charge-coupled devices, IEEE Trans. Electron Dev., vol. ED-11, p. 701, 1974. [12] I. H. Hopkins, G. Hopkinson, and B. Johlander, Proton-induced charge transfer degradation in CCDs for near-room temperature applications, IEEE Trans. Nucl. Sci., vol. 41, p. 1984, 1994. [13] T. Hardy, R. Murowinski, and M. J. Deen, Charge transfer efciency in proton damaged CCDs, IEEE Trans. Nucl. Sci., vol. 45, p. 154, 1998. [14] A. Sopczak, LCFI charge transfer inefciency studies for CCD vertex detectors, in Proc. IEEE Nuclear Science Symp., San Juan, PR, 2005, p. 1494, IEEE Nuclear Science Symp. Conf. Record N37-7. [15] A. Sopczak, Radiation hardness of CCD vertex detectors for the ILC, in Proc. IEEE Nuclear Science Symp., San Diego, CA, Oct. 29Nov. 4 2006, p. 576, IEEE Nuclear Science Conf. Record N14-215 (2006). [16] A. Sopczak et al., Simulations of the temperature dependence of the charge transfer inefciency in a high-speed CCD, IEEE Trans. Nucl. Sci., vol. 54, p. 429, 2007, and references therein. [17] A. Sopczak et al., Radiation hardness studies in a CCD with highspeed column parallel readout, in Proc. IEEE Nuclear Science Symp., Honolulu, HI, Oct. 27Nov. 3 2007, p. 2278, IEEE Nuclear Science Conf. Record N48-2 (2007). [18] A. Sopczak et al., Radiation hardness studies in a CCD with highspeed column parallel readout, JINST, vol. 3, p. 5007, 2008. [19] T. Maruyama, Private Communication. 2006, Stanford (Linear Accelerator Center, SLAC). [20] A. Vogel, Private Communication. 2006, (DESY Hamburg). [21] J. W. Walker and C. T. Sah, Properties of 1.0-MeV-electron-irradiated defect centers in silicon, Phys. Rev., vol. B7, p. 4587, 1972. [22] N. S. Saks, Investigation of bulk electron traps created by fast neutron irradiation in a buried n-channel CCD, IEEE Trans. Nucl. Sci., vol. NS-24, p. 2153, 1977. [23] J. R. Srour, R. A. Hartmann, and S. Othmer, Transient and permanent effects of neutron bombardment on a commercially available N-buriedchannel CCD, IEEE Trans. Nucl. Sci., vol. NS-27, p. 1402, 1980. [24] R. M. Ambrosi et al., The impact of low energy proton damage on the operational characteristics of EPIC-MOS CCDs, Nucl. Instrum. Meth. B, vol. 207, p. 175, 2003. [25] A. F. Abbey et al., Cooling out the radiation damage on the XMMNewton EPIC MOS CCDs, Nucl. Instrum. Meth. A, vol. 513, p. 136, 2003. [26] ROOT Homepage [Online]. Available: http://root.cern.ch/

Fig. 17. CTI versus temperature at 8 MHz readout frequency. The CTI is shown resulting from ts to average ADC codes, with and without baseline removal. The shown error bars result from the precision of the ts.

is due to the difculty in determining the baseline. The baseline is affected by the RC effect that decreases with the pixel number and the dark current that increases at high temperatures. This uncertainty in determining the baseline, which is subtracted from the X-ray peak, gives a negative value especially at high temperatures when the noise is high and closer to the X-ray peak. Also the low X-ray hit rate leads to a weak X-ray peak which leads to a bad t and hence an additional uncertainty in the X-ray centroid determination. The accuracy of CTI calculation can be improved by putting the Fe source in a position that it hits uniformly the CCD, carefully choosing the gain to use the ADC in its maximum dynamic and acquiring data in a large number of frames. A few measurements where a fraction of the data was lost in the sampling have been discarded. VII. CONCLUSIONS AND OUTLOOK An un-irradiated CPCCD is operated in a range of temperaC to C (liquid nitrogen cooling) with tures from different readout frequencies 2, 4 and 8 MHz. The spectrum of a Fe source is measured with this device. The CTI is analysed for different readout frequencies and operating temperatures. A clear X-ray signal is extracted by identifying isolated hits (3 3 method). The baseline is subtracted. The CTI value is small and compatible to the order of magnitude that can be expected for an un-irradiated CPCCD. Further CTI measurements with a CPCCD after irradiation and renement of the analysis method are planned. ACKNOWLEDGMENT The authors would like to thank S. Aoulmit, C. Bowdery, A. Chilingarov, L. Dehimi, and N. Sengouga for discussions and comments on the manuscript. This work was supported by the Science and Technology Facilities Council (STFC) and Lancaster University. K. Bekhouche would like to thank the Algerian Government for nancial support and Lancaster University for their hospitality. A. Sopczak would like to thank the Faculty of Science and Technology at Lancaster University for nancial