You are on page 1of 64

Del DRV Indico 100 Operators Manual

Indico 100

II

III

This manual contains information For the operation of the Indico 100 X-Ray generator
Del Medical Imaging Corp. 11550 W. King Street Franklin Park, IL 60131 Technical Support: 847-288-7000 Fax: 847-288-7911 Email: tech_support@delmedical.com Manual Part Number: 8000-DRVOPR Revision: A May 23, 2003

(This page intentionally left blank)

INTRODUCTION

1 2

Del DRV INDICO OPERATOR MANUAL

SAFETY AND SPECIFICATIONS

CONSOLE CONTROLS

3 4 5 6 7 8 9 10

REMOTE FLUORO CONTROLLER

kV -mAs REDUCTION FOR PERIPHERAL ANGIOGRAPHY

PROGRAMMING THE CONSOLE

ERROR CODES AND MESSAGES

GENERATOR EXPOSURE TABLES

MAINTENANCE SCHEDULE

X-RAY TUBE DATA

(This page intentionally left blank) 2003 Del Medical Imaging Corp. Revision A Page 3

2003 Del Medical Imaging Corp.

Revision A

Page 4

INTRODUCTION
Congratulations!

You have acquired the Del Indico 100 series high frequency X-ray generator. This generator features stateof-the-art computer-based control to ensure minimum patient dose, excellent repeatability, and superior contrast. The operator control functions are designed to be simple and user-friendly. The AEC (Automatic Exposure Control) and APR (Anatomical Programming) features give you controlled exposure factors, automatically optimized for the radiological study selected.

INDICO 100 MAIN FEATURES


Models available in maximum output power of 32 to 80 kW, in both radiographic and fluoroscopic models. A radiographic-only 100 kW model is also available. Full 150 kV capability for all models. Smaller, lighter modular design. Constant dose output due to kV and mA regulation during exposures. User-friendly controls. Large LCD panel display for APR and utility routines. User-friendly system configuration. APR techniques may be modified by the operator. ms and mAs limits for AEC may be operator set. Provisions for APR / tube downloading. Extensive self-diagnostics with operator prompt messages. Serial RS232 communications interface. May be programmed for APR operation only or both APR and manual technique selection. Time and mA / mAs selections are to IEC standards. Programmable setup, calibration and APR routines (using an external computer). Choice of ten fluoro kV to rad kV transfer curves (R & F units only). Choice of four kV vs. mA curves for ABS (automatic brightness stabilization) circuits (R & F units only).

OPTIONS

Dual speed starter for automatic high / low speed selection. Two-tube capability. Pedestal mount or wall mount bracket for operator control console. Automatic brightness stabilization (ABS) for consistent imaging (AUTO kV / mA or kV only). Remote fluoro controller for R & F tableside operation. Digital imaging interface. Remote exposure hand switch. High level fluoro for therapy simulators. Pulsed fluoro with last image hold. kV / mAs reduction for peripheral angiography. DAP (Dose Area Product).

SAFETY NOTICE:

This manual contains important operating and safety information. An understanding of this information is critical to the safe operation and long term reliability of your equipment. Please ensure that you read the warning notices before using this equipment. Revision A Page 5

2003 Del Medical Imaging Corp.

Introduction

GENERATOR DUTY CYCLE LIMIT NOTE: THE FOLLOWING SECTION CONTAINS IMPORTANT INFORMATION. PLEASE READ AND UNDERSTAND THIS MATERIAL BEFORE CONTINUING.

Internal X-ray generator components will heat up during normal use of the generator. This is similar to X-ray tube heating during normal generator operation. The amount of heat produced is proportional to the product of kV, mA, and time Modern X-ray generators are designed to operate with the majority of X-ray tubes over their rated power ranges. They are designed for operating duty cycles consistent with practical patient examination routines that allow for reasonable cooling intervals between X-ray exposures. Insufficient cooling time between exposures may lead to excessive heat build - up in the generator, which may cause serious generator damage. This X-ray generator has internal duty cycle monitoring to warn of excessive heat build - up. If the generator calculates that the next exposure will exceed the rated generator duty cycle limit, a warning message GEN DUTY WARNING will be displayed. Exposures will not be inhibited when this message is displayed, but it should be understood that continuing to make exposures might cause generator damage due to overheating. If possible, the generator should be allowed to cool sufficiently such that this message is no longer displayed.

DAILY X-RAY TUBE WARM-UP PROCEDURE


CAUTION: THE FOLLOWING PROCEDURE PRODUCES PRECAUTIONS TO PROTECT PERSONNEL. X-RAYS. OBSERVE ALL SAFETY

Use this procedure when the generator is first turned on for the day, or when it has not been used for several hours. This procedure provides for exposures at medium power before the tube is used at maximum mA or kV values. This will reduce the possibility of damaging the anode and high voltage components. No test setup is required. For maximum stability and reliability, use the following techniques at start up: Select the following:

Large focal spot 80 kV Normal 50/60 Hz anode rotation

For a 300 kHU to a 400 kHU tube, use approximately 200 mAs per exposure. For a 200 kHU to 300 kHU tube use approximately 150 mAs per exposure. Depending on the X-ray tube power rating, select either 100 mA or 200 mA. Make one (1) to three (3) exposures (depending on tube loading) at 30 second intervals.

2003 Del Medical Imaging Corp.

Revision A

Page 6

SA FET Y A N D SPECI FI CA T I ON S

Keep this operators manual with the equipment at all times, and periodically review the operating and safety instructions.

SAFETY / WARNING SYMBOLS


Warning symbol used to indicate a potential hazard to operators, service personnel or to the equipment. It indicates a requirement to refer to the accompanying documentation for details. Radiation exposure symbol used on operator console. Lights to indicate that an exposure is in progress. This is accompanied by an audible tone from the console. Fluoro radiation exposure symbol used on operator console and on optional remote fluoro control unit. Lights to indicate that a fluoro exposure is in progress. This is accompanied by an audible tone from the console. Radiation warning label on console, used in certain jurisdictions. Never allow unqualified personnel to operate the Xray generator.

WARNING

THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.

WARNING:

PROPER USE AND SAFE OPERATING PRACTICES WITH RESPECT TO X-RAY GENERATORS ARE THE RESPONSIBILITY OF THE USERS OF SUCH GENERATORS. DEL MEDICAL IMAGING CORP. PROVIDES INFORMATION ON ITS PRODUCTS AND ASSOCIATED HAZARDS, BUT ASSUMES NO RESPONSIBILITIES FOR AFTER-SALE OPERATING AND SAFETY PRACTICES. DEL ACCEPTS NO RESPONSIBILITY FOR ANY GENERATOR NOT MAINTAINED OR SERVICED ACCORDING TO THE SERVICE AND INSTALLATION MANUAL OR ANY GENERATOR THAT HAS BEEN MODIFIED IN ANY WAY. DEL ALSO ASSUMES NO RESPONSIBILITY FOR X-RAY RADIATION OVEREXPOSURE OF PATIENTS OR PERSONNEL RESULTING FROM POOR OPERATING TECHNIQUES OR PROCEDURES.

WARNING:

THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.

2003 Del Medical Imaging Corp.

Revision A

Page 7

Safety and Specifications

X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending over periods of many months or even years. X-ray operators should avoid any exposure to the primary beam and take protective measures to safeguard against scatter radiation. Scatter radiation is caused by any object in the path of the primary beam and may be of equal or less intensity than the primary beam that exposes the film. No practical design can incorporate complete protection for operators or service personnel who do not take adequate safety precautions. Only authorized and properly trained service and operating personnel should be allowed to work with this X-ray generator equipment. The appropriate personnel must be made aware of the inherent dangers associated with the servicing of high voltage equipment and the danger of excessive exposure to X-ray radiation during system operation.

Wear protective clothing. Protective aprons with an equivalent of a minimum of 1/64 (0.35 mm) of lead are recommended. To protect the patient against radiation, always use radiation protection accessories in addition to devices which are fitted to the X-ray equipment. Keep as large a distance as possible away from the object being exposed and the X-ray tube assembly. This X-ray equipment may only be operated in medical rooms which meet IEC requirements. Never operate this X-ray equipment in areas where there is a risk of explosion. Detergents and disinfectants, including those used on patients, may create explosive mixtures of gases. Please observe the relevant regulations. The operator console, or anything electrically connected to it, must never be used within 6 ft (1.8 m) of the patient environment. Only trained maintenance staff may remove the covers of the generator cabinet and the control console. Do not connect unapproved equipment to the rear of the console: J5 is for the interconnect cable to the generator main cabinet, J2 is a serial port for use by an external computer, and J13 is for connection of an external hand switch and / or foot switch. INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN INJURY OR EQUIPMENT DAMAGE. CONSULT YOUR SERVICE REPRESENTATIVE IF EXTERNAL EQUIPMENT IS TO BE CONNECTED TO THIS SYSTEM. PERIPHERAL DEVICES SHOULD BE IEC 601-1 APPROVED IF LOCATED IN THE PATIENT AREA, AND IEC 950 APPROVED IF OUTSIDE THE PATIENT AREA.

CAUTION:

DO NOT EXCEED THE TUBE MAXIMUM OPERATING LIMITS SHOWN IN THE SECTION XRAY TUBE DATA AT THE END OF THE OPERATORS MANUAL. INTENDED LIFE AND RELIABILITY WILL NOT BE OBTAINED UNLESS GENERATORS ARE OPERATED WITHIN PUBLISHED SPECIFICATIONS.

2003 Del Medical Imaging Corp.

Revision A

Page 8

Safety and Specifications

APPLICABLE STANDARDS
The Indico 100 family of generators complies with the following regulatory requirements and design standards:

FDA Center for Devices & Radiological Health (CDRH) - 21 CFR title 21 subchapter J (USA). Radiation Emitting Devices Act - C34 (Canada). Medical Device Regulations (Canada). EC Directive 93/42/EEC concerning Medical Devices (European Community). IEC 601.1, IEC 601.2.7:1998, CSA 601.1, UL2601.1 -Type of protection against electric shock: Class I equipment. -Degree of protection against electric shock: Not classified. -Degree of protection against harmful ingress of water: Ordinary equipment. -Mode of operation: Continuous operation with intermittent loading (standby - exposure). -Equipment not suitable for use in presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE. IEC 601.1.2 Immunity: IEC1000-4-2 IEC1000-4-3 IEC1000-4-4 IEC1000-4-5 IEC1000-4-6 IEC1000-4-8 IEC1000-4-11

Electrostatic discharge Radiated RF field Electrical fast transient Surge Conducted RF immunity Magnetic field immunity Voltage dips, interrupts and variations

Emission: EN55011 (CISPR Publications II Emission Standards, Group 1 Class A).

ELECTROMAGNETIC COMPATIBILITY (EMC)


In accordance with the intended use, this X-ray generator complies with the European Council Directive concerning Medical Devices. The CE marking affixed to this product signifies this. One of the harmonized standards of this Directive defines the permitted levels of electromagnetic emission from this equipment and its required immunity from the electromagnetic emissions of other devices. It is not possible, however, to exclude with absolute certainty the possibility that other high frequency electronic equipment, which is fully compliant to the EMC regulations, will not adversely affect the operation of this generator. If the other equipment has a comparatively high level of transmission power and is in close proximity to the generator, these EMC concerns (the risk of interference) may be more pronounced. It is therefore recommended that the operation of equipment of this type such as mobile telephones, cordless microphones and other similar mobile radio equipment be restricted from the vicinity of this X-ray generator.

2003 Del Medical Imaging Corp.

Revision A

Page 9

Safety and Specifications

OUTPUT PARAMETERS

kV range: kV accuracy: mA range:

Radiography 40 to 150 kV. Fluoroscopy 40 to 125 kV.


(5% + 1) kVp.

Radiography 10 to 400 mA (32 kW), 10 to 500 mA (40 kW), 10 to 630 mA (50 kW), 10 to 800 mA (65 kW), and 10 to 1000 mA (80 and 100 kW). Fluoroscopy 0.5 to 6.0 mA, 0.5 to 20 mA optional. Normally used for therapy simulators only.
(5% +1.0) mA for all mA values.

mA accuracy: mAs range: Time range:

Tube dependent, max 1000 mAs. Radiography 1.0 to 6300 ms (40 to 100 kW), 1.25 to 6300ms (32 kW). Fluoroscopy 0 to 5 or 0 to 10 minutes. 0.5 mAs.

Lowest current - time product: Loading factors which will allow lowest current - time product:

All mA values 500 mA. Refer to table 1 in section 7.

* 20 mA maximum for fluoroscopy available with high level fluoro option only.

2003 Del Medical Imaging Corp.

Revision A

Page 10

Safety and Specifications

OUTPUT PARAMETERS (Continued)


OUTPUT PARAMETER Maximum X-ray tube voltage and highest X-ray tube current at that voltage MODE Radiographic (Intermittent) GENERATOR SERIES 100 kW 80 kW 65 kW 50 kW 40 kW 32 kW 80 kW 65 kW 50 kW 40 kW 32 kW 100 kW 80 kW 65 kW 50 kW 40 kW 32 kW 80 kW 65 kW 50 kW 40 kW 32 kW 100 kW 80 kW 65 kW 50 kW 40 kW 32 kW 80 kW 65 kW 50 kW 40 kW 32 kW 100 kW 80 kW 65 kW 50 kW 40 kW 32 kW All models 100 kW 80 kW 63 kW 50 kW 40 kW 32 kW LOADING FACTOR 150 kV, 630 mA 150 kV, 500 mA 150 kV, 400 mA 150 kV, 320 mA 150 kV, 250 mA 150 kV, 200 mA 125 kV, 6 / 20 mA 125 kV, 6 / 20 mA 125 kV, 6 / 20 mA 125 kV, 6 / 20 mA 125 kV, 6 / 20 mA 1000 mA, 80 kV 1000 mA, 80 kV 800 mA, 80 kV 630 mA, 80 kV 500 mA, 80 kV 400 mA, 80 kV 6 / 20 mA, 125 kV 6 / 20 mA, 125 kV 6 / 20 mA, 125 kV 6 / 20 mA, 125 kV 6 / 20 mA, 125 kV 1000 mA, 100 kV 1000 mA, 80 kV 800 mA, 80 kV 630 mA, 80 kV 400 mA, 100 kV 320 mA, 100 kV 6 / 20 mA, 125 kV 6 / 20 mA, 125 kV 6 / 20 mA, 125 kV 6 / 20 mA, 125 kV 6 / 20 mA, 125 kV (1000 mA, 100 kV, 0.1 s) (800 mA, 100 kV, 0.1 s) (630 mA, 100 kV, 0.1 s) (500 mA, 100 kV, 0.1 s) (400 mA, 100 kV, 0.1 s) (320 mA, 100 kV, 0.1 s) 15 ms

Maximum X-ray tube voltage and highest X-ray tube current at that voltage * See note on previous page Maximum X-ray tube current and highest X-ray tube voltage at that current

Fluoroscopic (Continuous)

Radiographic (Intermittent)

Maximum X-ray tube current and highest X-ray tube voltage at that current * See note on previous page Combination of X-ray tube current and X-ray tube voltage resulting in highest output power

Fluoroscopic (Continuous)

Radiographic (Intermittent)

Combination of X-ray tube current and X-ray tube voltage resulting in highest output power * See note on previous page Highest constant output power at 100 kV, 0.1 sec

Fluoroscopic (Continuous)

Radiographic (Intermittent)

Nominal shortest irradiation time (AEC exposures)

AEC

2003 Del Medical Imaging Corp.

Revision A

Page 11

Safety and Specifications

ENVIRONMENTAL SPECIFICATIONS
OPERATING Ambient temperature range: Relative humidity: Atmospheric pressure range: TRANSPORT AND STORAGE Ambient temperature range: Relative humidity: Atmospheric pressure range: -20 to 70 C. 10 to 95%, including condensation. 500 to 1060 hPa (375 to 795 mm Hg). 10 to 40 C. 30 to 75%. 500 to 1060 hPa (375 to 795 mm Hg).

This information is provided to help you establish safe operating conditions for both you and your DEL Xray generator. Do not operate this X-ray generator except in accordance with information included in this section, and any additional information provided by the X-ray generator manufacturer and/or competent safety authorities. Address any questions regarding X-ray generator operation to:

Del Medical Imaging Corp. 11550 W. King Street Franklin Park, IL 60131 Technical Support: (847) 288-7000 Fax: (847) 288-7911 Email: tech_support@delmedical.com

2003 Del Medical Imaging Corp.

Revision A

Page 12

CONSOLE CONTROLS
OPERATOR CONTROL PANEL

Indico 100

II

III

FILE: IN_CONS.CDR

1. 2. 3. 4. 5. 6.

Power ON, power OFF buttons. Radiography controls and display. Image receptor buttons. Anatomical programming display and control. Fluoroscopy controls and display. Prep and exposure buttons and exposure indicator LED.
THIS MANUAL REFERS TO THE INDICO 100 R & F CONSOLE. FOR RAD ONLY UNITS, IGNORE REFERENCES TO THE FLUOROSCOPY FUNCTION. ICONS ADJACENT TO BUTTONS ARE SHOWN AS THE CONSOLE IS SHIPPED FROM THE FACTORY. THESE MAY BE ALTERED BY THE INSTALLER TO MATCH THE ACTUAL CONFIGURATION OF THE GENERATOR.

NOTE:

NOTE:

2003 Del Medical Imaging Corp.

Revision A

Page 13

Console Controls

POWER AND EXPOSURE CONTROLS Power ON / Power OFF buttons Press to switch the X-ray generator on. The console will light up and a self-check will be performed. Press to switch the generator off.

Prep / Exposure Buttons and Exposure Indicator LED Press and hold the prep button to spin the rotor. While pressing the prep button, press and hold the exposure button to make an X-ray exposure. will light when an

The X-ray exposure indicator X-ray exposure is being taken.

Pressing the exposure button only will cycle the generator through prep and then exposure. When the prep button is pressed, SPINNING ROTOR will be displayed in the APR window. When the prep cycle is complete, X-RAY READY will be displayed in the APR window. During the X-ray exposure, X-RAY ON will be displayed in the APR window.

2003 Del Medical Imaging Corp.

Revision A

Page 14

Console Controls

RADIOGRAPHY CONTROLS AND DISPLAY

II

III

1. 2. 3. 4. 5. 6. 7. 8. 9.

Field select buttons and indicators (for AEC only). Film / Screen select button and indicators (for AEC only). Technique select button and indicators. Focus select button and indicators. kV up / down buttons. mA and mAs up / down buttons. Time (ms) up / down buttons. Density up / down buttons. Radiography display panel: kV, mA / mAs, Post mAs / ms, and density.

2003 Del Medical Imaging Corp.

Revision A

Page 15

Console Controls

Field Select Buttons and Indicators (for AEC only) Press the appropriate field select button(s) to select the desired field or combination of fields. The adjacent LED(s) will light to indicate the selected field(s). All three fields may not be deselected. Operates only when AEC is selected. Film / Screen Select Button and Indicators (for AEC only)
I II III

Press the film / screen select button to select the desired film / screen combination. The appropriate LED will light to indicate the desired selection.
I II III

for film / screen I. for film / screen II. for film / screen III.

Operates only when AEC is selected. Technique Select Button and Indicators This function may be disabled in programming. Press the technique select button until the appropriate LED lights to indicate the desired technique. for Automatic Exposure Control (1 point technique). mAs for mAs technique (2 point technique). mA/ms for mA and ms technique (3 point technique).

Focus Select Button and Indicators Press the focus select button until the appropriate LED lights to indicate the desired focal spot. for small focus. for large focus.

This function may be programmed for auto focal spot selection.

2003 Del Medical Imaging Corp.

Revision A

Page 16

Console Controls

Note:

The rate of scrolling for the kV, mA and time display increases if the buttons are pressed continuously. To return to the slowest scroll rate, release the button(s) and start again.

kV up / down buttons and display

To increase kV, press +. To decrease kV, press -. Displays demanded kVp of the exposure.

mA and mAs up / down buttons and display

To increase mA or mAs, press +. To decrease mA or mAs, press -. Displays mA when AEC or mA/ms selected. Displays mAs when mAs is selected.

Time (ms) up / down buttons and post-mAs / ms display

To increase time (ms), press +. To decrease time (ms), press -. Displays AEC, or shows backup mAs or backup ms when AEC selected. No display when mAs selected. Displays time (ms) when mA/ms selected. Displays POST mAs after an AEC exposure is completed.

Density up / down buttons and display

Buttons operate only in AEC mode. To increase dose, press +. To decrease dose, press -. The range is typically -5 to +5. (Installer programmable to have a range of up to -8 to +8). This varies the optical density by changing the dose. The % dose change per density step change is installer programmable.

2003 Del Medical Imaging Corp.

Revision A

Page 17

Console Controls

IMAGE RECEPTOR BUTTONS AND INDICATORS

1. 2. 3. 4. 5. 6.

Table, with Bucky, image receptor select Table, no Bucky, image receptor select Vertical image receptor select Linear tomographic receptor select Under-table R/F and remote R/F select Auxiliary image receptor select Image Receptors (1-6) are depicted as the console is shipped from the factory. The receptors may have been reprogrammed during installation. As a result, the receptors may contain installer - applied icons that differ from those shown in this manual.

Note:

Table with Bucky Image Receptor Press this button to select over-table X-ray tube with table image receptor. The adjacent indicator will light. Used for Bucky / AEC techniques.

Table no Bucky image Receptor Press this button to select over-table X-ray tube when no image receptor is required. The adjacent indicator will light. Used for table top and off table techniques.

2003 Del Medical Imaging Corp.

Revision A

Page 18

Console Controls

Vertical Image Receptor Press this button to select vertical (wall) image receptor or cassette holder. The adjacent indicator will light. Used for Bucky / AEC or cassette only techniques.

Linear Tomographic Receptor Press this button to select the linear tomographic receptor. The adjacent indicator will light. This receptor may be programmed to be remotely selected.

Under-Table R/F and Remote R/F Press this button to select the fluoroscopy function (radiographic / fluoroscopic models only). The adjacent indicator will light and the fluoroscopy display will light.

Auxiliary Press this button to select the auxiliary image receptor. The adjacent indicator will light. This may be programmed for R & F applications and digital applications. This receptor may be programmed to be remotely selected.

Note:

The image receptors may have been reprogrammed for your installation.

2003 Del Medical Imaging Corp.

Revision A

Page 19

Console Controls

FLUOROSCOPY CONTROLS AND DISPLAY

1. 2. 3. 4. 5. 6. 7. 8. 9. 10.

Fluoroscopy display: kV, mA, time (min), PPS Fluoro ON indicator Fluoro kV up / down buttons Fluoro mA up / down buttons Time zero and accumulated time buttons Pulsed fluoro frame rate buttons I.I. magnification select button Delta dose select button Automatic brightness stabilization (ABS) select button Pulsed fluoro ON / OFF button (Optionally selects high level fluoro)

2003 Del Medical Imaging Corp.

Revision A

Page 20

Console Controls

Fluoro ON Indicator When fluoro is on, the fluoro eye symbol will light. This is accompanied by an audible tone from the console.

Note:

The fluoro kV and fluoro mA buttons are only functional when ABS is switched OFF.

Fluoro kV up / down buttons and display


+

To increase kV, press +. To decrease kV, push -. lit indicates that ABS is selected, manual kV control is therefore not available.

Fluoro mA up / down buttons and display


+

To increase mA, press +. To decrease mA, press -. lit indicates that ABS is selected, manual mA control is therefore not available. Pulsed fluoro mode uses a fixed (not operator adjustable) mA value.

Time zero and accumulated time buttons and display To reset the time display to zero, press To display accumulated time, press . once. The

symbol will light to indicate that accumulated time is displayed. To reset the accumulated time, press the accumulated time is displayed. when

2003 Del Medical Imaging Corp.

Revision A

Page 21

Console Controls

Pulsed fluoro frame rate buttons and display


+

To increase the PPS (Pulse Per Second) rate, press +. To decrease the PPS rate, press -. Pulse rates are 3, 7.5, 15, and 30 PPS for 60 Hz based cameras, and 2.5, 6.25, 12.5, and 25 PPS for 50 Hz based cameras (6.25 is displayed as 6, and 12.5 is displayed as 12). Only active when pulsed fluoro is switched ON.

I.I. magnification select button Press to select I.I. magnification mode. The relative magnification value is shown in the lower right corner of the LCD display window (installer programmable). Delta dose select button Press to select delta dose control function. The delta dose value is shown in the lower right corner of the LCD display.

ABS (Automatic Brightness Stabilization) Select button Press to select ABS. The adjacent indicator will light. Switching off ABS will allow the operator to set the fluoro kV and mA manually.

2003 Del Medical Imaging Corp.

Revision A

Page 22

Console Controls

Pulsed fluoro ON / OFF button (Optional high level fluoro ON / OFF button) Press to select pulsed fluoro. The adjacent indicator will light. This is only active for optional pulsed fluoro. For optional high level fluoro (HLF): With ABS off, selects high level fluoro. The adjacent indicator will light to indicate that HLF is on. The fluoro mA range is 0.5 to 6.0 mA if HLF is off, and 5.0 to 20 mA if HLF is on. With ABS on, if full HLF mode is programmed, the fluoro mA range will be 0.5 to 20 mA. The generator will automatically switch between normal and HLF modes, and the adjacent indicator will light to indicate that the generator is in HLF mode. Manual HLF selection is not available if ABS is on. The frequency (pitch) of the tone from the console will increase during an exposure when high level fluoro is selected. Warning: High Level Fluoroscopy is compatible with therapy simulators only. High level Fluoro must not be used for other applications.

Exposure Indicator The LED adjacent to the prep symbol will light when the foot-switch is pressed for fluoroscopy exposure. The APR screen will display X-RAY ON. The operator control panel will also indicate the exposure by means of audible beeps. The rotor may continue to spin briefly after the foot-switch is released (installer programmable).

2003 Del Medical Imaging Corp.

Revision A

Page 23

Console Controls

ANATOMICAL PROGRAMMING CONTROLS AND DISPLAY

THE ABOVE IS A TYPICAL DISPLAY. YOUR UNIT MAY DIFFER DEPENDING ON MODEL, SET-UP, AND PROGRAMMING

1.

Anatomical section and anatomical view selection buttons. To select an APR technique: From the main APR menu (shown in the above figure), press the button next to the desired anatomical section. The APR display will show anatomical views within that section. Press the button for the required anatomical view. The radiography display will show the programmed X-ray technique factors. Press the function button again to toggle between normal, large and small patient sizes. The default is normal. You may override the programmed techniques by pressing the appropriate selection button (the ability to override the / mAs / mA/ms button may be disabled in programming). This will not change the programmed parameters. To make permanent changes to the APR, see CHAPTER 6: PROGRAMMING THE CONSOLE.

2. 3.

Tube 1 / tube 2 selection indicator (two tube units only). This will indicate which tube has been programmed for the selected receptor. Anode heat indicator (%HU) readout. Indicates the tube anode heat for the selected X-ray tube. An anode heat warning message will be displayed at an installer-programmable level, typically 80 %. Exposures that will exceed a value of typically 90 % will be inhibited (this is also installer programmable).

2003 Del Medical Imaging Corp.

Revision A

Page 24

Console Controls

4.

DAP (Dose Area Product) display. This feature is optional. 2 2 Displays accumulated dose (mGycm ) or dose rate (mGycm /s). Refer to DAP - DOSE AREA PRODUCT at the end of this chapter for further details. ms/mAs readout. The significance of the ms/mAs readout varies depending on mode selected: In AEC mode, with fixed backup time selected, the AEC backup time is displayed. In AEC mode, with either variable ms or variable mAs backup time, the backup mAs or ms value is displayed. In mAs mode, the time of the exposure will be displayed. In mA/ms mode, the calculated mAs of the exposure will be displayed (based on the mA and ms selected). Status and operator message display area. This area shows status, warning, and error messages. Time readout - 24 hour mode (RAD only). Displays MAG and DOSE selection in fluoro mode. The current time is displayed in 24 hour format when in RAD mode. The time display is replaced by a MAG selection indicator when a fluoro receptor is selected (the mag function may be disabled in programming; in this case DOSE will be displayed instead of MAG). The MAG number indicates the relative magnification of the I.I. The MAG display is replaced by the DOSE selection display for approximately 3 seconds after pressing the dose button. This indicates the relative dose; the actual dose value at each setting is determined by the installer. Back (<<) and next (>>) selection buttons. Press the back and next selection buttons to scroll forward or backwards through the display. The symbols << or >> will be displayed when these buttons are active. Reset button. button to return to the main menu when in one of the APR submenus. Press the Pressing this button will clear status and error messages. When the main APR menu is presented, this button accesses a submenu that allows RESET and TEST of the DAP function. Refer to DAP - DOSE AREA PRODUCT at the end of this chapter for further details.

5.

6. 7.

8.

9.

2003 Del Medical Imaging Corp.

Revision A

Page 25

Console Controls

DAP - DOSE AREA PRODUCT (OPTIONAL) The DAP function is only available if the optional DAP device is installed in the X-ray system, and if the DAP function has been enabled in programming. Any references made to tube 2 or the ability to switch tubes apply to two tube generators only. DAP Overview The DAP device(s) must be allowed to stabilize when the generator is switched on, and the first time when switching from tube 1 to tube 2 or from tube 2 to tube 1. During this time, when the DAP device is settling, the DAP display will indicate --------. X-Ray exposures may be made during this time, but the DAP function will be disabled during the settling period. The settling period may last up to a few minutes after the generator is switched on. The DAP device for the tube that is selected when the generator is switched on is automatically tested by the generator after the settling period. If the DAP device passes the self-test, the DAP readout will display its last setting. This indicates that the DAP is functional, and ready to measure the dose-area product. If the DAP self-test fails, an error message will be presented, and the DAP readout will display --------. This indicates that the DAP is not functional. When switching to the other tube for the first time in that session (from tube 1 to tube 2 or from tube 2 to tube 1), the DAP device associated with that tube will be tested. As in the previous step, the DAP readout will display its last setting to indicate that the DAP is functional, or -------- to indicate that the DAP is not functional. The installer may program the generator to present an audible warning, along with a visual alarm via the LCD display on the console, to indicate that preset limits for accumulated DAP or DAP rate have been reached. * DAP ACCUM WARN will display to indicate that the accumulated dose exceeds the programmed accumulated dose-area product limit. * DAP RATE WARN will display to indicate that the current dose rate exceeds the programmed dose rate limit. A reading of 0mGycm2 indicates that the DAP display has been manually reset as described later in this section, and that no exposures have been taken since it was reset. Always check local regulations to determine how frequently the DAP device must be tested. The procedure to test the DAP follows later in this section.

Note:

Note:

2003 Del Medical Imaging Corp.

Revision A

Page 26

Console Controls

DAP Display Before proceeding with initial dose-area product measurements, ensure that the 2 DAP display is 0mGycm . The DAP display may be manually reset to zero as described in DAP Reset / Test, below. The DAP measures and displays cumulative dose per unit area when in RAD mode. It may be programmed to measure and display the dose-area product for each tube individually, or to sum the readings from both tubes. This is installer programmable. In fluoroscopy mode, the DAP may be programmed to display accumulated dose2 2 area product (mGycm ), or DAP rate (mGycm /s). If programmed to display DAP rate, the DAP rate display will be in reverse video during fluoroscopic operation. When the fluoro run is finished, the DAP display will revert to normal video, and the accumulated dose will be displayed. The DAP rate feature for fluoro operation is installer programmable. The maximum accumulated reading on the DAP display is 9999999. An error message will be presented when this limit is reached.

DAP Reset / Test The DAP display(s) may be reset to zero, and a manual DAP functional test may be performed when desired, as described below. 1. In order to select the DAP device to test, select a receptor that is programmed for the desired tube / DAP device (tube 1 or tube 2). This will also select the DAP to reset if programmed for individual dose-area product readings for each tube. If the two DAP readings are summed, the reset function may be performed on either the tube 1 or tube 2 selection. The DAP display is common to both tubes, therefore the DAP display may be reset at either the tube 1 or tube 2 tube position. If the generator is programmed for individual DAP measurements for each tube, and you are intending to reset the DAP display, confirm that the correct tube / DAP device is selected before continuing. 2. Press the button at the main APR menu when DAP is displayed at the bottom center of the LCD screen. This will present the DAP reset / test menu. Press RESET to reset the DAP display to zero. Press TEST to test the DAP. The message DAP TEST: OK will be presented if the DAP has passed its test. Repeat for the other tube if desired.

2003 Del Medical Imaging Corp.

Revision A

Page 27

Console Controls

(This page intentionally left blank)

2003 Del Medical Imaging Corp.

Revision A

Page 28

REMOTE FLUORO CONTROLLER (optional)

The illustration below shows the remote fluoro controller. This unit allows the operator to select fluoroscopy functions at a location remote from the operator control console. The remote fluoro controller will be energized only when a fluoro technique is selected.

1. 2. 3. 4. 5. 6. 7. 8. 9. 10.

Fluoroscopy exposure indicator Fluoro display: kV, mA, time (min), PPS, MAG Remote ON indicator Decrease / Increase kV and mA buttons Time zero / accum time buttons Pulsed fluoro rate control Magnification control for image intensifier (I.I.) Aux button: selects delta dose Automatic brightness stabilization (ABS) select button Pulsed fluoro select button

2003 Del Medical Imaging Corp.

Revision A

Page 29

Remote Fluoro Controller

Fluoroscopy must be selected at the operator console to enable the remote fluoro control unit. The default kV and mA will display on both the operator control console and on the remote fluoro unit. REMOTE FLUORO CONTROLS AND DISPLAY There are four modes of operation for the unit: Manual: - operator controls the exposure ABS (Automatic Brightness Stabilization) PF (Pulsed Fluoro) optional PF with ABS (optional). kV selector and display In manual mode: To increase kV, press +. To decrease kV, push -. In ABS mode: The AUTO indicator lights and the kV select buttons are inactive. The display will indicate the kV during the exposure. mA selector and display In manual mode: To increase mA, press +. To decrease mA, press -. In ABS mode: The AUTO indicator lights and the mA select buttons are inactive. The display will indicate the mA during the exposure. Pulsed fluoro mode uses a fixed (not operator - adjustable) mA value.

2003 Del Medical Imaging Corp.

Revision A

Page 30

Remote Fluoro Controller

Time zero and accumulated time (minutes) display To reset the time display to zero, press ZERO. To display accumulated fluoro exposure time, press ACCUM. The ACC indicator will light. To reset the accumulated time, press ZERO when the accumulated time is displayed.

Pulsed fluoro frame rate selector and display To increase the PPS (Pulses Per Second) rate, press +. To decrease the PPS rate, press -. Pulse rates are 3, 7.5, 15, and 30 PPS for 60 Hz based cameras, and 2.5, 6.25, 12.5, and 25 PPS for 50 Hz based cameras (6.25 is displayed as 6, and 12.5 is displayed as 12).

PPS

Mag (magnification) selector and display To increase the I.I. (Image Intensifier) magnification, press +. To decrease magnification, press -. Aux Button (Delta Dose Selector) Press this button to select delta dose when ABS is used. The ABS indicator will indicate the selected dose: No flash indicates dose 0, slow flashing indicates dose 1, and fast flashing indicates dose 2. The remote fluoro control will beep at a slow rate on dose 1, and at a fast rate on dose 2 during a fluoroscopy exposure.

2003 Del Medical Imaging Corp.

Revision A

Page 31

Remote Fluoro Controller

ABS (Automatic Brightness Stabilization) Selector Press this button to select ABS. The adjacent indicator will light. Pressing the button to deselect ABS will extinguish the indicator, allowing fluoro kV and mA to be set manually.

PF (Pulsed Fluoro) Select Button Selects pulsed fluoro. Only active for optional pulsed fluoro function.

Fluoroscopy Exposure Indicator This indicator will light during fluoroscopy exposure. This is accompanied by an audible tone from the console. Remote ON Indicator When fluoro mode is selected, REMOTE will light.

2003 Del Medical Imaging Corp.

Revision A

Page 32

kV - mAs REDUCTION FOR PERIPHERAL ANGIOGRAPHY (optional)

When interfaced to a special procedures stepping table, this option enables the Indico 100 generator to sequentially select varying radiographic techniques as the patient is moved over the image receptor. The kV and / or the mAs may be set to track changes in the anatomy of the patient for each of the table steps. When kV / mAs stepping is enabled, five kV / mAs sequences will be available, with each sequence containing six kV / mAs steps. The kV and mAs is programmable for each step as described later in this section. SELECTING THE TABLE STEPPER FUNCTION 1. Select the image receptor that is configured for peripheral angiography. The radiographic display will indicate a preselected and possibly arbitrary kV and mAs value. The first exposure in the exposure sequence will not necessarily use those kV and mAs values. The kV and mAs used for the exposure sequence will be as per steps 1 to 6 in the console LCD display window. The LCD display window will display the first kV - mAs stepping screen (PER. SHIFT1) as shown. The sample screens below do not show the secondary displays i.e. tube number, %HU, time etc. Also, the kV and mAs values may differ from the sample screens. * PER. SHIFT1 * 1- 80kV / 2- 75kV / 3- 72kV / 4- 72kV / 25mAs 25mAs 16mAs 16mAs 5- 68kV / 8mAs 6- 64kV / 4mAs

2.

>>

3.

Press >> to select the second kV - mAs stepping screen (PER. SHIFT2) as shown below. This may be programmed for different kV / mAs steps from the previous screen. * PER. SHIFT2 * 1- 80kV / 2- 75kV / 3- 72kV / 4- 72kV / 25mAs 25mAs 16mAs 16mAs 5- 68kV / 8mAs 6- 64kV / 4mAs

>>

2003 Del Medical Imaging Corp.

Revision A

Page 33

5
4.

kV - mAs Reduction for Peripheral Angiography

Press >> to select the remaining kV - mAs stepping screens (PER. SHIFT3 to PER. SHIFT5). These screens may be programmed for different kV / mAs steps from the previous screens. Press << to scroll backwards through the available kV - mAs stepping screens.

5.

RUNNING THE TABLE STEPPER FUNCTION 1. Select the desired kV - mAs stepping screen PER. SHIFT1 to PER. SHIFT5 as previously described. If required, edit the kV / mAs values as described later in this section for the selected exposure sequence. Press and hold the prep / expose button on the stepping table. This will cause exposure number 1 in the console LCD display window to be highlighted, and the kV and mAs in the radiographic display window to be updated. Exposure number 1 will be made when the generator receives the expose command. After the table shifts, the next exposure will be highlighted in the LCD display. That exposure will be made using revised kV / mAs values, if appropriate, when the generator receives the expose command. Continue the exposure sequence by holding the prep / expose button. Do not release the prep button until finished, doing so will abort the exposure sequence. When finished the exposure sequence, release the prep button. This will reset the exposure step counter to number 1.

2.

3.

4. 5.

EDITING THE TABLE STEPPING kV / mAs 1. 2. 3. Select the kV or mAs value to change by pressing the button adjacent to the desired kV / mAs step. The desired selection will be framed. Select the new kV or mAs value by pressing the + or - kV or mAs buttons in the radiographic display window. Press SAVE under the LCD display window to save the changes, or to exit without saving the changes. Repeat to change the kV or mAs for other steps in the same screen, or in one of the other kV - mAs stepping screens.

2003 Del Medical Imaging Corp.

Revision A

Page 34

PROGRAMMING THE CONSOLE


MAKING CHANGES TO PRE-PROGRAMMED APR TECHNIQUES

The default techniques associated with the anatomical views in the APR display window may be edited if desired. To do this, follow the steps in this chapter. ENTERING INTO PROGRAMMING MODE 1. 2. 3. Start with the generator switched OFF. Press and hold the console. is displayed, press the button while pressing the power ON button on the operator

When the console finishes its power-on sequence, and the ENTER PASSWORD prompt button 4 times to access the following menu: * GENERATOR SETUP * APR EDITOR: DISABLED

UTILITY

EXIT SETUP

Access is now provided to the following groups of functions. These will be discussed in sequence. UTILITY Set Time & Date Error Log Statistics Console Enables/disables changes to the APR techniques Returns to normal operation i.e. non programming mode.

APR EDITOR EXIT SETUP

2003 Del Medical Imaging Corp.

Revision A

Page 35

Programming the Console

UTILITY MENU To select the UTILITY menu, press the selection button adjacent to UTILITY in the APR display window. This allows access to the following functions:

* UTILITY * SET TIME & DATE ERROR LOG STATISTICS CONSOLE EXIT SET TIME & DATE ERROR LOG STATISTICS Allows setting of the time and date Allows the error log to be reviewed Allows review of the tube exposure counter(s), the fluoro exposure counter if applicable, and the generator accumulated exposure counter. Allows console parameters to be set Returns to the GENERATOR SETUP menu (previous page)

CONSOLE EXIT

Set Time and Date * SET TIME & DATE * YEAR: 1997 MONTH: 1 DAY: 1 EXIT Use these steps to set the time and date. 1. 2. 3. From the UTILITY menu, select SET TIME & DATE. Select the parameter to change. Press the + or - buttons to set the desired value. Time is set in 24 hour format. Select EXIT to return to the UTILITY menu. HOUR: MIN: 16 35 + -

2003 Del Medical Imaging Corp.

Revision A

Page 36

Programming the Console

Error Log * ERROR LOG * ERROR # 10 OF 100 DATE: 09-12-2000 ERROR CODE: 30 ERROR MESSAGE: ROTOR FAULT EXIT Use these steps to view the error log. 1. 2. From the UTILITY menu select ERROR LOG. Select ERROR # and press the + or - buttons to scroll through the error log. The error code, error message, date and time of the error will be displayed in the APR display window, and the associated parameters will be displayed on the console displays. Select EXIT to return to the UTILITY menu. TIME: 13:09 + -

3.

Statistics * STATISTICS * TUBE 1 EXP: TUBE 2 EXP: FLUORO HOURS: TOTAL EXP: EXIT 0 0 0 0

The statistics menu shows the tube exposure count, accumulated fluoro hours, and accumulated generator exposure count. 1. 2. 3. From the UTILITY menu select STATISTICS. The statistics (tube 1 exposures, tube 2 exposures if applicable, fluoro hours if applicable, total generator exposures) may now be viewed. Select EXIT to return to the UTILITY menu.

2003 Del Medical Imaging Corp.

Revision A

Page 37

Programming the Console

Console * CONSOLE * SLOW KEY REPEAT: 200MS MED KEY REPEAT: 150MS FAST KEY REPEAT: 75MS SPEAKER VOLUME: 15 EXIT

LCD SCREEN APR MODE: NO + -

The CONSOLE menu displays console parameters that may be changed to suit operator preferences. 1. 2. 3. 4. 5. 6. 7. From the UTILITY menu select CONSOLE. Select SLOW KEY REPEAT. Press the + or - button to select the speed at which displays change while the button is pressed for the first 5 counts. Select MED KEY REPEAT. Press the + or - button to select the speed at which displays change while the button is pressed for the next 5 counts. Select FAST KEY REPEAT. Press the + or - button to select the speed at which displays change while the button is pressed after 10 counts. Select SPEAKER VOLUME. Press the + or - button to set the speaker volume in the range 1 to 15. Select LCD SCREEN. Press button to toggle between normal and reverse display for the LCD display screen. Select APR MODE. Press selection button to toggle between NO and YES. Selecting YES disallows changes to the / mAs / mA/ms selections when in normal operating / mAs / mA/ms in normal mode. This must be set to NO to allow switching between operating mode. 8. 9. Select EXIT to return to the UTILITY menu. Select EXIT to return to the GENERATOR SETUP menu.

2003 Del Medical Imaging Corp.

Revision A

Page 38

Programming the Console

APR EDITOR * GENERATOR SETUP * APR EDITOR: DISABLED

UTILITY

EXIT SETUP

Use these steps to edit the programmed APR techniques. NOTE: BEFORE MAKING CHANGES TO THE DEFAULT APR TECHNIQUES, PLEASE RECORD ALL CURRENT DEFAULTS. THESE DEFAULTS MAY BE RECORDED IN COPIES OF THE APR PROGRAMMING WORKSHEET AT THE END OF THIS CHAPTER. 1. 2. From the GENERATOR SETUP menu select APR EDITOR. Press the APR EDITOR button to toggle between ENABLED and DISABLED. DISABLED does not allow editing and then saving the changed APR technique(s). ENABLED allows the operator to change the default APR techniques, and then save the changes. The APR will then default to the newly saved techniques. The generator stores the last APR editor setting prior to being switched OFF. If the APR editor was previously ENABLED, APR changes may subsequently be made and then saved in normal operating mode without the need to manually set the APR editor to ENABLED. To disable this feature, the APR editor must be set to DISABLED. With the APR EDITOR set to ENABLED, return to normal operation by pressing EXIT SETUP. The following menu is displayed. This is a typical menu, actual anatomical views may have been reprogrammed, and the screen may differ slightly depending on generator model. TUBE 1 0%HU SKULL THORAX ABDOMEN PELVIS 160MS * APR TABLES * LOWER EXTREMITY UPPER EXTREMITY G.I. SPINE

3. 4.

2003 Del Medical Imaging Corp.

Revision A

Page 39

6
5. 6. 7. 8.

Programming the Console

Perform the APR technique change as outlined below. Select the desired APR menu to be edited; for example SKULL. Select the desired APR view to edit, for example SKULL AP/PA, vertical image receptor. Select the patient size to edit within that view, for example medium. All techniques with the exception of kV will be common to all patient sizes. The common techniques for all patient sizes within that view should be programmed first, then the desired kV for the selected patient size. After these techniques are saved per the remainder of this section, kV values may be programmed if desired for the other patient sizes in the same APR view. The default selections are shown on the console, for example. 70 kV, 100 mA, 20 ms, technique = mA/ms, large focus.

9.

10. Change the technique to the desired value(s). Parameters that may be edited are: kV, mA/mAs, time (ms), technique select ( / mAs / mA/ms), focus, image receptor, plus film screen, density and field selection in AEC mode. The APR view being edited will be highlighted with a rectangular outline as soon as any parameter is changed within that view. Also the word SAVE will appear in the display window above the reset button. 11. Continue making all desired APR changes within that view. 12. When finished, press the button. This is the first step necessary to save the changes. This also allows the option of canceling the changes in the next step. 13. If you do not wish to save the changes, press EXIT to abort the SAVE function. To save the changes, press the button under SAVE. The message SAVING DATA... will briefly appear. 14. Press to return to the main APR screen.

15. You may now make changes to other APR techniques if desired by repeating the above steps 6 to 14. 16. It is suggested that the APR editor be set to DISABLED as described earlier in this section, in order to avoid unintended APR changes in the future. MAKING CHANGES TO THE APR TEXT The anatomical sections and views as displayed in the APR window may be altered to suit your requirements, but you will need to contact your service representative to have this done. Use copies of the worksheet on the next page to record your desired changes. One sheet should be used per APR view, the master copy should be photocopied as many times as necessary to record the required changes.

2003 Del Medical Imaging Corp.

Revision A

Page 40

Programming the Console APR PROGRAMMING WORKSHEET (THIS IS A MASTER, MAKE WORKING COPIES AS NEEDED)

__________________ __________________ __________________ __________________

__________________ __________________ __________________ __________________

MENU 1 APR TABLES MAX 16 CHARACTERS

__________________ __________________ __________________ __________________

__________________ __________________ __________________ __________________

MENU 2 ANATOMICAL VIEW MAX 16 CHARACTERS

kV mA/mAs ms DENSITY

Table with Bucky Receptor Table non Bucky Receptor Vertical Image Receptor Tomography R/F (Fluoro)

Patient Size

Small Medium Large

Auxiliary Image Receptor

Tech Select mAs mA/ms Film Screen

Focus

Field Select

2003 Del Medical Imaging Corp.

Revision A

Page 41

Programming the Console

(This page intentionally left blank)

2003 Del Medical Imaging Corp.

Revision A

Page 42

ERROR CODES AND MESSAGES

The Indico 100 console will display status messages on the APR display during normal and abnormal operation of the generator. This chapter contains tables of those messages and suggests actions to be taken should any malfunctions occur. Messages may be cleared by pressing the RESET button. Operator Messages These messages indicate the status of the generator. No action is required. MESSAGE INITIALIZATION SPINNING ROTOR X-RAY READY X-RAY ON DAP NOT READY DESCRIPTION Displayed during power up sequence. Displayed when prep state is active. Displayed when generator is ready to expose Displayed during both a rad and fluoro exposure. The optional DAP is in its warm up state, and not ready to make DAP measurements.

Limit Messages These messages indicate that an exposure has been requested that exceeds one or more limits. MESSAGE TUBE KV LIMIT GEN KV LIMIT TUBE MA LIMIT GEN MA LIMIT TUBE KW LIMIT GEN KW LIMIT TUBE MAS LIMIT GEN MAS LIMIT GEN MS LIMIT PROBLEM Requested kV not allowed as tube kV limit has been reached. Requested kV not allowed as generator kV limit has been reached. Requested mA not allowed as tube mA limit has been reached. Requested mA not allowed as generator mA limit has been reached. Requested parameter not allowed as tube kW limit has been reached. Requested parameter not allowed as generator kW limit has been reached. Requested mAs not allowed as tube mAs limit has been reached. Requested mAs not allowed as generator mAs limit has been reached. Requested ms not allowed as generator ms limit has been reached. None. None. None. None. None. None. None. None. None. ACTION

2003 Del Medical Imaging Corp.

Revision A

Page 43

Error Codes and Messages

MESSAGE CAL LIMIT GEN. PPS LIMIT AEC DENSITY LIM

PROBLEM Requested parameter not calibrated. PPS limit has been reached. Requested density not programmed.

ANODE HEAT WARN FL TIMER WARN INVALID PARAM GEN DUTY WARNING

Anode has exceeded programmed warning level. Fluoro interval timer 5.0 mins. Generator detected invalid parameter within received message, message ignored. The X-ray generator has reached its duty warning limit.

ACTION Consult your service representative. None. Select another density or request those steps be programmed. Consult your service representative for this. Wait for anode to cool. Reset fluoro timer. Select valid parameter. Re-evaluate technique factors. Allow generator to cool if possible. If exposures are continued, serious generator damage may result due to overheating. Reset the DAP. Reduce the dose.

DAP ACCUM WARN DAP RATE WARN

The accumulated DAP value has reached the programmed DAP limit. The current DAP rate exceeds the programmed DAP rate limit.

2003 Del Medical Imaging Corp.

Revision A

Page 44

Error Codes and Messages

Error Messages These messages indicate that an error has occurred.


ERROR CODE E001 E003 E004 MESSAGE GEN EPROM ERR GEN NVRAM ERR GEN RTC ERROR Generator CPU real time clock is not functioning. PROBLEM ACTION Consult your service representative. Consult your service representative. Reset time and date. Consult your service representative. Consult your service representative. Power unit off and retry rotor start. Consult your service representative. Consult your service representative. Consult your service representative. Retry exposure. Consult your service representative. Consult your service representative. Consult your service representative. Re-take exposure if necessary. Consult your service representative. Check exposure technique settings. Consult your service representative. Check exposure technique settings. Consult your service representative.

E005 E006

PS CONTACT ERR ROTOR FAULT

E007 E008 E009

FILAMENT FAULT MA FAULT PS NOT READY Power supply is not ready to make an exposure.

E011 E012 E013

HIGH MA FAULT LOW MA FAULT MANUAL TERMIN Operator released exposure switch during exposure.

E014

AEC BUT ERROR

AEC exposure exceeded allowed back up time.

E015

AEC BU MAS ERR

AEC exposure exceeded allowed back up mAs.

2003 Del Medical Imaging Corp.

Revision A

Page 45

Error Codes and Messages

ERROR CODE E016

MESSAGE TOMO BUT ERROR

PROBLEM Tomo exposure exceeded back up time.

ACTION 1. Check exposure technique settings. 2. Increase tomo back up time if necessary. Consult your service representative. Reduce length of time in prep state. Reduce parameters or wait for tube to cool. Wait for X-ray tube # 1 to cool. Wait for X-ray tube # 2 to cool. Close door. Check collimator. Check cassette. Check II safety. Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative. Turn power off and then on to Reset Generator. Consult your service representative. Turn power off and then on to reset generator. Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative.

E017 E018 E019

NOT CALIBRATED PREP TIMEOUT ANODE HEAT LIMIT Generator has been in prep state too long. Selected parameters will cause X-ray tube to exceed its programmed anode heat limit. X-ray tube # 1 too hot and its thermal switch has opened. X-ray tube # 2 too hot and its thermal switch has opened. Door is open. Collimator is not ready. Cassette is not ready. II safety is not ready. Spare input is not ready.

E020 E021 E022 E023 E024 E025 E026 E028 E029 E030 E031

THERMAL INT #1 THERMAL INT #2 DOOR INTERLOCK COLLIMATOR ERR CASSETTE ERROR II SAFETY INT SPARE INT PREP SW CLOSED X-RAY SW CLOSED FLUORO SW CLOSED REMOTE COMM ERR

Communication error detected with remote fluoro control unit.

E032

CONSOLE COMM ERR

Generator has detected error in communication to console.

E033 E034 E035

GEN BATTERY LOW +12VDC ERROR -12VDC ERROR

2003 Del Medical Imaging Corp.

Revision A

Page 46

Error Codes and Messages

ERROR CODE E036 E037 E038 E039 E040 E041 E042 E043

MESSAGE +15VDC ERROR -15VDC ERROR CAL DATA ERROR AEC DATA ERROR FLUORO DATA ERROR REC DATA ERROR TUBE DATA ERR KV ERROR

PROBLEM

E044 E045 E046 E047 E048 E049 E050 E051

COMM ERROR NOT SUPPORTED MODE INHIBITED FL TIMER LIMIT FOCUS MISMATCH NOT ENABLED GEN DATA ERROR AEC DEVICE ERR

Received communication message not valid and ignored. Received message valid, but not supported by this system. Received message valid, but not allowed during present state. Fluoro Timer has exceeded time limit. Requested function has not been programmed to be enabled. Generator has detected no feedback signal from AEC device.

ACTION Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative.. Switch OFF generator. Prevent further use of generator, then. consult your service representative. Reset error. Reset error. Reset error. Reset Fluoro timer. Consult your service representative. Consult your service representative. Consult your service representative. Check that X-ray tube is pointing at correct AEC device. Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative. Select AEC field(s).

E052 E053 E054 E055

HIGH SF CURRENT HIGH LF CURRENT AEC OUT OF RANGE NO FIELDS ACTIVE AEC enabled but no fields are selected.

2003 Del Medical Imaging Corp.

Revision A

Page 47

Error Codes and Messages

ERROR CODE E056 E057 E058 E059 E060 E061 E062 E063 E066 E070 E071 E072 E073

MESSAGE NO TUBE SELECTED AEC STOP ERROR CONSOLE BUT ERR HOUSE HEAT LIMIT EXP KV HIGH EXP KV LOW EXP_ SW ERROR FACTORY DEFAULTS NO SYNC PULSE SOFTWARE KEY ERR DAP OVERFLOW DAP DEVICE ERROR DAP DATA ERROR

PROBLEM

X-ray tube housing has exceeded limit.

ACTION Consult your service representative. Consult your service representative. Consult your service representative. Wait for tube to cool.

Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative. Consult your service representative. The accumulated DAP value Reset the DAP. exceeds the display limit. The DAP device is not Consult your service functional. representative. Consult your service representative.

2003 Del Medical Imaging Corp.

Revision A

Page 48

GENERATOR EXPOSURE TABLES

Table 1 following shows mAs values resulting from preselected mA and ms values (3 point operation). This table also shows the range and interrelation of these loading factors. For example, if in AEC operation it is known that the preset backup mAs limit is 50 mAs and 100 mA is chosen for that technique, then the maximum length of exposure possible (at which time the backup mAs limit is reached) is 500 ms. In the above example, read down the 100 mA column to the 50 mAs value. Then read across to the left side of the page to the 500 ms value, this will be the maximum time allowed for the stated conditions. Please review the note at the bottom of page 56. This details AEC backup options that are available. IEC 60601-2-7 requires that this information be provided in a form suitable to be displayed at a prominent location on or near the control console of the generator. To this end, an extra copy of table 1 is provided on pages 52 and 53 which may be photocopied as required, or removed from this manual and placed in a suitable location as per local requirements.

2003 Del Medical Imaging Corp.

Revision A

Page 49

Generator Exposure Tables INDICO 100 GENERATOR TECHNIQUE SELECTION (3 Point Selection: mA/TIME & mAs per IEC) Time (ms) 1.0 1.2 1.6 2 2.5 3.2 4 5 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400 500 25 32 40 50 63 80 100 125 mA Selected 160 200 250 320 400 .5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 500 .5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 630 .63 .8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 800 .8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400 1000 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400 500

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160

Table 1: mAs values vs. mA & time selected Table 1 continued on next page

2003 Del Medical Imaging Corp.

Revision A

Page 50

Generator Exposure Tables

Time (ms) 630 800 1000 1250 1600 2000 2500 3200 4000 5000 6300

25 16 20 25 32 40 50 63 80 100 125 160

32 20 25 32 40 50 63 80 100 125 160 200

40 25 32 40 50 63 80 100 125 160 200 250

50 32 40 50 63 80 100 125 160 200 250 320

63 40 50 63 80 100 125 160 200 250 320 400

80 50 63 80 100 125 160 200 250 320 400 500

100 63 80 100 125 160 200 250 320 400 500 630

125 80 100 125 160 200 250 320 400 500 630 800

mA Selected 160 200 100 125 125 160 160 200 200 250 250 320 320 400 400 500 500 630 630 800 800 1000 1000

250 160 200 250 320 400 500 630 800 1000

320 200 250 320 400 500 630 800 1000

400 250 320 400 500 630 800 1000

500 320 400 500 630 800 1000

630 400 500 630 800 1000

800 500 630 800 1000

1000 630 800 1000

Table 1 (Cont): mAs values vs. mA & time selected

Note:

32 kW Generators 40 kW Generators 50 kW Generators 65 kW Generators 80 kW Generators 100 kW Generators

Maximum mA = 400 Maximum mA = 500 Maximum mA = 630 Maximum mA = 800 Maximum mA = 1000 Maximum mA = 1000

Note:

1) 2)

mA/TIME values are tube rating dependent. For certain tubes, some mA/TIME selections are not available at higher kV selections. The maximum power output of the Indico 100 family of generators is 32 kW, 40 kW, 50 kW, 65 kW, 80 kW, or 100 kW respectively. The generator software does not allow a kV and mA combination to be selected that exceeds these maximum values.

2003 Del Medical Imaging Corp.

Revision A

Page 51

Generator Exposure Tables

Time (ms) 1.0 1.2 1.6 2 2.5 3.2 4 5 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400 500

25

32

40

50

63

80

100

125

mA Selected 160 200

250

320

400 .5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125

.5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160

500 .5 .63 .8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250

630 .63 .8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320

800 .8 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400

1000 1.0 1.2 1.6 2.0 2.5 3.2 4.0 5.0 6.3 8 10 12.5 16 20 25 32 40 50 63 80 100 125 160 200 250 320 400 500

Table 1: mAs values vs. mA & time selected Table 1 continued on next page

2003 Del Medical Imaging Corp.

Revision A

Page 52

Generator Exposure Tables

Time (ms) 630 800 1000 1250 1600 2000 2500 3200 4000 5000 6300

25 16 20 25 32 40 50 63 80 100 125 160

32 20 25 32 40 50 63 80 100 125 160 200

40 25 32 40 50 63 80 100 125 160 200 250

50 32 40 50 63 80 100 125 160 200 250 320

63 40 50 63 80 100 125 160 200 250 320 400

80 50 63 80 100 125 160 200 250 320 400 500

100 63 80 100 125 160 200 250 320 400 500 630

125 80 100 125 160 200 250 320 400 500 630 800

mA Selected 160 200 100 125 125 160 160 200 200 250 250 320 320 400 400 500 500 630 630 800 800 1000 1000

250 160 200 250 320 400 500 630 800 1000

320 200 250 320 400 500 630 800 1000

400 250 320 400 500 630 800 1000

500 320 400 500 630 800 1000

630 400 500 630 800 1000

800 500 630 800 1000

1000 630 800 1000

Table 1 (Cont): mAs values vs. mA & time selected

Note:

32 kW Generators 40 kW Generators 50 kW Generators 65 kW Generators 80 kW Generators 100 kW Generators


1) 2)

Maximum mA = 400 Maximum mA = 500 Maximum mA = 630 Maximum mA = 800 Maximum mA = 1000 Maximum mA = 1000

Note:

mA/TIME values are tube rating dependent. For certain tubes, some mA/TIME selections are not available at higher kV selections. The maximum power output of the Indico 100 family of generators is 32 kW, 40 kW, 50 kW, 65 kW, 80 kW, or 100 kW. respectively. The generator software does not allow a kV and mA combination to be selected that exceeds these maximum values.

2003 Del Medical Imaging Corp.

Revision A

Page 53

Generator Exposure Tables

(This page intentionally left blank)

2003 Del Medical Imaging Corp.

Revision A

Page 54

MAINTENANCE SCHEDULE

This section is organized into two parts. The first part of this section details tests by which the operator can verify that the AEC and ABS circuits are functioning, and that there are no apparent defects in the console or associated equipment. These tests should be done monthly as a minimum, or more frequently if required by local regulations.
CAUTION: THE AEC AND ABS VERIFICATION PROCEDURES REQUIRE THE PRODUCTION OF XRAYS. OBSERVE CORRECT OPERATING PROCEDURES, AND TAKE APPROPRIATE PRECAUTIONS AGAINST X-RADIATION.

The second part of this section is the recommended maintenance schedule for your generator. This maintenance schedule is recommended for all Indico 100 series X-ray generators.
WARNING: REFER ALL SERVICING TO APPROPRIATE SERVICE PERSONNEL WHO ARE TRAINED TO SERVICE THIS EQUIPMENT AND WHO ARE FAMILIAR WITH THE POTENTIAL HAZARDS ASSOCIATED WITH THIS EQUIPMENT. MAINTENANCE SCHEDULE FREQUENCY MAY BE DETERMINED BY CERTAIN REGULATORY REQUIREMENTS OF THE COUNTRY OR STATE IN WHICH THE INSTALLATION IS LOCATED. ALWAYS CHECK THE LOCAL CODES AND REGULATIONS WHEN DETERMINING A MAINTENANCE SCHEDULE.

NOTE:

2003 Del Medical Imaging Corp.

Revision A

Page 55

Maintenance Schedule

AEC FUNCTIONAL CHECK The following procedure may be used to verify that the AEC circuits are functioning on generators equipped with AEC (automatic exposure control). 1. 2. Switch the generator ON, and select an appropriate radiographic image receptor. Align the X-ray tube and the selected image receptor such that the central ray is directly over the center field of the AEC pickup device. Set the focal spot to film plane distance to 40 in. (1 m). Select AEC mode of operation. Select center field, large focus. MINIMUM EXPOSURE TIME: With no object in the radiation field, adjust the collimator or beam limiting device to project a 10 in. X 10 in. (24 cm X 24 cm) field at the image receptor. Select 80 kVp, 100 mA, and a backup mAs of 50 if it is operator selectable. If this is not operator selectable, the default AEC backup settings must be used. Refer to the note at the end of this page to determine the AEC backup mode that has been programmed. Make an exposure and verify that the POST mAs reading is 2 mAs. MAXIMUM EXPOSURE TIME: Close the collimator or beam limiting device completely. Place a folded lead apron over the image receptor. Select 60 kVp, 100 mA, and a backup mAs of 50 if it is operator selectable. If this is not operator selectable, the default backup settings must be used. Refer to the note at the end of this page to determine the AEC backup mode that has been programmed. Make an exposure and verify that the error message AEC BUT ERROR or AEC BU MAS ERR is indicated in the APR display window after the exposure has terminated. This confirms that the exposure has continued until it was terminated by the AEC backup timer.

3. 4. 5.

6. 7. 8.

9.

NOTE: The AEC backup mode is installer programmable. Three modes are available, FIXED, MAS, and MS. Only the MAS mode allows the operator to set the backup mAs for an AEC exposure. FIXED: The generator will determine the maximum AEC backup time, not to exceed preset AEC backup mAs/ms values or system limits. The characters AEC will be displayed in the time window of the LED display during AEC operation. MAS: Allows the operator to adjust the AEC backup mAs, not to exceed preset AEC backup mAs/ms values or system limits. The mAs value will be displayed in the time window of the LED display during AEC operation. MS: Allows the operator to adjust the AEC backup ms, not to exceed preset AEC backup mAs/ms values or system limits. The ms value will be displayed in the time window of the LED display during AEC operation.

2003 Del Medical Imaging Corp.

Revision A

Page 56

Maintenance Schedule

ABS FUNCTIONAL CHECK The following procedure may be used to verify that the ABS circuits are functioning on generators equipped with ABS (automatic brightness stabilization). 1. 2. 3. Switch the generator ON, and select an appropriate fluoroscopic image receptor. Select ABS mode. The indicator LED adjacent to the ABS selector button on the control console or on the remote fluoro control will be lit to indicate that ABS is ON. Open the collimator or beam limiting device shutters to maximum. With no object in the image field, press the fluoro footswitch. Confirm that the collimator or beam limiting device is fully open by observing the monitor. While continuing to press the fluoro footswitch, note the fluoro kV display at the control console or remote fluoro control. This should be less than 55 kV. Close the collimator or beam limiting device and press the fluoro footswitch. With no object in the image field, ensure that the radiation is fully blocked by observing the monitor. If necessary, cover the image intensifier with lead aprons to block all radiation. While continuing to press the fluoro footswitch, note the fluoro kV display at the control console or remote fluoro control. This should be the maximum available fluoro kV, 110 kV or 125 kV (depending on whether the upper fluoro kV limit has been set to 110 kV or 125 kV). Place an absorber (20 cm of water, or 1.5 in. (3.8 cm) of pure aluminum, or equivalent) in the center of the image field. Press the fluoro footswitch. While viewing the absorber on the monitor, open the collimator or beam limiting device sufficiently to cover as much of the absorber as possible. Ensure that the X-ray beam does not extend beyond the sides of the absorber as stray radiation will adversely affect the kVp reading in the next step. While continuing to press the fluoro footswitch, note the fluoro kV display at the control console or remote fluoro control. This should stabilize between 70 and 80 kVp.

4. 5.

6.

7. 8.

9.

CONSOLE AND MISCELLANEOUS GENERATOR CHECKS The user must routinely check the X-ray equipment for apparent defects or damage. Report any apparent defects or irregular operation of any equipment to service personnel immediately, and discontinue use of the suspected faulty equipment until repairs are made. Continuing operation with faulty equipment may present various safety hazards, including risk of increased radiation. 1. 2. 3. Check all visual displays (warning and status lights, technique displays and indicators, etc) for normal operation. Check all audible indicators (speakers, beepers, etc) for normal operation, and check that the loudness settings are adequate for the environment. Check all interconnect cables and connectors for damage.
Revision A Page 57

2003 Del Medical Imaging Corp.

Maintenance Schedule

MAINTENANCE ACTIVITIES IN THE TABLE BELOW ARE TO BE PERFORMED BY AUTHORIZED SERVICE PERSONNEL ONLY. HOWEVER THE OWNER OR OPERATOR OF THE EQUIPMENT MUST SCHEDULE THE SUGGESTED MAINTENANCE ACTIVITY WHEN REQUIRED BY CONTACTING THEIR LOCAL SERVICE AGENCY. Maintenance Frequency As governed by local regulations Every 6 Months AND whenever a related certifiable X-ray component is replaced: Description of Preventative Maintenance Check and recalibrate the DAP meter in the generator. 1. 2. Clean and re-grease all HV connections using vapour proof compound. Clean the control console, remote fluoro control (if used) and main cabinet as needed. REFER TO NEXT PAGE, CLEANING, BEFORE PROCEEDING Perform the X-ray tube auto calibration routine. Verify the calibration of the generator. Perform any additional tests required by laws governing this installation. Examine the following for any visible damage and replace any damaged components: 2. The exterior of the control console and remote fluoro control if used, including the membrane switch assembly. The cable between the control console and the generator main cabinet and between the remote fluoro control (if used) and generator main cabinet. The hand switch and fluoro footswitch (if used) and the cables connecting these to the console.

3. 4. 5. Every 12 months: 1.

Open the generator cabinet and examine the unit for any visible damage: missing or loose ground connections, oil leaks, damaged cables etc.

3. Every 5 years:

Ensure that there are no obstructions blocking the ventilation holes at the top of the cabinet, or the louvers on the front door panels. Replace the lithium batteries on the CPU board in the control console and on the generator CPU board in the main cabinet with the same manufacturer and type as identified on the original batteries.

2003 Del Medical Imaging Corp.

Revision A

Page 58

Maintenance Schedule

CLEANING Never use anything other than soap and water to clean plastic surfaces. Other cleaners may damage the plastic. Never use any corrosive, solvent or abrasive detergents or polishes. Ensure that no water or other liquid can enter any equipment. This precaution prevents short circuits and corrosion forming on components. Methods of disinfection used must conform to legal regulations and guidelines regarding disinfection and explosion protection. If disinfectants are used which form explosive mixtures of gases, these gases must have evaporated before switching on the equipment again. Disinfection by spraying is not recommended because the disinfectant may enter the X-ray equipment. If room disinfection is done with an atomizer, it is recommended that the equipment be switched OFF, allowed to cool down and covered with a plastic sheet. When the disinfectant mist has subsided, the plastic sheet may be removed and the equipment be disinfected by wiping.

2003 Del Medical Imaging Corp.

Revision A

Page 59

Maintenance Schedule

(This page intentionally left blank)

2003 Del Medical Imaging Corp.

Revision A

Page 60

X-RAY TUBE DATA


CAUTION:

10

NOTE TO THE INSTALLER: THE GENERATOR MUST BE PROGRAMMED NOT TO EXCEED THE MAXIMUM OPERATING LIMITS OF THE X-RAY TUBE.

The following pages contain data for the X-ray tube(s) used with your X-ray generator at the time of installation.

2003 Del Medical Imaging Corp.

Revision A

Page 61

10

X-Ray Tube Data

(This page intentionally left blank)

2003 Del Medical Imaging Corp.

Revision A

Page 62

X-Ray Tube Data

10

NOTE TO INSTALLER: INSERT TUBE DATA HERE.

2003 Del Medical Imaging Corp.

Revision A

Page 63

10

X-Ray Tube Data

(This page intentionally left blank)

2003 Del Medical Imaging Corp.

Revision A

Page 64

You might also like