Z Axis Calibration:
Z=1 Before .99992
Tolerance: +/-.0004 Z=4 Before After 4.00004 same Tolerance: +/- .0004 N/A Before After N/A same N/A same N/A same N/A same N/A same N/A same N/A same N/A same N/A same N/A same
Range: 1-7" X=5.0" Before After N/A same N/A same N/A same N/A same N/A same N/A same N/A same N/A same N/A same N/A same
N/A
N/A
The above Video System was calibrated by comparison to working standards which were compared to or reproduced from reference standards traceable to the N.I.S.T. This machine was calibrated to the OEM Procedure and specification. All clauses in the specification are met. This certificate shall not be reproduced except in full, without the approval of Certified Comparator Products, Inc. The machine was calibrated in compliance with ISO/IEC 17025, ISO 10012.1, and ANSI/NCSL Z540-1. Master Used For Calibration:P/N: CCP14707 S/N: 101810 The due date for this machine was set at the discretion of the customer requesting the calibration.
Measured Temperature during Calibration: Min; 74.2F / Max; 74.6F. Humidity N/A%RH
_______________________________________
Customers Signature Comments:
_______________________________________
Technicians Signature
Form: CPF-002 MFR, Rev. 1, 9/21/09 Page 1 of 3
The machine is within the limits of probability and the product conforms with the specification The machine is within the limits of probability and the product does not conform with the specification Certified Comparator Products, Inc. does not make a declaration of conformity; established by the customer