DPD CHLORINE TEST KIT


MODEL LP-1-NMC CODE 6901-NMC
QUANTITY CONTENTS CODE

50 50 50 50 3 1

*Chlorine DPD #1R Tablets (*6999A) Chlorine DPD #3R Tablets (*6905A) *Chlorine DPD #2R Tablets (6904A)

FMC6905A6999ABOX FMC6904A6906ABOX

*Chlorine DPD #4 Instrument Grade Tablets (*6906A) Test Tubes, plastic, w/caps 0106 Chlorine DPD Comparator, 0.1-1.0 ppm 6978

*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www.lamotte.com. To obtain a printed copy, contact Lamotte by e-mail, phone or fax.

To order individual reagents or test kit components, use the specified code number. USE OF THE OCTET COMPARATOR The Octet Comparator contains eight permanent color standards. A test sample is inserted into the openings in the top of the comparator. The sample can then be compared to four color standards at once, and the value read off the comparator. For optimum color comparison, the comparator should be positioned between the operator and a light source, so that the light enters through the special light-diffusing L Mott screen in the back of the comparator. Avoid viewing the comparator against direct sunlight or an irregularly lighted background.

PROCEDURE FREE AVAILABLE CHLORINE 1. Rinse the test tube (0106) with sample water. Fill to 10 mL line with sample water. 2. Add one *Chlorine DPD #1R Tablet (6999A). Cap and mix until tablet disintegrates. 3. Immediately insert test tube into Chlorine DPD Comparator (6978). Match sample color to a color standard. Record as Reading #1. Do not discard this sample if testing for Monochloramine, Dichloramine or Total Residual Chlorine. NOTE: Reading should be made within one minute of adding reagent. Reading #1 = ppm Free Available Chlorine MONOCHLORAMINE 4. Add one Chlorine DPD #2R Tablet (6904A) to sample from Step 3. Cap and mix until tablet disintegrates. 5. Insert test tube into Chlorine DPD Comparator (6978). Match sample color to a color standard. Record as Reading #2. Reading #2 - Reading #1 = ppm Monochloramine DICHLORAMINE AND TOTAL RESIDUAL CHLORINE 6. Add one *Chlorine DPD #3R Tablet (6905A) to sample from Step 5. Cap and mix until tablet disintegrates. 7. Insert test tube into Chlorine DPD Comparator (6978). Match sample color to a color standard. Record as Reading #3. Reading #3 - Reading #2 = ppm Dichloramine Reading #3 = ppm Total Residual Chlorine

TOTAL RESIDUAL CHLORINE AND COMBINED CHLORINE 1. Rinse test tube (0106) with sample water. Fill to 10 mL line with sample water. 2. Add one *Chlorine DPD #1R Tablet (6999A). Cap and mix until tablet disintegrates. 3. Immediately insert test tube into Chlorine DPD Comparator (6978). Match sample color to a color standard. Record as ppm Free Available Chlorine. 4. Add one *Chlorine DPD #3R Tablet (6905A). Cap and mix until tablet disintegrates. 5. Insert test tube into Chlorine DPD Comparator (6978). Match sample color to a color standard. Record as ppm Total Residual Chlorine. Combined Chlorine, ppm = Total Residual Chlorine - Free Available Chlorine TOTAL RESIDUAL CHLORINE *Chlorine DPD #4 Tablet determines Total Residual Chlorine and is used where a differentiation in the type of chlorine is not required. 1. Rinse test tube (0106) with sample water. Fill to 10 mL line with sample water. 2. Add one *Chlorine DPD #4 Instrument Grade Tablet (6906A). Cap and mix until tablet disintegrates. 3. Insert test tube into Chlorine DPD Comparator (6978). Match sample color to a color standard. Record as ppm Total Residual Chlorine.

LaMOTTE COMPANY
Helping People Solve Analytical Challenges PO Box 329 Chestertown Maryland 21620 USA 800-344-3100 410-778-3100 (Outside U.S.A.) Fax 410-778-6394 Visit us on the web at www.lamotte.com