Fusion of Ultra Fast MRM and Ultra Fast Polarity Switching An Ultra Fast Scan Speed of 15,000 u/sec 500 MRMs per Second, the Most Ever Possible
High Reliability
Robust System Provides Long-term Data Stability UFsweeperTM Technology Dramatically Minimizes Cross Talk Excellent Linearity with Wide Dynamic Range
User-Friendly
Easy Maintenance for Minimized Instrumetn Downtime Single-vendor Solution Provides Seamless Operation Quantitation Browser for Effective Multi-analyte Quantitation
CE - 20V - 40V
26 pharmaceutical compounds were analyzed using Shimadzus Synchronized Survey ScanTM (SSS) shown in Figure 1. In this mode, full scan measurement are rapidly followed by automated product ion scanning. High-speed polarity switching (15 msec) and rapid scan rates (15000u/sec) allow multiple collision energies to be employed for unknowns even with narrow peak widths. This enables molecular weight confirmation from the Q3 scan data and also generates structural fragmentation information from the same peak. A total of 26 pharmaceutical compounds were evaluated. Figure 2 shows the analytical result for Cilostazol and Warfarin. As illustrated in Table 1, all compounds were detected in either positive mode, negative mode, or both, demonstrating the LCMS-8030s effectiveness for drug discovery and synthesis confirmation.
40000000 40000000 30000000 20000000 10000000 0 0.0 Inten. (x10,000,000) 1.0 2.0 min 1.0 10000000 1:TIC(+) 2:TIC(+) 3:TIC(+) 4:TIC(-) 5:TIC(-) 6:TIC(-) 1:TIC(+) 2:TIC(+) 3:TIC(+) 4:TIC(-) 5:TIC(-) 6:TIC(-)
20V 40V
Cilostazol
30000000
Warfarin
20000000
370.2
1.0
307.2
Q3 Scan(+)
C20H27N5O2 MW :369.46
0.5
Q3 Scan(-)
C19H16O4 MW :308.33
2.0
125.1
5.0
307.1
1.0
55.2
2.5
55.1
1.0
57.4 117.0
0.5
83.1
0.0 100
146.0
0.0 200 300 m/z 100
Fig. 2 Analytical results (Cilostazol, Warfarin) Table 2 Analytical Conditions : Shim-pack XR-ODSII (2.0 mmI.D. 50 mmL.) : 5 mmol/L ammonium formate-water : acetonitrile : 5 %B (0 min) 95 %B (3-3.50 min) 5 %B (3.51-5 min) Flow rate : 0.3 mL/min Injection volume : 5 uL CDL temperature : 250 C Block heater temperature : 400 C Column Mobile phase A Mobile phase B Gradient program
26 Nitrendipine
Nebulizing gas flow Drying gas flow Column temperature Collision energy Q1/Q3 resolution Scan type Scan range Ionization mode Scan speed
: 1.5 L/min : 10 L/min : 40 C : pos 20V, 40V neg 20V, 40V : Unit/Unit : Q3 scan - product ion scan : m/z 200-500 : ESI positive and ESI negative : 3333 u/sec