Professional Documents
Culture Documents
!
! "!
ABSTRACT:
nitrophenyl ethanoic acid were analysed. The organic acids are cited in ascending order
of retention times, thus descending order of polarity. It was determined that an 80:20 ratio
of the potassium hydrogen phosphate – phosphoric acid buffer and acetonitrile (as mobile
phase) yielded the best balance between desired low retention times and high resolutions.
Its polarity index was 9.32; it was also calculated that a 83:17 ratio of the buffer:methanol
would be required to produce the same polarity index. The experiment also involved
comparison of isocratic versus gradient elution; results show that isocratic elution was the
better option, though this may be at odds with normalcy. A less than optimum choice of
By combining the methods of standard addition and linear regression analysis, the
concentration of p-nitrophenyl ethanoic acid was confirmed to be (~25 ppm) 24.46 ppm ±
!
! #!
INTRODUCTION:
organic acids: phtalic acid, salicylic acid, benzoic acid and p-nitrophenyl ethanoic acid;
absorption for identification and quantification). The technique involved running the
analyte sample through a column packed with siloxane-coated silica particles; this was
the stationary phase. The eluent-mixture (mobile phase) was a mixture of aqueous buffer
and acetonitrile (methyl cyanide). The column was sealed under high pressure and the
carrier fluid/eluent was liquid, hence the name high performance/pressure liquid
chromatography. The use of the polar mobile phase and the relatively non-polar
The components of the analyte sample mixture separate out by their different and
characteristic affinities for the polar mobile phase and non-polar stationary phase. The
more polar components exit the column before the less polar ones, thus having a lesser
retention time, due to a greater affinity for the polar mobile phase than the stationary
phase.
EXPERIMENTAL PROCEDURE:
benzoic and p-nitrophenyl ethanoic acid of 100 ppm; thereby reducing the concentration
of each component to 25 ppm. This test mixture was then run through the column at a
rate of 1.5 mL/min (constant throughout the experiment) in a 80:20 mixture of the buffer
and acetonitrile as mobile phase; the test mixture was then run at the same speed but
under different ratios of buffer and acetonitrile, 90:10, 70:30 and 60:40. Each run was
terminated once all the components had eluted. Then, the 100 ppm standard of each
The next series of experimental runs were conducted under a gradient of buffer-
acetonitrile ratios. One such run is cited here (refer to pg ?? for the data): 65:35 from 0 to
0.6 min., 80:20 from 0.6 min. to 0.8 min. and 90:10 from 0.8 min. to 3 min.
The second section of the experiment was commenced with the preparation of 50
phthalic acid at a concentration of 25 ppm each. This was done by adding 12.5 mL of 100
ppm standard solutions of each substance along with 12.5 mL of water. Using the 100
ppm p-nitrophenyl ethanoic acid standard as the new analyte, five solutions (10 mL) were
prepared of varying concentrations, 0, 10, 20, 30 and 40 ppm. These solutions were
(corresponding to 0, 10, 20, 30 and 40 ppm) of the analyte; the total volume was brought
to 10 mL using water. Each of the five solutions was then run through the column under a
! Table and graph of retention times (ret. time or tR) of 100 ppm standards
peak
peak width
peak # ret. time (min.) width area height area %
(half-height)
(proper)
phthalic acid
!
! '!
!
! (!
!
!
!
!
!
!
!
!
!
!
!
!
! )!
! Isochratic separation
and
! LC 1: Q1 – Q2
width
buffer:aceton
peak # tR (min.) (half- width (proper) area % k' Rs
-itrile
height)
! the order of elution of the components does not change and so is not shown
! LC 1: Q3
longest tR Rs :longest tR
buffer:acetonitrile Rs
(min.) ratio
! the 80:20 mixture yields the best compromise between resolution and ret. time.
A: aqueous buffer
B: acetonitrile
and x + y = 1
peak
sample buffer:aceto time tR area Rs/longest tR
peak # width Rs
run -nitrile (min.) (min.) (%) ratio
(proper)
!
! the higher resolutions are of reduced significance since an instrument is as good
combined with equal or better resolution since the mobile phase composition…
! *#!
processed data suggests, isocratic elution was more efficient with a higher Rs /
longest tR ratio compared to the highest provided by gradient elution (0.648 vs.
0.543).
! *$!
! LC 2: Q1
buffer:acetonitrile
! LC 2: Q2
!"#$%&',!-',(&-'(1'233'445'46$"7)(48*$9&'
)%! !"#$%&'()'%)*%' *78%$(":'%:";'!7%$;%);'%;;*;'
+'
)+!
(%!
,!-!%.'*"/!0!&).('!
(+! 12!-!+.($'!
'%!
'+!
&%!
,(&-'.!/0'
&+!
+! +.%! *! *.%! "! ".%! #! #.%! $! $.%!
0 65.177
1 79.232
2 84.878
3 87.69
4 89.511
gradient 5.713
y-intercept 69.872
N 5
Sxx 10
y bar 81.298
! LC 2: Q3:
DISCUSSION
sample elute as that would require alteration of the polarity of the components’
molecules. Since there is no physical nor chemical alteration to the molecules of the
Advantages: allows for analysis of complex samples with potentially significant matrix
effects.
standard.
! *)!
CONCLUSION:
The buffer:acetonitrile composition ratio of 80:10 was found to offer the best
compromise of resolution and retention times. Its polarity index was calculated to be
9.32, and that a 83:17 composition of the buffer:methanol would be required to produce
the same polarity index. Secondly, it is concluded that isocratic elution provided better
results than gradient elution; this may be attributed to less than optimum choice of
of linear regression analysis and standard additions, calculated to be 24.46 ppm ± 28.7 %.