Users Manual

OxySense Permeation Chamber

I.

Introduction

Congratulations! You are about to use one of the most exciting innovations in Film OTR measurement in many years. Accurate, easy to use and flexible, the OxySense Permeation Chamber, with the OxySense OTR software package allows you to use your OxySense instrument to perform flat film OTR studies and still have the instrument available for finished package OTR studies, head space analysis and dissolved oxygen studies. II. The Chamber

The OxySense Permeation Chamber consists of two inner chambers (upper and lower) attached together by a permanent hinge. Each inner chamber has two gas valves (an inlet and an outlet) attached to them. There are two large thumb screws that are used to secure the Chamber halves for measurement. On the inside there is a bottom inner chamber; factory calibrated black gasket surrounding the and on the top inner chamber, a OxyPerm Dot.

On the top of the Chamber you will find two labels. One is the serial number bar code label that you will use to identify this Chamber to the system. The second label has the calibration factors for the factory installed OxyPerm Dot. You will need these numbers to register the Chamber in the system before performing your first test. You will also find a pen well and alignment mark for the placement of the OxySense EasAlign Reader Pen when you take a measurement.

2 of 29 7/13/2009

I I I . Setting Up Your OxySense Permeation Chamber For A Test a. Open the Permeation Chamber by unscrewing the black thumb screws. b. Your OxySense Permeation Chamber comes with a factory installed OxyPerm Dot already installed. If the OxyPerm Dot is damaged or is older then six months old, you should replace it. (See Appendix___, Replacing the OxyPerm Dot). c. Prepare Your Film Sample Cut a sample square 2.5X2.5 (6.3cmX6.35cm) from the film you want to test. d. Using your finger or suitable applicator, apply vacuum grease to the area around the upper chamber (the one with the OxyPerm Dot in the middle) making sure the grease is all around the chamber, but do not put grease on the wall or bottom of the chamber.

e. Place the film sample square onto the greased surface of the upper chamber making sure the film is taught, without wrinkles and that it covers the cavity.

3 of 29 7/13/2009

f. .

Close the Chamber carefully and insert and tighten the thumb screws. The thumb screws should be hard hand tight to insure the proper seal. DO NOT USE PLIERS OR OTHER TOOLS TO TIGHTEN THE THUMBSCREWS .

g. Place the OxySense EasAlign Pen from the 4000B into the slot on top of the chamber making sure that red marks on the pen and chamber are aligned and that it is fully seated in the well.

The Permeation Chamber is now ready for calibration and testing.

4of 29
7/13/2009

IV.

Calibration

Your OxySense Permeation Chamber comes with a factory installed OxyPermDot. The factory calibration numbers can be found on a label on the top of the Chamber

. However, because changes may occur it is recommended that you recalculate the calibration numbers for the chamber prior to starting a new permeation test. The calibration procedure is simple: a. Set up your sample film and Chamber as described in Section IV above. b. Turn on your OxySense 4000B c. Click on the Calibration tab on the OxySense software and confirm that the box next to Low has a 0 in it and the box next to High has the percentage of oxygen equal to the percentage you want to set as the high end of the range: (21, if your ranges is 0% to 21%, or 5 if your range is -% to 5%). If they do not, change to numbers in the boxes to reflect the range you want to calibrate over. 1. If the high end of your range is 21%, as the chamber was open to air when you closed it, the oxygen concentration inside the chamber will be 21% (oxygen concentration in ambient air) and this can be used as the upper limit for calibration purposes. Go to step 5. If the high end of your range is less than 21%, you will need to flush the chamber with certified gas with an oxygen content equal to the top end of your range (this should be the same number you entered in the box next to High above), follow steps 6 through 9 for the flushing proceedure, below, before clicking the Get High button. Then return to step 6, using your pure nitrogen source to get the Low reading.

5 of 13 7/13/2009

d. With the EasAlign Pen in the Chamber pen well, reconfirm that the signal level is set at the 0 level (if it is not, adjust the gain, up or down, to align the end of the green signal bar at 0). Click the Get High button. e. Now Open the inlet and outlet valves of the upper chamber.( these will be the valves to the left with the Chamber set with the hinge away from you). f. Attach your nitrogen source to the Inlet valve, turn on the gas and begin flushing the upper chamber. g. You may verify the progress of the flush by periodically. clicking the Capture button found at the bottom of the calibration screen. A number will appear in the Last Oxygen Reading box, found in the upper left corner of the Calibration Screen. DO NOT BE CONCERNED ABOUT THE NUMBER ITSELF. You are only interested in when the number stabilizes (allowing for small variations). When the number is stable, the Chamber is flushed. This can take between 1 and 3 minutes depending on your nitrogen source. h. When the chamber is flushed first close the inlet valve and then close the outlet valve and turn off the nitrogen supply. i. Click the Get Low button j. Click the Calculate New dA/dB button and write down the dA and dB numbers that appear in the appropriate boxes. Be sure to include the + or sign and the decimal point. DO NOT click the Save Calculated dA/dB box. k. Now return to the Film Permeation tab and continue with setting up your test file.

6 of 13 7/13/2009

V. 5.0

Permeation Measurements

If the Permeation feature is enabled from the Configuration window, the Gen III-P logo will be displayed on the Top with the Film and Package Permeation Tab, if Unlocked is selected for Permeation within the Configuration Window. If the Permeation feature is not enabled, the Gen III logo will be displayed on the Top and the Permeation features (Film and Package) will not be enabled.

5.1 Film Permeation


Adding Chamber Test. Adding a New Test for Chamber Opening Existing Test 2. 3. 4. 5. 6. 7. Fig. 5.1: Film Permeation Screen 8. a. b. Browse Active Browse All View Log File & Graph Setup Close Test Print Report Capture Chamber Serial Test Name an existing

1.

a &

New New

5.1.1Adding a New Chamber and New Test Name

7 of 29 7/13/2009

To enter the Serial Number of a new Chamber and Test Name into the system, enter the new Chamber Serial Number in the Chamber Serial Number field by scanning the Chamber Barcode or enter through the keyboard and click the Open button. Since the Serial Number of the Chamber does not exist in the database, a message box will be displayed Chamber Serial Number not found. Do you want to add a New Chamber Serial Number: A0000-10099 If you click the Cancel button, the Chamber will not be added to your database. You must click the OK button and complete the process to add the first Test in order for the Chamber to be added. Clicking the OK button from the message box will allow you to add the new Chamber to your database by entering the first Test Name and Test Description into the following Add New Test screen. You must add the first Test to complete the addition of the Chamber to the database. a. Chamber Serial Number b. c. d. e. Fig. 5.2: Adding a New Chamber & New Test Name a. Chamber Serial Number: The previously entered new Chamber Serial Number will be displayed. b. Test Name: Enter the Test Name (maximum 17 characters) in the Test Name field. The Test Name must be unique in the system to create a New Test and cannot be blank. c. Description: Enter the Test Description (maximum 110 characters) in the Test Description field. The Test Description field is optional. d. Add: After entering the Test Name and Test Description, click the Add button to add the New Test for the new Chamber. Upon clicking Test Name Description Add Cancel

8 of 29 7/13/2009

the Add button, the existing OxyDot Calibration confirmation message box will be displayed: The Current OxyDot Calibration is dA=1.13 and dB=0.0. Is the existing OxyDot Calibration correct?. If you click the Yes button from the message box, the Add New Test window will be closed and the New Test Name with the Serial Number of the new Chamber will be opened in the Film Permeation tab with OxyDot calibration dA=1.13 and dB=-0.0. If you click the No button from the message box, the following screen will be displayed: 1. OxyDot Calibration 2. Confirm OxyDot Calibration 3. 4. Fig. 5.3: Adding OxyDot Calibration 1. 2. OxyDot Calibration: Enter the OxyDot Calibration dA and dB fields. Confirm OxyDot Calibration: Enter the same OxyDot Calibration dA and dB fields in the Confirm OxyDot Calibration fields. If the Calibration (dA/dB) and Confirm Calibration (dA/dB) do not match, an error message will be displayed. The Calibration Numbers will be entered into the application database for that Test and will utilized for all subsequent Tests for that Chamber until changed. 3. OK: When you enter the Calibrations and click the OK button, the Add New Test window will be closed and the New Test Name with the Chamber Serial Number will be opened in the Film Permeation tab. OK Cancel

9 of 29 7/13/2009

4.

Cancel: Clicking the Cancel button will close the Add New Test window without adding a New Test and the previous screen will be displayed to specify a different Chamber or utilize the Browse Active or Browse All buttons.

e. Cancel: If the Cancel button is clicked, the New Chamber Serial and Test will not be added to the database and will return to the Main screen.

5.3 Adding a New Test for an Existing Chamber To create a New Test for an existing Chamber, the prior Test must be closed. Please make note that once you have closed a test, you can no longer add additional readings to that Test. If the Open button displays an Oxygen Permeation Test Name when the Chamber Serial Number is entered, the previous Test has not been closed. You must close the prior Test before attempting to open a New Test using the same Chamber. Enter the Unique Serial Number of the Chamber by scanning the Chamber Barcode or enter the Serial Number of the Chamber through the keyboard. Click the Open button. If there are no open Tests for the Chamber, a message box will be displayed: Do you want to add New Test for Chamber Serial Number: A0000-100009?. If you click the Yes button, the following Add New Test window will open: a. Chamber Serial Number b. c. d. e. Test Name Description Add Cancel

10 of 29 7/13/2009

a. Chamber Serial Number: The entered Chamber Serial Number will be displayed. b. Test Name: Enter the Test Name (maximum 17 characters) in the Test Name field. The Test Name must be unique in the system to create a New Test and cannot be blank. c. Description: Enter the Test Description (maximum 110 characters) in the Test Description field. The Test Description field is optional. d. Add: After entering the Test Name and Test Description, click the Add button to add the New Test for the Chamber. Clicking the Add button will check whether the Test Name entered is already in the database. If the Test Name is found, an error message will be displayed: Test Name already exists. A duplicate Test Name is not allowed.. If the Test Name field is blank, the application will display an error message: Test Name cannot be Blank. Please type proper Test Name.. If the Test Name is correct, the application will display the following confirm existing OxyDot Calibration message: The Current OxyDot Calibration is dA=1.13 and dB=0.0. Is the existing OxyDot Calibration correct? with Yes and No buttons. If you click the Yes button from the message box, the Add New Test window will be closed and the New Test Name with the Serial Number of the new Chamber will be opened in the Film Permeation tab with OxyDot calibration dA=1.13 and dB=-0.0. If you click the No button from the message box, the following screen will be displayed: 1. 2. OxyDot Calibration Confirm OxyDot Calibration 3. OK

11 of 29 7/13/2009

4.

Cancel

1.

OxyDot Calibration: Enter the OxyDot Calibration dA and dB fields. If the Calibration dA field is blank, an error message will be displayed: Enter the valid value for the Calibration dA. If the Calibration dB field is blank, the application will display: Enter the valid value for the Calibration dB error message.

2.

Confirm OxyDot Calibration: Enter the same OxyDot Calibration dA and dB fields in the Confirm OxyDot Calibration fields. If the Calibration dA and Confirm Calibration dA do not match, an error message will be displayed: The Calibration dA and Confirm Calibration dA do not match. If the Calibration dB and confirm Calibration dB do not match, the application will display the following error message: The Calibration dB and Confirm Calibration dB do not match. The Calibration Numbers will be entered into the application database for that Test, and will be utilized for all subsequent Tests for that Chamber until changed.

3.

OK: When you enter the Calibrations and click the OK button, the Add New Test window will be closed and the New Test

12 of 29 7/13/2009

Name with the Chamber Serial Number will be opened in the Film Permeation tab. 4. Cancel: Clicking the Cancel button will close the Add New Test window without adding a New Test and the previous screen will be displayed to specify a different Chamber or utilize the Browse Active or Browse All buttons. e. Cancel: If the Cancel button is clicked, the New Test will not be added to the database and will return to the Main screen.

5.4 Opening an Existing Test a. Enter the unique Serial Number of the Chamber (by scanning the Chamber Barcode or enter the Serial Number of the Chamber through the keyboard). b. After entering the unique Serial Number of the Chamber or the Oxygen Permeation Test Name in the appropriate fields, click the Open button to open the Test. If a Test is Open for the Chamber, the Test will be Open and you will be able to utilize the various buttons to View Log File, View Log Graph, Capture a measurement Setup the test settings, Print Report or Close Test.

5.5 Browse Active Tests Click the Browse Active button to display the following window containing a list of Open (Active) Tests (Test Name, Chamber Number, Open Date & Time and Close Date & Time) for Film Permeation. 1. 2. 3. 4. 5. Open Test Update Description Update Calibration Close Test Close

13 of 29 7/13/2009

1.

Open Test: To open a Test, select the Test from the Open Tests list and click the Open Test button. Clicking the Open Test button closes the Browse Active window and displays the selected Test Name and Chamber Serial Number in the Film Permeation Tab screen. Clicking the Open Test button without previously selecting a Test will display the following error message: Please select the Test from the Active Test List to open.

2.

Update

Description:

Clicking the Update Description button

displays the selected Test Name with Chamber Number in a new window with the Test Description in the following screen to allow you to update the Description field in the data base: a. Test Name & Chamber b. c. d. Fig. 5.7: Film Permeation Update Description a. b. c. The selected Test Name and Chamber Number will be displayed. Update the Test Description in the Test Description field. Clicking the Update Description button updates the Test Description for the Selected Test Name and Chamber Number. d. 3. Clicking After the updating the database, will close the the Update Update Description window will be closed. Cancel button Description window without updating the Test Description. Update Calibration: Clicking the Update Calibration button displays the selected Test Name and Chamber Number in a new window with the existing OxyDot Calibration. If the Calibration dA/dB is changed, Test Description Update Description Cancel

14 of 29 7/13/2009

the previous data points for the Selected Test Name and Chamber Number will be recalculated based upon the New Calibration dA/dB. a. Test Name & Chamber b. c. d. Fig. 5.8: Film Permeation Update OxyDot Calibration a. b. The selected Test Name with Sample Number will be displayed. Enter the new OxyDot Calibrations in the dA and dB fields. If the Calibration dA/dB fields are empty, an error message will be displayed. c. Clicking the Update Calibration button will ask you for confirmation: Are you sure you want to update Calibration dA and dB. If Yes, the previous Data Points will be recalculated.. If you click the Yes button, the previous data points will be recalculated with the updated dA and dB for the selected Test Name and Chamber Number. If you click the No button, the Calibration dA and dB will not be updated. d. 4. Clicking the Cancel button will close the Update Calibration window. Close Test: Please make note that once you have closed a test, you can no longer add additional readings to that Test. Test Description Update Description Cancel

To Close a Test, select the Test from the Active Test List and click the Close Test button. Clicking the Close Test button will display a message box: Are you sure you want to Close the Test: For Chamber Serial Number : ? You will not be able to capture additional data

15 of 29 7/13/2009

points for a Closed Test.. from the Open Tests list.

Clicking the OK button will close the

selected Test Name and Chamber Number and remove the Test Name Clicking the Close Test button without previously selecting a Test will display an error message: Please select the Test from the Active List to close. 5. Close: Click the Close button to close the Browse Active window.

5.6 Browse All Tests Click the Browse All button to display the following window containing a list of Open (Active) Tests (Test Name, Chamber Number, Open Date & Time and Close Date & Time) followed by all Closed Tests for Film Permeation. 1. 2. 3. 4. 5. 6. Open Test Update Description Update Calibration Close Test Delete Test Close

Fig. 5.9: Film Permeation Browse All Tests

1.

Open Test: To Open a Test, select the Test from the Open Tests list or Closed Tests list and click the Open Test button. Clicking the Open Test button closes the Browse All window and displays the selected Test Name and Chamber Serial Number in the Film Permeation Tab screen.

16 of 29 7/13/2009

Clicking the Open Test button without previously selecting a Test will display an error message: Please select the Test and Chamber Number from the Active or Closed List to open. 2. Update Description: Clicking the Update Description button displays the selected Test Name with Chamber Number in a new window as per Fig. 5-7 with the Test Description. You are allowed to update the Test Description if the selected Test is closed. 3. Update Calibration: Clicking the Update Calibration button displays the selected Test Name with Chamber Number in a new window as per figure Fig. 5-8 with the existing OxyDot Calibration. If the Calibration dA/dB is changed, the previous data points for the Selected Test will be recalculated based upon the new Calibration dA/dB. Calibration for a Closed Test cannot be updated. 4. Close Test: To Close a Test, select the Test from the Active Test List and click the Close Test button. Clicking the Close Test button will display a message box Are you sure you want to Close the Test: For Chamber Serial Number: ? You will not be able to capture additional data points for a Closed Test.. Clicking the OK button will Close the selected Test and remove that specific Test from the Open Tests list and display the Test and Chamber Serial Number in the Closed Test list. Clicking the Close Test button without previously selecting a Test from the Active Test List will display an error message: Please select the Test from the Active Test List to close. 5. Delete Test: To delete a Test, select the Test from the Closed Test List and click the Delete Test button. Clicking the Delete Test button will display a message box with: Are you sure you want to Delete the Test: For Chamber Serial Number:?. Tests list. Clicking the OK button will Delete the selected Test and remove that specific Test from the Closed The

17 of 29 7/13/2009

Clicking the Delete Test button without previously selecting a Test from the Closed Test List will display an error message: Please select the Test from the Close Test List to delete. 6. Close: Click the Close button to close the Browse All window.

5.7 View Log File and Log Graph To open the Film Permeation Log or Graph screens, identify a specific Chamber Serial Number based upon the options noted above. If the View Log File or View Log Graph button is clicked without entering anything in the Chamber Serial Number field, the application will display an error message: Chamber Serial Number cannot be blank. If the View Log File or the View Log Graph button is clicked after entering a Chamber Serial that is not in the database, the application will display an error message: Could not find the entered Chamber Serial Number. Clicking the View Log File button will display the following Film Permeation Log screen with the list of Data Points, OTR, Starting and Final Data Points: Clicking the View Log Graph button will display the Film Permeation Graph screen (Fig. 5.11) with the Graph of Data Points, OTR, Starting and Final Data Points: 1. 2. 3. 4. 5. 6. 7. 8. OTR Test Name Chamber Starting Data Point Final Data Point Calibration Log Delete Data Point

18 of 29 7/13/2009

9. 10. 1.

Select Log View Graph

OTR: Displays the OTR Reading for the Starting and Final Data Points. If the OTR reading is not in the range specified, a message box will be displayed: OTR should be greater than 0.0000 and less than 2000.0000, with the OTR reading in red.

2. 3. 4.

Test Name: Displays the selected Test Name. Chamber Number: Displays the selected Chamber Number. Starting Data Point: Displays the selected starting data point (Date, Time and Oxygen (mBar) for the OTR calculation. If the Oxygen (mBar) reading is not in the range specified, a message box will be displayed: Oxygen Reading (mBar) should be greater than 0 and less than 300, with the Oxygen (mBar) reading in red. Clicking the Reset button will blank the starting data points (Date, Time and Oxygen (mBar) and the OTR value will be 0.000. Select the new starting data point by clicking an entry in the Log or the Graph.

5.

Final Data Point: Displays the selected final data point (Date, Time and Oxygen (mBar) for the OTR calculation. If the Oxygen (mBar) reading is not in the range specified, a message box will be displayed: Oxygen Reading (mBar) should be greater than 0 and less than 300, with Oxygen (mBar) reading in red. Clicking the Reset button will blank the final data point (Date, Time and Oxygen (mBar) and the OTR value will be 0.000. Select the new final data point by clicking an entry in the Log or the Graph.

6. 7.

Calibration: Displays the Calibration dA/dB for the selected Test and Chamber Number. Log: Displays the Data Points for the selected Test and Chamber Number. If the View By is selected in PPB or Percent, the user can select the Data Point in PPB or Percent. However, selecting the Data Point in PPB or Percent will actually utilize the associated mBar value

19 of 29 7/13/2009

for the same Data Point and enter within the empty Starting or Final Data Point. 8. Delete Data Point: To delete a Data Point from the Log List, select the Data Point from the Log List and click the Delete Data Point button. Clicking the Delete Data Point button displays the following message: Are you sure you want to Delete the Selected Data Point. Clicking the OK button deletes the selected data point and removes from the Log List and Log Graph. If a data point has not been selected prior to the click of the Delete Data Point button, an error message box will be displayed: Select the Data Point from the Log List. 9. Select Log: To select another Test, click the Select Log button. Clicking the Select Log button displays the Film Permeation screen Fig.5-10. 10. View Graph: Clicking the View Graph button will display the following Film Permeation Graph screen instead of the Log of Data Points: 1. View Log File 2. X-Axis Scale

Fig. 5.11: Film Permeation View Graph.

20 of 29 7/13/2009

1. View Log File: Clicking the View Log File button will display the Film Permeation Log screen Fig. 5.10 instead of the Graph of the Data Points. If the View By is selected in PPB or Percent, the user can select the Data Point in PPB or Percent. However, selecting the Data Point in PPB or Percent will actually utilize the associated mBar value for the same Data Point and enter within the empty Starting or Final Data Point. 2. X-Axis Scale a. If the Radio button Days in the X-Axis Scale is checked, the Graph will be displayed in Days. b. If the Radio button Hours in the X-Axis Scale is checked, the Graph will be displayed in Hours. c. If the Radio button Minutes in the X-Axis Scale is checked, the Graph will be displayed in Minutes. d. If the Radio button Seconds in the X-Axis Scale is checked, the Graph will be displayed in Seconds. 5.8 Setup (Setting) To open the Setup window, identify a specific Chamber Serial Number based upon the options noted above. If the Setup button is clicked without entering anything in the Chamber Serial Number field, the application will display an error message: Chamber Serial Number cannot be blank. If the Setup button is clicked and the Chamber Serial entered is not in the database, the application will display an error message: Could not find the entered Chamber Serial Number.. After identifying a specific Chamber Serial Number, click the Setup button to open the following Setup window: 1. 2. Password Change Password

21 of 29 7/13/2009

3. 4.

Cancel OK

1. 2.

Password: Enter the Password in the Password field. The factory set password is oxysense. Change Password: To change the Password, click the Change Password link to display the following screen: a. b. c. d. e. Old Password New Password Confirm Password OK Cancel

Fig. 5.13: Film Permeation Change Password Screen. a. b. c. d. Old Password: Enter the old Password. New Password: Enter the new Password. Confirm Password: Enter the same new Password in the Confirm Password field. OK: Clicking the OK button will check the Old Password matches the database entry and the New Password and Confirm

22 of 29 7/13/2009

Password are identical.

If the Old Password matches and the

New Password and Confirm Password are identical, a message box will display: Password updated. If the Old Password does not match the database entry, an error message will be displayed. If the New Password and Confirm Password are not identical, an error message will be displayed. e. 3. 4. Cancel: Clicking the Cancel button will display the Setup Password screen (Fig. 5.12).

Cancel: Clicking the Cancel button closes the Setup Window and displays the Film Permeation Tab screen. OK: After entering the correct Password and clicking the OK button, the following Permeation setting screen will be displayed. The Setup screen contains Pressure, Volume, Area, View By and View OTR By for the Chamber Serial Number and Test Name opened in the Film Permeation screen.

1. 2. 3. 4. 5. 6. 7.

Default Pressure Probe Pressure Pressure (mBar) View By View OTR By OK Cancel

23 of 29 7/13/2009

1.

Default Pressure: If Always Use Default Pressure check box is selected, the calculation of Oxygen will use the Default Pressure value.

2.

Probe Pressure: If Always Use Probe Pressure check box is selected, the calculation of Oxygen will use the real time Pressure value from the Gen III device and will not use the Default Pressure value.

3.

Default Pressure (mBar): Enter the Default Pressure value in the Default Pressure (mBar) field. If the Test is closed, changing the Default Pressure value will display an error message: Default Pressure value cannot be changed. Test is closed. If the Test is active and you enter a new default Pressure value in the Default Pressure (mBar) field and the value is not within the expected range, an error message will be displayed specifying the minimum and maximum range for the Default Pressure.

4. 5. 6.

View By: To change the View By, select Percent, ppb or mBar. View OTR By: To change the View OTR By, select ml/m2/day or ml/100in2/day. OK: If the Test is active and you select the Always Use Default Pressure check box and enter a new Default Pressure, Volume or Area value in their appropriate field and click the OK button, the following message will be displayed: Are you sure you wish to change these setup values? The setup values should not If you click the require changes for an existing Chamber.

Cancel button, the values will not be updated. If you click the OK button and the Volume field is changed, the following message box will be displayed: Changing the Volume of an

24 of 29 7/13/2009

existing Chamber is not advised. The values will be updated in the database for the open Test and displayed in the Film Permeation screen and the Setup window will be closed. If the values are updated and an OTR has been calculated, the OTR must be recalculated based upon the new Pressure, Volume or Area. If the View By is changed and you click the OK button, the Setup window will be closed and the Graph and the Log List Data Points will be displayed in the new selection in the Film Permeation Log or Graph screen. If the View OTR By is changed and you click the OK button, the Setup window will be closed and the OTR value will be calculated and displayed per the new selection in the Film Permeation Log or Graph screen. 7. Cancel: Clicking the Cancel button will close the Setup window.

5.9 Close Test Please make note that once you have closed a test, you can no longer add additional readings to that Test. 1. To close a Test, identify a specific Chamber Serial Number based upon the options noted above and click the Close Test button. If the Close button is clicked without entering anything in the Chamber Serial Number field, the application will display an error message: Chamber Serial Number cannot be blank. If the Close button is clicked and the Chamber Serial entered is not in the database, the application will display an error message: Could not find the entered Chamber Serial Number. 2. If the selected Chamber Serial Number and Test Name is active, clicking the Close Test will display a message box: Are you sure you want to Close the Test : For Chamber Serial Number : ? You will not

25 of 29 7/13/2009

be able to capture additional data points for a Closed Test.. If the Yes button is clicked within the message box, the Chamber Serial Number and the Test will be closed. 3. If the selected Test Name is already closed, clicking the Close Test button will display an error message: Test is already closed.

5.10 Print Report To Print the Permeation Report, identify a specific Chamber Serial Number based upon the options noted above. If the Close button is clicked without entering anything in the Chamber Serial Number field, the application will display an error message Chamber Serial Number cannot be blank. If the Close button is clicked after entering a Chamber Serial that is not in the database, the application will display an error message: Could not find the entered Chamber Serial Number.. If the selected Test is active, clicking the Print Report button displays the following message box: 1. Print Interim Report 2. Cancel

Fig. 5.15: Film Permeation Print Report Message.

1.

Print Interim Report: Clicking the Print Interim Report button opens the following Print Report window. The Print Report window for an Interim Report does not contain the Close Date because the

26 of 29 7/13/2009

Test is active.

If the Test is closed the above message will not be

displayed and the Print Report window will be opened and it will contain the Close Date. 1. 2. 3. 4. Print Preview Print PFD Report Close

Fig. 5.16: Film Permeation Print Report.

1.

Print Preview: To view how the report will be printed, click the Print Preview button. Clicking the Print Preview button will open the Print Preview window to display how the report will be printed.

2. 3.

Print: To Print the Report, click the Print button. Clicking the Print button will send the report directly to the printer. PDF Report: To create a PDF document of the Report, click the PDF Report button. Clicking the PDF Report button will send

27 of 29 7/13/2009

the report to the PDF writer software for generating the PDF report. The PDF Report button will generate the PDF report with the selected Test Name as the File Name with Date and Time and save in a Film folder in the Permeation Report folder within the Gen III folder in the My Documents directory and display a message box with the file saved path. 4. Close: Clicking the Close button will close the Print Report window.

Print Report window Data a. Displays the selected Chamber Number. b. Test Open Date. c. Test Close Date. If the Test is Active the Close Date will be blank. d. Displays the Selected Test Name. e. Displays the selected Test Name Description. f. Displays the OTR Value and the selected OTR View By. used for the OTR Calculation in Seconds, Minutes, Hours or Days as per the selection of the X-Axis Scale in the Film Permeation screen. h. Displays the Permeation Setting (Atmospheric Pressure, Volume and Area) for the selected Test. i. j. Displays the Data Points for the Selected Test in the Log. Displays the Data Points for the Selected Test in a Graph in Seconds, Minutes, Hours or Days as per the selection of the X-Axis Scale in the Film Permeation screen

g. Displays the Starting and Final Data Points and the Duration

28 of 29 7/13/2009

2.

Cancel: Clicking the Cancel button will not open the Print Report window and cancel the printing.

5.11 Capture To Capture an Oxygen reading, identify a specific Chamber Serial Number based upon the options noted above. If the Capture button is clicked without entering anything in the Chamber Serial Number field, the application will display an error message: Chamber Serial Number cannot be blank. If the Capture button is clicked after entering a Chamber Serial that is not in the database, the application will display an error message: Could not find the entered Chamber Serial Number. If the selected Chamber Serial Number and Test is active, clicking the Capture button will capture a new Oxygen reading, update the Log, update the Graph and save the Data Point in the database. If the selected Chamber Serial Number and Test is closed, clicking the Capture button will display an error message: Could not take Oxygen Measurement for Test Name. Test is Closed..

29 of 29 7/13/2009