Professional Documents
Culture Documents
Targets
Describe the physical principles behind each imaging modality. Overview of the engineering solutions. Future trends. Take away knowledge for our personal health and have some fun.
Imaging Modalities
X-ray
CT
PET
Ultrasound
MRI
Digital X-rays
Spectrum
~124keV ~511keV ~124eV 6-8um
30um
X-ray Generation
Courtesy of Siemens
X-ray Machine
DETECTOR
Courtesy of Hologic
Indirect
Scintillator
ADC Readout
PFL- Aachen, M. Overdick, 11 Sept 2002, IWORID 2002
Addressing
AFE-XR0064 operation
1. CDS samples offset. 2. The panel control turns on the TFTs of a new column of pixels. 3. The charge is integrated (needs about 14us). 4. The CDS takes the integrated values and subtracts the offsets. 5. We can now RST the integrators. CDS still holds the analog values. 6. Analog values are muxed to the ADC inputs.
Readout time
142mW For 128 lines (1 pixel/line): Tmin = 27.8us
142mW 130mW
Example: 1536 * 1536 panel Divide panel on 2 blocks of 768 columns, each with 24 AFEXR0064: 768*27.8us = 21.35ms FR >30fps
CT Computer Tomography
The machine
3 revolutions per second 1000 profiles per revolution
3KSPS/pixel
GE
1s
PQS T
GE
Siemens X-ray
Photodiodes
Example
Flux * T * Exray * Yieldscintillator * Ephoton * * S * 1/T
<1.9Gp/(mm2.s) 100keV >15kphotons/MeV 2.6eV 0.7 0.3 6k # of x-ray photons 3e5 photons Incident energy 3e10eV # of photons from scintillator 475M light photons Visible light energy out from scintillator 1.2GeV Visible light at the photodiode 0.86GeV Charge 40pC Current 250nA
The spec
Sampling rate: 5 - 10KSPS. Maximum charge/current: 150pC/150nA Noise: <1fCrms, <1 to 1.5pArms Linearity: +/- 1ppm of FSR + % of reading Null offset drift with temperature mWs/channel
Switched Integrator
CINT
Photodiode Current + +
A
ADC
20bits 6KSPS (x2) 7mW/channel
CINT
FS 100s pC
DDC232
The machine
The detector
Example (from Derenzo): NaI(TI) - 3.3 cm Light output: 50k (38k?) photons/MeV Principal Decay time: 230 ns Index of refraction: 1.85 15000 photons at photocathode 3000 photoelectrons at first dynode 3.109 electrons at anode
Centroid
To ADCs Position Anger logic
4 1 2
1 2 3 4 3
Xm
x E = E
i i i
mi
mi
Front-end
SE DIFF
SE DIFF
Xm
SE DIFF
x E = E
i i i
mi
mi
SE DIFF
DSP
AMP
Radiation
Natural background: 2.4mSv/year Air travel crew: 3mSv/year Radiation worker federal limit: 50mSv/year Dental radiography: 0.01mSv Chest radiography: 0.1mSv Mammography: 0.7mSv PET/SPECT : 7mSv Chest CT: 8mSv Pelvic/abdomen CT: 10mSv CT Angiography: 15mSv 50% of cases die in 30 days: 3Sv
References: http://www.xrayrisk.com/
Ultrasound
*@?!?
Principle
Physics (I)
Substance Fat Muscle Compact bone Air c [m/s] 1470 1568 3600 331 [g/cm3] 0.97 1.04 1.7 0.0013 Z [105 Rayl] 1.42 1.63 6.12 43.10-5
Absorption [dB/MHz.cm] 0.5 2 4-10
Reflections
Depth
Physics (II)
c = 1560m/s
Frequency [MHz] 2 5 10 15 Wavelength [mm] 0.78 0.31 0.16 0.1
RAxial
c c = = 6 dB 2.FBW 2 2f
Lateral resolution [mm] 3 1.2 0.6 0.4 Axial resolution [mm] 0.8 0.35 0.2 0.15
RLateral
c 2r = f w. cos
Scan
Electrical
Mechanical Mechanical
Courtesy of GE
B-mode
Ultrasound System
Transducer Transducer Transducer Transducer Transducer Transducer Transducer Transducer
Tx Buffer Amp
HV Mux/ Demux
T/R Switches
OPA695
DAC5652
ADC
AFE5805 ADS1605/06/25/26
ADS527x/8x
Tx beamformer
Transducer array 1 2 3 4
Rx beamformer
Delay
Transducer array 1 2 3 4
Apodization
Receive Beamforming
Transducer array 1 2 3 4 A1 ADC samples C1
Line 1
A1 A2
B1 B2
C2 Line 2
Interpolated
ADCs 43 21
5ns
CW VCA
LNA
N0_A N1_A N2_A GND -HV1 -HV2
ADC
12bit 40MSPS
30.00
IR Noise (nV/rt(Hz))
TI AFE5805
0.4
0.6
0.8
1.2
Vcontrol (V)
The machine
Liquid Nitrogen Liquid Helium Magnet Coils RF Volume Coils
Surface Coils
Bo
Y X Z
www.e-mri.org
Matter
www.e-mri.org
Magnetization vector
www.e-mri.org
Excitation
Relaxation
Gradient field
Bo
1. Slice: Gradient on excitation. Only the nucleuses with a Larmor frequency equal to the RF pulse frequency will be excited. 2. Frequency encoding: Gradient on relaxation. Looking at the spectrum of the received RF, the spatial position of the nucleuses can be identified.
RF Signal Characteristics
Carrier depending on Larmor frequency (depending on fix magnet). BW below 1MHz (depending on gradient coil). RF emissions from the human body have a decay time which depends on the nature of the tissue. Decay time can vary between 5ms and 200ms. The emissions SNR is dependent upon: Field strength and homogeneity RF coil design/tuning Distance of coil to the sample The sample itself For medical applications the typical dynamic range is ~80dB. Signal amplitude can range from Vs to several mVs.
Signal Chain
Slide 44
Volumes/Prices
Summary
X-ray CT PET Ultrasound MRI
Issues
Quality Radiation
Cost Radiation
Quality
Cost
References
www.e-mri.org http://www.cis.rit.edu/htbooks/mri/ http://www.nlm.nih.gov/medlineplus/tutorials/mri/ht m/index.htm http://www.mri-tutorial.com/ http://users.fmrib.ox.ac.uk/~stuart/thesis/chapter_ 2/contents.html