Professional Documents
Culture Documents
Copyright 2002 BP Pipelines (North America), Inc. In this book, the corporate entity BP Pipelines (North America) Inc. is usually referred to as BP Pipelines or BP.
BP Pipelines
Foreword
The BP Pipelines Measurement Manual is broken into three parts. Part I covers crude oil and petroleum products, Part II covers natural gas, and Part III covers chemicals and petrochemicals. This book, Part I, documents the current policies of BP Pipelines regarding measurement of crude oil and liquid petroleum products. BP Pipelines objective is to provide the basis for measurement that is fair and equitable for all parties. To this end, the manual is based on recognized industry standards, industry practices, and prudent operating procedures. The procedures and policies in this manual are to be used by BP Pipelines employees and contract personnel engaged in custody transfer. It is not intended to serve as a training manual. If you use the procedures and equipment described in this manual, you should be able to pass an audit. The measurement policies in this manual are to be used throughout BP Pipelines operations. BP Pipelines recognizes, however, that unique circumstances, such as regulatory or contractual requirements or unusual operating conditions, may require some variance from the standard policy or procedure. In such cases, exceptions will be considered; however, a request must be submitted to the BP Pipelines Measurement Team. This request must contain all relevant information pertaining to the particular operation and the basis for the exception request. Measurement Team Technical Services Engineering and Maintenance Department BP Pipelines (North America) Every effort has been made to ensure the accuracy and reliability of the information in this manual; however, the Measurement Team solicits feedback from users regarding perceived errors.
September 2002
Contents
List of Illustrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .iv List of Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii 1. Manual Sampling in Tanks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.12 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.17 2. Gravity and Temperature Measurement in Tanks . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.14 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.17 3. Gauging Lease Tanks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.13 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.14 4. Gauging Large Tanks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.16 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.17 5. Testing Crude Oil for Suspended Sediment and Water. . . . . . . . . . . . . . . . . . . . . . 5.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.4 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.15 6. Other Tests for Crude and Products. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2 Tests for Crude Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3 Tests for Products . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.13 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.17 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.18
September 2002
BP Pipelines
7. Tank Strapping. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.12 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.14 8. Seals and Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.3 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.8 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.9 9. LACT/ACT Verification and the ELM System. . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.4 Procedures for Checking the Operation of LACT and ACT Units . . . . . . . . . . 9.5 Overview of the ELM System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.11 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.14 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.16 10. Automatic Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.4 Procedures for Handling Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.5 Design and Operation of an Automatic Sampling System . . . . . . . . . . . . . . . 10.8 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.12 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14 11. Positive Displacement Meters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.3 Inspecting the Meter and Accessory Equipment . . . . . . . . . . . . . . . . . . . . . . . 11.4 Design and Operation of a Positive Displacement Meter . . . . . . . . . . . . . . . . 11.6 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.13 12. Turbine Meters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.3 Inspecting the Meter and Accessory Equipment . . . . . . . . . . . . . . . . . . . . . . . 12.4 Design and Operation of a Turbine Meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.5 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.14 13. Other Meters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.2 Orifice Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.3 Ultrasonic Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.5
September 2002
ii
BP Pipelines
Coriolis Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.7 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.9 14. Proving a Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14.4 When to Prove a Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14.5 Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14.6 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14.14 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14.16 15. Design of Prover Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15.2 Installation and Operating Requirements for Provers . . . . . . . . . . . . . . . . . . . 15.3 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15.9 16. Waterdraw Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.2 When To Calibrate a Prover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.3 Preparation Procedures for Waterdraw Calibration. . . . . . . . . . . . . . . . . . . . . 16.4 Procedures for Calibrating a Pipe Prover Using the Waterdraw Method . . . . 16.8 Calibration Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.11 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.12 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.18 17. Claims and Adjustments, Using SMART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17.1 Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17.3 Procedures for Making Claims and Adjustments . . . . . . . . . . . . . . . . . . . . . . 17.4 Procedures for Using the SMART Measurement Software for Ticketing and Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17.6 More About It . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17.8 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17.14 Abbreviations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G.1 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G.3 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .I.1
September 2002
iii
BP Pipelines
List of Illustrations
Figure 1.1. Figure 1.2. Figure 1.3. Figure 1.4. Figure 2.1. Figure 2.2. Figure 2.3. Figure 3.1. Figure 3.2. Figure 3.3. Figure 4.1. Figure 4.2. Figure 4.3. Figure 4.4. Figure 4.5. Figure 5.1. Figure 5.2. Figure 5.3. Figure 6.1. Figure 7.1. Figure 7.2. Figure 7.3. Figure 7.4. Figure 7.5. Figure 7.6. Thief . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.5 Examples of labels on sample containers . . . . . . . . . . . . . . . . . . . . . . .1.7 Sample bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.9 Where to take samples in a large tank . . . . . . . . . . . . . . . . . . . . . . . . .1.14 Using a thermohydrometer to measure API gravity . . . . . . . . . . . . . . .2.6 Reading a hydrometer for transparent liquids . . . . . . . . . . . . . . . . . . . .2.7 Locations for determining the temperature of the liquid in a large tank with liquid height more than 15 feet . . . . . . . . . . . . . . . . . . . . . . .2.8 Checking the tank before gauging . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.6 Checking the depth of settled S&W in the tank. . . . . . . . . . . . . . . . . . .3.9 Innage method of gauging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.10 Gauging an external floating roof tank from the platform. . . . . . . . . . .4.5 Gauging an external floating roof tank from the floating roof. . . . . . . .4.6 Gauging an internal floating roof tank . . . . . . . . . . . . . . . . . . . . . . . . . .4.7 Checking the tank before gauging . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.8 Innage method of gauging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.10 Free water in the bottom of centrifuge tube when preparing water-saturated toluene . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.6 Solubility of water in toluene. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.6 Sequence for determining S&W by the field centrifuge method using 100-ml (6-inch) centrifuge tubes . . . . . . . . . . . . . . . . . . . . . . . .5.10 Bar chart for determining haze. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6.16 Recommended tape paths for measuring tank circumference on a welded tank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7.5 Recommended tape paths for measuring tank circumference on a bolted tank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7.6 Vertical measurements of a tank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7.7 Measuring the shell height of a tank with a protruding bottom . . . . . . .7.8 Measuring the shell height of a tank with a recessed bottom . . . . . . . .7.8 Determining effective inside tank height . . . . . . . . . . . . . . . . . . . . . . .7.10
September 2002
iv
BP Pipelines
Figure 7.7. Figure 8.1. Figure 9.1. Figure 9.2. Figure 9.3.
Sample page of strapping report . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7.13 Boxcar seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8.8 LACT unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9.6 Sample LACT/ACT inspection form . . . . . . . . . . . . . . . . . . . . . . . . . .9.14 Blank LACT/ACT inspection form . . . . . . . . . . . . . . . . . . . . . . . . . . .9.15
Figure 10.1. Design of an automatic sampling system. . . . . . . . . . . . . . . . . . . . . . .10.8 Figure 10.2. Fixed sample receiver (stationary sample pot). . . . . . . . . . . . . . . . . . .10.9 Figure 10.3. Portable sample receiver (pot) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10.10 Figure 10.4. Sample mixing system for portable receivers . . . . . . . . . . . . . . . . . .10.11 Figure 11.1. Sealing points on a PD meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11.4 Figure 11.2. Typical positive displacement meter . . . . . . . . . . . . . . . . . . . . . . . . . .11.6 Figure 11.3. Design of a PD meter installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .11.8 Figure 12.1. Conventional turbine meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12.5 Figure 12.2. Helical turbine meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12.6 Figure 12.3. Typical turbine meter installation for refined products . . . . . . . . . . . .12.8 Figure 12.4. Turbine meter with preamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12.11 Figure 12.5. Turbine meter without preamp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12.11 Figure 13.1. Orifice meter installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13.3 Figure 13.2. Ultrasonic meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13.5 Figure 13.3. Path of the sound waves in an ultrasonic meter . . . . . . . . . . . . . . . . . .13.6 Figure 13.4. Coriolis meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13.7 Figure 13.5. Vibration of the flow tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13.8 Figure 13.6. Coriolis effect (twisting of the flow tubes) . . . . . . . . . . . . . . . . . . . . .13.8 Figure 14.1. Sample proving report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14.15 Figure 16.1. Waterdraw calibration unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16.7 Figure 16.2. Waterdraw calibration worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . .16.13 Figure 17.1. Example of a claim letter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17.9 Figure 17.2. Example of a notarized affidavit . . . . . . . . . . . . . . . . . . . . . . . . . . . .17.11 Figure 17.3. Example of a SMART tank custody transfer ticket . . . . . . . . . . . . . .17.12 Figure 17.4. Example of a SMART meter ticket for a lease meter . . . . . . . . . . . .17.13
September 2002
BP Pipelines
List of Tables
Table 1.1. Summary of Sampling Procedures and Applications. . . . . . . . . . . . . . . . . 1.13 Table 1.2. Samples for Determining S&W . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.14 Table 1.3. Number of Samples to Take from a Product Tank. . . . . . . . . . . . . . . . . . . 1.15 Table 1.4. Recommended Mixing Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.16 Table 2.1. Locations for Determining Temperature in Small and Large Tanks . . . . . 2.14 Table 2.2. Recommended Immersion Times for PETs and Cupcase Woodback Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.15 Table 2.3. Correction of API Gravity to 60F for Generalized Crude Oils. . . . . . . . . 2.16 Table 5.1. Size of Test Sample Based on the Expected Water Content . . . . . . . . . . . 5.12 Table 6.1. Typical Sulfur Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3 Table 6.2. Quality Guidelines for Metals in Crude Oils . . . . . . . . . . . . . . . . . . . . . . . 6.7 Table 6.3. Typical Distillation Fractions of a Crude Oil and Their Uses . . . . . . . . . . 6.10 Table 6.4. Typical Minimum Flash Points of Various BP Pipelines Products . . . . . . 6.15 Table 7.1. Fill Condition Required for Different Tanks while Strapping . . . . . . . . . . 7.12 Table 10.1. Recommended Mixing Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.12 Table 10.2. Typical Sizes of Sample Receivers for Various Operations . . . . . . . . . . 10.13 Table 11.1. Recommended Combinations of Meter Size, Measuring Element, and Gear Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.9 Table 11.2. Recommended Operating Range for Various Sizes of PD Meters. . . . . . 11.10 Table 11.3. Effect of Changes in the Conditions of the Fluid and PD Meter on Measurement Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11 Table 12.1. Recommended Operating Range for Various Sizes of Conventional Turbine Meters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9 Table 12.2. Recommended Operating Range for Various Sizes of Helical Turbine Meters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10 Table 12.3. Effect of Changes in the Conditions of the Fluid and Turbine Meter on Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.13 Table 14.1. Causes of Meter Factor Fluctuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14.10 Table 15.1. Effect of Various Criteria on the Design of a Prover System . . . . . . . . . 15.3 Table 15.2. Approximate Detector Error for Pipe Provers . . . . . . . . . . . . . . . . . . . . . 15.6
September 2002
vii
BP Pipelines
Table 15.3. Typical Actuator Times for Four-way Valves . . . . . . . . . . . . . . . . . . . . . 15.7 Table 15.4. Type of Sphere to Use for Various Liquids . . . . . . . . . . . . . . . . . . . . . . . 15.8 Table 16.1. Type of Sphere to Use for Various Liquids . . . . . . . . . . . . . . . . . . . . . . . 16.14 Table 16.2. Sphere Roundness Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.15 Table 16.3. Minimum Inflation Percentages for Given Pipe Sizes . . . . . . . . . . . . . . . 16.17 Table 17.1. When to File a Claim or Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17.4
September 2002
viii
Chapter 1
Scope
This chapter includes the procedures for taking manual samples in upright cylindrical tanks of all sizes containing crude oil or products.
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
Introduction
Manual sampling involves lowering a thief or bottle into the tank one or more times to get representative samples of the liquid. The purpose of sampling crude oil is to determine suspended and settled sediment and water, API gravity, and occasionally other properties like vapor pressure. Liquid petroleum products stored in tanks require sampling for API gravity, vapor pressure, and other properties like ash, haze, and color. Manual sampling is necessary in tanks that do not use a LACT or ACT unit. You take samples to determine S&W and API gravity in the eld and return them to the laboratory for other types of testing (see Chapter 6, Other Tests for Crude and Products for more information).
September 2002
1.3
BP Pipelines
Procedures
Table 1.1 is a summary of sampling procedures for different types of tanks and liquids. Table 1.2 and Table 1.3 show how many samples to take in different tanks. Table 1.4 shows how to mix samples for different tests. You will nd all these tables under More About It at the end of this chapter. Although sampling procedures are the same for crude oil and products, you must make sure the samples are representative of the entire volume in the tank, either by turning on mixers or taking a greater number of samples.
Be sure to wait at least 2 hours after turning off the mixer before gauging the height of the liquid.
Types of Samples
This chapter gives procedures for taking the following types of samples: Spot samplebottle or thief Composite spot samplebottle or thief All-levels samplebottle Running samplebottle Sample cocks (taps) on tanks or pipesbottle
September 2002
1.4
BP Pipelines
Hook
15 14 13 12 11 10 9 8 7
Petcocks
6 5 4 3 2
Trip Rod
Safety Reminder
Do not replace the sample cord with a cord containing synthetic materials. The cord must be 100% cotton to prevent a buildup of static electricity. Saturate new cotton cords with oil before using them the rst time to assure that they are conductive.
September 2002
1.5
BP Pipelines
Before sampling, assure that the conditions are safe, as listed under Safety on the Quick Reference page. If you nd any unsafe conditions or if the security of the tank has been compromised, do not take the samples and report these conditions to your supervisor.
2. Cock the valve at the bottom of the thief in the open position and trip the hook in the eye of the trip rod. Always lower the thief in the open position so that the thief lls from the bottom. Always start at the top and work down so you disturb the oil as little as possible. Do not disturb the cupcase thermometer hanging inside the tank, if you are using one (see Chapter 2, Gravity and Temperature Measurement in Tanks).
3. Lower the thief to the proper level. For lease tanks, take an upper sample just below the surface of the liquid, a middle sample from the center of the liquid, and a lower sample just above the suction line. For large tanks, see Table 1.2 and Figure 1.4 under More About It at the end of this chapter for proper sampling levels.
4. Jerk the cord sharply to close the bottom valve on the thief and trap the sample. 5. Pull the thief to the surface. 6. Pour about 6 inches of the liquid in the thief back into the tank. 7. Pour the sample into a small, clean sample container until it is about 3/4 full. Cap the sample container, wipe it clean, and label it as an upper or lower sample (see Figure 1.2). Put it into a compartment in your tool box.
September 2002
1.6
BP Pipelines
Sample bottles must be clean and dry be clear or brown glass be unpigmented linear polyethylene for certain oils have cork or glass stoppers or screw caps not have rubber stoppers be clearly labeled with bottle contents
Typical label
Figure 1.2. Examples of labels on sample containers 8. If you are compositing samples, measure out the proper amount of sample into a graduated cylinder and put it in the sample container. 9. Pour the remaining liquid back into the tank.
September 2002
1.7
BP Pipelines
7. Carefully raise the thief 2 to 4 inches and then allow it to bump the bottom hard enough to close the valve. 8. Pull the thief to the surface. 9. Slowly pour the contents over a glass plate or your gloved hand. Stop when you see S&W. 10. Return the thief to an upright position and measure the distance between the end of the trip rod and the surface of the S&W. This measurement is the depth of the S&W layer on the tank bottom. 11. Record the level of settled S&W. 12. Clean the thief.
September 2002
1.8
BP Pipelines
Lead weight
Figure 1.3. Sample bottle First assure that the conditions are safe, as listed under Safety on the Quick Reference page. If you nd any unsafe conditions or if the security of the tank has been compromised, do not take the samples and report these conditions to your supervisor.
September 2002
1.9
BP Pipelines
3. Attach a weighted line to the bottle or place the bottle in a sample cage. 4. Put the stopper in the bottle. 5. Lower the weighted, stoppered bottle to the proper depth. 6. When the container reaches the selected level, pull out the stopper with a sharp jerk of the line and allow the bottle to ll completely. 7. When you judge that the container is full, raise the bottle. 8. If the bottle is not at least 3/4 full, pour out the contents and repeat steps 47. 9. If taking only one sample, pour off a small amount from the full bottle, and put the stopper in it immediately. 10. Repeat steps 58 for each sample needed. 11. Close the sample container, and return it to your toolbox.
All-Levels Sample
1. Use clean, dry equipment. Before using a bottle, rinse it with Stoddard solvent or another naphtha of similar volatility. If necessary, use sludge solvents to remove all traces of sediment and sludge from previously used containers. Wash the bottle with a strong soap solution, rinse it thoroughly with tap water, then rinse it with distilled water. If you are sampling crude oil in the eld, you may omit this step. Dry the bottle either by blowing a current of clean, warm air into it or by placing it in a hot, dust-free cabinet at 100F or higher.
2. Lower the weighted, stoppered bottle as near as possible to the draw-off level. 3. Pull out the stopper with a sharp jerk of the line, then raise the bottle at a uniform rate so that it is about 3/4 full as it emerges from the liquid. For light products or deep tanks, a restricted opening may be needed to avoid lling the bottle before it reaches the surface of the liquid. The bottle should not be more than 3/4 full.
September 2002
1.10
BP Pipelines
4. If the container is full when it emerges from the liquid, pour the liquid back and try again. The goal is to get portions of sample from all levels of the tank. If the bottle is full, it did not sample any of the oil past the point that it lled. If you are unable to ll the sample container at the proper rate, use a different method, such as taking multiple spot samples, to obtain a representative sample.
Running Sample
1. Use clean, dry equipment. Before using a bottle, rinse it with Stoddard solvent or another naphtha of similar volatility. If necessary, use sludge solvents to remove all traces of sediment and sludge from previously used containers. Wash the bottle with a strong soap solution, rinse it thoroughly with tap water, then rinse it with distilled water. If you are sampling crude oil in the eld, you may omit this step. Dry the bottle either by blowing a current of clean, warm air into it or by placing it in a hot, dust-free cabinet at 100F or higher.
2. Lower the unstoppered bottle at a uniform rate as nearly as possible to the level of the bottom of the outlet connection or swing line. 3. Raise the bottle to the top of the oil at the same rate so that it is about 3/4 full when withdrawn from the liquid. For light products or deep tanks, you may need a notched cork or other restricted opening to avoid lling the bottle too quickly. The bottle should not be more than 3/4 full.
4. If the container is full when it emerges from the liquid, pour the liquid back and try again. The goal is to get portions of sample from all levels of the tank. If the bottle is full, it did not sample any of the oil past the point that it lled. If you are unable to ll the sample container at the proper rate, use a different method, such as taking multiple spot samples, to obtain a representative sample.
September 2002
1.11
BP Pipelines
More About It
Here you will nd tables and instructions for the following: Summary of manual sampling procedures for products and crude oil Number of samples of crude oil to take with a bottle or thief Number of samples to take from a product tank Mixing procedures
September 2002
1.12
BP Pipelines
Table 1.1.
Liquids of more than 16 lb. and not more than 26 lb. RVP (e.g., propane, butane, and pentane) Liquids of more than 16 lb. and not more than 26 lb. RVP (e.g., propane, butane, and pentane) Liquids of more than 2 lb. and not more than 16 lb. RVP (e.g., crude oil and products like gasoline, distillates, and kerosene) Liquids of 2 lb. RVP or less (e.g., heavy fuel oils and asphalt) Liquids of 2 lb. RVP or less (e.g., heavy fuel oils and asphalt) Liquids of 2 lb. RVP or less (e.g., heavy fuel oils and asphalt) Liquids and semi-liquids of 2 lb. RVP or less (e.g., heavy fuel oils and asphalt)
Storage tanks, ship and barge tanks, tank cars, and tank trucks
Bottle sampling
Tube sampling
Free or open-discharge streams, open tanks or kettles with open heads, tank cars, tank trucks, drums Pipelines
Dipper sampling
September 2002
1.13
BP Pipelines
Table 1.2.
Tank Capacity/Liquid Level Upper 1,000 barrels or less More than 1,000 barrels (no mixer) 1 Middle 1* Lower 1
Liquid level less than 15 feet Liquid level more than 15 feet
More than 1,000 barrels (with mixer)
* **
1 1 1
1* 1** 1**
1 1 1
The middle sample here is taken to determine the API gravity of the liquid. This sample is not included in the composite sample or in the tests for S&W. This middle sample may be used to determine the API gravity and then added to the composite sample for testing for S&W, but normally all three samples are tested for gravity and S&W and the results averaged. Alternately the three samples are composited and then tested.
In small tanks, test the upper and lower samples separately for suspended S&W. In addition to these samples, you may also take a bottom sample to determine the height of the settled S&W. Figure 1.4 shows where to take samples in a large tank.
UPPER midpoint of upper 1/3 MIDDLE midpoint of middle 1/3 LOWER midpoint of lower 1/3
September 2002
1.14
BP Pipelines
Table 1.3.
Over 15 feet
Note: Depending on the actual tests to be performed, additional samples may be required.
September 2002
1.15
BP Pipelines
Mixing Samples
A sample may or may not need to be mixed before testing it, depending on the type of test and how homogeneous the sample is. Because an automatic sampling system takes samples over a period of many hours, some settling almost always occurs. You can mix a sample with a stand-alone power mixer, with an internal mixer in the sample pot, or by shaking it. When using a stand-alone power mixer, be sure to use the correct type for the container or pot you have, as the mixer/container combination has been tested and proven effective. Table 1-4 lists the mixing recommendations for various tests.
Table 1.4.
Density for crude and heavy fuels Sediment and water Density for other hydrocarbons Vapor pressure Cloud point
*
X X X X X
Sample transferred from a container. For tests not listed, refer to the specic test procedure.
September 2002
1.16
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 3.1A Standard Practice for the Manual Gauging of Petroleum and Petroleum Products 2. API Manual of Petroleum Measurement Standards, Chapter 8.1 Standard Practice for the Manual Sampling of Petroleum and Petroleum Products 3. API Manual of Petroleum Measurement Standards, Chapter 18.1 Measurement Procedures for Crude Oil Gathered from Small Tanks by Truck 4. Pipelines (NA) Business Unit Safety Manual
September 2002
1.17
Chapter 2
Scope
This chapter includes the procedures for determining API gravity and temperature of crude oil and liquid petroleum products in nonpressurized tanks.
Use the upper, middle, and lower samples to measure gravity in tanks that hold more than 1,000 barrels. These samples may be composited for testing, or they may be tested separately and the results averaged. For tanks of less than 1,000 barrels (lease tanks), use a clean thermohydrometer and leave it in the thief for at least 3 minutes (longer in heavier oils or during extreme heat or cold). For tanks holding more than 1,000 barrels, pour the samples into a hydrometer cylinder in the laboratory for testing with a thermohydrometer. When the liquid is opaque, deduct 0.1API from the gravity reading to correct for the meniscus. Record the gravity to the nearest 0.1API.
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
Introduction
After sampling the liquid in the tank, you will gauge the tank to determine the amount of liquid in it. Part of the gauging process includes measuring the API gravity and temperature of the liquid. A tanks volume varies due to expansion and contraction of the liquid and the metal tank shell with changes in temperature. Each tank has a capacity table that is based on the volume at a certain temperature. Getting an accurate temperature reading allows you to correct the volume for the actual temperature of the liquid in the tank. Density, like volume, depends on temperature, and so you will also measure the temperature of the sample while determining the API gravity. One purpose of measuring API gravity is to allow conversion of the volume you measure by gauging to the volume at the standard temperature of 60F. API gravity is also a property of the oil that may affect the price paid for the oil.
September 2002
2.4
BP Pipelines
Procedures
The contents of large tanks containing crude oil should be thoroughly mixed before you make temperature measurements (lease tanks do not have mixers). Because different grades of crude require different mixing times, a facility may conduct tests to determine the ideal mixing time for each grade it handles. Unless otherwise instructed, follow these guidelines: Mix tanks that are less than 1/3 full for at least 2 hours. Mix tanks that are more than 1/3 full for 4 hours.
Be sure to wait at least 2 hours after turning off the mixer before gauging the height of the liquid to allow the liquid to stop moving.
Thermohydrometer Method
1. Use a sample from the middle of the column of liquid (small tanks) or 3 samples from the top, middle, and bottom of the column of liquid (large tanks). For small tanks, you may hang the thief in the hatch and determine the gravity before transferring the sample to a storage container. For large tanks, transfer the sample, without splashing, to a clean hydrometer cylinder for the test.
2. Insert a clean thermohydrometer into the thief or hydrometer cylinder until it oats freely, and then push it down another 1/4 inch (see Figure 2.1). 3. Leave the thermohydrometer in the thief or hydrometer cylinder for at least 3 minutes to allow the temperature to stabilize. Note: Do not allow the thermohydrometer to touch the side of the thief. 4. After at least 3 minutes, read the observed gravity at eye level. Take the measurement at the bottom of the meniscus for clear liquids (see Figure 2.1). 5. When the liquid is too opaque to see the meniscus (for example, crude oil), deduct 0.1API to correct for it.
September 2002
2.5
BP Pipelines
Stem
Liquid sample
Temperature scale
Thief
Figure 2.1. Using a thermohydrometer to measure API gravity 6. Record the gravity to the nearest 0.1API. 7. Read the temperature at eye level. Note: The thermohydrometer bulb must remain in the liquid while you are reading it. 8. Record the observed temperature to the nearest 1.0F.
Hydrometer Method
1. Transfer the sample to a clean hydrometer cylinder without splashing. Remove any air bubbles from the surface by touching them with a piece of clean paper towel or lter paper before inserting the hydrometer.
2. Place the cylinder upright in a location away from air currents to keep the temperature of the sample liquid from changing during the test. If the temperature of the sample is much above or below the ambient temperature, put the cylinder in a constant-temperature bath.
September 2002
2.6
BP Pipelines
3. Lower the hydrometer gently into the sample. Do not let the hydrometer stem get wet above the level where it oats in the liquid (see Figure 2.2). 4. Use the thermometer to stir the sample continuously. Keep the mercury completely immersed. Keep the hydrometer stem dry while stirring the sample.
5. Read and record the temperature to the nearest 0.5F. Immerse the thermometer as far as you can while still being able to read the scale. Wait about 30 seconds for the reading to stabilize. Keep the thermometer immersed while reading the temperature.
6. Push the hydrometer about 2 scale divisions into the liquid and release it. If the liquid has a low viscosity, spin the hydrometer slightly as you release it. Allow the hydrometer to come to rest and wait for all air bubbles to come to the surface.
7. Read the hydrometer scale to the nearest 0.0001 (relative density) or 0.05API. For transparent liquids, look at the scale with your eye just below the liquids surface, then raise your head until the surface appears to become a straight line cutting the scale (see Figure 2.2). For opaque liquids, read the scale at eye level and subtract 0.1 API to correct for the meniscus.
Meniscus
September 2002
2.7
BP Pipelines
8. Immediately after reading the hydrometer scale, stir the sample with the thermometer and record the temperature to the nearest 0.5F. If this temperature differs by more than 1F from your rst measurement, repeat steps 7 and 8 until the temperature stabilizes. 9. Record the nal hydrometer reading to the nearest 0.1API and the temperature reading to the nearest 1F. 10. Convert the observed API gravity to API gravity at standard temperature (60F) (see Table 2.3 under More About It).
Hatch
Upper temperature
Liquid height
} } }
1/3
1/3
1/3
Figure 2.3. Locations for determining the temperature of the liquid in a large tank with liquid height more than 15 feet
September 2002
2.8
BP Pipelines
5. Repeatedly raise and lower the probe about 1 foot in each direction until the temperature reading stabilizes; that is, when the readout varies by no more than 0.2F for at least 30 seconds. 6. After the temperature has stabilized, read and record the temperature. 7. Repeat steps 4, 5, and 6 in large tanks for each level needed. 8. When taking more than one reading, average the readings, round to the nearest 0.1F, and record the average. T1 + T2 + T3 ------------------------------- = average temperature 3 9. Clean the thermometer (and tape) with a solvent and dry it with a cloth.
September 2002
2.9
BP Pipelines
Safety Reminder
The temperature must be 120F or less. If the tank is above 120F, notify your supervisor.
4. After the temperature has stabilized, pull the thermometer from the liquid. Keep the assembly sheltered below the edge of the hatch to prevent the wind or air temperature from affecting the reading. Make sure the cup is full while you are reading the temperature. Do not hold the assembly by the brass cup, as heat from the hand may change the temperature of the oil. Read the temperature to the nearest 0.5F immediately after pulling the thermometer out.
5. For a custody transfer repeat steps 3 and 4. For inventory or for small tanks, usually only one middle temperature reading is all that is required.
September 2002
2.10
BP Pipelines
If the readings agree within 2F, average them and record the average to the nearest 0.1F. T1 + T2 ------------------ = average temperature 2
If the readings differ by more than 2F, take additional readings and average the results.
6. For tanks that hold 1,000 bbls or more, repeat steps 3, 4, and 5 at the top, middle, and bottom of the column of oil. Average the 3 temperature readings: T1 + T2 + T3 ------------------------------- = average temperature 3 7. When taking readings from more than one location in the tank, average the readings, round to the nearest 0.5F, and record the average. 8. After using the thermometer in a heavy, high-viscosity, or high-pour-point oil, clean all parts of the assembly. Rinse the cupcase thermometer with Stoddard solvent or another naphtha of similar volatility. If necessary, use sludge solvents to remove all traces of sediment and sludge.
Safety Reminder
The temperature must be 120F or less. If the tank is above 120F, notify your supervisor.
Verifying PETs
A PET must be calibrated in a laboratory before using it the rst time and recalibrated once a year thereafter.
September 2002
2.11
BP Pipelines
Spot-check the PET before each use or once a day, and make a more thorough check monthly. Do not use a PET that does not pass these checks. Document the monthly checks and keep the documentation on le at the location for at least two years.
Daily Check
1. Place the PET and an ASTM glass thermometer in a liquid. 2. After the temperatures have stabilized, compare the readings of the two thermometers. 3. If the readings differ by more than 0.5F, do not use the PET in a custody transfer until it has been recalibrated.
Monthly Check
1. Place the PET and a NIST-certied or equivalent thermometer side by side in a circulating hot-water bath, if available. Leave them undisturbed for at least 10 minutes. Compare the readings of the two thermometers. If they differ by more than 0.5F, do not use the PET.
2. Place the PET and a NIST-certied or equivalent thermometer side by side in ice water. Leave them undisturbed for at least 10 minutes. Compare the readings of the two thermometers. If they differ by more than 0.5F, do not use the PET.
3. If the PET is out of calibration, follow the manufacturers procedures for recalibration. Note: You will probably do this with a calibrator box attached to the PET. 4. Check the junction between the cable and the probe for damage. 5. Check the cable insulation for cuts, breaks, and abrasions. 6. Check the grounding cable for damage. 7. Check the case body for cracks or damage. 8. If the PET is damaged, do not use it until it is repaired.
September 2002
2.12
BP Pipelines
cause an inaccurate reading. 2. Check that the paint on the engraved scale is still present. Do not use a thermometer that has lost this paint. It is too difcult to see the reading. 3. Check that the liquid column has not separated. If the column separates, then rejoins, do not use the thermometer until it has been veried in a laboratory. 4. Compare the reading on the thermometer to that on a similar thermometer. If the readings differ by more than 1.0F, do not use the thermometer.
September 2002
2.13
BP Pipelines
More About It
Here you will nd the following tables and instructions for using them: Table 2.1. Locations for Determining Temperature in Small and Large Tanks Table 2.2. Recommended Immersion Times for PETs and Cupcase Woodback Assemblies Table 2.3 Correction of API Gravity to 60F for Generalized Crude Oils
Table 2.1.
Tank Capacity/Liquid Level 1,000 barrels or less More than 1,000 barrels Liquid level under 10 feet Liquid level 10 to 15 feet
1 2
Middle of liquid height 3 feet from top of liquid surface and 3 feet from bottom of tank Middle of top third Middle of middle third Middle of bottom third
September 2002
2.14
BP Pipelines
Table 2.2.
>50 40 to 49 30 to 39 20 to 29 <20
* **
While measuring, keep the sensor probe in motion by raising and lowering it 1 foot above and below the desired depth. Can be used in either an in-motion or a stationary mode. In motion is dened as repeatedly raising and lowering the assembly 1 foot above and below the desired depth. If additional mass is placed in the liquid near the thermometer (such as a weight to make the cupcase woodback assembly sink), the immersion time of the assembly will be longer than those listed in this table. Immersion times should be established by testing, and all parties involved should agree on the times.
Note: Immersion times are based on test procedures outlined in API MPMS Chapter 7. Failure to use these recommended times may result in incorrect temperature readings.
September 2002
2.15
BP Pipelines
The following table, which is an excerpt from API Table 5A, can be used to correct the observed API gravity at the observed temperature to the standard API gravity at 60F when gauging crude oil. API Table 6A can be used to correct the observed API gravity at the observed temperature to the standard API gravity at 60F when gauging products. Both tables are available at the eld locations or from the Measurement Team. These corrections are made automatically if you input the data into SMART (measurement ticketing software).
Table 2.3.
Observed Temperature
30.0
30.5
31.0
31.5
32.0
32.5
33.0
33.5
Example: If the observed gravity is 31.7 API (round it to 31.5API) and the observed temperature is 41.3F (round it to 41.5F), the API gravity is 32.8API. Take the difference between the observed gravity and API gravity in the table (31.7 - 31.5 = 0.2) The corrected API gravity is 32.8 + 0.2 = 33.0API. Note: API does not recommend interpolation of temperature.
September 2002
2.16
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 3.1A Standard Practice for the Manual Gauging of Petroleum and Petroleum Products 2. API Manual of Petroleum Measurement Standards, Chapter 7 Temperature Determination 3. API Manual of Petroleum Measurement Standards, Chapter 8.1 Manual Sampling of Petroleum and Petroleum Products 4. API Manual of Petroleum Measurement Standards, Chapter 9.1 Hydrometer Test Method for Density, Relative Density (Specic Gravity), or API Gravity of Crude Petroleum and Liquid Petroleum Products 5. API Manual of Petroleum Measurement Standards, Chapter 9.3 Thermohydrometer Test Method for Density and API Gravity of Crude Petroleum and Liquid Petroleum Products 6. API Manual of Petroleum Measurement Standards, Chapter 11.1, Volume I, Table 5A Generalized Crude Oils Correction of Observed API Gravity to API Gravity at 60F 7. API Manual of Petroleum Measurement Standards, Chapter 11.1, Volume I, Table 5B Generalized Products Correction of Observed API Gravity to API Gravity at 60F 8. Pipelines (NA) Business Unit Safety Manual
September 2002
2.17
Chapter 3
Scope
This chapter describes the procedures for measuring crude oil from upright cylindrical tanks with a capacity of less than 1,000 barrels.
The tanks oil level is above the maximum height stenciled on the side of the tank. The suspended S&W is greater than 1.0% (see Chapter 5, Testing Crude Oil for Suspended Sediment and Water). The free water in the tank is greater than 0.3% (see this chapter). (Note: Texas, New Mexico, and Louisiana allow 1.0% free water.) The settled S&W (tank bottoms) is less than 4 inches from the bottom of the pipeline (P/L) connection (see Chapter 1, Manual Sampling in Tanks). Leaks are present in the tank bottom or sides. The oil has not weathered (settled) for at least 2 hours before gauging.
BP Pipelines
September 2002
Exceptions
You may reject the tank if the oil height is less than 1/2 the normal height of the tank (with supervisors approval). The total S&W content must be less than 1.0%. Of this 1.0%, the free water content may be no more than 0.3%. In Texas, New Mexico, and Louisiana, the allowable free water content is 1.0%, with no S&W.
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
Introduction
This chapter includes the procedures for taking the opening and closing gauges of any lease tank that does not use a LACT unit. Gauging provides a measure of the amount of oil transferred from the tank.
September 2002
3.4
BP Pipelines
Procedures
Sampling and S&W testing are somewhat different in lease tanks than in large tanks.
September 2002
3.5
BP Pipelines
Hatch Oil level no greater than stenciled tank height Oil surface no foaming or boiling Tank 101 Lease 49203 Tank Ht. 15' Verify stenciled tank information
Fill lines that extend into top of tank should have notch to prevent siphoning
Opening Gauge
If the conditions are safe, run the tank as follows: 1. Before taking the opening gauge, inspect the tank (see Figure 3.1): Check the tank and valves for leaks or distortions. Check the seal on the tank stop valve for any signs of tampering. Check all bottom and side valves to make sure they are closed and have sealing devices for attaching seals. Make sure the ladders and catwalk are safe. Check the tank and lease numbers stenciled on the tank. Ground yourself before climbing on the tank. Be especially cautious on dry days, since the potential for static electricity is greater in low humidity. Check that the top valves are closed and can be sealed. Open the hatch using all safety precautions. Wear an approved breathing apparatus if H2S is a potential hazard. Make sure the ll lines at the top of the tank are notched to prevent siphoning.
September 2002
3.6
BP Pipelines
Safety Reminder
Do not run the tank if: The oils surface is foaming or boiling. The oil level is higher than the stenciled tank height.
2. Suspend the cupcase thermometer, if used, at the midpoint of the tank. 3. Take an upper sample for S&W testing (see Procedures for Testing Suspended S&W in Lease Tanks by the Field Centrifuge Method in Chapter 5, Testing Crude Oil for Suspended Sediment and Water). Use clean, dry equipment. Cock the valve at the bottom of the thief in the open position and trip the hook in the eye of the trip rod. Lower the thief to just below the surface of the liquid. Jerk the cord sharply to close the bottom valve on the thief and trap the sample. Pull the thief to the surface. Pour about 6 inches of the liquid in the thief back into the tank. Pour the sample into a small, clean sample container until it is about 3/4 full. Cap the sample container, wipe it clean, and label the sample. Put it into a compartment in your tool box. If you are compositing samples, measure out the proper amount of sample into a graduated cylinder and put it in the sample container. Pour the remaining liquid back into the tank.
4. Take and test the middle sample for API gravity (see Chapter 2, Gravity and Temperature Measurement in Tanks for details of testing). Hang the thief containing the sample on the inside of the gauge hatch. Determine the API gravity of the oil. Pour the remaining liquid back into the tank.
5. Follow the same procedures as under step 1 above to take a lower sample just above the suction line and test it for suspended S&W (see Procedures for Testing Suspended S&W in Lease Tanks by the Field Centrifuge Method in Chapter 5, Testing Crude Oil for Suspended Sediment and Water).
September 2002
3.7
BP Pipelines
6. Take an outlet/clearance sample (tank bottoms) to determine the level of settled S&W (see Figure 3.2). Adjust the trip rod so that it will trip the thief shut when it is bumped on the tank bottom (normally at 4 inches). Slowly lower the thief through the liquid and S&W until it touches the bottom of the tank. Let the thief rest on the bottom to allow the S&W to reach its natural level inside the thief. Do not use a pumping motion to force the thief through the S&W. (The length of time depends on the type and temperature of S&W.) Raise and trip the thief to trap an S&W sample. Make sure the distance between the tank bottoms and the bottom of the P/L connection is greater than 4 inches.
September 2002
3.8
BP Pipelines
Notes: Thief
15 14 13 12
1. If there is more S&W in the tank at closing, tank settling time before run was too short (2-hour minimum recommended). 2. If there is less S&W in the tank at closing, the balance has likely been pumped into the pipeline.
Tube
11 10 9 8 7 6 5 4 3
Tank contents
Distance between tank bottoms and bottom of P/L connection must be greater than 4 inches
Figure 3.2. Checking the depth of settled S&W in the tank 7. Take the opening gauge to the nearest 1/4 inch. Two consecutive readings must be the same before recording the opening gauge on the measurement ticket (see Figure 3.3). If necessary, apply a thin coat of water-indicating paste to the bob to read the free-water level. Hold the gauge line over the hatch and allow the plumb bob to sink through the oil.
September 2002
3.9
BP Pipelines
Be sure to keep the gauge tape in contact with the edge of the hatch to prevent dangerous sparking as the plumb bob enters the uid. If the tape has a grounding wire, attach it to the tank. Allow the plumb bob to penetrate the oil and S&W until it touches the tank bottom or datum plate. Dont let the bob tilt. (Use the known gauge height to estimate when the bob should reach bottom.) Always take the gauge reading from the same location or reference point on the hatch so the bob hits the same point on the tank bottom or datum plate. Using the same reference point for each gauging ensures a comparable gauge reading each time, even if the tank bottom is sloped. When using a water-indicating paste, leave the tape and bob in the liquid long enough for the paste to react with the water (usually 30 to 60 seconds). Slowly reel in the tape and stop when you see where the oil has wet the tape. Read and record the measured liquid level and free-water level. To verify the measurement, wipe off about 2 feet of the oil-wetted tape and lower the tape again, repeating the procedures under step 6 until you get two consecutive readings that agree. When you have completed the opening gauge, clean the gauge tape and bob thoroughly before putting it away.
Gauge hatch
Tape
10
11
12
Innage
Innage bob
4 3 2 1
Zero point
Innage Method
September 2002
3.10
BP Pipelines
8. Determine the temperature of the oil in the tank (see Chapter 2, Gravity and Temperature Measurement in Tanks for details). 9. Record all of your readings. 10. Seal the equalizer and ll lines. 11. Test the top and outlet samples for suspended S&W (see Chapter 5, Testing Crude Oil for Suspended Sediment and Water for details). 12. Accept or reject the oil. 13. If accepted, run the tank. Break the seal on the tank stop valve to the pipeline and remove it. Enter the number of the Off Seal on the ticket. Slowly open the tank stop valve. If the lease requires a pump to inject the oil into the pipeline, set the proper time on the pumps time clock and start the pump.
Closing Gauge
After the oil has been transferred, follow these steps for taking the closing gauge: 1. Before taking the closing gauge, inspect the tank: Close and seal the tank stop valve. Record the seal number. Inspect the pipeline, tank connections, seams, and the ground around the bottom of the tank for leaks or distortions. Check all seals that were put on when the tank was run for tampering. Make sure the seal numbers match the seal numbers written on the ticket. Make sure that the ladders and catwalk are safe. Check the tank and lease numbers stenciled on the tank to be sure you are turning off the correct tank. Ground yourself before climbing on the tank. Be especially cautious on dry days, since the potential for static electricity is greater in low humidity. Open the hatch using all safety precautions, including an approved breathing apparatus if H2S is a potential hazard. Use a mirror to reect light into the tank. Check for any buildup of sediment or encrustations above the level of the remaining oil in the tank. If excessive buildup is present, ask the producer to clean the tank.
2. Determine the temperature of the oil remaining in the tank (see Chapter 2, Gravity and Temperature Measurement in Tanks for details). 3. Gauge the height of the remaining oil and free water (repeat the same steps as under step 6 of Opening Gauge).
September 2002
3.11
BP Pipelines
4. Measure the depth of the settled S&W in the tank with a thief (see Chapter 1, Manual Sampling in Tanks for details). The depth of the tank bottoms should remain about the same before and after the tank is run. If the settled S&W in the tank is deeper at closing than at opening, you did not allow the tank to settle long enough before running it (a minimum of 2 hours). If the settled S&W is lower after you shut the tank off, some of it has most likely moved down the pipeline. Check around the tank for temporary gas rolling lines. If the tank was rolled while the tank was running, excessive S&W may have entered the pipeline. Report this to your supervisor.
If you are satised with the closing gauge: 5. Break all the seals you placed on the tank. Dont break any seals put on by other parties. 6. Make sure that all proper information is on the run ticket before obtaining the witness signature. 7. Sign the ticket printed from the SMART program and include the name of the producers representative (witness).
September 2002
3.12
BP Pipelines
More About It
To calculate the volume of transferred crude, use the opening and closing gauge readings, temperature, gravity, and S&W readings you have taken. You will also need a tank capacity table for the specic tank you measured and tables for volume correction.
September 2002
3.13
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 3.1A Standard Practice for the Manual Gauging of Petroleum and Petroleum Products 2. API Manual of Petroleum Measurement Standards, Chapter 7 Temperature Determination 3. API Manual of Petroleum Measurement Standards, Chapter 11.1, Volume I, Table 5A Generalized Crude Oils Correction of Observed API Gravity to API Gravity at 60F 4. API Manual of Petroleum Measurement Standards, Chapter 11.1, Volume I, Table 6A Generalized Crude Oils Correction of Volume to 60F Against API Gravity at 60F 5. API Manual of Petroleum Measurement Standards, Chapter 12.1 Calculation of Static Petroleum Quantities, Part 1 Upright Cylindrical Tanks and Marine Vessels 6. API Manual of Petroleum Measurement Standards, Chapter 18.1 Measurement Procedures for Crude Oil Gathered from Small Tanks by Truck 7. Pipelines (NA) Business Unit Safety Manual
September 2002
3.14
Chapter 4
BP Pipelines
September 2002
Scope
This chapter describes the procedures for measuring the level of crude oil or liquid petroleum products in upright cylindrical tanks with a capacity of more than 1,000 barrels with xed or oating roofs. Tank gauges may be used for inventory or custody transfer purposes, but the preferred method is to use meters for custody transfer.
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
Introduction
This chapter includes the procedures for taking gauges of a large-volume tank.
September 2002
4.4
BP Pipelines
Procedures
To gauge large tanks with internal or external oating roofs, follow the guidelines below for safety and accuracy.
Gauge tape
Seal (typical) Gauge tube Innage bob Datum plate Floating roof Roof support (typical)
Figure 4.1. Gauging an external oating roof tank from the platform
September 2002
4.5
BP Pipelines
Gauge tape Seal (typical) Gauge tube Innage bob Datum plate Floating roof Roof support (typical)
Figure 4.2. Gauging an external oating roof tank from the oating roof To get an accurate reading, follow these precautions: For both the opening and closing gauges, the roof must be either oating freely or fully supported. For example, do not take the opening gauge while the roof is oating or the closing gauge while the roof is partly or completely resting on the bottom. If the roof is not oating, its legs must be resting directly on the tank bottom, not on the settled S&W. Do not gauge the tank for custody transfer if the roof is in the critical zone (the vertical range in which the level of the stored oil is high enough to lift a tanks oating roof off the tank oor but too low to make the entire roof oat freely in a level position). The leg settings must match the settings listed on the tank table. The roof must be free of ice, snow, water, dirt, and scale. If possible, do not gauge on windy days. If the gauges will be used for custody transfer, the tank must have a recent tank calibration table calculated in accordance with API standards.
September 2002
4.6
BP Pipelines
Seal (typical) Gauge tube Innage bob Datum plate Floating roof Roof support (typical)
Figure 4.3. Gauging an internal oating roof tank To get an accurate reading, follow these precautions: Since an internal oating roof is lighter than an external oating roof, the oil will continue to move longer than in an external oating roof tank. Therefore, leave plenty of settling time. The roof must be either oating freely or fully supported for both gauges. For example, do not take the opening gauge while the roof is oating or the closing gauge while the roof is partly or completely resting on the bottom. If the roof is at rest, it must be resting directly on the tank bottom, not on the settled S&W. Do not gauge the tank for custody transfer if the roof is in the critical zone (the vertical range in which the level of the stored oil is high enough to lift a tanks oating roof off the tank oor but too low to make the entire roof oat freely in a level position).
September 2002
4.7
BP Pipelines
Opening Gauge
If the conditions are safe, gauge the tank as follows: 1. Before taking the opening gauge, inspect the tank (see Figure 4.4): Check the tank and valves for leaks or distortions. Check all bottom and side valves to make sure they are closed. Make sure the ladders and catwalk are safe. Check the tank number stenciled on the tank. Ground yourself before climbing on the tank. Be especially cautious on dry days, since the potential for static electricity is greater in low humidity. Open the hatch using all safety precautions. Wear an approved breathing apparatus if H2S is a potential hazard.
Hatch
If tank has mixer, turn mixer off at least 2 hours before gauging the tank
September 2002
4.8
BP Pipelines
Safety Reminder
Do not gauge the tank if: The liquids surface is foaming or boiling. The liquid level is higher than the stenciled tank height.
2. Suspend the cupcase thermometer, if used, at the midpoint of the tank. 3. Sample the liquid. The number and location of samples depends on the level of oil in the tank (see Chapter 1, Manual Sampling in Tanks for details). 4. Take the opening gauge to the nearest 1/8 inch. Two consecutive readings must be the same before recording the opening gauge on the measurement ticket (see Figure 4.5). Apply a thin coat of water-indicating paste to the bob to read the free-water level. If necessary, apply a thin coat of a gasoline-indicating paste to the tape (product tanks). Hold the gauge line over the hatch and allow the plumb bob to sink through the liquid. Be sure to keep the gauge tape in contact with the edge of the hatch to prevent dangerous sparking as the plumb bob enters the uid. If the tape has a grounding wire, attach it to the tank. Allow the plumb bob to penetrate the liquid and S&W until it touches the tank bottom or datum plate. Dont let the bob tilt. (Use the known gauge height to estimate when the bob should reach bottom.) Read the tape at the reference point. Compare this reading to the reference gauge height (which should be indicated on the top of the tank). A difference of more than 1/2 inch could indicate that the bob has not reached the bottom or datum plate. Always take the gauge reading from the same location or reference point on the hatch so the bob hits the same point on the tank bottom or datum plate. Using the same reference point for each gauging ensures a comparable gauge reading each time, even if the tank bottom is sloped. If the tank does not have an indicated reference gauge point, gauge opposite the gauge hatch hinge.
September 2002
4.9
BP Pipelines
When using a water-indicating paste or gasoline-indicating paste, leave the tape and bob in the liquid long enough for the paste to react with the water (usually 30 to 60 seconds). Slowly reel in the tape and stop when you see where the oil has wet the tape. Read and record the measured liquid level and free-water level. To verify the measurement, wipe off about 2 feet of the wetted tape and lower the tape again, repeating the procedure above. Repeat the procedure until you get two consecutive readings that agree. When you have completed the opening gauge, clean the gauge tape and bob thoroughly before putting it away.
Gauge hatch
Tape
10
11
12
Innage
Innage bob
4 3 2 1
Zero point
Innage Method
Figure 4.5. Innage method of gauging 5. Determine the temperature of the liquid in the tank (see Chapter 2, Gravity and Temperature Measurement in Tanks for details). For custody transfer, the use of a portable electronic thermometer (PET) is preferred. Take the recommended number of temperature readings (see Chapter 2, Gravity and Temperature Measurement in Tanks for details).
September 2002
4.10
BP Pipelines
For greater than 10 feet of product, take 3 temperatures middle of the upper, middle, and lower thirds of the liquid level. For less than 10 feet of product, take 1 temperature in the middle of the liquid level. Read a PET to the nearest 0.1F, or read a cupcase thermometer to the nearest 0.5F. If you have taken multiple temperature readings, average the readings to the nearest 0.1F.
6. Record all of your readings. 7. Test the samples for suspended S&W (see Chapter 5, Testing Crude Oil for Suspended Sediment and Water for details).
Closing Gauge
After the liquid has been transferred, follow these steps for taking the closing gauge: 1. Before taking the closing gauge, inspect the tank: Close the tank isolation valve. Inspect the pipeline, tank connections, seams, and the ground around the bottom of the tank for leaks or distortions. Make sure that the ladders and catwalk are safe. Check the tank number stenciled on the tank to be sure you are turning off the correct tank. Ground yourself before climbing on the tank. Be especially cautious on dry days, since the charge of static electrical sparks is greater in low humidity. Open the hatch using all safety precautions, including an approved breathing apparatus if H2S is a potential hazard.
2. Sample the liquid remaining in the tank (see Chapter 2, Gravity and Temperature Measurement in Tanks for details). 3. Gauge the height of the remaining liquid and free water (repeat the same steps as under step 4 of Opening Gauge). If you are satised with the closing gauge: 4. Make sure that all proper information is on the measurement ticket before obtaining the witness signature. 5. If the gauging was for custody transfer, sign the ticket printed from the SMART program and include the name of the producers representative (witness).
September 2002
4.11
BP Pipelines
Monthly Procedure
1. Grade Select appropriate grade from pull-down list. Verify with Tulsa Control Center (TCC) if necessary.
2. Preferred gauging option Handline gauge at 7:00 a.m. on the rst of the month per this chapter of the Measurement Manual for Crude and Petroleum Products. Tanks inactive at 7:00 a.m. may be gauged either before or after 7:00 a.m. if no activity occurs between 7:00 a.m. and the time the tank is gauged.
3. Sideline gauge option Record sideline gauge reading taken at 7:00 a.m. Tanks inactive at 7:00 a.m. may be gauged either before or after 7:00 a.m. if no activity occurs between 7:00 a.m. and the time the tank is gauged. This option is only acceptable if appropriate maintenance activities on tank gauging hardware have been performed prior to month end with necessary calibrations made. In extenuating circumstances, an adjustment can be made to the sideline gauge reading if recalibration could not be completed on a timely basis (note this in the Remarks eld of the SMART ticket). Recalibration is necessary if the sideline gauge and handline gauge vary by more than 1 inch.
4. Temperature options, in order of preference Take temperature while handline gauging at 7:00 a.m. on the rst of the month in accordance with Chapter 2, Gravity and Temperature Measurement in Tanks of the Measurement Manual for Crude and Petroleum Products.
September 2002
4.12
BP Pipelines
Record and report a temperature taken while performing tank gauging maintenance activities prior to month end, if the tank has been inactive since the hand gauging. For a tank inactive over the entire month, it may be acceptable (if noted in the Remarks eld) to use the prior months gauge and temperature readings.
5. Tank bottom S&W level SMART will automatically ll in this eld with a default value acquired from the corresponding tank screen in the SMART Reference Module. This value is updated every October as part of the annual inventory procedure. If signicant amounts of free water are placed or identied in a tank, a monthly entry in SMART Operations (not SMART Mobile) will be required to override the default value (also note this in the Remarks eld of the SMART ticket). If tank cleaning is completed during the month, an entry will be required to override the default value (also note this in the Remarks eld of the SMART ticket), and the corresponding tank screen in the SMART Reference Module must be updated.
6. Observables crude sampling options in order of preference Sample and analyze crude oil from each tank on the rst of the month in accordance with Chapter 2, Gravity and Temperature Measurement in Tanks and Chapter 5, Testing Crude Oil for Suspended Sediment and Water of the Measurement Manual for Crude and Petroleum Products. Sample and analyze crude oil for each grade on site near the end of the month. In sampling and analyzing crude oil, follow all safety advisories specied in Chapter 1, Chapter 2, and Chapter 4 of the BP Pipelines Measurement Manual for Crude and Petroleum Products.
September 2002
4.13
BP Pipelines
Notification
On or before September 1 of each year, the BP Pipelines Measurement Team will issue a formal advisory to each BP Pipelines core team leader of the mandatory physical tank inventory, which occurs on October 1.
Annual Procedure
1. Grade Select appropriate grade from pull-down list. Verify with TCC, if necessary.
2. Gauging Designated tanks will be handline gauged at 7:00 a.m. on October 1st in accordance with this chapter of the Measurement Manual for Crude and Petroleum Products. Tanks not gauged at 7:00 a.m. can be gauged either before or after 7:00 a.m. However, these tanks must remain inactive between the time of gauging and 7:00 a.m. (before) or 7:00 a.m. and the time of gauging (after). Exceptions to this requirement must be approved by TCC. Appropriate maintenance personnel are to be advised of any variances between the handline gauge, sideline gauge, and/or SCADA that exceed 1 inch. Field specialists must zero out empty tanks that have been taken out of service. Maintenance personnel will issue written conrmation when the cause of these variances has been corrected.
3. Temperature While handline gauging, take and record crude oil temperatures in accordance with Chapter 2, Gravity and Temperature Measurement in Tanks of the Measurement Manual for Crude and Petroleum Products.
4. Tank bottom S&W level The depth of sediment and water at the bottom of each tank must be measured using a thief in accordance with Chapter 1, Manual Sampling in Tanks of the Measurement Manual for Crude and Petroleum Products. The new tank bottom S&W levels must be updated in the appropriate eld on all tank inventory tickets with an October 1 close date. Prior to the next close date, a designated eld specialist(s) must update the S&W level on the corresponding tank screen in the SMART Reference Module. Subsequent months will automatically default to this value unless otherwise handled. If possible, the S&W level obtained for the annual inventory should be used on the
September 2002
4.14
BP Pipelines
monthly inventory report dated October 1 (Septembers closing inventory). 5. Observables crude sampling and analysis Sample and analyze crude oil from each tank in accordance with Chapter 2, Gravity and Temperature Measurement in Tanks and Chapter 5, Testing Crude Oil for Suspended Sediment and Water of the Measurement Manual for Crude and Petroleum Products. Insert sample gravity, sample temperature, and S&W percent in the appropriate elds on the monthly tank inventory tickets. Unless a more frequent sampling program is in place, insert the annually determined values as monthly observables. Exceptions need to be made in a month when very wet crudes are suspected. In sampling and analyzing crude oil, follow all safety advisories specied in Chapter 1, Chapter 2, and Chapter 4 of the Measurement Manual for Crude and Petroleum Products.
6. Remarks To differentiate this monthly tank ticket from other months, it is essential that the responsible eld specialist type Annual Physical Inventory and his/her name in the Remarks eld.
September 2002
4.15
BP Pipelines
More About It
To calculate the volume of transferred liquid, use the opening and closing gauge readings, temperature, gravity, and S&W readings you have taken. You will also need a tank capacity table for the specic tank you measured and tables for volume correction.
September 2002
4.16
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 3.1A Standard Practice for the Manual Gauging of Petroleum and Petroleum Products 2. API Manual of Petroleum Measurement Standards, Chapter 7 Temperature Determination 3. API Manual of Petroleum Measurement Standards, Chapter 11.1, Volume I, Table 5A Generalized Crude Oils Correction of Observed API Gravity to API Gravity at 60F 4. API Manual of Petroleum Measurement Standards, Chapter 11.1, Volume I, Table 6A Generalized Crude Oils Correction of Volume to 60F Against API Gravity at 60F 5. API Manual of Petroleum Measurement Standards, Chapter 12.1 Calculation of Static Petroleum Quantities, Part 1 Upright Cylindrical Tanks and Marine Vessels 6. API Manual of Petroleum Measurement Standards, Chapter 18.1 Measurement Procedures for Crude Oil Gathered from Small Tanks by Truck 7. Pipelines (NA) Business Unit Safety Manual
September 2002
4.17
Chapter 5
Dispose of all samples and security seals properly. Follow all applicable safety rules in the Pipelines (NA) Business Unit Safety Manual.
Scope
This chapter includes the procedures for determining the amount of sediment and water suspended in crude oil.
BP Pipelines
September 2002
Exceptions
Heat samples to greater than 140F if the crude is waxy or parafnic (with your supervisors approval). In locations where the free-water content is greater than 0.3%, you may only extend the allowable free water if the total S&W content remains below 1.0% and you get your supervisors approval before doing so. In Texas, New Mexico, and Louisiana, the total allowable S&W content is 1.0%. You may have free water up to 1.0% if no sediment is present. You may use other solvents (Stoddard solvent, Varsol, etc.) instead of toluene, if it can be shown that they produce the same results.
Start the titrator and allow it to titrate any moisture in the titrator. Homogenize the sample. Inject a measured amount of sample (usually 1.0 ml or less) into the titrator. When the instrument indicates that the titration is completed (usually less than 1 minute), record the water content indicated on the digital display. Analyze a second, duplicate sample and average the results. The two results must agree within 0.01%. If they do not agree, run additional samples until the results do agree.
BP Pipelines
September 2002
BP Pipelines
Introduction
You may test for suspended S&W by one of two methods that use a centrifuge, the eld method or the laboratory method. Another test method is the Karl Fischer test for suspended water content. The amount of sediment and water in crude oil is important because the buyer pays only for the crude and subtracts the S&W content from the total volume.
If using the laboratory centrifuge method: Two certied or veried 8-inch centrifuge tubes Water-saturated toluene as the diluent Demulsier solution Sample heater Bimetal pocket thermometer Centrifuge
If using the Karl Fischer titration method: Nonaerating, high-speed shear mixer Clean glass syringes Karl Fischer reagents anode solution and cathode solution Karl Fischer check solution Karl Fischer titrator
Testing Crude Oil for Suspended Sediment and Water September 2002
5.3
BP Pipelines
Procedures
Before beginning, you must prepare the water-saturated toluene and demulsier solution.
Follow these steps to water-saturate the toluene: 1. Adjust the heating bath to 140 5F. 2. Fill the bottle with 700 to 800 milliliters of toluene. Add 25 ml of water. Screw on the cap and shake vigorously for 30 seconds. 3. Loosen the cap and place the bottle in the bath for 30 minutes. 4. Remove the bottle, tighten the cap, and shake cautiously for 30 seconds. 5. Repeat these steps 3 times.
Testing Crude Oil for Suspended Sediment and Water September 2002
5.4
BP Pipelines
4. Write the date on the dropper bottle. 5. The mixed solution should be replaced on an annual basis. The pure DMO-46X should be replaced every two years. Using the demulsier solution: 1. Shake the dropper bottle before each use. 2. Add one or two drops (with a maximum of four drops) to the centrifuge tube after the solvent is added to the tube. Excessive demulsier may show up after centrifuging as a separate immiscible component in the bottom of the tube and may be misinterpreted as water, overstating the S&W.
Testing Crude Oil for Suspended Sediment and Water September 2002
5.5
BP Pipelines
Water-Saturated Toluene
100 ml Toluene
Free water
Figure 5.1. Free water in the bottom of centrifuge tube when preparing water-saturated toluene
0.200 0.175
Volume Percent
20
40
60
80
100
120
140
160
Temperature F
Testing Crude Oil for Suspended Sediment and Water September 2002
5.6
BP Pipelines
Procedures for Testing Suspended S&W in Lease Tanks by the Field Centrifuge Method
These procedures apply to the upper and lower samples from lease tanks. You will use 100-ml centrifuge tubes (Figure 5.3). Some companies use 200-part centrifuge tubes, but BP Pipelines prefers 100-ml tubes. If you encounter 200-part tubes, contact the Measurement Team for training on how to use and read these tubes. 1. Fill one centrifuge tube with exactly 50 milliliters of the upper sample. Fill another centrifuge tube with 50 milliliters of the lower sample. 2. Fill each tube with solvent to the 100-ml mark. 3. Add a few drops of demulsier solution, if needed. 4. Stopper the tubes tightly and invert them 10 times to mix the sample and solvent. 5. Loosen the stoppers and immerse the tubes in a sample heater. Heat the samples to 140 5F. 6. Stopper the tubes tightly and turn them over 10 times. 7. Put the tubes in the centrifuge and spin for at least 5 minutes. Note: Make sure the tubes are balanced in the centrifuge. 8. Immediately after the centrifuge comes to rest, use a sample thermometer to verify that the sample temperature is 125F or above. If the temperature is 125F or above, continue to the next step. If the temperature is below 125F, repeat step 4, raise the temperature of the sample higher than it was the rst time, and continue with steps 57.
9. Record the volume of water and sediment in each tube. These amounts will not necessarily agree. Add the readings from the two tubes and report this sum as the percentage of S&W. 10. Multiply the reading for the top tube by 2; then multiply the reading for the bottom tube by 2. Record these values. 11. If any of these readings (top plus bottom, top times 2, or bottom times 2) is greater than 0.3% (1.0% in Texas, New Mexico, and Louisiana), reject the tank. Note: The test is valid only if you see a clear separation between the oil and water layers. No emulsion should be present above the oil/water interface. If present, see the Exception, below. Exception If an identiable emulsion layer is present immediately above the oil/water interface: Shake the mixture just enough to disperse the emulsion, and repeat the test.
Testing Crude Oil for Suspended Sediment and Water September 2002
5.7
BP Pipelines
Use a different demulsier or an increased amount of demulsier. (The demulsier should not, however, contribute to the volume of sediment and water.) Use a different solvent.
After a satisfactory procedure for a particular type of oil has been developed, it will normally be suitable for all samples of the same crude oil.
Procedures for Testing Suspended S&W in Samples from Large Tanks and Automatic Samplers by the Field Centrifuge Method
Samples from large tanks and from automatic sampling systems are normally tested for suspended S&W by the eld centrifuge method. These procedures apply to 100-milliliter centrifuge tubes (Figure 5.3). Some companies use 200-part centrifuge tubes, but BP Pipelines prefers 100-ml tubes. If you encounter 200-part tubes, contact the Measurement Team for training on how to use and read these tubes. 1. Fill each of two centrifuge tubes with exactly 50 milliliters of one of the samples. 2. Fill each tube with solvent to the 100-ml mark. 3. Add a few drops of demulsier solution, if needed. 4. Stopper the tubes tightly and invert them 10 times to mix the sample and solvent. 5. Loosen the stoppers and immerse the tubes in a sample heater. Heat the samples to 140 5F. 6. Stopper the tubes tightly and turn them over 10 times. 7. Put the tubes in the centrifuge and spin for at least 5 minutes. Note: Make sure the tubes are balanced in the centrifuge. 8. Immediately after the centrifuge comes to rest, use a sample thermometer to verify that the sample temperature is 125F or above. If the temperature is 125F or above, continue to the next step. If the temperature is below 125F, repeat step 4, raise the temperature of the sample higher than it was the rst time, and continue with steps 57.
9. Read and record the volume of water and sediment at the bottom of each tube. 10. Reheat the tubes to the initial spin temperature and return them to the centrifuge. 11. Do not agitate them this time. 12. Spin the tubes for at least 5 minutes more. Read and record the volume of water and sediment at the bottom of each tube.
Testing Crude Oil for Suspended Sediment and Water September 2002
5.8
BP Pipelines
13. Continue to repeat steps 8 and 9 until two consecutive S&W readings are the same for each tube. If the difference between the readings for two tubes of the same sample is more than one subdivision on the tube, the test is invalid and you must repeat it.
Note: The test is valid only if you see a clear separation between the oil and water layers. No emulsion should be present above the oil/water interface. If present, see the Exception, below. 14. Record the nal volume of water and sediment in each tube. Add the readings from the two tubes and report this sum as the percentage of S&W. Exception If an identiable emulsion layer is present immediately above the oil/water interface: Shake the mixture between spins in the centrifuge just enough to disperse the emulsion. Use a different demulsier or an increased amount of demulsier. (The demulsier should not, however, contribute to the volume of sediment and water.) Use a different solvent.
After a satisfactory procedure for a particular type of oil has been developed, it will normally be suitable for all samples of the same crude oil.
Testing Crude Oil for Suspended Sediment and Water September 2002
5.9
BP Pipelines
Shake sample bottle, then pour Add liquid sample to 50 ml Add toluene to 100 ml
Heat to 140F
100-ml Tube
1.0 ml 0.50 ml
0.60 ml
1 1/2
0.80 ml X 2 = 1.6 ml
Read S&W
0.25 ml
Figure 5.3. Sequence for determining S&W by the eld centrifuge method using 100-ml (6-inch) centrifuge tubes
Testing Crude Oil for Suspended Sediment and Water September 2002
5.10
BP Pipelines
Testing Crude Oil for Suspended Sediment and Water September 2002
5.11
BP Pipelines
10. Repeat steps 8 and 9 until the combined volume of sediment and water is the same for two consecutive readings. If the difference between the 2 readings for the same sample is more than one subdivision on the tube or more than 0.025 ml for readings of 0.10 ml or less, the test is invalid and you must repeat it.
11. Record the nal volume of water and sediment in each tube. Add the readings and report this sum as the percentage of S&W.
Procedures for Testing for Water in Crude Oil by Karl Fischer Titration
Karl Fischer titration determines water content only. In this method, you will use a test sample of a certain volume taken from the sample you have already collected. Table 5.1 shows how much sample to use, depending on the water content you expect to nd in the sample. If you have no idea what the water content is, start with the smallest amount of sample as a test. You may then repeat the titration with a larger amount if needed.
Table 5.1.
Testing Crude Oil for Suspended Sediment and Water September 2002
5.12
BP Pipelines
2. Add the Karl Fischer cathode solution to the cathode (inner) compartment to a level 2 to 3 millimeters (1/16 to 1/8 inch) below the level of the solution in the anode compartment, normally 5 ml of solution. Note: The cathode solution normally used by BP Pipelines is supplied in premeasured 5-ml portions. 3. Seal all joints and connections to the titrator with the appropriate grease to prevent moisture in the air from entering. 4. Turn on the titrator and start the magnetic stirrer. Allow the moisture that may be in the titrator to be titrated until the end point is reached. Do not continue with testing until the background current or background titration rate is constant.
5. Homogenize (mix) the sample using a nonaerating mixer. During mixing, the temperature of the oil should not increase more than 20F, otherwise a loss of water may occur. Note: You may test a sample taken from the circulating system of an automatic sampler receiver without remixing if it you test it within 5 minutes of drawing it from the system. 6. Immediately after mixing, use a clean, dry syringe to withdraw at least 3 samples from the tank sample and discard them. 7. Then withdraw the test sample and expel the excess (see Table 5.1 for amount to leave in the syringe). Note: It is very important to leave exactly the right amount in the syringe and to have no air bubbles in the syringe. Minimize the air bubbles by keeping the tip of the needle below the surface of the sample and pulling the sample into the syringe slowly. Before expelling the excess sample, point the syringe up to allow any air to collect in the top and slowly expel it (covering the end of the needle with a cloth or tissue).
8. Clean the needle with a tissue. Record the volume of the sample. Verify that the titrator is set for the same volume of sample. 9. Insert the needle through the inlet port, start the titration, and inject all of the sample into the titration cell. Place the tip of the needle just below the surface of the liquid.
10. Take the needle out and wipe it clean. 11. When the titrator reaches the end point, record the percentage of water from the digital readout.
Testing Crude Oil for Suspended Sediment and Water September 2002
5.13
BP Pipelines
12. Take a second sample and inject it into the titrator after pushing the rest button. The two results should agree within 0.01%. If they do not agree, continue testing additional samples until they do agree.
13. Average the two results, rounding to the nearest 0.01%. 14. Every day (or before each test, whichever is longer), verify that the titrator is functioning properly by accurately injecting a very small amount (10 l) of pure water or an approved check solution.
Testing Crude Oil for Suspended Sediment and Water September 2002
5.14
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 10.3 Determination of Water and Sediment in Crude Oil by the Centrifuge Method (Laboratory Procedure) 2. API Manual of Petroleum Measurement Standards, Chapter 10.4 Determination of Sediment and Water in Crude Oil by the Centrifuge Method (Field Procedure) 3. API Manual of Petroleum Measurement Standards, Chapter 10.9 Standard Test Method for Water in Crude Oils by Coulometric Karl Fischer Titration 4. Pipelines (NA) Business Unit Safety Manual
Testing Crude Oil for Suspended Sediment and Water September 2002
5.15
Chapter 6
Scope
This chapter gives an overview of the procedures for quality testing of crude and products. Most of these tests are done in a nonpipeline laboratory, but some are done on site.
Summary of Tests
Crude Oil
Sulfur content Mercaptan content Organic chloride content Reid vapor pressure Metals content Types and amount of light ends Hydrogen sulde content Viscosity and pour point Boiling point range Neutralization number Nitrogen content
BP Pipelines
September 2002
Introduction
Both crude oil and petroleum products are tested for composition and characteristics other than gravity, S&W, and temperature that affect their quality and therefore their value. Since most of these tests are done in a (nonpipeline) laboratory, this chapter contains a general discussion of these tests rather than giving specic procedures. If you do the tests on site, follow local procedures.
BP Pipelines
September 2002
BP Pipelines
Sulfur
The test method for sulfur (ASTM D-4294) determines total sulfur in a crude oil from all its various forms. Sulfur is primarily present as organic sulfur (attached to hydrocarbon molecules) but can also be present as hydrogen sulde, mercaptans (see Mercaptans below), and inorganic salts (sulfates).
Test Summary
The common method used today to test samples for sulfur is by x-ray. A sample of crude oil is placed into a cup with a clear plastic Mylar window at the bottom. The cup is placed into the x-ray instrument and the sample is irradiated with x-rays. The signal received by the detector, coming from the sample, indicates the level of sulfur. The test result is total sulfur in weight percent.
Table 6.1.
Crude Designation Sweet crude A crude Light sour Medium sour Heavy
September 2002
6.3
BP Pipelines
recover sulfur from the rening process to comply with environmental regulations on plant emissions. Many governmental regulations limit the sulfur content of rened products, thus limiting the types of crude oils that some reneries can process. In the case of contamination, increased sulfur levels can jeopardize a renerys ability to comply with these regulations. In general, high-sulfur crudes tend to be heavier (lower API gravity).
Mercaptans
Mercaptans are naturally occurring sulfur compounds. Ethyl mercaptan is used to add odor to natural gas. Higher-molecular-weight mercaptans are a problem in jet fuel and gasolines due to specication levels of not more than 30 ppm in jet fuel product and not more than 40 ppm in gasolines. Levels of mercaptans in crude oil vary and depend on the eld from which the crude is produced.
Test Summary
A sample is diluted with isopropanol and titrated with silver nitrate solution.
September 2002
6.4
BP Pipelines
Organic Chlorides
Organic chlorides are primarily chlorinated hydrocarbons that remain in the hydrocarbon phase and are not removed in the renery desalting process. Also, certain renery processes inject chloride chemicals for operational needs. They do not occur naturally in crude oil and indicate contamination of the crude stream, typically from solvents such as dry cleaning uids, carburetor cleaners, and other halogenated hydrocarbons as well as transformer PCBs, degreasing agents, and used motor oils.
Test Summary
Testing for organic chlorides is a multistep process. First the crude oil is distilled to fractionate out the naphtha cut (initial boiling point to 400F). Next the naphtha fraction is washed with caustic and then with water to remove inorganic chlorides, which can interfere in the detection step. The nal step is to quantify the chloride level by either titration or combustion/coulometric analysis, both of which detect the level of organic chloride in ppm. The organic chloride level of the whole crude is then calculated from the chloride level in the naphtha fraction and the percentage of the naphtha fraction in the crude oil.
Typical Levels
Organic chlorides do not occur naturally in crude oils and indicate contamination. BP Pipelines has a specication for organic chloride of not more than 1 ppm by weight in the whole crude and not more than 5 ppm by weight in the naphtha fraction.
September 2002
6.5
BP Pipelines
Test Method
Testing for RVP is done by placing the sample in a xed volume cell at 100F. When stabilized, an automated analyzer uses a transducer to obtain the pressure. If the test is done manually, the sample is shaken several times and the pressure gauge is read.
Metals
Metals content generally refers to heavy metals in crude oil such as arsenic (As), iron (Fe), vanadium (V), and nickel (Ni). It may also refer to such metals as sodium, magnesium, and calcium. Table 6.2 shows the maximum amount of some of these metals allowed in crudes.
Test Summary
Metals in crude oil can be analyzed directly by x-ray spectroscopy, similar to the sulfur analysis, or by an alternative lab method. A more detailed analysis for metals would include distilling the crude to isolate either the 650F+ reduced crude fraction or 1050F+ resid fraction before doing the lab analysis.
September 2002
6.6
BP Pipelines
being run on the uid catalytic cracker unit, in platformers and magnaformers used in gasoline production, in hydrocrackers used in jet and gasoline production, and in hydrotreaters producing jet, gasoline, and diesel these all use catalysts in the processing stage. Some vanadium compounds can damage turbine blades and refractory furnaces. BP Pipelines concern about metals in crude is to prevent contamination of low-metal crudes (West Texas intermediate/West Texas sour) with heavy high-metal crudes such as Mayan. This is one reason for small buffer batches of West Texas sour surrounding Mayan batches on the Cushing Chicago Pipeline System.
Table 6.2.
Typical Levels
Metals content in crude oils varies widely. Mayan heavy has 700 ppm vanadium in the resid fraction, which equates to 260 ppm on a whole crude basis (37% resid). Arabian heavy crude oil has much less vanadium, 180 ppm in the resid fraction, 50 ppm whole crude basis (28% resid). To make a valid comparison of metals content of two or more crude oils, make sure you know whether the amounts are for the whole crude or for the resid fraction.
Light Ends
Light ends are the gaseous light hydrocarbons in crude oil: ethane (C2); propane (C3); isobutane and normal butane (C4); and isopentane and normal pentane (C5).
Test Summary
Light ends are generally analyzed in the laboratory by gas chromatography. This method essentially separates the C2 through C5 hydrocarbons from the crude oil and individually quanties each component.
September 2002
6.7
BP Pipelines
Significance
Light ends in crude oils are an important component to control. Since they are volatile, they can lead to a pressure over the pipeline specication of 10 psi (see Reid Vapor Pressure). A large amount of light ends in the pipeline can lead to pumping problems, metering errors, and noncompliance with above-ground storage tank regulations. In the past, some producers added low-cost butane to the crude oil to increase volumes and therefore increase prots. Extensive sampling and testing programs have essentially stopped this type of activity. At times in the past, some producers have created what is called a dumbbell crude oil. This refers to an articial crude oil mixture created from the blending of light ends/condensate with 650+/1050+ heavy-end bottoms to produce a material that meets gravity and sulfur specications but is not a valid full-boiling-range virgin crude oil.
Refinery Impact
The renerys biggest concern with light ends is with unanticipated high levels of them, which lead to several problems: higher than predicted light-end yields (vs. computer models, which help plan and schedule renery operations); exceeding gas compressor capacity limits; and potential distillation tower upsets. A crude with a high light-end content loads up renery units such as the isomerization unit and the light ends unit that processes C3 to C5 hydrocarbons, which necessitates placing the excess in storage/inventory. This adds to processing costs.
Hydrogen Sulfide
Hydrogen sulde (H2S) is a poisonous gas that occurs naturally in crude oil and can break out from it. Higher H2S levels are generally found in heavy sour crudes. You must use respiratory protection when levels in the air reach 10 ppm. At 300 ppm H2S is an Immediate Danger to Life and Health (IDLH).
Test Summary
In the laboratory, H2S can be determined by gas chromatography or the wet chemical method. The wet chemical method involves driving the H2S from the crude oil with nitrogen gas and heat, chemically trapping the H2S, and then nishing with a titration to quantify the level. Hydrogen sulde levels in the air can be determined using Draeger tubes or other H2S detectors.
September 2002
6.8
BP Pipelines
is vital to prevent accidents and can aid a renery in its ability to process the sulfur as a by-product from the rening process. Determination of the H2S content is now a standard test when crude oils are analyzed in the laboratory for economic assays.
Test Summary
Viscosity is tested by measuring how long it takes a crude oil to move through a calibrated capillary tube at a specied temperature. This is done in a constant temperature bath. Pour points of petroleum products are tested in a low-temperature refrigerated bath. The sample is tested until solid. The pour point is the temperature at which ow was last observed generally 5F above the solidication point.
September 2002
6.9
BP Pipelines
Distillation
A distillation analysis obtains data on the boiling point range of a crude oil and nished petroleum products. It determines the percentages or yields of crude fractions over their typical boiling point ranges.
Table 6.3.
Test Summary
Distillation of a crude oil can be carried out in two ways. An economic assay distillation, sometimes called a crude assay, which mimics a renery distillation operation, starts with distillation at atmospheric pressure (pot still) of the crude oil to fractionate out the naphtha. Then the bottoms are transferred to a vacuum still, where lower pressure allows separation of the distillate, gas oil, and resid portions. In both processes, temperatures are monitored as each fraction distills. The laboratory quanties the percentage of each fraction, combines this with the temperature data, and creates a boiling point curve for the crude oil being analyzed. A second method is high-temperature simulated distillation, called Sim-Dis (a special type of gas chromatography, or GC). A small quantity of crude oil is diluted into a solvent and injected into a GC analyzer. The crude oil components are separated by boiling point on a long capillary tube (column), and data is obtained as the ever higher boiling portions of the crude oil emerge from the analyzer. A computer program then creates the boiling point curve. Either method may be used. Distillation allows further tests to be conducted on the fractions, whereas Sim-Dis gives only a boiling point curve.
September 2002
6.10
BP Pipelines
dumbbell crudes both cause incorrect product yields and result in shortages or overages for market demands, which leads to prot losses for the renery. Similarly, distillation determines resid content (fraction boiling at 1050F or more, in vol%). High resid levels may indicate fraudulent blending.
Neutralization Number
Neutralization number is also referred to as total acid number. It is an indication of the amount of acidic components present in the crude oil.
Test Summary
The crude oil is diluted into a solvent and titrated with a standard base (potassium hydroxide, or KOH). The base neutralizes the acidic components. The test result, or neutralization number, is given in the unit mg KOH/gram sample.
Nitrogen
Nitrogen in crude refers to the amount of organic nitrogen that is bound (attached) to hydrocarbon molecules.
Test Summary
The test for organic nitrogen is performed by rst diluting the crude oil into a solvent. This mixture is then injected into a combustion furnace, which burns the sample under controlled conditions. The products of this process are swept by a owing gas stream into a reaction chamber where the level of nitrogen is determined by the response of the detector module.
September 2002
6.11
BP Pipelines
September 2002
6.12
BP Pipelines
Test Method
Testing for RVP may be done by placing the sample in a xed volume cell at 100F. When stabilized, an automated analyzer uses a transducer to obtain the pressure. If the test is done manually, the sample is shaken several times and the pressure gauge is read. Most gasoline samples are now tested for RVP using a Grabner RVP analyzer. A small portion of sample is injected into the analyzer and the resulting RVP is displayed on a digital readout.
September 2002
6.13
BP Pipelines
Test Summary
Viscosity is tested by measuring how long it takes a liquid to move through a calibrated capillary tube at a specied temperature. This is done in a constant temperature bath. Pour points of petroleum products are tested in a low-temperature refrigerated bath. The sample is tested until solid. The pour point is the temperature at which ow was last observed generally 5F above the solidication point.
Flash Point
The ash point of a petroleum product is the lowest temperature at which the vapors above it ignite in air.
Test Summary
A liquid sample is poured into a cup, a thermometer inserted, and the cup covered. The sample is then heated slowly until a ash, or small explosion, occurs that consumes the vapors inside the closed cup. The temperature when the ash occurs is the ash point.
September 2002
6.14
BP Pipelines
Table 6.4.
Jet A Diesel Supreme Diesel (high- and low-sulfur/fuel oil) Kerosene (K-1)
Test Summary
One liter of sample in a transparent, colorless bottle is inspected by placing the bottle in front of a white card with several black lines marked on it (Figure 6.1). The visibility of the lines through the sample describes its haze. ASTM (American Society for Testing and Materials) provides a numbered standard (1 to 6) for describing haze so that no one has to rely on anothers description of pretty cloudy or almost clear. Another test for water and contaminants is millipore sampling. In this test, a liquid sample of a distillate fuel is ltered through a millipore membrane and then assessed visually. The water separator index modied (WSIM) test measures the ability of water to separate from fuel. A fuel sample in a syringe is emulsied with water using a high-speed mixer and then expelled from the syringe through a berglass coalescer. The efuent is analyzed for uncoalesced water by light transmission measurement.
September 2002
6.15
BP Pipelines
Oxygenate Content
An oxygenated fuel is a gasoline that is blended with compounds containing oxygen (commonly MTBE or ethanol). Oxygenate content is determined by a number of procedures and instruments, including portable analyzers, on-line analyzers, and detailed laboratory testing.
September 2002
6.16
BP Pipelines
More About It
See the BP Crude Quality Manual for more detailed information about testing crude oil.
September 2002
6.17
BP Pipelines
Reference Documents
The following list includes some of the ASTM standards used to analyze crude oil and petroleum products. 1. ASTM D-56 Tag closed cup for jet fuel 2. ASTM D-93 Pensky Marten for diesel fuels 3. ASTM D-97 Pour point of petroleum products 4. ASTM D-323 Vapor pressure of petroleum products (manual method) 5. ASTM D-396 Heating oil specications 6. ASTM D-445 Kinematic viscosity 7. ASTM D-664 Total acid number (potentiometric method) 8. ASTM D-974 Neutralization number (colorimetric method) 9. ASTM D-975 Diesel fuel oil specications 10. ASTM D-1655 Jet fuel specications 11. ASTM D-1796 Water content
September 2002
6.18
BP Pipelines
12. ASTM D-2276 Particulate contamination of petroleum products 13. ASTM D-2276 Millipore color for jet fuel 14. ASTM D-2622 Sulfur in petroleum products by X-ray spectroscopy 15. ASTM D-2624 Electrical conductivity of petroleum products 16. ASTM D-2709 Water content 17. ASTM D-2887 Simulated distillation for petroleum products 18. ASTM D-2892 Distillation of crude petroleum (to 400F) 19. ASTM D-3227 Mercaptan sulfur in gasoline, kerosene, aviation turbine, and distillate fuels 20. ASTM D-3237 Lead content 21. ASTM D-3240 Aqua Glo for jet fuel 22. ASTM D-3710 Boiling point range 23. ASTM D-4176 Haze of petroleum products
September 2002
6.19
BP Pipelines
24. ASTM D-4629 Trace nitrogen in liquid hydrocarbons by syringe injection 25. ASTM D-4814 Gasoline specications 26. ASTM D-4929 Determination of organic chloride content in crude oil 27. ASTM D-5191 Vapor pressure of petroleum products (Grabner method) 28. ASTM D-5236 Distillation of heavy hydrocarbon mixtures (vacuum potstill method) 29. ASTM D-5307 Determination of boiling range distribution of crude petroleum by gas chromatography 30. ASTM D-5762 Nitrogen in petroleum and petroleum products by the Boat-Inlet method 31. ASTM D-5949 Pour point Other references: 32. ATA 103 manual from airlines for pipeline requirements 33. BP Crude Quality Manual 34. BP General Operations Manual 35. Pipelines (NA) Business Unit Safety Manual
September 2002
6.20
Chapter 7
Tank Strapping
Quick Reference
Safety
Get the required permits and clearances before entering any tank. Report any tank conditions that appear unsafe. Post warning signs in strategic locationsnear all ladders and stairways and around the lower tank shellbefore you leave the area. Keep the outer tank surfaces clean to reduce the risk of slipping and re. Always keep one hand on the handrail when climbing tank stairsfor maintaining balance and preventing the buildup of static electricity. Before taking any measurements, ground your bare hands and tools by touching the handrail. Avoid causing sparks, especially when opening gauge hatches. Sparks can be caused by metal tools, cleats, rings and watches, and other objects. Watch out for ice and oil on tank stairs and platforms. Do not walk on any ladder, stairway, or platform if you suspect that it is weakened by corrosion. Stand upwind and turn your face away when opening the tank hatch. Monitor the atmosphere for H2S, combustible vapors, and adequate oxygen content while strapping. Never strap a tank during an electrical storm or in strong winds. Keep the area around the tank clear of obstructions and tripping hazards. Some petroleum products are extremely ammable and/or poisonous. Follow the applicable safety procedures in the Pipelines (NA) Business Unit Safety Manual when strapping tanks that contain these dangerous liquids. Follow all applicable safety rules in the Pipelines (NA) Business Unit Safety Manual.
Scope
This chapter describes the procedures for calibrating small (less than 1,000 barrels) upright, cylindrical tanks by the manual tank strapping method. Larger tanks and tanks with oating roofs will be strapped by third-party contractors following procedures similar to these.
BP Pipelines
September 2002
Good Practices
Take all strapping measurements on the same day without interruption. Take strapping measurements while the liquid level in the tank is static. Allow the tank contents to settle at least 24 hours before strapping.
BP Pipelines
September 2002
BP Pipelines
Introduction
Strapping is a method of measuring the exact dimensions of a tank to develop a capacity table for that tank and increments for the measurement ticketing software. The capacity table allows the eld specialist to calculate the volume of liquid in the tank. These measurements include the circumference of each tank ring and shell plate thickness; the height of the shell, gauging height, and effective inside tank height; and the dimensions of any obstructions inside the tank (deadwood) that reduce its nominal capacity.
Tank Strapping
September 2002
7.3
BP Pipelines
Procedures
When you arrive at a tank location, follow these procedures before beginning strapping: 1. Check for hazardous atmosphere using a hydrogen sulde monitor. Also use a monitor to analyze the atmosphere for combustible vapors and adequate oxygen content. During the strapping, keep both monitors activated so that an alarm will sound if atmospheric conditions change.
2. Make sure no ice or snow has accumulated on the tank. If it has, postpone the strapping until the ice or snow has melted. 3. Clean any incrustations (for example, ice, mud, etc.) that will affect the circumference measurements. 4. Check the ll level. In general, welded tanks with a capacity of 500 barrels or less can be strapped in any condition of ll as long they have been lled at least once at their present location. Welded tanks with capacities over 500 barrels and all bolted and riveted tanks should be at least two-thirds full when strapped.
Note: The amount of liquid in the tank affects strapping, as do the size and type of tank. Table 7.1 under More About It shows what the ll condition must be for each type of tank. 5. Be sure that the tank is level by measuring the tanks tilt with a pitch indicator and level or a protractor. Dont strap the tank if the amount of tilt is greater than 1 part in 70, which equals 0.82 (about 1) from vertical.
6. Look for the strapping identication number stenciled on the side of the tank. If there is no number, locate the builders nameplate on the side of the tank. Record the information on the strapping report. 7. Determine the ambient temperature as well as the gravity and temperature of the tanks contents (see Chapter 2, Gravity and Temperature Measurement in Tanks).
Plate Thickness
For small lease tanks, get the plate thickness from the nameplate on the tank. For larger tanks (greater than 500 barrels), measure the thickness of the shell plate with an ultrasonic device or a depth gauge. Make at least two measurements to the nearest 1/64 inch on each ring, and record the average on the strapping report (see Figure 7.7 under More About It). Do not measure plate thickness where the edges have been distorted by caulking.
Tank Strapping
September 2002
7.4
BP Pipelines
Circumference Measurements
Measure the circumference of each ring of the tank. Do not take circumference measurements over insulation. 1. Calibrate the working (strapping) tape against a NIST-certied standard tape. Using the standard tape, measure the bottom ring of the tank (see steps 38 below) and score a vertical reference mark at the zero reading, at 100-foot intervals, and where the tape laps over the starting point. Use a tape clamp and a tension gauge to apply the tension specied in the NIST report. Using the working tape, measure the tank along the same path and apply the same amount of tension that you applied to the standard tape. Note any differences in the measurements of the working tape and the standard tape. If the working tape circumference is the same as the standard tape circumference, the tape is in calibration and no correction is needed. If there is a difference, add or subtract this amount to the measurements taken with the working tape. If the correction is more than 0.01 foot (approximately 3/32 inch) per 100 feet of tape, the working tape is not within the recommended tolerances. Use another tape.
2. Begin strapping with the bottom ring (ring 1). 3. Mark the tank shell to indicate the tape path with the aid of the strapping pole. For welded tanks, scribe a path for the tape at a position one-fth (or 20%) of the ring's height, measured from the top of the weld down (see Figure 7.1). For bolted tanks, the position for scribing is 25% and 75% above the bottom of each ring (see Figure 7.2).
20%
Figure 7.1. Recommended tape paths for measuring tank circumference on a welded tank
Tank Strapping
September 2002
7.5
BP Pipelines
Ring
75% 25%
Figure 7.2. Recommended tape paths for measuring tank circumference on a bolted tank 4. If you cannot follow the recommended path because of obstructions on the tank, mark a substitute path that is as near to the recommended path as possible. If an obstruction is unavoidable, use a stepover caliper to span the obstruction. Place the caliper in the horizontal position, determined with a level, against the tank shell near the center of the ring. Scribe marks on the tank shell with the points of the caliper. Place the working tape under the correct tension around the tank, and measure the distance between the scribed marks to the nearest 0.005 foot.
5. Make sure to get the tape straight along the scribed path. 6. When you have pulled the tape to the correct tension, slide it back and forth to spread the tape tension evenly around the shell. If friction causes the tape to cling to any point along the path, the measurement will not be accurate. Note: The part of the tape that laps over the starting point (where you read the measurement) must be at least 2 feet from an upright joint. 7. Holding the tape at against the side of the tank, read it to the nearest 0.005 foot, slide it back and forth for a few inches, then read it again. If these 2 measurements do not agree within 0.005 foot, repeat until you obtain the same reading for two consecutive measurements. Correct the measurement as required by the tape calibration, if needed. Record the average of the 2 readings for this ring on the strapping report (see Figure 7.7 under More About It at the end of this chapter).
Tank Strapping
September 2002
7.6
BP Pipelines
8. If butt straps or lap joints cause uniform voids between the tape and the tank shell at each joint, measure and record the width and thickness of the butt straps or lapped plates and the number of butt straps or lap joints in each ring. If possible, measure the shell plate thickness of each ring at 2 locations using an ultrasonic measurement device. If no direct measurement is possible, use the shell thickness on the tanks nameplate or the construction prints for the tank.
Reference point
Gauging tape
Datum plate
Tank Strapping
September 2002
7.7
BP Pipelines
Roof
Shell
Bottom
Figure 7.4. Measuring the shell height of a tank with a protruding bottom
Roof
Shell
Bottom
Figure 7.5. Measuring the shell height of a tank with a recessed bottom 1. If the tank bottom does not protrude beyond the shell, dig out the dirt at the base of the tank so that you can get a straight-edge or level underneath the tank bottom. 2. Measure near the reference point on the gauge hatch and at 3 other points equidistant from the rst around the shell with an innage gauge tape, using a level as a point of measurement.
Tank Strapping
September 2002
7.8
BP Pipelines
3. Check the amount of tilt in the shell height at each measuring point with a theodolite, optical plummet, or plumb bob. Note: The tilt must be less than 1 part in 70 (less than 1.3). If it is more, stop the strapping and notify your supervisor and the Measurement Team. The tank will need to be leveled before it can be restrapped and used. 4. Measure the height of each ring to the nearest 1/16 inch. The combined ring heights should add up to the overall shell height. If they do not match, remeasure the heights. 5. Record your measurements on the strapping report (see Figure 7.7 under More About It at the end of this chapter).
Tank Strapping
September 2002
7.9
BP Pipelines
Anchor line
Gauging height
Figure 7.6. Determining effective inside tank height 1. Anchor a line at the reference point on the hatch. 2. Extend the line at a right angle from the tank wall (parallel to the ground). 3. Level the anchored line, and measure the distance between the line and the top angle. 4. Subtract this measurement from the gauging height to get the effective inside tank height. 5. Record your measurements on the strapping report (see Figure 7.7 under More About It at the end of this chapter).
Overflow Location
Some tanks have an overow line near the top of the tank shell. Note the size and location of this line on the strapping report and whether it can be closed. If the line cannot be closed, measure the effective inside tank height as the distance from the gauge striking point or datum plate to the bottom of the overow line.
Connection Height
Connection height is the point near the base of the tank shell where liquid exits the tank and enters the pipeline. 1. Use the gauge tape to measure the distance from the tank oor to the bottom of the nipple where it protrudes from the shell. 2. Add 1/4 inch to this measurement to allow for the approximate thickness of the nipple, and record the measurement on the strapping report.
Tank Strapping
September 2002
7.10
BP Pipelines
Deadwood Measurements
Deadwood includes internal piping, roof supports, ladders, and other xtures that occupy volume inside the tank. If the tank contains liquid, it may not be possible to measure deadwood. For most tanks, you can nd the necessary dimensions on construction drawings if actual measurement is impossible. If the tank is empty, follow these procedures for measuring deadwood. 1. Inspect the tank for cleanliness. The inside vertical surface and roof supporting members should be clean and free of any foreign matter, such as liquid residues, rust, dirt, emulsion, and parafn. Clean the tank if needed.
2. Measure the highest and lowest levels where deadwood affects the tanks capacity. Measure from the tank bottom next to the shell.
3. Clearly identify work sheets containing the details of deadwood locations, shapes, and dimensions, and make these worksheets part of the strapping record.
Manhole Dimensions
Manhole measurements involve specialized tools and procedures depending on the size and shape of the manhole. These are presented in on-the-job training.
Tank Strapping
September 2002
7.11
BP Pipelines
More About It
Here you will nd the following: Table 7.1. Fill Condition Required for Different Tanks while Strapping Figure 7.7. Sample page of strapping report
The amount of liquid in the tank affects strapping, as do the size and type of tank. Table 7.1 shows what the ll condition must be for each type of tank.
Table 7.1.
Fill Condition <500 Filled at present location before strapping Any ll during strapping Must be 2/3 full during strapping Small amount of inow or outow allowed during strapping
Note: Tank capacity is in barrels.
>500 X
Tank Strapping
September 2002
7.12
BP Pipelines
Tank Strapping
September 2002
7.13
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 2.2A Measurement and Calibration of Upright Cylindrical Tanks by the Manual Tank Strapping Method 2. BP Self-Study Guide, Field Specialist III, Module 1 Tanks and Tank Strapping 3. Pipelines (NA) Business Unit Safety Manual
Tank Strapping
September 2002
7.14
Chapter 8
Scope
These procedures apply to lease tanks and LACT/ACT unit systems (including the ELM).
Note: At some locations, the seals may be installed by the other party. Also some ACT meters, depending on location, manning, and products, may not be sealed. Flow computers shall be password protected.
BP Pipelines
September 2002
When you nd a seal on a LACT/ACT system broken: Replace the boxcar seal with a wire seal. Do not transfer oil to/from that facility until your supervisor instructs you to do so. Conduct an immediate investigation and report your ndings to your supervisor.
BP Pipelines
September 2002
BP Pipelines
Introduction
Seals protect the company and the operator by assuring that no one has tampered with the oil run. They are placed on any part of a lease tank or LACT unit that could be tampered with. Seals are also part of assuring safety when working on tanksfor example, when welding a line. Electronic liquid measurement (ELM) ow computers require security in the form of software password protection, as well as physical locking of the area where they are located.
September 2002
8.3
BP Pipelines
Procedures
The purpose of using seals and security procedures is to make sure that the equipment containing and monitoring crude or product has not been not tampered with. The following will give you an overview of where you should check for seals and how to use them.
Using Seals
The two types of seals are boxcar seals and wire seals (see Figure 8.1 under More About It at the end of this chapter). Boxcar seals are numbered; record the numbers as a reference on the run ticket. Use a wire seal if you encounter a defective boxcar seal. Attach the defective boxcar seal to the wire seal, and record both seal numbers on the run ticket. If you nd that a boxcar seal has been used where a wire seal should have been used, replace the boxcar seal with a wire seal or use the boxcar seal with a wire seal attached to it. Replace any seals that were broken during LACT/ACT checkout, as well as any missing seals and any seals showing signs of deterioration.
2. When oil is being delivered by tank truck: Seal pipeline connections on lease tanks used for receiving oil delivered by tank truck. Close and seal all draw-off connections when the tank is on line. Request the operator to make sure that the tanks gauge and thieving hatches can be sealed while still allowing venting. Seal the tanks gauge and thieving hatches.
3. When the tank is on line to make sure that unwanted hydrocarbons or contaminants cannot be introduced or removed from the pipeline system undetected do the following: Seal all closed valves on draw-off connections, lling lines, overow lines, and equalizer lines. Seal steam lines.
September 2002
8.4
BP Pipelines
BP Pipelines does not require you to seal off re protection lines. Break all BP Pipelines seals when a tank is returned to the operator.
2. S&W monitor and probe Seal each connection on the probe and detector. Also, seal the electronic module at the hinges, cover latch, and at any other connections.
3. Sampler system Place a seal at any point where tampering may affect the systems accuracy. All container openings, valves, pipe unions, pump drains, and other locations on the system should be sealed.
5. Meter and meter accessory stack Seal the following on the meter and meter stack: the main meter body bolting, temperature and pressure averagers, right-angle drive dust cover, and meter totalizer.
6. Control panel and power panels All electronic devices are wired through the control box and can be disabled by entry into the panels. Seal all electronic devices at hinges, latches, and other points of access.
7. Tank head switches and level controls Seal the high- and low-level switches as well as the low-level limit switch just above the sales outlet.
8. Diverter valves Seal the diverter valve in such a way as to prevent tampering and the sale of bad oil.
September 2002
8.5
BP Pipelines
9. Piping and control conduits Seal all pipe couplings and screwed plugs located downstream from the meter to guard against the removal of metered liquid from the pipeline.
Signs of Tampering
Look for the following signs of tampering each time you check the LACT/ACT or gauge a tank: Forced entry Broken BP Pipelines security seals or seals that have been tampered with Electrical connections that have been tampered with (e.g., jumper wires) Small holes drilled in the weatherproof box housing the electronics (possible signs of an attempt to insert wires for manipulating controls) Monitor alarm and trip set points that dont agree with previous settings
In addition to checking the unit for signs of tampering, be sure to pick up and dispose of all security seals and wires you have removed.
The supervisor may authorize you to run oil from other tanks on the lease, but do not run any oil from the tank with the broken seal until you have permission from the District Manager. When you nd a broken seal on a pipeline connection: Replace the boxcar seal with a wire seal. Do not run oil from that tank or any other tank on the lease until your supervisor instructs you to. Conduct an immediate investigation and report your ndings to your supervisor. Specics of the investigation will depend on the circumstances, location, etc.
September 2002
8.6
BP Pipelines
Software Security
Passwords provide security for software. The rst level of security allows measurement personnel to enter routine data from gauging, metering, and proving. Higher levels of security allow only the appropriate personnel to change basic data (for example, tank strapping data, meter data, and prover calibration data), computational procedures, and/or algorithms.
Physical Security
The computer should be maintained in a secure location by locking the cabinet, house, or area where it is located. If this is not possible, then other means should be used to make the equipment inaccessible to tampering. Some other equipment, like temperature/pressure averagers, may be sealed to prevent tampering.
September 2002
8.7
BP Pipelines
More About It
4 0 0 - 1 B P
September 2002
8.8
BP Pipelines
Reference Documents
1. Pipelines (NA) Business Unit Safety Manual
September 2002
8.9
Chapter 9
Scope
This chapter lists procedures to verify the operation of LACT units on lease tanks containing crude oil and of ACT units measuring weathered crude oil or liquid petroleum products. This chapter also describes the components and operation of the electronic liquid measurement (ELM) system.
BP Pipelines
September 2002
ELM functions: Meter ticketing Meter proving control Sampling system control Communication link to the SCADA system
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
Introduction
In addition to verifying the operation of LACT/ACT units, you may also prove the meter and densitometer on ACT units, calibrate temperature and pressure devices, take samples, and test for S&W (in crude). See the appropriate chapters in this manual for these procedures. The ELM system uses a computer to simplify meter ticketing and to control proving and sampling by actuating the proving sequence and signaling the PLC to operate the appropriate valves. It also signals the sampler to sample the liquid.
September 2002
9.4
BP Pipelines
September 2002
9.5
BP Pipelines
High-level switch
Control panel
Temperature probe
Back pressure valve
Densitometer
Meter
Low-level switch
Figure 9.1. LACT unit Some of the following procedures apply only to LACT units. 1. Check for signs of leaks. Check the entire unit, particularly downstream from the meter and in the sampling system, for signs of leaks. If you nd any leaks, shut the unit down and have the leaks xed.
2. Check for tampering. Check the seals for signs of tampering or missing seals. Also check the entire unit for any signs of tampering or any indication of repair since the last inspection (see Chapter 8, Seals and Security).
3. Remove necessary seals (see Chapter 8, Seals and Security). 4. Record the meter reading. 5. Read and record the total hours run on the elapsed time counter (hour meter), if so equipped. 6. Check the level of the liquid in the sales tank. Make sure there is enough liquid in the tank to check the LACT operation.
September 2002
9.6
BP Pipelines
7. Check the ow rate. Calculate the current ow rate in barrels per hour. Do this by timing the ow with a stopwatch for 36 seconds and multiplying by 100. Check that the ow rate has not changed more than 15% from the last meter proving. If it has changed more than 15%, readjust the ow rate or reprove the meter. If the change in ow rate is the result of equipment malfunctioning, shut down the unit until the problem is repaired. Calculate the volume run through the meter since the last checkout and ensure that it is reasonable.
8. Check the programmable clock on the control panel. If the LACT is equipped with a clock to control on-off scheduling, check it for proper setting and operation. Follow the manufacturers instructions.
9. Check the height of the low-level switch on the LACT tank. If the sales tank is equipped with an adjustable pressure-activated level switch, check that the shutdown level is at least two feet above the connection height.
10. Check the LACT operation. If the LACT unit is not running and there is enough liquid in the tank, turn on the LACT by switching the Hand/Off/Auto switch to Hand. If the unit is running, perform the following checks without adjusting the Hand/Off/Auto switch. Observe the units operation. Check for smooth startup and operation. Check for leaks, excessive vibrations, high pressure, and unusual noises. Check the level of sample in the sample pot to make sure it has increased since the previous observation. The pot should be lling the entire time the LACT unit is running. If it is not, the sample probe may be plugged or broken and may need to be repaired or replaced. Check the sampler operation. Note whether the sampler is sampling at the proper interval. The interval frequency is based on the size of the parcel, extractor grab size, and the sample receiver (pot) volume. The ow meter volumetrically paces sampler operation. The interval may be preset in a ow computer, PLC, Durant counter, or sampler controller. Check that the sampler is completing its cycle and not hanging in the open or closed position. To do this, you may need to disconnect the sample line from the extractor and catch one or two samples from the sampler in a 100-ml centrifuge tube.
September 2002
9.7
BP Pipelines
Measure the amount of sample being taken per grab. Then verify that the correct amount of sample is being taken for the amount of liquid being transferred. Use the number of milliliters per grab to calculate how many gallons per transfer (batch or parcel) are being taken: bbls per transfer -------------------------------------------------------- = grabs per transfer number of bbls per grab grabs per transfer ml sample per grab = ml sample per transfer ml sample per transfer ----------------------------------------------------- = gallons of sample per transfer 3,785 Check the pressure in the sample storage pot, if one is available and operable. The sample pot should have some pressure in it, but no more than 15 psi. Check the S&W monitor. After the LACT has run for two minutes (to purge the liquid in the line), obtain a line sample and grind it out in the centrifuge. Reset the S&W Limit knob to the grindout value obtained to verify that the bad oil indicator light comes on. Check that the timer times out after 30 seconds and the diverter valve moves to the divert position. After the valve diverts, make sure the meter has stopped completely. Turn the S&W Limit knob back to the maximum allowable percentage of water. Check that the merchantable oil light comes on, the timer times out, and the diverter valve returns to the pipeline position. If this does not happen, grind out another sample to check that the S&W level has not gone above the maximum allowable percentage of water. If the monitor does not divert or diverts at the wrong S&W level, then it must be repaired or recalibrated. Follow the manufacturers recommendations for repairing or calibrating the monitor. Record the S&W percentage from the line sample on the LACT/ACT checklist. If the Hand/Off/Auto switch is set to Auto, check the meter monitor on the control panel. Close a suction valve on the LACT pump to stop the meter. Check that the unit shuts down on meter failure within 5 minutes. After shutdown, open the suction valve and reset the LACT control panel. Make sure the unit starts operating again. Check the temperature and pressure averagers. Insert a glass thermometer into the thermowell adjacent to the meter. Read the temperature on the glass thermometer and the temperature shown on the temperature averager (TA) display. Record these two temperatures on the form or checklist.
September 2002
9.8
BP Pipelines
Check the pressure averager by comparing that reading with the pressure indicated on the pressure gauge on the unit. Record the two pressure readings on the form or checklist. If the temperatures disagree by more than 0.2F and/or the pressure disagrees by more than 5 psi, shut down the equipment and repair or recalibrate it. Check the mechanical meter register (PD meter) for jerky or erratic movement. Monitor the register for at least 10 barrels to make sure it is operating properly. If the system also has a ow computer, compare the mechanical registers total barrels to the ow computers total barrels. If they do not agree, shut down the system and arrange for repairs. If the meter has a local electronic register, record the batch total barrels and compare this to the batch total barrels on the ow computer display. If these do not agree, shut down the system and arrange for repairs. Note any unusual observations in the Remarks section on the checklist. Sign or initial the checklist. Return the LACT to normal operation. If you encountered no problems during the LACT checkout and you had previously turned the Hand/Off/Auto switch to Hand, turn the switch back to the Auto position.
11. Replace any seals that were broken during the LACT/ACT checkout, as well as any missing seals and any seals showing signs of excessive age (see Chapter 8, Seals and Security).
September 2002
9.9
BP Pipelines
If during verication a temperature is out of tolerance (greater that 0.2F), check both the RTD and the transmitter to determine which is in error. Check or calibrate the RTD and transmitter independently in accordance with the manufacturer's instructions.
If during verication, a pressure is out of tolerance (greater that 5 psig), use a certied deadweight tester for calibration.
September 2002
9.10
BP Pipelines
Use three- or four-wire 100-ohm platinum RTDs to eliminate error due to changes in ambient temperature. The ow computer must be secure from unauthorized access. See Chapter 8, Seals and Security for more information about security for an ELM system.
Manual Input
You may need to enter some or all of the following information into the ow computer: Product name and grade Products physical properties API gravity at 60F Vapor pressure S&W content (crude only)
Month, day, year, and start and nish of each transfer/batch K factor (pulses per barrel) Meter factor Batch volume required
September 2002
9.11
BP Pipelines
Live Input
The computer will receive the following input data: Meter pulses Pressure transducer signal Temperature transducer signal Densitometer signal, where applicable
September 2002
9.12
BP Pipelines
Metered volume (in barrels to the nearest tenth) Temperature compensated (yes or no) Meter factor Compressibility factor Indicated (metered) barrels Gross standard barrels Net standard barrels Station totals (if multiple meters)
Name of BP Pipelines representative, electronic ID, date on meter volume Name of BP Pipelines representative, electronic ID, date off meter volume
Data from the ELM is usually transmitted electronically via the SCADA system to the Tulsa Control Center.
September 2002
9.13
BP Pipelines
More About It
In this section you will nd Figure 9.2, an example of a LACT/ACT inspection form, and Figure 9.3, a blank inspection form that you can copy and use.
September 2002
9.14
BP Pipelines
September 2002
9.15
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 6.1 Metering Assemblies Lease Automatic Custody Transfer (LACT) Systems 2. API Manual of Petroleum Measurement Standards, Chapter 6.6 Pipeline Metering Systems 3. API Manual of Petroleum Measurement Standards, Chapter 21.2 Flow Measurement Electronic Liquid Measurement 4. BP Self-Study Guide, Field Specialist IV Module 4 LACT/ACTs and their Components 5. Pipelines (NA) Business Unit Safety Manual
September 2002
9.16
Chapter 10
Automatic Sampling
Quick Reference
Safety
Monitor H2S while sampling. Some petroleum products are extremely ammable and/or poisonous. Follow the applicable safety procedures in the Pipelines (NA) Business Unit Safety Manual when sampling these dangerous liquids. Wear safety glasses or goggles at all times while handling pressurized receivers. Keep containers closed when not in use. Keep work areas as clean as possible and well-ventilated. Clean up spills promptly and in accordance with safety, health, and environmental regulations. Observe established exposure limits, and wear suitable protective clothing and equipment. Dispose of all samples and security seals properly. Follow all applicable safety rules in the Pipelines (NA) Business Unit Safety Manual.
Summary of Measurement Procedures for Flowing Liquids 1. Check the operation of the LACT/ACT unit and ELM system or meter installation (Chapter 9). 2. Check meter operation (Chapter 11, Chapter 12, Chapter 13). 3. Mix and withdraw samples (Chapter 10). 4. Analyze the samples as required for the specic type of transaction and product (Chapter 2, Chapter 5, Chapter 6). 5. Prove the meter (Chapter 14). 6. Calibrate temperature and pressure devices (Chapter 9). 7. Verify/calibrate the densitometer, as appropriate (Chapter 9). 8. Record all results (Chapter 17).
Scope
These procedures apply to crude oil and products owing through LACT or ACT units.
BP Pipelines
September 2002
Mix samples for the correct amount of time. Label each sample before taking it to the laboratory for testing. Test for S&W, temperature, and gravity, as required.
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
Introduction
An automatic sampler, which is part of a meter installation, periodically withdraws small amounts of liquid from a pipeline and delivers it to a stationary or portable sample receiver (pot). The amount it removes is proportional to the owthrough the pipeline. If the automatic system is set up and maintained properly, it provides a more representative sample of the liquid in the pipeline than manual sampling. After taking the sample from the automatic sampler, your job is to continue to handle it in such a way that the sample you deliver for analysis is still representative of the owing uid.
Automatic Sampling
September 2002
10.4
BP Pipelines
See More About It at the end of this chapter for more information about what type of mixing procedure to use for different tests.
Automatic Sampling
September 2002
10.5
BP Pipelines
6. Take the marked container(s) immediately to the laboratory for analysis, as gravity testing must take place within 2 hours of taking the sample. 7. Empty the sample receiver, and ush it with solvent. Take extra care to clean the receiver after sampling a high water content batch of crude oil. 8. Periodically open the receiver and wipe it clean with a lint-free rag. Make sure the top is clean and dry.
Automatic Sampling
September 2002
10.6
BP Pipelines
Receiver Testing
The sample receivers are also tested to assure that they properly mix the oil collected in the receiver. Tests are conducted at both low and high levels in the receiver, as well as low and high water concentrations. An independent third party usually performs the testing of the sample and receiver, in accordance with API procedures. This testing is required by U.S. Customs for sampling systems associated with foreign trade zones (FTZ).
Automatic Sampling
September 2002
10.7
BP Pipelines
The following equipment is part of the sampling system (see Figure 10.1): Conditioning equipment upstream of the sampling location to mix the liquid Sample probe and extractor to collect the sample from the center of the pipe Automatic controller to control the frequency (number of grabs) based on the ow and thus the volume of sample taken per batch One or more sample receivers (pots) to collect and hold the extracted sample grabs
Conditioning equipment
(pump, vertical pipe, elbows, tees, reduced-diameter pipe)
Flow
Sample extractor and probe
Sample receiver
Automatic Sampling
September 2002
10.8
BP Pipelines
Flow
Mixing system
Automatic Sampling
September 2002
10.9
BP Pipelines
Portable receivers (Figure 10.3) vary in size from 5 to 10 gallons. Normally they are not pressurized; if they are, the pressure is less than 15 psig.
Quick connection
Vacuum relief
Level gauge
Pressure relief
Pressure gauge
Downcomer
Quick connection
Figure 10.3. Portable sample receiver (pot) Constant-pressure receivers may be used for high-vapor-pressure crude oils and products. They may also be used for low-pressure liquids if loss of light ends is a problem. Automatic sampling systems on stabilized crude oil pipelines usually use nonpressurized receivers.
Automatic Sampling
September 2002
10.10
BP Pipelines
Mixing Equipment
Fixed sample receivers come equipped with mixing elements so you can properly mix the composite sample before testing for S&W or other components. Different types of mixers are available. Portable receivers require a separate mixing system (Figure 10.4).
Motor
Figure 10.4. Sample mixing system for portable receivers BP Pipelines prefers a centrifugal pump installed as close as possible to the sample pot outlet connection. This pump must be capable of displacing the entire sample volume at least twice per minute. The piping from the pump back to the sample pot must also be arranged so that no sediment and water can be trapped in low spots and so that you can withdraw a sample while mixing is taking place. A pressure gauge on the return line to the pot indicates that the pump is working properly while mixing. When emptying the container, watch the pressure gauge, which will indicate when the container is empty by a decrease in pressure.
Automatic Sampling
September 2002
10.11
BP Pipelines
More About It
This section includes the following tables and instructions for using them. Table 10.1. Recommended Mixing Procedures Table 10.2. Typical Sizes of Sample Receivers for Various Operations
Mixing Samples
A sample may or may not need to be mixed before testing it, depending on the type of test and how homogeneous the sample is. Because an automatic sampling system takes samples over a period of many hours, some settling almost always occurs. You can mix a sample with a stand-alone power mixer (see Figure 10.4), with an internal mixer in the sample pot, or by shaking it. When using a stand-alone power mixer, be sure to use the correct type for the container you have, as the mixer/container combination has been tested and proven effective. Table 10.1 lists the mixing recommendations for various tests.
Shaking
None
X X X X X
Automatic Sampling
September 2002
10.12
BP Pipelines
Automatic Sampling
September 2002
10.13
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 8.2 Automatic Sampling of Petroleum and Petroleum Products 2. API Manual of Petroleum Measurement Standards, Chapter 8.3 Mixing and Handling of Liquid Samples of Petroleum and Petroleum Products 3. API Manual of Petroleum Measurement Standards, Chapter 21.2 Flow Measurement Electronic Liquid Measurement 4. BP Self-Study Guide, Field Specialist III Module 1 Sediment and Water Determination 5. Pipelines (NA) Business Unit Safety Manual
Automatic Sampling
September 2002
10.14
Chapter 11
Scope
This chapter lists procedures for inspecting PD meters used for custody transfer of crude oil or liquid petroleum products. It also describes the design and operation of positive displacement meters.
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
Introduction
Displacement meter systems normally perform well for long periods with a minimum of maintenance. In addition to proving the meter to establish its accuracy (see Chapter 14, Proving a Meter), a positive displacement meter and its accessory equipment need to be inspected on a routine basis. The frequency of inspections depends on volume through the meter, but should be at least weekly.
September 2002
11.3
BP Pipelines
Meter head
4 6 3 7 0
Typical seals
Flow
Figure 11.1. Sealing points on a PD meter To repair a meter, disassemble it, inspect it, clean and adjust it, reassemble it, and prove it. Perform all maintenance according to the manufacturers instructions and procedures. Keep records of all maintenance and testing you do. Inspect the equipment routinely (at least weekly, depending on throughput) following the list below. 1. Meter Listen for noisy operation, which may indicate that the meter is wearing and may fail soon. If necessary, shut the meter down and use alternative measurement methods until it can be taken apart and repaired. Inspect the meter for vibrations and leaks. Inspect the lubrication system. Verify the pulse transmitters by comparing the barrels on the mechanical register with the barrels on the electronic register. Meter proving veries meter performance. Watch for and report meters that fail the repeatability requirements or that have a meter factor that varies by more than 0.0015 for ACT meters or 0.0025 for LACT meters since the last proving.
September 2002
11.4
BP Pipelines
Maintenance is necessary when the meter factor varies by more than 0.0050 from the original meter factor. Meter factors should be trended to monitor the condition of the metering system.
2. Strainer and air eliminator Inspect the strainer and air eliminator before every custody transfer. Clean the strainer whenever the differential pressure is greater than normal. Repair the strainer and eliminator as needed.
3. Flow computer Follow the manufacturers instructions for verifying the status of batch information on the ow computer before every custody transfer.
4. Counters, indicators, and volume-averaging devices Follow the manufacturers recommendations for maintenance, if any. Inspect the device for defects before every custody transfer and notify your supervisor if any part of the device appears defective.
5. Double block-and-bleed valves Check these valves for leaks before every custody transfer. Drain or monitor the cavity between the dual seals. If you see a leak, notify your supervisor and initiate repairs as soon as possible.
When you suspect measurement error with a meter, use the following checklist to determine the specic problem: 1. Check for air or vapor in the system. 2. Determine whether there have been any changes in the physical properties of the metered liquid, especially its viscosity. 3. Check pressure, temperature, and density-sensing devices for error. 4. Check all electrical equipment for any failures, including the pulse generator, counters, coil, preampliers, signal transmission system, power supply, and all readout devices. 5. Check for leakage from all isolation and diversion valves. 6. Check for proper gearing in the meter accessory drive.
September 2002
11.5
BP Pipelines
Meter head
4 6 3 7 0
Flow
September 2002
11.6
BP Pipelines
Pulse-driven (electronic) accessories: Readout that indicates the volume Flow computer that receives signals from the meter and other sensors to calculate volume or mass ow quantity and displays, transmits, and prints the data Preset totalizer to stop the ow at a preselected quantity Proving counter that displays the pulsed output from the meter Flow-rate indicator that converts an electrical signal to a visual display of ow rate
Protection and control equipment: Strainers and lters to prevent solids from entering the meter Air or vapor eliminator to remove gases from the liquid ow Valve for starting, stopping, and controlling the ow of liquid Pressure-reducing valves Temperature, pressure, density, and viscosity monitors Thermometer Temperature and pressure averagers Pressure gauges
September 2002
11.7
BP Pipelines
Pulse transmitter
RTD transmitter
Flow scoop
Flow
Sample controller
September 2002
11.8
BP Pipelines
API Manual of Petroleum Measurement Standards, Chapters 5.2 and 5.4, provides guidelines for choosing the appropriate accessory equipment and designing the piping conguration for a PD meter installation. It is important that an installation use the accessory equipment that the manufacturer recommends, that the ow rates are within those that the manufacturer species, and that meters are operated only with the liquids the installation was designed for. For example, the combination of measuring element, gear train, and stack accessories must be carefully matched to provide accurate volume registration. Use 100% gearing with a dummy calibrator. Table 11.1 offers some recommended combinations.
Table 11.1. Recommended Combinations of Meter Size, Measuring Element, and Gear Ratio
Meter Size 2" 3" 4" 6" 6" 8" 10" 12" 16" Typical Measuring Element (volume per revolution) 0.364 gal 1.4 gal 2 gal 3 gal 0.0714 bbl 0.1428 bbl 0.236 bbl 0.5 bbl 1 bbl Gear Ratio of Meters Gear Train 13.7:1 3.57:1 5:2 5:3 14:1 7:1 4.24:1 2:1 1:1 Output to Calibrator (volume per revolution) 5 gal 5 gal 5 gal 5 gal 1 bbl 1 bbl 1 bbl 1 bbl 1 bbl
September 2002
11.9
BP Pipelines
PD meters should be sized to operate between 40% and 80% of the manufacturers rated capacity. Operating a meter in this range improves accuracy and extends meter life. The following table gives examples of recommended operating ranges for PD meters typically used by BP Pipelines.
September 2002
11.10
BP Pipelines
Operation of a PD Meter
The accuracy of a meter depends on the properties of the liquid being measured, the condition of the meter and its accessories, the meters operating conditions, and variations, if any, between operating and proving conditions. Table 11.3 shows the results of changes in some of these factors.
Table 11.3. Effect of Changes in the Conditions of the Fluid and PD Meter on Measurement Accuracy
Factor Fluid properties Change in Conditions Increase in temperature viscosity gravity Result Increased slippage underregistration Meter Factor Change Increases
Decreases
Air or gas measured along with the hydrocarbons overregistration Increased wear on meter parts underregistration Possible damage to meter parts underregistration Increased slippage underregistration Increased slippage underregistration Coating on meter surface overregistration
Decreases
Increases
Increases
Meter conditions
Wear
Increases
Friction
Increases
Decreases
September 2002
11.11
BP Pipelines
Table 11.3. Effect of Changes in the Conditions of the Fluid and PD Meter on Measurement Accuracy (continued)
Factor Operating conditions Change in Conditions Control valve chatter Result Pulsating ow overregistration Pulsating ow overregistration Meter Factor Change Decreases
Decreases
September 2002
11.12
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 5.1 General Considerations for Measurement by Meters 2. API Manual of Petroleum Measurement Standards, Chapter 5.2 Measurement of Liquid Hydrocarbons by Displacement Meters 3. API Manual of Petroleum Measurement Standards, Chapter 5.4 Accessory Equipment for Liquid Meters 4. BP Self-Study Guide, Field Specialist IV Module 4 Meters and Meter Proving 5. Pipelines (NA) Business Unit Safety Manual
September 2002
11.13
Chapter 12
Turbine Meters
Quick Reference
Safety
Follow all applicable safety rules in the Pipelines (NA) Business Unit Safety Manual.
Summary of Measurement Procedures for Flowing Liquids 1. Check the operation of the LACT/ACT unit and ELM system or meter installation (Chapter 9). 2. Check meter operation (Chapter 11, Chapter 12, Chapter 13). 3. Mix and withdraw samples (Chapter 10). 4. Analyze the samples as required for the specic type of transaction and product (Chapter 2, Chapter 5, Chapter 6). 5. Prove the meter (Chapter 14). 6. Calibrate temperature and pressure devices (Chapter 9). 7. Verify/calibrate the densitometer, as appropriate (Chapter 9). 8. Record all results (Chapter 17).
Scope
This chapter lists procedures for inspecting turbine meters used for custody transfer of crude oil or liquid petroleum products. It also describes the design and operation of turbine meters.
BP Pipelines
September 2002
BP Pipelines
September 2002
BP Pipelines
Introduction
Turbine meter systems normally perform well for long periods with a minimum of maintenance. In addition to proving the meter to establish its accuracy (see Chapter 14, Proving a Meter), a turbine meter and its accessory equipment need to be inspected on a routine basis. The frequency of inspections depends on volume through the meter, but should be with each batch or at least weekly.
Turbine Meters
September 2002
12.3
BP Pipelines
2. Strainer and air eliminator Inspect the strainer and air eliminator before every custody transfer. Clean the strainer whenever the differential pressure is greater than normal. Repair the strainer and eliminator as needed.
3. Flow computer Follow the manufacturers instructions for verifying the status of batch information on the ow computer before every custody transfer.
4. Counters, indicators, and volume-averaging devices Follow the manufacturers recommendations for maintenance, if any. Inspect the device for defects before every custody transfer and notify your supervisor if any part of the device appears defective.
5. Double block-and-bleed valves Check these valves for leaks before every custody transfer. Drain or monitor the cavity between the dual seals. If you see a leak, notify your supervisor and initiate repairs as soon as possible.
6. When you suspect measurement error with a meter, check it immediately. See Chapter 14, Proving a Meter for a checklist to determine the specic problem.
Turbine Meters
September 2002
12.4
BP Pipelines
Upstream stator
Turbine Meters
September 2002
12.5
BP Pipelines
Pickup coil
Upstream stator
Downstream stator
Rotor
Turbine Meters
September 2002
12.6
BP Pipelines
Turbine Meters
September 2002
12.7
BP Pipelines
Instrument air
Relief valve Sample loop Sample removal valve Sample vessel Densitometer proving connections Nitrogen precharge vessel
Pipeline Densitometer
Flow computer Signal preamplifier Flow conditioner Strainer Flow Turbine meter Local totalizer
Thermowell
Temperature transducer Pressure gauge Detector switch 4-way block & bleed valve with electric motor operator
Pressure transducer
Turbine Meters
September 2002
12.8
BP Pipelines
Conventional turbine meters should be sized to operate between 50% and 100% of the manufacturers rated capacity. Helical turbines can be sized to operate between 30% and 100%. Operating the meters in these ranges will improve accuracy and extend meter life. Table 12.1 and Table 12.2 give examples of recommended operating ranges for both types of turbine meters used by BP Pipelines.
Table 12.1. Recommended Operating Range for Various Sizes of Conventional Turbine Meters
Flow Rate (bph) Meter Size 50% 2" 3" 4" 6" 8" 10" 12" 16" 160 465 890 2,070 3,710 5,700 8,550 12,850 100% 320 930 1,780 4,140 7,420 11,400 17,100 25,700 7,560 2,016 3,000 1,000 500 250 200 100 Pulses/bbl
Note: This data will vary depending on the specic manufacturer of the meter.
Turbine Meters
September 2002
12.9
BP Pipelines
Table 12.2. Recommended Operating Range for Various Sizes of Helical Turbine Meters
Flow Rate (bph) Meter Size 30% 2" 3" 4" 6" 8" 10" 12" 16" 57 208 566 1,132 1,887 3,774 5,661 7,548 100% 189 692 1,887 3,774 6,290 12,580 18,869 25,159 3,964.9 696.7 264 113.2 71.7 25.8 13.1 7.6 Pulses/bbl
Note: This data will vary depending on the specic manufacturer of the meter.
A secure and reliable pulse transmission system is also critical to a turbine meters accuracy. The transmission lines should be as short as possible. Transmission lines should be routed away from sources of electrical interference. Transmission lines should be continuous, if possible. If they are not, the shielding should be continuous. A ground shield at the receiving end (ow computer/PLC) prevents ground-loop effects.
Turbine Meters
September 2002
12.10
BP Pipelines
Figure 12.4 and Figure 12.5 are schematics of transmission lines to turbine meters with and without preamps.
3, 4, or 5/c shielded cable (external DC power supply required) Remote electronics Meter
Remote electronics
Meter
Turbine Meters
September 2002
12.11
BP Pipelines
Turbine Meters
September 2002
12.12
BP Pipelines
Table 12.3. Effect of Changes in the Conditions of the Fluid and Turbine Meter on Measurement Accuracy
Factor Fluid properties Change in Conditions Entrained air in the liquid Result Air or gas measured along with the hydrocarbons overregistration Increased wear on meter parts underregistration Possible damage to meter parts underregistration Distorted velocity prole of owing uid overregistration Increased velocity of uid through meter overregistration Reduced ow area overregistration Worn bearings underregistration Excessive mechanical friction between meter parts underregistration Pulsating ow overregistration Pulsating ow overregistration Meter Factor Change Decreases
Increases
Increases
Meter conditions
Upstream obstructions
Decreases
Decreases
Decreases
Increases
Friction
Increases
Operating conditions
Decreases
Decreases
Turbine Meters
September 2002
12.13
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 5.1 General Considerations for Measurement by Meters 2. API Manual of Petroleum Measurement Standards, Chapter 5.3 Measurement of Liquid Hydrocarbons by Turbine Meters 3. API Manual of Petroleum Measurement Standards, Chapter 5.4 Accessory Equipment for Liquid Meters 4. BP Self-Study Guide, Field Specialist IV Module 4 Meters and Meter Proving 5. Pipelines (NA) Business Unit Safety Manual
Turbine Meters
September 2002
12.14
Chapter 13
Other Meters
Quick Reference
Scope
This chapter briey describes the design and operation of orice, ultrasonic, and Coriolis meters, which may be used for custody transfer of natural gas, crude oil, and liquid petroleum products.
Summary of Measurement Procedures for Flowing Liquids 1. Check the operation of the LACT/ACT unit and ELM system or meter installation (Chapter 9). 2. Check meter operation (Chapter 11, Chapter 12, Chapter 13). 3. Mix and withdraw samples (Chapter 10). 4. Analyze the samples as required for the specic type of transaction and product (Chapter 2, Chapter 5, Chapter 6). 5. Prove the meter (Chapter 14). 6. Calibrate temperature and pressure devices (Chapter 9). 7. Verify/calibrate the densitometer, as appropriate (Chapter 9). 8. Record all results (Chapter 17).
BP Pipelines
September 2002
BP Pipelines
Introduction
Meters other than PDs and turbines, such as orice meters, are not routinely used in crude oil service. They are more commonly used to measure natural gas, LPGs, dense-phase uids (such as ethylene), and some chemicals. Ultrasonic meters and Coriolis meters are additional types of meters that are starting to achieve recognition for both crude and products applications. These are inference meters, so called because they do not measure the liquid volume directly but infer it by measuring other properties of the liquid. See BP Measurement Manual, Part II Natural Gas for more detailed information on orice and ultrasonic meters, and BP Measurement Manual, Part III Chemicals and Petrochemicals for more information on Coriolis meters.
Other Meters
September 2002
13.2
BP Pipelines
Orifice Meter
The orice meter measures the ow rate and thus the volume by creating and recording a pressure differential in the uid.
Temperature recorder
LCD Computer
219815
Pen
Transducers
Chart
Portable hand unit
Secondary element
Orifice plate
Primary element
Figure 13.1. Orice meter installation The uid passes through a plate with a small hole in it, the orice plate (Figure 13.1), which restricts the ow and reduces the owing pressure of the uid. Small gauge lines in pressure taps near the orice plate connect the tting to a differential pressure sensor and mechanical ow recorder or to pressure transducers and a computer. The recorder or computer records the pressure drop and other variables, such as temperature and static pressure, that are necessary for calculating ow rate.
Other Meters
September 2002
13.3
BP Pipelines
Orice plates are available in different sizes to t different pipeline diameters. The orice also varies in size to adjust the ow rate of the uid. The thickness of the plate is another variable. The diameter, orice size, and plate thickness are all important factors in designing an orice meter installation. Orice meters are often used for custody transfer measurement of gases and some petrochemicals, but not liquids. They are often used in reneries and chemical plants to measure both liquids and gases, but not for custody transfer.
Other Meters
September 2002
13.4
BP Pipelines
Ultrasonic Meter
Ultrasonic meters are inferential meters that derive the uid ow rate by measuring the transit times of high-frequency sound pulses. Transit times are measured for sound pulses traveling diagonally across the pipe, downstream with the ow, and also upstream against the ow. The difference in these transit times is related to the average uid ow velocity along the acoustic paths. Numerical calculation techniques are then used to compute the average axial ow velocity and the volume ow rate at line conditions through the meter. The primary element is a tube containing at least one pair of transducers that generate and receive sound waves, and cables to connect them to an electronic processor (Figure 13.2). The processor operates the transducers, measures the travel speed of the sound waves, processes the data, and records and displays the calculated volume and other data.
Meter body
Transducers
Other Meters
September 2002
13.5
BP Pipelines
An electrical voltage vibrates the crystal or ceramic transducers, which creates sound waves in the uid (see Figure 13.3). Each transducer can also convert sound waves to an electric signal. The processor measures the time it takes for sound to travel from one transducer to the other and back again.
Path 1 Path 2
0.5r
0.5r
View A-A
A Sonic path
Other Meters
September 2002
13.6
BP Pipelines
Coriolis Meter
Coriolis meters measure the mass, or weight, of the uid in the line, as well as the uid density. From these measurements the volume can then be calculated.
Drive coil
Flow
Velocity detectors
Figure 13.4. Coriolis meter The Coriolis meter has a sensor element and a transmitter (Figure 13.4). The sensor element consists of one or two ow tubes, a drive coil, and two pick-up coils (one for the left side and one for the right side). The product ows through the tube(s). The drive coil, when energized, causes the tubes to vibrate, and the ow of liquid through the tubes causes the tubes to twist (Figure 13.5 and Figure 13.6). This twisting is called the Coriolis effect. Each pick-up coil produces a sine wave that is representative of the twisting action. The phase shift between the two sine waves is directly proportional to the products mass ow (the greater the ow, the greater the phase shift). A temperature sensor accounts for changes in the elasticity of the tube due to temperature. As the density of the uid increases, the frequency of tube oscillation decreases because of the dampening effect of the heavier uid. The transmitter measures this frequency. The drive coil, pick-up coils, and the temperature sensor connect to the transmitter. The transmitter processes the signals from the sensor, determines the density and the mass and/or volumetric ow, and provides necessary inputs and outputs. Coriolis meters have been in service for measuring petrochemicals, but with the publication of two Draft API Standards in 2001, the industry is beginning use them for custody transfer measurement of crude oil and rened products.
Other Meters
September 2002
13.7
BP Pipelines
Fluid force
Vibrating frequency
Fluid force
Twist angle
Other Meters
September 2002
13.8
BP Pipelines
Reference Documents
1. BP Measurement Manual, Part II Natural Gas 2. BP Measurement Manual, Part III Chemicals and Petrochemicals 3. API Manual of Petroleum Measurement Standards, Chapter 14.3 Concentric, Square-Edged Orice Meters 4. API Manual of Petroleum Measurement Standards, Draft Standards Measurement of Single-Phase Intermediate and Finished Hydrocarbon Fluids by Coriolis Meters Measurement of Crude Oil by Coriolis Meters
Other Meters
September 2002
13.9
Chapter 14
Proving a Meter
Quick Reference
Safety
Make sure the portable prover is grounded to prevent static electricity buildup. Follow all applicable safety rules in the Pipelines (NA) Business Unit Safety Manual.
Summary of Measurement Procedures for Flowing Liquids 1. Check the operation of the LACT/ACT unit and ELM system or meter installation (Chapter 9). 2. Check meter operation (Chapter 11, Chapter 12, Chapter 13). 3. Mix and withdraw samples (Chapter 10). 4. Analyze the samples as required for the specic type of transaction and product (Chapter 2, Chapter 5, Chapter 6). 5. Prove the meter (Chapter 14). 6. Calibrate temperature and pressure devices (Chapter 9). 7. Verify/calibrate the densitometer, as appropriate (Chapter 9). 8. Record all results (Chapter 17).
Scope
These procedures apply to proving meters on crude and product pipelines, using pipe provers or small-volume provers.
The new meter factor must be within 0.0025 of the previous meter factor (under the same operating conditions). In most locations, apply the meter factor to the entire ticket period (often the entire month). For leases under the jurisdiction of the Bureau of Land Management (BLM) or Minerals Management Service (MMS), apply the meter factor from the time of the proving forward and not retroactive to the start of the ticketing period.
BP Pipelines
September 2002
BP Pipelines
The meter factor is calculated from 5 consecutive runs, out of a maximum of 10 runs, that agree (highest to lowest) to within 0.05%. The new meter factor must be within 0.0015 of the previous meter factor (under the same operating conditions).
Every six months: Pull the prover sphere and inspect it for damage. Verify that the size is still the same as during the last prover calibration.
Proving a Meter
September 2002
14.2
BP Pipelines
September 2002
BP Pipelines
Introduction
No meter can be absolutely accurate. To know the correct amount of oil being measured by a meter, it is necessary to determine how well a particular meter is performing under real operating conditions by proving it. Proving involves measuring the liquid ow through the meter and through a prover and comparing the two volumes. The two volumes are used to calculate the meter factor. All parties involved in a custody transfer use the meter factor to correct any inaccuracy in the meters measurements until the next scheduled proving. The type of prover used depends on the volume of liquid to be measured and the nature of the transfer.
Proving a Meter
September 2002
14.4
BP Pipelines
Proving a Meter
September 2002
14.5
BP Pipelines
Procedures
Calibrate provers regularly, normally every year to three years (see Chapter 16, Waterdraw Calibration). Keep provers in top mechanical condition to ensure accurate results. See Troubleshootingbelow for details about maintaining provers. During proving, the uid properties (for example, API gravity), ow rate, pressure, and temperature should be similar to those during normal operating conditions for the meter. The specic procedure for proving a meter varies according to the type of proving device. For each proving, however, follow the same procedure under the same operating conditions for each proving run so that you can produce repeatable results. Use the meter proving tool in the SMART program to enter proving data. SMART will then calculate the new meter factor.
4. After diverting the oil into the prover, allow time for the prover temperature to stabilize. The temperature is considered to be stabilized when the prover outlet temperature is constant and agrees closely with the meter temperature. 5. To adjust the ow rate back to normal, lower the back-pressure setting on the meter. Reset the back-pressure when you are nished proving. 6. If the meter is new or has been recently repaired, verify that the number of pulses generated per barrel matches the meter manufacturers specications as indicated on the Meter Proving Report. Reset the electronic pulse counter to zero. Start the electronic pulse counter with the remote switch attached to the counter. Open the ow through the meter. After 10 barrels have been registered, stop the electronic pulse counter. Read the counter.
Proving a Meter
September 2002
14.6
BP Pipelines
Using the following formula, calculate the pulses per barrel: total pulses --------------------------------------------------------------------------------------------------------------- = pulses per barrel closing meter reading opening meter reading
7. Take a sample of the owing oil and determine its gravity using a hydrometer (see Chapter 2, Gravity and Temperature Measurement in Tanks and Chapter 10, Automatic Sampling). Record the results so you can enter them later on the Meter Proving Report. 8. Reset the electronic pulse counter to zero. 9. Set the four-way valve and launch the sphere for the rst pass. 10. For a unidirectional prover, record the number of pulses at the end of the sphere (or displacer) pass. 11. For a bidirectional prover, reverse the four-way valve to allow the sphere to change directions, and record the number of pulses for the round trip. 12. Note the temperature and pressure of the prover and record the readings for this run. If both inlet and outlet thermometers are available, average the two readings. 13. For a bidirectional prover, record the meter temperature (unless temperature-compensated) and meter pressure. 14. Reset the prover counter and repeat steps 9-13 until you have accumulated data for 5 consecutive runs. If the results of the 5 runs meet BP Pipelines requirements of 0.05% repeatability (see Repeatability Requirementsbelow), stop. If not, continue the proving runs until you obtain 5 consecutive runs that are repeatable. If you make 10 runs without meeting the repeatability requirements, stop and determine the problem (see Troubleshooting below).
15. Return the ow through the LACT or ACT to normal. 16. Reset the back-pressure setting on the meter. 17. Complete the Meter Proving Report on your laptop. 18. Compare the new meter factor with the previous one. If the new meter factor varies from the previous one by more than 0.0015 for ACTs or more than 0.0025 for LACTs, identify the problem (see Troubleshooting below).
Proving a Meter
September 2002
14.7
BP Pipelines
3. Establish ow through the prover and check for leaks. 4. Launch the prover sphere and bleed air from the vent valves while the sphere is in motion. Be sure to bleed both prover chambers. 5. Verify that the prover counter is receiving signals from the pulse generator. Launch the sphere a few times to make sure that the detector switches are gating properly. 6. Clear the counter and then follow steps 7-16 under Proving with Stationary Provers above. 7. Block in and bleed pressure from the prover. 8. Disconnect the prover and reset the counter. 9. Stow all equipment and bleed pressure from the prover. 10. Clean the area. 11. Complete the Meter Proving Report on your laptop. 12. Compare the new meter factor with the previous one. If the new meter factor varies from the previous one by more than 0.0015 for ACTs or more than 0.0025 for LACTs, identify the problem (see Troubleshooting below).
Repeatability Requirements
The results of a meter proving must meet specic criteria that demonstrate repeatability. These requirements are as follows: Calculate a meter factor using the average number of pulses from a series of 5 consecutive round trips. Within the series, the total pulse counts must not vary by more than 0.05% between the highest and lowest reading.
Proving a Meter
September 2002
14.8
BP Pipelines
If you cannot meet the criteria for repeatability in 10 round trips, stop and determine the problem before continuing (see Troubleshooting below). For ACT locations, the new meter factor must be within 0.0015 of the last meter factor for this product. For LACT meters in crude oil service, the tolerance is 0.0025. The previous meter factor is shown on the Meter Proving Report. The meter factors should be trended, using the utility program in SMART, to monitor meter performance and condition.
See also Chapter 15, Design of Prover Systems for design considerations that affect repeatability.
Troubleshooting
When results from a meter proving are unacceptable or when there is other evidence that the meter or the proving equipment is malfunctioning, you must determine the cause of the problem. This section describes some of the most common difculties that you may encounter in the proving process and suggests ways of correcting them. Proving problems tend to fall into two broad categories: 1. The pulse counts vary by more than 0.05% between the highest and lowest reading in 5 consecutive runs (nonrepeatability). 2. The new meter factor varies from the previous meter factor by more than 0.0025 for LACTs or more than 0.0015 for ACTs. (These tolerances are valid only if the meter was proved under the same operating conditions.)
Nonrepeatability
Nonrepeatability can have the following causes: Air or gas in the prover Unstable temperature Unstable ow rates Bad detector switches A nearby electrical eld, two-way radios, or radio repeater stations, which can produce stray pulses A defective prover sphere A leaking four-way valve (bidirectional prover) A leaking diverter valve (unidirectional prover)
The procedures for these checks are given below in Procedures for Troubleshooting.
Proving a Meter
September 2002
14.9
BP Pipelines
Use good judgment in evaluating widely divergent meter factors at a particular location. Repairing or replacing a meter is expensive, so evaluate the meter thoroughly and perform extensive troubleshooting procedures to prevent unnecessary costs. Table 14.1 summarizes some of the most common causes of a uctuating meter factor.
Proving a Meter
September 2002
14.10
BP Pipelines
2. Prover displacers Inspect prover displacers regularly for over- or underination. A deated sphere causes liquid to leak past the sphere, and the meter factor drops. Overinated spheres may cause the sphere to jump, which will produce uneven meter pulses. The sphere should be 2-3% oversized. Overination causes increased wear and distortion of the sphere as well as jumping of the sphere, which produces erratic pulse counts. Underination causes high pulse counts and results in a meter factor that is too low.
3. Four-way diverter valve Check the four-way diverter valve prior to proving and then at least once during proving. If the valve is sealing properly, the gauge should remain at a pressure signicantly lower than the prover pressure. Any rise in pressure indicates that the valve is leaking. Leakage through the four-way valve causes an increase in pulse counts and thus a lower meter factor.
4. Detector switches If you remove either switch on a prover, contact the Measurement Team, as the prover must be recalibrated (waterdrawn) before using it again. Consult the manufacturers instructions before attempting any adjustments or repairs to the switches.
5. Prover isolation valve (block-and-bleed valve) During proving operations, check all connecting valves for leaks. Make sure the total ow from the meter (and only that ow) is owing through the prover while you are proving the meter. Any leakage causes higher pulse counts, resulting in a lower meter factor.
Proving a Meter
September 2002
14.11
BP Pipelines
6. Meter back-pressure API recommends that the back-pressure be no less than 2 times the pressure drop across the meter at the maximum operating ow rate plus 1.25 times the equilibrium vapor pressure at measurement temperature. For crude oil, this means that the system should be operated above 20 psi. Improper back-pressure causes the liquid to separate or ash, producing cavitation in the meter. The meter will become very erratic under these conditions.
7. Check for air or vapor in the system. 8. Check pressure, temperature, and density-sensing devices for error. 9. Check all electrical equipment for any failures, including the pulse generator, counters, coil, preampliers, signal transmission system, power supply, and all readout devices. 10. Check for leakage from all isolation and diversion valves. 11. Check for proper gearing in the meter accessory drive.
Proving a Meter
September 2002
14.12
BP Pipelines
3. For bidirectional provers, verify that the four-way valve is sealing properly by checking the differential pressure or by opening the body bleeder. 4. For unidirectional provers, verify that the interchange is sealing properly by checking the differential pressure or opening the cavity bleeder. Every six months: 1. Pull the prover sphere and inspect it for damage. Replace it with a new sphere if damage is extensive. 2. Verify that the size is still the same as during the last prover calibration. Measure the sphere around two axes (the seam and ll ports). The two readings should agree within 1%. See Table 16.2 for the acceptable tolerance for various sizes of spheres. If the sphere needs to be resized, use a 50-50 mix of water and antifreeze.
3. Reinstall provers with adjustable sphere ramps according to the manufacturers recommendations to prevent sphere damage.
Proving a Meter
September 2002
14.13
BP Pipelines
More About It
Here you will nd the following gure: Figure 14.1. Sample proving report
Before you prove a meter for the rst time, examine a sample Meter Proving Report to familiarize yourself with the information that SMART provides on the form and the kinds of proving data that you are expected to provide.
Proving a Meter
September 2002
14.14
BP Pipelines
Proving a Meter
September 2002
14.15
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 4 Proving Systems 2. BP Self-Study Guide, Field Specialist IV Module 4 Meters and Meter Proving 3. BP Self-Study Guide, Field Specialist III Module 1 Meter Prover Calibrations 4. Pipelines (NA) Business Unit Safety Manual
Proving a Meter
September 2002
14.16
Chapter 15
BP Pipelines
September 2002
BP Pipelines
Introduction
This chapter outlines the design of systems using bidirectional and unidirectional provers. Using the correct size and type of prover for the operating conditions, uid characteristics, and repeatability requirements is crucial to accurate meter proving.
September 2002
15.2
BP Pipelines
As prover diameter increases, prover pressure drops. All provers should have a baked-on phenolic internal coating. Resolution of meter pulse generator Resolution of detectors Resolution of prover counter
Fluid characteristics
Lubricity
Operation
Installation
Locate prover as close to meter as practical. Space limitations may affect prover diameter, distance between detectors, and prerun length. Ease in calibration Minimum volume of 50 gallons between detectors. Larger prover volumes should be rounded up to the next 25-gallon increment, for example, 75 gal., 100 gal., 125 gal., etc.*
Size
The minimum volume and rounding requirements do not apply to small-volume ballistic provers.
September 2002
15.3
BP Pipelines
Repeatability Considerations
Prover repeatability (see Chapter 14, Proving a Meter) relies on these design considerations: Resolution of the meter pulse generator how many pulses per unit volume are generated by the meter being proven? Resolution of the detectors within what distance can the detectors detect the presence of the prover sphere? Resolution of the prover counter electronic counters can indicate pulse intervals to within 1 interval.
September 2002
15.4
BP Pipelines
Prover Sizing
The prover must be sized so that the following considerations are met. Prover system repeatability must be 0.02% (see Chapter 16, Waterdraw Calibration). The meter must produce at least 10,000 pulses per proving pass. Recommended displacer velocity is a maximum of 5 feet per second (fps) for bidirectional pipe provers, a maximum of 10 fps for unidirectional pipe provers, a minimum of 0.5 fps for nonlubricating uids.
Volume between detector switches is at least 50 gallons (in nominal 25-gallon increments; that is, the volume should be 50 gallons, 75 gallons, 100 gallons, etc.). Fluid velocity through the four-way valve is a maximum of 15 fps for bidirectional pipe provers.
Prover Diameter
The diameter of the prover for a given ow rate is also important. As the prover diameter increases, the displacer velocity drops, the distance between detectors decreases, the prerun length decreases, and the size and cost of the prover increase.
d =
where d = the estimated prover diameter in inches F = ow rate in barrels per hour v = displacer velocity in feet per second
0.3F ---------v
September 2002
15.5
BP Pipelines
Approximate Detector Error* 0.0366 0.0392 0.059 0.0792 0.12 0.168 0.2208 0.309
Verify approximate detector error and all prover calculations with the manufacturer.
September 2002
15.6
BP Pipelines
Typical Actuator Time* (seconds) 2.15 2.6 3.55 5.0 5.7 5.25 or 9.45 8.5
Verify four-way valve closing times and all prover calculations with the manufacturer.
For unidirectional provers, C = sphere interchange seal time, in sec + 0.5 sec (safety factor) The time for a sphere interchange to seal depends on the type and manufacturer of the interchange.
September 2002
15.7
BP Pipelines
Sphere Type
The sphere material used depends on the type of liquid in the pipeline. The numbers for the polyurethane spheres in Table 15.4 indicate the hardness of the material, with higher numbers being harder.
-20 to 170F (in oil) -20 to 140F (in water) -20 to 170F (in oil) -20 to 140F (in water)
-20 to 225F
September 2002
15.8
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 4 Proving Systems 2. BP Self-Study Book, Field Specialist IV, Module 4 Meters and Meter Proving 3. Pipelines (NA) Business Unit Safety Manual
September 2002
15.9
Chapter 16
Waterdraw Calibration
Quick Reference
Safety
Follow all applicable safety rules in the Pipelines (NA) Business Unit Safety Manual.
Scope
These procedures for calibrating provers apply to conventional unidirectional and bidirectional provers, as well as small-volume provers.
Safety Reminder
The proper safe handling and disposal of the water used in prover calibrations is the responsibility of the local eld personnel.
BP Pipelines
September 2002
BP Pipelines
Introduction
Calibrating, or verifying, a provers volume at standard conditions ensures that it is measuring accurately. BP Pipelines uses the waterdraw method to calibrate pipe provers and small-volume provers (see BP Self-Study Book Meter Prover Calibrations for more information about the waterdraw method). BP Pipelines generally contracts with outside vendors to perform the waterdraw calibrations. BP Pipelines personnel prepare the prover for the waterdraw, witness the procedure, and must make sure that the calibrations are done properly and that the contractor meets high standards of quality and safety.
Waterdraw Calibration
September 2002
16.2
BP Pipelines
If the prover does not meet this requirement, check it to determine why it does not meet calibration standards. It may be necessary to repair the prover equipment or revise the conditions under which the prover is calibrated. After it is repaired or after the conditions are changed, follow the policies for a newly installed stationary prover. If either detector switch is removed for repair or replacement, the prover must be recalibrated before being placed back into service since this may change how the switches operate and thus affect the calibrated volume if the prover. This includes adjusting the reed switch inside the detector switch. Contact the Pipeline Measurement Team to schedule a recalibration. If it is impractical to perform the calibration at that time (for example, midwinter in cold climates), the Measurement Team may allow the prover to be placed back into service until a calibration can be scheduled. However, you must recalculate every meter ticket generated from the time the prover is repaired until the recalibration and generate correction tickets. Clean the prover and ll it with clean, deaerated water.
Reminder
If either the detector switch or one of its components is removed from the prover for repair, replacement, or adjustment, the prover must be recalibrated.
Waterdraw Calibration
September 2002
16.3
BP Pipelines
3. Make sure the detector switches are in perfect condition and are operating properly. The sphere must move freely and trip each detector as it passes so that the volume measured on each run is accurate. Remove and clean all mechanical switches and plunger parts. Lubricate them freely and work them several times to assure freedom of movement. Reinstall and adjust each part, making and breaking switch connections with an ohmmeter. Make all repairs and adjustment according to the manufacturers recommendations.
Waterdraw Calibration
September 2002
16.4
BP Pipelines
4. The water inlet and outlet of the prover should have 2-inch NPT threaded connections with a male or female end. 5. Isolate the prover from the pipeline. Insert isolation blinds or blind anges as close to the prover as possible. Double block-and-bleed-type valves do not need to be blinded if you can verify their sealing integrity before the calibration.
6. Clean the inside of the prover. For a crude oil prover, ush with gasoline, solvent, or any agent that is soluble with crude. DO NOT USE SOAP TO CLEAN THE PROVER! Immediately following, ush several times with water (saltwater can be used, but fresh water is preferred). Flush the prover until no oil or lm appears. During the ushing process, shift the four-way valve several times as oil may collect here and cause problems during the calibration procedure. Vent the four-way valve cavity by removing the pressure gauge if necessary. Vent all lines, if possible, to clear other points where oil or product may be trapped. Allow the prover to sit overnight with water in it. Flush again the next day until oil scum no longer appears. During this ushing, be sure to shift the four-way valve several times to run the sphere around the prover loop from side to sideleft to right and right to leftto clean all foreign material from the interior of the prover. Steam cleaning a prover is a good method of ushing if steam is available. For provers with nished products, ushing with water is usually adequate as they are normally clean.
7. Check the water. Use only clean, deaerated, potable water to calibrate the prover. The prover may be cleaned and ushed with saltwater if fresh water is not available, but the nal rinse should be with fresh water. The amount of water needed for ushing and calibrating will vary according to the size of the prover. For example, 1,500 gallons are needed for an eight-barrel, bidirectional prover after it has been cleaned1,000 gallons for the reservoir in the waterdraw trailer and 500 gallons for the prover.
Waterdraw Calibration
September 2002
16.5
BP Pipelines
Other Preparations
In addition to these steps, there may be procedures specic to certain brands of provers. For all types of provers, the waterdraw contractor will do the following: 1. The contractor levels the trailer containing the water reservoir and test measures and attaches it to a power supply. 2. The contractor wires the provers detector switches to an alarm in the trailer to indicate when the prover sphere rst makes contact and when it moves past each detector switch. 3. The contractor connects the prover to the trailer with inlet and outlet hoses. A pump on the trailer withdraws water from the trailers reservoir and circulates it through the connecting hoses through the prover and test measures and back to the reservoir once all preparations for proving have been completed. 4. The contractor will verify the size of the prover sphere.
Waterdraw Calibration
September 2002
16.6
BP Pipelines
Control panel
PI
Gauge glass
Test Measure #5
Test meas. #4 Test measure #3
Test measure #2
Test measure #1
Centrifugal pump
Waterdraw Calibration
September 2002
16.7
BP Pipelines
4. To start the calibration, the contractor cycles the four-way diverter valve on the prover and places the sphere in a launching position. 5. Next, the contractor begins drawing water through a bypass valve to bypass the test measures and deliver water directly into the reservoir. 6. The water continues to ow into the reservoir tank until the sphere makes contact with the rst detector switch. At this point, the contractor stops the ow of water and reverses the four-way valve. The contractor will withdraw water to bring the sphere past the switch to a point where the switch is de-energized (the sphere has moved just off the switch). 7. The contractor then reverses the four-way valve and moves the sphere toward the switch by withdrawing water through one of the solenoid valves. As the water moves through the solenoid valve, the contractor reads and records the system pressure indicated on the pressure gauge on the waterdraw manifold. 8. When the sphere contacts the detector switch, the solenoid valve closes, which stops all ow through the prover and the test measures. At this point, the contractor closes the block valve downstream of the solenoid valve to prevent additional ow through the valve.
Waterdraw Calibration
September 2002
16.8
BP Pipelines
9. The contractor then opens the main valve to allow water to ow into the rst test measure. When the sphere moves off the switch, the solenoid valve reopens. During the withdrawal of the rst one-third of the volume of the prover, the contractor reads and records the temperature at the outlet of the prover. 10. The contractor continues withdrawing water until the sphere is approximately 1 gallon short of the second detector switch. The contractor determines how close the sphere is to the switch by reading how much water has been withdrawn from the prover and delivered into the test measures. 11. At this point, the contractor will stop drawing water through the main outlet manifold valve and will continue lling the nal test measure by drawing water through the solenoid valve line. 12. During this rst run, the contractor will time the ll rate of one of the test measures. The contractor will determine this ll rate in gallons per minute (gpm) by taking the volume in the test measure and dividing that number by the time required to ll it. The contractor reads and records the gpm on the calibration worksheet. The contractor determines this rate to use as a point of reference. During subsequent runs, the contractor will decrease or increase the rate by at least 25% to verify that there is no leakage around the sphere. 13. The contractor also reads and records the average prover pressure during the waterdraw through the solenoid valve so that, if necessary, a correction for pressure on steel can be calculated. 14. When the sphere energizes the second switch, the run is over. The contractor has completed a one-way trip through the prover. This one-way trip may be left-to-rightor right-to-left, depending on how the contractor launched the sphere at the start of the run. 15. In a bidirectional prover, the contractor reverses the four-way valve and repeats the above operation from step 5 to step 14 in the opposite direction to complete a round trip, or run. In a unidirectional pipe prover, the run is only a one-way trip. 16. After each one-way trip, the water levels in each test measure must be read and recorded on the calibration worksheet (see Figure 16.2 under More About It). 17. The contractor reads the bottom of the meniscus on the scale on the test measures gauge glass. The contractor then drains the test measures to prepare for the next trip. While the water is draining, the contractor determines the temperature of the water draining from each test measure with a certied thermometer and records it. 18. After each run, the ll rate (gpm) should be changed (faster or slower than the previous one) by at least 25% (BP Pipelines prefers a 50% change if that is possible). Normally, the second run is slower than the rst run. The third run is then set at the same rate as the rst run, and so on. The preferred sequence of runs is fast, slow, fast. At the discretion and concurrence of all parties, the contractor may alter the
Waterdraw Calibration
September 2002
16.9
BP Pipelines
preferred order in which the ow rate is changed. However, the ow rate for at least one trip for a unidirectional prover and at least one round trip for a bidirectional prover must be changed by at least 25%. Note: At least one of the runs is a slow run to assure that there is no leakage around the sphere. 19. With a bidirectional prover, the contractor will continue making runs until 3 consecutive round trip volumes, corrected for temperature and pressure, agree within 0.020%. The corrected volume for consecutive passes in the same direction making up these round trips must also agree within 0.020%. The calibration passes in opposite directions do not necessarily have to agree. 20. With a unidirectional prover, the 3 consecutive one-way trips must agree within 0.020%. 21. The average of the 3 round trip corrected volumes is considered the base volume of a bidirectional prover at standard conditions. The average of the 3 consecutive one-way trips is the base volume of a unidirectional prover.
Waterdraw Calibration
September 2002
16.10
BP Pipelines
Calibration Calculations
Almost all calibration contractors prefer to use computers to calculate waterdraw calibrations. The computer calculations can be accepted, provided the Pipeline Measurement Team veries the computation software with periodic manual calculation checks. BP Pipelines employees who witness a calibration must sign and verify a worksheet that shows the calibration data (test measure readings, prover and test measure temperatures, pressures, etc.). This worksheet (see Figure 16.2 under More About It) must be completed in ink. Keep a copy of this sheet in the le along with the calculation data.
Waterdraw Calibration
September 2002
16.11
BP Pipelines
More About It
Here you will nd the following gures and tables: Figure 16.2. Waterdraw calibration worksheet Table 16.1. Type of Sphere to Use for Various Liquids Table 16.2. Sphere Roundness Verication Table 16.3. Minimum Ination Percentages for Given Pipe Sizes
Waterdraw Calibration
September 2002
16.12
BP Pipelines
Waterdraw Calibration
September 2002
16.13
BP Pipelines
Sphere Type
The sphere material used depends on the type of liquid in the pipeline. The numbers for the polyurethane spheres indicate the hardness of the material, with higher numbers being harder.
-20 to 170F (in oil) -20 to 140F (in water) -20 to 170F (in oil) -20 to 140F (in water)
-20 to 225F
Waterdraw Calibration
September 2002
16.14
BP Pipelines
Sphere Roundness
Verify the roundness of the sphere by determining its circumference around two separate axes perpendicular to each other. A circumference variation in the sphere (that is, the difference in length around these two perpendicular axes) of more than 1% of the nominal circumference is considered out-of-round. Measuring the sphere, rst around its equator and then around its polar axis, usually across the two valve holes, and comparing the difference between the two measurements according to Table 16.2 will verify sphere roundness.
Example
According to the table below, a 6-inch sphere with a nominal circumference of 19 inches would be expected to have a circumference variation around two axes no greater than 3/16 inch to be in round. Alternatively, a 30-inch sphere with a nominal circumference of 94 inches, and with a circumference variation around two axes greater than 15/16 inch, would be considered out-of-round.
Waterdraw Calibration
September 2002
16.15
BP Pipelines
Sphere Sizing
The sphere is oversized by inating it during calibration to compensate for piping irregularities, such as an oval pipe cross section, coating problems, or mandrel marks in the 90 bends and in the 180 returns. However, too much oversizing may cause the sphere to chatter. Undersized or oversized spheres may cause leakage past the sphere or erratic detector switch activation. If the prover is constructed with openings in the calibrated section (such as vents, drains, detector holes, or ange separations, etc.) that are wider than the surface sealing width of the sphere, a momentary leak path can exist. For example, an 8-inch standard-wall pipe prover with a 1/2-inch vent opening will require a prover sphere oversized by approximately 4% (0.66 inch in this example) to span the opening width by a comfortable margin. Table 16.3 shows the minimum additional sealing width (sphere ination) required to span openings in the measuring section. Other criteria for sizing the sphere must also be considered, such as pipe openings.
Waterdraw Calibration
September 2002
16.16
BP Pipelines
6 8 10 12 14 16 18 20 22 24 26 28 30 36 42 48
6.065 7.981 10.02 12 13.25 15.25 17.25 19.25 21.25 23.25 25.25 27.25 29.25 35.00 41.00 47.00
2% 2% 2% 2% 2% 2% 2% 2% 2% 2% 3% 3% 3% 3% 3% 3%
0.25 0.33 0.41 0.49 0.54 0.62 0.70 0.79 0.87 0.95 1.56 1.68 1.81 2.16 2.53 2.91
0.31 0.41 0.51 0.74 0.82 0.94 1.07 1.19 1.31 1.44 2.10 2.27 2.43 2.91 3.41 3.91
3% 3% 3% 4% 4% 4% 4% 4% 4% 4% 5% 5% 5% 5% 5% 5%
0.37 0.49 0.62 1.00 1.10 1.27 1.44 1.60 1.77 1.94 2.65 2.86 3.07 3.68 4.31 4.94
4% 4% 4% 5% 5% 5% 5% 5% 5% 5% 6% 6% 6% 65 6% 6%
0.50 0.66 0.83 1.26 1.39 1.60 1.81 2.02 2.23 2.44 3.22 3.47 3.72 4.46 5.22 5.99
5% 5% 5% 6% 6% 6% 6% 6% 6% 6% 7% 7% 7% 7% 7% 7%
0.64 0.84 1.05 1.53 1.69 1.94 2.20 2.45 2.71 2.96 3.79 4.09 4.39 5.25 6.15 7.05
Waterdraw Calibration
September 2002
16.17
BP Pipelines
Reference Documents
1. API Manual of Petroleum Measurement Standards, Chapter 4.2 Conventional Pipe Provers 2. API Manual of Petroleum Measurement Standards, Chapter 4.3 Small-Volume Provers 3. API Manual of Petroleum Measurement Standards, Chapter 4.9, Part 1 Determination of the Volume of Displacement and Tank Provers by the Waterdraw Method of Calibration 4. BP Self-Study Guide, Field Specialist III Module 1 Meter Prover Calibrations 5. Pipelines (NA) Business Unit Safety Manual
Waterdraw Calibration
September 2002
16.18
Chapter 17
Scope
This chapter tells you when and how to make claims and adjustments and how to use SMART (software you will use to create measurement tickets and reports). This chapter applies to custody transfer of all types of petroleum liquids in tanks and pipelines.
Procedure for making claims or adjustments: Provide full details of the claim and recommendation for fair settlement on a claim memorandum. The lease producer must provide a notarized afdavit when ling a claim. After investigating and verifying the claim, retain all original paperwork in the eld ofce.
BP Pipelines
September 2002
BP Pipelines
September 2002
17.2
BP Pipelines
Introduction
An adjustment is a correction to a measured volume in a custody transfer. An adjustment may be necessary due to a security breach, equipment failure, or human error. A claim is a demand for compensation in the event of such a problem. Any involved party may make a claim and the parties may agree to make an adjustment to settle the claim. Report writing and record keeping are important aspects of your job. Collecting accurate data and entering the data correctly will help ensure measurement integrity and aid in the settlement of claims or making of adjustments. Reports and records also provide a valuable source of information about the specications and performance history of pipeline equipment. BP Pipelines SMART software generates and stores most of the reports and tickets that the company has developed to meet industry requirements for reporting and record keeping. In addition, the SMART system is a repository of meter and tank records that you can consult as a reference.
September 2002
17.3
BP Pipelines
LACT or ACT measurement equipment malfunctions (meter, S&W monitor, diverter valve, etc.)
When a lease producer les a claim, a notarized afdavit is required to satisfy royalty and various government agency regulations. An afdavit is not normally required from other pipeline companies or reneries ling claims.
September 2002
17.4
BP Pipelines
Claims
Procedures for ling a claim: 1. Fully investigate all aspects of the claim in cooperation with the producer or other involved parties. 2. Record the details of your investigation. 3. Obtain a notarized afdavit from the producer, if required (see Table 17.1 above). 4. Submit the memorandum, afdavit, and run ticket to your supervisor for review. Keep the originals of these documents in the local eld ofce. A claim must include the following information: Full details of the claim, including but not limited to the gross barrels, temperature, S&W, API gravity, meter factor, and tank number. Include the API gravity and agreed net barrels when the exact temperature and S&W are not available.
Figure 17.1 under More About It at the end of this chapter shows a sample claim.
Adjustments
A notarized afdavit from the producer to BP Pipelines must include the following: Full details of the claim, including but not limited to the gross barrels, temperature, S&W, API gravity, meter factor, and tank number. The API gravity and agreed net barrels when the exact temperature and S&W are not available must be included. Send copies of all documentation supporting claims to the Manager of Oil Movements for reconciliation in SMART. Figure 17.2 under More About It at the end of this chapter shows a sample afdavit.
September 2002
17.5
BP Pipelines
Procedures for Using the SMART Measurement Software for Ticketing and Reports
You will use the SMART measurement software for these four tasks: Downloading the latest data about the locations you are responsible for Updating or creating new tank or meter records Creating tickets and reports Determining meter factor trends
See the BP Self-Study Book Ticket Writing (SMART) for more detailed instructions than are provided in this manual.
Corrections to Tickets
To correct the information on a custody transfer ticket, open the original ticket and make the change to the necessary elds. After making these changes, SMART will prompt you to give a reason for the change. This explanation will then be printed on the ticket. To correct a ticket for the current month or the past two months, use the SMART Mobile Module. To correct a ticket from an earlier cycle, you must use the Operations Module.
September 2002
17.6
BP Pipelines
When you print out copies of the document, make at least two copies one for the eld ofce and one for the witness. Some eld specialists also like to make a third copy for their own records. Obtain the signature of the witness on the paper copies. Turn in a signed paper copy and all supporting documents (such as records of calibration checks) to the eld ofce.
The electronic version of all tickets and reports is stored in SMARTs Operations Module for the life of the system. A paper copy of each ticket and report must stay on le at the designated eld ofce for at least 5 years. Supporting documents must be led with the tickets and also kept for at least 5 years.
September 2002
17.7
BP Pipelines
More About It
Here you will nd the following gures: Figure 17.1. Example of a claim letter Figure 17.2. Example of a notarized afdavit Figure 17.3. Example of a SMART tank custody transfer ticket Figure 17.4. Example of a SMART meter ticket for a lease meter
September 2002
17.8
BP Pipelines
Manager Revenue Accounting BP Pipelines Supply & Logistics P.O. Box 00000 Houston, TX 77000 Dear Sir:
Enclosed is a letter from Jack Producer, District Manager of XYZ Company at Lovington, NM regarding unregistered oil delivered to BP Pipelines from 7:00 a.m. September 19, 2001 to 9:00 p.m. September 20, 2001. The meter involved was No. 95545, LACT No. 4 at Painter Reservoir. The malfunction was caused by a set screw, located above the meter failure transmission switch, coming loose and letting a gear fall off, which caused the meter counter to stop turning. The gures presented to us by Mr. Producer have been checked with the Field Specialist in Lovington and we are in agreement that the 10,566 net barrels is a fair adjustment. Please issue a ticket for this amount to the XYZ Company account for September.
September 2002
17.9
BP Pipelines
Volume Adjustment, Painter Reservoir Lease #xxxx, Meter Serial #xxxxx On September 19, 2001, XYZ Company experienced a gear failure above the meter failure transmission switch, resulting in 500 barrels of merchantable oil being shipped that were not recorded on LACT #4. The volume of oil run through the meter but not registered by the LACT unit was veried by use of BP Pipelines telemetry totalizer. Calculations used to determine the unmetered oil loss were as follows: BP Pipelines meter reading at 7:00 a.m. 19 September 2001 = 236,059 BP Pipelines meter reading at 9:00 p.m. 20 September 2001 = 254,124 Difference = 18,605 Oil metered by the LACT unit from 5:00 a.m. 19 September 2001 to 9:00 p.m. 20 September 2001 was 8,387 gross bbl. Rate at which LACT meter was proved = 496 bph 496 bph 2 hours = 992 bbl shipped from 5:00 a.m. to 7:00 a.m. 19 September 2001 8,397 992 = 7,392 bbl metered on the LACT unit from 7:00 a.m. 19 September 2001 to 9:00 p.m. 20 September 2001 BP Pipelines metered gross bbl = 18,065 LACT unit metered bbl = 7,395 Unmetered gross bbl = 10,670 Meter factor established 15 September 2001 = 1.0031 Meter factor 1.0031 10,670 = 10,703.077 corrected bbl The API gravity @ 60F was 50.3 at last proving on 15 September 2001. The average temperature of this oil was 82F, based on the average temperature of crude shipped through other LACTs at this facility, resulting in a temperature correction factor of 0.9872. 0.9872 10,705.077 = 10,566 bbl corrected volume
September 2002
17.10
BP Pipelines
The cause of the aforementioned unregistered run was due to a malfunction in the lease automatic custody transfer meter. The stated quantity of oil run was calculated as accurately as possible with the information available. XYZ Company requests credit for the net volume of 10,566 barrels, gured at API gravity @ 60F of 50.3.
STATE OF NEW MEXICO COUNTY OF LEA On this 28th day of Sept. 2001, before me the undersigned Notary Public in and for said County and State, personally appeared Jim Honcho whose name is subscribed to the foregoing instrument and acknowledged that he executed the same as a free and voluntary act and deed for the purposes and considerations therein mentioned and set forth. Witness my hand and ofcial seal on the 28th day of Sept. 2001. My commission expires: June 10, 2002 Date ______________________ John Smith, Notary Public
September 2002
17.11
BP Pipelines
September 2002
17.12
BP Pipelines
September 2002
17.13
BP Pipelines
Reference Documents
1. BP Self-Study Book, Field Specialist IV, Module 4 Ticket Writing (SMART) 2. Pipelines (NA) Business Unit Safety Manual
September 2002
17.14
BP Pipelines
Abbreviations
ACTautomatic custody transfer ANSIAmerican National Standards Institute APIAmerican Petroleum Institute ASTMAmerican Society for Testing and Materials ATCautomatic temperature compensator BPVbase prover volume bblbarrel(s) bphbarrels per hour CPLcorrection factor for the effects of pressure on a liquid CPScorrection factor for the effects of pressure on steel CPUcentral processing unit CTLcorrection factor for the effects of temperature on a liquid. CTScorrection factor for the effects of temperature on steel. ELM electronic liquid measurement GOVgross observed volume GSVgross standard volume IDLHimmediate danger to life and health IPInstitute of Petroleum. ISOInternational Organization of Standardization. LACTlease automatic custody transfer. MFmeter factor MPMSManual of Petroleum Measurement Standards NISTNational Institute for Standards and Technology
Glossary
September 2002
G.1
BP Pipelines
NSVnet standard volume OSHAOccupational Safety and Health Administration PD meterpositive displacement meter PLC programmable logic controller psiapounds per square inch absolute psigpounds per square inch gauge psipounds per square inch RTDresistance temperature detector RVPReid vapor pressure S&Wsediment and water SCADASupervisory Control and Data Acquisitions SIInternational System of Units (metric system) TCCTulsa Control Center TOVtotal observed volume VCFvolume correction factor
Glossary
September 2002
G.2
BP Pipelines
Glossary
A
absolute pressure (psia)a measure of pressure that includes atmospheric pressure. For
example, if a pressure gauge on a tank shows a reading of 4.0 psi, the absolute pressure at that point would be 18.7 psia (4.0 psi + 14.7 psia = 18.7 psia).
ACTsee automatic custody transfer. adjustmenta correction to a measured volume in a custody transfer. An adjustment may be
necessary due to a security breach, equipment failure, or human error.
affidavita sworn statement in writing. algorithma formula or set of steps for solving a particular problem. An algorithm always has
a set of unambiguous rules and a clear stopping point.
ambient airthe air we breathe. American National Standards Institute (ANSI) a group of organizations and
agencies that provides information about and approval for American National Standards for industry, engineering, safety, design, and other applications. Many of these standards have been adopted as OSHA standards.
API gravityan arbitrary (U.S.) scale that measures the gravity or density of liquid petroleum
products (that is, weight per unit volume). API gravity is expressed in degrees. The lower the number, the denser the oil. See also relative density.
Glossary
September 2002
G.3
BP Pipelines
APIAmerican Petroleum Institute. The leading standardization organization for oileld drilling
and producing equipment.
applicationsoftware that performs a specic set of tasks. ASTMAmerican Society for Testing and Materials. An organization that sets guidelines for
testing and use of materials.
automatic custody transfer (ACT)a system for automatically sampling and measuring
crude oil at a point of receipt or delivery.
automatic line samplea sample withdrawn automatically from oil owing in a pipeline
which is representative of the total batch passed through the line. If the rate the sample is taken is proportional to the rate of ow it is called a ow-proportional sample.
automatic samplera device used to extract a representative sample from the liquid owing
in a pipe. The automatic sampler generally consists of a probe, a sample extractor, an associated controller, and a sample receiver.
B
back-pressure valvea valve that automatically regulates pressure on its inlet side to a
preset value.
barrelthe common unit of volume measurement in the oil eld. A barrel of oil is equivalent to
42 gallons of oil.
base prover volume (BPV)the calibrated volume of a provers proving section corrected
to standard conditions (0 psi and 60F). The base volume is indicated on the provers calibration certicate. Water is often used in calibrating provers because of its universal availability and its well-known properties.
batchthe total volume of oil shipped in a single custody transfer. Also called a parcel. bblbarrel(s) bidirectionalreferring to ow that may be in either direction; for example, a bidirectional
pipe prover or meter.
Glossary
September 2002
G.4
BP Pipelines
block-and-bleed valvea high-integrity valve with double seals and with a bleeder valve
for determining if either seal leaks.
bottom samplea sample obtained from the material on the bottom surface (or as close to
the bottom as possible) of the tank or container. Also called outlet sample or clearance sample.
boxcar seala numbered, metal sealing device for detecting tampering on a tank or metering
system. Compare wire seal.
BP Amoco Lease Numberthe companys internal number for a lease. BP Pipelines (North America), Inc.a common carrier pipeline, subject to the rules and
regulations of the Federal Energy Regulatory Commission or an applicable state commission, in the business of transporting liquid hydrocarbons for hire, without discrimination or preference, to the extent of available facilities, for anyone who requests such services and meets tariff requirements.
C
calibration certificatea certicate issued by a standardization laboratory expressing the
relationship between the reading indicated by an instrument and the true value. Knowing this relationship allows the instrument reading to be adjusted to give a closer approximation to the true value. The determination of the true value must be traceable to a national standard. See NIST-certied.
capacity tablesee tank calibration table. central processing unit (CPU)a device made up of one or more microprocessors and
associated components. The CPU controls certain system activities including interpreting and executing specic programs. The CPU uses arithmetic logic units (ALU), timing and control circuitry, an accumulator, a program counter, and other components to control certain processes within a closed system.
centrifugea device used to spin samples of crude oil to determine the oils suspended
sediment and water (S&W) content.
claim lettera letter from a eld representative to his or her supervisor reporting the
investigation of a claim and recommending a fair settlement.
Glossary
September 2002
G.5
BP Pipelines
clearance samplea spot sample taken 4 inches below the level of the tank outlet. Also
called bottom sample or outlet sample.
closing gaugethe measurements taken (oil height, temperature, and bottom S&W level) on
a manually gauged tank to determine the oil quantity and quality after transferring the tanks oil into the pipeline system. Also called closing a tank.
cloud pointthe temperature at which a waxy solid material appears as a diesel fuel is cooled. composite spot samplea blend of spot samples mixed in proportion to the volume of
material from which the spot samples were obtained.
connection heightthe point near the base of the tank shell where liquid exits the tank and
enters the pipeline
Coriolis metera meter that measures the mass ow of a uid and its density. The volume can
be calculated from these measurements. Previously used only for measuring petrochemicals, the industry is beginning use them for custody transfer measurement of crude oil and rened products.
CPScorrection factor for the effects of pressure on steel; used in calculations for determining
net standard volume in a tank.
CPUsee central processing unit. critical zonethe vertical range in which the level of the stored oil is high enough to lift a
tanks oating roof off the tank oor but too low to make the entire roof oat freely in a level position. Gauging measurements taken while the oil is within this zone will not be accurate because the height of the oil conforms to the slant of the oating roof.
CTLcorrection factor for the effects of temperature on a liquid; used in calculations for
determining net standard volume in a tank.
CTScorrection factor for the effects of temperature on steel; used in calculations for determining net standard volume in a tank.
Glossary
September 2002
G.6
BP Pipelines
custody transfer a change in ownership or responsibility for a quantity of crude oil or other
pipeline commodity. At a lease site, custody transfer takes place between the lease operator and the pipeline company. A custody transfer facility is responsible for measuring the commodity and accounting for charges to the buyer.
D
datum platea level metal plate attached to the tank shell or bottom. It is located directly
below the gauge hatch and is used as a xed reference point for all measurements of oil height made in the tank. The tip, or end, of the plumb bob will just make contact with the gauge plate during hand gauging procedures. Also called gauge plate or striking plate.
deadwoodany obstruction inside a tank that would decrease the tanks liquid volume. demulsifier solutiona liquid that destroys, or breaks, an unwanted emulsion; added to
samples that are being tested for suspended S&W.
densitythe mass (weight in a vacuum) of liquid per unit volume. See relative density, API
gravity.
differential pressurethe pressure drop across a device, such as an orice plate/lter, in a owing stream. The amount of this pressure drop varies with ow rate. digital signalthe representation of the magnitude of a variable in the form of discrete values or pulses of a measurable physical quantity. digitalof or pertaining to data in the form of digits, especially electronic data stored in the
form of a binary code.
displacera spherical or cylindrical object that moves inside the prover pipe and displaces a known volume of uid. dissolved waterwater that is in solution in the oil at a dened temperature and pressure. distillation analysisa method of obtaining data on the boiling point range of a crude oil. It
determines the percentages or yields of crude fractions over their typical boiling point ranges.
distillationa method used to determine the water content of oil. In this method, a sample of oil
is mixed with a solvent until it boils. The water and solvent vapors condense and continuously
Glossary
September 2002
G.7
BP Pipelines
separate with the water falling into a trap with graduated sections and the solvent returning to a ask. The amount of water can be read from the graduations on the trap.
diverter valveon a LACT unit, a valve that diverts a liquid back to the tank or storage unit
rather than shipping it downstream, due to an unacceptable level of S&W.
double chronometry draw-off levelthe level of liquid in a tank where the water draw-off valve is located. driftan observed change, usually uncontrollable, in meter performance, meter factor, etc. over
time.
E
electronic liquid measurement (ELM)a metering system using electronic calculation
equipment with API liquid measurement algorithms and security/auditing features, on-line temperature and pressure inputs, and linear meter pulse inputs. ELM provides real-time, on-line measurement.
emulsionan oil and water mixture that does not readily separate.
F
flash pointthe temperature at which a liquid will give off enough ammable vapor to ignite or
ash in the presence of a ame.
flashinga process where highly volatile components come out of solution, usually due to a
reduction in pressure.
floating roofa tank roof which oats freely on the surface of the liquid contents except at low levels, when it is partially or wholly supported by legs. This is not the same as the lightweight oating cover frequently installed in xed-roof product tanks to minimize evaporation loss. flow straightenera length of straight pipe containing straightening vanes or the equivalent,
which is installed at the inlet of a ow meter to eliminate swirl in the liquid, thereby reducing measurement errors.
flow-proportional samplea sample of liquid taken from a owing stream during the
entire custody transfer. The rate of sample collection is proportional to the ow of the liquid in the pipe over a given time period.
fluida substance that ows; liquids and gases are uids. four-way valvea ow-reversing valve used with bidirectional provers.
Glossary
September 2002
G.8
BP Pipelines
free waterthe measured volume of water present in a container of liquid hydrocarbons; free
water is not in suspension in the hydrocarbons at the observed temperature. Compare suspended water.
G
gasoline-indicating pastea paste applied to a gauge tape that changes color when it
comes into contact with gasoline.
gate valvea full-bore valve with a sliding (gate-type) closure. gathering systema pipeline, usually of small diameter, that brings oil, gas, or both from a
group of leases (i.e., wells in a eld) to a point for delivery to the main pipeline or other transport system. In SMART, a gathering system consists of facilities between a connection point and a base system.
gauge tubea pipe (usually slotted) that extends down from the gauge hatch in a oating-roof
tank to allow measurement through the roof of a oating roof tank.
gaugeto manually determine the quantity of a liquid in a tank. gauging heightthe distance from the datum plate at the tank bottom (or, with lease tanks,
from a striking point on the tank oor) to a designated reference point at the top of the tank.
gradethe type of crude oil based on its quality and composition. Grade is an important factor
in commodity pricing. The higher the sulfur content, the lower the quality of the crude. Crude oil is generally classied as sweet (containing little or no sulfur), intermediate, or sour (containing hydrogen sulde or another acid gas). Light crudes (those with relatively high API gravity and therefore low density) command a higher price than heavy crudes.
gross observed volume (GOV)a volume of oil, including dissolved water, suspended
water, and suspended sediment but excluding free water and bottom sediment, measured at the prevailing oil temperature and pressure. Compare total observed volume.
gross standard volume (GSV)a volume of oil, including dissolved water, suspended
water, and suspended sediment but excluding free water and bottom sediment, calculated at standard conditions. This may be either the volume in a tank or the difference between the volume before and after a transfer. Compare net standard volume.
Glossary
September 2002
G.9
BP Pipelines
H
hand gaugingmanually measuring the quantity and quality of the liquid contents of an
upright cylindrical storage tank.
handline gauginghand gauging. hand-run tanka tank whose oil content is transferred to a pipeline using manual
measurement methods.
hazea cloudy appearance in a transparent liquid or solid; an indication of contamination. high bottoma level of S&W that is higher than 4" below the pipeline connection. hydrocarbonsa term that describes a wide range of hydrogen and carbon molecules. On the light end, the hydrocarbon methane, a natural gas component, consists of one carbon and four hydrogen atoms. On the heavier end, some crude oils are made up of many carbon and hydrogen atoms. hydrogen sulfide (H2S)a poisonous gas which can evolve from crude oil. Higher H2S
levels are generally found in the heavy sour crude oils. Respiratory protection is required when levels in the air reach 10 ppm. The IDLH (immediate danger to life and health) level is 300 ppm.
hydrostatic pressurethe pressure that a stationary column of uid, such as oil or water,
exerts on its container in proportion to the height of the liquid. In a full tank, the hydrostatic pressure is greatest at the tank bottom and decreases uniformly to the top of the liquid level.
I
innagethe preferred method of gauging a tank. IPInstitute of Petroleum. ISOInternational Organization of Standardization. An organization that sets standards in many
businesses and technologies.
K
Karl Fischer titrationa method of determining the water content of an oil sample,
sometimes referred to as water content by titration.
K factordenotes the number of pulses emitted by a meter while a certain volume of uid ows
through it; usually expressed in pulses per barrel. See pulses per unit volume.
Glossary
September 2002
G.10
BP Pipelines
L
LACTsee lease automatic custody transfer. lease automatic custody transfer (LACT)a system for automatically sampling,
measuring, and transferring crude oil from a lease to a connected pipeline.
light endsgaseous light hydrocarbons in crude oil: ethane (C2); propane (C3); isobutane and
normal butane (C4); and isopentane and normal pentane (C5). Light ends may easily be lost by evaporation.
lower samplea spot sample obtained at the midpoint of the lower third of the tank contents
in a large tank, or just above the suction line in a lease tank.
M
mass flowthe measure of ow in mass (for example, pounds) per unit of time (for example,
hour). Coriolis meters measure the mass ow of a liquid.
massthe mass of a substance is the quantity of matter it contains. It is, therefore, independent
of external conditions such as the buoyancy of the atmosphere. In oil measurement, mass is frequently referred to as the weight-in-vacuo (weight in a vacuum).
measurement ticketan ofcial document that records any quantity of crude oil received
into or delivered out of the pipeline system. A measurement ticket is the bill of sale for this type of transaction. Also called run ticket or custody transfer ticket.
meniscusthe curved upper surface of the oil that adheres to the stem of the hydrometer due to
surface tension forces in the oil. The curvature can be either concave or convex.
mercaptansnaturally occurring sulfur compounds. metals contentheavy metals in crude oil such as arsenic (As), iron (Fe), vanadium (V), and
nickel (Ni).
meter factor (MF)a number used to correct a meters inaccuracy. The factor is derived in the
proving process by dividing the gross standard volume of oil owing through the prover by the gross standard volume registered by the meter.
meter provera vessel of known volume that is used as a volumetric reference standard for
checking the accuracy of pipeline meters.
Glossary
September 2002
G.11
BP Pipelines
meter provingthe procedure that determines the relationship between the volume of oil
indicated by the meter and the true volume being measured under the same set of conditions.
meter pulsean electronically generated signal. The number of pulses is proportional to the
volume being measured; the frequency of pulses is proportional to the ow rate.
metera device for measuring volumes, quantities, or ow rates of the liquid owing through
the meter.
middle samplea spot sample obtained from the middle of the tank contents (a point
halfway between the upper and lower sample points).
MPMSManual of Petroleum Measurement Standards; published by API. multiple tank composite samplea mixture of individual samples from several
compartments (on ships, barges, and so forth), each of which contains the same grade of petroleum material. The mixture is blended in proportion to the volume of material in each compartment. Compare single tank composite sample.
N
net standard volume (NSV)a volume of oil, excluding total water and total sediment,
calculated at standard conditions. Compare gross standard volume.
nitrogenin crude, the amount of organic nitrogen which is bound (attached) to hydrocarbon
molecules.
O
observed gravitythe API gravity of the product not temperature-corrected to 60F. off sealthe seal removed from the tank stop valve prior to opening the valve to transfer the
product into the pipeline.
Glossary
September 2002
G.12
BP Pipelines
opening gaugethe measurements taken (oil height, temperature, gravity, suspended S&W
content, and bottom S&W level) on a hand-run tank to determine the oil quantity and quality prior to transferring the tanks oil into the pipeline system. Also called opening a tank.
orifice metera meter that measures the velocity of the uid (ow rate) by creating and
recording a pressure differential in the uid and then uses that rate to calculate the volume.
outlet samplea spot sample taken at the level of the bottom of the tank outlet (either xed or
swing pipe) but not higher than three feet above the bottom of the tank. Also called bottom sample or clearance sample.
P
parcelsee batch pipelinea system of connected lengths of pipe, usually buried in the earth or laid on the
seaoor.
pour pointthe lowest temperature at which a liquid will ow when its container is inverted. powered mixera type of mixer that uses an external source of power (usually electricity) to
drive a motor and pump to mix the oil.
prerunthe piping on a bidirectional prover between the launch chamber and the rst detector
switch. The prerun length must be long enough to allow the four-way valve to be fully seated before the displacer (sphere) reaches the rst detector switch.
pressureforce per unit area measured in pounds per square inch (psi).
Glossary
September 2002
G.13
BP Pipelines
probea small device that extends into a section of pipe. The probe contains an opening
through which small discrete units of liquid are pulled and diverted to a sample storage container. See sample extractor.
production tanka tank used in the eld to receive crude oil as it comes from the well. Also
called a ow tank or lease tank.
prover isolation valvea valve in line with the meter that diverts ow through the prover
when closed.
prover passa single movement of the displacer between the detector switches. prover round tripa complete forward and reverse trip of the displacer in a bidirectional
prover.
proversee meter prover. proving runa single pass of the displacer in a unidirectional prover; a single round trip in a
bidirectional prover.
proving sectionthe calibrated area of a prover, located between the detector switches,
where actual measurement takes place. Also called the measuring section.
psiapounds per square inch absolute; that is, pressure measured in a vacuum. It equals the sum
of gauge pressure (psig) and atmospheric pressure.
psigpounds per square inch gauge; that is, pressure recorded by a gauge and measured with
respect to that of the atmosphere. It refers to the amount of pressure a uid exerts on the inside walls of a vessel containing the uid.
psipounds per square inch; a measure of pressure. pulsating flowow with periodic pressure uctuations. pulse generatoran accessory device on a meter that produces a series of repeating electrical signals (pulses). The number of pulses is proportional to the volume measured, and their frequency is proportional to the ow rate. pulse interpolationa technique for counting the whole number of pulses between two
events (such as the activation of detector switches) and then calculating any remaining fraction of a pulse.
pulses per round tripthe number of pulses or counts generated by a meter during a proving
run. With a properly designed conventional pipe prover, the meter will generate at least 10,000 pulses between detector switches.
Glossary
September 2002
G.14
BP Pipelines
pulses per unit volumethe number of pulses for a given unit (barrels, gallons) that the
meter will produce. This number, also known as the K factor, is provided by the meter manufacturer and depends on meter gearing, size, and construction.
R
reagenta substance used in a chemical reaction to detect, measure, examine, or analyze
another substance; for example, in Karl Fischer titration to determine the water content in oil samples.
reference pointa point near the gauge hatch used during hand gauging. This point is
marked, and the distance between it and the gauge plate is normally written nearby on the side of the tank.
remotea term used to refer to devices that are not part of or located near the main computer. repeatabilitythe closeness of successive measurement results for the same quantity under
the same conditions and with the same equipment and methodology.
representative samplea small portion extracted from the total parcel or batch that contains the same proportion of owing components as the total parcel being transferred. The precision of extraction must be equal to, or better than, the method used to analyze the sample. reproducibilitythe closeness of the agreement between the results of measurements of the
same quantity, where the individual measurements are made by different operators using different equipment at different locations.
roof correctionin gauging a tank, the number that corrects for the oil displaced by the
oating roof.
RTDsee resistance temperature detector. run a tankto transfer oil from an upright cylindrical lease tank to a pipeline. run ticketsee measurement ticket.
Glossary
September 2002
G.15
BP Pipelines
S
S&Wsediment and water; that portion of the crude oil that is unmerchantable. saltin crude, a contaminant resulting principally from either production practice in the oil elds
or handling in tankers transporting the crude.
sample conditioningthe process of mixing a sample to prepare it for analysis. sample containera can, bottle, or thief containing a liquid sample for testing. sample controllerthe device in an automatic sampler that governs the operation of the
extractor either in proportion to time or to the ow rate.
sample extractorin an automatic sampler, the device used to pull small discrete units of
liquid, called sample grabs, from the owing stream (through the probe). The extractor diverts these units of liquid to a nearby storage container.
sample grabthe volume of liquid extracted from the pipeline by a single actuation of the
sample extractor. The sum of all grabs results in a sample.
sample potsee sample receiver. sample provinga technique used to validate an automatic sampling system. An acceptable validation can be made by injecting a known quantity of water ahead of the sampling system, taking samples, analyzing the samples for sediment and water content, and comparing those results with the known water injection. sample receivera container on an automatic sampler in which all sample grabs are
collected. A receiver may be xed (stationary) or portable.
samplea small quantity of liquid obtained from a tank or pipeline for analysis. See
representative sample.
Glossary
September 2002
G.16
BP Pipelines
SCADASupervisory Control and Data Acquisitions. A software system that monitors and
controls pipeline or renery status and provides logging capabilities. SCADA systems are highly congurable and usually interface to a facility via PLCs.
sediment and water (S&W)materials coexisting with, yet foreign to, petroleum liquid
that require separate measurement for sales, accounting, and other reasons. These foreign materials include free water and sediment, and emulsied or suspended water and sediment.
sedimentany foreign substance of no value in the oil that will not dissolve and remains suspended, such as ne grains of rock, clay, metal, etc. shakeoutsee grindout. shell heightthe distance between the bottom of the bottom angle of the tank and the top of the
top angle.
SIInternational System of Units (metric system). sideline gaugingautomatic gauging of a tank. single tank composite samplea blend of the upper, middle, and lower samples taken
from a single tank. For a tank of uniform cross sections, such as upright cylindrical tanks, the blend consists of an equal part of the three samples. For other types, the blend may require different proportions from each sample. Compare multiple tank composite sample.
sludge solventa solvent, like Stoddard solvent, used to remove sludge from sample
containers, bottles, thiefs, test tubes, or other vessels.
sludgea mixture of sediment, dense hydrocarbons and water which can settle out on the
bottom of a tank or vessel containing a sample.
SMARTBP Pipelines software for creating measurement tickets and reports. spot samplea sample taken at a specic location in a tank or from a pipe at a specic time
during a pumping operation.
Glossary
September 2002
G.17
BP Pipelines
static mixera mixer with no moving parts. The kinetic energy of the moving uid provides
the power required for mixing (e.g., xed objects in a conduit that uses the kinetic energy of the moving uid to mix the hydrocarbons). Compare powered mixer.
Stoddard solventa solvent used to clean sludge from sample containers, bottles, thiefs, test
tubes, or other vessels; may also used to dilute samples in S&W testing.
storage tanka tank in which a hydrocarbon is stored pending transfer by pipeline, truck, or
other means to a point where it is ultimately sold.
strainera device placed upstream of a meter to remove from the stream foreign material which
is likely to damage the meter or interfere with its operation. The strainer element is generally coarser than that of a lter designed to remove solid contaminants.
strappingsee tank strapping. stream conditioningthe mixing of the pipeline contents, upstream of the sampling
location, which is necessary for delivery of a representative sample.
subsamplea sample of a sample. sulfur in crude oil, a contaminant primarily present as organic sulfur (attached to hydrocarbon
molecules) or as hydrogen sulde, mercaptans (see mercaptans), and inorganic salts (sulfates).
suspended watersmall droplets of water that do not quickly separate from the oil due to
the forces of gravity. Compare free water.
T
tank calibration tablea table that tells the quantity of liquid contained in a tank at any
given level. Also called tank table.
tank stop valvea valve that controls the ow of oil into or out of a tank. The term
specically applies to valves connected to lease tanks.
tank tablesee tank calibration table. tapa pipe welded to the shell of a tank that serves as a conduit or port to the inside of the tank.
Also known as a nozzle.
Glossary
September 2002
G.18
BP Pipelines
tape cutthe level at which the oil level in the tank crosses the gauge tape when the plumb bob
is properly resting on the datum plate.
test measurein a waterdraw calibration unit, a vessel of known volume calibrated by NIST. thermowella metal protective pocket installed in the pipe wall into which the sensing element
(bulb) of a thermometer is inserted.
titrationa process of chemical analysis by which drops of a standard solution (Karl Fischer
reagent, in the testing for water in oil by titration) are added to another solution (the oil sample) to obtain a desired response.
top samplesee upper sample. total observed volume (TOV)a volume of oil, including total water and total sediment,
measured at the prevailing oil temperature and pressure. Compare gross observed volume.
total waterthe sum of the dissolved, entrained, and free water in the cargo or batch of oil. traceabilitythe relation of a prover or a transducer calibration, through a step-by-step
process, to the fundamental standards of mass, length and time, certied and maintained by national or international standardization laboratories. See also NIST-traceable.
transducera device that is actuated by power from one system and supplies a different form
of power to another system For example, the transducers in an ultrasonic meter receive electrical power, which causes them to vibrate (mechanical power), and in turn they convert received sound waves back into an electrical signal.
trendto chart meter factors from one meter over time to monitor the meters performance and condition. SMART includes a utility program for this purpose. turbine metera ow meter with a bladed rotor that turns at a speed approximately
proportional to the mean velocity of the owing liquid and therefore to the volume rate of ow.
Glossary
September 2002
G.19
BP Pipelines
U
ultrasonic meteran inferential meter that derives the velocity of the uid (ow rate) by
measuring the transit times of high-frequency sound pulses; the ow rate is then used to calculate volume.
upper samplea spot sample taken at the midpoint of the upper third of the tank contents in a large tank, or just below the surface of the liquid in a lease tank.
V
vapor pressurethe pressure in an enclosed space that is due to the vapor of the substance
occupying that space. Since liquid evaporates more quickly as temperature rises, vapor pressure increases as the temperature of the liquid increases. Thus, vapor pressure is a measure of the tendency of the substance to evaporate.
Varsolthe brand name for a solvent that may be used in the centrifuge test, also referred to as
Stoddard solvent.
volatilitythe tendency of a liquid to vaporize. volume correction factor (VCF)the product of the temperature (CTL) and pressure
(CPL) factors for a liquid; used in calculations for determining net standard volume in a tank.
W
water cut of dip(1) the depth of free water lying on the bottom of a tank.; (2) the line of demarcation of the oil/water interface. waterdraw calibrationa method of certifying the volume of a prover. water-indicating pastea paste applied to a gauge tape that changes color when it comes
into contact with water.
Glossary
September 2002
G.20
BP Pipelines
wire seala metal sealing device used to replace a boxcar seal that is defective or shows signs
of tampering. Compare boxcar seal.
witnessa person who observes the measurements of a meter or tank custody transfer or meter
proving and veries the measurement activities by signing the accompanying ticket or proving report. This person must be a representative of the connecting carrier or shipper and should be knowledgeable about the information contained in the ticket or proving report.
Glossary
September 2002
G.21
BP Pipelines
Index
Page numbers in italics refer to gures.
A
ACT unit see LACT/ACT units adjustments procedure for filing, 17.5 sample affidavit, 17.11 when to file, 17.4 API gravity determination equipment list, 2.4 sample to use in lease tanks,
11.7
flow rate in LACT unit, 9.7 meter factor, 14.1, 14.4, 14.6 net standard volume, 3.13,
4.16
prover diameter, 15.5 prover prerun length, 15.7 pulses per barrel, 14.7 sample grabs per transfer, 9.8 tank volume, 7.3 waterdraw, 16.11 calibration PET, 2.12 pressure transducers, 9.10 prover, see waterdraw calibration strapping tape, 7.5 capacity table see tank capacity table centrifuge tube cleaning, 5.5 types, 5.7, 5.11 claims procedure for filing, 17.5 sample letter, 17.9 when to file, 17.4 cleaning pipe prover, 16.5 sample bottle, 1.9 sample receiver, 10.6 tape, 2.9 thermometer, 2.9, 2.11 thief, 1.6 closing gauge large tank, 4.11 lease tank, 3.113.12 cloud point testing, 6.136.14 compositing samples, 1.7, 1.10 connection height see tank strapping Coriolis meter design, 13.713.8
1.7, 3.5
API gravity determination in tanks hydrometer method, 2.62.8 thermohydrometer method,
2.52.6
automatic sampling see also sample receivers equipment list, 10.4 labeling sample container,
uses, 13.7 crude oil manual sampling mixing tank contents, 1.4 number of samples from tanks, 1.14 reasons for sampling, 1.3 temperature determination mixing tank contents, 2.5 tests distillation, 6.106.11 hydrogen sulfide, 6.86.9 light ends, 6.76.8 mercaptans, 6.4 metals, 6.66.7 neutralization number,
6.11
nitrogen, 6.116.12 organic chlorides, 6.5 Reid vapor pressure,
10.6
mixing samples, 10.12 procedures handling portable receivers, 10.5 handling samples, 10.4,
6.9
water content, 5.125.14 cupcase thermometer accuracy limits, 2.2 cleaning, 2.11 immersion time, 2.15 use of, 2.2
10.5, 10.6
maintaining receivers,
10.5
automatic sampling system see also sample receivers design, 10.810.11 testing receivers, 10.7 total system, 10.7
D demulsifier, 5.4
detectors calibration, 14.11 distance between, 15.6 maintenance, 16.4 role in nonrepeatability, 14.9 role in prover calibration,
C
calculations average tank temperature, 2.8 detector resolution, 15.4 distance between prover detectors, 15.6 ELM system functions, 9.11
16.3
role in prover sizing, 15.5 role in prover system design,
15.3
Index
September 2002
I.1
BP Pipelines
6.106.11 E
effective inside tank height see tank strapping ELM system described, 9.11 flow computer live input, 9.12 manual input, 9.11 ticket output, 9.129.13 security, 8.7
differences between lease tank and large tank, 3.5 innage, 3.5 opening gauge, 3.63.11 gauging height see tank strapping
H
haze testing, 6.156.16 hydrogen sulfide testing, 6.86.9
I
inferential meter see turbine meter, ultrasonic meter, Coriolis meter
F
field centrifuge testing see suspended S&W testing flash point testing, 6.146.15 forms LACT/ACT inspection blank, 9.15 sample, 9.14 proving report, 14.15 sample adjustment affidavit, 17.11 sample claim letter, 17.9 sample lease meter ticket, 17.13 sample tank ticket, 17.12 strapping report, 7.13 waterdraw calibration worksheet, 16.13
K
Karl Fischer titration see suspended S&W testing
L
labeling automatic samples, 10.6 demulsifier bottle, 5.5 sample container, 1.6, 1.7,
10.2
laboratory testing see suspended S&W testing LACT/ACT units checking calibrating pressure transducers, 9.10 calibrating temperature devices, 9.10 checkout list, 9.5 equipment list, 9.4 procedures, 9.69.9 programming run time,
G
gauging calculations, 3.13, 4.16 equipment list, 3.4, 4.4 for inventory annually, 4.134.15 monthly, 4.124.13 procedures, large tank closing gauge, 4.11 differences between lease tank and large tank, 4.5 external floating roof, 4.5 innage, 4.8 internal floating roof, 4.7 opening gauge, 4.84.11 procedures, lease tank closing gauge, 3.113.12
9.5
inspection form blank, 9.15 sample, 9.14 meter proving, 14.5 reporting malfunctions, 17.1 where to seal, 8.5 light ends testing, 6.76.8
large tank, 4.5 lease tank, 3.5 middle large tank, 3.5, 4.5 lease tank, 1.7, 3.5 number to take large tank, 1.14 small tank, 1.14 types listed, 1.4 upper large tank, 4.5 lease tank, 3.5 manual sampling see also API gravity determination, suspended S&W testing, settled S&W determination, manual samples compositing, 1.7, 1.10 equipment list, 1.3 mixing samples, 1.16 mixing tank contents, 1.4 number of samples to take crude tank, 1.14 product tank, 1.15 procedures all-levels, 1.10 bottle, 1.91.11 running, 1.11 spot with bottle, 1.91.10 spot with thief, 1.51.8 summary of, 1.13 where to sample large tank, 4.5 lease tank, 3.5 mercaptans testing, 6.4 metals testing, 6.66.7 meter factor accuracy limits, 11.411.5,
14.7
calculating, 14.1, 14.2, 14.4,
14.6, 14.8
causes of fluctuations, 14.10 effect of operating conditions,
11.11
trending, 11.5 troubleshooting,
M
manual samples clearance, 1.7, 3.8, 3.9 labeling sample container, 1.7 lower
14.1114.12
meter prover calibration schedule, 14.6 calibration, see waterdraw
Index
September 2002
I.2
BP Pipelines
calibration cleaning 16.5 maintenance, 14.13 meter proving procedures general, 14.6 portable prover, 14.8 stationary prover,
14.614.7
proving report, 14.15 proving schedule, 14.5 repeatability requirements,
inspection procedures auxiliary equipment, 11.5 meter, 11.411.5 troubleshooting, 11.5 inspection schedule, 11.3 maintenance, 11.3 PD meter installation design, 11.711.10 equipment, 11.711.8 factors affecting accuracy,
proving see meter proving proving system see meter proving system
R
reference height see tank strapping Reid vapor pressure testing in crude oil, 6.56.6 in products, 6.13 repeatability and detectors, 15.4 and pulse generator, 15.4 causes of nonrepeatability,
11.1111.12
operating ranges of meter,
14.8
troubleshooting meter factor, 14.10 nonrepeatability, 14.9 procedures, 14.1114.12 meter proving system design criteria, 15.315.4 detector resolution, 15.4 prover sizing diameter, 15.5 distance between detectors, 15.6 prerun length, 15.7 valve actuator time, 15.7 prover sphere type, 15.8 pulse generator, 15.4 mixing samples, 1.16, 10.6 tank contents, 1.4
11.10
specifications, 11.811.10 PET accuracy limits, 2.2 immersion time, 2.15 keeping in motion, 2.2, 2.15 petroleum products manual sampling mixing, 1.4 number of samples, 1.15 reasons for sampling, 1.3 summary of procedures,
14.9
requirements, 14.8 troubleshooting,
14.1114.12
reports distributing, 17.7 types, 17.6 using SMART to create, 17.6
S
S&W see settled S&W determination, suspended S&W testing sample bottle cleaning, 1.9 sample container labeling, 1.7 sample pots see sample receivers sample receivers cleaning, 10.6 maintaining, 10.5 portable described, 10.10 handling, 10.5 mixing equipment, 10.12 sizes, 10.13 stationary described, 10.9 mixing equipment, 10.11 samples, see manual samples, automatic sampling sampling, see manual sampling,
1.13
tests flash point, 6.146.15 oxygenate content, 6.16 Reid vapor pressure, 6.13 viscosity, pour point, cloud point, 6.136.14 water content and haze,
N
neutralization number, 6.11 nitrogen testing, 6.116.12
6.156.16
pour point testing in crude oil, 6.9 in products, 6.136.14 prover see meter prover prover sphere inflation, 16.1616.17 preparing for calibration,
O
opening gauge large tank, 4.84.11 lease tank, 3.63.11 organic chlorides testing, 6.5 orifice meter design, 13.3 orifice plates, 13.4 uses, 13.4 oxygenate content, 6.16
16.4
role in changing meter factor,
14.10, 14.11
role in nonrepeatability, 14.9 roundness tolerances,
P
PD meter design, 11.6 equipment for checking, 11.3
16.1516.16
types, 15.8, 16.14 when to check, 14.13
Index
September 2002
I.3
BP Pipelines
automatic sampling seals see also security boxcar seal, 8.8 broken, 8.6 during closing gauge, 3.11,
3.12
during opening gauge, 3.6 equipment for working with,
maximum allowable, 5.1 number of samples to take large tank, 1.14 small tank, 1.14 preparing demulsifier, 5.4 preparing samples, 5.5 preparing water-saturated toluene, 5.4 water content by volume,
tests for crude oil, see crude oil, tests for products, see petroleum products, tests thermometers verifying glass thermometer,
2.122.13
verifying PET, 2.112.12 thief, 1.5 cleaning, 1.6 cord, 1.5 procedures for lowering, 1.6 tickets correcting, 17.6 distributing, 17.7 output from ELM system,
8.3
how to use, 8.4 signs of tampering, 8.6 when and where to use on tanks, 8.48.5 where to use on LACT/ACT units, 8.58.6 security see also seals on ELM systems, 8.7 on LACT/ACT units, 8.58.6 on tanks, 8.7 settled S&W determination procedure, 1.71.8, 3.9, 3.12 rejecting tank for, 3.1 SMART for meter factor trending,
5.125.14 T
tank calibration table see tank capacity table tank capacity table development, 7.3, 7.9 use, 2.4, 3.13, 4.16 tank strapping calibration of tape, 7.5 equipment list, 7.3 fill condition, 7.12 procedures checks before strapping,
9.129.13
sample lease meter, 17.13 sample tank, 17.12 types, 17.6 using SMART to create, 17.6 turbine meter design conventional, 12.5 helical, 12.6 inspection procedures auxiliary equipment, 12.4 meter, 12.4 maintenance, 12.3 operation, 12.5, 12.12 turbine meter installation design, 12.712.11 factors affecting accuracy,
7.4
circumference, 7.57.7 connection height, 7.10 deadwood measurement,
14.9
for proving report, 14.6 for reports, 17.6 for settled S&W determination, 4.13, 4.14 for tickets, 4.12, 4.13, 17.6 sample lease meter ticket, 17.13 sample tank ticket, 17.12 sphere see prover sphere strapping see tank strapping sulfur testing, 6.36.4 suspended S&W testing equipment list, 5.3 field centrifuge method large tanks and automatic samplers, 5.85.10 lease tanks, 5.75.8 sediment only, 5.11 Karl Fischer titration, 5.12 laboratory method,
7.11
effective inside tank height, 7.97.10 height measurements, 7.7 manhole dimensions, 7.11 overflow line, 7.10 plate thickness, 7.4 reference height, 7.9 shell height and tilt,
12.13
operating range conventional meter, 12.9 helical meter, 12.10 pulse transmission accuracy,
7.87.9
strapping report, 7.13 when to strap, 7.3 temperature determination in tanks cupcase thermometer procedures, 2.102.11 for custody transfer, 2.10 for inventory, 2.10 immersion times, 2.15 mixing tank contents, 2.5 PET procedures, 2.92.10 where to measure, 2.8, 2.8,
12.1012.11 U
ultrasonic meter, 13.513.6
V
viscosity testing in crude oil, 6.9 in products 6.136.14
5.115.12
Index
2.14
September 2002
I.4
BP Pipelines
W
water content testing in crude oil, 5.125.14 in products, 6.156.16 waterdraw calibration accuracy limits, 16.10 calculations, 16.11 calibration schedule, 16.3 procedures general, 16.6 pipe prover, 16.416.5,
16.816.10
small-volume prover,
16.6, 16.10
prover sphere types, 16.14 sphere inflation, 16.1616.17 roundness, 16.1516.16 worksheet, 16.13 water-saturated toluene, 5.4
Index
September 2002
I.5