You are on page 1of 16

RTP Link and MLC Export Customization and Use

Revision History
Revision XiO_CST_RTPMLC_430_A.pdf Date August 2005 Changes New Document

All terms mentioned in this manual that are known trademarks, registered trademarks, or service marks have been appropriately labeled as such. Other products, services, or terms that are referred to in this manual may be trademarks, registered trademarks, or service marks of their respective owners. CMS makes no claim to these trademarks. Use of a term in this manual should not be regarded as affecting the validity of any trademark, registered trademark, or service mark. CMS and XiO are registered trademarks of CMS, Inc. Plan for More is a trademark of CMS, Inc. Adobe, Acrobat, and the Acrobat logo are trademarks of Adobe Systems Incorporated, which may be registered in certain jurisdictions. DICOM is a trademark of NEMS, National Electrical Manufacturers Association Hewlett-Packard is a trademark of Hewlett-Packard Corporation. Netscape and Netscape Navigator are registered trademarks of Netscape Communications Corporation in the United States and other countries. Netscape's logos and Netscape product and service names are also trademarks of Netscape Communications Corporation, which may be registered in other countries. UNIX is a registered trademark in the United States and other countries, licensed exclusively through X/Open Company Ltd. Red Hat, the Red Hat "Shadow Man" logo, RPM, Maximum RPM, the RPM logo, Linux Library, PowerTools, Linux Undercover, RHmember, RHmember More, Rough Cuts, Rawhide and all Red Hat-based trademarks and logos are trademarks or registered trademarks of Red Hat, Inc. in the United States and other countries. Linux is a registered trademark of Linus Torvalds. Mozilla is a Copyright 1998-2004 of The Mozilla Organization. Copyright 2005 by CMS All rights reserved. Performance characteristics given in this manual are for reference only and are not intended as guaranteed specifications. XiO RTP Link and MLC Export Customization and Use Document ID: XiO_CST_RTPMLC_430_ A.pdf Part Number: 98930

Manufacturer: CMS, Inc. 1145 Corporate Lake Drive St. Louis, Missouri 63132 USA Telephone: 1-314-993-0003 Toll Free Telephone: 1-800-878-4267 Facsimile: 1-314-993-0075 Email: sales-na@cms-stl.com European Union: CMS GmbH Sundgauallee 25 D-79114 Freiburg Germany Telephone: +49-761-88188-0 Facsimile: +49-761-88188-11 Email: sales-europe@cms-euro.com European Union Authorized Representative: Paul Shane Bennetts M Devices Group Marlborough House, Riding Street Southport PR8 1EW UK Telephone: (0) 1704 544 944 Asia: CMS Worldwide Corp. 25th Floor, Suntime International Mansion #450 Fushan Road Pudong, Shanghai 200122 P. R. China Telephone: +86-215058-1041 Facsimile: +86-215058-1045 Email: PacRim@cms-stl.com Japan: CMS, Japan, K.K. Heiwa Dai ichi Bldg., 5/F., 1-4-5, Hirakawacho, Khiyoda-ku, Tokyo, Japan 102-0093 Telephone: 81-3-3556-5391 Facsimile: 81-3-3556-5393

RTP Link and MLC Export - Customization and Use

Table of Contents

Table of Contents

Introduction.....................................................................................................................................1 Send Data Via RTP Link.............................................................................................................3 Send MLC Control Files Via Multileaf Collimator ....................................................................7 Steps for Varian Users.............................................................................................................7 Steps for Elekta Javelin Users..............................................................................................9

iii

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use


Introduction
RTP Link is an export program that allows you to digitally send plan parameters to another device, primarily a record and verify system. The kind of data that is sent is demographic (name), field sizes, angles, MU, MLC and block/compensator/cone/wedge IDs. The exported file is in an ASCII format and can be written to floppy or to XiO's hard disk for network retrieval by the receiving system. NOTE: For bolus beams, the receiving system expects one MU value per beam; therefore, a MU value will only be transferred if the beam is 100% bolused, or 100% unbolused. For all bolus beams, XiO reports the thickness of the bolus in centimeters.

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use

Before using RTP Link for the first time, you will want to customize the mapping of machine names and default tolerance tables. The RTP Link, Multileaf Collimator and Customization options are located under the Data Export heading on the main XiO window.

1.

Click Data Export on the main XiO window. XiO displays the following output selections.

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use


2. Click the Customization option. XiO displays the following output customization dialog.

3. 4.

Select an XiO Machine ID and create a name for the Remote Machine ID that is recognized by the receiving system. Enter a Tolerance Table Number that will be used by the receiving system. The tolerance tables are used on the receiving system to set up tolerances for table and gantry movements. Click Done. (Optional) Repeat the customization for as many machines as necessary.

5. 6.

Send Data Via RTP Link


Siemens and Varian Users (excluding Varian 120 leaf) can use RTP Link to export the patient, plan, and MLC controller files to a receiving system. 1. Click Data Export on the main XiO window. XiO displays the following output selections.

2. 3.

Click the RTP Link option. Select the Patient ID for the patient you want to export.

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use

Send Data Via RTP Link (cont.)


4. Select the Plan ID for the patient.

5.

Click Done. XiO displays the following dialog box.

6. 7. 8.

Enter the RTP Link Filename. XiO automatically adds a .rtp extension to this filename. (Optional) If necessary, edit the Remote Patient ID to match the format used on the receiving system. Enter a Course ID. XiO uses this number to indicate what fraction of the treatment you are sending. This information must be entered.
4 CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use

Send Data Via RTP Link (cont.)


9. Enter a description in the Rx Site: field for the Prescription Site.

10. Select a Transfer Method. The options are for 3.5 inch floppy disk and network. 11. (Optional) If necessary, edit the Remote Beam ID. Each beam number must be unique for the selected patient. The Beam Description and Gantry Angle fields appear automatically from the plan. 12. (Optional) If necessary, edit the Machine ID. The default value is the mapped machine ID you entered in RTP Link Customization. 13. (Optional) If necessary, edit the Tolerance Table value. XiO maps the default value to the Tolerance Table value you entered in RTP Link Customization. 14. Edit the Wedge ID. The default value is the XiO Wedge ID; however, you can edit this value to match the appropriate value used by the receiving system. 15. Click the Next Beam button. XiO displays the editable values for the next beam. NOTE: The RTP Link filename, Remote Patient ID, Course ID, Treatment Site, and Transfer Methods are now non-editable unless you go back to Beam 1. 16. Repeat the steps for editing specific beam parameters for each beam and click the Next Beam button between beams. 17. (Optional) Click the Previous Beam button to go back to edit a previous beam's parameters.

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use

Send Data Via RTP Link (cont.)


18. After you edit the last beam, XiO displays a Send File button at the bottom of the dialog box.

19. Click the Send File button when you are ready to export. XiO displays a yellow message when the file export is complete and prompts you for another Patient ID to export. OR If you are finished, click the Cancel button.

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use


Send MLC Control Files Via Multileaf Collimator
Varian and Elekta "Javelin" workstation users who have MLCs and do not have DICOM can use this option to send MLC Control files. Siemens users who do not have DICOM should use RTP Link to export MLC Control files.
Steps for Varian Users

1.

Click Data Export on the main XiO window. XiO displays the following output selections.

2.

Click the Multileaf Collimator option.

3. 4. 5.

Select the Patient ID for the patient you want to export. Select the Plan ID for the patient. Click Done. XiO displays the following dialog box.

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use


Send MLC Control Files Via Multi-leaf Collimator
Steps for Varian Users (cont.)

6. 7. 8. 9.

Enter a File name for the MLC Controller. Select the Beam Numbers representing the MLC fields you want to export. Select the Varian Revision File. (Rev G or H is needed for Varian 120 leaf files.) Select the file Transfer Method. The options are network and 3.5 inch floppy.

10. Click the Send File button when you are ready to send the MLC Controller file. XiO displays a system message to notify you when the file has been sent. NOTE: XiO only display the Send File button after you have fill in all the required fields. 11. When you are finished, click Cancel to exit Data Export.

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use


Send MLC Control Files Via Multileaf Collimator (cont.)
Steps for Elekta Javelin Users

12. Click Data Export on the main XiO window. XiO displays the following output selections.

13. Click the Multileaf Collimator option. 14. Select the Patient ID for the patient you want to export. 15. Select the Plan ID for the patient.

16. Click Done. XiO displays the following dialog box.

CMS, Inc.

RTP Link and MLC Export - Customization and Use

RTP Link and MLC Export Customization and Use


Send MLC Control Files Via Multileaf Collimator
Steps for Elekta Javelin Users (cont.)

17. Enter the MLC Node Name. Valid node names range from MLC_01 to MLC_08. 18. Enter the Group number. 19. Select the Beam Numbers representing the MLC fields you want to export. 20. Click the Send File button when you are ready to send the MLC Controller file. NOTE: XiO ONLY displays the Send File button after you have filled in all the required fields. 21. After you are finished, click Cancel to exit Data Export.