Professional Documents
Culture Documents
Column
High separation, secondary interaction, lot variation, temperature control
Detection
Sensitivity, selectivity, ambient temperature fluctuation affect
Sample
Extraction method selection, sample solvent selection, dissolved oxygen
Instrument
Performance, stability, durability, ease-of-use
Evaluation overall
Stability of analytical technique, robustness, quantitative accuracy
Mobile Phase
Mobile Phase
Water
Ethanol
Ethanol
1L
Water
Mobile Phase
Column
:Shim-Pack VP-ODS
(150mmL. x 4.6mmI.D.)
Flow rate :1mL/min
Temp.
:40C
Detection :UV-VIS (210nm)
2
3
1
A
1
3
B
min
10
Mobile Phase
Buffer Solution Notation and Preparation (1)
Preparation: 20mM Phosphoric acid buffer solution (pH2.5)
Mobile Phase
Buffer Solution Notation and Preparation (2)
Mobile phase A: phosphoric acid interpreted as 20mM
Mobile phase B: sodium interpreted as 20mM
Mobile phase C: phosphoric acid, sodium both
interpreted as 20mM,
pH adjusted with
perchloric
acid
Column :Shim-Pack VP-ODS
1,
Peaks
1:
acetaminophen
2: dihydrocodeine
3: caffeine
(150mmL. x 4.6mmI.D.)
Flow rate :1mL/min
Temp.
:40C
Detection :UV-VIS (210nm)
3
A
If interpreted wrong
Dihyrocodeine elution position
is different!!
B
1
3
C
min
10
Mobile Phase
pH Buffer Capacity and Peak Shape
Left figure: citric acid buffer solution (pH 4.5) used as mobile phase
Right figure: phosphoric acid buffer solution (pH 4.5) used as mobile phase
Column
Mobile
phase
Flow rate
Temp.
Detection
: STR ODS-II
(150mmL. x 4.6mm I.D.)
: 20mM buffer solution (pH 4.5)
/ acetonitrile (3/1, v/v)
: 1mL/min
: 40C
: UV-VIS (240nm)
Mobile Phase
Differences between Acetonitrile and Methanol
Absorbance HPLC grade acetonitrile has low absorbance at short UV wave
lengths
Mobile Phase
Effect of Flow Rate Change (1)
Retention time change
Peak area change Quantitation Error!!
Detector that responds absolute quantity of compound
Peak area is unaffected by flow rate.
Detector that responds to concentration of compound
Peak area is affected by flow rate.
*Almost all LC detectors are concentration-response type
Mobile Phase
Effect of Flow Rate Change (2)
Peak Area
Area Ratio
A (1.00mL/min)
B (0.99mL/min)
B/A
methylparaben
3901107
3938213
1.0095
ethylparaben
2744605
2771087
1.0096
n-propylparaben
2626362
2652087
1.0098
n-butylparaben
2886547
2914591
1.0097
Column
: Shim-Pack VP-ODS (150mm L. x 4.6mm I.D.)
Mobile phase : methanol / water = 6/4 (v/v)
Flow rate
: A ; 1.00mL/min
B ; 0.99mL/min
Temperature : 40C
Detection
: UV-VIS (260nm)
Mobile Phase
Peaks
1: methylparaben
2: ethylparaben
3: propylparaben
1
2
3
A : Pre-mixed solutions (2) delivered using one pump
B : Each solvent delivered with a separate pump and mixed
A
B
Mobile Phase
Area Ratio
B/A
methylparaben
4100127
4151876
1.013
ethylparaben
2885602
2922556
1.013
n-propylparaben
2761054
2797180
1.013
n-butylparaben
3035538
3075259
1.013
Column
Mobile phase
Flow rate
Temperature
Detection
Mobile Phase
Troubles caused by Bubbles Formed in Mobile Phase
Causes of bubble occurrence
Occurs when dissolved air exceeds air dissolution saturation point.
Causes: Solution warming, decreasing pressure, solvent mixing, etc.
Column
Peak deformation, column degradation
Detector
Noise generation
Mobile Phase
Influence of Dissolved Air in Mobile Phase on Detection
Fluorescence detection
Quenching due to oxygen
Changes in peak area
Absorption detection
Changes in background absorbance
Baseline fluctuation (especially with gradient)
Appearance of peaks originating from oxygen
Mobile Phase
Degassing Influence in Fluorescence Detection (1)
400
Degassed
mAU
* Analysis of bisphenol A
by fluorescence detection
BPA
300
Not degassed
200
100
0.0
0.5
1.0
1.5
2.0
2.5
min
3.0
3.5
4.0
4.5
5.0
Mobile Phase
Degassing Influence in Fluorescence Detection (2)
Gas-liquid membrane separation degasser takes time
to get degassing result.
Peak Area Value (106)
Analyte : Pyrene
Detection : Fluorescence
(Ex 310nm, Em 390nm)
Degassing:Gas-liq. separation membrane
4
3
2
Degasser OFF
Column
: Shim-Pack VP-ODS
(150mm L x 4.6mm I.D.)
Mobile Phase: acetonitrile/H2O = 4/1 (v/v)
Flow rate
: 1mL/min
Temperature : 40C
Degasser ON
1
0
0 10 20 30 40
Time from start of degassing (min)
50
60
Mobile Phase
Background Changes in UV Detection
Air saturation
Degassed
Mobile Phase
Dissolved Air-derived Peaks in UV Detection
Peak appears with mobile phase solution injection!?
The difference in dissolved air quantities in mobile phase and
sample solvent solutions can be expressed as peaks.
Mobile phase
not degassed
Degassed
mobile phase
Oxygen bubbling
Oxygen bubbling
Air saturation
Air saturation
He bubbling
He bubbling
Detection: UV210nm
Mobile Phase
Dissolved Air-derived Peaks with RID Detector
15
0.0
mRIU
10
mRIU
- 0.2
- 0.4
Magnified
0
10
15
20
25
30
min
10
15
20
25
30
min
Column
: SCR-101C
Mobile phase : Water
Flow rate
: 0.5mL/min
Temperature : 85C
Detection
: RID detector
Degasser
: Gas-liq. Membrane Separ. (DGU-14A)
Sample
: 0.5% mannitol
Detection
Cell Temperature Influence in UV Detection (1)
In general, spectrum shape is affected by ambient temperature.
Spectrum at low
temperature
Absorbance
Spectrum at high
temperature
Absorbance difference due to
temperature change
Measurement wavelength
Detection
Cell Temperature Influence in UV Detection (2)
Comparison of peak area values without temperature control (34C)
Peak Area Ratios Compared with Areas at 34C
40C
50C
Phenol
0.994
0.991
Benzoic acid
0.991
0.986
p-toluic acid
0.985
0.968
Column
: Shim-Pack VP-ODS (150mm L. x 4.6mm I.D.)
Mobile phase : 10mM acetic acid (sodium) buffer solution (pH 4.5) /
Methanol = 7/3, (v/v)
Flow rate
: 1mL/min
Temperature : 40C
Detection
: UV-VIS (260nm)
Detection
Cell Temperature Influence in Fluorescence Detection (1)
Normally, the higher the temperature, the lower the fluorescence intensity.
Analysis result of acridine (Temp. effect on fluorescence intensity)
Areas 2025C
13.5% decrease
Areas at 2025C
2.2% decrease
Detection
Cell Temperature Influence in Fluorescence Detection (2)
Peak Area
20C
40C
Area Ratio
A40/A20
ASP
1500427
1165757
0.78
THR
1478447
1143925
SER
1570743
GLU
Peak Area
20C
40C
Area Ratio
A40/A20
TYR
1156850
850976
0.74
0.77
PHE
1315479
995832
0.76
1204273
0.77
HIS
1819675
1431665
0.79
1315462
1017768
0.77
LYS
487433
336837
0.69
PRO
505373
436522
0.86
ARG
1105530
839183
0.76
GLY
2440772
1924900
0.79
ALA
1558560
1203529
0.77
CYS
177078
138426
0.78
VAL
1565440
1226042
0.78
MET
1026839
790013
0.77
ILE
1457340
1126683
0.77
LEU
1185514
911017
0.77
Detection
Solvent Influence on Fluorescence Intensity
Fluorescence Spectrum of Bisphenol A in Each
Solvent
(Excitation wavelength 270nm, with background
correction)
Methanol
Acetonitrile
Wavelength(nm)
Sample
Cooling
No cooling
Column
Mobile
phase
Flow rate
Temp.
Detection
: 1mL/min
: 40C
: UV-VIS (245nm)
Sample
Influence of Sample Solvent on Peak Shape (1)
Aqueous solvent
Methanol solvent
Analyte
Liquid flow
Methanol
solvent
Theoretical
Plate
Number
Aqueous solvent
Analyte
Liquid flow
Methanol / H2O
Sample
Influence of Sample Solvent on Peak Shape (2)
If sample solvent elution is stronger than that of mobile phase, peak shape deteriorates.
Column
: STR ODS-II
(150mm L. x 4.6mm I.D.)
Mobile
: 20mM citric acid (sodium) buffer
Phase
: solution (pH 4.5)
/ acetonitile = 3/1 ( v/v)
Flow rate : 1mL/min
Temp.
: 40C
Detection : UV-VIS (240nm)
Sample
Advantages of Automated Sample Pretreatment
Sample
Column Switching HPLC System
for Direct Injection of Blood Plasma (Co-Sense for BA)
Analytical column
Detector
Mobile phase
Autosampler
Pump
Dilution bypass
Pump
Mobile phase
Sample
Automated Pretreatment Example using Co-Sense for BA
Analysis of Plasma Spiked with Isopropylantipyrine
protein
Isopropylantipyrine
8.5
Time (min)
Injection volume
Mobile phase (Smpl. Inj. side)
Detection
: 100L
: 0.1% phosporic acid / acetonitrile = 95/5 (v/v)
: Isopropylantipyrine 275nm
Plasma matrix 280nm
17
Sample
Automatic Pretreatment
Using Co-Sense for BA
Sample
Causes and Control of Carryover
Causes and Means of Control in Autosampler
Cause: Adsorption to metallic materials due to ionic
interaction or coordination isomerism interaction (ionic
compounds or basic compounds)
Remove adsorbed components by needle washing
Control adsorption by making needle inert
Sample
Needle Wash Mechanism (Total Volume Injection)
Sample loop
Needle
Wash port
Pump
Column
Ready
Readystate
state
Needle
Washing
Washingstate
state
Washing this surface is important.
Needle seal
Sample
Selection of Needle Wash Solution
To control adsorption of substances due to ionic
interaction or coordination isomerism (ionic compounds
or basic compounds)
Acidify.
Make ion pair with large radius counter ion.
(Example) 100mM perchloric acid aqueous solution /
methanol (or acetonitrile)
Sample
Minimizing Contamination due to Basic Substances (1)
Standard
Solution
Blank
W/ Wash
0.000730
0.0730
W/O Wash
0.000425
0.0425
Column
Shim-pack VP-ODS (150mmL.x 2.0mm I.D.)
Mobile phase 10mM phosphoric acid buffer solution (pH 2.6) containing
100mMNaClO4 / acetonitrile
= 55 / 45 (v/v)
Flow rate
0.2mL/min
Temperature 40C
Detection
UV (260nm using semi micro cell)
Sample
Minimizing Contamination due to Basic Substances (2)
Standard
Solution
Blank
SUS
0.000425
0.0425
Teflon coating
0.000023
0.0023
PEEK coating
0.000021
0.0021
0.000009
0.0009
Sample
Minimizing Contamination Due to Basic Substances (3)
chlorohexidin
chlorohexidin
Sample
Minimizing Contamination of Hydrophobic Substances (1)
Sample loop
Pump
Needle
Column
Rotor seal
Adsorption of hydrophobic
compounds is thought to occur
mainly in rotor seal groove
through which samples pass.
It is believed that because the
adsorptive substance is washed
with mobile phase, it elutes
gradually.
Sample
Minimizing Contamination of Hydrophobic Substances (2)
Blank
[%]
10 minutes aging
0.000051
0.0051
20 minutes aging
0.000136
0.0136
Column
Sample
Minimizing Contamination of Hydrophobic Substances (3)
1.0
9412178
0.5
0.0
10
12
14
min
Sample
High-Throughput Autosampler SIL-HT
Reduced carryover
Multi-sample processing