Professional Documents
Culture Documents
Brian W Woodget
Owner
Title
Classification
Keywords
ukoer, sfsoer, oer, open educational resources, metadata, analytical science, cpd
training resource, analytical chemistry, measurement science, chromatographic
methods, thin layer chromatography
Description
This chapter reviews the variety of separation techniques which come under the
banner of Chromatography, compares the Plate and Rate theories of separation
and finally considers in more detail that of Thin Layer Chromatography. Some of the
images show a degree of animation.
http://creativecommons.org/licenses/by-nc-nd/2.0/uk/
Language
English
File size
1.4 Mbytes
File format
1
Royal Society of Chemistry 2010
Analytes are usually introduced into the mobile phase stream. As this stream
moves past the stationary phase, some of the analytes will interact with this fixed
phase, resulting in them being retained. Other analytes remain in the mobile
phase and move on. The retained analytes can then move back into a fresh
portion of mobile phase as it flows past them.
To achieve separation each analyte needs to spend different lengths of time
being retained or held by the stationary phase.
History of chromatography
The first chromatographic separation was reported in 1903 by
Tswett, who separated plant pigments using a stationary phase
of calcium carbonate and a mobile phase of petroleum ether.
The separation as illustrated in figure (6.1)
shows the two coloured pigments separating
as they progress through the column allowing
them to be collected separately as they emerge
(elute) from the column.
As the pigments formed distinct
coloured bands Tswett coined
the phrase chromatography
(colour writing) from chroma
colour and graphien - writing
Figure 6.1
In 1941 Martin and Synge published a paper describing the principles of partition
chromatography, in which the stationary phase was a high boiling liquid coated
onto a solid support. In 1952 they were jointly awarded the Nobel prize for
chemistry.
In 1952, Martin and James invented gas-liquid chromatography (glc), a
technique which uses a gas as the mobile phase. This is now a widely used
technique for separating volatile or semi-volatile organic substances, for example
herbicide pollutants in water or the components in a petroleum fraction.
Although research into high pressure (now termed high performance) liquid
chromatography (hplc) began in the late 1960s, the technique did not get
widely used until about a decade later with the advent of instrumental
components that could withstand pressures as high as 6000 psi and pumps that
could deliver these pressures in a smooth rather than a pulsed flow mode. More
recently, further advances have enabled the development of ultra performance
liquid chromatography (uplc) where pressures of 15000 psi are used.
In the 1990s electrochromatography was developed. The mobile phase flow is
driven through a packed capillary column by an electric field rather than a pump.
These developments are summarised in table (6.1) on the next slide.
7
Principles of chromatography
In column chromatography analytes are injected into the mobile phase stream
and are swept on to the column containing the stationary phase. Those
analytes which do not interact in any way with the stationary phase pass
straight through the column, they are the first to elute (flow out of the column).
Some analytes will interact so strongly with the stationary phase that they
remain on the column, never to be seen again. In between these two
extremes, are analytes which spend some time interacting with the mobile
phase, during this time they move through the column, and some time bonded
to the stationary phase. Each analyte spends a slightly different length of time
in the two phases and hence by the end of the column, the individual
components of the analyte sample, which had been injected into the mobile
phase stream, are separated and can be detected.
Remember the analogy of the boats on the river (slide 5)
A finely divided solid spread as a thin layer onto a sheet of glass or plastic;
A finely divided solid packed tightly into a glass or stainless steel tube,
referred to as a column;
A finely divided solid coated with a thin layer of organic liquid and again
packed into a glass or stainless steel column.
A glass or silica capillary whose inner surface is coated with a low volatile
organic liquid
Detector response
10 minutes
Retention time
Figure 6.2 - an example of a chromatogram
Ideal results are not always easily achieved, however the following notes provide
the basic information to enable simple chromatographic separations to be carried
out.
11
Column chromatography
Most chromatography is performed on
columns, figure (6.3) on the right is a
schematic of a packed column, the
arrows at the top of the column are
indicative of the mobile phase.
12
Modes of chromatography
The particles which fill a chromatographic column are not all the same
hence the types of interactions between analytes and stationary phases are
not all the same.
In liquid-solid chromatography the stationary phase is a solid such as
calcium carbonate or more likely these days silica or alumina (Al2O3).
In gas-liquid chromatography the stationary phase is a low volatility liquid
coated on to a solid support eg Carbowax 20M - a polyethylene glycol with
a molecular mass of 20,000 g/mole.
In liquid-liquid chromatography, the stationary phase is a liquid, eg a C18
silane, covalently bonded to an inert silica support.
Movement between the stationary and mobile phase can be summarised
by the following equation, where A is an analyte
Amobile
Astationary phase
13
Adsorption
Partition
Bonded phase
Ion exchange
Affinity
Size exclusion
The mode describes the way in which analytes interact with the stationary
phase.
14
Adsorption chromatography
If silica, a silicon-oxygen polymer as illustrated
in figure (6.4) is the stationary phase, the
surface hydroxyl groups interact with the
analytes, the more the interaction the more the
analytes are retained and therefore the longer
they take to pass through the column.
Si OH
O
Si OH
Figure 6.4 O
structure of silica
Applications of adsorption
chromatography
Separation of isomers
17
Partition chromatography
In partition chromatography a solid
support with a high surface area
such as crushed firebrick or
keiselguhr is coated with a high
boiling liquid which acts as the
stationary phase. Separation occurs
because of the differences in
solubility for the analytes in the
stationary and mobile phases.
The partition coefficient is defined as:
K =
Solid
support
Stationary phase
Analyte
18
Applications of partition
chromatography
There are an immense number of possible applications of partition
chromatography. The list below gives just a few examples of where this
technique is routinely applied:
19
Si OH + Cl SiR
Si O SiR + HCl
20
There are two types of partition chromatography normal phase and reversed phase, they are defined by the relative polarities of the mobile and stationary phases
For this reason, the use of silica (a polar molecule) as the stationary phase (as in adsorption chromatography) is also considered to be a normal phase separation method.
Because of its versatility and wide range of applicability, reversed-phased chromatography is the most frequently used hplc method.
21
NR3+ OHPolymeric
stationary phase
Exchanging ion
23
Cl- Br- OH
Separating column OH Cl
Cl OH
Br-- OH Br
Cl
Cl-OH-
Suppressor column
Conductance detector
OHH2O
OHD
NR3+OH- + NaX
NR3+X- + NaOH
SO3-Na+ + H2O
Comparatively low
conductivity
24
Figure (6.10) on the previous slide, shows that analyte ions literally exchange with
similarly charged ions from the stationary phase. Each analyte ion will interact with
the stationary phase according to its charge, size and to a certain extent shape,
thereby bringing about a separation. As the analytes are ions, a convenient way of
detecting them is to use a conductance detector.
The problem with using a conductance detector is that it will also measure the
conductance of the mobile phase. As the mobile phase is an ionic solvent it will
itself have a high conductance against which the comparatively small analyte signal
needs to be measured.
For a long time the problem described above, meant that ion chromatography was
not sensitive enough to detect low (ppm) levels of ionic analytes. The problem was
solved by a company called Dionex (their name is often synonymous with ion
exchange chromatography), which introduced a suppressor column after the
separating column. The suppressor column also contains an ion exchange resin.
The ions from this resin react with the ions of the mobile phase to produce
molecules which have a lower conductance, e.g. water, than the ions themselves.
The analyte ions will also react with this resin but the molecules formed will be of
equal conductance to those they replace. In the previous slide NaCl (eluting
analyte) would be replaced with HCl.
25
Separation of vitamins
Analysis of serum
Analysis of drugs
26
Affinity chromatography
Affinity chromatography involves the interaction of a ligand (bound to a solid
support with a spacer molecule) with the analyte.
activation
loading
the analytes to be separated are introduced into the
mobile phase stream.
binding
the analytes of interest are retained due to interaction
with the ligand of the stationary phase.
washing
elution
the analyte(s) of interest are washed from the column
by changing the mobile phase composition.
These five steps are illustrated in figure (6.12) shown on the next slide
29
activation
load
binding
washing
elution
Applications of affinity
chromatography
Purification of proteins
Immunoassays
31
32
6.13
6.14
33
Figure (6.14) on the previous slide, illustrates how each size exclusion column
is only capable of separating molecules within a limited molecular weight range
(103.5 - 105.6 on the graph shown). All molecules with a mass larger than this,
will pass straight through the column and be eluted together at the start of the
chromatogram ie: they will provide tm the retention time for un-retained
analytes. Molecules which have a mass smaller than the range will be trapped
within the pores of the stationary phase and never seen again.
In order for molecules to be separated they need mass differences of
approximately 10%.
34
Analysis of sugars
35
Chromatographic terms
Chromatography is a method used for separating a mixture of compounds.
Separation takes place on the column containing the stationary phase. By
washing eluent or mobile phase through the column the sample is separated
into its individual components and is eluted from the column.
A chromatograph is a sophisticated instrument used to separate analytes. The
components of different types of chromatograph will be considered in a later
chapters.
A chromatogram is a recording of the detector response versus time.
36
Signal
The chromatogram
Equation (6.2)
trA
tref
Equation (6.3)
38
39
Equation (6.4)
40
Resolution -Rs
t r2 t r1
Rs
0.5( w 2 w 1 )
Equation (6.5)
where:
tr1 is the retention time of one analyte
and tr2 is the retention time of the next
analyte to elute
w1 and w2 are the widths of the peaks at
base when approximated to triangles.
Note: both tr measurements are made from
that of the unretained peak.
k'
'
t rA
tm
Equation (6.6)
42
43
The analyte (64 parts) partitions itself between the two phases as they come
into contact.
Mobile phase
64 32 32
64 16 12 16 12 16 12 8
2
8 4 8 4
8 4 8 4 2
Cs
1
Cm
4
4
8 2 12 8 12 8
8
16 32
16
Stationary phase
44
In figure (6.18) the analyte started on one plate of the mobile phase but after 5
equilibrations it was spread across five plates. This demonstrates how dividing the
phases into theoretical plates explains the spreading of analyte molecules during a
chromatographic separation. It also demonstrates why chromatographic peaks
are Gaussian, or bell shaped, in shape rather than being triangular. In reality the
analyte molecules also spread due to
other factors which will be discussed
when the deficiencies or limitations
of the plate theory are considered.
45
solvent
Concentration of analyte
The figure (6.20) below illustrates the effect of the analyte partition
coefficients, the parts of the analyte have been plotted as the concentration
of the analyte. Analytes with different partition coefficients separate from one
another as they partition between the theoretical plates.
50
150
250
300
Plate number
350
400
450
The nature of the analyte and the mobile and stationary phases
determines the partition coefficients;
How sensitive a detector is to a particular analyte - the response factor
changes the size of the peak observed;
The number of equilibrations which can take place in a column ie the
number of theoretical plates determines how good the final separation of
the analytes will be.
47
Column efficiency
H or HETP is the length of column which
corresponds to a theoretical plate. This
needs to be a minimum for efficient
separation
N is the number of plates per column
L is the length of the column
From a chromatographic peak, such as
the one shown in figure (6.21), it is
possible to evaluate H and N.
Figure 6.21- chromatographic
peak showing tr and w
Continued on the next slide
48
L NH
Equation (6.7)
16t r2
N 2
w
Equation (6.8)
L Lw 2
H
N 16t r2
Equation (6.9)
For this reason H and N are often quoted, e.g. in suppliers catalogues, to
give an indication how good (efficient) a column is at separating particular
analytes. The higher the value of N and conversely the lower the value for
H, the more efficient the column. This enables a comparison of similar
chromatographic columns from different suppliers. However it is important
to remember that when selecting a column to perform a particular
separation, like separates like i.e. in gas chromatography a polar column
is needed to separate polar analytes. In practise this means that if a
column is efficient at separating one class of analytes (eg alcohols) it will
not necessarily be as efficient as separating those of a different nature (eg
alkanes).
49
The plate theory assumes K is linear and therefore that the plates are
symmetrical;
It assumes rapid equilibration of the analyte between mobile and
stationary phases;
All peak broadening is not accounted for;
The effect of the mobile phase is ignored;
The dimensions (eg the thickness) of the phases are not taken into
account
50
Eddy diffusion
6.22
The fact that the analyte molecules do not all take the same path through a
packed column means that they do not all reach the detector at the same
time hence peak broadening is observed.
51
Longitudinal broadening
(blotting paper effect)
Mobile phase
flow
Narrow at point of
introduction on the
column
If the mobile phase velocity is high the worse the broadening becomes.
53
As will be shown at the end of the Rate theory section of these notes, this
apparent contradiction results in an optimum value for the mobile phase flow
rate.
54
H Au
B
Cu
u
Equation (6.10)
Where:
u is the average mobile phase velocity
A takes into account eddy diffusion
B longitudinal diffusion
C mass transfer
The effect of the three terms in the van Deemter equation are shown
graphically in figure (6.25) below.
6.25
Remember that we want H to be a minimum for efficient separations. So the
above plot illustrates that there is an optimum mobile phase flow rate to achieve
this, the range is shown by the red arrows. The red line shows the combination
of the three broadening terms.
56
The plate theory does not take into account all these variables nor does it take
into account the mobile phase flow rate which as has been shown has a
significant effect on the peaks obtained during a separation. However the
plate theory does explain why separation is achieved and the shape of the
peaks.
57
Types of chromatography
This chapter of this teaching & learning programme has outlined the theoretical
principles of a number of chromatographic techniques. Three of these
techniques will be considered in more detail. These are:
58
6.26
60
Heptane / toluene
(non-polar)
The mobile
phase is too
weak compared
to the stationary
phase
toluene
Methanol / toluene
(polar)
The mobile
phase is too
strong compared
to the stationary
phase
Tlc visualisation
In tlc there are a number of ways of observing or visualising the separated
(and mostly colourless) analytes. These divide into two categories physical
and chemical methods.
Physical Methods
The physical methods are based on the absorbance or emission of radiation,
usually in the ultraviolet (UV) range of the electromagnetic spectrum. Many of
the compounds separated will absorb uv light. Therefore placing a developed
tlc plate under a UV lamp will show up these compounds as darker spots (the
light having been absorbed) against a lighter background. If the separated
analytes are capable of emitting radiation (fluorescing), then placing the plate
under a suitable light source will results in bright, fluorescing spots being
shown.
Some analytes are incapable of absorbing UV light or fluorescing. In this case
a fluorescent compound, such as cadmium or zinc sulphides, is incorporated
into the stationary phase so the whole plate appears bright yellow/green when
held under a uv light source. Once this type of plate is developed the analytes
appear as dark spots as they quench the background fluorescence from the
stationary phase.
63
Chemical Methods
Chemical methods of visualisation can be selective and highlight a few
analytes, or be universal and highlight nearly every analyte. To use a
chemical visualiser the reagent is sprayed on to the plate as a fine mist or the
plate is quickly dipped into a tank of the reagent vapour. Universal visualisers
include iodine and concentrated sulphuric acid.
Iodine forms complexes with organic compounds which are brown in colour
and hence the analytes on the tlc plate appear brown. The advantage of this
method is that the complexes are not very stable and so the spots fade within a
short period of time leaving the analyte unchanged. Sulphuric acid also
produces brown spots but this is because any organic material has been
charred and therefore destroyed.
Selective visualisers include ninhydrin which reacts with amines, peptides and
amino acids to produce purple/pink compounds. Another reagent, fluorescein
can be used to convert hydrocarbons, barbiturates and many other compounds
into fluorescent derivatives which then fluoresce under UV light.
64
Equation (6.11)
Solvent front
Figure 6.28 - diagram of tlc plate to show distance to the solvent front to enable
calculation of the Rf value.
Continued on the next slide
65
66
Tlc - applications
The main disadvantage of tlc is that it does not readily lend itself to quantitative
analyses, unlike hplc or gc and thus is used mainly for identification purposes
(qualitative analysis)
67
Question 6.1 Sketch and label an idealised chromatogram. How can the resolution between
two peaks be evaluated from such a plot?
Question 6.2 What are the causes of peak broadening and which of the terms in the rate
equation take them into account?
The A, B and C terms may be defined in terms of a number of variables which the
chromatographer can control; such as column length, diameter and thickness of the
stationary phase and the nature of both the stationary and mobile phases.
In addition the mobile phase flow rate, as shown by the rate equation, also affects the
peak broadening.