Professional Documents
Culture Documents
Prior to sensitivity calibration lower angles will detect the target at a higher
amplitude than the higher angles due to a longer sound path and more
attenuation.
Without sensitivity calibration defects can be missed and all focal laws
cannot be used for code required amplitude based acceptance criteria such
as ASME section V and similar.
In the examples below, the MX2 general gain of 24.1dB is applied to all focal
laws in the group.
After completion of the sensitivity calibration wizard each focal law applies a
calibrated gain offset that is added to the general gain to achieve the
required sensitivity, typically 80% amplitude.
20mm
The target used for sensitivity calibration should be in the general range of
the area of interest for the inspection and specified in the work procedure.
A general rule of thumb for weld inspection is to use the closest calibration
target available that is at least 1.5 times the skip thickness in true depth.
SDH selection for sensitivity calibration that is too shallow or too deep
results in poor inspection results and may impede TCG construction.
6mm
12mm
After completing the group setup wizard process, position the probe so
that the calibration target can be seen in the S-scan. (15mm SDH from IIW
as pictured below)
Ensure that the range is sufficient for the target on all focal laws. (45-70
degrees)
Enter the wizard to calibrate the S-scan for sensitivity (ACG) by selecting
Wizard>Calibration>Type Ultrasound>Mode Sensitivity>All Angles.
Select start.
Enter the reference amplitude. (default 80%) All focal laws will be
adjusted to this amplitude by addition of a focal law gain offset after
sensitivity wizard completion.
Enter tolerance. The tolerance is a visual indicator for amplitude
verification after the sensitivity calibration correction. In the example
below a tolerance of 5% on an 80% calibration would place the window
from 77.5% to 82.5%.
Select Next.
5% tolerance window
Amplitude envelope
Select the last focal law to be calibrated for sensitivity. When there are no
obstructions from corners or adjacent calibration targets the entire sector scan can
be calibrated at one time.
Excluding a portion of the S-scan as pictured to the right will allow the calibration to
be completed in sections where the gate can be repositioned to avoid interference
from corners of the block or adjacent calibration targets if necessary.
Select Next.
Focal law
45-70
Focal law
45-60
Focal law
61-70
Set the start and width of gate A to ensure that the calibration target can be detected
by all focal laws in the S-scan.
Set the gate A threshold as low as possible to ensure the last focal law can be
detected. Gate thresholds below approximately 10% may have an amplitude error
beyond the tolerance when the focal law gain offset is calculated to correct to 80%.
Select Next.
Gate A
Gate A
Gate A
10
45
2dB
e
gr
de
2dB
es
2dB
70
deg
ree
s
2dB
11
2dB
Move the probe across the calibration block with consistent pressure and coupling.
As the calibration target travels through the gate of each focal law the green
envelope records the maximum amplitude detected.
The MX2 will calculate the required gain offset in dB for each focal law to correct to
80%. Focal laws with amplitude over 80% but less than 100% will receive a
negative gain offset.
Do not select Calibrate yet.
Continue to next slide.
12
The green envelope should be smooth and consistent without any radical changes
in amplitude between focal laws.
No focal law can be higher than 100% amplitude and the lowest focal law must be
higher than approximately 10% amplitude prior to selecting calibrate.
This ratio equates to approximately 20dB. The highest and lowest focal laws must
be within approx. 20dB of each other for the sensitivity wizard to function.
If the amplitude envelope cannot be completed with a <100% and >10% ratio the
angular range of the S-scan must be reduced. (Reduce S-scan from 30-75 degrees
to 45-70 degrees)
The longer the sound path and/or smaller the calibration target, the greater dB
difference between the first and last focal laws of an S-scan calibration.
If the green envelope was not created
properly select clear envelope and repeat the
probe movement.
Select Calibrate. After sensitivity calibration
a focal law offset is created in
UT>Beam>GainOffset for every focal law.
13
14
The gain offset for individual focal laws can be manually edited in
UT>Beam>GainOffset.
15
16
17