You are on page 1of 61

CMU200 Training

Mar 2010 | CMU200 training R&D | RSVN | 2

General Information
Specification
l

Temperature (Specification)
-

l
l
l

Operating Temperature Range : +5oC ~ +45oC


Storage Temperature Range : -25oC ~ +60oC
Humidity : 80% relative humidity

Frequency Range : 10MHz to 2.7GHz


Level uncertainty : <0.5dB (range of +5oC ~ +45oC, RF2 port)
For more detailed information, you can refer to the Specification material for
CMU200 from R&S website

(http://www.rohde-schwarz.com, Search keyword : CMU200).

Mar 2010 | CMU200 training R&D | RSVN | 3

Rolling of CMU200
CMU200 is a measurement instrument which designed to use for Developing, Producing
and Maintain the Cell phone.

All those Cell Phone was made based on the Specification like in below.
GSM/GPRS : ETSI(European Telecommunication Standard Institute)
WCDMA : 3GPP(3rd Generation Partnership Project)
For CDMA, AMPS and TDMA also having such a specification.

CMU200 was designed based on that specification, and do the measurement as


mentioned in that specification. Especially during the Call Testing, CMU200 working as a
Base Station, and do measuring.

Mar 2010 | CMU200 training R&D | RSVN | 4

3 kinds of CMU200
CMU200v02
Original CMU200
Can be used for all kind of use.

CMU200v30
CMU200v30 is identical to
Calibration procedure.
Only can be used for Calibration :
Can not perform Final Test

CMU200v10
Dedicated to Service Market.
Good for the place which not
a time critical field, like service center.

Mar 2010 | CMU200 training R&D | RSVN | 5

CMU Platform : User Benefit!!


Speed!
Up to ten times faster than previous
generations of cellular radio
instruments
Two-Channel concept plus additional
RF-Out & RF-In

Modularity!
CMU200 is designed to support 2G, 2.5G and 3G
Flexible option concept allows to minimize costs

Accuracy!
Guaranteed specifications over the entire frequency and dynamic range
of the instrument
Very high downlink accuracy with highest repeatability
Wide selection of highly accurate power measurements

Mar 2010 | CMU200 training R&D | RSVN | 6

Platform Features/Benefits

Mar 2010 | CMU200 training R&D | RSVN | 7

Platform Concept
HardwareOptions
HardwareOptions

SoftwareOptions
SoftwareOptions

CMUBasicUnit
CMUBasicUnit
CDMA
CDMA

Standardindependenttesting
Standardindependenttesting

cdmaOne
cdmaOne

WCDMA
WCDMA
HSDPA
HSDPA

cdma20001x
cdma20001x

HSUPA
HSUPA

1xEVDO
1xEVDO

GPRS
GPRS

1xEVDO
1xEVDORel.A
Rel.A

GSM
GSM

Bluetooth
Bluetooth

AMPS
AMPS

TDMA
TDMA

AGPS
AGPS

Mar 2010 | CMU200 training R&D | RSVN | 8

ECSD
ECSD
EGPRS
EGPRS

CMU200 Base Unit


Power
Supply

RF Front End

Front Module
Controller

Digital Board

RxTx-Board 1

Mar 2010 | CMU200 training R&D | RSVN | 9

CMU construction concept (HW)


Application-independent hardware platform
Some possible extensions:
l Up to two Signalling Units
-One universal for networks like GSM/IS136/AMPS/Bluetooth/WCDMA
-One dedicated for selected networks like CDMA (IS95)/CDMA2000
l Reference Board
l Audio board
l Baseband Interface
l Exchange DSP- Module(s) and/or additional add-on DSPs for

RX/TX and acquisition board(s)


l ....

Mar 2010 | CMU200 training R&D | RSVN | 10

CMU200 possible Extensions


Space for
WCDMA Layer 1
CoPro-Board

Audio Board
CMU-B41
Reference Board
CMU-B11/B12
Space for
future extensions

Signalling Unit 2nd


Space for Space for 2nd
CMU-B21
Signalling Unit I/Q-IF-Board RxTx-Board
Mar 2010 | CMU200 training R&D | RSVN | 11

Applications
l RF development
l Module design
l Module test in production
l Adjustment of mobile phones
l Final test in production
l Functional test
l Feature test
l High-end service
l Quality inspection
l Basis for test systems
l Base station simulation

Mar 2010 | CMU200 training R&D | RSVN | 12

HSDPA/HSUPA transmission rates


The data transmission rates in tests of WCDMA mobile phones extend into the megabit range
in the downlink (highspeed downlink packet access, HSDPA) and uplink (highspeed uplink
packet access, HSUPA). As always, all relevant RF parameters can be measured and
applications can also be tested in these two cases

Mar 2010 | CMU200 training R&D | RSVN | 13

IP based services

Mar 2010 | CMU200 training R&D | RSVN | 14

Protocol analysis
The R&SCMU-Z46 WCDMA message analyzer and recorder option allows all
universa terrestrial radio access network (UTRAN)protocol layers to be recorded,
which can then be used for more detailed analysis. This powerful tool permits indepth analyses, including transport layer analyses

Mar 2010 | CMU200 training R&D | RSVN | 15

Video telephony
l Video telephony is the most spectacular new

WCDMA application. It is unique in that it is cicuitswitched, and not IP-based like the previously
described applications.
l The test is performed in echo mode, where the

transmission and reception of video and audio


signals can be checked with just one mobile phone.
l The video telephony signals transmitted by the

phone to the R&S CMU200 are looped back by the


radio tester and displayed by the phone as would-be
video and audio signals of a called station

Mar 2010 | CMU200 training R&D | RSVN | 16

I/Q and IF interfaces for the CMU200


l The CMU-B17 option allows access to

analog I/Q and IF signals in both


communication directions (uplink and
downlink). Once a radio link has been
established, complex I/Q signals can be
applied or transmitted for further
analysis
l During normal operation without
access to I/Q or IF signals, the interface
module can be set to the by
pass mode
l Receiver tests under fading conditions
l Testing of mobile radio modules

Mar 2010 | CMU200 training R&D | RSVN | 17

Mar 2010 | CMU200 training R&D | RSVN | 18

Front Panel
Menu Select :

PRINT button

PCMCIA slot :

To change the Test


Mode of CMU200

SETUP button

For external
Storage device

Numeric Keypad,
On/Off switch

Variation button :
Change and set the
Parameter

Mar 2010 | CMU200 training R&D | RSVN | 19

RF port
Default : RF2

SYSTEM : SETUP
1. By Pressing SETUP button, you can go to SETUP display.

Mar 2010 | CMU200 training R&D | RSVN | 20

Setup Menus
1.

SETUP
1)
2)

3)
4)
5)

Print
:
Shows
the
parameter to set the
Printer configuration.
Remote : Configuration for
Remote Controlling. Can
set the GPIB address of
the instruments.
Comm. : Configuration for
RS232 port(in case of use
for Serial Connection).
Options : Shows the
Option configuration of
instrument.
Time : Time setting for
instrument.

Mar 2010 | CMU200 training R&D | RSVN | 21

Setup menus : Remote Menu


2.

Remote
1)
2)

3)

SCPI Connection : Remote


Configuration(GPIB or Serial)
Primary Address
:
GPIB
Address
for
instrument.
This address is used to
communicate with Controller
PC when test program is
running.
Others Test Program controls
automatically.

Mar 2010 | CMU200 training R&D | RSVN | 22

Setup menus : Options Menu


3.

Options
1)
2)
3)

Serial Number
SW Verson[Basesystem] :
Main FW version
Software Options/Hardware
Optons
List of the SW options and
HW options which installed.
Also can checks the
version of each SW options.

Mar 2010 | CMU200 training R&D | RSVN | 23

Option Configuration
For WCDMA/HSDPA*****)
1.

2.

HW options
1) CMU-B21v14 : Link Handler unit*)
2) CMU-B56 : Signaling unit for WCDMA signaling
3) CMU-B68 : L1 Unit for WCDMA signaling**)
4) CMU-U65 : Additional DSP unit
SW Options
1) CMU-K65, K66 : Signaling SW Option for WCDMA
2) CMU-K61~K63, K67~K69 : Signaling SW Option for WCDMA band.
3) CMU-K64, K60 : Signaling SW Option for HSDPA.
4) CMU-K56: Signaling SW Option for HSUPA
5) CMU-K48 : SW option for IQ vs SLOT measurement****)

*) Old Link Handler Board(CMU-B21) cant support WCDMA signaling.


**) Actually another HW option CMU-B66 is needed, but this was normally displayed as Included in CMU-B68.
***) CMU-K64 supports HSDPA signaling, up to 3.6Mbps data throughput, CMU-K60 supports up to 14 Mbps
****) CMU-K48 option is related to the calibration item so called Polar Modulator Calibration process. Unless the RX calibration will not work.
*****) To perform correctly for WCDMA Calibration procedure, you shouldnt use the FW version 4.2X series.

Mar 2010 | CMU200 training R&D | RSVN | 24

Mar 2010 | CMU200 training R&D | RSVN | 25

CMU200 - Home Menu


Hit [MENU select] key to recall the home menu screen

Mar 2010 | CMU200 training R&D | RSVN | 26

CMU200 - Using Function Group


l Function groups are dedicated to a

specific technology.
l A function group is a software option
(CMU-Kxx package).
l Each function group is independent from
each other.
l Function group contains 2 applications
(Non-Signaling and Signaling.)
l The Non-Signaling application combines
the CMU hardware and software resources
in a way that CMU behaves as general
purpose equipments (Power Meter, SA,
etc).
l The Signaling application combines the
CMU hardware and software resources in a
way CMU behaves as a base station
simulator.

Mar 2010 | CMU200 training R&D | RSVN | 27

CMU200 - Navigation
l FUNCTION Keys

[Menu Select] key recalls the CMU home


menu screen.
[DATA] function key recalls the SAVE and
RECALL screen
[CRTL] key is sometime used as wizard key
l SYSTEM keys
[Help] Help on line
[Setup] Instrument settings
[Print] Print or Hardcopy
l Keyboard and Knob
[Knob] Value variation in input fields and
parameters, line selection in tables, field
selection
in
popup
menus.
Press
to
expand/compress tables and pull-down lists and
to confirm entries and selections.
l Control keys
[Reset] Reset to default value

Mar 2010 | CMU200 training R&D | RSVN | 28

CMU200 - Navigate Between Environments


l

The Connection environment is the


environment where the user can access to
setup the CMU resources.
(e.g RF generator, RF Analyzer,signaling,
network)
l The Measurement environment is the

environment used by the


performing measurements.

CMU

when

l [Connect Control] tab allows to toggle

between the connection environment and the


measurement environment.

Mar 2010 | CMU200 training R&D | RSVN | 29

CMU200 Setting Up Measurement Object


l

In a measurement screen the tab below


the Connect Control tab is a tab to control
the measurement object.
l The tab is labeled with the measurement
type and on the left of the tab you can see
the status of the measurement.
l Hit twice the measurement tab to open the
configuration screen.
l Hit the [Control] tab
l Repetition set the measurement mode to
continuous or single shot.
l Stop condition setup when the DSP
measurement application will stop.
l Display Mode setup which mode will be
used to display the measurement results.
l Statistic Count set the number of sample
which are taking in account per
measurement .

Mar 2010 | CMU200 training R&D | RSVN | 30

CMU200 - Measurement Object Status

Mar 2010 | CMU200 training R&D | RSVN | 31

CMU - Icons

Mar 2010 | CMU200 training R&D | RSVN | 32

Menu Structure
MENU
SELECT

Connect
Control

Mar 2010 | CMU200 training R&D | RSVN | 33

Menu Structure

Mar 2010 | CMU200 training R&D | RSVN | 34

CMU200 Menu Structure (GSM / GPRS/ EGPRS)


Root(Menu)
Overview

Power

Modulation

Spectrum

RXQuality

SubMenu(Application)
P/TNorm.
GMSK
Ext.Phase
ErrorGMSK
Overview
8PSK

P/TGMSK

Overview8PSK

P/T8PSK

EVM8PSK

P/TMultiSlot

Magn.Error
8PSK

P/Frame
P/SlotGraph

Ext.PhaseError
GMSK

P/SlotTable

PhaseError8PSK

P/PCL
P/AccessBurst
Mar 2010 | CMU200 training R&D | RSVN | 35

Modulation
GMSK
Modulation
8PSK
Switching
GMSK
Switching
8PSK

BER
BERAverage
NeighborCells
BLER

CMU200 Menu Structure (WCDMA / HSDPA)


Root(Menu)
Power

Modulation

Spectrum

CodeDomianPower

RXQuality

SubMenu(Application)
Maximum
Power

Overview
WCDMA

ACLR
Filter

CDP/Code
Auto

BER

Minimum
Power

EVM
WCDMA

ACLR
FFT/OBW

CDP/Code
Manual

UEReport

OffPower
InnerLoop
TPC
P/SlotTable
On/OffTime
Mask

Magn.Err.
WCDMA
PhaseErr.
WCDMA
I/QAnalyz.
WCDMA

Emision
Mask

CDP/Code
RhoAuto
CDP/Code
RhoManual
CDP/Time
HSDPCCH
TimeMask

Mar 2010 | CMU200 training R&D | RSVN | 36

RLCBLER
HSDPA
Ack
HS
DPCCH
Logging

Mar 2010 | CMU200 training R&D | RSVN | 37

CMU200 : State Diagram for WCDMA(RMC)

Calibrationhavingdifferentwayforworking.
AlloftheFinalTestitemwasdoneinConnectedstate.

Mar 2010 | CMU200 training R&D | RSVN | 38

CMU200 : State Diagram for HSDPA

UE Attach
UE Dettach

Attach
UE Accept
Packet
Connectio
n

TocomparetoWCDMA(RMC)connection,HSDPAneed
onemorestate;ATTACH.
ButthisATTACHwasnormallydonewithRegistration.
ToconnectHSDPA,shouldusetheConnectUE(PS).
Mar 2010 | CMU200 training R&D | RSVN | 39

Network Configuration : MCC, MNC

This MCC, MNC code is


affect to Registration.
Basically this is the
default parameter as in
Test Specification, but
also its need to be match
to the SIM card.

Mar 2010 | CMU200 training R&D | RSVN | 40

Network Configuration : MCC, MNC

Secret Key also should be


matching to your SIM card.
If this is not correctly set as
the SIM card, it will disturb
the Registration.

Mar 2010 | CMU200 training R&D | RSVN | 41

Network Configuration : MCC, MNC

This Parameter should be


set as Output Channel
Power.
Unless you should calculate
this everytime to set the
correct Downlink Power
Level.
Mar 2010 | CMU200 training R&D | RSVN | 42

Network Configuration : MCC, MNC

Mar 2010 | CMU200 training R&D | RSVN | 43

CMU200 Parameter Setting

Mar 2010 | CMU200 training R&D | RSVN | 44

Mar 2010 | CMU200 training R&D | RSVN | 45

Screen Capture : Make Screen Capture


1. By using PRINT button, load the PRINT popup menu.

Mar 2010 | CMU200 training R&D | RSVN | 46

Screen Capture : Make Screen Capture


2. Set the Destination on PRINT Popup menu.
1) Printer : Print the screen with Printer.
2) Internal/Internal WMF : Store the picture file at internal
HDD.
3) External/External WMF : Store the picture file at external
storage material.
Items

Printer

Internal

Internal WMF

External

External WMF

Mar 2010 | CMU200 training R&D | RSVN | 47

Screen Capture : Make Screen Capture


3. Set the File name.
1) If you dont put a specific name, CMU200 will use file name
automatically such as PRINT00.WMF, PRINT01.WMF.

4. Store the picture by OK button.

Mar 2010 | CMU200 training R&D | RSVN | 48

Screen Capture : External / External WMF


1. Using external Storage material.

Use the PCMCIA Memory card.


You should have a PCMCIA Memory card.
Choose the External WMP for Popup menu, and press
OK.

PCMCIA Port

Mar 2010 | CMU200 training R&D | RSVN | 49

Screen Capture : Internal / Internal WMF


2. How to copy with external storage device?
1) Put a Memory Card into
the PCMCIA slot.
2) DATA
Arrange
3) Choose the directory
PRINT

Mar 2010 | CMU200 training R&D | RSVN | 50

Screen Capture : Internal / Internal WMF


4) Choose the file
in the Print directory
5) Press Copy

6) Locate the cursor


to External
7) Press Paste

Mar 2010 | CMU200 training R&D | RSVN | 51

Screen Capture : Internal / Internal WMF


3. Internal WMP

First, you should store the picture file manually.


Need to copy that picture file to PC.
Way to copy
GPIB : use Remote Service Tool Program.
PCMCIA Memory card : copy by memory card.

< Remote Service Tool >

Mar 2010 | CMU200 training R&D | RSVN | 52

Screen Capture : Internal / Internal WMF


4. How to copy to PC?
1) Run
Remote Service Tool
Program
2) Application File Copy
3) Choose Print Directory
At right hand side.
(Userdata Print)
4) Choose the file and copy
to PC

Mar 2010 | CMU200 training R&D | RSVN | 53

Mar 2010 | CMU200 training R&D | RSVN | 54

Firmware Reset
1. Soft Reset

Use the RESET button on the front panel.

Mar 2010 | CMU200 training R&D | RSVN | 55

Firmware Reset
2. Soft Reset

Will display the screen regarding RESET.

Mar 2010 | CMU200 training R&D | RSVN | 56

Firmware Reset
1. By Version Manager
1)

2.

This RESET will make all the parameters to be default state.

How to?
1) Loading Version Manager
I. After booting CMU200, You
can see the Logo of CMU200
with Beep sount.
II. At this moment, press
Menu Select button.
2) After loading of Version Manager,
run the Firmware Update after
board Change.
3) Then remaining process
automatically done.

Mar 2010 | CMU200 training R&D | RSVN | 57

Mar 2010 | CMU200 training R&D | RSVN | 58

Self Calibration & Maintenance


1. Menu

Menu Select

Basic Functions

Mar 2010 | CMU200 training R&D | RSVN | 59

Base

Self Calibration & Maintenance


1. Selection of Maintenance Item
2. By On/Off button, run the Maintenance Item.

Mar 2010 | CMU200 training R&D | RSVN | 60

Thank you for your attention

Mar 2010 | CMU200 training R&D | RSVN | 61