Prepared by:
Ronak vasava
Endoscopy
GI Tract
Endoscopy—Using an
instrument to
visually examine the
interior of a hollow
organ or cavity
Gastrointestinal
Endoscopy
Methods of GI Endoscopy
Upper Intestinal Endoscopy
Colonoscopy
X-Rays
CT Scans
MRI
Surgery
Capsule Camera is the Solution
Patient feel discomfort by
above mentioned
techniques
Sensor Array
Data Recorder
Belt
RAPID Workstation
Features
Size:2.5 cm long and 1.1 cm wide
Dimension :11 mm * 26 mm
Weight: 4 gm
Frame Rate: 2 images/second
50,000 images are obtained during an 8 hour exam
Magnification: 8x
Capsule coating: non-adherent
Disposable
Price: US$450/ each
Power Source:Battery
Made in : ISRAEL
Inside the capsule
Schematic diagram
Functions
Optical Dome:Allows tissue to be in focus even if
they are to be in contact
Lens:Used to increase the Depth of the field
CMOS Image Sensor:Detects the images
White LEDS:To illuminate the area around the
capsule
Batteries:Power the capsule
ASIC:Allows the integration of video transmitter
of sufficient Power output ,Efficiency,bandwidth
of small size into capsule
Block Diagram
LIGHT
SOURCE
(LED)
DATA
RECO-
VIDEO RF RDER
CMOS DAC TRANSMITTER
IMAGE
SENSOR
BATTERY MICRO-
CONTROLLER
Continue
Combination of 3 technologies: Improvement of
signal to noise ratio,White LED development,
ASIC circuits development
CMOS Detectors: Active Pixel Resolution ,Each
pixel is equipped with a buffer amplifier
White Color LEDS distinguish disease tissue by
color
Analog Transmission:Requires high data rate
with lower frequency , so DAC is used
Analog Video transmitter transmits signal at a
30-800 MHz frequency
Sensor Array
Several Wires are
attached to the abdomen
like ECG leads to obtain
images by radio
frequency.
These wires are
connected to a
lightweight data recorder
worn on a belt.
Used to calculate and
indicate the position of
capsule in the body.
Data Recorder
Data recorder receives and
records signals transmitted
by the camera to an array
of sensor placed on the
patient’s body.
It is of the size of walkman
It recieves and stores
50000 to 60000 JPEG
images on a 9 GB hard
drive
Images take several hours
to download through serial
connection.
Belt
A patient wear a
receiver belt around
his or her waist over
clothing
A belt is applied
around the waist and
holds a recording
device and a battery
pack
RAPID Workstation
Reporting and Processing
of Images and Data
Image data from the
data recorder is
downloaded to a
computer equipped with
software called RAPID
Application software
Helps to convert images
into a movie and allows
the doctor to view the
color 3D images .
Fantastic Voyager
Fantastic Voyager
Clinical Image
Small Intestine Imaging in
3 Easy Steps
1. Ingesting the M2A® capsule
Stomach : 30 to 60 minutes
Questions ?