Professional Documents
Culture Documents
Yanna Indrayana
EVOLUTION
Last generations
Smaller, last
longer, MRI
Late 1920s compatible, self
1958
early 1930s 1990s adjustment,
First 1970s
First external Microprocessor- internal
pacemaker Passive and active
cardiac pacemaker driven pacemaker defibbrilator
implantation fixation
Right Atrium
Right Ventricle
The Pulse Generator
Contains a battery to
provide energy for sending
electrical impulses to the Connector
heart Block
Houses the circuitry that Circuitry
controls pacemaker
operations
Battery
Includes a connector to
join the pulse generator to
the lead(s)
Leads
Insulated Wires
Deliver electrical impulses Lead
-
conduction is lost
Single ventricular pacing does not provide AV synchrony
Dual Chamber Pacing
AV synchrony
- High skill
The pacing stimulus
The pacemaker delivers the electrical current between two points, called electrodes.
UNIPOLAR BIPOLAR
Anode
+
Anode
+
When pacing, the impulse stimulates large area of When pacing, the impulse flows through the tip electrode
the body between the tip of the lead and the pulse located at the end of the lead wire
generator Pacing artifacts may be very difficult to see on the surface
Usually exhibit larger pacing artifacts on the surface ECG
ECG May have a larger diameter lead body than unipolar leads
Electrical Concepts
V=IxR
I=V/R
R=V/I
Programming :
Two settings to ensure capture :
Amplitude (V)
The amplitude of the impulse must be large
enough to cause depolarization / to capture the
heart
The amplitude of the impulse must be sufficient
to provide an appropriate pacing safety margin
x3
1.5 V
Lead impedance
Amplitude and pulse width setting
Percentage paced vs. intrinsic events
Rate responsive modes programmed ON
Concepts in pacemaker function
Pacin
goutput of electrical current for depolarizing the cardiac tissue in adjacent of the lead
resulting propagation of a wave of depolarization throughout that chamber
Sensin
gresponse of a pacemaker to intrinsic electrical activity (intrinsic heartbeats)
VVI / 60
Concepts in pacemaker function
Inhibition of the
output
Pacemaker can inhibit pacing if it senses intrinsic activity pacemaker will not deliver a
stimulus if it senses an intrinsic beat at the proper time
Triggered pacing
Pacemakers deliver a pacing stimulus whenever intrinsic activity is sensed. Most often
used in dual chamber pacemakers
Triggered pacing. (a) and (b) both show atrial sensing, and ventricular pacing, in a tracking mode. The
pacemaker settings are the same in both panels. The difference is that in (b) the intrinsic atrial rate is faster.
Concepts in pacemaker function
Rate
responsiveness
Pacemakers programmed to vary the pacing rate in response to the patients level of
activity
AV delay
Time interval between an atrial
paced or sensed event, and the
delivery of a ventricular
pacingstimulus analogous to
the intrinsic PR interval
The Revised NASPE/BPEG Generic Code for Antibradycardia, Adaptive-Rate, and Multisite Pacing, 2002
Common Pacemaker Modes
Response Programmability
Pacing Sensing to Sensing & Rate Response
V O O
- Ventricular pacing
- No sensing
- No response to sensing
Common Pacemaker Modes
Response Programmability
Pacing Sensing to Sensing & Rate Response
V V I
- Ventricular pacing
- Ventricular sensing
- Inhibited when sensing a ventricular event
Common Pacemaker Modes
Response Programmability
Pacing Sensing to Sensing & Rate Response
V V I R
- Ventricular pacing
- Ventricular sensing
- Inhibited when sensing a ventricular event
- Rate response capabilities
Common Pacemaker Modes
Response Programmability
Pacing Sensing to Sensing & Rate Response
D D D
Response Programmability
Pacing Sensing to Sensing & Rate Response
D D D R
I IIa IIb III Pacing is not indicated in patients with AV block which is
due to reversible causes
C
I IIa IIb III In patients with sinus rhythm, dual-chamber PM should be
preferred to single chamber ventricular pacing for avoiding
A PM syndrome and improving quality of life.
Indication for pacing : Permanent AF and AVB
Cause :
Inappropriately programmed sensitivity
Lead dislodgment
Lead failure: Insulation break; conductor fracture
Lead maturation
Change in the native signal
Oversensing
Sensing of an inappropriate signal can be physiologic or nonphysiologic
Cause :
Lead failure
Poor connection at connector block
Exposure to interference
Lost of capture
No evidence of depolarization after pacing artifact
Cause :
Poor connection at connector block
Lead failure
Battery depletion
Circuit failure
Pacemaker syndrome
The occurence of overt symptoms due to adverse hemodynamic impact
from the loss of AV synchrony and occurence of ventriculoatrial
conduction or atrial contraction againts closed AV valves in patient
with an implanted pacemaker
Symptoms include:
Dizziness Confussion
Presyncope Malaise
Chest tightness Fatigue
Shortness of breath
Cause :
Loss of capture Single chamber system
A-V intervals of long duration Absence of rate increase with exercise
Onset of 2:1 block
Thank you...