Objectives continued…
3
Mean = average of data = 𝑋ത
SD =
5
Coefficient of Variation (CV)
8
Defining QC ranges
+3 SD
+2 SD
+1 SD
MEAN
-1 SD
-2 SD
-3 SD
10
SD limits too large!
11
SD limits too large!
↓↓ QC failures
+3 SD
+2 SD
+1 SD
MEAN
-1 SD
-2 SD
-3 SD
12
SD limits too small !!
13
SD limits too small !!
↑↑ QC failures
+3 SD
+2 SD
+1 SD
MEAN
-1 SD
-2 SD
-3 SD
14
What are acceptable QC Values?
15
Why not use the manufacturer’s QC limits?
• Manufacturer’s limits are often 2-3 times
too large –Not sensitive to your laboratory
conditions
• They are general guidelines that include
several different instrument/method types
• If the QC range is too large you will not
find problems
16
Lactate U/L Roche Cobas C700
5.376
4.928
Lactate Value
Mean = 4.48
SD = 0.448 4.48
CV = 10%
4.032
3.584
3.136
0 5 10 15 20 25
QC run
Lactate U/L Roche Cobas C700
4.92
4.7
Mean = 4.48
Lactate Value
4.48
SD = 0.22
CV = 4.9% 4.26
4.04
3.82
0 5 10 15 20 25
QC Run
How do I determine the SD limits
that are correct?
24
Glucose Example of 1 month glucose QC data
1 88
2 89
3 86
20 data points of new QC Material
4 84 93
5 89
6 90
92
7 92 91
8 87
90
glucose value
9 86
10 91 89
11 88
12 87
88
13 86 87
14 84
15 89
86
16 86 85
17 87
18 90
84
19 91 83
20 88 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Average 87.9 Each QC Assay
SD 2.2
CV 2.49 SMILE 25
Track CVH over time
Normal
Jan 2.1
Feb 2.5
Jul 2.2 2
h
Aug 2.1
%CV
1.5
Sep 2.2
Oct 2.5 1
Nov 2.4
0.5
Dec 2.5
0
CVH 2.34 Jan Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec
26
Monitor CVH to alert for problems
Increasing CV
Jul 2.9
%CV
2
Aug 3.1
1.5
Sep 3.2
1
Oct 3.4
0.5
Nov 2.5
0
Dec 2.3
1 2 3 4 5 6 7 8 9 10 11 12
CVH 2.64
Month
SMILE 27
Monitor CVH to alert for problems
Random CV
Jan 2.1
Feb 2.3
12 month plot CVH with
Mar 2.2
4 SPIKE
Apr 2.5
3.5
May 2.2
3
Jun 2.3
2.5
Jul 2.2
2
CVH
Aug 2.1
1.5
Sep 3.4
1
Oct 3.6
0.5
Nov 2.3
0
Dec 2.2
1 2 3 4 5 6 7 8 9 10 11 12
CVH 2.45
Month
SMILE 28
Things that increase your CVH
– Day to day instrument differences
– Electrical and power quality
– Different persons operating the
instrument
– Different reagent lots
– QC material preparation
– Reagent Quality
29
How do you determine if your
CVH is an acceptable value?
30
Standard 1:
Instrument/Method
manufacturer’s value
• The instrument manufacturer determines
and publishes the instrument/reagent
method CV (precision)
• If you can not achieve the precision (CV)
that the manufacturer claims on your
instrument, contact the manufacturer for
service
31
Standard 2:
The External Quality Control (EQA)
survey method CV
• CAP & Accutest (OWA) materials are
considered an External Quality Assurance
(EQA) quality indicator. (Between labs)
• EQA providers publish instrument/method
peer CV data with survey results. Your lab
CVH should be lower than the CVEQA
published
32
Calculating CVEQA
34
Demonstration of establishing
sensitive SD limits using CVH
35
Establishing your new mean
1. Ensure that your old lot of QC material is
running inside of your current range with
no bias, shifts or trends
2. Run new normal QC material for at least
20 data points with old QC material for at
least 5 days. Ensure that your old QC
material is within acceptable range for
each run.
3. Calculate SD, MEAN & CV from data
4. Is the CV ≤ CVH and CVMAN? 36
20 data points of Normal QC data -Glucose
1 88
2 89 94.5
3 86
4 84
5 89 92.3
6 90
7 92
90.1
8 87
Glucose Value
9 86
10 91
87.9
11 88
12 87
13 86 85.7
14 84
15 89
16 86 83.5
17 87
18 90
19 91 81.3
20 88 0 2 4 6 8 10 12 14 16 18 20
MEAN 87.9 QC Run
SD 2.2
CV 2.55%
37
Example normal glucose QC data
MEAN = 87.9
SD = 2.2
CV = 2.55 from new precision data
38
Example normal glucose QC data
39
Use CVREF to calculate SD limits
SD = MEAN x (CVREF/100)
SD = 87.9 x (3.0/100)
SD = 2.6
40
QC range limits are defined by CVREF
95.7
2.6
3 SD
units
93.1
2 SD
90.5
1 SD
87.9 MEAN
-1 SD
85.3
-2 SD
82.7
3 SD
80.1
41
Patient Safety Monitoring in International Laboratories (SMILE)
42
Patient Safety Monitoring in International Laboratories (SMILE)
References