You are on page 1of 101

Flow Doctor Program

Level 1
Introduction to Product Data
Sheets

Module 108
Learning Objectives
When you finish this module, you will be able to do
the following :

 Understand and explain the performance specifications


for Flow and Density products

 Know what product information must be committed to


memory

 Know where to find the general product information most


often requested by customers
Why Should You Learn This Material ?
Customer Expectations
 Sales person can quickly provide general information
about products
– Provides customer with confidence
– Supports our claim as the leader in Flow and Density
knowledge and expertise

 Sales person must be able to speak to a certain level


about our products without having to use reference
materials
Micro Motion
General Sensor Specifications
Product Data Sheet
“Specification Summary”
Sensor Attributes to Commit to Memory
 Base Flow Accuracy (Liquid/Gas)
– +/- 0.10% of Rate ( +/- 0.35% of Rate)
– +/- 0.05% of Rate for Liquid Only (Optional)
 Base Density Accuracy
– +/- 0.0005 gm/cc
– +/- 0.0002gm/cc(Optional)
 Base Model Names & Nominal Sizes
– CMF010, 025, 050, 100, 200, 300, 400 (1/8” to 6”)
 Wetted Materials Available
– 316L SS, 304L SS & Nickel Alloy
 Operating Temperature Limits (Standard Sensors)
– -400 to 400 oF (-240 to 204 oC)

Note : All Examples are for Standard ELITE Sensors


Sensor Attributes to
Commit to Memory – Cont…
 Max Operating Pressure Rating (Standard Sensors)
– Varies from sensor model and tube material (refer to updated product
data sheets)
 Most Common Process Fittings Available
– WNRF(150#, 300#, 600#), Union, Sanitary, DIN, Wafer
 Case Options
– Standard Secondary Containment, Optional-Purge Connections or
Rupture Disks
 Hazardous Area Classification - Sensors
– UL & CSA Class I, Div 1, Groups C & D (Memorize)
Class I, Div 2, Groups A, B, C & D
– SAA Meets (Ref. PDS for Specifics)
– ATEX Meets (Ref. PDS for Specifics)

Note : All Examples are for Standard ELITE Sensors


Micro Motion High
Precision Multivariable
Sensors

Elite (CMF)
F-Series
H-Series
T-Series
Micro Motion ELITE® Series
• Highest Flow and Rangeability
• Highest Density Performance
• Unsurpassed Immunity to Field Effects such as
Pressure, Temperature, and Vibration
• Secondary Pressure Containment is Standard
• Nominal Line Sizes from 1/8” to 6” (3 to 150 mm)
Models : CMF010, CMF025, CMF050, CMF100,
CMF200, CMF300, CMF400
BRIEF SPECIFICATIONS :
Liquid Accuracy +/-0.10% of Flow Rate (0.05% of Rate Optional)
Gas Accuracy +/-0.35% of Flow Rate (Legacy +/-0.5%)
Density Accuracy +/-0.0005 g/cc (0.0002 gm/cc Optional)
Wetted Parts 316L Stainless Steel or Nickel Alloy
Temp. Rating -240 to 204 C
Pressure Rating 100 Bars for SS and 148 Bars for HY Sensors
Micro Motion F-Series
• Flow Accuracy to 0.2% on Liquids and 0.5% on
Gases, with 0.15% and 0.1% Accuracy Option
Available on Liquids with MVD Transmitters
• Optional Secondary Containment
• Nominal Line Sizes from ¼” to 3” (6 to 75 mm)
Models : F025, F050, F100, F200, F300

BRIEF SPECIFICATIONS :
Flow Accuracy +/-0.20% of Flow Rate
Gas Accuracy +/-0.50% of Flow Rate (Legacy +/-0.7%)
Density Accuracy +/-0.002 g/cc (+/-0.001g/cc Optional)
Wetted Parts 316L Stainless Steel / Hast - C for F300H
Temperature Rating -100 to 180 C
Pressure Rating - 100 Bars / High Pressure Models also Available
(F025P, F050P and F300P)
Micro Motion H-Series
• Flow Accuracy to 0.15% on Liquids and 0.5% on Gases
• 3-A Authorized & EHEDG Approved
• 32 Ra Internal Surface Finish is Std (15 Ra Optional)
• Smooth, Easy-to-Clean Exterior Finish
• Optional Secondary Containment
• Nominal Line Sizes from ¼” to 3” (6 to 75 mm)
Models : H025, H050, H100, H200

BRIEF SPECIFICATIONS :
Flow Accuracy +/-0.15% of Flow Rate (option +/- 0.10% of flow rate)
Gas Accuracy +/-0.50% of Flow Rate
Density Accuracy +/-0.002 g/cc (option +/- 0.001 g/cc)
Wetted Parts 316L Stainless Steel
Temperature Rating -100 to 180 C
Pressure Rating 100 Bars
Micro Motion T-Series
• Best Single Straight -Tube Design
• 3-A Authorized and EHEDG Approved
• Improved Surface Finish Option 20 Ra or Better for
Hygienic Applications (Standard is 32 Ra)
• Optional Display
• Secondary Pressure Containment of 50 Bars is Standard
• Nominal Line Sizes from 1/4” to 2” (6 to 50 mm)
Models : T025, T050, T075, T100, T150

BRIEF SPECIFICATIONS :
Flow Accuracy +/-0.15% of Flow Rate
Density Accuracy +/-0.002 g/cc
Wetted Parts Titanium ASTM Grade 9
Temperature Rating -50 to 150º C
Pressure Rating 100 Bars
Review Questions :
 Name the “High Precision, Multivariable MMI Sensors” ?
 What are 4-5 Main Attributes of an Elite Sensor ?
 Which Sensor Series are Supplied with Standard Secondary
Containment ? Which have Optional Secondary Containment ?
 Which Sensor Line(s) Meet All of the Requirements of 3-A and
EHEDG ?
 What are Model Base Model Numbers of F-Series & H-Series
Sensors ?
 Which Sensor Line has a Base Accuracy for Liquids of 0.1% ?
 What is the Density Accuracy of All CMF Sensors ?
Micro Motion General
Purpose, Flow-Only
Sensor
R- Series
Micro Motion R-Series
• Simple, Reliable Flowmeter
• Optional Integrally-Mounted Electronics
and Display
• Nominal Line Sizes from 1/4” to 3” (6 to 75 mm)
Models : R025, R050, R100, R200

BRIEF SPECIFICATIONS :

Flow Accuracy +/-0.5% of Flow Rate


Gas Accuracy +/-0.75% of Flow Rate (Legacy +/-1%)
Wetted Parts 316 Stainless Steel
Temperature Rating -40 to 125º C
Pressure Rating 100 Bars
Micro Motion
Specialty Sensors
High Temperature Sensors
High Pressure Sensors
High Capacity Sensors
CNG Sensor
Model DL
Tefzel Lined Sensors
High Temperature Sensors ( > 200 degC)
 CMF200A, CMF300A ,
F300A & CMF400A – Max
Temp. Rating 350 DegC
(427 DegC Optional on
Certain Models – Consult
Factory)

 Mat’l. of Construction (CMF


HT) - SS316L & Hast-C

 DT065,DT100 & DT150 –


Max. Temp Rating 427DegC
High Pressure Sensors ( > 150 bar)

– CMF010P, F025P,
F300P, R025P, DH150
and DH300

– Pressuring Rating from


158 Bar to 413 Bar

– Refer PDS for


Individual Rating
CNG Sensor
• Designed to Meet Compressed Natural Gas Industry
Standards
• For Use in Automobile or Light and Heavy Duty Vehicle
Dispenser Design
• NTEP and OIML Approved
• 1/2” (12 mm) Nominal Line Size
Models : CNG050

BRIEF SPECIFICATIONS :
Flow Accuracy ±0.5% of Batch from 1 to 100 kg/min
Max Std Volume Flow Rate 4444 SCFM (7550 Nm3/hr)
170 lb/min

Wetted Parts 316L Stainless Steel


Temperature Rating -40 to 125º C

Pressure Rating 345 Bars


D-Series …“High Capacity Sensor”
 High Flows
– (-) 240 to 204°C, 43 bar, 1362 ton/hr

 Custody Transfer Approvals


– Singapore, China, India, Indonesia, Malaysia,
Thailand, Philippines, etc.
 Accuracy
– Liquid : 0.1% of Rate (flat)
– Gas : 0.65% of Gas (flat)
– Density : 0.0005 gm/cc
Model DL
• Continuous Single Tube Design
• Approved for Use in 3-A Sanitary Applications
• Self Draining
• Nominal Line Sizes from 1/4” to 2” (6 to 50 mm)
Models : DL65, DL100, DL200

BRIEF SPECIFICATIONS :

Flow Accuracy +/-0.15% of Flow Rate


Density Accuracy* +/-0.001 g/cc
Wetted Parts 316 Stainless Steel
Temperature Rating -240 to 204 C
Pressure Rating* up to 103 Bars
Tefzel Lined Sensors
 Tefzel Lined Sensors
Available in D Series
Sensor

 Models : DS150Z and


DS300Z

 Refer PDS for Detail


Specification
LF Series Flow Meters
 Low Flow Meters
 Model : LF2M, LF3M & LF4M
 Flow Range
– LF2M 0.001 to 0.948 Kg/Hr
– LF3M 0.01 to 3.307 Kg/Hr
– LF4M 0.1 to 60.845 Kg/Hr
 Flow Accuracy : 1.0% of Rate
(0.5% of Rate – Optional)
 Liquid Density Accuracy :
0.005gm/cc
Micro Motion Density
Specifications
Coriolis Density Meters
Micro Motion Coriolis Sensors
Density Performance
 MMI coriolis meters only measure densities of
liquids and slurries… not gas
 Density accuracy of sensor is often dependent
on the sensor/transmitter combination
 Dependent on :
– Sensitivity to changes in tube frequency relative to
changes in density
• Increasing tube wall thickness – deceases density accuracy
High pressure vs. std. pressure sensors
– Precision of calibration performed and ability of
transmitter to resolve signal
Micro Motion Coriolis Sensors
Density Specifications
 Accuracy expressed as: + g/cc value
– ELITE
• All sensors (except high press) = + 0.0005 g/cc (+ 0.5 kg/m3)
• CMF010P (High Pressure) = + 0.0020 g/cc (+ 2.0 kg/m3)

 What does the density spec. mean ?


– An ELITE sensor indicating a density reading of
1.0505 g/cc may actually have a fluid passing through
the senor with a density range of:
• 1.0510 g/cc to 1.0500, a spread of 0.0010 g/cc
Micro Motion Density
Specifications
Tube and Fork Densitometers
Micro Motion Densitometers
- formerly from Solartron
Complementary Products that Solartron brings :
1. Tube Made of NiSpanC for Fiscal Density Applications
2. High Accuracy Straight Tube Densitometer
3. Forks - Insertion Density for Pipelines and Tanks
4. Gas Density - Line Density or SGU
Micro Motion Tube and Fork
Density Meters
– Tube Made of NiSpan C : Meets API, Fiscal Metering
Standards

– Density Frequency Output for Flow Computer Connection

– Single Straight Tube Densitometer


• Low Pressure Drop for By-Pass Installation
• Easy to Clean
• Shear Sensitive Fluids

– Forks : Insertion Density for Pipes up to 48” and Tanks

– New Wetted Materials (Monel and soon Zirconium)

– Gas Density
Micro Motion Solartron
Tube Densitometers

Type Material Features

7835 Ni Span C High accuracy

7845 316L St. St General purpose

Chemical
7846 Hastelloy
resistance

7847 316L St. St Hygienic


Micro Motion Solartron Tube
Densitometers Specification – cont’d
 7846
– Hastelloy C tube material
– Base Accuracy ±0.00035 g/cc over density range 0.6 -1.2 g/cc
– Base Accuracy ±0.00010 g/cc with UKAS calibration (limited temp
range – 30oC)

 7847
– Hygienic Densitometer
– Stainless steel 316 tube material
– Base Accuracy ±0.00035 g/cc over density range 0.6 – 1.2 g/cc
– Base Accuracy ±0.00010 g/cc with UKAS calibration (limited temp
range – 30oC)
Micro Motion Solartron
Tube Densitometers Specifications
 7835
– Ni-span C tube material
– Base Accuracy ±0.00015 g/cc
– Base Accuracy ±0.00010 g/cc with UKAS calibration

 7845
– Stainless steel 316L tube material
– Base Accuracy ±0.00035 g/cc
– Base Accuracy ±0.00010 g/cc with UKAS calibration
(limited temp range – 30oC)
Micro Motion Solartron Tube
Densitometers Specifications Summary
Parameter Type 7835 Type 7845/46/47
Accuracy ± 0.00015 g/cc ± 0.00035 g/cc
Density range 0-3 g/cc 0-3 g/cc
Factory calibrated range 0.3-1.1 g/cc 0.6-1.6 g/cc
Repeatability 0.00002 g/cc 0.0001 g/cc
Temperature effect ± 0.000005 g/cc/°C ± 0.00005 g/cc/°C
(± 0.000003 g/cc/°F) (± 0.00003 g/cc/°F)
Max operating pressure 150 bar 100 bar (1450 psi) - 7845
or flange limit (2175 psi) 50 bar (725 psi) – 7846
20 bar (290 psi) - 7847
Temperature range -50 to +110 °C 50 to +160 °C
(-58 to +230 °F) (-58 to +320 °F)
Wetted parts Ni-span C & 316L SS 316L SS & Hast C (7846)
Micro Motion Solartron Densitometer
- Short & Long Stem Forks

Long Stem
Short Stem w/ Flange
Short Stem
w/ Cone Fitting w/ Flange
Comparison of Performance
- w/ Short & Long Tines

7826 7828
Long Tines Short Tines

500 cp 20000 cp
Viscosity Limit Viscosity Limit
7828 7826 More tolerant to Less tolerant to
entrained gas in entrained gas in
the liquid the liquid

Standard or
Advanced
7828 Advanced
Amplifier Only !
w/PTFE Coating Amplifier
Fork Wetted Materials

316 L SS
304 SS
Hastelloy B
Hastelloy C
Monel 400
Titanium
Zirconium (in the works)

New Opportunity

MMI Proprietary/Confidential
Board Meeting – 1/11/06
Coating and Electro Polish

 PTFE (Teflon) Coating is available


– Not intended for corrosion resistance
– Helps prevent buildup of sticky products

 Electro Polish is available


– Also helps prevent buildup of sticky products
– Unlike PTFE, less susceptible to erosion

7828
w/PTFE Coated
Tines
Micro Motion Fork Densitometer
Specifications (Short Stem)
Long Tines Short Tines

Parameter Type 7826 & 7828


Density Range 0 - 3g/cc (0 - 3,000kg/m3)
Calibrated Range 0.6 - 1.25g/cc (600 - 1,250kg/m3)
Density Accuracy ± 0.001 g/cc (± 1kg/m3)
Repeatability ± 0.0001 g/cc (± 0.1kg/m3)

Temperature Range (-58 to +390°F) -50 to +200 °C

Max Pressure Range (3000 psi) 207bar


Viscosity Range 0-500cP (7826) & 0-20,000cP (7828)
Long Tines Short Tines
Environmental Rating IP66
Safety (All Units) EEx d IIC T4, Class 1 Div 1 Group C
Micro Motion Solartron Fork Densitometer
Specifications (Long Stem)
Long Tines Short Tines
Parameter Type 7826 & 7828
Density Range 0 - 3g/cc (0 - 3,000kg/m3)
Calibrated Range 0.6 - 1.25g/cc (600 - 1,250kg/m3)
Density Accuracy ± 0.001 g/cc (± 1kg/m3)
Repeatability ± 0.0001 g/cc (± 0.1kg/m3)
Temperature Range (Safe Area) (-58 to +390°F) -50 to +200 °C
Temperature range (Hazardous Area) (-58 to +300°F) -50 to +150 °C
Max Pressure range (1450 psi) 100bar
Viscosity range 0-500cP (7826) & 0-20,000cP (7828)
Long Tines Short Tines
Environmental rating IP66
Safety (all units) EEx d IIC T4, Class 1 Div 1 Group C
Micro Motion Density Performance

+/- 0.0001 - 0.0002 g/cc


Best Real World Performance
– 7835 NiSpanC
– 7845 316SS with UKAS Calibration Calibration Traceability
– Enhanced Elite
Density&Flow Full stream
+/- 0.00035 – 0.0005 g/cc Hygienic Net Flow
– Elite
– 7845 316SS Single Path

+/- 0.001 g/cc Insertion Easy to Install


– F Series Monel
– Forks
Micro Motion Gas Density and
Specific Gravity Products
Micro Motion Solartron Gas Densitometer
- Model 7812
 Product Range – 7812 Gas Density sensor
 Operating Principle – Vibrating Cylinder in fixed
volume
 Model 7812 Gas Densitometers
– Ni-span C cylinder material
– NACE compliant pressure retaining parts
– 4 density ranges (1.5-10kg/m3, 9-90kg/m3, 25-250kg/m3, 40-400kg/m3)
– Fiscal/Custody transfer market
– Accuracy ±0.1% on Nitrogen, ±0.15% Natural Gas
– Traceable calibration with UKAS calibration available
– Dedicated Ethylene Gas densitometer version
Micro Motion Solartron Gas Specific Gravity
- Model 3098
 Product Range – 3098 Specific Gravity sensor
 Operating Principle – Vibrating Cylinder in fixed
volume, P & T controlled to reference levels
 Model 3098
– Ni-span C cylinder material
– SG range 0-5 (typical)
– Accuracy up to ±0.1%
– Fiscal/Custody transfer markets
– Customer calibrates unit
Power and Temperature
Specifications
Micro Motion Coriolis Meters
Power Consumption
 All Elite meters with core processors
– 3 Watts

 Power is supplied by the transmitter or with MVD


Solo, the power is supplied to the core processor.

 Common Question : Is the meter loop powered ?


- No.
Temperature Specifications
Temperature Accuracy :
 Refers to the ability of the sensor to continuously
read the process fluid temperature from the RTD
attached to the sensor.
 Accuracy Statement :
+ 1oC + 0.5% of reading in oC
Repeatability :
+ 2oC (all models)
Temperature Specifications
 Why does Coriolis need to measure
temperature ?
 Isn’t mass flow rate independent of
temperature ?
 RTD measures tube wall temperature
 Young’s Modulus of Elasticity
 Other Uses
– Enhanced Density
– Petroleum Measurement
Temperature Specifications
Process Fluid Limits :
 Defines the operating temperature limits of
the sensor and its components to withstand
continuous exposure to the heat/cold radiated
by the process fluid(s) being measured by the
sensor
– Limiting components are coils, wiring, RTD, wire
feed-through, J-box terminals, location of core
processor and other electronic components on
the sensor
Temperature Specifications
Ambient Limits
 These are specifications provide the temperature range
the area in which the sensor will be installed must be
maintained in order to meet the electrical approvals
required by the customer

 For example, in order for an Elite sensor to meet UL


approval, the sensor can not be exposed to temperatures
beyond –40oF to 104oF (w/either a core processor or J-
box)

 Ambient temperature and fluid temperatures limit each


other for integral core processors
Temp. Limits for Elite Sensor
Practice Questions
 If the customer wants to use a CMF025 with an
integral (not extended) core processor on a liquid
with a temperature of 125°C, what is the
maximum ambient temperature it can withstand ?

 If the customer needs to operate at higher


ambient temperatures, what are their options ?
Pressure
Specifications
Micro Motion Coriolis Meters
Pressure Ratings
Flow Tube Rating
Represents the maximum sustained operating pressure that
the sensor tubing can be exposed to from the process
fluid
 Ratings are based on materials of construction and tube
wall thickness
 Tube rating must be > maximum process pressure
– Large safety factor employed in design per ASME codes
 ELITE Sensor Examples:
– 316L and 304L stainless steel 1500 psi
– Nickel-alloy (Hastelloy) 2755 psi
– High-pressure CMF010P 6000 psi
Pressure Ratings
Housing Rating
 The pressure rating of any std. or optional secondary
containment of the sensor. Customers purchase
secondary containment when there is concern about
safety should there be a tube failure
 There are two numbers shown for secondary containment
– Containment per ASME B31.3
• CMF050 1421 psi
– Burst Pressure per ASME B31.3 (point of failure)
• CMF050 5286 psi

 Other Options/Techniques :
– Purge Fittings
– Rupture Disks
Pressure De-rating
 Above certain temperatures, the pressure rating
of a sensor (case and housing) decreases.

 For example, all Elite sensors are derated by


7.2% for 316L SS flow tubes at 400°F and by
5.4% for housing

 Calculate the pressure rating for flow tubes and


housing of a CMF050M at 400°F.
Environmental
Effects
Micro Motion Basics for Now

More Detail in Future Advanced


Class
Environmental Effects
 Process Temperature Effect on Zero
– Recommend customer zeros sensor at or near
operating conditions
– Factory zero is good on most of the applications for
new Elite with 800 Core Processor (Flow rate is within
5:1)

 Pressure Effect
– Very Predictable
– Can Compensate
– No Effect on Small Meters
Pressure and Temperature Effects

Pressure Effect Temperature Effect

Calibration Change : Zero Uncertainty :


Size % per psi % per bar lb/min/°F kg/hr/°C

CMF025 None None ± 0.00006 ± 0.0027

CMF050 None None ± 0.00017 ± 0.0085

CMF100 - 0.0002 - 0.003 ± 0.00069 ± 0.034

CMF200 - 0.0008 - 0.012 ± 0.009 ± 0.43

CMF300 - 0.0006 - 0.009 ± 0.028 ± 1.36


Materials of
Construction
Micro Motion Coriolis Meters
Materials of Construction
 Wetted Parts
– Any portion of the sensor that comes into contact with
the process fluid

 Non-Wetted Parts
– Housing
– Junction Box
– Core Processor

 316L SS, 304L SS, Hastelloy C-22, Tantalum,


Titanium, Tefzel Lined, Epoxy-Coated Aluminum
Fittings, Flanges and
End Connections
Micro Motion Coriolis Meters
Fitting and Flange Options
 Flanges
– Weld Neck Raised Face Flange
– Lap Joint
– RTJ
– ANSI, JIS, DIN

 Fittings
– Tube Compression
– NPT Cajon Union Fitting
– Sanitary Fitting (Tri-Clamp Compatible)
Fitting and Flange Options - contd
 Other
– Grayloc High Pressure Hubs
– Wafer Style
– CEQ

 Common Customer Words


– Threaded
– Swagelok
Ordering Information
Micro Motion Product Matrix
Product Description
 First Several Letters of a Model Number
– Describes the base model
– Includes designation for high pressure or high
temperature
– Includes wetted material designation

 Examples
– DT065H
– CMF010P
– CMF025L
– CMF300A
Product Description – cont’d
Densitometer Model Selection Example

7828 A C A A C B L B B Z
Process Connections, Case Options
 Process Connections
– End Connection Choice
– Three Digit Code
– Example : 323 – ¼” NPT Female CAJON Size 4 VCO Fitting

 Case Options
– Describes the case
• Standard Case
• Compact Case (F-Series Only)
• Hygenic Case
• Purge Fittings
• Rupture Disks
• Containment Options
Electronics Interface
 Core Processor
– Extended or Not
– Aluminum or Stainless Housing
 Junction Box
– Extended or Not

 MVD Solo
– Extended or Not
– Aluminum or Stainless Housing

 Integral Transmitter
– Not Available for Elite
Conduit Connections, Approvals
 Conduit Connections
– NPT or M20 (½” NPT for MVD, ¾” NPT for J-box)
– Glands are Optional

 Approvals
– Many Available – Discussed in Product Approval
Module
– MVD Solo with Approvals includes a Barrier
Language and Options
 Language
 Future Option
 Measurement Application Software
– Petroleum Measurement Software - Only used for
MVD Solo

 Factory Options
– Z for Standard
– X for CEQ
– R for Restocked
Rosemount 8800D
Vortex Meter
General Sensor Specifications
Rosemount 8800D Vortex Meters
8800DF (Flanged) , 8800DW (Wafer) Vortex
Reliable in Tough Applications
Designed for Reliability
• Eliminates Impulse Lines
• Only Vortex on the Market with No Ports or Gaskets
• All-Cast, Non-Clog Design
• Vibration Immunity with ADSP Filters & Mass-Balanced Sensor
Wafer
• Fast Filter Optimization with ‘Auto Filter Adjust’
Flanged • Replaceable Sensor Without Breaking Process Seals
• Simple, ‘In the Field” Troubleshooting through Device Diagnostics

8800DR Reducer Vortex


Extends Low Flow Coverage at a Reduced Cost
Designed to Reduce Cost and Risk
• Doubles Low Flow Range Compared to Standard Vortex
• Avoid Costly Field Welding by Using Reducer Vortex where Downsized
Vortex Meters were Historically Used.
• Reducer Vortex and Standard Vortex have the same Face-To-Face
Dimension Eliminating Pipe Change Costs and Risk
• Simplified Installation with No Field Welding of Reducers
and Expanders
Rosemount 8800D Vortex Meters – cont’d
8800DD Dual Vortex
Safety Integrated Systems (SIS)
Designed for Safety
• Redundant Flow Signals
• Two (2) Complete Vortex Meters : Sensor, Electronics and Shedder
Bar
- 10” and 12” Dual Style Vortex Meters uses Single Shedder Bar
Rosemount 8800D MultiVariable
(T-Comp) Vortex Meter
 Temperature sensor added to 8800D as
option
 Allows vortex to measure both flow and
temperature
 Provide a temperature compensated
mass flow for saturated steam
 Great for saturated steam applications
– Increased accuracy by calculating density
from actual process temperature
– Reduces cost of installing vortex and
temperature separately
– Provides excellent rangeability
Rosemount 8800D MultiVariable (MV)
Vortex Meter Design
 Electronics
– Processes two signals – vortex shedding for flow,
temperature for density
– Calculates and outputs temperature compensated
mass flow for saturated steam
– Can output two simultaneous variables (4-20 and
Pulse): Flow, Steam Mass flow, Temperature
 Temperature Sensor
– Type N thermocouple
– Isolated from process like vortex sensor
– uses shedder bar as thermowell
– 2-wire, Conserves power for future enhancements
 The only MV Vortex with temperature and
vortex sensor that can be replaced
independently and without exposing user to
process fluid
Rosemount 8800D MV Vortex Meter
- Why Type-N Thermocouple ?

 Why a Thermocouple ?
– 1/8” diameter temperature sensor is required to use shedder bar
as thermowell
– Thermocouples have been shown to have better performance
and reliability in small diameters.
– This is not true for larger diameters (¼” diameter) where RTDs
will be more accurate

 Why Type N ?
– Type-N special limits gets us the same level of accuracy as small
diameter RTD with less drift
• Type N is ni-cor
• Not a base-metal thermocouple (Type J or K), which makes it less
susceptible to drift.
Rosemount 8800D MV Vortex Offering
 Materials
– Stainless Steel
 Line Sizes
– Flanged >=2”
– Reducer >=3”
– Not available on Dual or Wafer
 Flange Ratings
– ANSI Class 150-600, DIN PN 10-100, JIS 10K-40K
 HART or HART w/ Pulse
 Called out by MTA option code in 8800D model
structure
 Lead-time is 15-30 working days for most
configurations
 Remote Mount and FOUNDATION fieldbus not
available at initial release
Rosemount 8800D Vortex Meter Line Sizes

 Wafer – 8 sizes (code W)


– ½, 1, 1½, 2, 3, 4, 6 and 8 inches

 Flanged & Dual Design – 10 sizes (codes F and D)


– ½, 1, 1½, 2, 3, 4, 6, 8, 10 and 12 inches

 Reducer – 9 sizes (code R)


– 1, 1½, 2, 3, 4, 6, 8, 10 and 12 inches

 Note : all meters are compatible with pipes with


10, 40 and 80 schedule
Reynold’s Number & Velocity Limits
 Minimum Reynold’s Number
– 10,000 for sizes ½” through 4”
– 20,000 for sizes 6” through 12”

 Minimum Measurable Velocity


– Liquid application : 0.22 m/s or 54/
– Gas application : 2.00 m/s or 54/

 Maximum Measurable Velocity


– Liquid application : 7.6 m/s or 134,000/
– Gas application : 76 m/s or 134,000/
Temperature & Pressure Specs
 Process Temperature Limits
– Standard : -40 ºC to 232 ºC
– Extended : -200 ºC to 427 ºC

 Pressure Limits
– Flanged, Reducer & Dual : Rated to ASME B16.5 for classes
150, 300, 600 and 900
– Wafer : Rated to ASME B16.5 for classes 150, 300 and 600
Performance Specs
 Accuracy
– Includes Linearity, Hysteresis and Repeatability
– For Liquids :  0.65% of Rate
• For 6” to 12” Reducer Vortex, Accuracy is 1.0% of Rate

– For Gas/Steam : 1.35% of Rate


• For ½” and 1” Wafer, Flanged and Reducer Vortex, Max. Velocity is
67 m/s
• For Dual Style (all sizes) Max. Velocity is 30.5 m/s
• For 6” to 12” Reducer Vortex, Accuracy is 1.5% of Rate

 Repeatability : 0.1% of rate


MV Vortex Accuracy in Saturated Steam
 Volumetric Flow
– ± 1.35% of Rate for Gas/Steam

 Temperature
– ± 2.2°F (1.2°C) or 0.4% of Reading (in °C), whichever
is Greater

 Temperature Compensated Mass Flow


– ± 2.0% of Rate Nominal
 Nominal conditions include pressures 150 psig (10 bar-g) or higher.
Add 0.08% of uncertainty for every 15 psi (1 bar) below 150 psig (10 bar-g).
8800 Ordering Information

8800D-W-020-S-A1-N-1-D-1-I5-M5-Q4

Cal Cert
LCD Indicator
FM I/S
4-20 mA
¼” NPT Conduit Entry
Std Sensor
ANSI 150#RF
SST Wetted Parts
2” Size
Wafer Design
8700 Series Magmeter
General Sensor Specifications
8705 Flanged Flowtube
8707 Hi-Signal Flowtube
General Sensor Specifications
8705/8707 Flowtubes
 Available Sizes
– 8705 Standard : ½” to 36”
– 8707 High Signal : 3” to 36”

 Measurement Electrodes
– Code SA : 316L SSt
– Code HA : Hastelloy C-276
– Code TA : Tantalum
– Code PA : 90% Platinum & 10% Iridium
– Code NA : Titanium

 Third Grounding Electrodes


– Same as above but Code is “ _E”
– Not available on 8707 High Signal Mag Flow Tube
8705/8707 Flow Tubes
 Liners
– Code A : PFA (3” to 12” only)
– Code T : Teflon
– Code F : Tefzel (1/2” to 16” only)
– Code P : Polyurethane (1.5” to 36”)
– Code N : Neoprene (1.5” to 36”)
– Code L : Linatex (1.5” to 36”)

 Flanges
– Carbon Steel/304 SSt : ANSI 150#, 300#, 600# and 900#
– Carbon Steel/304 SSt : DIN, PN 10, 16, 25 and 40
– Note : For DIN flanges refer to price book as not all sizes are
available
Temperature Limits
 Temperature Limit is Liner-Dependent
– Teflon Liner : -29 ºC to 177 ºC
– Tefzel Liner : -29 ºC to 149 ºC
– PFA Liner : -29ºC to 177 ºC
– Polyurethane Liner : -18ºC to 60 ºC
– Neoprene Liner : -18 ºC to 85 ºC
– Linatex Liner : -18 ºC to 70 ºC
Flow Tube Pressure vs Temperature Limits

Flange -29 ºC to 93 ºC 149 ºC 177 ºC


Rating 38 ºC

CS Class 285 260 230 215


150 psi psi psi psi
CS Class 740 675 655 645
300 psi psi psi psi
CS Class 1000 800 700 650
600 psi psi psi psi
Performance Specs
 Accuracy
– 8705 w/ 8712D, 8732C or 8742C : ±0.5% or, 0.3 to 10 m/s
– 8707 w/ 8712H : ±0.5% or, 1 to 10 m/s

 Conductivity Requirement
– For 8705 : min 5 microseimens
– For 8707 with 8712H : min 50 microseimens
– When using code _E : min 100 microseimens
8705/8707 Ordering Information

8705-T-SA-020-C1-W0-N0-G1-B3

Integral mount 8732C


316L sst grounding rings
FM Cl1 Div2
Std Housing
ANSI 150# RF
2” Size
SST Electrodes
Teflon Liner
Flanged Design
8711 Wafer Style
Flowtube
General Sensor Specifications
8711 Wafer Mag
 Available Liner & Sizes
– With Tefzel Lining : 4, 8, 15, 25, 40, 50 80, 100, 150 and 200 mm
– With Teflon Lining : 15, 25, 40, 50 80, 100, 150 and 200 mm

 Measurement Electrodes
– Code SA : 316L sst
– Code HA : Hastelloy C-276
– Code TA : Tantalum
– Code PA : 90% Platinum & 10% Iridium
– Code NA : Titanium

 Mounting Kit (2 alignment rings, sst studs and nuts)


– Code 1 : ANSI 150#
– Code 2 : DIN PN10/16
– Code 3 : ANSI 300#
– Code 4 : DIN PN 25/40
Performance Specs
 Accuracy
– ±0.5% of rate from 1 to 10 m/s
– ±0.0045 m/s from low flow cutoff to 1 m/s

 Conductivity
– Min of 5 microsiemens/cm

 Pressure Limits
– With Tefzel Lining : Full vacuum to 51 bar (15mm to 200mm)
Full vacuum to 19.6 bar (4mm and 8mm)
– With Teflon Lining : Consult Factory
8711 Ordering Information

8705-T-SA-020-U-1-N0-G1

316L SST Grounding Rings


FM Cl1 Div2
Mounting Kit, ANSI 150#
Integral Mount
2” Size
SST Electrodes
Teflon Liner
Flanged Design
8721 Sanitary Flowtube
General Sensor Specifications
Rosemount 8721 Hygienic Mag
True Hygienic Design from the Rosemount 8721 Sensor
Most Reliable Name in with Remote Terminal
Junction Box
Measurement
 Sizes to Fit Most Applications :
– 1”, 1.5”, 2”, 3”, 4”
(25mm, 40mm, 50mm, 80mm, 100mm)
– Future 1/2” (15mm) and 2.5” (65mm)

 Sanitary/Hygienic Approvals :
– 3-A Authorized
– EHEDG Certification PFA Liner and
SS Meter Body
Rosemount 8721 Hygienic Mag
 FDA Approved Materials for Product
Rosemount 8721 Sensor
Contact Surfaces : with Remote Terminal
– PFA Liner Junction Box

• Ra< 0.8µm Surface Finish (180 grit)


– 316 SS Electrodes
• Ra< 0.8µm Surface Finish (180 grit)
– 316L Process Fittings
• Ra< 0.8µm Surface Finish (180 grit)

 Suitable Exterior Surfaces :


– 304 Stainless Steel Flow Tube and PFA Liner and SS
Tube Adapter Construction Meter Body
Rosemount 8721 Hygienic Mag
 Full Diameter Design Rosemount 8721 Sensor
– Meter ID Matches Process Piping with 8732 Transmitter

– “Piggable”

 Flexible Process Connections :


– Tri-Clamp
– IDF
– Direct Weld Nipple
– Can be Customized to Meet
User Needs
– Variety of Gasket Materials to Tri-clamp
Process
Weld
Nipple
Meet Process Requirements Connection
• EPDM, Viton, Buna-N, Silicone, IDF
Teflon Process
Connection
Rosemount 8721 Transmitters
 8721 Design is Optimized for all
Rosemount Transmitters
– 8732C for Integral Mount transmitter
• Also available in remote mount
configuration

– 8742C Foundation Fieldbus


Transmitter for integral mount
• Also available in remote mount
configuration

– 8712C\U Remoter Mount Transmitter


8721 Ordering Information

8721-A-SA-020-U-A-1-N0-AH

Electro Polish to 20 Ra

FM, Cl1, Div1


Silicone Gasket
Tri-Clamp Connection
Integral Mount
2” Size
SST Electrodes
Teflon Liner
Sanitary Mag